K Number
K191595

Validate with FDA (Live)

Date Cleared
2019-09-13

(88 days)

Product Code
Regulation Number
862.1215
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARCHITECT STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnl) in human plasma (dipotassium [K2] EDTA) on the ARCHITECT i2000SR System.

The ARCHITECT STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

Device Description

The ARCHITECT STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of cardiac troponin I (cTnI) in human plasma (dipotassium EDTA) using chemiluminescent microparticle immunoassay technology with flexible assay protocols, referred to as Chemiflex. The reagent kit contains anti-troponin I (mouse, monoclonal) coated microparticles and anti-troponin I (mouse-human chimeric, monoclonal) acridinium-labeled conjugate. The procedure involves combining sample and antibody-coated microparticles, incubation and wash, adding acridinium-labeled conjugate, another wash cycle, and then adding Pre-Trigger and Trigger Solutions. The resulting chemiluminescent reaction is measured as relative light units (RLUs), which are directly related to the amount of cTnI in the sample. The cTnI concentration is determined relative to a standard curve.

AI/ML Overview

The provided text describes the Abbott Laboratories ARCHITECT STAT High Sensitivity Troponin-I assay (K191595). Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or categorical format for diagnostic performance (e.g., "Sensitivity must be >X%"). Instead, it presents the performance characteristics observed in the clinical study, implying that these demonstrated values were deemed acceptable for substantial equivalence. The non-clinical performance (precision, linearity, limits of measurement, and analytical specificity) results are presented as the device performance and are implicitly accepted as demonstrating substantial equivalence to the predicate.

For the purpose of this response, I will consider the reported "Sensitivity" and "Specificity" values at different time points and cutoffs as the "reported device performance." The document states that the conclusion drawn from nonclinical laboratory studies and clinical performance is that the candidate assay (ARCHITECT STAT High Sensitivity Troponin-I) performance is substantially equivalent to the predicate Elecsys Troponin T Gen 5 STAT Immunoassay (K162895). This implies the presented performance metrics met the FDA's criteria for substantial equivalence.

Reported Device Performance (Clinical Study - Myocardial Infarction Diagnosis):

MetricSexTime PointCutoff (ng/L)Reported Performance (%)95% CI (%)
SensitivityFemaleBaseline28 (Overall)91.773.0 - 99.0
Female2 - 4 Hours28 (Overall)94.472.7 - 99.9
Female4 - 9 Hours28 (Overall)94.171.3 - 99.9
MaleBaseline28 (Overall)81.870.4 - 90.2
Male2 - 4 Hours28 (Overall)91.781.6 - 97.2
Male4 - 9 Hours28 (Overall)93.784.5 - 98.2
FemaleBaseline17 (Female)95.878.9 - 99.9
Female2 - 4 Hours17 (Female)94.472.7 - 99.9
Female4 - 9 Hours17 (Female)94.171.3 - 99.9
MaleBaseline35 (Male)78.867.0 - 87.9
Male2 - 4 Hours35 (Male)90.079.5 - 96.2
Male4 - 9 Hours35 (Male)93.784.5 - 98.2
SpecificityFemaleBaseline28 (Overall)92.088.9 - 94.5
Female2 - 4 Hours28 (Overall)89.385.8 - 92.1
Female4 - 9 Hours28 (Overall)87.083.1 - 90.4
MaleBaseline28 (Overall)81.577.6 - 84.9
Male2 - 4 Hours28 (Overall)83.579.8 - 86.7
Male4 - 9 Hours28 (Overall)81.076.9 - 84.6
FemaleBaseline17 (Female)87.683.9 - 90.7
Female2 - 4 Hours17 (Female)85.381.4 - 88.6
Female4 - 9 Hours17 (Female)82.878.5 - 86.6
MaleBaseline35 (Male)84.580.9 - 87.8
Male2 - 4 Hours35 (Male)86.182.6 - 89.1
Male4 - 9 Hours35 (Male)84.380.5 - 87.6
AUCFemaleBaselineN/A0.94580.8738-1.0000
Female2 - 4 HoursN/A0.94020.8381-1.0000
Female4 - 9 HoursN/A0.94040.8339-1.0000
MaleBaselineN/A0.91360.8818-0.9453
Male2 - 4 HoursN/A0.93880.9028-0.9747
Male4 - 9 HoursN/A0.94790.9105-0.9854

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size:

    • Clinical Study: 1065 subjects presenting to the ED with symptoms consistent with ACS.
      • 248 specimens with serial sampling from 116 MI subjects (31 female, 85 male).
      • 2488 specimens with serial sampling from 949 non-MI subjects (440 female, 509 male).
    • Reference Range Study: 1531 apparently healthy individuals.
  • Data Provenance:

    • Clinical Study: Multi-center prospective study from 11 geographically diverse EDs in the US.
    • Reference Range Study: US population.
    • The document implies the studies are prospective as subjects were enrolled for acute clinical presentation (ACS symptoms) and for reference range establishment. Specimens were collected and frozen for subsequent evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Three.
  • Qualifications of Experts: Board-certified cardiologists.

