K Number
K192937

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-11-16

(29 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles), STERRAD® 100NX (Standard, Flex , Express and Duo cycles), STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX with ALLClear™ Technology (Standard, Flex , Express and Duo cycles), AMSCO® V PRO® 1 (Lumen cycle), AMSCO® V PRO® 1 Plus (Lumen and Non Lumen cycles), AMSCO® V PRO® maX Low Temperature Sterilization System (Lumen , Non Lumen, and Flexible cycles), AMSCO® V-PRO® 60 Low Temperature Sterilization System (Lumen , Flexible cycles), and AMSCO® V-PROTM maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles) sterilizers. The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

Device Description

The 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248 consists of a noncellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.

AI/ML Overview

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Test Method / NameAcceptance CriteriaReported Device Performance (Result)
Color Change in Health Care Facility CycleColor change from blue toward pinkPass (Color changed from blue toward pink)
Minimum Exposure ParametersDetermination of the minimum time for color to change from blue toward pinkPass (Minimum time for color change was determined)
End Point Color StabilityNo significant color change after exposure to fluorescent light for a minimum of 1 month (4 weeks).Pass (No significant color change observed after 1 month)

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample size for the test set or the data provenance (country of origin, retrospective/prospective). It generally refers to "nonclinical testing" performed in specific sterilizers and cycles.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The device's performance (color change) is an objective, physical characteristic, not requiring expert interpretation for ground truth.

4. Adjudication method for the test set

This information is not applicable as the device's performance is determined by an objective color change, not subjective evaluation requiring adjudication among experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a chemical indicator, not an AI-powered diagnostic tool, and therefore no human reader studies with or without AI assistance were conducted or would be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical chemical indicator, not an algorithm. Its performance is inherent to its material properties and chemical reaction.

7. The type of ground truth used

The ground truth for the performance of the chemical indicator is the objective chemical reaction leading to a color change from blue toward pink upon exposure to vaporized hydrogen peroxide. For the "End Point Color Stability" test, the ground truth is the absence of significant color change after a specified period. These are physical/chemical observations rather than expert consensus, pathology, or outcomes data.

8. The sample size for the training set

This is not applicable. The device is a physical chemical indicator, not a machine learning model, so there is no training set in the context of AI.

9. How the ground truth for the training set was established

This is not applicable, as there is no training set for a chemical indicator. The performance characteristics are based on established chemical principles and empirically determined through laboratory testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 16, 2019

3M Company Mary Fretland Senior Regulatory Affairs Associate 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144

Re: K192937

Trade/Device Name: 3M Comply Hydrogen Peroxide Chemical Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: October 17, 2019 Received: October 18, 2019

Dear Mary Fretland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192937

Device Name

3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248

Indications for Use (Describe)

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 1 00S, STERRAD® NX (Standard and Advanced cycles), STERRAD® 100NX (Standard, Flex , Express and Duo cycles), STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX with ALLClear™ Technology (Standard, Flex , Express and Duo cycles), AMSCO® V PRO® 1 (Lumen cycle), AMSCO® V PRO® 1 Plus (Lumen and Non Lumen cycles), AMSCO® V PRO® maX Low Temperature Sterilization System (Lumen , Non Lumen, and Flexible cycles), AMSCO® V-PRO® 60 Low Temperature Sterilization System (Lumen , Flexible cycles), and AMSCO® V-PROTM maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles) sterilizers. The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The "3" has a horizontal line cut out of the top right corner, creating a unique design element. The logo is simple and recognizable.

510(k) Summary for 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 K192937

Sponsor Information:

3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000

Contact: Mary Fretland Senior Regulatory Affairs Associate Phone Number: (651) 737-2296 Fax Number: (651) 737-5320

Date of Summary: November 04, 2019

{4}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248

1. Device Name and Classification:

Common or Usual Name:Chemical Indicator
Proprietary Name:3M™ Comply™ Hydrogen Peroxide Chemical Indicator1248
Classification Name:Indicator, physical/chemical sterilization process
Device Classification:Class II, 21 CFR § 880.2800
Product Code:JOJ

2. Predicate Device:

K183211 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248

3. Description of Device:

The 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248 consists of a noncellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.

4. Indications for Use

Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100, STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles). STERRAD® 100NX (Standard, Flex, Express and Duo cycles), STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express and Duo cycles), AMSCO® V-PRO® 1 (Lumen cycle), AMSCO® V-PRO® 1 Plus (Lumen and Non Lumen cycles), AMSCO® V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles), AMSCO® V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible cycles) and AMSCO® V-PRO™ maX 2 Low Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast Non-Lumen cycles) sterilizers. The chemical indicator bar turns from blue toward pink after exposure to vaporized hydrogen peroxide.

