K Number
K203521

Validate with FDA (Live)

Device Name
FineCross M3
Date Cleared
2021-03-03

(92 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The product (FineCross M3) is intended to be percutaneously into blood vessels and support a guide wire while performing PCI (percutaneous coronary intervention). The product is also intended for injection of radiopaque contrast media for angiography. The product should not be used in cerebral and peripheral vessels.

Device Description

FineCross M3 is a single use, ethylene oxide sterilized device that is intended to be percutaneously introduced into blood vessels and support a guide wire while performing PCI (percutaneous coronary intervention). The product is also intended for injection of radiopaque contrast media for angiography. FineCross M3 features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyester elastomer and an inner layer of polytetrafluoroethylene. The outer surface of the catheter is coated with hydrophilic polymer.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the FineCross M3 device, structured to answer your specific questions.

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a list of performance tests conducted on the FineCross M3 device. For each test, it states that "Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life." However, it does not provide specific numerical or qualitative values for the acceptance criteria, nor does it detail the specific reported device performance values for each test. Instead, it offers a general statement of compliance.

Test ItemAcceptance CriteriaReported Device Performance
Radio-detectabilityPredetermined acceptance criteria (not specified)Met acceptance criteria
SurfacePredetermined acceptance criteria (not specified)Met acceptance criteria
Peak tensile forcePredetermined acceptance criteria (not specified)Met acceptance criteria
Freedom from leakagePredetermined acceptance criteria (not specified)Met acceptance criteria
FlowratePredetermined acceptance criteria (not specified)Met acceptance criteria
Distal tipPredetermined acceptance criteria (not specified)Met acceptance criteria
Fluid leakagePredetermined acceptance criteria (not specified)Met acceptance criteria
Sub-atmospheric pressure air leakagePredetermined acceptance criteria (not specified)Met acceptance criteria
Stress crackingPredetermined acceptance criteria (not specified)Met acceptance criteria
Resistance to separation from axial loadPredetermined acceptance criteria (not specified)Met acceptance criteria
Resistance to separation from unscrewingPredetermined acceptance criteria (not specified)Met acceptance criteria
Resistance to overridingPredetermined acceptance criteria (not specified)Met acceptance criteria
Torque strengthPredetermined acceptance criteria (not specified)Met acceptance criteria
Kink strength of catheter shaftPredetermined acceptance criteria (not specified)Met acceptance criteria
Compatibility with guide wirePredetermined acceptance criteria (not specified)Met acceptance criteria
Product dimensionPredetermined acceptance criteria (not specified)Met acceptance criteria
Exterior sliding characteristics (early phase)Predetermined acceptance criteria (not specified)Met acceptance criteria
Particulate evaluationPredetermined acceptance criteria (not specified)Met acceptance criteria
Coating IntegrityPredetermined acceptance criteria (not specified)Met acceptance criteria
Butting resistancePredetermined acceptance criteria (not specified)Met acceptance criteria
Strength of distal partPredetermined acceptance criteria (not specified)Met acceptance criteria
Simulated use Usability testPredetermined acceptance criteria (not specified)Met acceptance criteria

Biocompatibility Testing:

Test ItemAcceptance CriteriaReported Device Performance
CytotoxicityCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
SensitizationCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Intracutaneous ReactivityCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Acute Systemic ToxicityCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
PyrogenicityCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
HemolysisCompliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Thrombogenicity (with and without anticoagulant agent)Compliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Complement Activation (C3a and SC5b-9)Compliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Physicochemical Profile (Physicochemical and FT-IR)Compliance with ISO 10993-1 and FDA GuidanceDemonstrate device is biocompatible
Accelerated-aged (2 years) CytotoxicityCompliance with ISO 10993-1 and FDA Guidance (to show biocompatibility maintained throughout shelf life)Demonstrate device is biocompatible throughout shelf life
Accelerated-aged (2 years) HemolysisCompliance with ISO 10993-1 and FDA Guidance (to show biocompatibility maintained throughout shelf life)Demonstrate device is biocompatible throughout shelf life
Accelerated-aged (2 years) Physicochemical ProfileCompliance with ISO 10993-1 and FDA Guidance (to show biocompatibility maintained throughout shelf life)Demonstrate device is biocompatible throughout shelf life

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for each of the performance or biocompatibility tests. It only states that tests were performed on "non-aged and accelerated aged samples" for performance testing (except Radio-detectability and Simulated Use Usability) and on "non-aged, sterile, whole device" and "accelerated-aged (2 years), sterile, whole device" for biocompatibility.

