K Number
K203223
Manufacturer
Date Cleared
2021-01-28

(87 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as bronchoscopes including duodenoscopes, and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the endoscope processing cycle (4 minute generation sequence followed by a 6-minute exposure sequence).

Device Description

The Reliance Endoscope Processing System is a high level disinfection system that can wash and high level disinfect up to two manually precleaned, immersible, reusable, heatsensitive, semi-critical devices such as GI flexible endoscopes and related accessories. The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.

AI/ML Overview

The provided text is a 510(k) Summary for the STERIS Reliance Endoscope Processing System. This document details the device, its intended use, and a comparison to a predicate device, along with a summary of non-clinical testing performed to demonstrate substantial equivalence.

Based on the provided information, I will answer the questions regarding acceptance criteria and the study that proves the device meets them.

1. A table of acceptance criteria and the reported device performance

TestAcceptance CriteriaReported Device Performance
High level disinfection efficacy of selected worst-case duodenoscope models, representing multiple manufacturers and distal tip designs.>6 log reduction of Mycobacterium terrae per site under worst case processing conditions.Pass
Biocompatibility of Reliance Dry Germicide made with a proprietary component from an alternate sourceMedical devices shall be non-cytotoxic after exposure to Reliance DG made with a proprietary component from an alternate sourcePass
Stability of Reliance Dry Germicide made with a proprietary component from an alternate sourceReliance Dry Germicide shall meet acceptance criteria after 18 months storage.Pass

2. Sample sized used for the test set and the data provenance

The document does not specify the exact sample sizes for the "selected worst-case duodenoscope models" or the number of medical devices used in biocompatibility testing. It also does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. However, based on the context of a 510(k) submission for a medical device and the type of tests performed (e.g., efficacy, biocompatibility, stability), these would typically be controlled laboratory studies conducted prospectively.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide information on the number of experts or their qualifications for establishing ground truth in these non-clinical tests. These types of tests (disinfection efficacy, biocompatibility, stability) are generally conducted by laboratory technicians and scientists following established protocols (e.g., ISO standards, AOAC methods) and do not typically involve human expert consensus for "ground truth" in the way a diagnostic AI system would. The "ground truth" for these tests is based on the measurable outcomes of the chemical and biological assays.

4. Adjudication method for the test set

Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies involving interpretation by multiple human readers. For the non-clinical laboratory tests described here (efficacy, biocompatibility, stability), an adjudication method is not applicable as the results are quantitative and based on predefined measurement thresholds.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. The device in question is an automated endoscope processing system, which performs automated cleaning and high-level disinfection. It is not an AI-assisted diagnostic tool that aids human readers. Therefore, the concept of human readers improving with or without AI assistance does not apply here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The primary tests reported are for the performance of the automated system itself (efficacy of disinfection, biocompatibility of the germicide, stability of the germicide). In this context, the "standalone" performance is precisely what was evaluated. The device functions automatically without a human actively making decisions based on its output during the disinfection process.

7. The type of ground truth used

For the efficacy test, the ground truth is the microbiological log reduction of Mycobacterium terrae following the disinfection process. For biocompatibility, the ground truth is cellular response (e.g., non-cytotoxicity) based on ISO 10993-5 standards. For stability, the ground truth is the chemical and functional properties of the germicide maintaining within acceptance criteria over time. These are all objective, measurable laboratory outcomes.

8. The sample size for the training set

The document describes non-clinical performance testing for a physical device (an automated endoscope reprocessor and its germicide). It does not involve machine learning or AI models with "training sets." Therefore, the concept of a training set size is not applicable.

9. How the ground truth for the training set was established

As there is no training set involved for this type of device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

January 28, 2021

Steris Marcia Benedict Senior Director, Regulatory Affairs 5960 Heislev Rd Mentor, Ohio 44060

Re: K203223

Trade/Device Name: Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NZA Dated: October 30, 2020 Received: November 2, 2020

Dear Marcia Benedict:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K203223

Device Name

Reliance Endoscope Processing System

Indications for Use (Describe)

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as bronchoscopes including duodenoscopes, and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the endoscope processing cycle (4 minute generation sequence followed by a 6-minute exposure sequence).

