K Number
K203617

Validate with FDA (Live)

Device Name
MaxFOV 2
Date Cleared
2021-03-02

(81 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The deep learning based Max Field-of-View (MaxFOV 2) is a CT image reconstruction method intended to produce images of the head and whole-body using Axial, Helical, and Cine acquisitions.

MaxFOV 2 is designed to extend the nominal display field of view (DFoV) for cases where patient size and positioning requirements result in a portion of the patient's body to be outside of the nominal DFoV.

These extended FoV images are intended for use in radiation therapy planning and are clinically useful for the simulation and planning of radiation therapy for the treatment of cancer for patients. They can also be used for visualization of patient anatomy for cases not involving therapy planning. MaxFOV 2 is intended for patients of all ages, especially bariatric patients.

Device Description

The MaxFOV 2 is an enhanced Extended Field of View (EFOV) reconstruction option for GE's CT scanners. The MaxFOV 2 utilizes a new deep learning algorithm to extend the display field of view (DFOV) beyond the CT system's scan field of View (SFOV) of 50cm to up to 80cm depending on the bore size of the CT system. CT scanners use the EFOV reconstruction algorithms to visualize tissue truncated due to large patient habitus and/or off-center patient positioning. Same as the Wide View option on the predicate, the MaxFOV 2 is designed to enable a clinically useful visualization of the skin line and CT Number of human body parts located outside of the SFOV. EFOV images are especially useful for radiation therapy planning and they can also be used for visualization of patient anatomy outside of the SFOV for routine CT imaging. This DL enabled new MaxFOV2 EFOV reconstruction process offers improved performance over the existing WideView option on the predicate device.

The DL MaxFOV2 algorithm was designed and tested for GE's multiple CT scanner platforms of various bore sizes from 70cm to 80cm. These CT systems with the integrated MaxFOV 2 option remain compliant with the same standards as base CT systems.

This option is commercially marketed as MaxFOV2.

AI/ML Overview

The provided text describes the MaxFOV 2, a CT image reconstruction method that utilizes deep learning to extend the display field of view. The document outlines the device's indications for use, technological characteristics, and the summary of non-clinical and clinical testing performed to support its substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a table format with pass/fail thresholds. Instead, it describes various tests and a clinical reader study designed to demonstrate performance that is "consistent and acceptable," does not "raise new or different questions of safety and effectiveness," and supports "substantial equivalence and performance claims." The performance is evaluated relative to the predicate device, "Wide View (K023332)".

Assuming the core acceptance criteria revolve around image quality, accuracy of patient contour (skin line), and CT number accuracy in the extended FOV, the reported performance is qualitative but positive:

Acceptance Criterion (Inferred from documentation)Reported Device Performance
Image Quality Performance"A suite of engineering bench testing using phantoms was performed to evaluate image quality performance of MaxFOV 2... all test results demonstrated MaxFOV 2's consistent and acceptable performance."
MaxFOV 2 Patient contour (Skin line) accuracyTested as part of engineering bench tests. Results demonstrated "consistent and acceptable performance."
MaxFOV 2 CT Number accuracyTested as part of engineering bench tests. Results demonstrated "consistent and acceptable performance."
Does not raise new/different safety & effective questions compared to predicate"The complete testing and results did not raise different questions of safety and effectiveness than associated with predicate device." "MaxFOV2's design, verification, validation and risk management processes did not identify any new hazards, unexpected results, or adverse effects stemming from the changes to the predicate."
Clinical Acceptability (Reader Study)"The results of the study support substantial equivalence and performance claims." Images were scored using a 5-point Likert scale for "depiction of the patient's skin surface; depicted tissue densities in the extended FOV region; and overall image quality." The implicit acceptance is that a majority of readers, for a majority of cases, rated the images as clinically acceptable.

Study Details

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 49 CT exams were used in the clinical reader study.
  • Data Provenance: The exams were "acquired from different GE CT system platforms." The text does not specify the country of origin or whether the data was retrospective or prospective. It states the exams "represent typical and challenging RTP-relative scenarios where the MaxFOV2 will likely be used."

