K Number
K203215

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-06-11

(221 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Radius T wearable thermometer is intended for single-use, continuous noninvasive measurement of body temperature on the upper chest via wireless communication to a smart device application or compatible patient monitor (i.e., Masimo Root, Masimo Rad-97).

The Radius T is indicated for single-use, continuous body temperature measurements of persons 5 years of age or older in hospitals, hospital-type facilities, and home environments.

Device Description

Radius T is a wearable, battery powered thermometer that is applied to the body to continuously provide body temperature measurements. The Radius T continuously approximates based upon the surface temperature at the application site. The body temperature data from the Radius T is transferred wirelessly to a monitoring device (e.g., Root, Rad-97) or smart device application for continuous display and monitoring.

As part of this submission, the software of the Root (K191882) and Rad-97 (K193626) is being modified to support the communication and display compatibility for the Radius T.

AI/ML Overview

The provided text is a 510(k) Premarket Notification from Masimo Corporation for their Radius T Wearable Thermometer. While it mentions clinical testing and provides performance specifications, it does not contain the detailed information required to fully answer all aspects of your request, particularly regarding the specifics of the clinical study, sample size for test datasets, expert qualifications, or adjudication methods.

The document primarily focuses on establishing "substantial equivalence" to a predicate device (FeverScout Continuous Monitoring Thermometer, K181013) rather than a comprehensive, detailed report of a study proving the device meets specific acceptance criteria in the way an AI/ML device study might.

However, based on the information available, here's what can be extracted:


1. Table of Acceptance Criteria and Reported Device Performance

The document specifies performance characteristics for the Radius T. It does not explicitly state these as "acceptance criteria" in a dedicated section with pre-defined pass/fail thresholds for clinical efficacy, but rather as "specifications" and the results from a clinical validation study in accordance with a standard.

Feature / Acceptance Criteria (as per ISO 80601-2-56)Reported Device Performance
Temperature Accuracy - Laboratory$\pm$ 0.1°C ($\pm$ 0.18°F) in the range of 25°C to 43°C (77°F to 109.4°F)
Temperature Accuracy - ClinicalClinical bias of -0.2°C (-0.36°F) with limits of agreement $\leq$ 1.0°C (1.8°F)
Intended Population5 years or older
Application SiteUpper Chest

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document states that a "clinical validation study was performed" to support the performance of the Radius T and evaluated its clinical performance to a reference clinical thermometer in accordance with ISO 80601-2-56. However, the exact sample size for this clinical study (test set) is not provided in the supplied text. It only mentions the "indicated population of the subject device was narrowed to align to the clinically tested population, age group C (older than 5 years) in accordance with ISO 80601-2-56."
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). However, the nature of a clinical validation study for a wearable thermometer implies a prospective collection of data from enrolled subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: This information is not provided in the document.
  • Qualifications of Experts: This information is not provided in the document. For a thermometer, "ground truth" would typically refer to a highly accurate reference thermometer measurement, not necessarily expert radiologist interpretation.

4. Adjudication Method for the Test Set

  • Adjudication Method: This information is not applicable/not provided as the ground truth for a clinical thermometer study would be objective measurements from a reference device, not expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study design is typically used for AI/ML diagnostic imaging devices where human interpretation is a key component. This document describes a clinical electronic thermometer, where the device itself provides the measurement. AI assistance for human "readers" (in this context, users) is not relevant for a direct temperature measurement device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Yes, the core clinical validation study described is a standalone performance study of the Radius T Wearable Thermometer. It evaluates the device's ability to measure body temperature itself against a reference clinical thermometer. The specifications provided (clinical bias and limits of agreement) are measures of this standalone performance. The device provides a direct measurement, and its "algorithm" refers to its internal logic for deriving body temperature from skin surface temperature.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the clinical study was established using a reference clinical thermometer in accordance with ISO 80601-2-56. This is a form of objective, gold-standard measurement.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: This information is not applicable/not provided. This device is a traditional medical device (thermometer) and does not appear to employ machine learning that would require a separate "training set" for an algorithm in the way an AI/ML imaging device would. Its "principle of action" is based on established biophysical principles (heat flux, skin surface temperature correlation to body temperature) and uses a thermistor. Any internal calibration would be based on engineering design rather than a large data-driven training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: This information is not applicable/not provided as there is no mention of an ML training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

