K Number
K202525

Validate with FDA (Live)

Date Cleared
2022-05-19

(625 days)

Product Code
Regulation Number
862.1215
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alinity i STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnI) in human plasma (lithium heparin) on the Alinity i system.

The Alinity i STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

Device Description

The Alinity i STAT High Sensitivity Troponin-I Reagent Kit contains:

  • . Microparticles: 1 bottle (6.6 mL per 100 test cartridge / 33.8 mL per 600 test cartridge). Anti-troponin I (mouse, monoclonal) coated microparticles in TRIS buffer with protein (bovine) stabilizer. Minimum concentration: 0.035% solids. Preservative: ProClin 300.
  • . Conjugate: 1 bottle (6.1 mL per 100 test cartridge / 33.8 mL per 600 test cartridge). Anti-troponin I (mouse-human chimeric, monoclonal) acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer and human IgG. Minimum concentration: 0.1 mg/L. Preservative: ProClin 300.

The Alinity i STAT High Sensitivity Troponin-I assay is an automated, two-step immunoassay for the quantitative determination of cTnI in human plasma (lithium heparin) using CMIA technology.

Sample and anti-troponin I antibody-coated paramagnetic microparticles are combined and incubated. The cTnI present in the sample binds to the anti-troponin I coated microparticles. The mixture is washed. Anti-troponin I acridinium-labeled conjugate is added to create a reaction mixture and incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are added. The resulting chemiluminescent reaction is measured as a relative light unit (RLU). There is a direct relationship between the amount of cTnI in the sample and the RLU detected by the system optics.

AI/ML Overview

Acceptance Criteria and Study for Alinity i STAT High Sensitivity Troponin-I

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state formal acceptance criteria in a tabular format for the clinical performance. However, typical performance metrics for diagnostic assays like this include:

MetricAcceptance Criteria (Implicit/Assumed)Reported Device Performance (Alinity i STAT High Sensitivity Troponin-I)Notes
Precision (Reproducibility)Generally low %CV across different concentrations. Specific thresholds are not provided in the document.Reproducibility (Overall %CV): Ranges from 2.7% to 12.7% across different concentrations (3.5 ng/L to 2871.4 ng/L). Highest %CV seen at lower concentrations.Good precision observed, especially at higher concentrations. The higher %CV at lower concentrations is typical for immunoassays near their detection limits.
Within-Laboratory Precision (%CV)Generally low %CV. Specific thresholds are not provided.Within-Lab %CV: Low Control: 4.1%; Medium Control: 3.6%; Bio-Rad Level 2: 4.2%.Consistent and good precision over 20 days.
Lower Limits of Measurement (LoD, LoQ)Levels should be adequately low for early detection of MI. Often compared to established guidelines.LoB: 0.0 ng/L; LoD: 0.9 ng/L; LoQ: 2.7 ng/L.These values demonstrate the assay's ability to detect very low levels of troponin I, which is crucial for high-sensitivity assays in MI diagnosis.
Linearity (Analytical Measuring Interval)Wide range that covers clinically relevant concentrations.2.7 to 3600.0 ng/L (pg/mL).A broad linear range ensures accurate measurements across a wide spectrum of troponin concentrations encountered in clinical practice.
Analytical Specificity (Interference)Interference within ±10% for common substances/drugs. Identified interferences should be noted.Endogenous: Bilirubin, Hemoglobin, Intralipid: no significant interference (within ±10%). Total Protein > 8.8 g/dL showed interference (up to -16.3%).Drugs: Most tested drugs showed no significant interference (within ±10%). Fibrinogen at 1000 mg/dL showed 24.3% interference at 15 ng/L.Other Conditions: HAMA > 150 ng/mL showed up to -11.0% interference. RF > 1200 IU/mL showed up to -18.9% interference.Most common interferents are within acceptable limits. Identified interferences (high total protein, fibrinogen, HAMA, RF) are noted, prompting caution in interpretation for affected patients.
Cross-Reactants≤ 1% cross-reactivity for related cardiac/muscle proteins.≤ 1% for Actin, Cardiac troponin C, Cardiac troponin T, CK-MB, Myoglobin, Myosin, Skeletal troponin I, Tropomyosin.Excellent specificity for cardiac troponin I, minimizing false positives from other muscle proteins.
Diagnostic Accuracy (Sensitivity, Specificity, PPV, NPV) for MI diagnosis (at various time points relative to ED presentation)These values should align with clinical needs for an "aid in diagnosis of MI" especially regarding high NPV for ruling out MI and acceptable sensitivity/specificity. Generally, high sensitivity and NPV are critical for MI rule-out. The document presents ranges and 95% CIs. Specific numerical acceptance criteria are not explicitly stated, but the performance is presented to demonstrate clinical utility.Sex-specific cutoffs (Female 14 ng/L, Male 35 ng/L):- Sensitivity: 85.29% to 98.46% (Female), 72.20% to 92.06% (Male) depending on time point.- Specificity: 69.13% to 85.64% (Female), 74.22% to 88.98% (Male).- PPV: 27.02% to 42.74% (Female), 33.78% to 50.00% (Male).- NPV: 98.80% to 99.85% (Female), 97.05% to 99.19% (Male).Overall cutoff (27 ng/L):- Sensitivity: 78.43% to 94.44% (Female), 76.45% to 94.44% (Male)- Specificity: 66.44% to 92.57% (Female), 66.44% to 88.34% (Male).- PPV: 35.09% to 47.66% (Female), 29.06% to 44.07% (Male).- NPV: 98.37% to 99.46% (Female), 97.40% to 99.35% (Male).The high NPV across all time points and cutoffs is a strong indicator of the assay's ability to rule out MI. Sensitivities are also generally high. The lower PPV reflects the prevalence of non-MI conditions that can elevate troponin and emphasizes the need for combining results with clinical data. The study notes that lower specificity at ≥6 hours is due to study design and patient flow in the ED.
AUCNot explicitly stated as an acceptance criterion, but higher values indicate better diagnostic performance. Generally, AUC > 0.9 is considered excellent.AUC: 0.9257 to 0.9777 (Female), 0.8994 to 0.9489 (Male) across different time points.Consistently high AUC values, indicating excellent overall diagnostic accuracy for both sexes and at different time points.

