K Number
K203458

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-12-11

(17 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the following sterilizers and cycles: STERRAD 100® Sterilization System, STERRAD 100S® Sterilization System, STERRAD NX® Sterilization System (Standard and Advanced cycles), STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles), STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles), V-PRO® 1 Low Temperature Sterilization System (Lumen cycle), V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles), V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles), V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles), V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles). The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

Device Description

The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 consists of a noncellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.

AI/ML Overview

This document describes the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248, a Class II medical device, and its acceptance criteria for showing substantial equivalence to a predicate device (K192937). The device is a chemical indicator that changes color (blue to pink) upon exposure to vaporized hydrogen peroxide, verifying exposure to sterilization conditions.

The context of this document is a 510(k) premarket notification for a medical device that is not an AI/ML powered device. The questions provided in the prompt are designed for AI/ML powered devices. Therefore, a direct answer to each question as it pertains to an AI/ML study is not possible. However, the document does contain information about how the device's performance was evaluated and accepted.

Here's an interpretation based on the provided document and the nature of the device:

Device Acceptance Criteria and Performance:

The device's acceptance is based on demonstrating substantial equivalence to a legally marketed predicate device (K192937). Since no changes were made to the device's fundamental design, materials, or performance specifications, the primary acceptance criteria revolve around its functionality as a chemical indicator in newly claimed sterilizers and cycles.

1. A table of acceptance criteria and the reported device performance:

The acceptance criteria for a chemical indicator like this are typically based on its ability to accurately change color under specified sterilization conditions (Pass) and not change under non-sterilization conditions (Fail), as well as maintaining the changed color for a defined period.

Acceptance Criteria CategorySpecific CriteriaReported Device Performance (3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248)
Functionality in Healthcare Facility CyclesColor change from blue to pink when exposed to vaporized hydrogen peroxide in specified sterilizers and cycles.Pass (for all newly claimed V-PRO® s2 cycles)
Minimum Exposure ParametersIndicator must show color change at or above the minimum effective exposure parameters.Pass (for all newly claimed V-PRO® s2 cycles)
End Point Color StabilityThe changed color (pink) must remain stable for a specified duration (e.g., at least one month).Pass (Identical to predicate, which has 1 month stability)
Shelf LifeDevice maintains its performance for its stated shelf life.Two (2) years (Identical to predicate)
BiocompatibilityExposure to healthcare professionals is minimal and below toxic thresholds.Identical (to predicate, which deemed acceptable)

2. Sample size used for the test set and the data provenance:

The document states that "nonclinical testing was performed in accordance with the FDA Guidance for Industry and Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003." This guidance would specify the number of indicators to be tested for each condition. While the exact sample size isn't explicitly stated in this summary, the "Pass" results imply that the required sample sizes per the FDA guidance were met.

  • Sample Size: Not explicitly stated, but implied to be sufficient as per FDA guidance for chemical indicators.
  • Data Provenance: The testing was "nonclinical testing," meaning it was conducted in a laboratory setting, likely at 3M's facilities. The location (country) is not specified, but 3M is a US-based company, so it's most likely in the US. This would be considered prospective testing, as it was specifically conducted to validate the device for the new indications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

For a chemical indicator, the "ground truth" is a direct, observable, and objective chemical reaction (color change) based on exposure to a known concentration of hydrogen peroxide for a defined time/temperature. It does not typically involve human expert interpretation in the same way an AI diagnostic image analysis would. The ground truth is established by the controlled sterilization cycles themselves. The "experts" would be the laboratory personnel or engineers who set up and monitor the sterilization cycles and visually inspect the chemical indicators according to defined pass/fail criteria. Their qualifications would be in laboratory practices, quality control, and understanding of sterilization processes.

4. Adjudication method for the test set:

Not applicable in the context of a chemical indicator. The determination of "pass" or "fail" is based on the visible color change according to a reference, not on human interpretation or consensus of complex data.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI/ML device involving human readers or interpretation of complex data.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI/ML algorithm. The device's "performance" is its physical color change, which is then visually interpreted by a human user (e.g., healthcare professional) to verify sterilization.

