Infrared Thermometer (Model K9)

K203497 · Kangzhan Communication Electronics Co., Ltd. · FLL · Oct 8, 2021 · General Hospital

Device Facts

Record IDK203497
Device NameInfrared Thermometer (Model K9)
ApplicantKangzhan Communication Electronics Co., Ltd.
Product CodeFLL · General Hospital
Decision DateOct 8, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Intended Use

The Infrared Thermometer, Model K9 is a non-contact infrared thermometer intended for the intermittent measurement and monitoring of human body temperature from forehead for clinical and home use environments. It can be used for anybody, e.g. for infant, children and adults (Excluding neonates). The device does not require sterilization and can be reused.

Device Story

Non-contact infrared thermometer; measures infrared energy radiated from forehead. Input: infrared energy collected via lens; processed by thermopile sensor and ASIC. Features CapSense proximity technology to ensure measurement within 5 cm distance; device triggers automatic measurement when correct distance detected. Output: temperature value displayed on LED screen with buzzer alert. Used in clinical and home environments by clinicians or patients. Provides temperature readings to assist in clinical decision-making; benefits patient through non-invasive, reusable temperature monitoring.

Clinical Evidence

Clinical accuracy validation conducted per ISO 80601-2-56:2017 (including Amendment 1). Bench testing confirmed compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Biocompatibility testing (cytotoxicity, skin sensitization, skin irritation) performed per ISO 10993-1. Usability testing conducted per FDA human factors guidance.

Technological Characteristics

Materials: ABS+PC (housing/buttons), PC (lens). Sensing: Thermopile detector (TU05) with integrated thermistor for target/ambient compensation. Proximity: Capacitance-touch sensor. Power: 2x AAA batteries (3V). Connectivity: None. Sterilization: None (reusable). Software: Moderate level of concern. Standards: ISO 80601-2-56, ASTM E1965, IEC 60601-1, IEC 60601-1-2.

