(29 days)
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the endoscope processing cycle (4 minute generation sequence followed by a 6-minute exposure sequence).
The Reliance Endoscope Processing System is a high level disinfection system that can wash and high level disinfect up to two manually precleaned, immersible, reusable, heatsensitive, semi-critical devices such as GI flexible endoscopes and related accessories. The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.
The provided text describes a 510(k) premarket notification for a medical device called the "Reliance Endoscope Processing System." The submission is to demonstrate substantial equivalence to a legally marketed predicate device (K102244) and specifically addresses a change to the ink used on the label of the Reliance DG Dry Germicide.
Therefore, the study described is not for the primary performance of an AI/ML medical device, but rather a biocompatibility study related to a minor modification in a non-AI medical device's labeling. As such, many of the requested elements for an AI/ML device (e.g., sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth establishment) are not applicable (N/A) to this specific submission.
However, I can extract the relevant information regarding the acceptance criteria and the biocompatibility study performed.
Acceptance Criteria and Device Performance (Biocompatibility Study)
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Biocompatibility with alternate ink used on the label of Reliance DG Dry Germicide | Extracts from devices processed in the Reliance EPS with Reliance DG with the alternate ink on the label shall meet the acceptance criteria for cytotoxicity in ISO 10993-5:2009, Annex A Neutral Red Assay | PASS |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated in terms of number of samples, but refers to "Extracts from devices processed in the Reliance EPS with Reliance DG with the alternate ink on the label." This is a laboratory-based biomechanical/biocompatibility test, not a clinical study with patients.
- Data Provenance: The study is a non-clinical, laboratory test (biocompatibility). Country of origin is not specified, but it's a verification activity performed by the manufacturer, STERIS Corporation. It is prospective testing for this specific change.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A. This is a non-clinical biocompatibility test with a defined chemical/biological assay (ISO 10993-5:2009, Annex A Neutral Red Assay). The "ground truth" is measured objectively by the assay, not by expert consensus.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A. Not applicable for a laboratory biocompatibility test. The result is a direct measurement against a predefined standard.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- N/A. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- N/A. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For this biocompatibility study: The "ground truth" is the objective measurement of cytotoxicity as defined by the ISO 10993-5:2009, Annex A Neutral Red Assay.
8. The sample size for the training set:
- N/A. This is not an AI/ML device, and no training set is described or relevant for this type of non-clinical testing.
9. How the ground truth for the training set was established:
- N/A. As no training set is relevant, this question is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 14, 2020
STERIS Corporation Jennifer Nalepka Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K200989
Trade/Device Name: Reliance Endoscope Processing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NZA Dated: April 14, 2020 Received: April 15, 2020
Dear Jennifer Nalepka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K200989
Device Name
Reliance Endoscope Processing System
Indications for Use (Describe)
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the endoscope processing cycle (4 minute generation sequence followed by a 6-minute exposure sequence).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is in a bold, serif font at the top of the image. Below the word is a graphic of several horizontal, wavy lines in blue. The lines are stacked on top of each other, creating a visual effect of water or waves.
510(k) Summary For Reliance Endoscope Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Telephone: 440-354-2600 Fax No: 440-357-9198
Contact: Jennifer Nalepka Lead Regulatory Affairs Specialist Telephone: 440-392-7458 Fax No: 440-357-9198
Summary Date: May 7, 2020
Premarket Notification Number: K200989
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
1. Device Name
Trade Name: Reliance Endoscope Processing System Device Classification: Class II Common/usual Name: Automated Endoscope Reprocessor Classification Name: Accessories, Germicide, Cleaning, For Endoscopes Classification Number: 21 CFR 876.1500 Product Code: NZA
2. Predicate Device
Reliance Endoscope Processing System, K102244
3. Description of Device
The Reliance Endoscope Processing System is a high level disinfection system that can wash and high level disinfect up to two manually precleaned, immersible, reusable, heatsensitive, semi-critical devices such as GI flexible endoscopes and related accessories.