4. Adjudication method for the test set

  • Adjudication Method: The final adjudicated diagnosis (ground truth) was made by majority agreement of the three board-certified cardiologists. This is often referred to as a 3+1 method where consensus among the human readers determines the outcome.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study concerning human readers improving with or without AI assistance was not done. This device is an in vitro diagnostic (IVD) assay, a laboratory test system, not an AI or imaging diagnostic software that assists human readers.


6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the clinical study assessed the standalone performance of the ARCHITECT STAT High Sensitivity Troponin-I assay. The assay itself provides a quantitative measurement and these measurements, when compared against the adjudicated clinical diagnosis, determine its diagnostic accuracy (sensitivity, specificity, etc.). There is no "human-in-the-loop" explicitly described for interpreting the assay result itself, although clinicians use the results in conjunction with other clinical signs.


7. The type of ground truth used

  • Clinical Study: The ground truth for the diagnosis of myocardial infarction (MI) was established by the adjudicated diagnosis of three board-certified cardiologists according to the 2007 universal definition of MI. The document also states that troponin results should always be used in conjunction with clinical signs and symptoms in accordance with the fourth universal definition of MI.
  • Reference Range Study: Ground truth for "apparently healthy individuals" was based on participants having normal levels of cardiac B-type natriuretic peptide (BNP) and HbA1c, and glomerular filtration rate (GFR) values ≥ 60 mL/min.

8. The sample size for the training set

The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an immunoassay (hardware and reagents), not an AI algorithm. Therefore, the concept of a "training set" as it relates to AI is not applicable here. The focus is on the analytical and clinical validation of the assay.


9. How the ground truth for the training set was established

As noted above, the concept of a "training set" as related to AI is not applicable to this device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the text "U.S. FOOD & DRUG ADMINISTRATION" is written in blue.

September 13, 2019

Abbott Laboratories Diagnostic Division Judi Wallach Regulatory Affairs Project Manager Dept. 09AA. Bldg. AP8-1. 100 Abbott Park Road Abbott Park, IL 60064-6038

Re: K191595

Trade/Device Name: ARCHITECT STAT High Sensitivity Troponin-I Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI Dated: June 13, 2019 Received: June 17, 2019

Dear Judi Wallach:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191595

Device Name ARCHITECT STAT High Sensitivity Troponin-I

Indications for Use (Describe)

The ARCHITECT STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnl) in human plasma (dipotassium [K2] EDTA) on the ARCHITECT i2000SR System.

The ARCHITECT STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (Summary of Safety and Effectiveness)

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

I. 510(k) Number

K191595

II. Applicant Name

Abbott Laboratories Diagnostics Division Dept. 9AA, AP8-1 100 Abbott Park Road Abbott Park, IL 60064

Primary contact person for all communications:

Judi Wallach, ADD, Regulatory Affairs Project Manager Telephone Number: (224) 667-1132 Fax Number: (224) 667-4836 E-Mail: judith.r.wallach@abbott.com Date summary prepared: August 14, 2019

Secondary contact person for all communications:

Grace LeMieux ADD, Director, Regulatory Affairs Telephone Number: (224) 668-0409 E-Mail: grace.lemieux@abbott.com

III. Device Name

ARCHITECT STAT High Sensitivity Troponin-I

Reagents

Trade Name: ARCHITECT STAT High Sensitivity Troponin-I Device Classification: Class II Classification Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Governing Regulation: 862.1215

Code: MMI

{4}------------------------------------------------

IV. Predicate Device

Reagents

Elecsys Troponin T Gen 5 STAT Immunoassay (K162895)

V. Description of Device

Reagents

The ARCHITECT STAT High Sensitivity Troponin-I reagent kit contains:

  • . Microparticles: 1 bottle (6.6 mL per 100 test bottle / 29.0 mL per 500 test bottle) Anti-troponin I (mouse, monoclonal) coated microparticles in TRIS buffer with protein (bovine) stabilizer. Minimum concentration: 0.035% solids. Preservative: ProClin 300.
  • . Conjugate: 1 bottle (5.9 mL per 100 test bottle / 28.5 mL per 500 test bottle). Anti-troponin I (mouse-human chimeric, monoclonal) acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer and human IgG. Minimum concentration: 0.1 mg/L. Preservative: ProClin 300.

Principles of the Procedure

The ARCHITECT STAT High Sensitivity Troponin-I assay is a two-step immunoassay for the quantitative determination of cardiac troponin I in human plasma (dipotassium EDTA) using chemiluminescent microparticle immunoassay technology with flexible assay protocols, referred to as Chemiflex.

    1. Sample and anti-troponin I antibody-coated paramagnetic microparticles are combined. The cTnI present in the sample binds to the anti-troponin I coated microparticles.
    1. After incubation and wash, anti-troponin I acridinium-labeled conjugate is added.
    1. Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture.
    1. The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of cTnI in the sample and the RLUs detected by the ARCHITECT iSystem optics.

The cTnI concentration is read relative to a standard curve established with calibrators of known cTnI concentrations.

{5}------------------------------------------------

VI. Intended Use of the Device

The ARCHITECT STAT High Sensitivity Troponin-I assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of cardiac troponin I (cTnI) in human plasma (dipotassium [K₂] EDTA) on the ARCHITECT i2000SR System.