The data provenance is industrial (manufacturer-conducted testing) and likely combines both novel testing for this specific device and potentially established testing protocols based on industry standards. It is not patient or clinical data, so terms like "retrospective" or "prospective" are not applicable in this context. The country of origin for the manufacturing and testing is Japan (Ashitaka Factory of Terumo Corporation).


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes non-clinical performance and biocompatibility testing of a medical device (a microcatheter). It does not involve diagnostic interpretation or patient data where "ground truth" would typically be established by human experts like radiologists. Therefore, this information is not applicable to the provided document. The ground truth for these tests is based on objective, measurable physical and chemical properties and engineering standards.


4. Adjudication Method for the Test Set

As this document describes non-clinical performance and biocompatibility testing, an "adjudication method" in the context of expert consensus (like 2+1 or 3+1 for clinical interpretations) is not applicable. The results of these tests are determined by adherence to pre-defined scientific and engineering protocols and acceptance criteria.


5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no AI component mentioned in this 510(k) submission. This K-submission is for a medical device (microcatheter), not an AI/software-as-a-medical-device.


6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable as the device is a physical medical instrument (microcatheter), not an algorithm or AI system.


7. The Type of Ground Truth Used (expert concensus, pathology, outcomes data, etc)

For the performance tests, the "ground truth" is defined by engineering specifications, material science principles, and established industry standards. For example, the freedom from leakage is tested against a standard preventing fluid escape, and material biocompatibility is tested against ISO 10993 standards and FDA guidance. This is not a "ground truth" derived from expert consensus, pathology, or outcomes data in a clinical sense.


8. The Sample Size for the Training Set

There is no training set in this context. This is a physical medical device, not a machine learning model.


9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no training set mentioned in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.

March 3, 2021

Terumo Medical Corporation Vaibhav Sivaramakrishan Regulatory Affairs Specialist II 265 Davidson Ave., Suite 320 Somerset, New Jersey 08873

Re: K203521

Trade/Device Name: FineCross M3 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 30, 2020 Received: December 1, 2020

Dear Vaibhav Sivaramakrishan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203521

Device Name FineCross M3

Indications for Use (Describe)

The product (FineCross M3) is intended to be percutaneously into blood vessels and support a guide wire while performing PCI (percutaneous coronary intervention). The product is also intended for injection of radiopaque contrast media for angiography. The product should not be used in cerebral and peripheral vessels.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the Terumo logo. The logo consists of a red swoosh above the word "TERUMO" in green. The swoosh is curved and points to the right, and the word "TERUMO" is in a bold, sans-serif font.

510(K) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by: Vaibhav Sivaramakrishan Regulatory Affairs Specialist II Terumo Medical Corporation Tel. (609) 613-3958 Fax (410) 398-6079

Prepared for:Owner/Operator
Terumo Corporation
44-1, 2-Chome, Hatagaya
Shibuya-Ku, Tokyo
Japan 151-0072
Owner/Operator Number: 8010026

Manufacturer and Sterilization Facility (Applicant) Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya, Shizuoka 418-0015, Japan Registration Number: 9681834

  • Contact Person: Vaibhav Sivaramakrishan Regulatory Affairs Specialist II Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, NJ 08873 Tel. (609) 613-3958 Fax (410) 398-6079 E-mail: vai.sivaramakrishan@terumomedical.com
    Date prepared: November 30, 2020

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red curved line above the word "TERUMO", which is written in green. The red curved line is positioned above and to the left of the word "TERUMO", creating a dynamic and modern look.

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:FineCross M3
Common Name:Microcatheter
Classification Name:Catheter, Percutaneous
Classification Panel:Cardiovascular
Regulation:21 CFR 870.1250
Product Code:DQY
Classification:Class II

C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device to which substantial equivalence is claimed is:

Predicate Device: K082519 - FINECROSS MG Coronary Micro-Guide catheter, manufactured by Ashitaka Factory of Terumo Corporation.

Reference Devices:

    1. K152447: ASAHI Caravel, Asahi Intecc co., ltd.
    1. K191560: Turnpike LP Catheter, Vascular Solution LLC.

D. REASON FOR 510(k) SUBMISSION

This traditional 510(k) for FineCross M3 is being submitted for the new device for the purposes of establishing substantial equivalence to a legally marketed predicate device.

E. DEVICE DESCRIPTION (807.92(a)(4))

Principle of Operation Technology

FineCross M3 submitted in this 510(k) and its predicate (K082519) are operated by a manual process.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red arc above the word "TERUMO" in green, block letters. The arc is positioned above and to the left of the word, creating a sense of motion.

Design/Construction

FineCross M3 is a single use, ethylene oxide sterilized device that is intended to be percutaneously introduced into blood vessels and support a guide wire while performing PCI (percutaneous coronary intervention). The product is also intended for injection of radiopaque contrast media for angiography.

