K Number
K202544

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-01-07

(127 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.

Device Description

The Instylla Delivery Kit is comprised of two delivery syringes, a syringe holder, and a plunger clip (link).

AI/ML Overview

The provided document describes the Instylla Delivery Kit (K202544), which is a medical device intended for the administration of radiographic contrast media, saline, and other aqueous solutions. The submission is a 510(k) premarket notification, aiming to demonstrate substantial equivalence to a predicate device (K191659). This particular submission focuses on a modified version of the Instylla Delivery Kit with 3mL syringes, whereas the predicate device used 1mL syringes.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of specific, quantitative acceptance criteria. Instead, it states that "The Instylla Delivery Kit with 3mL syringes met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device."

Here's a summary of the performance tests conducted and the general outcome:

Performance TestReported Device Performance
Visual InspectionMet predetermined acceptance criteria
Dimensional VerificationMet predetermined acceptance criteria
Functional Testing: Syringe Holder and Plunger Clip Disconnection ForceMet predetermined acceptance criteria
Functional Testing: Syringe CompatibilityMet predetermined acceptance criteria
Functional Testing: Syringe Glide Force and Break ForceMet predetermined acceptance criteria
Functional Testing: Fluids CompatibilityMet predetermined acceptance criteria
Functional Testing: Delivery IntegrityMet predetermined acceptance criteria
Particulate Matter Testing per USP <788>Met predetermined acceptance criteria
BiocompatibilityDeemed acceptable by referencing predicate device results due to identical materials and contact duration.
Sterilization (Ethylene Oxide)Validated to a Sterility Assurance Level (SAL) of 10^-6, with acceptable EO and ECH levels.
Bacterial Endotoxin Test (BET) / LAL testValidated to establish endotoxin level <20 EU/device.
Shelf LifeDemonstrated device performance and sterile barrier maintenance for 6 months.

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes used for the bench and functional testing. It mentions "Performance testing of the final, sterilized Instylla Delivery Kit with 3mL syringes included bench testing and functional testing." Without specific numbers, the sample size remains unknown.

Data provenance: All testing appears to be retrospective bench and functional testing conducted by the manufacturer, Instylla, Inc. There is no information provided regarding the country of origin of the data beyond the company's location (Bedford, Massachusetts).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the document describes a 510(k) submission for a piston syringe device, not an AI or diagnostic imaging device that typically requires expert-established ground truth for performance evaluation. The "ground truth" here is defined by engineering specifications, material properties, and regulatory standards.

4. Adjudication method for the test set

This information is not applicable for the same reasons as #3. Adjudication methods like 2+1 or 3+1 are used for reconciling discrepancies in expert interpretations, which is not relevant for the type of testing described here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This type of study is relevant for AI-powered diagnostic devices, which is not the subject of this 510(k) submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

There was no standalone (algorithm-only) performance study conducted. This device is a physical medical device (syringe kit), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the performance specifications (e.g., disconnection force, glide force, particulate matter), the "ground truth" is based on predetermined engineering specifications, validated test methods, and compliance with recognized industry standards like ISO 7886-1, ISO 594-2, USP <788>, ISO 11135, and ISO 10993-7. For biocompatibility, the ground truth was established by prior testing of the predicate device against ISO 10993-1.

8. The sample size for the training set

This information is not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 7, 2021

Instylla, Inc. Jennifer Greer Regulatory Affairs Manager 201 Burlington Rd, North Building Bedford, Massachusetts 01730

Re: K202544

Trade/Device Name: Instylla Delivery Kit Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: December 7, 2020 Received: December 8, 2020

Dear Jennifer Greer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rumi Young, M.S., RAC. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202544

Device Name Instylla Delivery Kit

Indications for Use (Describe)

The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "instylla" in black font. There is a red dot above the "i" and a red curved line that goes over the "t" and under the "lla". The background is white.

510(k) Summary

Submitter Information:

Instylla, Inc. 201 Burlington Rd North Building Bedford, MA 01730

Contact Person:

Jennifer Greer Regulatory Affairs Manager Phone: 781-622-9293 E-mail: jennyg@instylla.com

Date Prepared:

December 7, 2020

Subject Device:

Proprietary Name:Instylla Delivery Kit
Common Name:Syringe, Piston
Classification Name:Piston syringe(21 CFR 880.5860, Product Code FMF)
Device Classification:Class II
Classification Panel:General Hospital

Predicate Devices:

Proprietary Name:Instylla Delivery Kit
Manufacturer:Instylla, Inc.
510(k) Number:K191659

Device Description:

The Instylla Delivery Kit is comprised of two delivery syringes, a syringe holder, and a plunger clip (link).

{4}------------------------------------------------

Indications for Use:

The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.

The indications for use statement for the modified Instylla Delivery Kit with 3mL syringes remain unchanged from the cleared predicate device.