The ARCHITECT STAT High Sensitivity Troponin-I assay is to be used as an aid in the diagnosis of myocardial infarction (MI).

VII. Comparison of Technological Characteristics

The ARCHITECT STAT High Sensitivity Troponin-I assay (candidate assay) utilizes a CMIA methodology for the quantitative in vitro determination of cTnI and is intended for use on the ARCHITECT i2000SR System.

The similarities and differences between the candidate assay and the predicate assay are presented in the following table.

CharacteristicsCandidate DeviceARCHITECT STAT High SensitivityTroponin-IPredicate DeviceRoche cobas Elecsys Troponin TGen 5 STAT(K162895)
PlatformARCHITECT i2000SRcobas e 411 and e 601 immunoassayanalyzers
MethodologyCMIAElectrochemiluminescenceImmunoassay (ECLIA)
Intended Use andIndications for UseThe ARCHITECT STAT High SensitivityTroponin-I assay is a chemiluminescentmicroparticle immunoassay (CMIA) usedfor the quantitative determination ofcardiac troponin I (cTnI) in human plasma(dipotassium [K2] EDTA) on theARCHITECT i2000SR System.The ARCHITECT STAT High SensitivityTroponin-I assay is to be used as an aid inthe diagnosis of myocardial infarction(MI).Immunoassay for the in vitroquantitative determination of cardiactroponin T (cTnT) in lithium heparinplasma. The immunoassay is intendedto aid in the diagnosis of myocardialinfarction.The electrochemiluminescenceimmunoassay "ECLIA" is intended foruse on the cobas system analyzers.
Specific AnalyteDetectedcTnIcTnT
CharacteristicsCandidate DeviceARCHITECT STAT High SensitivityTroponin-IPredicate DeviceRoche cobas Elecsys Troponin TGen 5 STAT(K162895)
Specimen TypePlasma (K2 EDTA)Plasma (lithium heparin)
Reagent ComponentsMicroparticles – Anti-troponin I (mouse,monoclonal) coated microparticles in TRISbuffer with protein (bovine) stabilizer.Minimum concentration: 0.035% solids.Preservative: ProClin 300.Conjugate - Anti-troponin I(mouse-human chimeric, monoclonal)acridinium-labeled conjugate in MESbuffer with protein (bovine) stabilizer andhuman IgG. Minimum concentration:0.1 mg/L. Preservative: ProClin 300.M - Streptavidin-coated microparticles(transparent cap), 1 bottle, 6.5 mL:Streptavidin-coated microparticles0.72 mg/mL; preservative.R1 - Anti-troponin T-Ab~biotin (graycap), 1 bottle, 8 mL: Biotinylatedmonoclonal anti-cardiac troponinT-antibody (mouse) 2.5 mg/L;phosphate buffer 100 mmol/L, pH 6.0;preservative; inhibitors.R2 - Anti-troponinT-Ab-Ru(bpy)2+(black cap), 1 bottle,8 mL: Monoclonal chimericanti-cardiac troponin T-antibody(mouse/human) labeled with rutheniumcomplex 2.5 mg/L; phosphate buffer100 mmol/L, pH 6.0; preservative.
Limit of Quantitation(LoQ)The LoQ is 3.5 ng/L.The LoQ at ≤ 20 %CV is 6.0 ng/L.
Measurement RangeAnalytical measuring interval (AMI):3.5 to 5000.0 ng/L (pg/mL)6.0 to 10,000 ng/L
PotentiallyInterferingEndogenousSubstances andClinical ConditionsSamples targeted to 15 and 500 ng/L cTnIwere evaluated with unconjugated andconjugated bilirubin (20 mg/dL),hemoglobin (500 mg/dL), total protein(9.3 g/dL) and triglycerides (3000 mg/dL).No significant interference (interferencewithin ± 10%) was observed.Samples from approximately 13 to8500 ng/L cTnT were evaluated withbilirubin (25 mg/dL), hemoglobin(100 mg/dL), lipemia / Intralipid(1500 mg/dL), human serum albumin(7 g/dL), cholesterol (310 mg/dL),biotin (20 ng/mL), rheumatoid factor(RF) (900 IU/mL), and humananti-mouse antibodies (HAMA)(322 µg/L). Bias less than 10% wasobserved.
Drug Interferences(General Drug andCardiac Drug Panel)Commonly used pharmaceuticals andcardiac-specific drugs were evaluated withsamples targeted to 15 and 500 ng/L cTnI.No significant interference (interferencewithin ± 10%) was observed at therapeuticlevels, and no significant interference(interference within ± 10%) was observedat high levels, with the exception offibrinogen.Commonly used pharmaceuticals wereevaluated in samples with cTnTconcentrations of 15 ng/L and9800 ng/L. Cardiac-specific drugs weretested in samples with cTnTconcentrations of 15 ng/L and1900 ng/L. Bias less than ± 10% wasobserved.
CharacteristicsCandidate DeviceARCHITECT STAT High SensitivityTroponin-IPredicate DeviceRoche cobas Elecsys Troponin TGen 5 STAT(K162895)
99th PercentileCutoff / ExpectedValues fromApparently HealthyIndividualsFemale: 17 ng/L (pg/mL)Male: 35 ng/L (pg/mL)Overall: 28 ng/L (pg/mL)Female: 14 ng/LMale: 22 ng/LOverall: 19 ng/L