FineCross M3 features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyester elastomer and an inner layer of polytetrafluoroethylene. The outer surface of the catheter is coated with hydrophilic polymer.

Materials

The materials for FineCross M3 are provided in Table 5.1.

No.Raw materialPatient Contact
1*Polyester elastomerColorantDirect
2Stainless steelNon-contact
3*PolytetrafluoroethyleneDirect
4Au (Gold)Non-contact
5*Dimethyl acrylamide-glycidylmethacrylate copolymerDirect
6Acrylic resinNon-contact
7*PolyamideIndirect
8Polyester elastomerColorantNon-contact

Table 5.1: List of Materials

  • Blood contacting material (External communicating Device, Circulating Blood, Limited Contact (<24 hours)

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Terumo logo. The logo consists of a red arc above the word "TERUMO" in green. The arc is positioned above and to the left of the word "TERUMO". The font of the word "TERUMO" is sans-serif and bold.

Specifications

The specifications for FineCross M3 are provided in Table 5.2.

PartSpecifications
Catheter Size:Distal 1.8 Fr. / Proximal 2.5Fr.
Catheter OD:Distal OD 0.61mmProximal OD 0.84mm
Catheter Effective Lengths*:1300mm, 1500mm
Hydrophilic Coating Length:750mm, 950mm
Maximum guidewire outer diameter:0.014"
Table 5.2: FineCross M3 Specifications
--------------------------------------------

*The length from the proximal anti-kink protector to distal end of catheter shaft.

F. INDICATIONS FOR USE (807.92(a)(5))

The product (FineCross M3) is intended to be percutaneously introduced into blood vessels and support a guide wire while performing PCI (percutaneous coronary intervention). The product is also intended for injection of radiopaque contrast media for angiography. The product should not be used in cerebral and peripheral vessels.

The indications for use are equivalent to the predicate (K082519).

G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

FineCross M3, the subject of this Traditional 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K082519 - FINECROSS MG Coronary Micro-Guide catheter, manufactured by Ashitaka Factory of Terumo Corporation.

In addition to the above-listed primary predicate, Terumo has identified the following reference devices. These are market leading devices with the same intended use and basic design as the subject device. Since these devices are frequently used in clinical practice, Terumo felt it was appropriate to use them as references when setting the acceptance criteria for FineCross M3 performance testing.

    1. Asahi Intecc co., Itd. ASAHI Caravel (K152447)
    1. Vascular Solution LLC., Turnpike LP Catheter (K191560)

The comparison of the technological characteristics is summarized in Table 5.3.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red arc above the company name. The company name "TERUMO" is written in green, block letters.