Study Proving Acceptance Criteria:

The study described is a multi-center prospective clinical study designed to assess the diagnostic accuracy of the Alinity i STAT High Sensitivity Troponin-I assay in patients presenting with chest discomfort or equivalent ischemic symptoms.

2. Sample size used for the test set and the data provenance:

  • Sample Size (Clinical Study):
    • Total Subjects: 6174
    • MI Subjects: 432 (124 female, 308 male)
    • Non-MI Subjects: 5742 (2128 female, 3614 male)
    • Total Specimens (serial sampling):
      • 891 from MI subjects
      • 8975 from non-MI subjects
  • Data Provenance:
    • Country of Origin: United States (implied by the FDA 510(k) submission context for a US population reference range study). The study was conducted at 23 Emergency Departments (EDs) in the US, reflecting regional, urban, and rural patient populations.
    • Retrospective or Prospective: Prospective. Specimens were collected prospectively from subjects presenting to the ED.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: A "panel of board-certified cardiologists" was used. The exact number is not explicitly stated, but "panel" implies more than one.
  • Qualifications of Experts: Board-certified cardiologists.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Adjudication Method: The subject diagnoses (MI or non-MI) were adjudicated by a panel of board-certified cardiologists based on the third universal definition of MI. The adjudicators were blinded to the Alinity i STAT High Sensitivity Troponin-I assay results. This indicates an expert consensus method, likely involving all panel members reaching a decision, but the specific voting scheme (e.g., 2+1, 3+1) is not detailed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a MRMC comparative effectiveness study was not done. This document describes the performance of a lab assay (IVD - In Vitro Diagnostic), not an AI-assisted diagnostic tool that would typically involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance study was done. The entire clinical study described assesses the performance of the Alinity i STAT High Sensitivity Troponin-I assay (the "algorithm" or device in this context) as a standalone diagnostic tool for aid in MI diagnosis. The results (sensitivity, specificity, PPV, NPV, AUC) are presented for the direct output of the assay. The wording "aid in the diagnosis" implies it's used in conjunction with other clinical information, but its individual performance is what's being evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Expert Consensus and Clinical Definition: The ground truth for MI diagnosis was established by a panel of board-certified cardiologists based on the third universal definition of MI. This definition incorporates clinical observations, ECG, and other diagnostic information, which can include pathology and outcomes data indirectly but is primarily an expert consensus application of a standardized clinical definition.

8. The sample size for the training set:

  • Not Applicable / Not Explicitly Stated for the Clinical Study. This document describes an IVD device, not an AI/machine learning model in the traditional sense that requires an explicit "training set" of patient data for model development. The "training" for such an assay involves the optimization of reagents, antibodies, and detection protocols during product development. The non-clinical studies (precision, linearity, interference) represent the internal testing and validation that inform the assay's operational parameters. The clinical study described functions as a validation/test set for the final device performance.

9. How the ground truth for the training set was established:

  • Not Applicable. As noted above, for this IVD device, there isn't a "training set" of patient data with ground truth in the way there is for AI/ML algorithms. The ground truth for the assay's development and calibration would typically be established using manufactured controls, calibrators, and characterized reference materials with known concentrations of cTnI, following industry standards and guidelines (e.g., CLSI documents referenced in the non-clinical section).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 19, 2022

Abbott Laboratories Diagnostic Division Judi Wallach Regulatory Affairs Project Manager Dept. 09AA, Bldg. Ap8, 100 Abbott Park Road Abbott Park, Illinois 60064-6038

Re: K202525

Trade/Device Name: Alinity i STAT High Sensitivity Troponin-I Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine Phosphokinase/Creatine Kinase Or Isoenzymes Test System Regulatory Class: Class II Product Code: MMI Dated: February 1, 2022 Received: February 2, 2022

Dear Judi Wallach:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202525

Device Name

Alinity i STAT High Sensitivity Troponin-I

Indications for Use (Describe)

The Alinity i STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnI) in human plasma (lithium heparin) on the Alinity i system.

The Alinity i STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Federal Food, Drug, and Cosmetic Act and 21 CFR 807.92.