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: 5 external, clinical readers.
  • Qualifications of Experts: The specific qualifications (e.g., years of experience, subspecialty) are not explicitly stated, beyond them being "external, clinical readers."

4. Adjudication Method for the Test Set

The text indicates that readers scored images using a 5-point Likert scale. It does not mention an explicit adjudication method (e.g., 2+1, 3+1 consensus) for establishing a single "ground truth" score per image. It seems the readers' individual scores were aggregated and analyzed to support the claims.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done?: A clinical reader study was performed, which is a type of MRMC study, though the primary goal seems to be demonstrating substantial equivalence rather than a direct comparison of human readers with and without AI assistance to quantify an "improvement effect size." The study evaluates the output of the AI model (the reconstructed images) through human reader assessment.
  • Effect Size of Human Improvement with AI vs. without AI assistance: Not explicitly quantified in terms of human reader improvement. The study assesses the quality of images produced by MaxFOV2, which uses a deep learning component. The improvement is implied for the images produced by MaxFOV2 over the predicate device. "This DL enabled new MaxFOV2 EFOV reconstruction process offers improved performance over the existing WideView option on the predicate device." However, this statement refers to the device's performance, not specifically how human readers improve their diagnostic accuracy or efficiency when using the AI-assisted images compared to reading images generated without the AI.

6. Standalone Performance (Algorithm Only)

The text describes "engineering bench testing using phantoms." These tests (MaxFOV 2 Patient contour (Skin line) accuracy and CT Number accuracy, MaxFOV 2 IQ Performance Evaluation using a very large phantom, MaxFOV 2 Performance Evaluation Using an anthropomorphic phantom) appear to be standalone performance assessments of the algorithm and its output, independent of a human reader's interpretation in a diagnostic context. These tests would evaluate the algorithm's output directly against known phantom characteristics.

7. Type of Ground Truth Used for Test Set

  • For Bench Testing: The ground truth for bench testing (e.g., image quality, skin line accuracy, CT number accuracy) would be established by the known physical properties and measurements of the phantoms used.
  • For Clinical Reader Study: The "ground truth" for clinical acceptability in the reader study is based on the expert consensus/opinion of the 5 external clinical readers using a Likert scale. This is a subjective assessment of image quality and clinical utility rather than an objective "pathology" or "outcomes" ground truth. The study evaluates how well the generated images aid human interpretation.

8. Sample Size for the Training Set

The document does not explicitly state the sample size used for the training set of the deep learning algorithm ("CNN"). It only mentions that the MaxFOV2 "uses a CNN which is trained on multiple CT scanners."

9. How the Ground Truth for the Training Set Was Established

The document does not provide details on how the ground truth for the training set was established. It simply states that the CNN was "trained on multiple CT scanners." For deep learning image reconstruction, the training might involve paired data (e.g., truncated vs. full FOV images, or images generated with existing algorithms as a reference), or simulated data, but the specific method of ground truth establishment is not described.

{0}------------------------------------------------

Image /page/0/Picture/10 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized eagle, and the text on the right is in blue. The words "U.S. FOOD & DRUG" are in a larger font size than the word "ADMINISTRATION".

GE Medical Systems, LLC. % Ms. Amy Yang Regulatory Affairs Manager 3000 N Grandview Blvd WAUKESHA WI 53188

March 2, 2021

Re: K203617

Trade/Device Name: MaxFOV 2 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: December 9, 2020 Received: December 11, 2020

Dear Ms. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203617

Device Name MaxFOV 2

Indications for Use (Describe)

The deep learning based Max Field-of-View (MaxFOV 2) is a CT image reconstruction method intended to produce images of the head and whole-body using Axial, and Cine acquisitions.

MaxFOV 2 is designed to extend the nominal display field of view (DFoV) for cases where patient size and positioning requirements result in a portion of the patient's body to be outside of the nominal DFoV.