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature from the forehead in clinical and home environments. Suitable for infants, children, and adults; excluding neonates.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 8, 2021 Kangzhan Communication Electronics Co., Ltd. Su Baotong General Manager No. 1008, songbai road, yangguang community, xili street, Nanshan District Shenzhen, 518055 China Re: K203497 Trade/Device Name: Infrared Thermometer (Model K9) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 30, 2021 Received: September 8, 2021 Dear Su Baotong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K203497 Device Name Infrared Thermometer (Model K9) Indications for Use (Describe) The Infrared Thermometer, Model K9 is a non-contact intended for the intermittent measurement and monitoring of human body temperature from forehead for clinical and home use environments. It can be used for anybody, e.g. for infant, children and adults (Excluding neonates). The device does not require sterilization and can be reused. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) > Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for KANGZHAN. The logo consists of a blue rectangle with an orange parallelogram on top. The text "KANGZHAN" is written in a sans-serif font below the rectangle. # K203497 - 510(K) SUMMARY This summary of 510(K) safety and effective information is being submitted in accordance with the requirement SMDA and 21 CFR 807.92. # 1. Submitter of 510(K): | Date of Prepared: | October 8, 2021 | |-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Name: | Kangzhan Communication Electronics Co., Ltd. | | Address: | 4th floor, building F, qihang kechuang industrial park,<br>No. 1008, songbai road, yangguang community, xili street, nanshan<br>district, shenzhen City,China | | Contact person: | Mr. Su Baotong | | TEL: | 0755-88377262/<br>13602582167 | | FAX: | / | | Email: | kzs168@126.com | # 2. Proposed Device and code: | | Device Trade Name: | Infrared Thermometer (Model K9) | |-------------------|--------------------|---------------------------------| | Common Name: | | Clinical Electronic Thermometer | | Product Code: | | FLL | | Regulation Name: | | Clinical electronic thermometer | | Regulation number | | 21 CFR 880.2910 | | Device Class | | 2 | # 3. Predicate Device: | 510(K) | Trade or Proprietary or Model Name | Manufacturer | |--------|------------------------------------|--------------| |--------|------------------------------------|--------------| {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for KANGZHAN. The logo is a blue and orange geometric shape, with the word "KANGZHAN" written in blue below it. The geometric shape is a square with a triangle cut out of the top, with the top triangle being orange and the bottom square being blue. | K191668 | Infrared Forehead Thermometer (Model IR-FT) | Comper Chuangxiang<br>(Beijing) technology Co.,<br>Ltd. | |---------|---------------------------------------------|---------------------------------------------------------| |---------|---------------------------------------------|---------------------------------------------------------| # 4. Description of Proposed Device: The Infrared Thermometer. Model K9 is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value. The device uses CapSense Technology to detect distance this technology is used to assist measurement. The concept of proximity sensor detects human electrical proximity level to transform to distance. The distance control feature is added to the user interface to identify the measurement distances are in the correct parameter. In other words, the device will enter into measurement mode after the correct measurement distance is detected. This device can take a measurement automatically when the device detects the distance is appropriate within 5 cm. The Infrared Thermometer, Model K9, consists of the following parts: - a) Thermopile Sensor - b) Application-Specific Integrated Circuitry - c) Erasable Programmable Read-Only Memory Integrated Circuit - d) Capacitance-touch Integrated Circuit - e) LED and Backlight - f) Alkaline batteries; size AAA, 2 x 1.5 V - g) Lens ## 5. Indications for Use The Infrared Thermometer, Model K9 is a non-contact infrared thermometer intended for the intermittent measurement and monitoring of human body temperature from forehead for clinical and home use environments. It can be used for anybody, e.g. for infant, children and adults (Excluding neonates). The device does not require sterilization and can be reused. #### Technical and Performance 6. The following table compares the device to the predicate device with basic technological characteristics. Both subject and predicate devices use infrared technology to measure and monitor the body temperature by the site of Forehead. Infrared Thermometer K9 has been compared to the Infrared Forehead Thermometer Infrared Forehead Thermometer Models: IR-FT (K191668) as a predicate device for substantial equivalence. A table comparing the two devices is provided as follows: {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for KANGZHAN. The logo consists of a blue trapezoid with an orange trapezoid on top of it. Below the trapezoid is the word "KANGZHAN" in a simple sans-serif font. | Elements of<br>Comparison | Subject device | Predicate device | Similar<br>or<br>Different | |----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------| | 510(k) Number | K203497 | K191668 | / | | Indication for<br>use | The Infrared Thermometer, Model K9<br>is a non-contact infrared thermometer<br>intended for the intermittent<br>measurement and monitoring of human<br>body temperature from forehead for<br>clinical and home use environments. It<br>can be used for anybody, e.g. for<br>infant, children and adults (Excluding<br>neonates). The device does not require<br>sterilization and can be reused. | This device is a non-sterile, reusable,<br>contact Infrared Forehead<br>Thermometer (Model IR- FT)<br>intended for intermittent<br>determination of human body<br>temperature through a touch on the<br>center of the forehead as the<br>measurement site on people of all<br>ages. The Infrared Forehead<br>Thermometer (Model IR- FT) can be<br>used inclinical and home<br>environments. | Same | | Thermometer<br>type | Infrared forehead | Infrared forehead | Same | | Components | The product is mainly<br>composed of a temperature sensor, a<br>housing, a circuit board and a battery<br>compartment. | The