K Number
K202440

Validate with FDA (Live)

Date Cleared
2021-06-17

(296 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The infrared thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It can be used by consumers in the household environment and by healthcare providers.

Device Description

HT series infrared thermometers (includes model HT-101, HT-102, HT-103 and HT-104) measure the body temperature through receiving infrared energy radiation via the ear or forehead. These thermometers have the capability to measure temperature via forehead temperature mode or ear temperature mode, and the temperature is directly shown on the LCD display. These thermometers have the following features:

  • The device is intended to be used by consumers in the household environments and by healthcare providers;
  • The device is mainly composed of infrared sensor, signal receiving processor, buttons, buzzer, LCD display. It is powered by 2*1.5 AAA batteries;
  • It focuses the infrared energy received from the human's forehead or ear by using the Fresnel lens of the thermometer;
  • Switching between Celsius and Fahrenheit;
  • Multi-functional, can measure ear and forehead temperature;
  • Automatic range selection; resolution is 0.1 Celsius (0.1 Fahrenheit);
  • The latest 35 measurement data can be memorized and stored; the user can view the previous measurement results;
  • High temperature alarm function, displayed in green, orange and red backlight;
  • Switching between mute and un-mute mode;
  • Low battery indication, and auto shut-down.
AI/ML Overview

The document describes the acceptance criteria and the study conducted for the Infrared Thermometer (Models HT-101, HT-102, HT-103, HT-104).

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Infrared Thermometer are primarily based on established medical device standards and comparison to a predicate device (K191570).

Feature / CriterionAcceptance Criteria (from predicate or standards)Reported Device Performance (Subject Devices)Comments
Product CodeFLLFLLSame
Regulation Number880.2910880.2910Same
Indications for UseIntermittent measurement of body temperature from auditory canal or central forehead skin surface on people of all ages; for consumers and healthcare providers.Intermittent measurement of body temperature from auditory canal or central forehead skin surface on people of all ages; for consumers and healthcare providers.Same
Thermometer typeDigital thermometerDigital thermometerSame
SensorThermopileThermopileSame
Operational PrincipleInfrared radiation detectionInfrared radiation detectionSame
Display typeLCDLCDSame
Measurement siteForehead and earForehead and earSame
Measurement RangeEar & Forehead: 32.0°C43.0°C (89.6°F109.4°F)Ear & Forehead: 32.0°C43.0°C (89.6°F109.4°F)Same
Measurement AccuracyPredicate: ±0.4°F (0.2°C) within 95107.6°F (35.042.0°C), ±0.5°F (0.3°C) for other range.Standard: ISO80601-2-56 and ASTM E1965-98 requirements.34.0°C43.0°C (89.6°F109.4°F): ±0.2°C(0.4°F)32.0 °C34.0°C(89.6°F93.2°F): ±0.3°C(0.5°F)Comparable (meets ISO80601-2-56 and ASTM E1965-98)
Display Resolution0.1°C (0.1°F)0.1°C (0.1°F)Same
Power Supplyd.c. 3.0V (2pcs AAA batteries)d.c. 3.0V (2pcs AAA batteries)Same
Measurement Time1 second1 secondSame
Measurement data memories20 sets35 setsComparable (software verified, no safety/effectiveness issues)
Beeper settingYesYesSame
Date and time settingNoNoSame
BacklightYesYesSame
Auto-off time30s15s±1sSimilar (meets IEC 60601-1:2005+AMD 1:2012, no safety/effectiveness issues)
Operation Condition (Temperature)10°C~40°C10°C~40°CSame
Operation Condition (Relative Humidity)≤85%≤95%Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98)
Operation Condition (Atmospheric Pressure)70kPa~106kPa70kPa~106kPaSame
Storage & Transportation Temp.-20°C~55°C-20°C~50°CComparable (no safety/effectiveness issues)
Storage & Transportation Humidity15%~93%≤95%Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98)
Storage & Transportation Pressure70kPa~106kPa70kPa~106kPaSame
Protection against electric shockExternally power supply, Type BFExternally power supply, Type BFSame
IP ClassIP22IP22Same
Mode of operationContinuous OperationContinuous OperationSame
Materials of skin-contacting componentsABSABS/ PA757(+); PC/PC-110Comparable (tested per IEC 60601-1-2 EMC, IEC 60601-1, ISO 80601-2-56, ASTM E1965-98 and ISO 10993-1)
Physical DimensionsNot available (for predicate)HT-101: 1653850 mm; HT-102: 1563744mm; HT-103: 1555238 mm; HT-104: 1543950mmComparable (no new safety/effectiveness issues)
WeightNot available (for predicate)HT-101: 62.5g (Battery excluded); HT-102: 58g (Battery excluded); HT-103: 64.5g (Battery excluded); HT-104: 63g (Battery excluded)Comparable (no new safety/effectiveness issues)
Electromagnetic CompatibilityMeets IEC 60601-1-2 requirementsPassed testing according to IEC 60601-1-2:2014, IEC 60601-1:2005+AMD 1:2012, IEC 60601-1-11:2015Compliance demonstrated
Electrical SafetyMeets IEC 60601-1 requirementsPassed testing according to IEC 60601-1:2005+AMD 1:2012Compliance demonstrated
BiocompatibilityMeets ISO 10993 requirementsPassed testing according to ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2010 (irritation and skin sensitization)Compliance demonstrated
Performance (Bench)Meets ISO 80601-2-56 and ASTM E1965-98 requirementsPassed testing according to ISO 80601-2-56:2017+AMD2018 and ASTM E1965-98 (Reapproved 2016)Compliance demonstrated
Software Verification & ValidationModerate level of concern, all software requirements met, hazards mitigated to acceptable risk levels.Software documentation consistent with moderate level of concern, system validation testing demonstrates all software requirement specifications met and all software hazards mitigated to acceptable risk levels.Compliance demonstrated
Clinical AccuracyMeet requirements for clinical bias and clinical repeatability per ASTM E1965-98 (Reapproved 2016)Clinical accuracy validation test report demonstrates that the subject devices met the requirements for clinical bias and clinical repeatability per ASTM E1965-98(Reapproved 2016).Compliance demonstrated

