K Number
K201428
Date Cleared
2021-06-24

(391 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Black Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

AI/ML Overview

This document is a 510(k) Pre-Market Notification for a Powder Free Black Nitrile Examination Glove (K201428) and does not describe a study involving an AI/ML powered device. Instead, it details the substantial equivalence of the new nitrile glove to a predicate device (K153562) based on non-clinical performance and biological safety testing.

Therefore, I cannot provide the requested information about acceptance criteria for an AI/ML device, sample sizes for test/training sets, expert qualifications, or MRMC studies.

However, I can extract the acceptance criteria and reported performance for the nitrile examination glove based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance (for the Nitrile Examination Glove):

Characteristic/ParameterAcceptance Criteria (Predicate or Standard)Reported Performance (Proposed Device K201428)Discussion / Outcome
Product CodeLZALZASame
Intended UseTo prevent contamination between patient and examinerTo prevent contamination between patient and examinerSame
ClassificationClass 1Class 1Same
Raw Rubber MaterialNitrile (Acrylonitrile-butadiene)Nitrile (Acrylonitrile-butadiene)Same
Surface AppearanceBlue, Ambidextrous, Finger Textured (Predicate)Black, Ambidextrous, Finger TexturedDifferent color; same design/texture
Freedom of HolesMeet AQL 1.5 at G1 (Predicate) / AQL 2.5 at G1 (ASTM)Meet AQL 1.5 with G1Similar
Overall LengthMinimum 230mmAverage: 242mmSimilar
Width (S, M, L)S: 75-95mm, M: 85-105mm, L: 100-120mmS: 85mm, M: 95mm, L: 104mm (Average)Similar (Meeting specification)
Palm ThicknessMinimum 0.05mmAverage: 0.06mmSimilar
Finger ThicknessMinimum 0.05mmAverage: 0.11mmSimilar
Tensile Strength (before age)Minimum 14 MPaAverage: 17.44 MPaSimilar
Tensile Strength (after age)Minimum 14 MPaAverage: 16.37 MPaSimilar
Ultimate Elongation (before age)Minimum 500%Average: 559%Similar
Ultimate Elongation (after age)Minimum 400%Average: 508%Similar
Residual Powder TestLess than 2mg/gloveS: 0.43 mg/glove, M: 0.31 mg/glove, L: 0.47 mg/glove (Average)Similar (Contained less than 2mg/glove)
Primary Skin IrritationNot an irritantUnder the conditions of study, not an irritantSimilar
Dermal SensitizationNot a sensitizerUnder the conditions of study, not a sensitizerSimilar
CytotoxicityN/A (Predicate)Not CytotoxicDifferent (Pred. info N/A)
Acute Systemic ToxicityN/A (Predicate)Not induce systemic toxicityMeeting requirements per ISO 10993-11

As there is no AI/ML component to the device described in this document, the following points are not applicable:

2. Sample size used for the test set and the data provenance: Not applicable. (Performance claims are based on physical and biological tests of the glove material, not a data-driven test set.)
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. (Ground truth not established in the context of an AI/ML test set.)
4. Adjudication method for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
7. The type of ground truth used: Not applicable in the context of AI/ML. For the glove, the "ground truth" or reference for performance is established by ASTM standards and ISO biological evaluation guidelines.
8. The sample size for the training set: Not applicable.
9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 24, 2021

Vietglove Corporation Terence Lim Quality Assurance & Regulatory Affairs Manager Cau Sat Hamlet, Lai Hung Commune, Bau Bang District, Binh Duong Province Bingh Duong, Bingh Duong Province 72600 Viet Nam

Re: K201428

Trade/Device Name: Powder Free Black Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 25, 2021 Received: May 27, 2021

Dear Terence Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Ryan Ortega, Ph. D Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201428

Device Name POWDER FREE BLACK NITRILE EXAMINATION GLOVE

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (K) SUMMARY

K201428

1.0SubmitterVietglove Corporation
Cau Sat Hamlet, Lai Hung Commune

Cau Sat Hamlet, Lai Hung Commune Bau Bang District Binh Duong Province

Tel:+84-650-591220
Fax:+84-650-591220

2.0

Name of Contact Person: Terence Lim Email Address: limsinkooi@@gmail.com Date of Summary Prepared: May 17, 2021