7. The type of ground truth used:

The ground truth used for this device is based on controlled exposure to specific sterilization parameters. In a laboratory setting, the chemical indicators are subjected to precisely defined concentrations of vaporized hydrogen peroxide under specific temperature and time conditions that are known to achieve sterilization, and also to conditions that are known not to achieve sterilization (e.g., minimum exposure parameters). The "pass" or "fail" is determined by whether the indicator exhibits the expected color change given the known exposure conditions.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font. The logo is simple and professional, and it is easily recognizable.

December 11, 2020

3M Company Yumi Wackerfuss Senior Regulatory Affairs Associate 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144

Re: K203458

Trade/Device Name: 3M Comply Hydrogen Peroxide Chemical Indicator 1248 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: November 20, 2020 Received: November 24, 2020

Dear Yumi Wackerfuss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203458

Device Name

3M Comply Hydrogen Peroxide Chemical Indicator 1248

Indications for Use (Describe)

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hyrdogen peroxide in the following sterilizers and cycles.

STERRAD 100® Sterilization System

STERRAD 100S® Sterilization System

STERRAD NX® Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles)

STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles)

V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)

V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)

V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles)

V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles)

V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles)

V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles)

The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the 3M logo, which is a well-known corporate symbol. The logo consists of the number "3" followed by the letter "M", both rendered in a bold, sans-serif typeface. The color of the text is a vibrant red, which stands out against the white background. The overall design is simple, clean, and easily recognizable.

K203458 510(k) Summary for 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248

Sponsor Information:

3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000

Contact: Yumi Wackerfuss Senior Regulatory Affairs Associate Phone Number: (651) 736-5286 Fax Number: (651) 737-5320

Date of Summary: November 20, 2020

{4}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248

1. Device Name and Classification:

Common Name:Chemical Indicators
Proprietary Name:3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248
Classification Name:Physical/chemical sterilization process indicators
Device Classification:Class II, 21 CFR 880.2800(b)
Product Code:JOJ

2. Predicate Device:

K192937, 3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248

3. Description of Device:

The 3MTM Comply™ Hydrogen Peroxide Chemical Indicator 1248 consists of a noncellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.

4. Indications for Use

Proposed indication for use for this submission is:

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the following sterilizers and cycles:

STERRAD 100® Sterilization System

STERRAD 100S® Sterilization System

STERRAD NX® Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles)

STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles)

V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)

V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)

V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles)

V-PRO® 60 Low Temperature Sterilization System (Lumen, and Flexible cycles)

V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles)

V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles)

The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

{5}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248

5. Technological Characteristic Comparison

The 3M™ Comply ™ Hydrogen Peroxide Chemical Indicator 1248 is the same design as the previously cleared device of same model number (predicate device: K192937). No change has been made to the device materials, performance specifications, or fundamental technology.