Similarities and Differences

{6}------------------------------------------------

{7}------------------------------------------------

VIII. Summary of Nonclinical Performance

A. Precision

Reproducibility

A study was performed using 1 lot of the ARCHITECT STAT High Sensitivity Troponin-I reagent, 1 lot of the ARCHITECT STAT High Sensitivity Troponin-I Calibrators, and 1 lot of the ARCHITECT STAT High Sensitivity Troponin-I Controls. The study was performed to include K2 EDTA plasma specimens within each of 4 concentration ranges (> LoQ to 6 ng/L, 10 to 20 ng/L, 30 to 50 ng/L, and 150 to 200 ng/L). Only one specimen per concentration range was collected in a single day. The study was performed over a minimum of 3 days. Each specimen was stored at room temperature and tested in duplicate, twice in one day, on each of 3 instruments (for a total of 12 replicates) within 8 hours of collection.

Within-RunBetween-RunWithin-LaboratoryªReproducibilityb
SamplenMean(ng/L)SD%CVSD%CVSD%CVSD%CV
Sample 1125.30.122.20.234.20.254.80.254.8
Sample 21211.20.474.20.000.00.474.20.625.5
Sample 31217.50.502.90.231.30.553.10.814.6
Sample 41218.80.603.20.221.20.643.40.754.0
Sample 51234.60.842.40.000.00.842.41.103.2
Sample 61238.81.002.61.102.81.483.81.925.0
Sample 71245.01.523.41.423.22.084.63.167.0
Sample 812163.75.003.16.233.87.994.911.687.1
Sample 912167.56.824.12.821.77.384.411.096.6
Sample 1012179.66.763.80.000.06.763.88.194.6

4 Includes within-run and between-run variability.

b Includes within-run, between-run, and between-instrument variability.

{8}------------------------------------------------

Within-Laboratory Precision

A study was performed based on guidance from Clinical and Laboratory Standards Institute (CLSI) EP05-A2. Testing was conducted using 3 lots of the ARCHITECT STAT High Sensitivity Troponin-I reagent, 2 lots of the ARCHITECT STAT High Sensitivity Troponin-I Calibrators, 1 lot of the ARCHITECT STAT High Sensitivity Troponin-I Controls, 1 lot each of Bio-Rad Liquichek Cardiac Markers Plus Control LT (Level Low, 2, and 3), and 2 instruments. Five controls were tested in duplicate, twice per day on 20 days, following the manufacturers' storage and handling requirements.

Note: Patient samples can only be stored for 8 hours at room temperature; therefore, 20-day precision was conducted with quality controls. Bio-Rad controls were thawed and tested each day of the precision study.

Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. CLSI Document EP05-A2. Wayne, PA: CLSI; 2004.

{9}------------------------------------------------

ReagentLotnMean(ng/L)Within-RunWithin-Laboratory(Total)a
SampleInstrumentSD%CVSD%CV
Low Control118019.30.613.20.723.7
28020.30.613.00.783.9
38019.70.643.30.783.9
218020.40.844.10.854.1
28020.20.643.20.834.1
38020.00.663.30.874.3
Medium Control1180190.74.212.25.542.9
280195.03.571.84.132.1
380191.24.272.24.492.3
2180197.85.262.75.762.9
280196.84.652.45.362.7
380194.34.432.35.953.1
Bio-Rad Level118043.31.272.91.463.4
Low28046.11.483.21.573.4
38045.41.272.81.513.3
218045.21.363.01.824.0
28046.41.803.91.844.0
38046.01.403.01.483.2
Bio-Rad Level 211801198.031.132.633.802.8
2801281.333.382.640.953.2
3801267.127.572.231.722.5
21801260.138.343.042.333.4
2801309.128.252.242.683.3
3801309.635.682.743.813.3
Bio-Rad Level 311802812.364.562.380.502.9
2803023.083.522.893.823.1
3803015.394.133.195.143.2
21802978.380.342.7102.423.4
2803103.983.932.796.253.1
3803138.255.491.884.502.7

4 Includes within-run, between-run, and between-day variability.

{10}------------------------------------------------

B. Lower Limits of Measurement

A study was performed based on guidance from CLSI EP17-A2. T Testing of zero-analyte samples was conducted using 4 lots of the ARCHITECT STAT High Sensitivity Troponin-I reagent kit across 5 instruments over a minimum of 3 days. Testing of low-analyte samples was conducted using 2 lots of the ARCHITECT STAT High Sensitivity Troponin-I reagent kit on each of 2 instruments over a minimum of 3 days. The limit of blank (LoB), limit of detection (LoD), and LoQ values are summarized below.

ng/L (pg/mL)
LoBa0.9
LoDb1.7
LoQc3.5

a The LoB represents the 95th percentile from n ≥ 60 replicates of zero-analyte samples.

b The LoD represents the lowest concentration at which the analyte can be detected with 95% probability based on n ≥ 60 replicates of low-analyte level samples.

c The LoQ presented in the table is in alignment with the low end of the AMI for the ARCHITECT STAT High Sensitivity Troponin-I assay. The observed LoQ on the ARCHITECT i2000SR System was 2.3 ng/L (2.3 pg/mL). This LoQ is defined as the lowest concentration at which a maximum allowable precision of 20 %CV was met and was determined from n ≥ 60 replicates of low-analyte level samples.

