K Number
K201536

Validate with FDA (Live)

Date Cleared
2021-02-10

(247 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The infrared forehead thermometer is non-sterile, reusable, non-contact and handheld device for the intermittent measurement and monitoring of human body temperature from the center of the forehead from a measurement distance of 3-5 cm. It can be used by consumers in homecare environment and doctors in clinic. It is intended for measuring human body temperature of people of all ages.

Device Description

The infrared forehead thermometer, Model of QY-EWQ-02 is a hand-held, battery powered IR thermometer. It is intended to measure the temperature of human body from forehead without contact to human body. It is indicated for use by people of all ages in the home. The infrared forehead thermometer has the following features: 1) This device can be used for measuring the forehead temperature; 2) Body and object temperature switching 3) Two temperature unit conversion, °C or °F; 4) LCD display with backlight; 5) Automatic range selection; resolution is 0.1℃ (0.1℃F) 6) The latest 20 measurement data can be memorized and stored; Three color backlight display (Red, Orange, Green); 7) 8) Low battery detection; Turn on/off the prompt tone; 9) 10) High temperature alarm limit setting; 11) Ambient temperature detection; 12) Malfunction indication; 13) Low and high temperature alarm

AI/ML Overview

Here's an analysis of the provided text regarding the Infrared Forehead Thermometer (Model QY-EWQ-02).

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the infrared forehead thermometer are primarily based on established international standards for clinical thermometers. The performance claims for the subject device are listed in comparison to a predicate device.

Acceptance Criteria (from Standards)Reported Device Performance (Subject Device K201536)
Measurement Accuracy:Clinical Accuracy (ASTM E1965-98): Clinical bias and clinical repeatability met the acceptance criteria of the clinical study protocol. Bench Accuracy: ±0.2 °C (32.0°C ~ 43.0 °C); ±0.4 °F (89.6°F~109.4°F)
Clinical Repeatability:≤0.2 °C (0.4 °F)
Measurement Range:32.0°C 43.0°C (89.6°F109.4°F)
Display Resolution:0.1°C ( 0.1°F)
Measurement Time:≤ 1 second
Biocompatibility:Passed ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and skin sensitization)
Electrical Safety:Passed IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11
Performance (General):Passed ISO 80601-2-56:2017+AMD2018 and ASTM E1965-98 (Reapproved 2016) Bench Testing

Study Proving Device Meets Acceptance Criteria:

A clinical accuracy validation study was conducted according to ASTM E1965-98 (Reapproved 2016). This study demonstrated that the clinical data, represented by clinical bias and clinical repeatability, met the acceptance criteria defined in the clinical study protocol.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: 150 subjects.
    • 50 infants (newborn to one year)
    • 50 children (greater than one to five years)
    • 50 adults (greater than five years old)
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study data. It is a "randomization, simple blind homologous control, pairing design of clinical investigation," which implies a prospective study.

3. Number of Experts and Qualifications for Ground Truth of Test Set

The document does not provide details on the number of experts used to establish the ground truth or their specific qualifications for the clinical study referenced in section 8. For thermometer accuracy studies, ground truth typically involves a highly accurate reference thermometer measurement, often taken rectally or orally, which would be performed by trained medical staff.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set. The study is described as a "simple blind homologous control, pairing design," which focuses on comparing the device's readings to a reference standard rather than an expert consensus process requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document describes a clinical accuracy validation study for a medical device (thermometer), not an AI-powered diagnostic tool that would typically involve human readers. Therefore, an MRMC study and a human reader improvement effect size are not applicable in this context.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone performance study was conducted. The "Clinical Accuracy Validation Study" (Section 8) and "Performance Test-Bench" (Section 7.3) directly assess the device's performance based on its inherent functionality, without human interpretation or intervention in the temperature measurement itself. The device measures and displays a temperature value.

7. Type of Ground Truth Used

The ground truth used for the clinical accuracy validation study would have been established by a reference clinical thermometer measurement (e.g., rectal or oral temperature) known for its high accuracy, as is standard practice in ASTM E1965-98. The document states the study "met the acceptance criteria of the clinical study protocol," which inherently relies on comparison to a gold standard of temperature measurement.