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 150 subjects were included in the clinical accuracy validation test.
    • Infants: 50 subjects (newborn to one year)
    • Children: 50 subjects (greater than one to five years)
    • Adults: 50 subjects (greater than five years old)
  • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. It is a clinical accuracy validation test, which implies prospective data collection for the purpose of validating the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide details on the number of experts or their qualifications for establishing ground truth during the clinical accuracy validation. It only states that the clinical accuracy validation testing was conducted according to ASTM E1965-98 (Reapproved 2016), which implies standard clinical measurement procedures and potentially comparison to a reference thermometer.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (such as 2+1 or 3+1). The testing focused on comparing the device's readings against established clinical accuracy standards (ASTM E1965-98), which inherently define acceptable differences from a reference method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an infrared thermometer, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable in the context of an infrared thermometer. The device itself is the standalone measurement tool. Its performance is evaluated both on the bench (bench testing) and in clinical settings (clinical accuracy validation), where it is operated by a human user to take a measurement. However, there is no separate "algorithm only" performance that needs to be assessed, as the device's function is direct temperature measurement.

7. The Type of Ground Truth Used

For the clinical accuracy validation, the ground truth was implied by the reference measurements taken in accordance with ASTM E1965-98 (Reapproved 2016). This standard typically defines the reference method for temperature measurement against which the test device's accuracy (bias and repeatability) is assessed. It's not expert consensus, pathology, or outcomes data in the traditional sense, but rather a standard measurement reference.

8. The Sample Size for the Training Set

The document does not mention a training set as this device is a hardware medical device (infrared thermometer), not an AI/machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

June 17, 2021

Shenzhen Sonida Digital Technology Co., Ltd % Yoyo Chen Consultant Shenzhen Joyantech Consulting Co.,Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town Shenzhen, Guangdong 518000 China

Re: K202440

Trade/Device Name: Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 8, 2020 Received: December 8, 2020

Dear Yoyo Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202440

Device Name Infrared Thermometer, Models HT-101, HT-102, HT-103, HT-104

Indications for Use (Describe)

The infrared thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It can be used by consumers in the household environment and by healthcare providers.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K202440 510(k) Summary

