K Number
K203316

Validate with FDA (Live)

Date Cleared
2021-10-27

(351 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The V.A.C.® Ulta Negative Pressure Wound Therapy System is an integrated wound management system that provides Negative Pressure Wound Therapy (V.A.C.® Therapy) with an instillation option ( Veraflo™ Therapy).

· V.A.C.® Therapy in the absence of instillation is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material.

· Veraflo™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

The V.A.C.® Ulta Negative Pressure Wound Therapy System with and without is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts.

· V.A.C.® Therapy in the absence of instillation may also be used for:

o The temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The abdominal dressing will most often be applied in the operating theater.

o The management of the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Device Description

The V.A.C. 8 Ulta Negative Pressure Wound Therapy System is a negative pressure wound therapy system with an instillation feature which allows controlled delivery and drainage of topical wound treatment solutions and suspensions. The unit is comprised of a vacuum pump and an instillation pump. The vacuum pump delivers negative therapy for the removal of wound exudate and instilled solutions when applied. The instillation pump provides controlled delivery of topical wound solutions and suspensions. Both pumps are software controlled. Instillations and negative pressure are delivered through tubing to foam dressings in the wound covered by an occlusive drape. Software monitors both negative pressure wound therapy as well as positive pressure during instillation of fluids to the wound bed. Software also provides controls for setting therapy parameters as well as help and alarm functions.

AI/ML Overview

The document describes the V.A.C.® Ulta Negative Pressure Wound Therapy System, which is a negative pressure wound therapy system with an instillation feature. The submission for this device is a 510(k) premarket notification.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present acceptance criteria in a tabular format with corresponding device performance values for all features. However, it states the performance specification for the new Smart Instill™ Feature and the general performance testing conclusion.

Feature/ParameterAcceptance CriteriaReported Device Performance
Veraflo™ Therapy Cycle
Smart Instill™ FeatureDressing saturation must be achieved for the instilled volume within 12 hours, without creating an overfill condition.Performance testing has documented that the Smart Instill™ Feature meets its design specification fill volumes that will saturate the applied dressing within 12 hours of initiation without underfilling or overfilling the dressed wound.
All Therapy CyclesVerification data confirms that the specified negative pressure and instillation performance specifications have been met.Performance testing has documented that the Smart Instill™ Feature meets its design specification fill volumes that will saturate the applied dressing within 12 hours of initiation without underfilling or overfilling the dressed wound.
Software Level of ConcernModerateModerate
CybersecurityRisk of patient harm from device exploitability at a controlled and acceptable level. Device is trustworthy, resilient to cybersecurity threats that can lead to patient harm, suited to perform intended functions, and provides a reasonable level of availability, reliability, and correct operation.Cybersecurity assessment indicates the risk is at a controlled level and acceptable. The device is considered sufficiently trustworthy, reasonably resilient, suited for intended functions, and provides a reasonable level of availability, reliability, and correct operation.
New Features Usability/SafetyThe new features could be safely and effectively used by all test subjects.Human factors engineering assessment with 30 subject nurses and doctors indicated that the new features could be safely and effectively used by all test subjects.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide details on the sample size or provenance for the "performance testing" that documented the Smart Instill™ Feature's compliance. This testing is described as non-clinical.

For the human factors engineering assessment:

  • Sample Size for Test Set: 30 subject nurses and doctors.
  • Data Provenance: Not specified, but generally, human factors studies for medical devices are prospective and conducted in a simulated or real-world use environment. The document does not specify the country of origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • The document does not mention the use of experts to establish "ground truth" in the conventional sense for the performance testing or the human factors study.
  • For the human factors study, 30 "subject nurses and doctors" were used as participants to assess usability and safety, but they were not establishing ground truth, rather their interactions with the device were the data. Their specific qualifications (e.g., years of experience) are not detailed beyond "nurses and doctors."

4. Adjudication Method

  • Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or studies where expert consensus is needed to establish ground truth from ambiguous cases. This type of adjudication is not described or indicated for either the non-clinical performance testing or the human factors assessment mentioned in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study is mentioned. The submission is for a medical device (Negative Pressure Wound Therapy System), not an AI-assisted diagnostic tool, so this type of study would not typically apply. The document focuses on performance specifications and human factors for the device itself.