8. Sample Size for the Training Set

The document does not mention a training set or its sample size. This device is an infrared thermometer, which is a hardware device with embedded firmware/software for measurement and display, rather than a machine learning or AI algorithm that typically requires a large training dataset. The "Software Validation" mentioned in Section 7.4 refers to standard software engineering validation practices, not machine learning model training.

9. How Ground Truth for the Training Set Was Established

As no training set is discussed or implied for an AI/ML algorithm, this question is not applicable. The device's functionality is based on established physics and pre-programmed algorithms, validated against physical standards and clinical performance, not by learning from a training dataset in the AI sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2021

Hangzhou Qingyuan Medical Equipment Technology Co., Ltd. % Yoyo Chen Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue Xili Town Shenzhen, Guangdong 518000 China

Re: K201536

Trade/Device Name: Infrared Forehead Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 28, 2020 Received: January 4, 2021

Dear Yoyo Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201536

Device Name Infrared Forehead Thermometer, Model QY-EWQ-02

Indications for Use (Describe)

The infrared forehead thermometer is non-sterile, reusable, non-contact and handheld device for the intermittent measurement and monitoring of human body temperature from the center of the forehead from a measurement distance of 3-5 cm. It can be used by consumers in homecare environment and doctors in clinic. It is intended for measuring human body temperature of people of all ages.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201536 510(k) Summary

This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.

1. Administrative Information

Submission DateFebruary 10, 2021
Manufacturer informationHangzhou Qingyuan Medical Equipment Technology Co., Ltd.Address:No.688 Dong'an 1st Road, Xintang Street, Xiaoshan District,Hangzhou, China 311200Contact person: JennyTEL: +86(0571) 82623931E-Mail: Jenny@aodicharger.com;
Submission CorrespondentShenzhen Joyantech Consulting Co., Ltd.1713A, 17th Floor, Block A, Zhongguan Times Square, LiuxianAvenue, Xili Town, Nanshan District, Shenzhen, GuangdongProvince, China. 518000Contact person: Ms. Yoyo ChenTEL:+86(755) 86069197E-Mail: yoyo@cefda.com; field@cefda.comImage: Logo
Establishment registration number3016658659

2. Device Information

Device Name:Infrared Forehead Thermometer
Model:QY-EWQ-02
Classification Name:Clinical Electronic Thermometer (Infrared Thermometer)
Review Panel:General Hospital
Device Class:2
Regulation Number:21 CFR 880.2910
Product Code:FLL

{4}------------------------------------------------

3. Predicate Device

ManufacturerMicrolife Intellectual Property GmbH
Device nameMicrolife Non-Contact Infrared Forehead Thermometer
ModelFR1DG1 (NC200)
510(K) Number:K191829
Product CodeFLL

4. Device Description

The infrared forehead thermometer, Model of QY-EWQ-02 is a hand-held, battery powered IR thermometer. It is intended to measure the temperature of human body from forehead without contact to human body. It is indicated for use by people of all ages in thehome.

The infrared forehead thermometer has the following features:

    1. This device can be used for measuring the forehead temperature;
    1. Body and object temperature switching
    1. Two temperature unit conversion, °C or °F;
    1. LCD display with backlight;
    1. Automatic range selection; resolution is 0.1℃ (0.1℃F)
    1. The latest 20 measurement data can be memorized and stored;
  • Three color backlight display (Red, Orange, Green); 7)
    1. Low battery detection;
  • Turn on/off the prompt tone; റി
    1. High temperature alarm limit setting;
    1. Ambient temperature detection;
    1. Malfunction indication;
    1. Low and high temperature alarm

5. Indication for Use

The infrared forehead thermometer is non-sterile, reusable, non-contact and handheld device for the intermittent measurement and monitoring of human body temperature from the center of the forehead from a measurement distance of 3-5 cm. It can be used by consumers in homecare environment and doctors in clinic. It is intended for measuring human body temperature of people of all ages.