C. Linearity

A study was performed based on guidance from CLSI EP06-A. * This assay is linear across the analytical measuring interval of 3.5 to 5000.0 ng/L (3.5 to 5000.0 pg/mL).

D. Measuring Interval

Based on representative data for the LoQ, the ranges over which results can be quantified are provided below.

آ Clinical and Laboratory Standards Institute (CLSI). Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. CLSI Document EP17-A2. Wayne, PA: CLSI; 2012.

* Clinical and Laboratory Standards Institute (CLSI). Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. CLSI Document EP06-A. Wayne, PA: CLSI; 2003.

{11}------------------------------------------------

ng/L (pg/mL)
AMIa3.5 - 5000.0
Extended Measuring Interval (EMI)b5000.0 - 50,000.0

a AMI: The AMI extends from the LoQ to the upper limit of quantitation (ULoQ).

b EMI: The EMI extends from the ULoQ x dilution factor. The value reflects a 1:10 dilution factor.

E. Analytical Specificity

1. Interference

Potentially Interfering Endogenous Substances

A study was performed based on guidance from CLSI EP07-A2. * Each substance was tested at 2 levels of the analyte (approximately 15 ng/L and 500 ng/L). No significant interference (interference within ± 10%) was observed at the following concentrations:

Potentially Interfering SubstanceInterferent Level
Unconjugated Bilirubin≤ 20 mg/dL
Conjugated Bilirubin≤ 20 mg/dL
Hemoglobin≤ 500 mg/dL
Total Protein≤ 9.3 g/dL
Triglycerides≤ 3000 mg/dL

Interference beyond ± 10% was observed at the concentrations shown below for

the following substance.

PotentiallyInterferingSubstanceInterferent LevelAnalyte Level% Interference
Total Protein12.4 g/dL15 ng/L-12.0%
Total Protein12.4 g/dL500 ng/L-18.4%

§ Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition. CLSI Document EP07-A2. Wayne, PA: CLSI; 2005.

{12}------------------------------------------------

Total protein at 12.4 g/dL decreases troponin values at 15 ng/L and 500 ng/L by -12.0% and -18.4%, respectively.

Potentially Interfering Drugs

A study was performed based on guidance from CLSI EP07-A2. Each drug was tested at 2 levels of the analyte (approximately 15 ng/L and 500 ng/L). No significant interference (interference within ± 10%) was observed at the following concentrations:

PotentiallyInterfering DrugInterferent LevelPotentiallyInterfering DrugInterferent Level
TherapeuticHighTherapeuticHigh
Abciximab4 µg/mL20 µg/mLIbuprofen40 µg/mL500 µg/mL
Acetaminophen20 µg/mL250 µg/mLLevodopa1.8 µg/mL20 µg/mL
Acetylsalicylic Acid260 µg/mL1000 µg/mLLow MW Heparin1.8 U/mL5 U/mL
Adrenaline60 ng/mL0.37 µg/mLMethyldopa4 µg/mL25 µg/mL
Allopurinol12 µg/mL400 µg/mLMethylprednisolone8 µg/mL80 µg/mL
Ambroxol0.1 µg/mL400 µg/mLMetronidazole23 µg/mL200 µg/mL
Ampicillin10 µg/mL1000 µg/mLNicotine37 ng/mL2 mg/dL
Ascorbic Acid12 µg/mL300 µg/mLNifedipine125 ng/mL60 µg/mL
Atenolol1 µg/mL10 µg/mLNitrofurantoin2.0 µg/mL64 µg/mL
Biotin10 ng/mL290 ng/mLNystatin2 µg/mL7.5 µg/mL
Bivalirudin11 µg/mL42 µg/mLOxytetracycline2 µg/mL5 µg/mL
Caffeine12 µg/mL100 µg/mLPhenobarbital25 µg/mL15 mg/dL
Captopril1.0 µg/mL50 µg/mLPhenylbutazone30 µg/mL400 µg/mL
Carvedilol5 µg/mL150 µg/mLPhenytoin12 µg/mL100 µg/mL
Cefoxitin120 µg/mL2500 µg/mLPrimidone10 µg/mL10 mg/dL
Cinnarizine4 µg/mL400 µg/mLPropranolol1 µg/mL5 µg/mL
Clopidogrel15 µg/mL75 µg/mLQuinidine4 µg/mL20 µg/mL
Cocaine0.1 µg/mL10 µg/mLRifampicin7 µg/mL60 µg/mL
Cyclosporine0.8 µg/mL5 µg/mLSalicylic Acid199 µg/mL600 µg/mL
Diclofenac2.5 µg/mL50 µg/mLSimvastatin4 µg/mL20 µg/mL
Digoxin1 ng/mL7.5 µg/mLSodium Heparin2 U/mL8 U/mL
Dopamine0.3 µg/mL900 µg/mLStreptokinase4 U/mL31.3 U/mL
Doxycycline10 µg/mL50 µg/mLTheophylline12 µg/mL75 µg/mL
Eptifibatide2 µg/mL7 µg/mLTPA0.52 µg/mL2.3 µg/mL
Erythromycin11 µg/mL200 µg/mLTrimethoprim12 µg/mL75 µg/mL
Fibrinogen100 mg/dLNAVerapamil325 ng/mL160 µg/mL
Fondaparinux1.2 µg/mL4 µg/mLWarfarin2 µg/mL30 µg/mL
Furosemide20 µg/mL400 µg/mL

MW = Molecular weight, NA = Not applicable, TPA = Tissue plasminogen activator

Interference beyond ± 10% was observed at the concentrations shown below for the following drug.