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word "STERIS" is written in a serif font in all capital letters. Below the word is a graphic of several horizontal wavy lines stacked on top of each other. The lines are blue.

510(k) Summary For Reliance Endoscope Processing System

STERIS 5960 Heisley Road Mentor, OH 44060 Telephone: 440-354-2600 Fax No: 440-357-9198

Contact: Marcia Benedict Senior Director, Regulatory Affairs Telephone: 440-392-7063 Fax No: 440-357-9198

January 15, 2021 Summary Date:

Premarket Notification number: K203223

STERIS ▪ 5960 Heisley Road ▪ Mentor, OH 44060-1834 USA ▪ 440-354-2600

{4}------------------------------------------------

1. Device Name

Trade Name: Reliance Endoscope Processing System Device Classification: Class II Common/usual Name: Automated Endoscope Reprocessor Classification Name: Accessories, Germicide, Cleaning, For Endoscopes Classification Number: 21 CFR 876.1500 Product Code: NZA

Predicate Device 2.

Reliance Endoscope Processing System, K200989

3. Description of Device

The Reliance Endoscope Processing System is a high level disinfection system that can wash and high level disinfect up to two manually precleaned, immersible, reusable, heatsensitive, semi-critical devices such as GI flexible endoscopes and related accessories.

The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.

  • The Reliance Endoscope Processor is an electromechanical washer/high level . disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor self-decontamination cycles.
  • . Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use oxidative chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor.
  • Optional washing is provided through the automated delivery of Klenzyme Enzymatic ● Presoak and Cleaner during the wash phase of the cycle.
  • . CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, is used in one of the two self-decontamination cycles provided by the processor.
  • . Various accessories are available to accommodate the processing needs of specific endoscopes and endoscopic accessories.
  • . VERIFY Reliance CI Process Indicator is available to monitor for the presence of the Reliance DG active ingredient, peracetic acid.

{5}------------------------------------------------

The Reliance Endoscope Processing Cycle has the following features:

  • The first part of this cycle is an optional programmable washing phase. This phase consists of a wash that uses Klenzyme, followed by a rinse. The washing phase can be programmed on or off. In the "on" mode, the user can choose either one or two washing phases per processing cycle, and the wash time can be adjusted to be between 5 and 10 minutes. The Reliance washing phase does not replace manual cleaning by the user.
  • . The second part is a high level disinfection phase that is non-optional and the parameters cannot be changed by the user. In this phase, the proprietary Reliance DG components, provided in a single use container, are dissolved with water at ~50°C for four minutes of generation time and circulated throughout the processor and through device lumens for a 6 minutes of high level disinfecting solution exposure time.
  • . Following the high level disinfection phase, the Reliance Endoscope Processor removes the high level disinfecting solution through a rinse phase which is nonoptional and the parameters cannot be changed by the user. The processor filters the rinse water (as well as all of the water used throughout the cycle) through a 0.2 micron bacterial-retentive filter. It also incorporates an automatic internal integrity check of this filter at the end of each processing cycle. If the integrity check fails, an alarm alerts the user, and the processor does not complete the cycle.
  • . The last step in the processing cycle is an air purge phase using HEPA-filtered air. The air purge helps to remove excess rinse water from the processed devices. The final air purge is preset to run for 4 minutes; additional air purge time may be selected by the operator.
  • . The processor will print a detailed cycle summary at the end of each cycle that includes information such as processor number, cycle date, start and stop times, as well as phase parameters. With an optional bar code reader, the printouts can also include identification numbers for the operator, patient, device, doctor and procedure.

The processor features two decontamination cycles that are to be used without endoscopes in the processor:

  • The first, called D-SHORT, consists of hot water circulating through the processor for ● 10 minutes, followed by a 10-minute hot air purge. This cycle is to be run every 54 hours. D-SHORT is intended to prevent biofilm from forming.
  • . The second, called D-LONG, consists of a cycle in which CIP 200 Acid-Based Process and Research Cleaner is added to hot water. The cleaning solution is then circulated through the processor for 20 minutes; this is followed by three rinses to remove the solution from the processor and a 10-minute hot air purge. D-LONG is to be used on those occasions when the D-SHORT cycle has not been run within the past 54 hours.

