K Number
K202761

Validate with FDA (Live)

Date Cleared
2021-04-11

(202 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The medical surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The proposed device is a three-layer, flat pleated mask. Each mask is composed of a mask body, a nose piece and two ear loops. The mask body is manufactured with three layers, the inner layer and the outer layer are made of spunbond polypropylene nonwoven fabric, and the middle layer is made of meltblown polypropylene nonwoven fabric. The model of the proposed device, ear-loop, is held in place over the user's mouth and nose by two elastic ear loops welded to the mask body. The elastic ear loops are made of knitted elastic loops (made of nylon and spandex). The nose piece is in the layers of face mask to allow the user to fit the face mask around his nose, which is a galvanized wire with polyethylene covering. The proposed device is provided non-sterile and is intended to be a single use, disposable device.

AI/ML Overview

This document describes the premarket notification (510(k)) for a Medical Surgical Mask (Non-sterile). The acceptance criteria and performance study details are provided in the "Non-clinical Testing" section, specifically in Table 2 "Performance Testing" and Table 3 "Biocompatibility Testing".

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The device is a medical surgical mask and its performance is evaluated against the ASTM F2100-19 standard for materials used in medical face masks. The acceptance criteria are specified for Level 1, Level 2, and Level 3 requirements, and the device reports performance that meets all these levels.

TestPurposeAcceptance Criteria (ASTM F2100-19)Reported Device Performance (Statistics of three lots)Verdict
Fluid Resistance (ASTM F1862)Verify the fluid resistance of the proposed device can simultaneously meet the requirements for Level 1, Level 2 and Level 3 specified in ASTM F2100-19.Level 1: ≥29 out of 32 pass at 80 mmHg32 out of 32 pass at 80 mmHgPass
Level 2: ≥29 out of 32 pass at 120 mmHg32 out of 32 pass at 120 mmHgPass
Level 3: ≥29 out of 32 pass at 160 mmHg32 out of 32 pass at 160 mmHgPass
Bacterial Filtration Efficiency (BFE) (ASTM F2101)Verify that bacterial filtration efficiency of the proposed device can simultaneously meet the requirements for Level 1, Level 2 and Level 3 specified in ASTM F2100-19.Level 1: ≥95%99.56%~99.99%Pass
Level 2: ≥98%(Same as above)Pass
Level 3: ≥98%(Same as above)Pass
Particulate Filtration Efficiency (PFE) (ASTM F2299)Verify that particulate filtration efficiency of the proposed device can simultaneously meet the requirements for Level 1, Level 2 and Level 3 specified in ASTM F2100-19.Level 1: ≥95%98.07%~99.24%Pass
Level 2: ≥98%(Same as above)Pass
Level 3: ≥98%(Same as above)Pass
Differential Pressure (Delta-P) (EN 14683)Verify that differential pressure of the proposed device can simultaneously meet the requirements for Level 1, Level 2 and Level 3 specified in ASTM F2100-19.Level 1: < 5.0 mmH₂O/cm²(3.4~3.9) mmH₂O/cm²Pass
Level 2: < 6.0 mmH₂O/cm²(Same as above)Pass
Level 3: < 6.0 mmH₂O/cm²(Same as above)Pass
Flammability (16 CFR 1610)Verify that Flammability of the proposed device can simultaneously meet the requirements for Level 1, Level 2 and Level 3 specified in ASTM F2100-19.Level 1: Class 1Class 1Pass
Level 2: Class 1(Same as above)Pass
Level 3: Class 1(Same as above)Pass
Biocompatibility Testing (ISO 10993-5, -10)In vitro Cytotoxicity: Verify that the proposed device extract is non-cytotoxic.The extract is non-cytotoxic under the research conditions.PassPass
Skin Irritation: Verify that the proposed device extract is non-irritating.The polar and non-polar extracts are non-irritating under the research conditions.PassPass
Skin Sensitization: Verify that the proposed device extract is non-sensitizing.The polar and non-polar extracts are non-sensitizing under the research conditions.PassPass

2. Sample Size Used for the Test Set and Data Provenance

The document states "Results (Statistics of three lots)". This implies that the tests were performed on samples from three different production lots of the medical surgical mask. For Fluid Resistance, a sample size of 32 units (e.g., masks or material sections) per lot was used (as per 29 out of 32 requirement). The exact sample size for other tests like BFE, PFE, Differential Pressure, and Flammability is not explicitly stated as a number, but the results are reported as percentages or ranges, indicating testing on a sufficient number of samples to generate these statistics for each of the three lots.