Comparison of Technological Characteristics to the Predicate Devices:

The Instylla Delivery Kit with 3mL syringes is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate device. The below table summarizes the similarities in design and configuration of the Instylla Delivery Kit with 3mL syringes compared with the predicate devices.

AttributeSubject Device:Instylla Delivery Kit with 3mLSyringesPredicate Device:Instylla Delivery Kit(K191659)SubstantialEquivalenceComparison
Indications for UseThe Instylla Delivery Kit isintended to be used for the intra-arterial or intra-venousadministration of radiographiccontrast media, saline and otheraqueous solutions.The Instylla Delivery Kit isintended to be used forthe intra-arterial or intra-venous administration ofradiographic contrastmedia, saline and otheraqueous solutions.Same
Basic Design/ComponentsTwo standard delivery pistonsyringesTwo standard deliverypiston syringesSame
ComponentsTwo syringes, syringe holder,plunger clipTwo syringes, syringeholder, plunger clipSame
MaterialPolycarbonate, silicone, ABSPolycarbonate, silicone,ABSSame
Principle ofOperationManually operated by advancingthe dual syringe plungerssimultaneously with the aid of aplunger clip.Manually operated byadvancing the dualsyringe plungerssimultaneously with theaid of a plunger clip.Same
Operation Volume3mL1mLThe operationalvolume of thesyringes differs.

{5}------------------------------------------------

AttributeSubject Device:Instylla Delivery Kit with 3mLSyringesPredicate Device:Instylla Delivery Kit(K191659)SubstantialEquivalenceComparison
GradationPrinted with accurate graduationlines that are compliant with ISO7886-1.Printed with accurategraduation lines that arecompliant with ISO 7886-1.Same
Method ofSterilizationEthylene Oxide (EO) to a SAL of10-6Ethylene Oxide (EO) to aSAL of 10-6Same
Shelf Life6 months6 monthsSame

The modified Instylla Delivery Kit differs from the legally marketed predicate device in that it includes two 3mL syringes instead of two 1mL syringes. The intended use, indications, principle of operation, and technological characteristics remain identical between the subject and predicate device.

Performance Data

Performance testing of the final, sterilized Instylla Delivery Kit with 3mL syringes included bench testing and functional testing to verify specifications fundamental to the design of the device. The following tests were conducted:

  • -Visual Inspection
  • Dimensional Verification .
  • -Functional Testing
    • o Syringe Holder and Plunger Clip Disconnection Force
    • Syringe Compatibility o
    • Syringe Glide Force and Break Force o
    • Fluids Compatibility o
    • o Delivery Integrity
  • Particulate Matter Testing per USP <788> Particulate Matter in Injections -

The Instylla Delivery Kit with 3mL syringes met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing. Further, the syringes of the subject device are complaint against FDA recognized standards ISO 7886-1 and ISO 594-2.

Biocompatibility Testing

A biocompatibility evaluation was conducted on the Instylla Delivery Kit with 3mL syringes in accordance with the FDA Guidance Document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process." The Instylla Delivery Kit with 3mL syringes is classified as an external

{6}------------------------------------------------

communicating device, indirectly contacting blood for a limited duration (≤24 hours). The patient contact, materials and manufacturing processes remain identical between the Instylla Delivery Kit with 3mL syringes and the legally marketed predicate device. Therefore, per ISO 10993-1, the biocompatibility results of the predicate device are representative of the subject device and no additional biocompatibility testing was conducted on the subject device.

Sterility

The Instylla Delivery Kit with 3mL syringes is sterilized via a validated ethylene oxide (EO) process to a Sterility Assurance Level (SAL) of 10°. The sterilization process was validated per ISO 11135 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices. The Instylla Delivery Kit with 3mL syringes was evaluated and adopted into the validated process per AAMI TIR28 Product adoption and process equivalence for ethylene oxide sterilization. The EO and ECH levels were determined to be acceptable in accordance with ISO 10993-7 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.

A bacterial endotoxin test (BET), also known as the Limulus amebocyte lysate (LAL) test, was also validated to establish that the delivery kit endotoxin level will be <20 endotoxin units (EU)/device.

Shelf Life

The Instylla Delivery Kit with 3mL syringes has a 6 month shelf life. Shelf life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life.

Clinical Performance Data

The technological characteristics, indications for use, material, manufacturing and sterilization processes are the same as the predicate devices and therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.

Conclusion

Instylla has demonstrated that the Instylla Delivery Kit with 3mL syringes is substantially equivalent in fundamental design, function, device materials, packaging, sterilization, operating principle, intended use/indication for use and technology as the legally marketed predicate device, Instylla Delivery Kit, which was cleared under 510(k) Premarket Notification K191659 on October 11, 2019.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).