K Number
K202926

Validate with FDA (Live)

Device Name
Micro Catheter
Date Cleared
2022-02-10

(499 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micro Catheter is intended for selective delivery of therapeutic devices and infusion of contrast media into the peripheral and neuro vasculature.

Device Description

The proposed device Micro Catheter is a single-lumen catheter designed to be introduced over a steerable guidewire into the peripheral and neuro vasculature. The proximal end of the catheter incorporates standard Luer adapter to facilitate the attachment of accessories. The outer surface of the catheter has a lubricious coating at the distal end of the Micro Catheter has a radiopaque marker at the distal end to facilitate fluoroscopic visualization. The device is intended for single use and is provided sterile.

AI/ML Overview

The provided document is a 510(k) Pre-market Notification for a medical device called "Micro Catheter". It details non-clinical performance and biocompatibility testing to demonstrate substantial equivalence to a predicate device.

However, the document does not contain any information about a study proving the device meets acceptance criteria related to an AI/Machine Learning model, nor does it discuss human reader performance, AI assistance, or the establishment of ground truth for such models. The document focuses on the physical and biological performance of a catheter device through standard engineering and biocompatibility tests.

Therefore, I cannot fulfill your request for the following sections as the information is not present in the provided text:

  • A table of acceptance criteria and the reported device performance: The document provides a table of test methods and results for non-clinical performance and biocompatibility, but these are for the physical device, not an AI model.
  • Sample sized used for the test set and the data provenance: Not applicable as no AI/ML model testing is described.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  • Adjudication method: Not applicable.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • The type of ground truth used: Not applicable.
  • The sample size for the training set: Not applicable.
  • How the ground truth for the training set was established: Not applicable.

The document pertains to the regulatory clearance of a physical medical device (a micro catheter) and its performance is evaluated through standard engineering and biological compatibility tests, not through the evaluation of an AI or machine learning algorithm.

{0}------------------------------------------------

February 10, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shanghai Heartcare Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 China

Re: K202926

Trade/Device Name: Micro Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: January 8, 2022 Received: January 11, 2022

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K202926

Device Name Micro Catheter

Indications for Use (Describe)

The Micro Catheter is intended for selective delivery of therapeutic devices and infusion of contrast media into the peripheral and neuro vasculature.

Type of Use (Select one or both, as applicable)
✘ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of 21 CFR 807.92.

510(k) Number: __ K202926

Date of Preparation: 02/10/2022

1. Sponsor Identification

Shanghai Heartcare Medical Technology Co., Ltd.

590 Ruiqing Rd, Building 4, Suite 201, East Zhangjing High-Tech Park, Shanghai, P.R. China

Establishment Registration Number: Not registered

Contact Person: Zongyu Xue Position: RA Director Tel: +86-18621683501 Fax: +86-21-68798512 Email: zyxue@strokemedical.com

2. Designated Submission Correspondent

Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

3. Identification of Proposed Device

Trade Name: Micro Catheter Common Name: Intravascular Catheter

Regulatory Information

{4}------------------------------------------------

Primary Product Code Classification Name: Catheter, Percutaneous, Neurovasculature Classification: II Product Code: QJP Regulation Number: 21CFR 870.1250 Review Panel: Neurology

Secondary Product Code Classification Name: Catheter, Percutaneous Classification: II Product Code: DQY Regulation Number: 21CFR 870.1250 Review Panel: Cardiovascular

Identification of Predicate Device and Reference Devices 4.

Predicate Device 510(k) Number: K192122 Product Name: Trevo Trak 21 Microcatheter

Reference Device 1 510(k) Number: K993672 Product Name: Rebar Micro Catheter

Reference Device 2 510(k) Number: K001966 Product Name: Rebar Micro Catheter

રં. Device Description

The proposed device Micro Catheter is a single-lumen catheter designed to be introduced over a steerable guidewire into the peripheral and neuro vasculature. The proximal end of the catheter incorporates standard Luer adapter to facilitate the attachment of accessories. The outer surface of the catheter has a lubricious coating at the distal end of the Micro Catheter has a radiopaque marker at the distal end to facilitate fluoroscopic visualization. The device is intended for single use and is provided sterile.

{5}------------------------------------------------

6. Indications for Use

The Micro Catheter is intended for selective delivery of therapeutic devices and into the peripheral and neuro vasulature.