The data provenance is China, as the applicant is Guangdong GoodFeeling Hygiene Material Tec Co., Ltd. located in China. The data is from non-clinical performance testing, which is inherently prospective testing specific to the device being submitted.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable as the study is a performance test of a physical device (medical mask) against established engineering and material standards, not a clinical study involving interpretation of data by human experts (like radiologists for AI models). The ground truth is the objective measurement defined by the respective ASTM, EN, and CFR standards.

4. Adjudication Method for the Test Set

This section is not applicable for the same reasons as #3. Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers establish ground truth, e.g., in medical image interpretation.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This section is not applicable. The device is a medical mask, not an AI-assisted diagnostic tool or an imaging device. No MRMC study would be relevant for this type of product.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

This section is not applicable. The device is a physical product (a medical mask), not an algorithm or software. Its performance is evaluated through laboratory tests, not algorithmic output.

7. The Type of Ground Truth Used

The ground truth for this medical device is based on established international and national consensus standards for medical face masks and biocompatibility:

  • ASTM F2100-19: Standard Specification for Performance of Materials Used in Medical Face Masks.
  • ASTM F1862/F1862M-17: Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood.
  • ASTM F2101-19: Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials.
  • ASTM F2299/F2299M-03(R2017): Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates.
  • EN 14683:2019+AC:2019: Medical Face Masks - Requirements and Test Methods (specifically for Differential Pressure).
  • 16 CFR 1610: Standard for the Flammability of Clothing Textiles.
  • ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010: Biological evaluation of medical devices (for biocompatibility).

The "ground truth" is therefore derived from the quantitative measurements and qualitative observations (e.g., pass/fail for cytotoxicity) against the specified criteria within these recognized standards.

8. The Sample Size for the Training Set

This section is not applicable. As this is a physical medical device (mask) evaluated through performance and biocompatibility tests, there is no "training set" in the context of machine learning or AI models. The testing is for product verification against standards.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 11, 2021

Guangdong GoodFeeling Hygiene Material Tec Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China

Re: K202761

Trade/Device Name: Medical Surgical Mask (Non-sterile) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 15, 2020 Received: September 21, 2020

Dear Grace Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rvan-O

Ryan Ortega, PhD Acting Assistant Director THT4B2: Personal Protective Equipment, Reprocessing & Disinfection Devices Team DHT4B: Division of Infection Control and Plastic Surgerv Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202761

Device Name Medical Surgical Mask (Non-sterile)

Indications for Use (Describe)

The medical surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Version: A/0

510(k) Summary

K202761

1. Contact Details

1.1 Applicant information

Applicant NameGuangdong GoodFeeling Hygiene Material Tec Co., Ltd.
AddressNo.10 Phoenix Avenue Eco-industrial Park, Danzao Town,Nanhai, Foshan City, Guangdong Province, China
Contact personZheng Zhijin
Phone No.+86-15860481335
E-mail/
Date PreparedApr. 2, 2021

1.2 Submission Correspondent

Image: Company LogoShenzhen Joyantech Consulting Co., Ltd
1713A, 17th Floor, Block A, Zhongguan Times Square,Nanshan District, Shenzhen, Guangdong Province, China
卓远天成
Phone No.86-755-86069197
Contact personGrace Liu; Field Fu;
Contact person's e-mailgrace@cefda.com; field@cefda.com
Websitehttp://www.cefda.com

2. Device Information

Trade nameMedical Surgical Mask (Non-sterile)
Common nameMedical Surgical Mask
ModelGFYY95
ClassificationII
Classification nameMask, Surgical
Product codeFXX
Regulation No.21 CFR 878.4040

3. Legally Marketed Predicate Device

Trade NameSurgical Face Mask, Ear Loops, Model101B, 101G, 136B, 136G, 137B, 137GSurgical Face Mask, Tie-on, Model145B, 145G, 143B, 143G, 138B, 138G, 142B, 142G, 151B, 151G
510(k) NumberK133070
Product CodeFXX
ManufacturerBH Medical Products Co., Ltd.