June 11, 2021

Masimo Corporation Sindura Penubarthi Regulatory Affairs Manager 52 Discovery Irvine, California 92618

Re: K203215

Trade/Device Name: Radius T Wearable Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 6, 2021 Received: May 17, 2021

Dear Sindura Penubarthi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203215

Device Name Radius T Wearable Thermometer

Indications for Use (Describe)

Radius T wearable thermometer is intended for single-use, continuous noninvasive measurement of body temperature on the upper chest via wireless communication to a smart device application or compatible patient monitor (i.e., Masimo Root, Masimo Rad-97).

The Radius T is indicated for single-use, continuous body temperature measurements of pears of age or older in hospitals, hospital-type facilities, and home environments.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark inside, followed by the company name "Masimo" in a bold, sans-serif font. The checkmark is stylized and appears to be part of the letter "M" in Masimo.

MASIMO CORPORATION
52 Discovery
Irvine, CA 92618

510(k) Summary - K203215

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7541FAX: (949) 297-7592
Date:June 8, 2021
Contact:Sindura PenubarthiRegulatory Affairs ManagerMasimo CorporationPhone: (949) 396-4041
Trade Name:Radius T Wearable Thermometer
Common Name:Electronic Clinical Thermometer
Classification Regulation/Product Code:21 CFR 880.2910, Class II/ FLL
Establishment RegistrationNumber:3011353843
Reason for PremarketNotification:New device
Predicate Device:K181013- FeverScout Continuous Monitoring Thermometer

1. Device Description

Radius T is a wearable, battery powered thermometer that is applied to the body to continuously provide body temperature measurements. The Radius T continuously approximates based upon the surface temperature at the application site. The body temperature data from the Radius T is transferred wirelessly to a monitoring device (e.g., Root, Rad-97) or smart device application for continuous display and monitoring.

As part of this submission, the software of the Root (K191882) and Rad-97 (K193626) is being modified to support the communication and display compatibility for the Radius T.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark inside, followed by the company name in black, stylized font. To the right of the logo is the company's address, which includes the number 52.

MASIMO CORPORATION
52 Discovery
Irvine, CA 92618

The performance specifications for Radius T are listed in Table 1.1-1 below:

Radius T Specifications
FeatureSpecification
Temperature Accuracy - Laboratory$\pm$ 0.1°C ( $\pm$ 0.18°F) in the range of 25°C to 43°C (77°F to 109.4°F)
Temperature Accuracy - ClinicalClinical bias of -0.2°C (-0.36°F) with limits of agreement $\leq$ 1.0°C (1.8°F).
Intended Population5 years or older
Application siteUpper Chest
Product use life/Battery lifeMinimum of 8 days (192 hours) of continuous run time
Environmental
Storage/Transport Temperature-20°C to 50°C (-4°F to 122°F)
Operating Temperature10°C to 40°C (50°F to 104°F)
Storage/Transport Humidity10% RH to 95% RH (non-condensing)
Operating Humidity10% RH to 95% RH (non-condensing)
Atmospheric Pressure700 to 1060 hPa @ ambient temperature and humidity
Wireless
TypeBluetooth Low Energy

2. Intended Use/ Indications For Use

Radius T wearable thermometer is intended for single-use, continuous noninvasive measurement of body temperature on the upper chest via wireless communication to a smart device application or compatible patient monitor (i.e., Masimo Root, Masimo Rad-97).

The Radius T is indicated for single-use, continuous body temperature measurements of persons 5 years of age or older in hospitals, hospital-type facilities, and home environments.

3. Technological Characteristics

Principle of Action

To achieve its intended purpose, the Radius T relies on the principle that the heat flux generated by the body temperature can be inferred by measuring the skin surface temperature to allow for the estimation of the body temperature.

Mechanism of Action for Achieving the Intended Effect

The Radius T achieves its intended effect through application of the battery powered thermometer to the skin surface on the patient's chest. The thermometer measures the skin surface temperature and estimates the body

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark-like symbol inside, followed by the company name "Masimo" in bold, black font. To the right of the name, there is some smaller text that appears to be the company's address, but it is too blurry to read.