I. 510(k) Number

K202525

II. Applicant Name

Date summary prepared: May 13, 2022

Abbott Laboratories Diagnostics Division Dept. 9AA, CP01-2 100 Abbott Park Road Abbott Park, IL 60064

Primary contact person for all communications:

Judi Wallach, ADD, Regulatory Affairs Project Manager Telephone Number: (224) 667-1132 Fax Number: (224) 667-4836

Secondary contact person for all communications:

Julian Braz, ADD, Director, Regulatory Affairs Telephone Number: (224) 330-9230

III. Device Name

Alinity i STAT High Sensitivity Troponin-I

Reagents

Trade Name: Alinity i STAT High Sensitivity Troponin-I Device Classification: Class II Classification Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Governing Regulation: 862.1215

Code: MMI

{4}------------------------------------------------

IV. Predicate Device

ARCHITECT STAT High Sensitivity Troponin-I (K191595)

V. Intended Use of the Device

The Alinity i STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnI) in human plasma (lithium heparin) on the Alinity i system.

The Alinity i STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

VI. Description of Device

The Alinity i STAT High Sensitivity Troponin-I Reagent Kit contains:

  • . Microparticles: 1 bottle (6.6 mL per 100 test cartridge / 33.8 mL per 600 test cartridge). Anti-troponin I (mouse, monoclonal) coated microparticles in TRIS buffer with protein (bovine) stabilizer. Minimum concentration: 0.035% solids. Preservative: ProClin 300.
  • . Conjugate: 1 bottle (6.1 mL per 100 test cartridge / 33.8 mL per 600 test cartridge). Anti-troponin I (mouse-human chimeric, monoclonal) acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer and human IgG. Minimum concentration: 0.1 mg/L. Preservative: ProClin 300.

Principles of the Procedure

The Alinity i STAT High Sensitivity Troponin-I assay is an automated, two-step immunoassay for the quantitative determination of cTnI in human plasma (lithium heparin) using CMIA technology.

Sample and anti-troponin I antibody-coated paramagnetic microparticles are combined and incubated. The cTnI present in the sample binds to the anti-troponin I coated microparticles. The mixture is washed. Anti-troponin I acridinium-labeled conjugate is added to create a reaction mixture and incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are added.

{5}------------------------------------------------

The resulting chemiluminescent reaction is measured as a relative light unit (RLU). There is a direct relationship between the amount of cTnI in the sample and the RLU detected by the system optics.

VII. Comparison of Technological Characteristics

The Alinity i STAT High Sensitivity Troponin-I assay (subject device) utilizes a CMIA methodology for the quantitative in vitro determination of cTnI and is intended for use on the Alinity i system.

The similarities and differences between the subject device and the predicate device are presented in the following table.

Subject DevicePredicate Device
DescriptionAlinity i STAT High SensitivityTroponin-IARCHITECT STAT High SensitivityTroponin-I (K191595)
General Device Characteristic Similarities
Intended Use /Indications for UseThe Alinity i STAT High SensitivityTroponin-I assay is a CMIA used for thequantitative determination of cTnI inhuman plasma (lithium heparin) on theAlinity i system.The Alinity i STAT High SensitivityTroponin-I assay is to be used as an aid inthe diagnosis of MI.The ARCHITECT STAT HighSensitivity Troponin-I assay is a CMIAused for the quantitative determinationof cTnI in human plasma (dipotassium[K2] EDTA) on the ARCHITECT2000SR System.The ARCHITECT STAT HighSensitivity Troponin-I assay is to beused as an aid in the diagnosis of MI.
Specific AnalyteDetectedcTnISame
General Device Characteristic Differences
Specimen TypePlasma (lithium heparin)Plasma (dipotassium EDTA)
99th PercentileCutoff / ExpectedValues fromApparently HealthyIndividuals(ng/L, pg/mL)Female: 14Male: 35Overall: 27Female: 17Male: 35Overall: 28

{6}------------------------------------------------

VIII. Performance Summary*

A. Nonclinical

1. Precision

Reproducibility

A study was performed using 1 lot of the Alinity i STAT High Sensitivity Troponin-I reagents, 1 lot of the Alinity i STAT High Sensitivity Troponin-I Calibrators, and 1 lot of the Alinity i STAT High Sensitivity Troponin-I Controls. The study was performed to include lithium heparin separator plasma specimens within each of 5 target concentration ranges (> 3 to 6 ng/L, 10 to 20 ng/L, 30 to 50 ng/L, 100 to 300 ng/L, and 1000 ng/L to near the upper limit of the analytical measuring interval [AMI]). Only one specimen per concentration range was collected in a single day. The study was performed over a minimum of 3 days. Each specimen was stored at room temperature and tested in duplicate, twice in one day, on each of 3 instruments (for a total of 12 replicates) within 8 hours of collection.

SamplenMean(ng/L, pg/mL)Within-RunSDWithin-Run%CVBetween-RunSDBetween-Run%CVWithin-LaboratoryaSDWithin-Laboratorya%CVReproducibilitybSDReproducibilityb%CV
1123.50.226.20.000.00.226.20.308.7
2125.20.305.70.000.00.305.70.6712.7
3128.00.253.20.232.90.344.30.678.4
41213.60.372.70.000.00.372.70.614.5
51213.90.372.70.000.00.372.70.372.7
61215.10.654.30.000.00.654.30.654.3
71218.20.824.50.000.00.824.50.824.5
81220.20.663.30.000.00.663.30.743.7
91236.90.732.00.000.00.732.01.173.2
101246.70.901.90.000.00.901.90.952.0
111251.61.222.41.072.11.633.21.633.2
1212172.64.512.65.533.27.144.17.144.1
1312233.05.002.19.183.910.454.510.454.5
  • Unless otherwise specified, all studies were performed on the Alinity i system.