{13}------------------------------------------------

Potentially InterferingSubstanceInterferent LevelAnalyteLevel% Interference
FibrinogenNA1000 mg/dL15 ng/L11.6

Specimens from individuals with elevated levels of fibrinogen may demonstrate falsely elevated values.

Potentially Interfering Clinical Conditions

Twenty-three specimens positive for HAMA and 23 specimens positive for RF were evaluated for potential interference.

Specimens from patients who have received preparations of mouse monoclonal antibodies for diagnosis or therapy may contain HAMA. Such specimens may show either falsely elevated or depressed values when tested with assay kits such as ARCHITECT STAT High Sensitivity Troponin-I or others that employ mouse monoclonal antibodies. ** ** Specimens containing HAMA may show either falsely elevated or depressed values when tested with the ARCHITECT STAT High Sensitivity Troponin-I assay.

Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or to animal serum products can be prone to this interference, and anomalous values may be observed. Additional information may be required for diagnosis. **

RF in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. * Specimens containing RF may show either falsely

Primus FJ, Kelley EA, Hansen HJ, et al. "Sandwich" type immunoassay of carcinoembryonic antigen in patients receiving murine monoclonal antibodies for diagnosis and therapy. Clin Chem 1988;34(2):261-264.

** Schroff RW, Foon KA, Beatty SM, et al. Human anti-murine immunoglobulin responses in patients receiving monoclonal antibody therapy. Cancer Res 1985;45(2):879-885.

## Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988;34(1):27-33.

{14}------------------------------------------------

elevated or depressed values when tested with the ARCHITECT STAT High Sensitivity Troponin-I assay.

Although the ARCHITECT STAT High Sensitivity Troponin-I assay is specifically designed to minimize the effects of HAMA, heterophilic antibodies, and RF, assay results may be impacted by these proteins.

Troponin autoantibodies have been reported to be present in approximately 10% to 20% of patients presenting to the emergency department (ED) and may lead to falsely low troponin assay results and delay in treatment of acute coronary syndrome (ACS). §§ · *** Therefore, a test result that is inconsistent with the clinical picture and patient history should be interpreted with caution.

2. Cross-Reactants

A study was performed based on guidance from CLSI EP07-A2. Samples with cTnI concentrations from 3.5 to 5000 ng/L containing the cross-reactants listed in the following table were tested with the ARCHITECT STAT High Sensitivity Troponin-I assay.

The observed % cross-reactivity was < 0.1% for skeletal troponin I and < 1% for all other cross-reactants.

Cross-ReactantCross-Reactant Concentration
Actin1,000,000 ng/L
Cardiac troponin T1,000,000 ng/L
Creatine kinase-muscle/brain (CK-MB)1,000,000 ng/L
Myoglobin1,000,000 ng/L
Myosin1,000,000 ng/L
Skeletal troponin I1,000,000 ng/L
Tropomyosin1,000,000 ng/L
Troponin C1,000,000 ng/L

89 Park JY, Jaffe AS. Troponin autoantibodies: from assay interferent to mediator of cardiotoxicity. Clin Chem 2017;63(1):30-32.

*** Nussinovitch U, Shoenfeld Y. Anti-troponin autoantibodies and the cardiovascular system. Heart 2010;96:1518-1524.

{15}------------------------------------------------

F. Expected Values

A reference range study was conducted based on guidance from CLSI EP28-A 3c. *** Specimens were collected from 1531 apparently healthy individuals in a US population with normal levels of cardiac B-type natriuretic peptide (BNP) and HbA1c, and glomerular filtration rate (GFR) values ≥ 60 mL/min. Each specimen was stored frozen, thawed, and evaluated in replicates of one using the ARCHITECT STAT High Sensitivity Troponin-I assay. The 99th percentiles described in the following table for this population were determined using the robust statistical method described in CLSI EP28-A3c.

Apparently HealthyPopulationNAge Range(years)99th Percentile(ng/L, pg/mL)90% CI*(ng/L, pg/mL)
Female76521 - 7517[14, 20]
Male76621 - 7335[27, 44]
Overall153121 - 7528[22, 33]
  • CI = Confidence Interval

IX. Summary of Clinical Performance

A multi-center prospective study was performed to assess diagnostic accuracy of the ARCHITECT STAT High Sensitivity Troponin-I assay. Specimens were collected at 11 EDs from 1065 subjects presenting to the ED with symptoms consistent with ACS. The specimen collection sites represented geographically diverse EDs associated with primary care hospitals and medical centers, reflecting regional, urban, suburban, and rural patient populations. All subject diagnoses were adjudicated by three board certified cardiologists according to the 2007 universal definition of MI *** before the ARCHITECT STAT High Sensitivity Troponin-I assay results were available. The final adjudicated diagnosis was made by majority agreement of the 3 board-certified cardiologists. The observed MI prevalence in this study was 10.89%.

