K Number
K203674

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-09-27

(285 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

B Recon Line is intended for transfer of solvent from IV bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line is intended for the transfer of solvent from a bag and liquid drug from a vial to a bag through the APOTECAbag automated system.

B Filling Line is intended for transfer of drug from a vial to a bag through the APOTECAbag automated system.

B Recon Line with needle is intended for transfer of solvent from a bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line with vial needle is intended for the transfer of solvent from a bag and liquid drug from a bottle to a bag through the APOTECAbag automated system.

B Filling Line for bag is intended for transfer of drug from a bag to another bag through the APOTECAbag automated system.

B Dispensing Line devices are only to be used with APOTECAbag pharmacy compounding device. These devices are not to be used for the compounding of chemotherapy and oncology drugs.

Device Description

B Dispensing Line is a disposable medical device used for the transfer of liquids from an initial container to a final one, by means of the action of a peristaltic pump. It is a sterile tube, with sections designed to be inserted in peristaltic pump and with specific terminal connectors for the connection of the device with the different containers.

The tube section to be inserted in the peristaltic pump presents a larger diameter than the other sections of tube.

The device is intended to be used with the purpose of pharmacy compounding in the automated system APOTECAbag.

The device is manufactured with biocompatible materials and it is provided sterile by EO sterilization method.

AI/ML Overview

This document describes the regulatory submission for the "B Dispensing Line" device (K203674). This device is an IV fluid transfer set designed for use with the APOTECAbag automated pharmacy compounding system. The provided information focuses on non-clinical testing to demonstrate substantial equivalence to a predicate device, the KIRO Set (K152441).

No AI/ML device is mentioned in this documentation. The device is a physical medical device (IV fluid transfer set). Therefore, questions related to AI/ML specific studies, such as sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, or standalone algorithm performance, are not applicable.

Here's the information regarding the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria / StandardReported Performance
Performance Testing
Performance test (in APOTECAbag)Acceptable dosage time and accuracy, No leakage, Right disconnection (no ruptures), Ability to maintain acceptable performances after prolonged used, Dosage accuracyPASS
Dose accuracy and repeatabilityAcceptable dosage accuracy, mean and maximum error in the limits imposed by US Pharmacopoeia, Ability to maintain performances after prolonged usePASS
Leakage testISO 8536-9PASS
Tensile stress testISO 8536-9PASS
Test for transparencyISO 8536-9PASS
Drug CompatibilityNo adverse compatibility effectsPASS
Biocompatibility Testing
Cytotoxicity MEM ElutionISO 10993-5:2009, ISO 10993-12:2012PASS
Acute Systemic ToxicityISO 10993-11:2017, ISO 10993-12:2012PASS
Guinea Pig Sensitization TestISO 10993-10:2010, ISO 10993-12:2012PASS
Haemolysis test direct and indirect contactISO 10993-4:2017, ISO 10993-12:2012PASS
Pyrogen Test on rabbitsUSP 41-NF36:2018 <151> Pyrogen Test (USP Rabbit Test)PASS
Rabbit Intracutaneous injection testISO 10993-10:2010, ISO 10993-12:2012PASS
Bacterial Endotoxins TestUSP 41-NF36:2019 <85> Bacterial Endotoxins TestPASS
Particulate contaminationISO 8536-4PASS
Chemical characterizationISO 8536-4PASS
Particulate Matter for InjectionsUSP <788> (Method 1 Light Obscuration Particle Count Test)PASS
Sterilization and Shelf Life Testing
Packaging ValidationEffective microbiological barrier, product sterility and integrity preservationPASS
Sterilization ValidationISO 11135:2014PASS
Shelf life, Real timeSterility and product integrity maintained over the entire shelf lifeOngoing
Labelling validationCorrectness and completeness of labelingPASS

2. Sample size used for the test set and the data provenance

The document states that "AEA srl has performed the following non-clinical/design verification testing." This refers to laboratory-based testing on the physical device itself, not a dataset in the context of an AI/ML model. Therefore, "sample size" and "data provenance" (country of origin, retrospective/prospective) as concepts related to data analysis for AI/ML are not applicable here. The tests were performed on physical samples of the B Dispensing Line.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable as this is a physical medical device and no AI/ML model requiring expert-established ground truth data is involved. The ground truth for the performance tests would be established by measuring the physical properties and functionality of the device against predefined engineering and quality standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for a physical medical device's non-clinical performance testing. The "adjudication method" traditionally refers to how conflicting expert opinions are resolved in establishing ground truth for evaluating diagnostic or prognostic algorithms.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical tests, the "ground truth" is defined by established engineering standards, regulatory guidelines (e.g., ISO, USP), and documented performance specifications for the device. For example:

  • Performance tests: Measured physical accuracy, absence of leakage, structural integrity.
  • Biocompatibility tests: Absence of adverse biological reactions as defined by ISO 10993 standards.
  • Sterilization tests: Validation against ISO 11135.