These extended FoV images are intended for use in radiation therapy planning and are clinically useful for the simulation and planning of radiation therapy for the treatment of cancer for patients. They can also be used for visualization of patient anatomy for cases not involving therapy planning. MaxFOV 2 is intended for patients of all ages, especially bariatric patients.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. There are three white flourishes or droplets evenly spaced around the perimeter of the circle, adding a dynamic element to the design.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:December 10, 2020
Submitter:GE Medical Systems, LLC3000 North Grandview BlvdWaukesha, WI 53188
Primary Contact:Amy YangRegulatory Affairs ManagerPhone: 414-514-3904Email: Amy.yang@ge.com
Secondary Contacts:John JaeckleChief Regulatory Affairs EngineerPhone: 262-424-9547Email: john.jaeckle@ge.com
Device Trade Name:MaxFOV 2
Device ClassificationClass II
Regulation Number/Product Code:21 CFR 892.1750 Computed tomography x-ray system / JAK
Predicate Device Information
Device Name:LightSpeed Ultra with Wide View Option
Manufacturer:GE Medical Systems, LLC
510(k) Number:K023332, Cleared on October 23, 2002
Regulation Number/Product Code:21 CFR 892.1750 Computed tomography x-ray system / JAK
Reference Devices Information
Device Name:SOMATOM DEFINITION AS OPEN

Manufacturer: Siemens Medical Systems, Inc. 510(k) Number: K130901 cleared on January 02, 2014 Regulation Number/ 21 CFR 892.1750 Computed tomography x-ray system / JAK

Product Code:

{4}------------------------------------------------

GE Healthcare 510(k) Premarket Notification Submission

Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are four white, teardrop-shaped elements surrounding the circle, positioned at the cardinal directions. The logo is simple and iconic, representing the long-standing brand identity of General Electric.

Device Description

The MaxFOV 2 is an enhanced Extended Field of View (EFOV) reconstruction option for GE's CT scanners. The MaxFOV 2 utilizes a new deep learning algorithm to extend the display field of view (DFOV) beyond the CT system's scan field of View (SFOV) of 50cm to up to 80cm depending on the bore size of the CT system. CT scanners use the EFOV reconstruction algorithms to visualize tissue truncated due to large patient habitus and/or off-center patient positioning. Same as the Wide View option on the predicate, the MaxFOV 2 is designed to enable a clinically useful visualization of the skin line and CT Number of human body parts located outside of the SFOV. EFOV images are especially useful for radiation therapy planning and they can also be used for visualization of patient anatomy outside of the SFOV for routine CT imaging. This DL enabled new MaxFOV2 EFOV reconstruction process offers improved performance over the existing WideView option on the predicate device.

The DL MaxFOV2 algorithm was designed and tested for GE's multiple CT scanner platforms of various bore sizes from 70cm to 80cm. These CT systems with the integrated MaxFOV 2 option remain compliant with the same standards as base CT systems.

This option is commercially marketed as MaxFOV2.

Intended Use

The MaxFOV 2 reconstruction software is intended for head and whole body CT scans.

Indications for Use

The deep learning based Max Field-of-View (MaxFOV 2) is a CT image reconstruction method intended to produce images of the head and whole-body using Axial, Helical, and Cine acquisitions.

MaxFOV 2 is designed to extend the nominal display field of view (DFoV) for cases where patient size and positioning requirements result in a portion of the patient's body to be outside of the nominal DFoV.

These extended FoV images are intended for use in radiation therapy planning and are clinically useful for the simulation and planning of radiation therapy for the treatment of cancer for patients. They can also be used for visualization of patient anatomy for cases not involving therapy planning. MaxFOV 2 is intended for patients of all ages, especially bariatric patients.

Technological Characteristics

The MaxFOV 2 employs the same fundamental technology as that of the WideView on the predicate device. It is used in the same clinical environment and by the same intended users. Use of the MaxFOV2 does not require change in the CT scanner hardware and does not alter the CT system's control mechanism, or energy type or operating principles.

The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for GE (General Electric). The logo is a blue circle with the letters "GE" in a stylized, cursive font in the center. There are three water droplet shapes surrounding the circle, one on the top, and one on each side. The logo is simple and recognizable, and it is associated with a well-known and established company.