6. Comparison with predicate device

The subject device Infrared Forehead Thermometer (Model QY-EWQ-02) is substantially equivalent to the predicate device (K191829). The comparison of technological characteristics between the subject device and predicate device is listed as follows:

{5}------------------------------------------------

ItemsSubject Device(K201536)Predicate Device(K191829)Comparison
Intended useThe infrared foreheadthermometer is intended forthe intermittent measurementand monitoring of humanbody temperature. The deviceis indicated for use by peopleof all ages in the home.The Microlife Non-ContactInfrared ForeheadThermometer, ModelFR1DG1 (NC200) is intendedfor the intermittentmeasurement and monitoringof human body temperature.The device is indicated for useby people of all ages in thehome.Same
ThermometertypeInfrared thermometerNon-contactInfrared thermometerNon-contactSame
DeviceMeasurementTechnologyInfrared radiation detectionInfrared radiation detectionSame
MeasurementlocationForeheadForeheadSame
MeasurementRange$32.0°C 43.0°C$$(89.6°F109.4°F);$$32.0°C 43.0°C$$(89.6°F109.4°F);$Same
Measurementaccuracy$±0.2 °C: 32.0°C ~ 43.0 °C$$±0.4 °F: 89.6°F~109.4°F$$±0.2 °C: 35.0 ~ 42.0 °C$$±0.3 °C: 34°C ~ 34.9°C,$$42.1°C 43°C,$$±0.4 °F: 95.0 ~ 107.6 °F,$$±0.5 °F: 93.294.8 °F,$$107.8~109.4 °F$Different(Note 1)
ClinicalRepeatability$≤0.2 °C (0.4 °F)$unknownDifferent(Note 1)
TemperatureMeasurementdistance3 to 5 cmAppropriate within 5 cmSimilar
Display TypeLCD DisplayLCD DisplaySame
Displayresolution$0.1°C( 0.1°F)$$0.1°C( 0.1°F)$Same
Power supplyDC 3V(2x 1.5V AAAbatteries)DC 3V (2x 1.5V AAA batteries)Same
Measurementtime$≤ 1second$$≤ 3 second$Different(Note 2)
Measurementdata memories20 sets memories30 sets memoriesDifferent(Note 3)
ItemsSubject Device(K201536)Predicate Device(K191829)Comparison
Beeper settingYesYesSame
Date and timesettingNoYesDifferent(Note 4)
BacklightGreen, yellow and redbacklight according to themeasured temperature;Green and red backlightaccording to the measuredtemperature;Different(Note 5)
Auto-off timeApprox. 13 seconds after lastmeasurement has been takenApprox. 1 minute after lastmeasurement has been takenDifferent(Note 6)
OperationConditionAmbient Temperature:$15°C40°C$ ( $59°F104°F$ );Relative humidity:$15%~95%RH$Ambient Temperature:$15°C40°C$ ( $59°F104°F$ );Relative humidity:$15%~95%RH$Same
Storage andtransportationconditionAmbient Temperature:$-25°C55°C$ ( $-13°F131°F$ );Relative humidity:$15%~95%RH$Ambient Temperature:$-25°C55°C$ ( $-13°F131°F$ );Relative humidity:$15%~95%RH$Same
IP ClassIP22IP22Same
ErrorDisplay Err when system hasmalfunctionDisplay Er0 or Er6 whensystem has malfunctionDifferent(Note 7)
Hightemperaturealarm5 short beeps and a red LCDbacklight alerts that thetemperature equal to orhigher than $37.5°C$10 short beeps and a red LCDbacklight alerts that thetemperature equal to or higherthan $37.5°C$Different(Note 8)
AutomeasurementNoneThe device can take ameasurement automaticallywhen the device detects thedistance is appropriate within5 cm.Different(Note 9)
Sensor typeSGXV02-100-000-100TPS336Different(Note 10)
HousingmaterialABS/PA-757ABS/PA 707Different(Note 11)
Button materialABS/PA-757PMMADifferent(Note 11)
IC (IntegratedCircuitry)STM8I101HY11P14Different(Note 10)
PhysicalDimensions15587.744mm156.74347 mmDifferent(Note 12)
ItemsSubject Device(K201536)Predicate Device(K191829)Comparison
Safety &PerformanceIEC 60601-1:2005+AMD 1:2012;IEC 60601-1-2:2014;IEC 60601-1-11:2015;ISO 80601-2-56: 2018;ASTM E1965-98.IEC 60601-1:2005+AMD 1:2012;IEC 60601-1-2:2014;IEC 60601-1-11:2015;ISO 80601-2-56: 2018;ASTM E1965-98.Same
BiocompatibilityCytotoxicity, ISO 10993-5Skin Irritation, ISO 10993-10Skin Sensitization,ISO 10993-10Cytotoxicity, ISO 10993-5Skin Irritation, ISO 10993-10Skin Sensitization,ISO 10993-10Same
Clinical StudySupportYesYesSame