8. The sample size for the training set

Not applicable. This is not an AI/ML device, so there is no training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device, so there is no training set or ground truth for it.

In summary, the provided document details the non-clinical performance and safety testing of a physical medical device, the B Dispensing Line. It does not involve any AI/ML components, and therefore, many of the requested specific details related to AI/ML study design are not relevant to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

September 27, 2021

AEA srl Michele Mengoni Quality Assurance & Regulatory Affairs Via Fiume 16 Angeli de Rosora, Ancona 60030 Italy

Re: K203674

Trade/Device Name: B Dispensing Line Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI, NEP Dated: August 5, 2021 Received: August 26, 2021

Dear Michele Mengoni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203674

Device Name B Dispensing Line

Indications for Use (Describe)

B Recon Line is intended for transfer of solvent from IV bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line is intended for the transfer of solvent from a bag and liquid drug from a vial to a bag through the APOTECAbag automated system.

B Filling Line is intended for transfer of drug from a vial to a bag through the APOTECAbag automated system.

B Recon Line with needle is intended for transfer of solvent from a bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line with vial needle is intended for the transfer of solvent from a bag and liquid drug from a bottle to a bag through the APOTECAbag automated system.

B Filling Line for bag is intended for transfer of drug from a bag to another bag through the APOTECAbag automated system.

B Dispensing Line devices are only to be used with APOTECAbag pharmacy compounding device. These devices are not to be used for the compounding of chemotherapy and oncology drugs.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

LOCCIONI

510k K203674 - Summary

The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92.

I. SUBMITTER

AEA SRL Via Fiume, 16 Angeli di Rosa 60030 Ancona Italy Phone: 0039-0731-816689 Contact Person: Michele Mengoni Date prepared: September 27th, 2021

II. DEVICE

Name of device:B Dispensing Line
Common or Usual Name:IV Fluid Transfer Set
Classification Name:Set, IV Fluid Transfer
Regulation Number:880.5440 - Intravascular administration set
Regulatory Class:II
Product Code:LHI – NEP (Secondary product code)
Panel Identification:General Hospital

III. PREDICATE DEVICE

KIRO Set (510k Number: K152441, Regulation number: 880.5440, Product code: LHI)

Subject device: B Dispensing Line
Traditional 510(k)

{4}------------------------------------------------

LOCCIONI

IV. DEVICE DESCRIPTION

B Dispensing Line is a disposable medical device used for the transfer of liquids from an initial container to a final one, by means of the action of a peristaltic pump. It is a sterile tube, with sections designed to be inserted in peristaltic pump and with specific terminal connectors for the connection of the device with the different containers.

The tube section to be inserted in the peristaltic pump presents a larger diameter than the other sections of tube.

The device is intended to be used with the purpose of pharmacy compounding in the automated system APOTECAbag.

The device is manufactured with biocompatible materials and it is provided sterile by EO sterilization method.

List of products/variantsCodeUDI-DI
B Recon LineBDL-015060304050830
B Double Filling LineBDL-025060304050847
B Filling LineBDL-035060304050854
B Recon Line with needleBDL-045060304050878
B Double Filling Line with vial needleBDL-055060304050885
B Filling Line for bagBDL-065060304050892

Six different variants of the device are available:

V. INDICATION FOR USE

B Recon Line (BDL-01)

B Recon Line is intended for transfer of solvent from IV bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line (BDL-02)

B Double Filling Line is intended for the transfer of solvent from a bag and liquid drug from a vial to a bag through the APOTECAbag automated system.

B Filling Line (BDL-03)

B Filling Line is intended for transfer of drug from a vial to a bag through the APOTECAbag automated system.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "LOCCIONI" in a bold, red, sans-serif font. The letters are evenly spaced and appear to be a logo or brand name. The background is plain white, which makes the red text stand out.

