K Number
K201582

Validate with FDA (Live)

Date Cleared
2021-01-15

(218 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

Device Description

Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is a hand-held, battery powered IR thermometer. It is intended to measure the temperature of human body from forehead without contact to human body. It is indicated for use by people of all ages in the home.

The work principle of Non-contact Electronic Forehead Infrared Thermometer, Model FT3010, is by using the infrared sensor converts the radiated power into an electrical signal, electrical signal is processed by the ambient temperature compensation circuit and an internal MCU circuit. The electrical signal is displayed in degrees Celsius (or Fahrenheit) on LCD.

The Non-contact Electronic Forehead Infrared Thermometer has the following features:

  • Measure the forehead temperature;
  • . Provide the prompt tone for high or low body temperature alert;
  • . Equip with 32 sets of measurement memories;
  • LCD digital displayer with backlight;
  • . Display unit of Fahrenheit or Celsius;
  • Automatic shutdown feature to save energy;
  • . Low battery detection;
AI/ML Overview

The provided document describes the acceptance criteria and the study that proves the performance of the "Non-contact Electronic Forehead Infrared Thermometer. Model: FT3010."

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document references conformity with several standards, which implicitly define the acceptance criteria. The performance is reported as meeting these standards.

Standard / CriteriaAcceptance CriteriaReported Device Performance
Measurement AccuracyAs per ISO 80601-2-56:2017 and ASTM E1965-98. Specifically: - $\pm 0.2$ °C for $35.0 \sim 42.0$ °C - $\pm 0.3$ °C for $34$°C $\sim 34.9$°C and $42.1$°C $\sim 43$°C (based on predicate criteria, assumed similar for subject device based on "does not affect performance and accuracy")Body temperature: $22.0$ °C ~ $43.0$ °C: $\pm 0.2$ °C $71.6$°F ~ $109.4$°F: $\pm 0.4$°F (The document states "The difference does not affect the performance and accuracy which was evaluated in the performance testing of ISO 80601-2-56 and ASTM E1965-98.")
Measurement RangeAs per ISO 80601-2-56:2017 and ASTM E1965-98. Specifically: Body mode: $34.0$ °C -$43.0$ °C ($93.2-109.4$ °F) (Based on predicate criteria)Body temperature: $22.0 \sim 43.0$ °C ($71.6 \sim 109.4$ °F) (The document states "Wider body temperature measurement range... The difference does not affect the performance and accuracy which was evaluated in the performance testing of ISO 80601-2-56 and ASTM E1965-98.")
Electrical SafetyIEC 60601-1:2005+AMD 1: 2012Passed safety testing in accordance to IEC 60601-1:2005+AMD 1: 2012
Electromagnetic CompatibilityIEC 60601-1-2:2014Passed safety testing in accordance to IEC 60601-1-2:2014
Home Healthcare EnvironmentIEC 60601-1-11:2015Passed safety testing in accordance to IEC 60601-1-11:2015
Biocompatibility - CytotoxicityISO 10993-5:2009Passed biocompatibility tests in accordance to ISO 10993-5:2009
Biocompatibility - Irritation & SensitizationISO 10993-10:2010Passed biocompatibility tests in accordance to ISO 10993-10:2010
Clinical Performance (Bias & Repeatability)Met the acceptance criteria of the clinical study protocol (as per ASTM E1965-98)Clinical data, represented by clinical bias and clinical repeatability, met the acceptance criteria of the clinical study protocol.
Software ValidationConsistent with moderate level of concern as per FDA guidanceSoftware documentation consistent with moderate level of concern was submitted.

2. Sample size used for the test set and the data provenance

  • Sample size: 150 subjects
    • 50 infants (newborn to 1 year)
    • 50 children (1 - 5 years old)
    • 50 subjects (> 5 years old)
  • Data provenance: Not explicitly stated (e.g., country of origin). The study is described as a "clinical investigation." It does not specify if it was retrospective or prospective, but clinical investigations are typically prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It mentions a "clinical study" conducted in accordance with ASTM E1965-98 and ISO 80601-2-56, which are standards for clinical thermometers and imply the use of reference thermometers for ground truth.