ItemsPredicate Device (K192937)Proposed device (this submission)
3MTM Comply™ HydrogenPeroxide Chemical Indicator 12483MTM Comply™ Hydrogen PeroxideChemical Indicator 1248
Indications foruseUse the 3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248 asan internal pack process indicator toverify exposure to vaporizedhydrogen peroxide in theSTERRAD® 100, STERRAD®100S, STERRAD® NX (Standardand Advanced cycles), STERRAD®100NX (Standard, Flex, Express andDuo cycles), STERRAD® NX withALLClear™ Technology (Standardand Advanced cycles), STERRAD®100NX with ALLClear™Technology (Standard, Flex, Expressand Duo cycles), AMSCO® V-PRO® 1 (Lumen cycle), AMSCO®V-PRO® 1 Plus (Lumen and NonLumen cycles), AMSCO® V-PRO®maX Low Temperature SterilizationSystem (Lumen, Non Lumen, andFlexible cycles), AMSCO® V-PRO®60 Low Temperature SterilizationSystem (Lumen, Non Lumen andFlexible cycles) and AMSCO® V-PRO™ maX 2 Low TemperatureSterilization System (Lumen, NonLumen, Flexible, and Fast NonLumen cycles) sterilizers. Thechemical indicator bar turns fromblue toward pink after exposure tovaporized hydrogen peroxide.Use the 3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248 asan internal pack process indicator toverify exposure to vaporized hydrogenperoxide in the following sterilizersand cycles:STERRAD® 100 SterilizationSystemSTERRAD® 100SSterilizationSystemSTERRAD® NX SterilizationSystem (Standard and Advancedcycles)STERRAD® 100NX SterilizationSystem (Standard, Flex, Express,and Duo cycles)STERRAD® NX with ALLClear®Technology Sterilization System(Standard and Advanced cycles)STERRAD® 100NX withALLClear® Technology SterilizationSystem (Standard, Flex, Express,and Duo cycles)V-PRO® 1 Low TemperatureSterilization System (Lumen cycle)V-PRO® 1 Plus Low TemperatureSterilization System (Lumen andNon Lumen cycles)V-PRO® maX Low TemperatureSterilization System (Lumen, NonLumen, and Flexible cycles)V-PRO® 60 Low TemperatureSterilization System (Lumen, NonLumen, and Flexible cycles)V-PRO® maX 2 Low TemperatureSterilization System (Lumen, NonLumen, Flexible, and Fast NonLumen cycles)V-PRO® s2 Low TemperatureSterilization System (Lumen, NonLumen, Flexible, and Fast cycles)
ItemsPredicate Device (K192937)Proposed device (this submission)
3M™ Comply™ Hydrogen Peroxide Chemical Indicator 12483M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248
The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.
Sterilizers and Sterilization Cycles(1) STERRAD® 100(1) STERRAD® 100
(2) STERRAD® 100S(2) STERRAD® 100S
(3) STERRAD® NX (Standard and Advanced cycles)(3) STERRAD® NX (Standard and Advanced cycles)
(4) STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles)(4) STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles)
(5) STERRAD® 100NX (Standard, Flex, Express, and Duo cycles)(5) STERRAD® 100NX (Standard, Flex, Express, and Duo cycles)
(6) STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express, and Duo cycles)(6) STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express, and Duo cycles)
(7) AMSCO® V-PRO™ 1 (Lumen cycle)(7) V-PRO™ 1 (Lumen cycle)
(8) AMSCO® V-PRO™ 1 Plus (Lumen and NonLumen cycles)(8) V-PRO™ 1 Plus (Lumen and NonLumen cycles)
(9) AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles)(9) V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles)
(10) AMSCO® V-PRO™ 60 (Lumen, Non Lumen and Flexible cycles)(10) V-PRO™ 60 (Lumen, Non Lumen and Flexible cycles)
(11) AMSCO® V-PRO™ maX 2 Low Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast Non-Lumen cycles)(11) V-PRO™ maX 2 Low Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast Non-Lumen cycles)
(12) V-PRO™ s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast cycles)
SubstratePolyethyleneIdentical
BiocompatibilityThe exposure to health care professionals is minimal and well below any identified toxic thresholds for the compounds.Identical
Color ChangeBlue toward pinkIdentical
DetectionHydrogen PeroxideIdentical
Stability of the endpoint reactionAt least one month (4 weeks)Identical
Shelf lifeTwo (2) yearsIdentical

Table 6.1: Technological Characteristic Comparison Table

{6}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248

{7}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248

6. Nonclinical Comparison to the Predicate Device

The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is identical to the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 (K192937). As no change has been made to the device materials, performance specifications, or fundamental technology, the biocompatibility and nonclinical testing provided in K192937 was referenced in this submission to support performance of the device in the claimed sterilizers.

To demonstrate performance in the newly claimed sterilizers and cycles for the V-PRO® s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast cycles), nonclinical testing was performed in accordance with the FDA Guidance for Industry and Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003. Reference Table 6.2 for testing completed in V-PRO® s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast cycles) sterilizer and the tests were resulted passed.

Table 6.2 Summarv of Nonclinical Testing

Test Method/NameResult
Color Change in Health Care Facility CyclePass
Minimum Exposure Parameters to Affect the Change of the Indicator in Health CareFacility CyclePass
End Point Color StabilityPass

7. Conclusion

Based on the non-clinical testing performance data, the 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248 is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the 3MTM Comply™ Hydrogen Peroxide Chemical Indicator 1248 cleared under K192937, Class II (21 CFR 880.2800), product code JOJ.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).