B Recon Line with needle (BDL-04)

B Recon Line with needle is intended for transfer of solvent from a bag to a vial with powder drug to be reconstituted, through the APOTECAbag automated system.

B Double Filling Line with vial needle (BDL-05)

B Double Filling Line with vial needle is intended for the transfer of solvent from a bag and liquid drug from a bottle to a bag through the APOTECAbag automated system.

B Filling Line for bag (BDL-06)

B Filling Line for bag is intended for transfer of drug from a bag to another bag through the APOTECAbag automated system.

B Dispensing Line devices are only to be used with APOTECAbag pharmacy compounding device.

These devices are not to be used for the compounding of chemotherapy and oncology drugs.

{6}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

B Dispensing Line (BDL-01,BDL-02, BDL-03, BDL-04, BDL-05, BDL-06)KIRO setDifferences
BRIEFDESCRIPTIONAND INTENDEDUSEB Dispensing Line is a single-usetransfer set device intended tobe used by pharmacists asdisposable in an automateddrug compounding systems,called APOTECAbag. Sixdifferent variants of the deviceexist:B Recon Line (BDL-01) and BRecon Line with vial needle(BDL-04) for the transfer ofsolvent inside drug vials withthe purpose ofreconstitution of lyophilizeddrugs B Double Filling Line (BDL-02) and B Double Filling Linewith vial needle (BDL-05) forthe transfer inside a bag ofliquid drug and solvent B Filling Line (BDL-03) for thetransfer of drug from a vialto a bag B Filling Line for bag (BDL-06) for the transfer of liquiddrug from a bag to anotherbag B Dispensing Line is provided ofa section compatible with aperistaltic pump.The KIRO Set is a sterile,single-use disposableancillary device used with theperistaltic pump in the KIROOncology pharmacycompounding device for thetransfer of fluids in thepreparation of finalmedication containers andthe reconstitution of drugvials in hospital pharmacies.Same intended use:transfer of fluids inthe preparation offinal medicationcontainers and thereconstitution of drugvials in hospitalpharmacies by meansof peristaltic pumps
PRODUCT CODELHI, NEP (Secondary ProductCode)LHI, NEP (Secondary ProductCode)Same
PHARMACYCOMPOUNDINGDEVICESPECIFIEDAPOTECAbagKIRO OncologyDifferent
USEENVIRONMENTHospital pharmacy, insideAPOTECAbag (PCD). ISO 5environmentHospital pharmacy inside theKIRO Oncology PCD ISO 5environmentSame useenvironment
B Dispensing Line (BDL-01,BDL-02, BDL-03, BDL-04, BDL-05, BDL-06)KIRO setDifferences
TARGET USERSHealth-care personnel,specifically trained in injectabledrug manipulation andcompoundingTrained health-carepersonnelSame
TUBINGMedical Grade Polyvinylchloride(PVC)Medical GradePolyvinylchloride (PVC) andMedicale Grade SiliconeDifferent
CONNECTORMATERIALABSABSSame
INFUSIONMETHODPeristaltic pumpPeristaltic pumpSame
STERILIZATIONEthylene OxideGamma RadiationDifferent
SINGLE USEYesYesSame
INPUTLINECONNECTOR(S)B Recon Line: Vented SpikeB Double Filling Line: Vented Spike and Vented MicrospikeB Filling Line: Vented MicrospikeB Recon Line with vial: Vented SpikeB Double Filling Line with vial needle: Vented Spike and needleB Filling Line for bag: Vented spikeVented Spike or Male LuerSame
OUTPUTLINECONNECTOR(S)B Recon Line: Vented MicrospikeB Double Filling Line: NeedleB Filling Line: NeedleB Recon Line with vial: NeedleB Double Filling Line with vial needle: NeedleB Filling Line for bag: NeedleMale Luer ConnectorDifferent

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "LOCCIONI" in all capital letters. The font is a blocky, sans-serif style, and the color is a dark red. The letters are evenly spaced and aligned horizontally.

Differences discussion

B Dispensing Line and KIRO set have the same purpose, both are used for the compounding of drugs and reconstitution of vials containing lyophilized drugs by means of peristaltic pumps. The liquid transfer method is therefore the same. Reference and subject device have product code. The use environment and target users are the same except the systems in which they have to be used. To address this difference, performance tests of B Dispensing Line with APOTECAbag system have been performed.