{6}------------------------------------------------

4. Intended Use

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semicritical devices such as bronchoscopes, GI flexible endoscopes including duodenoscopes, and their accessories. High level disinfection is achieved within the 50-57℃ HLD phase of the endoscope processing cycle (4-minute generation sequence followed by a 6-minute exposure sequence).

5. Technological Characteristic Comparison Tables

The Reliance Endoscope Processing System (EPS) is the same as the predicate device; this submission is provided to demonstrate the validity of the system's ongoing use to process duodenoscopes, and to qualify a new source for a proprietary component of Reliance DG Dry Germicide. There were no changes to the Reliance EPS.

A comparison between the predicate and proposed devices can be found in Tables 1 and 2.

FeatureProposed DeviceReliance EPSPredicate DeviceReliance EPSK200989Comparison
Intended UseThe Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as bronchoscopes, GIflexible endoscopes includingduodenoscopes, and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minuteexposure sequence).The Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minuteexposure sequence).Mentionsduodenoscopesspecifically foruser clarity.
OperatingPrinciples/TechnologyThe Reliance EPS and itsintegrated endoscope processingsupport provide for delivery ofsolutions and fluids toendoscopes and theiraccessories. Klenzyme andReliance DG are introduced intothe Reliance EPS to providewashing and high levelThe Reliance EPS and itsintegrated endoscope processingsupport provide for delivery ofsolutions and fluids toendoscopes and theiraccessories. Klenzyme andReliance DG are introduced intothe Reliance EPS to providewashing and high levelIdentical
FeatureProposed DeviceReliance EPSPredicate DeviceReliance EPSK200989Comparison
disinfection. Decontaminationcycles (D-Long with CIP-200and D-Short) are used toprevent biofilm formation in theReliance EPS Processor andensure effective processingfollowing periods of processorinactivity.disinfection. Decontaminationcycles (D-Long with CIP-200and D-Short) are used to preventbiofilm formation in theReliance EPS Processor andensure effective processingfollowing periods of processorinactivity.
ProcessParametersStandardized cycle parameterscannot be altered by operator.The critical process parametersare:• Contact Time• Use Dilution Temperature• Cleaning solution andReliance DG concentration• Water filter integrityStandardized cycle parameterscannot be altered by operator.The critical process parametersare:• Contact Time• Use Dilution Temperature• Cleaning solution andReliance DG concentration• Water filter integrityIdentical
ProcessMonitors• Control Handle Bootpressure alarms if pressuretoo low to process, or if toohigh and could potentiallydamage scopes(specification 6.5 - 10.5PSI).• Detection of a freshReliance DG container inevery processing cycleCleaning Solution levelmonitored; alarm indicateswhen container does nothave sufficient amount tocomplete cycle• Temperature alarms ifoutside of range• Water filter integrity test atend of each high leveldisinfection cycle.• Control Handle Bootpressure alarms if pressuretoo low to process, or if toohigh and could potentiallydamage scopes(specification 6.5 - 10.5PSI).• Detection of a freshReliance DG container inevery processing cycleCleaning Solution levelmonitored; alarm indicateswhen container does nothave sufficient amount tocomplete cycle• Temperature alarms ifoutside of range• Water filter integrity test atend of each high leveldisinfection cycle.Identical