{4}------------------------------------------------

4. Device Description

The proposed device is a three-layer, flat pleated mask. Each mask is composed of a mask body, a nose piece and two ear loops. The mask body is manufactured with three layers, the inner layer and the outer layer are made of spunbond polypropylene nonwoven fabric, and the middle layer is made of meltblown polypropylene nonwoven fabric.

The model of the proposed device, ear-loop, is held in place over the user's mouth and nose by two elastic ear loops welded to the mask body. The elastic ear loops are made of knitted elastic loops (made of nylon and spandex).

The nose piece is in the layers of face mask to allow the user to fit the face mask around his nose, which is a galvanized wire with polyethylene covering.

The proposed device is provided non-sterile and is intended to be a single use, disposable device.

5. Intended Use/Indication for Use

The medical surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

6. Substantial Equivalence Comparison

ItemProposed Device(K202761)Predicate Device (K133070)Comment
Product nameMedical Surgical Mask (Non-sterile)Surgical Face Mask, Ear Loops,Model101B, 101G, 136B, 136G, 137B,137GSurgical Face Mask, Tie-on,Model145B, 145G, 143B, 143G, 138B,138G, 142B, 142G, 151B, 151GNone
ManufacturerGuangdong GoodFeelingHygiene Material Tec Co., Ltd.BH Medical Products Co., Ltd.None
Product CodeFXXFXXSame
Regulation Number21 CFR § 878.404021 CFR § 878.4040Same
ClassificationClass IIClass IISame
OTC useYesYesSame
Indications for useThe medical surgical masks areintended to be worn to protectboth the patient and healthcarepersonnel from transfer ofmicroorganisms, body fluids andparticulate material. Thesemasks are intended for use ininfection control practices toreduce the potential exposure toThe surgical face masks areintended to be worn to protectboth the patient and healthcarepersonnel from transfer ofmicroorganisms, body fluids andparticulate material. These facemasks are intended for use ininfection control practices toreduce the potential exposure toSame

Table 1 Substantial Equivalence Comparison

{5}------------------------------------------------

Version: A/0

blood and body fluids. This is asingle use, disposable device(s),provided non-sterile.blood and body fluids. This is asingle use, disposable device(s),provided non-sterile.
Design featureEar-loopEar-loop/Tie-onSimilar
Mask style3 flat pleated3 flat pleatedSame
Single useYesYesSame
ColorBlueBlue, GreenSimilar
Specifications andDimensionsLength: 18cm±1cm(7.1"±0.39")Width: 9.5cm(3.7")±1cm(3.7"±0.39")Length: 6.8" ±0.25"Width:3.5"+/-0.25"4.2"+/-0.25"Similar
SterilityNon-SterileNon-SterileSame
MaterialsOuterlayerSpunbond polypropyleneSpunbond polypropyleneSame
MiddlelayerMeltblown polypropyleneMeltblown polypropyleneSame
InnerlayerSpunbond polypropyleneSpunbond polypropyleneSame
NosepieceGalvanized wire withpolyethylene coveringAluminum wireDifferent(Issue 1)
Ear loopNylon and spandexPolyesterDifferent(Issue 2)
ASTM Level (ASTMF2100-19)Level 1Level 2Level 3Level 1Level 2Level 3Same
Fluid ResistancePass at 80 mmHg, 120 mmHgand 160 mmHg(ASTM F1862)Meet ASTM F1862-07Similar
Bacterial FiltrationEfficiencyPass at 99.56%~99.99%(ASTM F2101)Meet ASTM F2101-07Similar
Particulate FiltrationEfficiencyPass at 98.07%~99.24%(ASTM F2299)Meet ASTM F2299-03Similar
Differential Pressure(Delta-P)Pass at (3.4~3.9) mmH2O/cm2(EN 14683)Meet MIL-M-36954CDifferent(Issue 3)
FlammabilityPass at Class 1(16 CFR 1610)Class 1Same
BiocompatibilityNon-cytotoxic, non-sensitizing,non irritatingNon-cytotoxic, non-sensitizer,non-irritantSame

lssue 1 and Issue 2: The differences in the materials do not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials.

lssue 3: The Differential Pressure test of the proposed device was conducted in accordance with the method specified in EN 14683:2019, Annex C that is the test method specified in ASTM F2100-19. ASTM F2100-19 is the recognized consensus standard [Rec# 6-425] which is the standard specification for performance of materials used in medical face masks. And the Differential Pressure performance of the proposed device meets the requirement of ASTM F2100-19.