4. Adjudication method for the test set

The document does not mention an adjudication method. It describes the clinical study as a "randomization, simple blind homologous control, pairing design of clinical investigation."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a standalone infrared thermometer, not an AI-assisted diagnostic tool that involves human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone study was done. The entire document describes the performance of the device itself (Non-contact Electronic Forehead Infrared Thermometer), which operates as an algorithm-only device without human intervention beyond taking the measurement and reading the display. The clinical study and non-clinical bench tests evaluated the device's performance directly.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the clinical study would have been established using reference temperature measurement devices as per the requirements of ASTM E1965-98 and ISO 80601-2-56. These standards typically require comparison against highly accurate clinical reference thermometers to determine bias and repeatability.

8. The sample size for the training set

Not applicable. This device is a traditional electronic thermometer, not an AI/ML-based device that requires a training set in the conventional sense. The "software validation" mentioned refers to traditional software engineering verification and validation steps, not machine learning model training.

9. How the ground truth for the training set was established

Not applicable, as there is no traditional "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 15, 2021

Zhongshan Jinli Electronic Weighing Equipment Co., Ltd. % Ms. Yoyo Chen Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town Shenzhen, Guangdong 518000 China

Re: K201582

Trade/Device Name: Non-contact Electronic Forehead Infrared Thermometer. Model: FT3010 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 10, 2020 Received: December 18, 2020

Dear Ms. Yoyo Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201582

Device Name

Non-contact Electronic Forehead Infrared Thermometer, Model: FT3010

Indications for Use (Describe)

The Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K201582

This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.

1. Administrative Information


Submission DateMay 27, 2020
ManufacturerinformationZhongshan Jinli Electronic Weighing Equipment Co., Ltd.Address:283rd South Min'an Road, Xiaolan Town, Zhongshan City,Guangdong Province, 528416, P.R.ChinaContact person: Mr. Jerry LiangTEL: +86(0760) 28133793; +86 13928121219FAX: +86(0760) 28133793E-Mail: jerry@kinleehealth.com;
SubmissionCorrespondentShenzhen Joyantech Consulting Co., Ltd.1713A, 17th Floor, Block A, Zhongguan Times Square, LiuxianAvenue, Xili Town, Nanshan District, Shenzhen, GuangdongProvince, China.Contact person: Ms. Yoyo ChenE-Mail: yoyo@cefda.com; field@cefda.com

2. Device Information

Device Name:Non-contact Electronic Forehead Infrared Thermometer
Model:FT3010
Classification Name:Clinical Electronic Thermometer
Review Panel:General Hospital
Device Class:2
Regulation Number:880.2910
Product Code:FLL

{4}------------------------------------------------

3. Predicate Device

ManufacturerMicrolife Intellectual Property GmbH
Device nameMicrolife Non-Contact Infrared Forehead Thermometer
ModelFR1DG1 (NC200)
510(K) Number:K191829
Product CodeFLL

4. Device Description

Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is a hand-held, battery powered IR thermometer. It is intended to measure the temperature of human body from forehead without contact to human body. It is indicated for use by people of all ages in the home.

The work principle of Non-contact Electronic Forehead Infrared Thermometer, Model FT3010, is by using the infrared sensor converts the radiated power into an electrical signal, electrical signal is processed by the ambient temperature compensation circuit and an internal MCU circuit. The electrical signal is displayed in degrees Celsius (or Fahrenheit) on LCD.

The Non-contact Electronic Forehead Infrared Thermometer has the following features:

  • Measure the forehead temperature;
  • . Provide the prompt tone for high or low body temperature alert;
  • . Equip with 32 sets of measurement memories;
  • LCD digital displayer with backlight;
  • . Display unit of Fahrenheit or Celsius;
  • Automatic shutdown feature to save energy;
  • . Low battery detection;

5. Indications for Use

The Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

6. Comparison with predicate device

The subject device Non-contact Electronic Forehead Infrared Thermometer, Model FT3010 is substantially equivalent to the predicate device (K191829). The comparison of technological characteristics between the subject device and predicate device is listed as follows:

{5}------------------------------------------------

ItemsSubject Device(K201582)Predicate Device(K191829)Comparison
Intended useThe Non-contact ElectronicForehead InfraredThermometer, ModelFT3010 is intended for theintermittent measurementand monitoring of humanbody temperature. Thedevice is indicated for useby people of all ages in thehome.The Microlife Non-ContactInfrared ForeheadThermometer, ModelFR1DG1 (NC200) isintended for the intermittentmeasurement andmonitoring of human bodytemperature.The device is indicated foruse by people of all ages inthe home.Same
ThermometertypeInfrared thermometerNon-contactInfrared thermometerNon-contactSame
DeviceMeasurementTechnologyInfrared radiation detectionInfrared radiation detectionSame
MeasurementlocationForeheadForeheadSame
MeasurementRangeBody temperature:$22.0\sim43.0$ °C(71.6~109.4°F);Body mode:$34.0$ °C -43.0 °C(93.2-109.4 °F);DifferentNote (1)
MeasurementaccuracyBody temperature:$22.0$ °C ~43.0 °C: ±0.2 °C71.6°F ~109.4°F: ±0.4°F±0.2 °C: 35.0 ~ 42.0 °C±0.3 °C: 34°C ~ 34.9°C,42.1°C ~43°C,±0.4 °F: 95.0 ~ 107.6 °F,±0.5 °F: 93.2 94.8 °F,107.8109.4 °FDifferentNote (1)
TemperatureMeasurementdistance30mm~50mmAppropriate within 5 cmSimilar
Display TypeLCD DisplayLCD DisplaySame
Displayresolution0.1°C (0.1°F)0.1°C (0.1°F)Same
Power supplyDC 3V (2 x AA LR6batteries)DC 3V (2x 1.5V AAAbatteries)DifferentNote (2)
Measurementtime≤ 1second≤ 3 secondDifferentNote (3)
Measurementdata memories32 sets memories30 sets memoriesDifferentNote (4)
ItemsSubject Device(K201582)Predicate Device(K191829)Comparison
Beeper settingYesYesSame
Date and timesettingNoYesDifferentNote (5)
BacklightWhite backlightGreen and red backlightaccording to the measuredtemperature;DifferentNote (6)
Auto-off timeApprox. 10 second after lastmeasurement has beentakenApprox. 1 minute after lastmeasurement has beentakenDifferentNote (7)
OperationConditionAmbient Temperature:16°C40°C (60.8°F104°F)Relative humidity:15%~95%RH (non-condense)Atmospheric pressure: 860-1060hPaAmbient Temperature:15°C40°C (59°F104°F);Relative humidity:15%~95%RHDifferentNote (8)
Storage andtransportationconditionAmbient Temperature:-20°C50°C (-4°F122°F)Relative humidity:15%~95%RH (non-condense)Atmospheric pressure: 860-1060hPaAmbient Temperature:-25°C55°C(-13°F131°F);Relative humidity:15%~95%RHDifferentNote (8)
IP ClassIP22IP22Same
ErrorDisplay Err when systemhas malfunctionDisplay Er0 or Er6 whensystem has malfunctionDifferentNote (9)
Hightemperaturealarm3 short beeps and LCDdisplay " H " while equal toor higher than 37.5°C10 short beeps and a redLCD backlight alerts that thetemperature equal to orhigher than 37.5°CDifferentNote (10)
AutomeasurementNoThe device can take ameasurement automaticallywhen the device detects thedistance is appropriate within5 cm.DifferentNote (11)
Sensor typeSTP583F55TPS336DifferentNote (12)
HousingmaterialABS/PA-757ABS/PA 707DifferentNote (13)
Button materialABS/PA-757PMMADifferent
ItemsSubject Device(K201582)Predicate Device(K191829)Comparison
IC (IntegratedCircuitry)HY11P13HY11P14DifferentNote (13)
PhysicalDimensions9043148mm156.74347 mmDifferentNote (12)
Safety &PerformanceIEC 60601-1:2005+AMD 1:2012;IEC 60601-1-2:2014;IEC 60601-1-11:2015;ISO 80601-2-56: 2017;ASTM E1965-98.IEC 60601-1:2005+AMD 1:2012;IEC 60601-1-2:2014;IEC 60601-1-11:2015;ISO 80601-2-56: 2017;ASTM E1965-98.SameNote (15)
BiocompatibilityCytotoxicity, ISO 10993-5Skin Irritation, ISO 10993-10Skin Sensitization,ISO 10993-10Cytotoxicity, ISO 10993-5Skin Irritation, ISO 10993-10Skin Sensitization,ISO 10993-10Same
Clinical StudySupportYesYesSame