{7}------------------------------------------------

SamplenMean(ng/L,pg/mL)Within-RunBetween-RunWithin-LaboratoryaReproducibilityb
1412300.011.003.75.611.912.354.112.354.1
15121228.329.032.445.233.753.744.453.744.4
16121974.649.462.557.972.976.213.976.213.9
17122871.472.862.585.313.0112.193.9112.193.9

ª Includes within-run and between-run variability.

b Includes within-run, between-run, and between-instrument variability.

Within-Laboratory Precision

A study was performed based on guidance from CLSI EP05-A3. * Testing was conducted using 3 lots of the Alinity i STAT High Sensitivity Troponin-I reagents, 2 lots of the Alinity i STAT High Sensitivity Troponin-I Calibrators, 1 lot of the Alinity i STAT High Sensitivity Troponin-I Controls, 1 lot of Bio-Rad Liquichek Cardiac Markers Plus Control LT (Level 2), and 2 instruments. Three controls were tested in duplicate, twice per day on 20 days (following the manufacturers' storage and handling requirements) on 6 reagent lot/calibrator lot/instrument combinations. The performance from a representative combination is shown in the following table.

Note: Patient samples can only be stored for 8 hours at room temperature; therefore, 20-day precision was conducted with quality controls.

Within-Run(Repeatability)Within-Laboratorya
SamplenMean(ng/L, pg/mL)SD%CVSD(Rangeb)%CV(Rangeb)
Low Control8019.90.733.70.81(0.76-1.13)4.1(3.9-5.1)
Medium Control80199.86.143.17.25(5.54-7.25)3.6(2.8-3.6)
Bio-Rad Level 2801442.745.913.260.02(49.50-69.39)4.2(3.3-4.5)

ª Includes within-run, between-run, and between-day variability.

* Clinical and Laboratory Standards Institute (CLSI). Evaluation of Quantitative Measurement Procedures: Approved Guideline-Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.

{8}------------------------------------------------

2. Lower Limits of Measurement

A study was performed based on guidance from CLSI EP17-A2. * Testing was conducted using 2 lots of the Alinity i STAT High Sensitivity Troponin-I reagents on each of 2 instruments (1 instrument for limit of blank [LoB]). The maximum observed LoB, limit of detection (LoD), and limit of quantitation (LoQ) values are summarized below. These representative data support the lower limit of the analytical measuring interval.

ng/L (pg/mL)
LoBa0.0
LoDb0.9
LoQc2.7

a The LoB represents the 95th percentile from n ≥ 60 replicates of zero-analyte samples.

b The LoD represents the lowest concentration at which the analyte can be detected with 95% probability based on n ≥ 60 replicates of low-analyte level samples.

C The LoQ is defined as the lowest concentration at which a maximum allowable precision of 20 %CV was met and was determined from n ≥ 60 replicates of low-analyte level samples.

3. Linearitv

A study was performed based on guidance from CLSI EP06-A. This assay is linear across the analytical measuring interval of 2.7 to 3600.0 ng/L (pg/mL).

" Clinical and Laboratory Standards Institute (CLSI). Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. CLSI Document EP17-A2. Wayne, PA: CLSI; 2012.

* Clinical and Laboratory Standards Institute (CLSI). Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI Document EP06-A. Wayne, PA: CLSI; 2003.

{9}------------------------------------------------

4. Analytical Specificity

a. Interference

Potentially Interfering Endogenous Substances

A study was performed based on guidance from CLSI EP07, 3rd ed. * Each substance was tested at 2 levels of the analyte (approximately 15 ng/L and 500 ng/L for bilirubin, total protein, and Intralipid and 15 ng/L and 200 ng/L for hemoglobin).

No significant interference (interference within ± 10%) was observed at the following concentrations.

Potentially Interfering SubstanceInterferent Level
Bilirubin (conjugated)40 mg/dL
Bilirubin (unconjugated)40 mg/dL
Hemoglobin1000 mg/dL
Total protein8.8 g/dL
Intralipid3000 mg/dL

Interference beyond ± 10% was observed at the concentration shown below for the following substance.

Potentially InterferingSubstanceInterferentLevelAnalyteLevel% Interference
Total Protein9.0 g/dL500 ng/L-10.9%

The Alinity i STAT High Sensitivity Troponin-I assay is susceptible to interference effects from total protein > 8.8 g/dL. Total protein from 9.0 to 12.0 g/dL decreased troponin values at 500 ng/L by up to -16.3%.

* Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry. 3rd ed. CLSI Guideline EP07. Wayne, PA: CLSI; 2018.

{10}------------------------------------------------

Potentially Interfering Drugs

A study was performed based on guidance from CLSI EP07-A2* and EP07, 3rd ed. Each drug was tested at 2 levels of the analyte (approximately 15 ng/L and 500 ng/L).