The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.
- The Reliance Endoscope Processor is an electromechanical washer/high level . disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor self-decontamination cycles.
- Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use oxidative . chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor.
- Optional washing is provided through the automated delivery of Klenzyme Enzymatic Presoak and Cleaner during the wash phase of the cycle.
- CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, is used ● in one of the two self-decontamination cycles provided by the processor.
- . Various accessories are available to accommodate the processing needs of specific endoscopes and endoscopic accessories.
- Verify Reliance CI Process Indicator is available to monitor for the presence of the ● Reliance DG active ingredient, peracetic acid.
The Reliance Endoscope Processing Cycle has the following features:
- The first part of this cycle is an optional programmable washing phase. This phase ● consists of a wash that uses Klenzyme, followed by a rinse. The washing phase can
{5}------------------------------------------------
be programmed on or off. In the "on" mode, the user can choose either one or two washing phases per processing cycle, and the wash time can be adjusted to be between 5 and 10 minutes. The Reliance cleaning phase does not replace manual cleaning by the user.
- . The second part is a high level disinfection phase that is non-optional and the parameters cannot be changed by the user. In this phase, the proprietary Reliance DG components, provided in a single use container, are dissolved with water at ~50°C for four minutes of generation time and circulated throughout the processor and through device lumens for a 6 minutes of high level disinfecting solution exposure time.
- Following the high level disinfection phase, the Reliance Endoscope Processor removes the high level disinfecting solution through a rinse phase which is nonoptional and the parameters cannot be changed by the user. The processor filters the rinse water (as well as all of the water used throughout the cycle) through a 0.2 micron bacterial-retentive filter. It also incorporates an automatic internal integrity check of this filter at the end of each processing cycle. If the integrity check fails, an alarm alerts the user, and the processor does not complete the cycle.
- . The last step in the processing cycle is an air purge phase using HEPA-filtered air. The air purge helps to remove excess rinse water from the processed devices. The final air purge is preset to run for 4 minutes; additional air purge time may be selected by the operator.
- . The processor will print a detailed cycle summary at the end of each cycle that includes information such as processor number, cycle date, start and stop times, as well as phase parameters. With an optional bar code reader, the printouts can also include identification numbers for the operator, patient, device, doctor and procedure.
The processor features two decontamination cycles that are to be used without endoscopes in the processor:
- The first, called D-SHORT, consists of hot water circulating through the processor for . 10 minutes, followed by a 10-minute hot air purge. This cycle is to be run every 54 hours. D-SHORT is intended to prevent biofilm from forming.
- . The second, called D-LONG, consists of a cycle in which CIP 200 Acid-Based Process and Research Cleaner is added to hot water. The cleaning solution is then circulated through the processor for 20 minutes; this is followed by three rinses to remove the solution from the processor and a 10-minute hot air purge. D-LONG is to be used on those occasions when the D-SHORT cycle has not been run within the past 54 hours.
4. Intended Use
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semicritical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50-57°C HLD phase of the endoscope processing cycle (4-minute generation sequence followed by a 6-minute exposure sequence).
{6}------------------------------------------------
5. Technological Characteristic Comparison Tables
The Reliance Endoscope Processing System (EPS) is the same as the predicate device; the specific modification described in this submission is for a change to the ink on the label of the Reliance DG Dry Germicide. A comparison between the predicate and proposed devices can be found in Tables 1 and 2.