#* Clinical and Laboratory Standards Institute (CLSI). Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline—Third Edition. CLSI Document EP28-A3c. Wayne, PA: CLSI; 2010.

## Thygesen K, Alpert JS, White HD. Universal definition of myocardial infarction. Eur Heart J 2007;28(20):2525-2538.

{16}------------------------------------------------

  • 248 specimens with serial sampling from 116 MI subjects (31 female subjects, . 85 male subjects)
  • . 2488 specimens with serial sampling from 949 non-MI subjects (440 female subjects, 509 male subjects)

The specimens were collected in K2 EDTA tubes and frozen. The specimens were thawed and evaluated using the ARCHITECT STAT High Sensitivity Troponin-I assay.

NOTE: The study population did not include type 4 or 5 MI subjects. Therefore, the ability of the assay to identify these patients was not evaluated.

The results were analyzed using the serial sampling time points collected during the ED visit.

An analysis for both females and males was performed using the overall 99th percentile cutoff (28 ng/L). The results are summarized in the following table.

SexTimePointaNbSensitivitycSpecificitydPPVeNPVf
%95% CI%95% CI%95% CI%95% CI
FemaleBaseline41291.7(22/24)73.0 -99.092.0(357/388)88.9 -94.541.5(22/53)28.1 -55.999.4(357/359)98.0 -99.9
2 - 4Hours41894.4(17/18)72.7 -99.989.3(357/400)85.8 -92.128.3(17/60)17.5 -41.499.7(357/358)98.5 -100.0
4 - 9Hours37294.1(16/17)71.3 -99.987.0(309/355)83.1 -90.425.8(16/62)15.5 -38.599.7(309/310)98.2 -100.0
MaleBaseline51981.8(54/66)70.4 -90.281.5(369/453)77.6 -84.939.1(54/138)30.9 -47.896.9(369/381)94.6 -98.4
2 - 4Hours52691.7(55/60)81.6 -97.283.5(389/466)79.8 -86.741.7(55/132)33.2 -50.698.7(389/394)97.1 -99.6
4 - 9Hours48993.7(59/63)84.5 -98.281.0(345/426)76.9 -84.642.1(59/140)33.9 -50.898.9(345/349)97.1 -99.7

a All time points are relative to ED presentation / ED triage; baseline is within 2 hours of ED presentation / ED triage.

b Some time points could not be collected for some subjects.

For footnotes c-f:

Diagnosis
ARCHITECT STAT High Sensitive Troponin-IMINon-MI
cTnI Value > cutpointAB
cTnI Value ≤ cutpointCD

c Sensitivity = A/(A + C) × 100

{17}------------------------------------------------

d Specificity = D/(B + D) × 100

  • e Positive Predictive Value (PPV) = A/(A + B) × 100
  • f Negative Predictive Value (NPV) = D/(C + D) × 100

The lower end of the CI for PPV demonstrated for female subjects using the established overall 99th percentile was as low as 15.5%. Taking into consideration the lower bound of the 95% CI, up to 71.9% (baseline), 82.5% (at 2 to 4 hours), and 84.5% (at 4 to 9 hours) of positive troponin results could come from females that are not having an MI.

The lower end of the CI for PPV demonstrated for male subjects using the established overall 99th percentiles was as low as 30.9%. Taking into consideration the lower bound of the 95% CI, up to 69.1% (baseline), 66.8% (at 2 to 4 hours), and 66.1% (at 4 to 9 hours) of positive troponin results could come from males that are not having an MI.

Troponin results should always be used in conjunction with clinical signs and symptoms in accordance with the fourth universal definition of MI§§§ requiring myocardial injury represented by a rise and/or fall of cTn values with at least one value above the 99th percentile URL and at least one of the following: symptoms of myocardial ischaemia, new ischaemic ECG changes, development of pathological O waves, imaging evidence of new loss of viable myocardium or new regional wall motion abnormality in a pattern consistent with an ischaemic aetiology, identification of a coronary thrombus by angiography or autopsy.

The results using the sex-specific 99th percentile cutoffs (female 17 ng/L, male 35 ng/L) are summarized in the following table.

§§§ Thygesen K, Alpert JS, Jaffe AS, et al. Fourth universal definition of myocardial infarction (2018). J Am Coll Cardiol 2018;72(18):2231-2264.