No significant interference (interference within ± 10%) was observed at the following concentrations.

Potentially InterferingDrugInterferentLevelPotentially InterferingDrugInterferentLevel
Acetaminophen250 µg/mLIbuprofen500 µg/mL
Acetylsalicylic Acid1000 µg/mLLevodopa20 µg/mL
Adrenaline0.37 µg/mLLow MW Heparin5 U/mL
Allopurinol400 µg/mLMethyldopa25 µg/mL
Ambroxol400 µg/mLMethylprednisolone80 µg/mL
Ampicillin1000 µg/mLMetronidazole200 µg/mL
Ascorbic Acid300 µg/mLNicotine2 mg/dL
Atenolol10 µg/mLNifedipine60 µg/mL
Biotin4250 ng/mLNitrofurantoin64 µg/mL
Bivalirudin42 µg/mLNystatin7.5 µg/mL
Caffeine100 µg/mLOxytetracycline5 µg/mL
Captopril50 µg/mLPhenylbutazone400 µg/mL
Carvedilol150 µg/mLPhenytoin100 µg/mL
Cefoxitin2500 µg/mLPrimidone10 mg/dL
Cinnarizine400 µg/mLPropranolol5 µg/mL
Clopidogrel75 µg/mLQuinidine20 µg/mL
Cocaine10 µg/mLRifampicin60 µg/mL
Cyclosporine5 µg/mLSalicylic Acid600 µg/mL
Diclofenac50 µg/mLSimvastatin20 µg/mL
Digoxin7.5 µg/mLSodium Heparin8 U/mL
Dopamine900 µg/mLStreptokinase31.3 U/mL
Doxycycline50 µg/mLTheophylline75 µg/mL
Entifibatide7 µg/mLTPA2.3 µg/mL

* Clinical and Laboratory Standards Institute (CLS). Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition. CLSI Document EP07-A2. Wayne, PA: CLSI; 2005.

{11}------------------------------------------------

Potentially InterferingDrugInterferentLevelPotentially InterferingDrugInterferentLevel
Erythromycin200 µg/mLTrimethoprim75 µg/mL
Fondaparinux4 µg/mLVerapamil160 µg/mL
Furosemide400 µg/mLWarfarin30 µg/mL

MW = Molecular weight

TPA = Tissue plasminogen activator

Interference beyond ± 10% was observed at the concentration shown

below for the following drug.

Potentially InterferingDrugInterferentLevelAnalyteLevel% Interference
Fibrinogen1000 mg/dL15 ng/L24.3%

Specimens from individuals with elevated levels of fibrinogen may demonstrate falsely elevated values.

Potentially Interfering Other Conditions

Specimens containing human anti-mouse antibodies (HAMA) were evaluated at 3 levels of the analyte (approximately 2.7, 15, and 500 ng/L) with the Alinity i STAT High Sensitivity Troponin-I assay. The individual differences for samples near 2.7 ng/L ranged from -0.2 to 0.1 ng/L. The individual % differences for samples near 15 ng/L and 500 ng/L ranged from -8.9% to -0.5%. No significant interference (interference within ± 10%) was observed with specimens containing HAMA at concentrations up to 150 ng/mL.

The Alinity i STAT High Sensitivity Troponin-I assay is susceptible to interference effects from HAMA > 150 ng/mL. HAMA at 225 ng/mL decreased troponin values up to -11.0%.

Samples at 3 levels of the analyte (approximately 2.7, 15, and 500 ng/L) were spiked with rheumatoid factor (RF) to concentrations of 1200 IU/mL

{12}------------------------------------------------

and 1495 IU/mL and tested with the Alinity i STAT High Sensitivity Troponin-I assay. For samples tested at an RF concentration of 1200 IU/mL, the difference for a sample near 2.7 ng/L was 0.3 ng/L, and the % differences for samples near 15 ng/L and 500 ng/L were -7.1% and -7.6%, respectively. For samples tested at an RF concentration of 1495 IU/mL. the difference for a sample near 2.7 ng/L was -0.1 ng/L, and the % differences for samples near 15 ng/L and 500 ng/L were -12.6% and -18.9%, respectively.

The Alinity i STAT High Sensitivity Troponin-I assay is susceptible to interference effects from RF > 1200 IU/mL. RF at 1495 IU/mL decreased troponin values up to -18.9%.

Although the Alinity i STAT High Sensitivity Troponin-I assay is specifically designed to minimize the effects of HAMA, heterophilic antibodies, and RF, assay results may be impacted by these proteins.

Troponin autoantibodies have been reported to be present in approximately 10% to 20% of patients presenting to the emergency department (ED) and may lead to falsely low troponin assay results and delay in treatment of acute coronary syndrome (ACS). ** Therefore, a test result that is inconsistent with the clinical picture and patient history should be interpreted with caution.

Park JY, Jaffe AS. Troponin autoantibodies: from assay interferent to mediator of cardiotoxicity. Clin Chem 2017;63(1):30-32.

* Nussinovitch U, Shoenfeld Y. Anti-troponin autoantibodies and the cardiovascular system. Heart 2010;96:1518-1524.