| Table 1: Technological comparison for Reliance EPS | |||
|---|---|---|---|
| Feature | Proposed DeviceReliance EPS | Predicate DeviceReliance EPSK102244 | Comparison |
| Intended Use | The Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence). | The Reliance EndoscopeProcessing System is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50-57°C HLD phase of theendoscope processing cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence). | Identical |
| OperatingPrinciples/Technology | The Reliance EPS and itsintegrated endoscope processingsupport provide for delivery ofsolutions and fluids toendoscopes and their accessories.Klenzyme and Reliance DG areintroduced into the Reliance EPSto provide washing and highlevel disinfection.Decontamination cycles (D-Longwith CIP-200 and D-Short) areused to prevent biofilmformation in the Reliance EPSProcessor and ensure effectiveprocessing following periods ofprocessor inactivity. | The Reliance EPS and itsintegrated endoscope processingsupport provide for delivery ofsolutions and fluids toendoscopes and their accessories.Klenzyme and Reliance DG areintroduced into the Reliance EPSto provide washing and highlevel disinfection.Decontamination cycles (D-Longwith CIP-200 and D-Short) areused to prevent biofilmformation in the Reliance EPSProcessor and ensure effectiveprocessing following periods ofprocessor inactivity. | Identical |
| ProcessParameters | Standardized cycle parameterscannot be altered by operator.The critical process parametersare:● Contact Time● Use Dilution Temperature | Standardized cycle parameterscannot be altered by operator.The critical process parametersare:● Contact Time● Use Dilution Temperature | Identical |
| Table 1. Technological comparison for Reliance EPS | ||
|---|---|---|
{7}------------------------------------------------
| Feature | Proposed DeviceReliance EPS | Predicate DeviceReliance EPSK102244 | Comparison |
|---|---|---|---|
| Cleaning solution and Reliance DG concentration Water filter integrity | Cleaning solution and Reliance DG concentration Water filter integrity | ||
| Control Handle Boot pressure alarms if pressure too low to process, or if too high and could potentially damage scopes (specification 6.5 to 10.5 PSI). Detection of a fresh Reliance DG container in every processing cycle Cleaning Solution level monitored; alarm indicates when container does not have sufficient amount to complete cycle Temperature alarms if outside of range Water filter integrity test at end of each high level disinfection cycle. | Control Handle Boot pressure alarms if pressure too low to process, or if too high and could potentially damage scopes (specification 6.5 to 10.5 PSI). Detection of a fresh Reliance DG container in every processing cycle Cleaning Solution level monitored; alarm indicates when container does not have sufficient amount to complete cycle Temperature alarms if outside of range Water filter integrity test at end of each high level disinfection cycle. | ||
| ProcessMonitors | Identical | ||
| DesignFeatures | Intended for use with Reliance DG only Microprocessor controlled Internal components constructed of stainless steel, silicone, polypropylene and PVDF. Processor provides 0.2 micron filtered water for washing, disinfection and rinsing Automated injection of cleaning solutions Automated generation and delivery of high level disinfecting solution Air intake for Air Purge is | Intended for use with Reliance DG only Microprocessor controlled Internal components constructed of stainless steel, silicone, polypropylene and PVDF. Processor provides 0.2 micron filtered water for washing, disinfection and rinsing Automated injection of cleaning solutions Automated generation and delivery of high level disinfecting solution Air intake for Air Purge is | Identical |
| Feature | Proposed DeviceReliance EPS | Predicate DeviceReliance EPSK102244 | Comparison |