{5}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248

5. Technological Characteristic Comparison Table
FeatureSubmission Device (K192937):3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248Predicate Device (K183211):3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248Comparison
Indications foruseUse the 3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248 asan internal pack process indicator toverify exposure to vaporizedhydrogen peroxide in theSTERRAD® 100, STERRAD®100S, STERRAD® NX (Standardand Advanced cycles), STERRAD®100NX (Standard, Flex, Express andDuo cycles), STERRAD® NX withALLClear™ Technology (Standardand Advanced cycles), STERRAD®100NX with ALLClear™Technology (Standard, Flex, Expressand Duo cycles), AMSCO® V-PRO® 1 (Lumen cycle), AMSCO®V-PRO® 1 Plus (Lumen and NonLumen cycles), AMSCO® V-PRO®maX Low Temperature SterilizationSystem (Lumen, Non Lumen, andFlexible cycles), AMSCO® V-PRO® 60 Low TemperatureSterilization System (Lumen, NonLumen and Flexible cycles) andAMSCO® V-PRO™ maX 2 LowTemperature Sterilization System(Lumen, Non-Lumen, Flexible, andFast Non-Lumen cycles) sterilizers.The chemical indicator bar turnsfrom blue toward pink after exposureto vaporized hydrogen peroxide.Use the 3M™ Comply™ HydrogenPeroxide Chemical Indicator 1248 asan internal pack process indicator toverify exposure to vaporizedhydrogen peroxide in theSTERRAD® 100, STERRAD®100S, STERRAD® NX (Standardand Advanced cycles), STERRAD®100NX (Standard, Flex, Express andDuo cycles), STERRAD® NX withALLClear™ Technology (Standardand Advanced cycles), STERRAD®100NX with ALLClear™Technology (Standard, Flex, Expressand Duo cycles), AMSCO® V-PRO® 1 (Lumen cycle), AMSCO®V-PRO® 1 Plus (Lumen and NonLumen cycles), AMSCO® V-PRO®maX Low Temperature SterilizationSystem (Lumen, Non Lumen, andFlexible cycles) and AMSCO® V-PRO® 60 Low TemperatureSterilization System (Lumen, NonLumen and Flexible cycles)sterilizers. The chemical indicator barturns from blue toward pink afterexposure to vaporized hydrogenperoxide.Similar
Sterilizers andSterilizationCyclesSTERRAD® 100STERRAD® 100SSTERRAD® NX (Standard andAdvanced cycles)STERRAD® NX with ALLClear™Technology (Standard and Advancedcycles)STERRAD® 100NX (Standard, Flex,Express, and Duo cycles)STERRAD® 100NX withALLClear™ Technology (Standard,Flex, Express, and Duo cycles)AMSCO® V-PRO™ 1 (Lumencycle)AMSCO® V-PRO™ 1 Plus (Lumenand Non Lumen cycles)AMSCO® V-PRO™ maX LowTemperature Sterilization System(Lumen, Non Lumen and Flexiblecycles)STERRAD® 100STERRAD® 100SSTERRAD® NX (Standard andAdvanced cycles)STERRAD® NX with ALLClear™Technology (Standard and Advancedcycles)STERRAD® 100NX (Standard, Flex,Express, and Duo cycles)STERRAD® 100NX withALLClear™ Technology (Standard,Flex, Express, and Duo cycles)AMSCO® V-PRO™ 1 (Lumencycle)AMSCO® V-PRO™ 1 Plus (Lumenand Non Lumen cycles)AMSCO® V-PRO™ maX LowTemperature Sterilization System(Lumen, Non Lumen and Flexiblecycles)Similar
Submission Device (K192937):Predicate Device (K183211):Comparison
Feature3M™ Comply™ HydrogenPeroxide Chemical Indicator 12483M™ Comply™ HydrogenPeroxide Chemical Indicator 1248
AMSCO® V-PRO™ 60 (Lumen,Non Lumen and Flexible cycles)AMSCO® V-PRO™ maX 2 LowTemperature Sterilization System(Lumen, Non-Lumen, Flexible, andFast Non-Lumen cycles)AMSCO® V-PRO™ 60 (Lumen,Non Lumen and Flexible cycles)
SubstratePolyethylenePolyethyleneIdentical
BiocompatibilityThe exposure to health careprofessionals is minimal and wellbelow any identified toxic thresholdsfor the compounds.The exposure to health careprofessionals is minimal and wellbelow any identified toxic thresholdsfor the compounds.Identical
Color ChangeBlue toward pinkBlue toward pinkIdentical
DetectionHydrogen PeroxideHydrogen PeroxideIdentical
Stability of theendpointreactionAt least one month (4 weeks)At least one month (4 weeks)Identical
Shelf lifeTwo (2) yearsTwo (2) yearsIdentical

5. Technological Characteristic Comparison Table

{6}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248

The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is identical to the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 (K18321). As no change has been made to the device materials, performance specifications, or fundamental technology, the biocompatibility and nonclinical testing provided in K183211 was referenced in this submission to support performance of the device in the claimed sterilizers.

6. Summary of Nonclinical Testing

To demonstrate performance in the newly claimed sterilizers and cycles, nonclinical testing was performed in accordance with the FDA Guidance for Industry and Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003. Reference Table 6.1 for testing completed in AMSCO® V-PRO™ maX 2 Low Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast Non-Lumen cycles) sterilizer.

Test Method / NamePurposeAcceptance CriteriaResult
Color Change inHealth CareFacility CycleTo demonstrate the color change ofthe device when used in theAMSCO® V-PRO™ maX 2 LowTemperature Sterilization SystemColor change from blue towardpinkPass
MinimumExposureParametersTo determine the minimum timerequired for the color change of thedevice when used in the AMSCO®Determination of the minimumtime for color to change fromblue toward pinkPass

Table 6.1 Summary of Nonclinical Testing

{7}------------------------------------------------

PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248

Test Method /NamePurposeAcceptance CriteriaResult
End Point ColorStabilityV-PROTM maX 2 LowTemperature Sterilization SystemNo significant color change afterexposure to fluorescent light for aminimum of 1 month (4 weeks).Pass
To demonstrate the post-sterilization color stability of thedevice after use in the AMSCO®V-PROTM maX 2 LowTemperature Sterilization System

7. Conclusion

Based on the non-clinical testing performance data, the 3MTM Comply™ Hydrogen Peroxide Chemical Indicator 1248 is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the 3MTM Comply™ Hydrogen Peroxide Chemical Indicator 1248 cleared under K183211, Class II (21 CFR 880.2800), product code JOJ.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).