K Number
K203548

Validate with FDA (Live)

Date Cleared
2022-02-21

(444 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EndoArmor™+ Surgical Gowns are single-use personal protective equipment intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker of microorganisms, body fluids, and particulate matter.

EndoArmor™+ Surgical Gowns meet the respective Level 3 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Device Description

EndoArmor™+ Surgical Gowns are intended to be worn by health care professionals as personal protective equipment to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter.

The EndoArmor™+ Surgical Gown is constructed from a polyethylene film laminated with nonwoven spunbond polypropylene that provides AAMI Level 3 liquid barrier performance in the critical zones. The EndoArmor™+ Surgical Gown back was designed to allow for airflow and breathability to support user comfort. The back of the gown is constructed from spunbond polypropylene and is non-protective.

The EndoArmor™+ Surgical Gown is single-use, one-size fits most personal protective equipment supplied non-sterile with sterilization instructions and validated sterilization parameters to enable the end-user to sterilize the gown prior to being used in sterile surgical procedures.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for the EndoArmor™+ Surgical Gown, which is a medical device. However, it does not describe an AI/ML device or a study involving human readers and AI assistance for medical imaging. Therefore, many of the requested points related to AI/ML devices, ground truth establishment for training sets, multi-reader multi-case studies, and expert qualifications are not applicable based on the provided document.

The document primarily focuses on the physical and biological performance characteristics of the surgical gown as a personal protective equipment.

Here's the information extracted from the provided text, addressing the applicable points:

Acceptance Criteria and Device Performance for EndoArmor™+ Surgical Gown

The device in question, the EndoArmor™+ Surgical Gown, is a surgical apparel, not an AI/ML diagnostic tool. Therefore, the acceptance criteria and performance data are entirely based on physical, mechanical, and biological safety standards for such garments.

1. A table of acceptance criteria and the reported device performance

Test MethodPurposeAcceptance CriteriaResults
ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityEvaluate the cytotoxicity endpoint.Under the test conditions, the subject device must be non-cytotoxic per ISO 10993-5.Pass.
ISO 10993-7, Biological Evaluation of Medical Devices-Part 7: Ethylene Oxide Sterilization ResidualsEvaluate the EO/ECH residuals.Under the test conditions, the EO/ECH residual levels of the subject device must be below the maximum allowable limits for adult patient populations based on body mass per ISO 10993-7.Pass.
ISO 10993-10, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitizationEvaluate the irritation and skin sensitization endpoint.Under the test conditions, the subject device must be non-irritating and non-sensitizing per ISO 10993-10.Pass.
ASTM F2407-20, Standard Specification for Surgical Gowns Intended for Use in Healthcare FacilitiesEvaluate the standard specification for surgical gown.Under the test conditions, the subject device must meet the required tensile strength, tear resistance, and seam strength per ASTM F2407-20.Pass.
AATCC TM42, Test Method for Water Resistance: Impact PenetrationEvaluate the impact penetration endpoint.Under the test conditions, the subject device must meet the required resistance to the penetration of water per ANSI/AAMI PB70.Pass.
AATCC TM127, Test Method for Water Resistance: Hydrostatic Pressure TestEvaluate the Hydrostatic Pressure Test endpoint.Under the test conditions, the subject device must meet the required resistance to the penetration of water under hydrostatic pressure per ANSI/AAMI PB70.Pass.
ASTM D5034-09, Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test)Evaluate the Grab Test endpoint.Under the test conditions, the subject device must meet the required tensile strength per ASTM F2407-20.Pass.
ASTM D5587-15, Standard Test Method for Tearing Strength of Fabrics by Trapezoid ProcedureEvaluate the Tearing Strength of Fabrics by Trapezoid Procedure.Under the test conditions, the subject device must meet the required tear resistance per ASTM F2407-20.Pass.
ASTM D1683 / D1683M-17, Standard Test Method for Failure in Sewn Seams of Woven FabricsEvaluate the Standard Test Method for Failure in Sewn Seams of Woven FabricsUnder the test conditions, the subject device must meet the required seam strength per ASTM F2407-20.Pass.
16 CFR Part 1610, Standard for the Flammability of Clothing TextilesEvaluate the Flammability of Clothing Textiles.Under the test conditions, the subject device must meet the Class I flammability standard per 16 CFR Part 1610.Pass.

2. Sample size used for the test set and the data provenance:
The document does not specify the exact sample sizes (e.g., number of gowns tested for each criterion) for the performance tests. It states that "The final proposed finished device was tested..." which implies a sufficient sample size was used to meet the standard requirements, but the specific numbers are not provided. Data provenance is not described in terms of country of origin or retrospective/prospective since it involves laboratory testing of physical samples, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is not an AI/ML device or a diagnostic device relying on expert interpretation of medical images. The "ground truth" is established by adherence to physical, chemical, and biological testing standards (e.g., ISO, ASTM, AATCC, CFR standards).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This is not a study involving human reader interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a surgical gown, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device is defined by internationally recognized and FDA-referenced standards for medical garments and biocompatibility (e.g., ISO 10993 series, ASTM F2407-20, AATCC TM42/TM127, 16 CFR Part 1610, and ANSI/AAMI PB70:2012). The determination of "Pass" or "Fail" is based on whether the device meets the quantitative and qualitative requirements set forth in these standards.