MASIMO CORPORATION
52 Discovery
Irvine, CA 92618

temperature continuously. The estimated body temperature is transmitted wirelessly through a secure Bluetooth connection to a compatible device or software application. Once displayed, the temperature data can be stored or used to trigger threshold alarms on the connected device or software application.

4. Summary of Technological Characteristics of a Subject Device to the Predicate

The subject device, Radius T, and the predicate device, FeverScout (K181013), have the following key similarities:

  • Both devices have the same intended use to continuously measure body temperature;
  • Both devices estimate body temperature from a sensor applied to the skin surface;
  • Both devices include single use adhesive sensors; ●
  • Both devices rely on wireless communication for display of the body temperature. .

The subject device, Radius T and the predicate device, FeverScout (K181013), have the following key differences:

  • Subject device is applied to the chest while the predicate device is applied to an axial site; ●
  • Subject device is intended for adults and pediatrics 5 years and older while the predicated ● for adults and pediatrics 29 days and older;
  • Subject device includes a disposable sensor while the predicate device includes reusable patch with ● disposable tabs.

The subject device and predicate device were found to be substantially equivalent based upon both devices having the same intended use and no technological differences that raise new concerns of safety and effectiveness.

Between to the subject device and predicate device, the main technological difference is that the predicate device estimates body temperature from the armpit while the subject device estimates the body temperature from the chest. To support that this difference along with other technological difference do not significant affect the safety and effectiveness of the Radius T clinical and non-clinical testing was conducted. The test results supported the substantial equivalence of the Radius T.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black letters. To the right of the logo is the text "MASIMO CORPORATION", followed by the address "52 Discovery, Irvine, CA 92618".