Table 5.3: Summary of Comparative Information
DeviceCharacteristicSubject Device:FineCross M3Predicate Device:FINECROSS MG CoronaryMicro-Guide catheter(K082519)Reference Device #1:ASAHI Caravel(K152447)Reference Device #2:Turnpike LP Catheter(K191560)
ManufacturerAshitaka Factory of TerumoCorporationSameAsahi Intecc co., ltd.Vascular Solution LLC.
Intended Use/Indications forUseThe product (FineCross M3) isintended to be percutaneouslyintroduced into blood vesselsand support a guide wire whileperforming PCI (percutaneouscoronary intervention). Theproduct is also intended forinjection of radiopaquecontrast media forangiography. The productshould not be used in cerebraland peripheral vessels.The product (Finecross™ MG)is intended to bepercutaneously introduced intoblood vessels and support aguide wire while performingPCI (percutaneous coronaryintervention. The product isalso intended for injection ofradiopaque contrast media forangiography.This microcatheter isintended to provide supportto facilitate the placementof guide wires in thecoronary and peripheralvasculatures, and can beused to exchange one guidewire for another.This microcatheter is alsointended to assist in thedelivery of contrast mediainto the coronary andperipheral vasculatures.Do not use thismicrocatheter other thanfor use in the coronary andperipheral vasculatures.The Turnpike catheters areintended to be used to accessdiscrete regions of thecoronary and/or peripheralvasculature.They may be used to facilitateplacement and exchange ofguidewires and tosubselectively infuse/deliverdiagnostic and therapeuticagents.
OperationPrincipleManualSameSameSame
DeviceCharacteristicSubject Device:FineCross M3Predicate Device:FINECROSS MG CoronaryMicro-Guide catheter(K082519)Reference Device #1:ASAHI Caravel(K152447)Reference Device #2:Turnpike LP Catheter(K191560)
Design/ConstructionThree layer constructioncatheter shaft with hydrophiliccoating, distal tip and hubSameMulti layers constructioncatheter shaft withhydrophilic coating, distaltip and hubMulti layers constructioncatheter shaft with hydrophiliccoating, distal tip and hub
Materials- Polyester elastomer w/pigment*- Stainless steel- Polytetrafluoroethylene*- Au (Gold)- Dimethyl acrylamide-glycidyl methacrylatecopolymer*- Polyamide*- Polyester elastomer w/pigment- Acrylic resin* blood contacting material- Polyester elastomer w/pigment*- Stainless steel- Polytetrafluoroethylene*- Au (Gold)- Dimethyl acrylamide-glycidyl methacrylatecopolymer*- Polyamide*- Polyester elastomer w/pigment- Acrylic resin*blood contacting materialInformation not publiclyavailable.Information not publiclyavailable.
Package• Individual package on whichthe product label and theSameSameSame
DeviceCharacteristicSubject Device:FineCross M3Predicate Device:FINECROSS MG CoronaryMicro-Guide catheter(K082519)Reference Device #1:ASAHI Caravel(K152447)Reference Device #2:Turnpike LP Catheter(K191560)
peel-off labels are attached• 1 unit per package
Specifications• Effective lengths:1300mm, 1500mm• Catheter size:Distal 1.8 Fr. / Proximal2.5Fr.• Catheter OD:Distal:0.61mmProximal:0.84mm• Maximum guidewire outerdiameter: 0.014"• Effective lengths:1300mm,1500mm• Catheter size:Distal 1.8 Fr. / Proximal2.6Fr.• Catheter OD:Distal:0.60mmProximal:0.87mm• Maximum guidewireouter diameter: 0.014"• Effective lengths:1350mm,1500mm• Catheter size:Distal 1.9 Fr. /Proximal 2.6Fr.• Catheter OD:Distal:0.62mmProximal:0.85mm• Maximum guidewireouter diameter:Unknown• Effective lengths:1350mm,1500mm• Catheter size:Distal 2.2 Fr. / Proximal2.9Fr.• Catheter OD:Distal:0.74mmProximal:0.97mm• Maximum guidewireouter diameter: 0.014"
SterilizationEthylene oxideSameSameSame
Shelf life24 monthsSameunknownunknown

of Ca ative Info Tahle 5 3. Cr noti

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red swoosh above the letters "TE". The font is sans-serif and the letters are bold.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the Terumo logo. The logo consists of the word "TERUMO" in green, with a red arc above the "T". The font is bold and sans-serif. The logo is simple and modern.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red arc above the company name, which is written in green, sans-serif font. The arc is positioned above and slightly to the left of the word "TERUMO."

H. NON CLINICAL TESTS (807.92(b)(1))

Performance Testing

Performance testing was conducted to ensure the substantial equivalence of FineCross M3 throughout the shelf life and to verify conformity to the applicable external and internal standards. With the exception of the Radio-detectability and Simulated Use Usability tests, the following performance tests were performed on non-aged and accelerated aged samples. Table 5.4 provides a list of performance tests that were performed on FineCross M3.

Test Item
Radio-detectability
Surface
Peak tensile force
Freedom from leakage
Flowrate
Distal tip
Fluid leakage
Sub-atmospheric pressure air leakage
Stress cracking
Resistance to separation from axial load
Resistance to separation from unscrewing
Resistance to overriding
Torque strength
Kink strength of catheter shaft
Compatibility with guide wire
Product dimension
Exterior sliding characteristics (early phase)
Particulate evaluation
Coating Integrity
Butting resistance
Strength of distal part
Simulated use Usability test

Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the Terumo company logo. The logo consists of a red arc above the company name, which is written in green, block letters. The arc is positioned above and to the left of the company name.

Biocompatibility

In accordance with ISO 10993-1, FineCross M3 is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. The table below provides a list of biocompatibility tests conducted on FineCross M3.

Non-aged, sterile, whole device
Cytotoxicity
Sensitization
Intracutaneous Reactivity
Acute Systemic Toxicity
Pyrogenicity
Hemolysis
Thrombogenicity (with and without anticoagulant agent)
Complement Activation (C3a and SC5b-9)
Physicochemical Profile (Physicochemical and FT-IR)
Accelerated-aged (2 years), sterile, whole device
Cytotoxicity
Hemolysis
Physicochemical Profile (Physicochemical and FT-IR)

Table 5.5: Summary of ISO 10993 Biocompatibility Testing

Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red swoosh above the letters "TE". The font is sans-serif and the letters are bold.

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2018, Sterilization of Health Care Products – Ethylene Oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6.

I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

J. CONCLUSION (807.92(b)(3))

In summary, FineCross M3, subject of this 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K082519 - FINECROSS MG Coronary Micro-Guide catheter, manufactured by Ashitaka Factory of Terumo Corporation.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).