{18}------------------------------------------------

Cutoff(ng/L)TimePointaNbSensitivitycSpecificitydPPVeNPVf
%95% CI%95% CI%95% CI%95% CI
17(Femaleonly)Baseline41295.8(23/24)78.9 - 99.987.6(340/388)83.9 - 90.732.4(23/71)21.8 - 44.599.7(340/341)98.4 - 100.0
2 - 4Hours41894.4(17/18)72.7 - 99.985.3(341/400)81.4 - 88.622.4(17/76)13.6 - 33.499.7(341/342)98.4 - 100.0
4 - 9Hours37294.1(16/17)71.3 - 99.982.8(294/355)78.5 - 86.620.8(16/77)12.4 - 31.599.7(294/295)98.1 - 100.0
35(Maleonly)Baseline51978.8(52/66)67.0 - 87.984.5(383/453)80.9 - 87.842.6(52/122)33.7 - 51.996.5(383/397)94.2 - 98.1
2 - 4Hours52690.0(54/60)79.5 - 96.286.1(401/466)82.6 - 89.145.4(54/119)36.2 - 54.898.5(401/407)96.8 - 99.5
4 - 9Hours48993.7(59/63)84.5 - 98.284.3(359/426)80.5 - 87.646.8(59/126)37.9 - 55.998.9(359/363)97.2 - 99.7

a All time points are relative to ED presentation / ED triage; baseline is within 2 hours of ED presentation / ED triage.

b Some time points could not be collected for some subjects.

For footnotes c-f:

Diagnosis
ARCHITECT STAT High Sensitive Troponin-IMINon-MI
cTnI Value > cutpointAB
cTnI Value ≤ cutpointCD

c Sensitivity = A/(A + C) × 100

d Specificity = D/(B + D) × 100

e PPV = A/(A + B) × 100

t NPV = D/(C + D) × 100

The lower end of the CI for PPV demonstrated for female subjects using the established female 99th percentile was as low as 12.4%. Taking into consideration the lower bound of the 95% CI, up to 78.2% (baseline), 86.4% (at 2 to 4 hours), and 87.6% (at 4 to 9 hours) of positive troponin results could come from females that are not having an MI.

The lower end of the CI for PPV demonstrated for male subjects using the established male 99th percentile was as low as 33.7%. Taking into consideration the lower bound of the 95% CI, up to 66.3% (baseline), 63.8% (at 2 to 4 hours), and 62.1% (at 4 to 9 hours) of positive troponin results could come from males that are not having an MI.

Troponin results should always be used in conjunction with clinical signs and symptoms in accordance with the fourth universal definition of MI requiring myocardial injury

{19}------------------------------------------------

represented by a rise and/or fall of cTn values with at least one value above the 99th percentile URL and at least one of the following: symptoms of myocardial ischaemia, new ischaemic ECG changes, development of pathological Q waves, imaging evidence of new loss of viable myocardium or new regional wall motion abnormality in a pattern consistent with an ischaemic aetiology, identification of a coronary thrombus by angiography or autopsy.

There are conditions other than MI that are known to cause myocardial injury and elevated troponin values. The ARCHITECT STAT High Sensitivity Troponin-I clinical trial enrolled all patients presenting to the ED with symptoms consistent with ACS. Some of these patients had an acute or chronic condition other than MI.

In the clinical trial, 16.5% of patients without an MI diagnosis had at least one ARCHITECT STAT High Sensitivity Troponin-I test result above the sex-specific 99th percentile on one or more serial draws.

Cardiac ConditionsNon-Cardiac Conditions
AnginaCardiac contusion related to a traumatic injury
Atrial fibrillationChronic lung disease
CardiomyopathyPneumonia
Coronary artery diseasePulmonary embolism
Heart failureRenal failure
Hypertensive urgencyShock
PericarditisSystemic sclerosis
Recent cardiac intervention
Severe valvular heart disease
Tachycardia

One or more of the following conditions were found in 71.3% of these patients:

The Area Under the Curve (AUC) results *** are summarized in the following table.

Obuchowski NA. Fundamentals of clinical research for radiologists: ROC analysis. Am J Roentgenol 2005;184(2):364-372.

{20}------------------------------------------------

SexTime PointaNbAUCStandard Error95% Wald CI
FBaseline4120.94580.0367[0.8738, 1.0000]
2 - 4 Hours4180.94020.0521[0.8381, 1.0000]
4 - 9 Hours3720.94040.0544[0.8339, 1.0000]
MBaseline5190.91360.0162[0.8818, 0.9453]
2 - 4 Hours5260.93880.0184[0.9028, 0.9747]
4 - 9 Hours4890.94790.0191[0.9105, 0.9854]

a All time points are relative to ED presentation / ED triage; baseline is within 2 hours of ED presentation / ED triage.

b Some time points could not be collected for some subjects.

X. Conclusion Drawn from Nonclinical Laboratory Studies and Clinical Performance

The results presented in this 510(k) premarket notification demonstrate that the candidate assay (ARCHITECT STAT High Sensitivity Troponin-I) performance is substantially equivalent to the predicate Elecsys Troponin T Gen 5 STAT Immunoassay (K162895).

The similarities and differences between the candidate assay and the predicate assay are presented in section VII. The results presented in this 510(k) provide reasonable assurance that the ARCHITECT STAT High Sensitivity Troponin-I assay is safe and effective for the stated intended use. Any differences between the candidate assay and the predicate assay shown in the tables do not affect the safety and effectiveness of the candidate assay.

There is no known potential adverse effect to the operator when using this in vitro device according to the ARCHITECT STAT High Sensitivity Troponin-I package insert instructions.

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.