510(k) Data Aggregation

    K Number
    K213844
    Device Name
    Surgical Gown
    Date Cleared
    2023-02-16

    (434 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K190950

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Surgical gowns are intended to be worn by operating room personnel procedure to protect both the surgical patient and the operating room persomel from transfer of microorganisms, body fluids, and particulate material. This device is provided sterile.

    Per ASNI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, the standard (non-reinforced) surgical gowns meet Level 2 classification and the reinforced surgical gowns meet Level 3 classification.

    Device Description

    Surgical gowns are intended to be worn by healthcare professionals to protect both the patient and the healthcare professional from the transfer or microorganisms, body fluids, and particulate material. The proposed devices are single use, disposable medical devices and ethylene oxide (EO) sterilized. They are available in six sizes: S, M, L, XL, XXL and XXXL.

    The proposed surgical gowns are constructed of a Spunbond (SMS) nonwoven material, with cuffs made of polyester and elastic, offered in standard (non-reinforced) and reinforced surgical gowns, in blue color.

    AI/ML Overview

    This is a 510(k) summary for a surgical gown, which is a low-risk medical device. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of AI algorithm performance. Therefore, many of the requested criteria, which are typically relevant for AI/ML device evaluations, are not applicable here.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document provides a summary of performance testing in Table 2, outlining the test method, purpose, acceptance criteria, and result (Pass/Fail or specific finding).

    Test MethodPurposeAcceptance CriteriaResult
    AATCC 127Hydrostatic pressureLevel 2 > 20 cm, Level 3 > 50 cmPass
    AATCC 42Impact penetration≤1gPass
    ASTM F 2407 (ASTM D 5034)Tensile StrengthLength Direction ≥ 115N, Width Direction ≥ 115NPass
    ASTM D 5587Tearing strength by TrapezoidLength Direction > 20 N, Width Direction > 20 NPass
    ASTM D 1683Seam strength≥ 20 NPass
    16 CFR Part 1610Flammability of TextilesClass 1Class 1
    ISO 10993-7Ethylene oxide residuesEO residual ≤ 4 mg/device, ECH residual ≤ 9 mg/devicePass
    ISO 9073-10LintingLog10