6. Standalone Performance (Algorithm Only)

  • The Smart Instill™ Feature involves software control, and its performance was "documented." This implies a standalone evaluation of the algorithm's ability to achieve saturation within the specified time frame. However, the exact methodology for this standalone testing is not detailed (e.g., if it was tested solely in simulation or on benchtop models without human interaction to verify the algorithm's output). The "performance testing" mentioned effectively serves as a standalone evaluation of the Smart Instill™ algorithm's functionality within the device.

7. Type of Ground Truth Used

  • For the Smart Instill™ Feature: The "ground truth" would be objective measurements of fill volume and dressing saturation achieved within the 12-hour timeframe. This would likely be established through physical measurements on test setups (e.g., measuring liquid volume and observing saturation on wound models).
  • For the human factors engineering assessment: The "ground truth" is established by observing user interactions and collecting feedback to determine if tasks can be performed safely and effectively. This is based on direct observation and qualitative/quantitative data from the participants.

8. Sample Size for the Training Set

  • The document does not mention a "training set" as this is not an AI/machine learning model that requires a distinct training phase. It describes new software features, which are typically developed based on design specifications and then verified/validated.

9. How the Ground Truth for the Training Set Was Established

  • Since there is no mention of a training set in the context of an AI/ML model, this question is not applicable. The device's software features are developed based on established engineering principles and clinical requirements, rather than being "trained" on a dataset with ground truth labels.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

October 27, 2021

KCI, a part of 3M Health Care Business Group Margaret Marsh Regulatory Affairs Advanced Specialist 6203 Farinon Drive San Antonio, Texas 78249

Re: K203316

Trade/Device Name: V.A.C. Ulta Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 23, 2021 Received: September 27, 2021

Dear Margaret Marsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203316

Device Name

V.A.C.® Ulta Negative Pressure Wound Therapy System

Indications for Use (Describe)

The V.A.C.® Ulta Negative Pressure Wound Therapy System is an integrated wound management system that provides Negative Pressure Wound Therapy (V.A.C.® Therapy) with an instillation option ( Veraflo™ Therapy).

· V.A.C.® Therapy in the absence of instillation is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material.

· Veraflo™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

The V.A.C.® Ulta Negative Pressure Wound Therapy System with and without is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts.

· V.A.C.® Therapy in the absence of instillation may also be used for:

o The temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The abdominal dressing will most often be applied in the operating theater.

o The management of the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date preparedOctober 27, 2021
Submitter information [21 CFR 807.929(a)(1)]
NameKCI, now a part of 3M Health Care Business Group
Address6203 Farinon Drive; San Antonio, Texas 78249
Fax number210 255-6727
EstablishmentRegistrationNumber3009897021
Name of contactpersonMargaret Marsh, Regulatory Affairs Advanced Specialist
Name of the device [21 CFR 807.92(a)(2)]
Trade orproprietary nameV.A.C.® Ulta Negative Pressure Wound Therapy System
Common or usualnameNegative pressure wound therapy system with an instillation option
ClassificationnameNegative Pressure Wound Therapy Powered Suction Pump(and components)
Legally marketed device(s) to which equivalence is claimed[21 CFR 807.92(a)(3)]

{4}------------------------------------------------

Device description [21 CFR 807.92(a)(4)]

The V.A.C. 8 Ulta Negative Pressure Wound Therapy System is a negative pressure wound therapy system with an instillation feature which allows controlled delivery and drainage of topical wound treatment solutions and suspensions. The unit is comprised of a vacuum pump and an instillation pump. The vacuum pump delivers negative therapy for the removal of wound exudate and instilled solutions when applied. The instillation pump provides controlled delivery of topical wound solutions and suspensions. Both pumps are software controlled. Instillations and negative pressure are delivered through tubing to foam dressings in the wound covered by an occlusive drape. Software monitors both negative pressure wound therapy as well as positive pressure during instillation of fluids to the wound bed. Software also provides controls for setting therapy parameters as well as help and alarm functions.

Indications for Use [21 CFR 807.92(a)(5)]

The V.A.C. 8 Ulta Negative Pressure Wound Therapy System is an integrated wound management system that provides Negative Pressure Wound Therapy (V.A.C.® Therapy) with an instillation option (Veraflo™ Therapy).

  • V.A.C.® Therapy in the absence of instillation is intended to create an environment that promotes wound healing by secondary or . tertiary (delayed primary) intention by preparing the wound bed for closure, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material.
  • . Veraflo™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

The V.A.C. 8 Uta Negative Pressure Wound Therapy System with and without is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts.