KIRO Set and B Dispensing Line are both for single use and provided sterile. Both devices are provided of vented spikes and mini-spikes as input line connector

Subject device: B Dispensing LineTraditional 510(k)Section: 510 (k) Summary Rev.04VOL#:005
-----------------------------------------------------------------------------------------------------

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The letters are all capitalized and are a dark red color. The background is white, which makes the text stand out.

The sterilization method is different, B Dispensing Line is provided sterile by means of EO sterilization cycle, while KIRO Set is sterilized using gamma radiations. A complete validation of sterilization method of B Dispensing Line has been performed according to standard ISO 11135.

The tube material is different, B Dispensing Line is made of PVC, while KIRO Set is made of PVC and silicone. To address this difference a drug compatibility analysis has been performed and a biocompatibility test campaign according to ISO 10993 has been performed.

The connector material (spikes, mini-spikes and luer connection) is the same.

As input connector B Double Filling Line with vial needle and a spike, KIRO Set is provided with spike and with male luer connectors. The material of connector in both devices is the same. As output line connector, KIRO Set is provided of male luer connector and B Dispensing Line is provided of a mini-spike (only B Recon Line) or needle (all the other models). The material of connector in both devices is the same. To address the difference of the output line connector performance tests according to ISO 8536-9 have been performed.

{9}------------------------------------------------

VII. PERFORMANCE DATA

Non-Clinical Testing

AEA srl has performed the following non-clinical/design verification testing based on the risk analysis conducted and the favourable outcome of these tests demonstrate that the AEA proposed devices perform in an equivalent manner to the predicate devices.

TESTCONCLUSIONSACCEPTANCE CRITERIA / STANDARD
Performance Testing
Performance test (in APOTECAbag)PASSAcceptable dosage time and accuracy• No leakage• Right disconnection, no ruptures• Ability to maintain acceptable performances after prolonged used• Dosage accuracy
Dose accuracy and repeatabilityPASSAcceptable dosage accuracy, mean• and maximum error in the limits imposed by US Pharmacopoeia• Ability to maintain performances after prolonged use
Leakage testPASSISO 8536-9
Tensile stress testPASSISO 8536-9
Test for transparencyPASSISO 8536-9
Drug CompatibilityPASSNo adverse compatibility effects
Biocompatibility Testing
Cytotoxicity MEM ElutionPASSISO 10993-5:2009, ISO 10993-12:2012
Acute Systemic ToxicityPASSISO 10993-11:2017, ISO 10993-12:2012
Guinea Pig Sensitization TestPASSISO 10993-10:2010, ISO 10993-12:2012
Haemolysis test direct and indirect contactPASSISO 10993-4:2017, ISO 10993-12:2012
Pyrogen Test on rabbitsPASSUSP 41-NF36:2018 <151> Pyrogen Test(USP Rabbit Test)
Rabbit Intracutaneous injection testPASSISO 10993-10:2010, ISO 10993-12:2012
Bacterial Endotoxins TestPASSUSP 41-NF36:2019 <85> BacterialEndotoxins Test;
Particulate contaminationPASSISO 8536-4
Chemical characterizationPASSISO 8536-4
Particulate Matter for InjectionsPASSUSP <788> (Method 1 Light ObscurationParticle Count Test)
Sterilization and Shelf Life Testing
Packaging ValidationPASSEffective microbiological barrier, productsterility and integrity preservation
Sterilization ValidationPASSISO 11135:2014
Subject device: B Dispensing LineTraditional 510(k)Section: 510 (k) Summary Rev.04VOL#:005
----------------------------------------------------------------------------------------------------

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The word is displayed in a dark red color. The letters are evenly spaced and appear to be capitalized.

TESTCONCLUSIONSACCEPTANCE CRITERIA / STANDARD
Shelf life, Real timeOngoingSterility and product integrity maintainedover the entire shelf life
Labelling validationPASSCorrectness and completeness of labeling

Clinical Testing

Clinical testing was not required for this submission.

Substantial Equivalence

B Dispensing Line is substantially equivalent to the predicate device in its intended use, principles of operation, technology, design, materials and performance.

VIII. CONCLUSION

Results of performance and biocompatibility testing conducted on the proposed devices demonstrate that B Dispensing Line is substantially equivalent to the predicate devices.

Subject device: B Dispensing LineTraditional 510(k)Section: 510 (k) Summary Rev.04VOL#:005
----------------------------------------------------------------------------------------------------

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.