{6}------------------------------------------------

{7}------------------------------------------------

Note 1: Measurement Range and Accuracy

Compare with the predicate device, the subject device has a wider body temperature measurement range, and the accuracy in the temperature range is different. The difference does not affect the performance and accuracy which was evaluated in the performance testing of ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new safety and effectiveness questions.

Note 2: Power supply

The electrical safety of the subject device complies with standard IEC 60601-1. The difference does not raise new issues on the device safety and effectiveness

Note 3: Measurement time

The measurement time of subject device is much quicker than predicate device, the accuracy of measurement has been validated during performance testing. The difference does not raise new issues on the device safety and effectiveness.

Note 4: Measurement data memories

This function has been verified during software verification. The difference does not raise new issues on the device safety and effectiveness.

Note 5: Date and time setting

Although there is no date and time setting functions for subject device, but both subject device and predicate device are meet the basic safety requirement of IEC 60601-1, ASTM E1965-98, and ISO 80601-2-56. The difference does not raise new issues on the device

{8}------------------------------------------------

safety and effectiveness.

Note 6: Backlight

The subject device and predicate device are all tested in according to ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

Note 7: Auto-off time

Both subject device and predicate device are met the basic safety requirement of ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

Note 8: Operation Condition, Storage and transportation condition

The subject device has been demonstrated to comply with the requirements of electrical safety IEC 60601-1, IEC60601-1-11, and ASTM E1965-98 standard. The difference does not raise new issues on the device safety and effectiveness.

Note 9: Error

Although the icons are different, but the purpose means are the same. It does not affect device's performance, safety and effectiveness.

Note 10: High temperature alarm

Even though the beeping time is different, but the purpose of the alarm is the same. It does not affect device's performance, safety and effectiveness.

Note 11: Auto measurement

The purpose of body temperature measurement can be achieved for subject device and predicate device. The subject device performance complies with ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

Note 12: Sensor type, IC

The subject device was tested to conform with same safety and performance standard IEC 60601-1, ASTM E1965-98, and ISO 80601-2-56. A clinical study was carried out on the subject device in accordance with ASTM E1965-98. The different will not arise new safety and effectiveness issues.

Note 13: Housing material, button material

Although the housing material and button material of the subject device and predicate device are different, but they are all compliance with the biocompatibility standards ISO 10993-5 and ISO 10993-10. The difference does not raise new issues on the device safety and effectiveness.

{9}------------------------------------------------

Note 15: Physical Dimensions

The appearance of the subject device and predicate device is different. The subject device has been tested and confirmed according to IEC 60601-1-2, IEC 60601-1, and ISO 80601-2-56 standards. The difference does not raise new issues on the device safety and effectiveness.

7. Non-Clinical Test Summary

7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject device has passed safety testing in according to following standards.

    1. IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    1. IEC 60601-1-11:2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

7.2. Biocompatibility Test

The subject device has passed biocompatibility tests in according to following standards.

    1. ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
    1. ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization

7.3. Performance Test-Bench

The subject device has passed performance testing in according to following standard.

    1. ISO 80601-2-56:2017 Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
    1. ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature

7.4. Software Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k) in according to FDA guidance - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.

{10}------------------------------------------------

8. Clinical Study

Clinical study was conducted in according to ASTM E1965-98(Reapproved 2016). This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consisting of 150 subjects, of which 50 subjects are infants (newborn to 1 year), 50 subjects are children (1 - 5 years old) and 50 subjects are > 5 years old. The clinical test report demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical study protocol.

9. Conclusion

Based on the performance testing, technological characteristics and analysis, the subject device is substantially equivalent to the predicate device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.