DesignFeatures• Intended for use withReliance DG only• Microprocessor controlled• Internal componentsconstructed of stainlesssteel, silicone,polypropylene and PVDF.• Processor provides 0.2micron filtered water for• Intended for use withReliance DG only• Microprocessor controlled• Internal componentsconstructed of stainlesssteel, silicone,polypropylene and PVDF.• Processor provides 0.2micron filtered water forIdentical
FeatureProposed DeviceReliance EPSPredicate DeviceReliance EPSK200989Comparison
CycleDescriptionwashing, disinfection andrinsing Automated injection ofcleaning solutions Automated generation anddelivery of high leveldisinfecting solution Air intake for Air Purge isHEPA filtered Optional: Wash 1 with 1rinse or Wash 2 with 1rinse Wash time can beadjusted to be between5 and 10 minutes, 40second rinse Additional "smart rinse"occurs after optional washif boot pressure < 7.0 psi isdetected during the heatingphase prior to generationphase: 40 second rinse HLD Generation: 4 minutes HLD Exposure: 6 minutes Rinse 1: 40 seconds Rinse 2: 40 seconds Air purge Can be adjusted to bebetween 4 and 30 minuteswashing, disinfection andrinsing Automated injection ofcleaning solutions Automated generation anddelivery of high leveldisinfecting solution Air intake for Air Purge isHEPA filtered Optional: Wash 1 with 1rinse or Wash 2 with 1rinse Wash time can beadjusted to be between5 and 10 minutes, 40second rinse Additional "smart rinse"occurs after optional washif boot pressure < 7.0 psi isdetected during the heatingphase prior to generationphase: 40 second rinse HLD Generation: 4 minutes HLD Exposure: 6 minutes Rinse 1: 40 seconds Rinse 2: 40 seconds Air purge Can be adjusted to bebetween 4 and 30minutesIdentical
MaintenanceRequirementsD-SHORT decontaminationcycle required every 54-hours D-LONG decontaminationcycle required if D-SHORTnot performed within past54 hours Periodic cleaning of debrisscreen and spray arms Periodic replacement ofwater and air filtersD-SHORT decontaminationcycle required every 54-hours D-LONG decontaminationcycle required if D-SHORTnot performed within past54 hours Periodic cleaning of debrisscreen and spray arms Periodic replacement ofwater and air filtersIdentical
Accessories
ChemicalIndicatorVERIFY Process Indicator forReliance EPS, K063285.VERIFY Process Indicator forReliance EPS, K063285.Identical
FeatureProposed DeviceReliance EPSPredicate DeviceReliance EPSK200989Comparison
Peracetic acid dose indicator forroutine monitoring of RelianceEPS using Reliance DG.Chemical reaction on indicatorpad to produce color change.Peracetic acid dose indicator forroutine monitoring of RelianceEPS using Reliance DG.Chemical reaction on indicatorpad to produce color change.
Flow UnitsFlow units are required only forscope channels that do not openin the endoscope control handleor mechanical action is requiredfor valve operation. ElevenFlow Units are available.Flow units are required only forscope channels that do not openin the endoscope control handleor mechanical action is requiredfor valve operation. ElevenFlow Units are available.Identical