{13}------------------------------------------------

b. Cross-Reactants

A study was performed based on guidance from CLSI EP07, 3rd ed. Samples with cTnI target concentrations of 2.7, 15, and 500 ng/L containing the cross-reactants at the concentrations listed below were tested with the Alinity i STAT High Sensitivity Troponin-I assay. The observed % cross-reactivity was ≤ 1% for all cross-reactants evaluated at each analyte level.

Cross-ReactantCross-Reactant Concentration
Actin1 000 000 ng/L
Cardiac troponin C1 000 000 ng/L
Cardiac troponin T1 000 000 ng/L
Creatine kinase-MB (CK-MB)1 000 000 ng/L
Myoglobin1 000 000 ng/L
Myosin1 000 000 ng/L
Skeletal troponin I1 000 000 ng/L
Tropomyosin1 000 000 ng/L

5. Expected Values

This study was performed on the ARCHITECT i2000SR System.

A reference range study was conducted based on guidance from CLSI EP28-A3c. Specimens were collected from 1531 apparently healthy individuals in a US population with normal levels of cardiac B-type natriuretic peptide (BNP) and HbA1c, and glomerular filtration rate (GFR) values ≥ 60 mL/min. Each specimen was stored frozen, thawed, and evaluated in replicates of one using the ARCHITECT STAT High Sensitivity Troponin-I assay. The 99th

* Clinical and Laboratory Standards Institute (CLSI). Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline-Third Edition. CLSI document EP28-A3c. Wayne, PA: CLSI; 2010.

{14}------------------------------------------------

ApparentlyHealthy PopulationnAge Range(years)99th Percentile(ng/L, pg/mL)90% CI(ng/L, pg/mL)
Female763a21 - 7514[12, 17]
Male76621 - 7335[27, 44]
Overall153121 - 7527[22, 32]

percentiles described in the following table for this population were determined using the robust statistical method described in CLSI EP28-A3c.

ª During the sex-specific analysis. 2 female subjects were identified as outliers. The subjects and results were excluded from the sex-specific analysis but were included in the overall analysis.

B. Clinical

The Alinity i STAT High Sensitivity Troponin-I results should be used in conjunction with other diagnostic information such as electrocardiogram (ECG), clinical observations and information, and patient symptoms to aid in the diagnosis of MI.

A multi-center prospective study was performed to assess diagnostic accuracy of the Alinity i STAT High Sensitivity Troponin-I assay. Specimens were collected at 23 EDs from 6174 subjects presenting to the ED with chest discomfort or equivalent ischemic symptoms consistent with ACS. The specimen collection sites represented geographically diverse EDs associated with primary care hospitals and medical centers, reflecting regional, urban, and rural patient populations. All subject diagnoses were adjudicated by a panel of board-certified cardiologists based on the third universal definition of MI. The adjudicators were blinded to the Alinity i STAT High Sensitivity Troponin-I assay results. The observed MI prevalence in this study was 7.0%.

  • . 891 specimens with serial sampling from 432 MI subjects (124 female subjects, 308 male subjects)
  • . 8975 specimens with serial sampling from 5742 non-MI subjects (2128 female subjects, 3614 male subjects)

" Thygesen K, Alpert JS, Jaffe AS, et al. Third universal definition of myocardial infarction. Eur Heart J 2012;33(20):2551-2567.

{15}------------------------------------------------

The specimens were collected in lithium heparin or lithium heparin separator tubes. The specimens were evaluated using the Alinity i STAT High Sensitivity Troponin-I assay.

NOTE: The study population did not include type 4 or 5 MI subjects. Therefore, the ability of the assay to identify these patients was not evaluated.

The results were analyzed using the serial sampling time points collected as part of the ED visit.

An analysis for both females and males was performed using the sex-specific 99th percentile cutoffs (female 14 ng/L, male 35 ng/L). The results are summarized in the following table.

Cutoff(ng/L)TimePointanbSensitivityc%95% CISpecificityd%95% CIPPVe%95% CINPVf%95% CI
14(Femaleonly)0 to < 1Hour157485.29(87/102)77.15-90.8884.04(1237/1472)82.08-85.8227.02(87/322)22.46-32.1298.80(1237/1252)98.03-99.27
1 to < 3Hours84193.62(44/47)82.84-97.8185.64(680/794)83.03-87.9127.85(44/158)21.45-35.3099.56(680/683)98.72-99.85
3 to < 6Hours85498.46(64/65)91.79-99.7382.00(647/789)79.17-84.5331.07(64/206)25.14-37.6999.85(647/648)99.13-99.97
≥ 6 Hoursg, h28498.15(53/54)90.23-99.6769.13(159/230)62.89-74.7542.74(53/124)34.38-51.5499.38(159/160)96.55-99.89
35(Maleonly)0 to < 1Hour291772.20(187/259)66.45-77.3088.98(2365/2658)87.73-90.1138.96(187/480)34.70-43.3997.05(2365/2437)96.30-97.65
1 to < 3Hours133989.29(75/84)80.88-94.2688.29(1108/1255)86.39-89.9533.78(75/222)27.88-40.2399.19(1108/1117)98.48-99.58
3 to < 6Hours148090.85(139/153)85.23-94.4784.33(1119/1327)82.27-86.1840.06(139/347)35.04-45.3098.76(1119/1133)97.94-99.26
≥ 6 Hoursg, h57692.06(116/126)86.01-95.6374.22(334/450)69.99-78.0550.00(116/232)43.62-56.3897.09(334/344)94.73-98.41

The results using the overall 99th percentile cutoff (27 ng/L) are summarized in the following table.