  • V.A.C.® Therapy in the absence of instillation may also be used for: .
    • The temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are ೧ necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The abdominal dressing will most often be applied in the operating theater.
    • The management of the environment of surgical incisions that continue to drain following sutured or stapled closure by O maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

{5}------------------------------------------------

Comparison To Predicate Table [21 CFR 807.92(a)(6)]
ComparatorSubject DevicePredicate Device
Trade nameV.A.C.® Ulta Negative Pressure Wound TherapySystemV.A.C.® Ulta Negative Pressure WoundTherapy System
Therapy UnitModel numberSame as predicateULTDEV 01/US
510(k)Submitter/HolderSame as predicate; however, KCI is now apart of 3MHealth Care BusinessKCI USA, Inc.
510(k) numberK203316K162790
Indications for UseSame as predicate:The V.A.C.® Ulta Negative Pressure Wound TherapySystem is an integrated wound management system thatprovides Negative Pressure Wound Therapy (V.A.C.®Therapy) with an instillation option (Veraflo™ Therapy).V.A.C.® Therapy in the absence of instillation isintended to create an environment that promoteswound healing by secondary or tertiary (delayedprimary) intention by preparing the wound bed forclosure, reducing edema, promoting granulationtissue formation and perfusion, and by removingexudate and infectious material.Veraflo™ Therapy is indicated for patients whowould benefit from vacuum assisted drainage andcontrolled delivery of topical wound treatmentsolutions and suspensions over the wound bedThe V.A.C.® Ulta Negative Pressure Wound TherapySystem is an integrated wound management system thatprovides Negative Pressure Wound Therapy (V.A.C.®Therapy) with an instillation option (Veraflo™ Therapy).V.A.C.® Therapy in the absence of instillation isintended to create an environment that promoteswound healing by secondary or tertiary (delayedprimary) intention by preparing the wound bed forclosure, reducing edema, promoting granulationtissue formation and perfusion, and by removingexudate and infectious material.Veraflo™ Therapy is indicated for patients whowould benefit from vacuum assisted drainage andcontrolled delivery of topical wound treatmentsolutions and suspensions over the wound bed
ComparatorSubject DevicePredicate Device
Indications for Use,continuedThe V.A.C.® Ulta Negative Pressure Wound TherapySystem with and without instillation is indicated forpatients with chronic, acute, traumatic, subacute anddehisced wounds, partial- thickness burns, ulcers (such asdiabetic, pressure and venous insufficiency), flaps andgrafts.V.A.C.® Therapy in the absence of instillation may also beused for:The temporary bridging of abdominal wall openingswhere primary closure is not possible and/or repeatabdominal entries are necessary and for openabdominal wounds with exposed viscera including,but not limited to, abdominal compartmentsyndrome. The intended care setting is a closelymonitored area within the acute care hospital, suchas the ICU. The abdominal dressing will most oftenbe applied in the operating theater.The V.A.C.® Ulta Negative Pressure Wound TherapySystem with and without instillation is indicated forpatients with chronic, acute, traumatic, subacute anddehisced wounds, partial- thickness burns, ulcers (such asdiabetic, pressure and venous insufficiency), flaps andgrafts.V.A.C.® Therapy in the absence of instillation may also beused for:The temporary bridging of abdominal wall openingswhere primary closure is not possible and/or repeatabdominal entries are necessary and for openabdominal wounds with exposed viscera including,but not limited to, abdominal compartmentsyndrome. The intended care setting is a closelymonitored area within the acute care hospital, suchas the ICU. The abdominal dressing will most oftenbe applied in the operating theater.
The management of the environment of surgicalincisions that continue to drain following sutured orstapled closure by maintaining a closed environmentand removing exudate via the application of negativepressure wound therapy.The management of the environment of surgicalincisions that continue to drain following sutured orstapled closure by maintaining a closed environmentand removing exudate via the application of negativepressure wound therapy.
Components of the indication
Care settingSame as predicateThe V.A.C.® Ulta Negative Pressure Wound Therapy Unit is for acute caresettings only; it is not intended for use in home care.
Duration oftherapySame as predicateDuration of therapy is determined by the instructions for use for the woundcare dressing applied.• For V.A.C.® Granufoam, V.A.C. Whitefoam and Veraflo™ Dressings,the typical duration is 3 days between dressing changes.• The AbThera™ Open Abdomen Dressing has a 3- day use life.• The Prevena™ Incision Management Dressings may be used for up to 7
ComparatorSubject DevicePredicate Device
Wound typesSame as predicate.V.A.C.® and Veraflo™ Therapies are intended for use with chronic, acute, traumatic, subacute and dehisced wounds, partial- thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts. The AbThera™ Open Abdomen Dressing is intended for use in abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The Prevena™ Incision Management Dressings are intended to be used on closed surgical incisions
Wound sizesSame as predicate, except for the addition of aLarge size of the Veraflo Cleanse ChoiceDressing.Dressings are provided in various sizes and shapes for wounds/incisions of various sizes and configurations.
Mechanisms ofactionSame as predicateNPWT (applies to all indicated wound types): Helps hold the wound edges together Removes wound exudates and infectious materials. Reduces edema V.A.C.® and Veraflo™ Therapies also promote granulation tissue formation and perfusion Veraflo™ Therapy also cleanses the wound with appropriate topical wound solutions
TechnologySame as predicateSoftware controlled application of negative pressure therapy via a vacuum pump and instillation therapy via an instillation pump.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