7. Comparison of Technological Characteristics

ItemSubject DevicePredicate Device K192122Reference Device 1 K993672Reference Device 2 K001966
ClassificationIIIIIIII
Product CodeQJP, DQYDQO, DQYKRAKRA
Regulation Number21 CFR 870.125021 CFR 870.120021 CFR 870.125021 CFR 870.121021 CFR 870.1210
Indication for UseThe Micro Catheter isintended for selectivedelivery of therapeuticdevices and infusion ofcontrast media into theperipheral and neurovasculature.The Microcatheter is indicatedfor use in the selectiveplacement of devices and/orfluids, such as contrast media,into the peripheral, coronary,and neuro vasculature duringdiagnostic and/or therapeuticprocedures.The Rebar Micro Catheter isintended for the controlledselective infusion of physician-specified therapeutic agents orcontrast media into thevasculature of the peripheraland neuro anatomy.The Rebar Micro Catheter isintended for the controlledselective infusion of physician-specified therapeutic agents orcontrast media into thevasculature of the peripheraland neuro anatomy.
ComponentRuhr HandleOuter Stress TubeInner Stress TubeCoil TubeOuter JacketShaft BraidStrain ReliefInner LayerCatheter HubRuhr HandleOuter Stress TubeInner Stress TubeCoil TubeRuhr HandleOuter Stress TubeInner Stress TubeCoil Tube
Inner Diameter0.021, 0.027 inch0.021 inch0.027 inch0.021 inch
Proximal OuterDiameter2.85 Fr (0.95 mm)2.85 Fr (0.95 mm)2.8 Fr (0.93 mm)2.8 Fr (0.93 mm)
Distal Outer Diameter2.55 Fr (0.85 mm),2.85 Fr (0.95 mm)2.4 Fr (0.80 mm),2.0 Fr (0.67 mm)2.8 Fr (0.93 mm)2.3 Fr (0.76 mm)
Effective length110, 130, 145, 153 cm162 cm110, 130, 145 cm110, 130, 153 cm
Maximum Guidewire0.010 inch0.018 inch0.021 inch0.018 inch
Radiopaque MarkerYesYesYesYes
Single UseYesYesYesYes
Materials
Inner Layer of CoilTubePolytetrafluoroethylenePTFEPolytetrafluoroethylenePolytetrafluoroethylene
Outer Layer of CoilTubePebaxPolyolefinPebaxPebax
CoatingPolyvinylpyrrolidone (PVP)Hydrophilic CoatingHyaluronic AcidHyaluronic Acid
Radiopaque MarkerPlatinum-iridium alloyPlatinum/iridiumPlatinum-iridium alloyPlatinum-iridium alloy
Ruhr HandlePolycarbonate (PC)Polyurethaneunknownunknown
AdhesiveDymax 204-CTH-FAcrylic (Acrylated Urethane)unknownunknown

Table 1 Comparison of Technological Characteristics

{6}------------------------------------------------

{7}------------------------------------------------

8. Non-Clinical Performance Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications and was substantially equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-3: 2014 Biological evaluation of medical devices-Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity

  • ISO 10993-4: 2017 Biological evaluation of medical devices-Part 4: Selection of tests for interactions with blood

  • ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitrocytotoxicity

  • ISO 10993-7: 2008 Biological evaluation of medical devices-Part 7: Ethylene oxide sterilization residuals

  • ISO 10993-10: 2010 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization

  • ISO 10993-11: 2017 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity

  • USP <85> Bacterial Endotoxins Test

  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

  • ASTM F88/88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials

  • ISO 594-1: 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements

  • ISO 594-2: 1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings

  • ISO 10555-1: 2013 Intravascular catheters - Sterile and single-use intravascular catheters - Part 1: General requirements

  • ISO 80369-7: 2106 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications

The results of verification and validation testing conducted on the Micro Catheter demonstrate that it performs as designed and is suitable for its intended use. A summary of the tests performed is provided in the table below:

TestTest Method SummaryResults and Conclusion
Surface EvaluationObserve the catheter by eye or under microscope at X 2.5 magnification for any structural or mechanical damage.Surface met acceptance criteria.
RadiopacityThe radiopaque marker on the catheter tip should be visible under X - ray.The radiopaque marker on the catheter tip is visible under X-ray.
DimensionalverificationVerify dimensions using specifiedmeasurement tools. Record measurements.Size verification met acceptancecriteria.
Liquid leakageThe proposed device was evaluated per ISO10555-1 to demonstrate that the devicemeets the liquid leakage under pressurerequirements.No liquid leakage.
Static burst pressureBurst pressure tests were performed atpressures greater than the manual syringeinjection pressures.Burst pressure met acceptancecriteria.
Peak tensile forceUse a tensile test machine to apply a tensileload to the sample and determine whether themaximum tensile force meets the acceptancecriteria.Peak tensile force met acceptancecriteria.
Torque StrengthFix the distal end of the catheter and rotatethe proximal end until the failure. Record thenumber of rotation and the failure mode.Torque strength met acceptancecriteria.
TorqueabilityRotate the proximal end of the catheter 90degrees to observe the distal end of thecatheter and calculate a proximal-to-distalrotational ratio for the device.Torqueability met acceptancecriteria.
Kink resistanceThe proposed device was evaluated perFDA guidance “Non-Clinical EngineeringTests and Recommended Labeling forIntravascular Stents and AssociatedDelivery Systems", section IV, C.9.Kink resistance met acceptancecriteria.
TrackabilityThe proposed device was evaluated per FDAguidance "Non-Clinical Engineering Testsand Recommended Labeling forIntravascular Stents and AssociatedDelivery Systems", section IV, C.2.Trackability met acceptancecriteria.
Particulate testingAfter simulated use testing with compatibledevices, determine the quantity and size ofthe particles generated.The number and size of theparticles were similar to that ofthe predicate device.
Coating integrityAfter simulated use testing with compatibledevices, the coating is dyed and checked forcoating defects at X 500 magnification.Coating integrity metacceptance criteria.
Tip flexibilityThe proposed device was evaluated per FDA guidance “Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling”, section IV, G.15.Tip flexibility met acceptance criteria.
Air leakageThe proposed device was evaluated per ISO 10555-1 to demonstrate that the device meets the hub aspiration air leakage requirements.No air leakage.
Corrosion resistanceThe proposed device was evaluated per ISO 10555-1 to demonstrate that the device is corrosion resistant.Corrosion resistance met acceptance criteria.
Connector performanceThe proposed device was evaluated per ISO 594-1, 594-2 and 80369-7 to demonstrate that the device meets the requirements for small bore connectors.Connector performance met acceptance criteria.
Compatibility testSimulated use testing with compatible devices in a vascular model was performed.The device can be used as intended.