8. The sample size for the training set:
Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:
Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 21, 2022

Boston Scientific Corporation Laura Meehan Principal Regulatory Affairs Specialist 100 Boston Scientific Wav Marlborough, Massachusetts 01752

Re: K203548

Trade/Device Name: EndoArmor™ + Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: January 20, 2022 Received: January 21, 2022

Dear Laura Meehan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203548

Device Name EndoArmor™ + Surgical Gown

Indications for Use (Describe)

EndoArmor™+ Surgical Gowns are single-use personal protective equipment intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker of microorganisms, body fluids, and particulate matter.

EndoArmor™+ Surgical Gowns meet the respective Level 3 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for EndoArmor™+ Surgical Gown K203548

1. Submitter

Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way Marlborough, MA 01752

Contact:

Laura Meehan Principal Regulatory Specialist Phone: (508) 382-0442 E-mail: Laura.Meehan@bsci.com

Date Prepared: January 20, 2022

2. Device

Trade Name:EndoArmor™+ Surgical Gown
Common Name:Surgical Gown
Classification Name:Surgical Apparel
Product Code:FYA
Device Class and Panel:Class II, General Hospital
Classification Regulation:21 CFR §878.4040

3. Predicate Devices

Trade Name:Medline Level 3 Surgical Gown (Sirus Non-Reinforced)
Manufacturer:Medline Industries, Inc.
Clearance Number:K190950
Common Name:Surgical Gown
Classification Name:Surgical Apparel
Product Code:FYA
Device Class and Panel:Class II, General Hospital
Classification Regulation:21 CFR §878.4040

4. Device Description

EndoArmor™+ Surgical Gowns are intended to be worn by health care professionals as personal protective equipment to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter.

The EndoArmor™+ Surgical Gown is constructed from a polyethylene film laminated with nonwoven spunbond polypropylene that provides AAMI Level 3 liquid barrier performance in

{4}------------------------------------------------

the critical zones. The EndoArmor™+ Surgical Gown back was designed to allow for airflow and breathability to support user comfort. The back of the gown is constructed from spunbond polypropylene and is non-protective.

The EndoArmor™+ Surgical Gown is single-use, one-size fits most personal protective equipment supplied non-sterile with sterilization instructions and validated sterilization parameters to enable the end-user to sterilize the gown prior to being used in sterile surgical procedures.

5. Indications for Use

EndoArmor™+ Surgical Gowns are single-use personal protective equipment intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter.

EndoArmor™+ Surgical Gowns meet the respective Level 3 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Technological Characteristics Comparison of the proposed device and predicate device 6.

DeviceCharacteristicProposed DevicePredicate DeviceComparisonAnalysis
Product NameEndoArmor™ + SurgicalGownMedline Level 3 SurgicalGown (Sirus Non-Reinforced)N/A
510(k) ReferenceK203548K190950N/A
Product OwnerBoston ScientificCorporationMedline Industries, Inc.N/A
RegulationNumber21 CFR §878.404021 CFR §878.4040Identical
Product CodeFYAFYAIdentical
DeviceCharacteristicProposed DevicePredicate DeviceComparisonAnalysis
Intended UseEndoArmorTM+ SurgicalGowns are single-usepersonal protectiveequipment intended to beworn by healthcareprofessionals to helpprotect both the patient andthe healthcare worker fromthe transfer ofmicroorganisms, bodyfluids, and particulatematter.EndoArmorTM+ SurgicalGowns meet the respectiveLevel 3 requirements ofANSI/AAMI PB70:2012Liquid barrier performanceand classification ofprotective apparel anddrapes intended for use inhealth care facilities.Medline Level 3 SurgicalGowns (Sirus Non-Reinforced) are sterile,single use surgical apparelintended to be worn byhealthcare professionals tohelp protect both thepatient and the healthcareworkerfrom the transfer ofmicroorganisms, bodyfluids, and particulatematter.Medline Level 3 SurgicalGowns (Sirus Non-Reinforced) meet therespective Level 3requirements ofANSI/AAMI PB70:2012Liquid barrier performanceand classification ofprotective apparel anddrapes intended for use inhealth care facilities.Similar
ColorBlueBlueIdentical
Design FeaturesFabric Non-ReinforcedHook and Loop Closure atneckBelt TiesKnit CuffsStandard SleevesFabric Non-ReinforcedHook and Loop Closure atneckBelt TiesKnit CuffsStandard SleevesIdentical
SizesOne-size fits mostSmall to XXXX-LargeSimilar
MaterialsFront/Sleeves:Polyethylene FilmLaminated with NonwovenSpunbond PolypropyleneBack: NonwovenSpunbond PolypropyleneNonwoven SMSPolypropylene/PolyolefinSimilar
Critical ZonePerformanceSpecificationsAAMI Level 3 BarrierProtectionAAMI Level 3 BarrierProtectionIdentical
DeviceProposed DevicePredicate DeviceComparison
CharacteristicAnalysis
Prescription vs.OTCOTCOTCIdentical
ContactDurationSurface, Intact, < 24 hoursSurface, Intact, < 24 hoursIdentical
Sterile vs.Non-SterileSold as single-use, non-sterile to end user withsterilization instructionsand validated EtOsterilization methodSterile (Ethylene Oxide) orsold as bulk single-use,non-sterile to re-packager/re-labelerestablishments for furtherpackaging and sterilizationusing the validated EtOsterilization methodaccording to ISO 11135-1prior to being provided tothe end userIdentical
Single Use vs.ReusableSingle UseSingle UseIdentical
BiocompatibilityUnder the test conditions,the subject device wasshown to be non-cytotoxic,non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10.Under the test conditions,the subject device wasshown to be non-cytotoxic,non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10.Identical
FlammabilityMeets requirements ofFlame Resistant CPSC1610 Class 1Meets requirements ofFlame Resistant CPSC1610 Class 1Identical