1. Administrative Information

Prepared DateJune 14, 2021
ManufacturerinformationShenzhen Sonida Digital Technology Co., LtdAddress:6F./3F-B., Building B, Zhengchangda Technopark, TangweiJianan Road, Fuhai Street, Baoan, Shenzhen, Guangdong,China, 518103
Contact person: Vivi Wey
TEL Number: +(86) 0755-29607298
FAX Number: +(86) 0755-85259480
E-Mail: 2609815717@qq.com
SubmissionCorrespondentShenzhen Joyantech Consulting Co., Ltd.1713A, 17th Floor, Block A, Zhongguan Times Square, LiuxianAvenue, Xili Town, Nanshan District, Shenzhen, GuangdongProvince, China.
Contact person: Ms. Yoyo Chen
E-Mail: yoyo@cefda.com; field@cefda.com
Image: Green logo with text "卓远天成" below it

2. Device Information

Type of 510(k) Submission:Traditional
Device Name:Infrared Thermometer
Model:HT-101, HT-102, HT-103, HT-104
Classification Name:Thermometer, Electronic, Clinical
Device Class:Class II
Regulation Number:21 CFR 880.2910
Product Code:FLL

{4}------------------------------------------------

3. Predicate Device

ManufacturerGuangzhou Berrcom Medical Device Co., Ltd.
Device nameInfrared Thermometer
ModelMD-H30
510(K) Number:K191570
Product CodeFLL

4. Device Description

HT series infrared thermometers (includes model HT-101, HT-102, HT-103 and HT-104) measure the body temperature through receiving infrared energy radiation via the ear or forehead. These thermometers have the capability to measure temperature via forehead temperature mode or ear temperature mode, and the temperature is directly shown on the LCD display. These thermometers have the following features:

  • The device is intended to be used by consumers in the household environments and 1) by healthcare providers;
    1. The device is mainly composed of infrared sensor, signal receiving processor, buttons, buzzer, LCD display. It is powered by 2*1.5 AAA batteries;
    1. It focuses the infrared energy received from the human's forehead or ear by using the Fresnel lens of the thermometer;
  • Switching between ℃ and ºF; 4)
    1. Multi-functional, can measure ear and forehead temperature;
  • Automatic range selection; resolution is 0.1℃ (0.1℃F); 6)
    1. The latest 35 measurement data can be memorized and stored; the user can view the previous measurement results;
    1. High temperature alarm function, displayed in green, orange and red backlight;
  • Switching between mute and un-mute mode; 9)
    1. Low battery indication, and auto shut-down.

5. Intended Use/Indication for Use

The infrared thermometer is intended for the intermittent measurement of body temperature from the auditory canal or central forehead skin surface on people of all ages. It can be used by consumers in the household environment and by healthcare providers.

6. Substantial Equivalence Comparison

ItemsSubject DevicesPredicate Device(K191570)Comments
Product CodeFLLFLLSame
ItemsSubject DevicesPredicate Device(K191570)Comments
Regulation number880.2910880.2910Same
ManufacturerShenzhen Sonida DigitalTechnology Co., LtdGuangzhou Berrcom MedicalDevice Co., Ltd.Same
Indications for useThe infrared thermometer isintended for the intermittentmeasurement of bodytemperature from the auditorycanal or central forehead skinsurface on people of all ages. Itcan be used by consumers in thehousehold environment and byhealthcare providers.The infrared thermometer isintended for the intermittentmeasurement of bodytemperature from the auditorycanal or central forehead skinsurface on people of all ages.It can be used by consumersin the household environmentand by healthcare providers.Same
Thermometer typeDigital thermometerDigital thermometerSame
SensorThermopileThermopileSame
OperationalprincipleInfrared radiation detectionInfrared radiation detectionSame
Display typeLCDLCDSame
Measurement siteForehead and earForehead and earSame
MeasurementRangeEar & Forehead: 32.0°C43.0°C(89.6°F109.4°F)Ear & Forehead:32.0°C43.0°C(89.6°F109.4°F)Same
Measurementaccuracy34.0°C43.0°C(89.6°F109.4°F): ±0.2°C(0.4°F)32.0 °C34.0°C(89.6°F93.2°F):±0.3°C(0.5°F)±0.4°F (0.2°C) within95107.6°F (35.042.0°C),±0.5°F (0.3°C) for otherrange.Comparable(Note 1)
Display resolution0.1°C (0.1°F)0.1°C (0.1°F)Same
Power supplyd.c.3.0V (2pcs AAA batteries)2pcs AAA batteriesSame
Measurement time1 second1 secondSame
Measurement datamemories35 sets20 setsComparable(Note 2)
Beeper settingYesYesSame
Date and timesettingNoNoSame
BacklightYesYesSame
Auto-off time15s±1s30sSimilar(Note 3)
OperationConditionTemperature: 10°C40°CRelative Humidity: ≤95%Atmospheric Pressure:70kPa106kPaTemperature: 10°C40°CRelative Humidity: ≤85%AtmosphericPressure:70kPa106kPaComparable(Note 4)
Storage andTemperature: -20°C~50°CTemperature: -20°C~55°CComparable
ItemsSubject DevicesPredicate Device (K191570)Comments
transportationconditionRelative Humidity: ≤95%Atmospheric Pressure:70kpa~106kpaRelative Humidity: 15%93%Atmospheric Pressure:70kpa106kpa(Note 4)
Protection againstelectric shockExternally power supply, Type BFExternally power supply, Type BFSame
IP ClassIP22IP22Same
Mode of operationContinuous OperationContinuous OperationSame
Materials of skin-contactingcomponentsABS/ PA757(+);PC/PC-110ABSComparable(Note 5)
PhysicalDimensionsHT-101:1653850 mm;HT-102:1563744mm;HT-103:1555238 mm;HT-104:1543950mmNot availableComparable(Note 5)
WeightHT-101: 62.5g (Battery excluded)HT-102: 58g (Battery excluded)HT-103: 64.5g (Battery excluded)HT-104: 63g (Battery excluded)Not availableComparable(Note 5)