{16}------------------------------------------------

SensitivitycSpecificitydPPVeNPVf
SexTime Pointanb%95% CI%95% CI%95% CI%95% CI
Female0 to < 1Hour157478.43(80/102)69.50 -85.3089.95(1324/1472)88.30 -91.3835.09(80/228)29.19 -41.4898.37(1324/1346)97.54 -98.92
1 to < 3Hours84191.49(43/47)80.07 -96.6492.57(735/794)90.53 -94.2042.16(43/102)33.03 -51.8599.46(735/739)98.62 -99.79
3 to < 6Hours85493.85(61/65)85.22 -97.5888.34(697/789)85.91 -90.4039.87(61/153)32.45 -47.7899.43(697/701)98.54 -99.78
≥ 6 Hoursg, h28494.44(51/54)84.89 -98.0975.65(174/230)69.71 -80.7547.66(51/107)38.45 -57.0498.31(174/177)95.14 -99.42
Male0 to < 1Hour291776.45(198/259)70.92 -81.2185.82(2281/2658)84.44 -87.0934.43(198/575)30.67 -38.4197.40(2281/2342)96.67 -97.97
1 to < 3Hours133991.67(77/84)83.78 -95.9085.02(1067/1255)82.94 -86.8929.06(77/265)23.92 -34.7999.35(1067/1074)98.66 -99.68
3 to < 6Hours148093.46(143/153)88.39 -96.4180.71(1071/1327)78.50 -82.7435.84(143/399)31.29 -40.6699.07(1071/1081)98.31 -99.50
≥ 6 Hoursg, h57694.44(119/126)88.98 -97.2866.44(299/450)61.96 -70.6544.07(119/270)38.28 -50.0497.71(299/306)95.35 -98.89

CI = confidence interval

a All time points are relative to ED presentation / ED triage.

b Some time points could not be collected for some subjects.

For footnotes c-f:

Alinity i STAT High SensitivityTroponin-IDiagnosis
MINon-MI
cTnI Value > cutpointAB
cTnI Value ≤ cutpointCD

6 Sensitivity = A/(A + C) × 100%

d Specificity = D/(B + D) × 100%

€ Positive predictive value (PPV) = A/(A + B) × 100%

f Negative predictive value (NPV) = D/(C + D) × 100%

  • 8 The study design followed the standard of care at each site where few specimens would be obtained at later time points because most patients would not typically require further serial cTnI testing after 6 hours. Therefore, the lower specificity at the ≥ 6 hour time point was the result of the disproportionate number of elevated and non-elevated specimens carried over from previous time points. The cTnI value should be used in conjunction with information available from clinical evaluation and other diagnostic procedures.
  • h Of the specimens collected at greater than or equal to 6 hours, all were collected within 19 hours from presentation to the ED, except for one specimen from a male collected at 35 hours.

&quot; Gulati M, Levy PD, Mukherjee D, et al. 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guideline for the evaluation and diagnosis of chest pain: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 2021;144:e368-e454.

{17}------------------------------------------------

Prevalence of adjudicated MI in the study population was 7.0% (5.5% for females and 7.9% for males).

In the Alinity i STAT High Sensitivity Troponin-I clinical study, the percent of false negatives for females using the sex-specific cutoff (14 ng/L) was up to 2.41% lower when compared to the false negative rate for females when using the overall cutoff of 27 ng/L. When using the female cutoff of 14 ng/L, 5.65% of females with MI had non-elevated Alinity i STAT High Sensitivity Troponin-I test results. When using the overall cutoff of 27 ng/L, 8.06% of females with MI had non-elevated Alinity i STAT High Sensitivity Troponin-I test results. Troponin results should always be used in conjunction with clinical data, signs, and symptoms.

In the Alinity i STAT High Sensitivity Troponin-I clinical study, the percent of false negatives for males using the sex-specific cutoff (35 ng/L) was up to 1.63% higher when compared to the false negative rate for males when using the overall cutoff of 27 ng/L. When using the male cutoff of 35 ng/L, 8.12% of males with MI had non-elevated Alinity i STAT High Sensitivity Troponin-I test results. When using the overall cutoff of 27 ng/L, 6.49% of males with MI had non-elevated Alinity i STAT High Sensitivity Troponin-I test results. Troponin results should always be used in conjunction with clinical data, signs, and symptoms.

When using the female cutoff of 14 ng/L, the lower bound of the 95% CI for the PPV was as low as 21.45%. Taking into consideration the lower bound of the 95% CI, up to 77.54% (at 0 to < 1 hour), 78.55% (at 1 to < 3 hours), 74.86% (at 3 to < 6 hours), and 65.62% (at ≥ 6 hours) of positive troponin results could come from females that are not having an MI. When using the overall cutoff of 27 ng/L, the lower bound of the 95% CI for PPV was as low as 29.19%. Taking into consideration the lower bound of the 95% CI, up to 70.81% (at 0 to < 1 hour), 66.97% (at 1 to < 3 hours), 67.55% (at 3 to < 6 hours), and 61.55% (at ≥ 6 hours) of positive troponin results could come from females that are not having an MI. Troponin results should always be used in conjunction with clinical data, signs, and symptoms.