Table 1. Technological comparison for Reliance EPS

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Table 2. Technological comparison of Reliance DG Dry Germicide
FeatureProposed DeviceReliance DG Dry GermicidePredicate DeviceReliance DG Dry GermicideComparison
Indicationsfor UseThe Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as bronchoscopes, GIflexible endoscopes includingduodenoscopes, and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence).The Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence).Mentionsduodenoscopesspecifically foruser clarity.
GermicideExposureTime (min)for HLD10 minutes(4-minute generation sequencefollowed by a 6-minute exposuresequence)10 minutes(4-minute generation sequencefollowed by a 6-minute exposuresequence)Identical
UseTemperature50 - 57°C50 - 57°CIdentical
ReuseSingle useSingle useIdentical
HumanFactorsDispensed ready to use.Container opened automaticallyby the processor, limiting userexposure to the germicide.Dispensed ready to use.Container opened automaticallyby the processor, limiting userexposure to the germicide.Identical
Physical and Chemical Properties
ContainerSingle-use,2 Compartment ContainerSingle-use,2 Compartment ContainerIdentical
FeatureProposed DeviceReliance DG Dry GermicidePredicate DeviceReliance DG Dry GermicideComparison
Composition,PackagedproductDry/DryDry/DryIdentical
ActiveIngredientPeracetic acid, generated in situPeracetic acid, generated in situIdentical
Mode ofActionIt is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydral and sulfurbonds in proteins and enzymes,particularly in the cell walls¹ -hydroxyl radicals produced fromPAA are bactericidal²-PAA damages the viral capsidand viral nucleic acid3,4It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydral and sulfurbonds in proteins and enzymes,particularly in the cell walls¹-hydroxyl radicals producedfrom PAA are bactericidal²-PAA damages the viral capsidand viral nucleic acid3,4Identical
Rinses2 rinses, 40 seconds each, with0.2µ filtered water.2 rinses, 40 seconds each, with0.2µ filtered water.Identical
Microbial Efficacy
SporicidalActivity ofDisinfectantsAOACOfficialMethod966.04Meets efficacy requirements.5Bacillus subtilisClostridium sporogenesMeets efficacy requirements.5Bacillus subtilisClostridium sporogenesIdentical
ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod966.04Meets efficacy requirements. 5Bacillus subtilisClostridium sporogenesMeets efficacy requirements. 5Bacillus subtilisClostridium sporogenesIdentical
FungicidalActivity ofDisinfectantsAOACOfficialMethod955.17Solution is fungicidal.Trichophyton mentagrophytesSolution is fungicidal.Trichophyton mentagrophytesIdentical
Use-DilutionMethodAOACSolution is bactericidal.Salmonella choleraesuisSolution is bactericidal.Salmonella choleraesuisIdentical
FeatureProposed DeviceReliance DG Dry GermicidePredicate DeviceReliance DG Dry GermicideComparison
OfficialMethods955.14,955.15, 964.02Staphylococcus aureusPseudomonas aeruginosaStaphylococcus aureusPseudomonas aeruginosa
EPA VirucidalTesting(DIS/TSS-7,Nov. 1981)Process conditions are virucidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Process conditions are virucidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Identical
TuberculocidalActivity ofDisinfectantsAOACOfficialMethod965.12Solution is tuberculocidal.Mycobacterium bovisSolution is tuberculocidal.Mycobacterium bovisIdentical
TuberculocidalActivityAscenziQuantitativeSuspensionTestSolution is tuberculocidalMycobacterium terraeSolution is tuberculocidalMycobacterium terraeIdentical
Simulated-UseTestMeets efficacy requirementMycobacterium terraeMeets efficacy requirementMycobacterium terraeIdentical
Biocompatibility
CytotoxicityDeviceExtractsNon-cytotoxic per ISO 10993-5Non-cytotoxic per ISO 10993-5Identical
ResidueReductionAutomatic within the RelianceEndoscope Processor, 2 x 15 L -0.2 μ filtered water rinses afterHLD cycle effectively reducesgermicide residues to safe levels.Automatic within the RelianceEndoscope Processor, 2 x 15 L -0.2 μ filtered water rinses afterHLD cycle effectively reducesgermicide residues to safe levels.Identical
DeviceMaterialCompatibilityCompatible with intendedflexible endoscopes andaccessories established throughtesting finished medical devices.No device functional changes.Some materials show cosmeticchanges such as fading ofexternal markings but allremained legible, and bleachingof black anodized aluminumwithout harm to the basematerialCompatible with intendedflexible endoscopes andaccessories established throughtesting finished medical devices.No device functional changes.Some materials show cosmeticchanges such as fading ofexternal markings but allremained legible, and bleachingof black anodized aluminumwithout harm to the basematerialIdentical

DG Dry Germicide Table 2 Technological comparison of Reliance

{10}------------------------------------------------

! Block, S. ed., Disinfection, Sterilization, and Preservation. 5ª edition, 2001
² Clapp et al., Free Rad. Res., (1994) 21:147-167
³ Maillard et al., J. Appl. Breriol. (1996)

{11}------------------------------------------------

{12}------------------------------------------------

Summary of Non-Clinical Testing 6.

The proposed device and its predicate have the same intended use and technological characteristics. New testing was performed to evaluate the modified device and the results are summarized in Table 3.

TestAcceptance CriteriaResult
High level disinfection efficacy ofselected worst-case duodenoscopemodels, representing multiplemanufacturers and distal tip designs.>6 log reduction ofMycobacterium terrae per siteunder worst case processingconditions.Pass
Biocompatibility of Reliance DryGermicide made with a proprietarycomponent from an alternate sourceMedical devices shall be non-cytotoxic after exposure toReliance DG made with aproprietary component from analternate sourcePass
Stability of Reliance Dry Germicidemade with a proprietary componentfrom an alternate sourceReliance Dry Germicide shallmeet acceptance criteria after 18months storage.Pass
Table 3. Summary of verification activities.
------------------------------------------------------

7. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K200989), Class II (21 CFR 876.1500), product code NZA.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.