| CycleDescription | Optional: Wash 1 with 1• rinse or Wash 2 with 1 rinseo Wash time can beadjusted to be between 5and 10 minutes, 40second rinseAdditional "smart rinse"• occurs after optional wash ifboot pressure < 7.0 psi isdetected during the heatingphase prior to generationphase: 40 second rinse• HLD Generation: 4 minutes• HLD Exposure: 6 minutes• Rinse 1: 40 seconds• Rinse 2: 40 seconds• Air purgeo Can be adjusted to bebetween 4 and 30minutes | Optional: Wash 1 with 1• rinse or Wash 2 with 1 rinseo Wash time can beadjusted to be between 5and 10 minutes, 40second rinseAdditional "smart rinse"• occurs after optional wash ifboot pressure < 7.0 psi isdetected during the heatingphase prior to generationphase: 40 second rinse• HLD Generation: 4 minutes• HLD Exposure: 6 minutes• Rinse 1: 40 seconds• Rinse 2: 40 seconds• Air purgeo Can be adjusted to bebetween 4 and 30minutes | Identical |
| MaintenanceRequirements | • D-SHORT decontaminationcycle required every 54-hours• D-LONG decontaminationcycle required if D-SHORTnot performed within past 54hours.• Periodic cleaning of debrisscreen and spray arms• Periodic replacement ofwater and air filters | • D-SHORT decontaminationcycle required every 54-hours• D-LONG decontaminationcycle required if D-SHORTnot performed within past 54hours.• Periodic cleaning of debrisscreen and spray arms• Periodic replacement ofwater and air filters | Identical |
| Accessories | |||
| ChemicalIndicator | Reliance CI Process Indicator,K063285. Peracetic acid doseindicator for routine monitoringof Reliance EPS using RelianceDG. Chemical reaction onindicator pad to produce colorchange. | Reliance CI Process Indicator,K063285. Peracetic acid doseindicator for routine monitoringof Reliance EPS using RelianceDG. Chemical reaction onindicator pad to produce colorchange. | Identical |
| Flow Units | Flow units are required only forscope channels that do not openin the endoscope control handleor mechanical action is requiredfor valve operation. 11 FlowUnits are available. | Flow units are required only forscope channels that do not openin the endoscope control handleor mechanical action is requiredfor valve operation. 11 FlowUnits are available. | Identical |
| Proposed Device | Predicate Device | Comparison | |
| Feature | Reliance DG Dry Germicide | Reliance DG Dry Germicide | |
| Indicationsfor Use | High level disinfectant for useonly in the Reliance EndoscopeProcessor. The REPS is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50 - 57°C HLD Phase of theEndoscope Processing Cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence). | High level disinfectant for useonly in the Reliance EndoscopeProcessor. The REPS is intendedfor washing and high leveldisinfection of up to twomanually pre-cleaned,immersible, reusable, heat-sensitive, semi-critical devicessuch as GI flexible endoscopes,bronchoscopes and theiraccessories. High leveldisinfection is achieved withinthe 50 - 57°C HLD Phase of theEndoscope Processing Cycle (4-minute generation sequencefollowed by a 6-minute exposuresequence). | Identical |
| GermicideExposureTime (min)for HLD | 10 minutes(4-minute generation sequencefollowed by a 6-minute exposuresequence) | 10 minutes(4-minute generation sequencefollowed by a 6-minute exposuresequence) | Identical |
| UseTemperature | 50 - 57°C | 50 - 57°C | Identical |
| Reuse | Single use | Single use | Identical |
| HumanFactors | Dispensed ready to use.Container opened automaticallyby the processor, limiting userexposure to the germicide. | Dispensed ready to use.Container opened automaticallyby the processor, limiting userexposure to the germicide. | Identical |
| Physical and Chemical Properties | |||
| Container | Single-use,2 Compartment Container | Single-use,2 Compartment Container | Identical |
| Composition,Packagedproduct | Dry/Dry | Dry/Dry | Identical |
| ActiveIngredient | Peracetic acid, generated in situ | Peracetic acid, generated in situ | Identical |
| Feature | Proposed DeviceReliance DG Dry Germicide | Predicate DeviceReliance DG Dry Germicide | Comparison |
| Mode ofAction | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydral and sulfurbonds in proteins and enzymes,particularly in the cell walls1 -hydroxyl radicals produced fromPAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4 | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydral and sulfurbonds in proteins and enzymes,particularly in the cell walls1-hydroxyl radicals produced fromPAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4 | Identical |