product is mainly<br>composed of a temperature sensor, a<br>housing, a circuit board and a battery<br>compartment. | Same | | Temperature<br>Measurement<br>Technology | The thermoneter uses a thermopile<br>sensor with integrated thermistor for<br>the target reading, a thermistor<br>mounted in the head of the<br>thermometer for ambient<br>temperature readings, and an<br>infrared proximity sensor for<br>detection of contact use and<br>compensation of the temperature<br>reading. | The thermometer uses a thermopile<br>sensor with integrated thermistor for<br>the target reading, a thermistor<br>mounted in the head of the<br>thermometer for ambient temperature<br>readings, and an infrared proximity<br>sensor for<br>detection of contact use and<br>compensation of the temperature<br>reading. | Same | | Key<br>Temperature<br>Sensor | High-end Thermopile Detector<br>TU05 manufactured by TE<br>Connectivity inc | Thermopile Sensor<br>OTP-336 manufactured by Oriental<br>system technology inc. | Same | | Power<br>requirements | DC 3V(2 AAA batteries) | D.C. 3 V (2 AAA dry batteries) | Same | | Materials | Patient contacting materials include<br>ABS+PC (Top cover, battery cover,<br>chamber, button power,front cover)<br>and PC(lens) | Patient contacting materials include<br>PC (Top cover, battery cover,<br>chamber, button power) and<br>ABS+TPU (front cover). | Similar<br>(note 1#) | | Temperature<br>range | 32.0°C ~42.9°C (89.6°F -109.2°F) | 32.0°C~43.0°C (89.6°F -109.4°F) | Similar<br>(note 2#) | | Accuracy for<br>body<br>temperature<br>measurement | ≤±0.2°C (0.4°F), for the range 35°C<br>-42°C (95°F -107.6°F); | ≤±0.2°C (0.4°F), for the range 35°C -<br>42°C (95°F -107.6°F); | Same | | | ≤±0.3°C (0.5°F),for the range 32°C -<br>34.9°C (89.6°F -94.8°F) and<br>42.1-42.9°C (107.8°F -109.2°F) | ≤±0.3°C (0.5°F),for the range 32°C -<br>34.9°C (89.6°F -94.8°F) and<br>42.1-43°C (107.8°F -109.4°F) | | | Contact/non<br>contact use | Non contact use | None contact use | Same | | Measurement<br>site | Forehead | Forehead | Same | | Measurement<br>distance | Within 3 cm | Within 3 cm | Same | | operating<br>mode | adjusted mode | adjusted mode | Same | | reference<br>body site | oral | oral | Same | | Resolution of<br>display | 0.1°C / 0.1°F | 0.1°C / 0.1°F | Same | | Signal output<br>And display | LED, Buzzer | LCD, Buzzer | Similar<br>(note 3#) | | Biocompatibility | ISO 10993-1 | ISO 10993-1 | Same | | Voluntary<br>standards for<br>Clinical<br>Electronic<br>Thermometers | ISO 80601-2-56<br>ASTM E1965 | ISO 80601-2-56<br>ASTM E1965 | Same | | Medical<br>Electrical Safety<br>and EMC | IEC 60601-1, IEC 60601-<br>1-2 | IEC 60601-1, IEC 60601-<br>1-2 | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo for KANGZHAN. The logo consists of a blue square with rounded corners. Inside the square is an orange shape that resembles a checkmark or an arrow pointing upwards and to the right. Below the square, the word "KANGZHAN" is written in a simple, sans-serif font. #### Discussion: From the comparison table1 and table2, the subject devices and predicate devices have the similar Intended use & Indications for Use, same measurement place, Scale selection, Display screen, Auto power-off while no operation & Conformance standard. There are slightly differences between the devices and predicate devices as follows,through the verification and validation process, it has been shown that the differences do not raise new questions of safety and effectiveness. #### Note 1#: Materials: The subject device and predicate device have both performed biocompatibility tests according to the ISO 10993- 1, both devices meet the requirements of ISO 10993-1. Therefore it is substantially equivalent on biocompatibility risk. #### Note 2#: Temperature range The measurement range of the subject device meets the requirements of ISO 80601-2-56 and ASTM E1965-98. Note 3#: Signal output and display Meets the requirement of IEC 60601-1 #### Conclusion: The subject device Infrared Thermometer (Model K9) has all features of the predicate device. The differences between them do not affect the safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device. ### 7. Performance Testing: {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a logo for KANGZHAN. The logo consists of a blue rectangle with an orange parallelogram on top. The word "KANGZHAN" is written in a simple sans-serif font below the rectangle. Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below. #### 7.1 Non-Clinical Data: The following performance data were provided in support of the substantial equivalence determination. #### 7.2 Biocompatibility testing The biocompatibility evaluation for the Non-Contact Infrared Forehead Thermometer were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the testing included the following tests: - Cytotoxicity - Skin Sensitization - Skin Irritation #### 7.3 Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Non-Contact Infrared Forehead Thermometer, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC. #### 7.4 Bench Testing Bench testing was conducted on the Non-Contact Infrared Forehead Thermometer, consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2015 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]. #### 7.5 Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. #### 7.6 Usability Testing Usability testing according to following FDA Guidance 1757. Applying Human Factors and Usability Engineering to Optimize Medical Device Design, was conducted. #### 7.7 Clinical data: Clinical testing is conducted per ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]. Sections 201.102 Clinical Accuracy Validation {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows a logo for KANGZHAN. The logo consists of a blue square with an orange parallelogram on top. The word "KANGZHAN" is written in a sans-serif font below the square. #### 7.7 Summary Based on the non-clinical and clinical performance as documented in the development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device. # 8. Conclusions: The proposed device has the same intended use and similar technological characteristics as the predicate device. The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. Based on the conducted performance and safety testing, the Infrared Thermometer (Model: K9) is substantially equivalent (SE) to the Infrared Forehead Thermometer (Model IR-FT) cleared under K191668.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%