{5}------------------------------------------------

{6}------------------------------------------------

Note 1: Measurement accuracy:

The measurement accuracy of the subject devices has been verified to meet the requirements of ISO80601-2-56 and ASTM E1965-98 standard. The difference does not raise different safety and effectiveness issues.

Note 2: Measurement data memories

The purpose of function of measurement data memories is intended to store and view the previous readings. This function has been verified during software verification. The difference does not raise any issues on the device safety and effectiveness.

Note 3: Auto-off time

Although the time of auto-off time is less than the predicate device, both the subject devices and predicate device meet the basic safety requirement of IEC 60601-1:2005+AMD 1: 2012. The difference itself does not raise any issues on the device safety and effectiveness.

Note 4: Operation, Storage and transportation condition

The differences in operating and storage does not raise any safety and effectiveness questions. The subject device has been tested according to IEC 60601-1-2 EMC, IEC 60601-1, and ISO 80601-2-56 and ASTM E1965-98 standards. Therefore, the difference does not raise any issues on the device safety and effectiveness.

Note 5: Materials of skin-contacting components, Physical Dimensions, and weight The subject devices and predicate device have different appearances, but the difference does not raise any new safety and effectiveness questions. This has been tested and confirmed according to IEC 60601-1-2 EMC, IEC 60601-1, and ISO 80601-2-56, ASTM E1965-98 and ISO 10993-1 standards.

{7}------------------------------------------------

7. Non-Clinical Test Summary

7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject devices have passed safety testing in according to following standards.

    1. IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    1. IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

7.2. Biocompatibility Test

The subject devices have passed biocompatibility tests in according to following standards.

  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro 1) cytotoxicity
    1. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

7.3. Performance Test-Bench

The subject devices have passed performance testing in according to following standards.

    1. ISO 80601-2-56:2017+AMD2018 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers 2) for Intermittent Determination of Patient Temperature

7.4. Software verification and validation

Software documentation consistent with moderate level of concern is submitted in this 510(k). System validation testing presented in this 510(k) demonstrates that all software requirement specifications are met and all software hazards have been mitiqated to acceptable risk levels

8. Clinical Accuracy Validation

Clinical accuracy validation testing was conducted according to ASTM E1965-98(Reapproved 2016). The clinical accuracy validation test report included temperature readings of 150 subjects, of which 50 subjects were infants, 50 subjects were children and the rest of the 50 subjects were adults (NOTE: Infants---newborn to one year; Children--greater than one to five years; Adults---greater than five years old.). The clinical accuracy validation test report demonstrates that the subject devices met the requirements for

{8}------------------------------------------------

clinical bias and clinical repeatability per ASTM E1965-98(Reapproved 2016).

9. Conclusion

The Infrared Thermometer (Model HT-101, HT-102, HT-103, HT-104) is substantially equivalent to the predicate device (K191570). This conclusion is based upon comparison of intended use, technological characteristics and applicable safety standards. Any differences in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.