| Rinses | 2 rinses, 40 seconds each, with0.2µ filtered water. | 2 rinses, 40 seconds each, with0.2µ filtered water. | Identical |
| Microbial Efficacy | |||
| SporicidalActivity ofDisinfectantsAOACOfficialMethod966.04 | Meets efficacy requirements.5Bacillus subtilisClostridium sporogenes | Meets efficacy requirements.5Bacillus subtilisClostridium sporogenes | Identical |
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod966.04 | Meets efficacy requirements.5Bacillus subtilisClostridium sporogenes | Meets efficacy requirements.5Bacillus subtilisClostridium sporogenes | Identical |
| FungicidalActivity ofDisinfectantsAOACOfficialMethod955.17 | Solution is fungicidal.Trichophyton mentagrophytes | Solution is fungicidal.Trichophyton mentagrophytes | Identical |
| Use-DilutionMethodAOAC,OfficialMethods955.14,955.15, 964.02 | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosa | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosa | Identical |
{8}------------------------------------------------
{9}------------------------------------------------
Table 2. Technological comparison of Reliance DG Dry Germicide
{10}------------------------------------------------
' Block, S. ed., Disinfection, Sterilization, and Preservation. 5ª edition, 2001
² Clapp et al., Free Rad. Res., (1994) 21:147-167
³ Maillard et al., J. Med Microbiol. (1996)
{11}------------------------------------------------
| Feature | Proposed DeviceReliance DG Dry Germicide | Predicate DeviceReliance DG Dry Germicide | Comparison | ||
|---|---|---|---|---|---|
| EPA VirucidalTesting(DIS/TSS-7,Nov. 1981) | Process conditions are virucidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1 | Process conditions are virucidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1 | Identical | ||
| TuberculocidalActivity ofDisinfectantsAOACOfficialMethod965.12 | Solution is tuberculocidal.Mycobacterium bovis | Solution is tuberculocidal.Mycobacterium bovis | Identical | ||
| TuberculocidalActivityAscenziQuantitativeSuspensionTest | Solution is tuberculocidalMycobacterium terrae | Solution is tuberculocidalMycobacterium terrae | Identical | ||
| Simulated-UseTest | Meets efficacy requirementMycobacterium terrae | Meets efficacy requirementMycobacterium terrae | Identical | ||
| Biocompatibility | |||||
| CytotoxicityDeviceExtracts | Non-cytotoxic per ISO 10993-5 | Non-cytotoxic per ISO 10993-5 | Identical | ||
| ResidueReduction | Automatic within the RelianceEndoscope Processor, 2 x 15 L -0.2 µ filtered water rinses afterHLD cycle effectively reducesgermicide residues to safe levels. | Automatic within the RelianceEndoscope Processor, 2 x 15 L -0.2 µ filtered water rinses afterHLD cycle effectively reducesgermicide residues to safe levels. | Identical | ||
| DeviceMaterialCompatibility | Compatible with intended flexibleendoscopes and accessoriesestablished through testingfinished medical devices. Nodevice functional changes.Some materials show cosmeticchanges such as fading ofexternal markings but allremained legible, and bleachingof black anodized aluminumwithout harm to the basematerial. | Compatible with intended flexibleendoscopes and accessoriesestablished through testingfinished medical devices. Nodevice functional changes.Some materials show cosmeticchanges such as fading ofexternal markings but allremained legible, and bleachingof black anodized aluminumwithout harm to the basematerial. | Identical |
{12}------------------------------------------------
6. Summary of Non-Clinical Testing
Shown in Table 3 is the new testing that was performed to evaluate the modified device.
| Test | Purpose | Acceptance Criteria | Result |
|---|---|---|---|
| Biocompatibilitywith alternate inkused on the labelof Reliance DGDry Germicide | To ensure the alternate newink used on the label of theReliance DG DryGermicide does not cause acytotoxic response. | Extracts from devicesprocessed in the Reliance EPSwith Reliance DG with thealternate ink on the label shallmeet the acceptance criteria forcytotoxicity in ISO 10993-5:2009, Annex A Neutral RedAssay | PASS |
Table 3. Summary of verification activities
7. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K102244), Class II (21 CFR 876.1500), product code NZA.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.