Table 5-1: Comparison of Proposed and Predicate Devices

{5}------------------------------------------------

{6}------------------------------------------------

7. Performance Data

The final proposed finished device was tested in accordance with Guidance on Premarket Notification [510(k)] Submission for Surgical Gowns and Surgical Drapes, August 1993, Guidance for Industry and FDA Staff - Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings issued by the FDA on December 9, 2015 and ANSI/AAMI PB70:2012, Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Test MethodPurposeAcceptance CriteriaResults
ISO 10993-5, Biologicalevaluation of medicaldevices - Part 5: Tests forin vitro cytotoxicityThe purpose of this testmethodology is toevaluate the cytotoxicityendpoint.Under the test conditions,the subject device must benon-cytotoxic per ISO10993-5.Pass.
ISO 10993-7, BiologicalEvaluation of MedicalDevices-Part 7: EthyleneOxide SterilizationResidualsThe purpose of this testmethodology is toevaluate the EO/ECHresiduals.Under the test conditions,the EO/ECH residuallevels of the subjectdevice must be below themaximum allowable limitsfor adult patientpopulations based on bodymass per ISO 10993-7.Pass.
Test MethodPurposeAcceptance CriteriaResults
ISO 10993-10, Biologicalevaluation of medicaldevices- Part 10: Tests forirritation and skinsensitizationThe purpose of this testmethodology is toevaluate the irritation andskin sensitizationendpoint.Under the test conditions,the subject device must benon-irritating and non-sensitizing per ISO 10993-10.Pass.
ASTM F2407-20,Standard Specification forSurgical Gowns Intendedfor Use in HealthcareFacilitiesThe purpose of this testmethodology is toevaluate the standardspecification for surgicalgown.Under the test conditions,the subject device mustmeet the required tensilestrength, tear resistance,and seam strength perASTM F2407-20.Pass.
AATCC TM42, TestMethod for WaterResistance: ImpactPenetrationThe purpose of this testmethodology is toevaluate the impactpenetration endpoint.Under the test conditions,the subject device mustmeet the requiredresistance to thepenetration of water perANSI/AAMI PB70.Pass.
AATCC TM127, TestMethod for WaterResistance: HydrostaticPressure TestThe purpose of this testmethodology is toevaluate the HydrostaticPressure Test endpoint.Under the test conditions,the subject device mustmeet the requiredresistance to thepenetration of water underhydrostatic pressure perANSI/AAMI PB70.Pass.
ASTM D5034-09,Standard Test Method forBreaking Strength andElongation of TextileFabrics (Grab Test)The purpose of this testmethodology is toevaluate the Grab Testendpoint.Under the test conditions,the subject device mustmeet the required tensilestrength per ASTMF2407-20.Pass.
ASTM D5587-15,Standard Test Method forTearing Strength ofFabrics by TrapezoidProcedureThe purpose of this testmethodology is toevaluate the TearingStrength of Fabrics byTrapezoid Procedure.Under the test conditions,the subject device mustmeet the required tearresistance per ASTMF2407-20.Pass.
ASTM D1683 / D1683M-17, Standard Test Methodfor Failure in Sewn Seamsof Woven FabricsThe purpose of this testmethodology is toevaluate the Standard TestMethod for Failure inSewn Seams of WovenFabricsUnder the test conditions,the subject device mustmeet the required seamstrength per ASTMF2407-20.Pass.
16 CFR Part 1610,Standard for theFlammability of ClothingTextilesThe purpose of this testmethodology is toevaluate the Flammabilityof Clothing Textiles.Under the test conditions,the subject device mustmeet the Class Iflammability standard per16 CFR Part 1610.Pass.

{7}------------------------------------------------

8. Conclusion

The conclusions drawn from the nonclinical tests that demonstrate that the EndoArmor™+ Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.