K203316

V.A.C.® Ulta Negative Pressure Wound Therapy System
-----------------------------------------------------------------
ComparatorSubject DevicePredicate Device
System designSame as predicateTherapy system consists of:Software controlled therapy unit 500 and 1000 mL exudate canisters Instillation cassette (Veraflo™ Therapy only) Single pad tubing set for delivery of NPWT Single and dual pad tubing sets for delivery of both instillation and negative pressure (Veraflo™ Therapy only)
Same as predicateDressings:V.A.C.® Granufoam, Granufoam Silver, Whitefoam Dressings, Veraflo™, Veraflo Cleanse™, and Veraflo Cleanse Choice™ Dressings Prevena™ Incision Management Dressings AbThera™ Open Abdomen Dressings
OperatingPrincipleSame as predicateTherapy unit delivers software controlled negative pressure to the wound site. The open cells of the foam dressing to which the therapy unit is connected enable distribution of the negative pressure across the surface of the wound, while the tubing transfers accumulated fluids to the canister.Therapy unit also provides automated delivery of instillation fluids into the wound bed via an instillation pump between negative pressure therapy cycles. After a user-selected soak time, fluids are removed into the canister via application of negative pressure in the next negative pressure wound therapy cycle.
The operating principle for the determination of the instill volume provided by the new Smart Instill™ Feature is based on the decay in pressure after application of negative pressure to the dressed wound site.
ComparatorSubject DevicePredicate Device
Performance Specifications
V.A.C. NegativePressure WoundTherapy(for V.A.C.Therapy Cycle)Same as predicateContinuous mode and intermittent (“DPC”) modes Continuous negative pressure range: -25 to -200 mmHg in 25 mmHg increments DPC mode: 1 to 10 minutes to rise up and 1 to 10 minutes to fall down (in 1 minute increments each); target negative pressure range: -50 to-200 mmHg in 25 mmHg increments
Prevena™Therapy CycleSame as predicateContinuous negative pressure at -125 mmHg,
AbThera™Therapy CycleSame as predicateContinuous negative pressure at -100, 125, and 150 mmHg
Veraflo™Therapy CycleSame as predicateSmart Instill™ Feature: Dressing Saturationmust be achieved for the instilled volume within12 hours, without creating an overfill condition.Volume delivered by pump: 6 mL to 500 mL Instill hold time (time in wound bed, called "soak time"): 1 sec to 30 min Instillation solution pressure limit: 3 psi Negative pressure time: 3 min to 12 hrs Negative pressure mode: continuous or DPC Negative pressure range: -50 to -200 mmHg in 25 mmHg increments Not provided
Performance testingSame as predicate.Verification data confirms that the specified negative pressure andinstillation performance specifications have been met.