Radius T Wearable ThermometerFeverScout(K181013)Comparison to Predicate Device
510(k) NumberSubject DevicePredicate Device
General Information
Indications for Use (IFU)Radius T wearable thermometer is intended forsingle-use, continuous noninvasive measurement ofbody temperature on the upper chest via wirelesscommunication to a smart device application orcompatible patient monitor (i.e., Masimo Root,Masimo Rad-97).The Radius T is indicated for single-use, continuousbody temperature measurements of persons 5 yearsof age or older in hospitals, hospital-type facilities,and home environments.The Wireless FeverScout TM continuousMonitoring thermometer is a non-invasive and re-usable electronic device for home use and non-invasive and single patient use in the hospital.This product is intended for non-urgentambulatory continuous armpit temperaturemonitoring from ages 29 days and olderSubject device has a narrower indicatedpopulation of persons 5 years of age orolder versus the predicate which isindicated for ages 29 days and older. Theindicated population of the subject devicewas narrowed to align to the clinicallytested population, age group C (older than5 years) in accordance with ISO 80601-2-56.The indication difference does not resultin a new intended use as both devices areintended to be used as continuousmonitoring thermometers.
Classification Regulation/Product Code21 CFR 880.2910/ FLL21 CFR 880.2910/ FLLSame
Regulation DescriptionClinical electronic thermometerClinical electronic thermometerSame
Principle of OperationThe Radius T relies on the principle that thetemperature measured at the skin surface can beextrapolated to the body temperature.The sensor within the Radius T detects the heat andmeasures the temperature using a resistance thatchanges with heat.FeverScout relies on the principle that thetemperature measured at skin surface can beextrapolated to the body temperature.The sensor within the FeverScout detects the heatand measures the temperature using a resistance thatchanges with heat.Similar. The subject device and predicatedevice are similar in that they both arebased upon the principle that the bodytemperature can be extrapolated from theskin surface temperature. Both devicesutilize a thermistor to detect the heat andmeasure the temperature based upon aresistance that changes with heat. Thepredicate device utilizes the adjustmentfrom the armpit temperature, while the
Radius T Wearable ThermometerFeverScout(K181013)Comparison to Predicate Device
510(k) NumberSubject DevicePredicate Device
subject device utilizes temperature at theupper chest.
To support the difference does notsignificantly affect the safety andeffectiveness of the subject device, thetest methods described in the ISO 80601-2-56 standard were used.
TechnologicalCharacteristics
Laboratory Accuracy$\pm$ 0.1°C in the range of 25°C to 43°C ( $\pm$ 0.18°F in therange of 77°F to 109.4°F)$\pm$ 0.1°C in the range of 37 to 39°C ( $\pm$ 0.18°F in therange of 98.6 to 102.2°F),$\pm$ 0.2°C in the range of 35 to 37°C and 39 to 42°C( $\pm$ 0.36°F in the range of 95 to 98.6°F and 102.2 to107.6°F)Similar. Subject device has a tighteraccuracy specification across a widerperformance range. Provided the widerperformance range there is no significantimpact to the safety and effectiveness ofthe subject device.
Application SiteUpper ChestAxillaryDifferent. The difference in applicationsite was address through performancetesting in accordance with ISO 80601-2-56. The testing supported the differencedoes not raise different questions of safetyand effectiveness.
Validation MethodConformed to ISO 80601-2-56Conformed to ASTM E1112Subject device is validated in accordancewith ISO 80601-2-56 while the predicatedevice was validated in accordance withASTM E1112. As the ISO 80601-2-56 isalso recognized by the FDA, the use of
Radius T Wearable ThermometerFeverScout(K181013)Comparison to Predicate Device
510(k) NumberSubject DevicePredicate Device
Type of Use (Sensor)DisposableDisposable for hospital use; Reusable for OTCSimilar. As both the subject and predicate devices can be used as a disposable, the difference was not found to result in a new intended use.
Type of SensorThermistorThermistorSame
TemperatureMeasurement IntervalsTransmits every 60 secondsTransmits every 15 secondsSimilar. Subject device transmits data less frequently than the predicate. Although the data transmission frequencies are different, the difference is not significant as compared the expected timeframe for meaningful change in body temperature.
Wireless communicationSupported DevicesMobile (Android, Apple), Masimo PatientMonitoring Devices (i.e., Rad-97, Root)Mobile (Android, Apple)Subject device includes communication to Masimo Patient Monitoring Devices (Root and Rad-97).Software integration testing performed supports the difference is not significant.
TypesBluetooth BLEBluetooth BLESame
MechanicalType of ApplicableWearableWearableSame
Overall Dimension5" x 5" x 0.5"2.4" x 1.6" x 0.2"Subject device is provided with a different form factor.Human factors and usability evaluation supports the difference in form factor.
Radius T Wearable ThermometerFeverScout(K181013)Comparison to Predicate Device
510(k) NumberSubject DevicePredicate Device
Weight30 gms7.2 gmsdoes not significantly affect the safety and effectiveness of the subject device.Subject device has a different weight.Human factors and usability evaluation supports the difference in form factor does not significantly affect the safety and effectiveness of the subject device.
BiocompatibilityConformed to ISO 10993-1, 5, 10Conformed to ISO 10993-1, 5, 10Same
Electrical
Power SourceInternal Battery (Lithium Coin Cell)Internal Battery (Lithium Rechargeable)Similar. Both devices are provided with internal batteries. Subject device is disposable and therefore a rechargeable battery is not applicable. Therefore, the difference was not significant.
Battery LifeMinimum of 8 days (192 hours) of continuous run timeMinimum of 7 days of continuous run timeSimilar. Subject device has a longer battery life. As the battery life was longer, the difference was not found to be significant.
Electrical SafetyConformed to IEC 60601-1Conformed to IEC 60601-1Same
Electromagnetic compatibilityConformed to IEC 60601-1-2Conformed to IEC 60601-1-2Same
Environmental
Operating Temperature10°C to 40°C50°F to 104°F10°C to 40°C50°F to 104°FSame
Operating Humidity10% RH to 95% RH (non-condensing)15% RH to 85% RHSimilar. Subject device is provided with a wider environmental specification;
Radius T Wearable ThermometerFeverScout(K181013)Comparison to Predicate Device
510(k) NumberSubject DevicePredicate Device
Atmospheric Pressure700 to 1060 hPa700 to 1060 hPatherefore, the difference was not found tobe significant.
Operating Temperature10°C to 40°C50°F to 104°F10°C to 40°C50°F to 104°FSame
Storage/ TransportTemperature-20°C to 50°C-4°F to 122°F-Predicate device does not disclose thestorage or transport environmentalconditions. The environmental conditiontesting conducted supports the subjectdevice specifications for reasonabletransport and storage conditions. Asresult, any difference was not found to besignificant.
Storage/ TransportHumidity10% RH to 95% RH (non-condensing)-Predicate device does not disclose thestorage or transport environmentalconditions. The environmental conditiontesting conducted supports the subjectdevice specifications for reasonabletransport and storage conditions. Asresult, any difference was not found to besignificant.
Mode of Operation perIEC 60601-1
Mode of OperationContinuousContinuousSame

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618".