GE Healthcare 510(k) Premarket Notification Submission

SpecificationPredicate DeviceProposed Device
Wide View (K023332)MaxFOV 2
Patient PopulationPatients of all agesSame
Intended usersDosimetristMedical PhysicistsRadiation OncologistsSame
Clinical UseRoutine Clinical UseSame
Targeted clinical condition,anatomyPart of scanned object is located out ofthe scan field of view.Same
Compatible Scan modesCompatible with Axial, Helical, and Cinescan modes.Same
Technology/PrinciplesAnalytical model - classic algorithm wherethe measured sinogram is expanded tocover a FOV larger than SFOV.MaxFOV 2 uses a CNN which is trained onmultiple CT scanners. MaxFOV 2'sreconstruction process (algorithm) containsa Deep Learning based component.
Extended Field of View SpecAllows visualization of up to 65cmAllows visualization of up to 80cm, same asreference device

The MaxFOV 2 option has same intended use, indications for use and substantially equivalent technological characteristics as its predicate and reference device. The changes and the different technological characteristics do not raise new or different questions of safety and effectiveness. The software was developed, verified, and validated under GE Healthcare's QMS including software development lifecycle.

Determination of Substantial Equivalence Summary of Non-Clinical Testing

MaxFOV 2 has successfully completed the design control activities per our quality system that conforms to the Quality System Regulations of 21CFR 820 and ISO 13485, including the Software Development Life Cycle process per IEC62304.

The following quality assurance measures have been applied to the development of the system:

  • Risk Analysis
  • Required Reviews
  • Design Reviews
  • . Software Development Lifecycle
  • . Testing on unit level (Module verification)
  • . Integration testing (System verification)
  • . Performance testing (Verification)
  • Safety testing (Verification)
  • . Simulated use testing (Validation)

{6}------------------------------------------------

GE Healthcare 510(k) Premarket Notification Submission

Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are decorative swirls around the letters and the outer edge of the circle. The logo is simple and recognizable, and it is associated with a well-known and established company.

A suite of engineering bench testing using phantoms was performed to evaluate image quality performance of MaxFOV 2. These tests include:

  • . MaxFOV 2 Patient contour (Skin line) accuracy and CT Number accuracy
  • . MaxFOV 2 IQ Performance Evaluation using a very large phantom
  • MaxFOV 2 Performance Evaluation Using an anthropomorphic phantom

This set of testing was repeated and performed on different GE CT system platforms, all test results demonstrated MaxFOV 2's consistent and acceptable performance.

The complete testing and results did not raise different questions of safety and effectiveness than associated with predicate device. We consider the proposed device is substantially equivalent to the predicate and reference devices, and hence is safe and effective for its intended use.

Summary of Clinical Testing

A clinical reader study with 49 CT exams was conducted. The exams were acquired from different GE CT system platforms and were scored by 5 external, clinical readers. The exams represent typical and challenging RTP-relative scenarios where the MaxFOV2 will likely be used. The readers used a 5 point Likert scale to score the images for each of the following aspects: depiction of the patient's skin surface; depicted tissue densities in the extended FOV region; and overall image quality. The results of the study support substantial equivalence and performance claims.

Substantial Equivalence

MaxFOV 2 was developed under GE Healthcare's quality system. MaxFOV2's design, verification, validation and risk management processes did not identify any new hazards, unexpected results, or adverse effects stemming from the changes to the predicate.

Software verification & validation, along with bench testing and the clinical reader study demonstrate that MaxFOV 2 is substantially equivalent and hence as safe and as effective as the legally marketed predicate device, hence is safe and effective for its intended use.

The substantial equivalence is also based on the submitted software documentation that is consistent with the appropriate Level of Concern as recommended by pre-market software guidance.

Conclusion

Based on development under GE Healthcare's quality system, the successful verification and validation, engineering bench and clinical testing, GE Healthcare believes that MaxFOV 2 is substantially equivalent to the predicate device and hence is safe and effective for its intended use.

N/A