{9}------------------------------------------------

{10}------------------------------------------------

ComparatorSubject DevicePredicate Device
Therapy Unit
• Dimensions –with emptycanister(inches)Same as predicateWidthLengthDepth
8.5510.257.25
• Weight –with empty500 mLcanisterSame as predicate8 lbs (without cassette)
• UserinterfaceSame as predicateTouch screen
• SoftwareLevel ofConcernSame as predicateModerate

{11}------------------------------------------------

K203316 510(k) SUMMARY

V.A.C.® Ulta Negative Pressure Wound Therapy System

ComparatorSubject DevicePredicate Device
• Software ControlsSame as predicateThe embedded software accepts user instructions for initiating andmodifying therapy via a touch screen.
Same as predicateSoftware controls include the user selection of negative pressure therapyparameters as well as negative pressure therapy application andmonitoring. It also controls user selection of instillation therapyparameters as well as application and monitoring of therapy.
Same as predicate, with the addition of:• Smart Instill™ Feature provides theclinician an additional option to select thesolution volume setpoint (applicable toVeraflo™ Therapy only).• On-Screen Help Animations are nowprovided for the most frequent alarm/alertstates (applicable to V.A.C.® Therapy,Veraflo™ Therapy, Prevena™ Therapy,and AbThera™ Therapy).• Postponement Feature allows for thepostponement of Veraflo™ Therapy for upto 60 hours while still providing V.A.C.®Therapy.• Therapy Inactive Alarm Delay allows fordressing changes to occur without triggeringa Therapy Inactive Alarm (applicable toboth V.A.C.® and Veraflo™ Therapies).The software also provides controls for alarms and a variety of optionalfeatures for the user, such as:• Seal Check Leak Detector• Dressing Soak Tool (Veraflo™ Therapy only)• Fill Assist Tool (Veraflo™ Therapy only)• Test Cycle Tool (Veraflo ™Therapy only)• Wound Imaging Tool for viewing uploaded wound images andcalculating wound area/volume• History Tool for alarm, therapy and patient history logs• Utilities Tool to set system preferences such as language, unitmeasures, dates, etc.• Help Tool to provide on screen help. This tool providesadditional information regarding alarms/alerts and how toresolve them, operation of therapies and cleaning instructions.• Information Tool to allow access to the current therapy settingsand the therapy summary tabs.
ComparatorSubject DevicePredicate Device
Alarms
• System Notifications
Same as predicateBattery Exhausted
• V.A.C. Therapy Alerts/Alarms
Same as predicateSystem Error Alarm (at Power ON)V.A.C.® Therapy Blockage Alarm (Therapy Interrupted)
V.A.C.® Therapy Canister Full Alarm (Therapy Interrupted)
V.A.C.® Therapy Canister Not Engaged Alarm
V.A.C.® Therapy – Therapy Inactive Alarm
V.A.C.® Therapy Leak Alarm
V.A.C.® Therapy Leak Alarm – Therapy Interrupted
V.A.C.® Therapy Low Pressure Alarm (Therapy Interrupted)
V.A.C.® Therapy Battery Low Alert
V.A.C.® Therapy Battery Critical Alarm
V.A.C.® Therapy Internal Temperature Alert
V.A.C.® Therapy System Error Alarm (Therapy Interrupted)
• PREVENA Therapy Alerts
Same as predicatePREVENA™ Therapy Blockage Alert*
PREVENA™ Therapy Blockage Alert (Therapy Interrupted)
PREVENA™ Therapy Canister Full Alert* (Therapy Interrupted)
PREVENA™ Therapy Canister Not Engaged Alert
PREVENA™ Therapy - Therapy Inactive Alert
PREVENA™ Therapy Leak Alert*
PREVENA™ Therapy Battery Low Alert*
PREVENA™ Therapy Battery Critical Alert*
PREVENA™ Therapy Internal Temperature Alert
PREVENA™ Therapy System Error Alert (Therapy Interrupted)
ComparatorSubject DevicePredicate Device
• ABTHERA Therapy AlertsSame as predicateABTHERA™ Therapy Blockage Alert*
ABTHERA™ Therapy Blockage Alert (Therapy Interrupted)
ABTHERA™ Therapy Canister Full Alert*
ABTHERA™ Therapy Canister Not Engaged Alert
ABTHERA™ Therapy - Therapy Inactive Alert
ABTHERA™ Therapy Leak Alert*
ABTHERA™ Therapy Battery Low Alert*
ABTHERA™ Therapy Battery Critical Alert
ABTHERA™ Therapy Internal Temperature Alert
ABTHERA™ Therapy System Error Alert (Therapy Interrupted)
• V.A.C. VERAFLO Therapy Alerts/AlarmsSame as predicate, exceptfor the addition of SmartInstill Inactive AlertVERAFLO™ Therapy Blockage Alert
VERAFLO™ Therapy Blockage Alarm(Therapy Interrupted
VERAFLO™ Therapy Canister Full Alarm(Therapy Interrupted)
VERAFLO™ Therapy Canister Not EngagedAlarm
VERAFLO™ Therapy - Therapy InactiveAlarm
VERAFLO™ Therapy Leak Alarm
VERAFLO™ Therapy Leak Alarm (Therapy Interrupted)
VERAFLO™ Therapy Low Pressure Alarm (Therapy Interrupted)
VERAFLO™ Therapy V.A.C. VERALINKNot Engaged Alert
VERAFLO™ Therapy Solution Bag / BottleEmpty Alert
VERAFLO™ Therapy Fill Assist InactiveAlert
VERAFLO™ Therapy Pressure DeviationAlarm (Therapy Interrupted)
VERAFLO™ Therapy Instill Tube BlockageAlert (Therapy Interrupted)
VERAFLO™ Therapy Battery Low Alert
VERAFLO™ Therapy Battery Critical Alarm
VERAFLO™ Therapy Internal TemperatureAlert
VERAFLO™ Therapy System Error Alarm(Therapy Interrupted)