{18}------------------------------------------------

When using the male cutoff of 35 ng/L, the lower bound of the 95% CI for the PPV was as low as 27.88%. Taking into consideration the lower bound of the 95% CI, up to 65.30% (at 0 to < 1 hour), 72.12% (at 1 to < 3 hours), 64.96% (at 3 to < 6 hours), and 56.38% (at ≥ 6 hours) of positive troponin results could come from males that are not having an MI. When using the overall cutoff of 27 ng/L, the lower bound of the 95% CI for PPV was as low as 23.92%. Taking into consideration the lower bound of the 95% CI, up to 69.33% (at 0 to < 1 hour), 76.08% (at 1 to < 3 hours), 68.71% (at 3 to < 6 hours), and 61.72% (at ≥ 6 hours) of positive troponin results could come from males that are not having an MI. Troponin results should always be used in conjunction with clinical data, signs, and symptoms.

Troponin results should always be used in conjunction with clinical data, signs, and symptoms in accordance with the fourth universal definition of MI requiring acute myocardial injury with clinical evidence of acute myocardial ischemia, detection of a rise and/or fall of cardiac troponin (cTn) values, at least one value above the 99th percentile upper reference limit (URL), and at least one of the following: symptoms of myocardial ischemia, new ischemic ECG changes, development of pathological Q waves, imaging evidence of new loss of viable myocardium or new regional wall motion abnormality in a pattern consistent with an ischemic etiology, identification of a coronary thrombus by angiography.

There are conditions other than MI that are known to cause myocardial injury and elevated troponin values. The Alinity i STAT High Sensitivity Troponin-I clinical trial enrolled all patients presenting to the ED with symptoms consistent with ACS. Some of these patients had an acute or chronic condition other than MI.

&quot; Thygesen K, Alpert JS, Jaffe AS, et al. Fourth universal definition of myocardial infarction (2018). J Am Coll Cardiol 2018;72(18):2231-2264.

{19}------------------------------------------------

A sex difference in 99th percentile has been reported. **

In the clinical trial, 14.1% of patients without an MI diagnosis had at least one Alinity i STAT High Sensitivity Troponin-I test result above the sex-specific 99th percentile on one or more serial draws.

One or more of the following conditions were found in 94.2% of these patients:

Cardiac ConditionsNon-Cardiac Conditions
Atrial fibrillationCerebrovascular accidents and subarachnoid bleeds
Cardiotoxic drugsChronic obstructive pulmonary disease
Heart failureChronic renal insufficiency with or without hemodialysis
HypertensionCocaine user
Infiltrative cardiomyopathiesDiabetes mellitus
Left ventricular hypertrophyPulmonary embolism
MyocarditisRhabdomyolysis
Recent cardiac intervention and/or surgerySepsis
Recent MIVigorous exercise

The Area Under the Curve (AUC) results are summarized in the following table.

SexTime PointanbAUCStandard Error95% Wald CI
Female0 to < 1 Hour15740.92570.0129[0.9003, 0.9510]
Female1 to < 3 Hours8410.95820.0179[0.9230, 0.9933]
Female3 to < 6 Hours8540.97770.0056[0.9666, 0.9888]
Female≥ 6 Hours2840.95000.0148[0.9210, 0.9790]
Male0 to < 1 Hour29170.89940.0106[0.8787, 0.9202]

Wu AHB, Christenson RH, Greene DN, et al. Clinical laboratory practice recommendations for the use of cardiac troponin in acute coronary syndrome: expert opinion from the Academy of the American Association for Clinical Chemistry and the Task Force on Clinical Applications of Cardiac Bio-Markers of the International Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem 2018;64(4):645-655.

T Thygesen K, Alpert JS, Jaffe AS, et al. Fourth universal definition of myocardial infarction (2018). J Am Coll Cardiol 2018;72(18):2231-2264.

* Apple FS, Ler R, Murakami MM. Determination of 19 cardiac troponin I and T assay 99th percentile values from a common presumably healthy population. Clin Chem 2012:58(11):1574-1581.

് Obuchowski NA. Fundamentals of clinical research for radiologists: ROC analysis. Am J Roentgenol 2005;184(2):364-372.

{20}------------------------------------------------

SexTime PointanbAUCStandard Error95% Wald CI
1 to < 3 Hours13390.94550.0094[0.9270, 0.9640]
3 to < 6 Hours14800.94890.0094[0.9306, 0.9673]
≥ 6 Hours5760.92430.0140[0.8969, 0.9516]

a All time points are relative to ED presentation / ED triage.

b Some time points could not be collected for some subjects.

IX. Conclusion Drawn from Nonclinical Laboratory Studies and Clinical Performance

The results presented in this 510(k) demonstrate that the subject device (Alinity i STAT High Sensitivity Troponin-I) performance is substantially equivalent to the predicate device (ARCHITECT STAT High Sensitivity Troponin-I, K191595).

The similarities and differences between the subject device and predicate device are presented in section VII.

There is no known potential adverse effect to the operator when using this in vitro device according to the Alinity i STAT High Sensitivity Troponin-I reagent package insert instructions.

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.