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a stylized font, with a red checkmark symbol to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

5. Performance Data

Performance Bench Testing

To support the temperature measurement performance of the Radius T, performance bench testing was conducted in accordance with ISO 80601-2-56 that compared the accuracy of the surface temperature measurements made by the Radius T against a reference temperature source to establish the laboratory accuracy. The performance bench testing supported the performance of the Radius T.

Biocompatibility Testing

The patient contacting materials of the Radius T were tested in accordance with ISO 10993-1, 5 and 10 found to have acceptable biocompatibility.

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

To support the safe use of the Radius T, testing to support EMC compliance with the IEC 60601-1-2 standard, Electrical Safety according to IEC 60601-1 standard, environmental, and mechanical requirements were provided as part of the submission. As the Radius T is disposable, considerations for cleaning were not applicable.

Software Verification and Validation Testing

Software verification and validation testing was conducted and the documentation was provided as recommended by FDA's Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005. The software for Radius T was considered as a "moderate" level of concern, because a failure or latent flaw in the software could directly result in minor to moderate injury to the patient. The testing was found to substantial equivalence of the subject device.

The software for Rad-97 and Root are considered as a "moderate" level of concern, because a failure or latent flaw in the software could directly result in minor to the patient. The testing to support the communication capability with the Rad-97 and Root software was provided to support this submission.

Wireless and Cybersecurity Testing

As the Radius T uses wireless communication to the host device, wireless and cybersecurity considerations were made in accordance with FDA Guidance for Industry and Food and Drug Administration Staff-Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, dated October 2, 2014 and draft guidance dated October 18, 2018.

In accordance with the FDA guidance for Cybersecurity, the Radius T was considered a Tier 2 cybersecurity risk device. To support the cybersecurity of the Radius T, a risk-based approach was used to

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark-like symbol inside, followed by the company name in black, bold letters. To the right of the name, there is some light gray text that appears to be an address, with the number 52 visible.

MASIMO CORPORATION
52 Discovery
Irvine, CA 92618

identify cybersecurity threats and mitigations. Cybersecurity testing was conducted to support the acceptability of the subject device's cybersecurity risk.

As the compatibility of the Radius T to the Rad-97 and Root did not affect the software architecture, the modification did not affect the acceptability of the cybersecurity risks of those devices.

Human Factors and Usability Testing

Human factors engineering/usability engineering (HFE/UE) activities for Radius T relied upon the representative testing of a previously cleared device with similar human factor and use characteristics. The HFE/UE activities were conducted in accordance with the FDA Guidance, Applying Human Factors and Usability Engineering to Medical Devices, February 3, 2016. The HFE/UE testing which used a riskbased approach accounted for the expected user related tasks to the Radius T. The representative testing was found to support acceptability of the human factors and usability risks of the subject device.

Non-clinical Testing

Non-clinical bench testing was included to support the Radius T. The non-clinical testing was conducted in accordance with Masimo requirements to ensure that the specifications of the subject device were met. The following non-clinical testing was performed:

  • Electrical safety testing per IEC 60601-1 .
  • EMC testing per IEC 60601-1-2 ●
  • Usability testing per FDA Human Factors and Usability Guidance ●
  • Software verification and validation testing per FDA Software Guidance ●
  • Biocompatibility testing per ISO 10993-1 .
  • . Mechanical testing per IEC 60601-1

The testing was found to support the subject device met its design requirements.

Clinical Testing

To support the performance of the Radius T. clinical validation study was performed. The study evaluated the clinical performance of the Radius T to a reference clinical thermometer in accordance with ISO 80601-2-56.

6. Conclusion

The non-clinical and clinical data was found to support the Radius T is substantially equivalent to the predicate device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.