{12}------------------------------------------------

K203316 510(k) SUMMARY

V.A.C.® Ulta Negative Pressure Wound Therapy System

{13}------------------------------------------------

{14}------------------------------------------------

ComparatorSubject DevicePredicate Device
• Energy sourceSame as predicateAC and battery
• Battery typeSame as predicateCustom lithium battery
• Use lifeSame as predicateIndefinite, with routine servicing and repair
CanistersSame as predicate• 500 mL capacity canister with charcoal (odor control) and membrane filters (fluid barrier)• 1000 mL capacity canister with charcoal (odor control) and membrane filters (fluid barrier)
InstillationCassetteSame as predicateThe instillation cassette is comprised of a solution container holder which connects to the V.A.C.® Ulta instillation pump via tubing on the back of the holder. The cassette has a spike that is used to connect to the solution container as well as tubing that delivers solution to the dressing.
Negative PressureTubingSame as predicateAll therapies have tubing sets constructed from PVC.
Instillation TubingSame as predicatePVC tubing; approximately 9 feet in length from dressing to canister.
Dressing SystemConfigurationSame as predicateV.A.C.®, Veraflo™ and AbThera™ Dressings: All have multiple dressing components designed for use over open wounds. In typical orderof application:• Foam dressing (s)• Polyurethane drape• Pressure sensing pad and tubing• Instillation pad and tubing (Veraflo™Therapy only)Prevena™ Dressings have similar components, except that the foam dressing has a bonded interface fabric that contacts the incision site.

{15}------------------------------------------------

K203316 510(k) SUMMARY

V.A.C.® Ulta Negative Pressure Wound Therapy System

Performance Data [21 CFR 807.92(b)]

Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]

  • Software has been assessed in accordance, Guidance, Guidance for the Content of Premarket Submissions for Software . Contained in Medical Devices Document issued on May 11, 2005.
  • A cybersecurity assessment of the V.A.C.® Uta Negative Pressure Wound Therapy Unit, based on FDA Draft of . Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, October 2018, indicates that the risk of patient harm resulting from device exploitability is at a controlled level and is considered acceptable and that the V.A.C. ® Uta Negative Pressure Wound Therapy Unit is sufficiently trustworthy, reasonably resilient to uncontrolled cybersecurity threats that can lead to patient harm, reasonably suited to perform its intended functions and provides a reasonable level of availability, reliability, and correct operation.
  • Performance testing has documented that the Smart Instill™ Feature meets its design specification fill . volumes that will saturate the applied dressing within 12 hours of initiation without underfilling or overfilling the dressed wound.

Summary of clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(2)]

No clinical tests were necessary. However, a human factors engineering assessment, per FDA Guidance. Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff, issued February 2016 was conducted with 30 subject nurses and doctors. The results indicated that the new features could be safely and by all test subjects.

Conclusions drawn [21 CFR 807.92(b)(3)]

  • There has been no change to the intended use of the V.A.C. ® Uta Negative Pressure Wound Therapy System. .
  • There has been no change to the technology delivering negative pressure and instillation therapy. Performance specifications for . provided therapies remain unchanged.
  • · Human factors engineering assessment documents that the changes to the therapy unit software and its associated labeling to provide ease of use features are safe and effective for their intended use.

The subject device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics. There are no different questions regarding safety or effectiveness.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.