{6}------------------------------------------------

{7}------------------------------------------------

Note 1: Measurement accuracy, Clinical Repeatability

Compare with predicate device, the measurement accuracy and clinical repeatability is different. But The difference does not affect the performance and accuracy which was evaluated in the performance testing of ISO 80601-2-56 and ASTM E1965-98. The different will not arise new safety and effectiveness issue.

Note 2: Measurement time

The measurement time of subject device is much quicker than predicate device, the accuracy of measurement has been validated during performance testing. The difference does not raise new issues on the device safety and effectiveness.

Note 3: Measurement data memories

This function has been verified during software verification. The difference does not raise new issues on the device safety and effectiveness.

Note 4: Date and time setting

Although there is no date and time setting functions for subject device, but both subject device and predicate device are meet the basic safety requirement of IEC 60601-1, ASTM E1965-98, and ISO 80601-2-56. The difference does not raise new issues on the device safety and effectiveness.

Note 5: Backlight

The subject device and predicate device are all tested in according to ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

Note 6: Auto-off time

Both subject device and predicate device are met the basic safety requirement of ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

{8}------------------------------------------------

Note 7: Error

Although the icons are different, but the purpose means are the same. It does not affect device's performance, safety and effectiveness.

Note 8: High temperature alarm

Even though the beeping time is different, but the purpose means are the same. It does not affect device's performance, safety and effectiveness.

Note 9: Auto measurement

The purpose of body temperature measurement can be achieved for subject device and predicate device. The subject device performance complies with ISO 80601-2-56 and ASTM E1965-98. The difference does not raise new issues on the device safety and effectiveness.

Note 10: Sensor type, IC

The subject device was tested to conform with same safety and performance standard IEC 60601-1, ASTM E1965-98, and ISO 80601-2-56. A clinical study was carried out on the subject device in accordance with ASTM E1965-98. The different will not arise new safety and effectiveness issues.

Note 11: Housing material, Button material

Although the housing material and button material for subject device and predicate device are different, but they are all compliance with the biocompatibility standards ISO 10993-5 and ISO 10993-10. The difference does not raise new issues on the device safety and effectiveness.

Note 12: Physical Dimensions

The appearance of the subject device and predicate device is different. The subject device has been tested and confirmed according to IEC 60601-1-2, IEC 60601-1, and ISO 80601-2-56 standards. The difference does not raise new issues on the device safety and effectiveness.

7. Non-Clinical Test Summary

7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject device has passed safety testing in according to following standards.

    1. IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

{9}------------------------------------------------

    1. IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

7.2. Biocompatibility Test

The subject device has passed biocompatibility tests in according to following standards.

  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro 1) cytotoxicity
    1. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

7.3. Performance Test-Bench

The subject device has passed performance testing in according to following standard.

  • ISO 80601-2-56:2017+AMD2018 Medical electrical equipment Part 2-56: Particular 1) requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
    1. ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature

7.4. Software Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k) in according to FDA guidance - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.

8. Clinical Accuracy Validation Study

Clinical study was conducted in according to ASTM E1965-98(Reapproved 2016). This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consisting of 150 subjects, of which 50 subjects are infants, 50 subjects are children and the rest 50 subjects are adults (NOTE: Infants---newborn to one year; Children--- greater than one to five years; Adults---greater than five years old.). The clinical test report demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical study protocol.

9. Conclusion

Based on the performance testing, technological characteristics and analysis, the subject device is substantially equivalent to the predicate device, Microlife Non-Contact Infrared Forehead Thermometer cleared under K191829.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.