7. Non-clinical Testing

Non-clinical testing was conducted to verify that the proposed device met all design specifications as

{6}------------------------------------------------

similar to the predicate device. The tests were conducted according to the following standards, and the results demonstrated that the proposed device complies with the following standards:

  • A ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a riskmanagement process
  • A ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for in Vitro cytotoxicity
  • A ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
  • A ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
  • A ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
  • ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus

  • A EN 14683:2019+AC:2019 Medical Face Masks - Requirements and Test Methods
  • A ASTM F2299/F2299M-03(R2017) Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
  • A 16 CFR 1610 Standard for the Flammability of Clothing Textiles

Table 2 Performance Testing

TestPurposeAcceptance CriteriaResults(Statistics ofthree lots)Verdict
FluidResistance(ASTM F1862)Verify the fluid resistance ofthe proposed device cansimultaneously meet therequirements for Level 1, Level2 and Level 3 specified inASTM F2100-19.Level 1: 29 out of 32pass at 80 mmHg32 out of 32 passat 80 mmHgPass
Level 2: 29 out of 32pass at 120 mmHg32 out of 32 passat 120 mmHgPass
Level 3: 29 out of 32pass at 160 mmHg32 out of 32 passat 160 mmHgPass
Bacterialfiltrationefficiency (BFE)(ASTM F2101)Verify that bacterial filtrationefficiency of the proposeddevice can simultaneouslymeet the requirements forLevel 1, Level 2 and Level 3specified in ASTM F2100-19.Level 1: ≥95%99.56%~99.99%Pass
Level 2: ≥98%Pass
Level 3: ≥98%Pass
Particulatefiltrationefficiency (PFE)(ASTM F2299)Verify that particulate filtrationefficiency of the proposeddevice can simultaneouslymeet the requirements forLevel 1, Level 2 and Level 3specified in ASTM F2100-19.Level 1: ≥95%98.07%~99.24%Pass
Level 2: ≥98%Pass
Level 3: ≥98%Pass
Differentialpressure (Delta-Verify that differential pressureof the proposed device canLevel 1: < 5.0mmH2O/cm²(3.4~3.9)mmH2O/cm²Pass

{7}------------------------------------------------

Version: A/0

P)(EN 14683)simultaneously meet therequirements for Level 1, Level2 and Level 3 specified inASTM F2100-19.Level 2: < 6.0mmH₂O/cm²Pass
Level 3: < 6.0mmH₂O/cm²Pass
Flammability(16 CFR 1610)Verify that Flammability of theproposed device cansimultaneously meet therequirements for Level 1, Level2 and Level 3 specified inASTM F2100-19.Level 1: Class 1Pass
Level 2: Class 1Class 1Pass
Level 3: Class 1Pass

Table 3 Biocompatibility Testing

TestPurposeAcceptance CriteriaResult
In vitroCytotoxicity(ISO 10993-5)Verify that the proposed deviceextract is non-cytotoxic.The extract is non-cytotoxic underthe research conditions.Pass
Skin Irritation(ISO 10993-10)Verify that the proposed deviceextract is non-irritating.The polar and non-polar extractsare non-irritating under theresearch conditions.Pass
Skin Sensitization(ISO 10993-10)Verify that the proposed deviceextract is non-sensitizing.The polar and non-polar extractsare non-sensitizing under theresearch conditions.Pass

8. Clinical Testing

No clinical study is included in this submission.

9. Conclusions

Indications for Use, materials, design feature, specifications and technological characteristics for the proposed device are similar to the predicate device (K133070). The non-clinical performance testing demonstrates that the proposed device is as safe and effective as the legally marketed predicate device (K133070). Therefore, the results show that it is Substantially Equivalent (SE) between the proposed device and the predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.