{8}------------------------------------------------

{9}------------------------------------------------

Biocompatibility

The device is categorized as Externally Communicating Device, Circulating Blood, Limited Contact (< 24 hours), per ISO 10993-1, the following testing was conducted:

TestReference StandardTest SummaryConclusion
CytotoxicityTested in accordance with ISO10993-5:2009, BiologicalEvaluation of Medical Devices –Part 5: Tests for in vitro toxicity.Verify the viability, ifviability is reduced to < 70%of the blank, it has acytotoxic potential.Non-cytotoxic
IrritationTested in accordance with ISO10993-10:2010, BiologicalEvaluation of Medical Devices –Part 10: Tests for Irritation andSkin Sensitization.Study animals tested with thesubject device were observedfor dermal sensitization.No irritation
SensitizationTested in accordance with ISO10993-10, Biological Evaluationof Medical Devices – Part 10Tests for Irritation and SkinSensitization.Study animals tested with thesubject device were observedfor dermal sensitization.No sensitization
Systemic ToxicityTested in accordance with ISO10993-11:2017,BiologicalEvaluation of Medical Devices –Part 11: Tests for SystemicToxicity.Study animals tested with thesubject device were observedfor abnormal clinical signsindicative of toxicity duringthe 72-hour test period.No systemic toxicity
HemolysisTested in accordance with ASTMF756-17, Standard Practice forAssessment of HemolyticProperties of Materials.The difference between thehemolytic indexes of thesubject device and thenegative control wasevaluated.Non-hemolytic
Material MediatedPyrogenicityTested in accordance with ISO10993-11:2017,BiologicalEvaluation of Medical Devices –Part 11: Tests for SystemicToxicity and USP 42 <151>Pyrogen Test.Study animals wereobserved for temperaturerise.Nonpyrogenic
InVivoThromboresistanceTested in accordance with ISO10993-4:2017,BiologicalEvaluation of Medical Devices –Part 4: Selection of Tests forInteractions with Blood.Study animals with subjectdevice were observed forthrombogenic potentials andtest results demonstratedsimilar thromboresistancecharacteristics with thecontrol device.Met thepredeterminedacceptance criteria,the test score is 0,i.e., No thrombosis.
ComplementActivationTested in accordance with ISO10993-4, Biological Evaluationof Medical Devices – Part4:2017: Selection of Tests forInteractions with Blood, SC5b-9Complement Activation.Comparison of the subjectdevice SC5b-9 value to thepredicate device for allexposure times wasperformed.No statisticaldifference from thepredicate device
PartialThromboplastin TimeTested in accordance with ISO 10993-4:2017, BiologicalEvaluation of Medical Devices –Part 4: Selection of Tests forInteractions with Blood, andASTM F2382-2018, StandardTest Method for Assessment ofIntravascular Medical DeviceMaterials on PartialThromboplastin Time.The clotting time wasobserved for both the subjectdevice and the predicate.No statisticaldifference from thepredicate device.
Bacterial ReverseMutationTested in accordance with ISO10993-3:2014 Biologicalevaluation of medical devices-Part 3: Tests for genotoxicity,carcinogenicity and reproductivetoxicity.The difference between thenumber of revertant coloniesof the subject device and thenegative control wasevaluated.No backwardmutation insalmonellatyphimurium.
In Vitro MammalianCell Gene MutationTested in accordance with ISO10993-3:2014 Biologicalevaluation of medical devices-Part 3: Tests for genotoxicity,carcinogenicity and reproductivetoxicity.The difference between themutation frequency of thesubject device and thenegative control wasevaluated.Non-mutagenic toTK+/--3.7.2CSubline of L5178Ycell.

{10}------------------------------------------------

{11}------------------------------------------------

Sterilization and Shelf Life

The Micro Catheter sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10th. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 2.15 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 2 year shelf-life.

9. Clinical Performance Testing

No clinical studies were necessary to demonstrate substantial equivalence.

10. Conclusion

The subject device has similar technological characteristics and intended use as the predicate device. The differences do not raise new questions of safety and effectiveness. The biocompatibility, sterility and bench performance testing demonstrate that the Micro Catheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).