3.0 Name of Device:

510(k) number K201428 Trade Name: Powder Free Black Nitrile Examination Glove Classification Name: Polymer Patient Examination Glove Device Classification: 21 CFR 880.6250 Regulation Number: General Hospital Panel: Product Code: LZA

4.0 Identification of The Legally Marketed Device

Predicate Device Name: Powder Free Blue Nitrile Examination Glove

Predicate 510(K) Number: K153562

Manufacturer's Name: VIETGLOVE CORPORATION

5.0 Description of Device

Powder Free Black Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

6.0 The Intended Use/Indications for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

7.0 Summary of the Technological Characteristic of the Device compared to the Predicate Device for Substantial Equivalent Discussion.

{4}------------------------------------------------

VIETGLOVE CORPORATION

Cau Sat Hamlet, Lai Hung Commune, Bau Bang District ,Binh Duong Province Vietnam

There is no difference in technology characteristic compared to the predicate device. Gloves are made from nitrile latex compound. Non-Sterile, Powder Free Blue Nitrile Examination Gloves has the below technological characteristic compared to ASTM or Equivalent standards.

Comparison On Technological Characteristic Between The Predicate Device And Subject Devices

Characteristics andParametersProposed Device -Powder Free NitrileExamination Gloves(K201428)Predicate Device -Powder Free BlueNitrile ExaminationGloves (K153562 )Discussion
Product CodeLZALZASame product Code
IntendedUse/Indications for UseA patient examinationglove is a disposabledevice intended formedical purposes thatis worn on the hand orfinger to preventcontamination betweenpatient and examiner.A patient examinationglove is a disposabledevice intended formedical purposes thatis worn on the hand orfinger to preventcontaminationbetween patient andexaminer.Same Intended Use.
ClassificationClass 1Class 1:Same Class.
Raw Rubber MaterialNitrile (Acrylonitrile-butadiene)Nitrile (Acrylonitrile-butadiene)Same synthetic rubbermaterial.
Surface Appearance1.Black2. Ambidextrous3.Finger Textured1.Blue2. Ambidextrous3.Finger Textured1. Different color2. Same ambidextrousdesign3. Same texture area.
Freedom of HolesMeet AQL 2.5 at G1Meet AQL 1.5 withG1Meet AQL 1.5 withG1.Similar
Overall LengthMinimum 230mmAverage : 242mmMore than 230mmSimilar
WidthS : 75mm - 95mmM: 85mm - 105mmL: 100mm – 120mmAverage :S : 85 mmM : 95 mmL : 104 mmMeetingspecificationSimilar
Palm Thickness(Minimum 0.05mm)Average : 0.06mmMore than 0.05mmSimilar
Finger Thickness(Minimum 0.05mm)Average : 0.11mmMore than 0.05mmSimilar
Tensile Strength(before age)Minimum 14 MPaAverage : 17.44 MPaMore than 14 MPaSimilar
Tensile Strength(After Age)Minimum 14 MPaAverage : 16.37 MPaMore than 14 MPaSimilar
Characteristics andParametersProposed Device -Powder Free NitrileExamination Gloves(K201428)Predicate Device -Powder Free BlueNitrile ExaminationGloves (K153562 )Discussion
Ultimate Elongationbefore age(Minimum 500%)Average : 559%Minimum 500%Similar
Ultimate Elongationafter age(Minimum 400%)Average : 508%Minimum 400%Similar
Residual powder test(Less than2mg/glove)Average powderresidue for each size:S : 0.43 mg/gloveM : 0.31 mg/gloveL : 0.47 mg/gloveContained less than2mg/gloveSimilar
Primary Skin IrritationUnder the conditionsof study, not anirritantUnder the conditionsof study, not anirritantSimilar
Dermal SensitizationUnder the conditionsof study, not asensitizer.Under the conditionsof study, not asensitizer.Similar
CytotoxicityCytotoxicN/ADifferent
Acute SystemicToxicityNot induce systemictoxicityN/AMeeting the requirementsper ISO 10993-11

{5}------------------------------------------------

VIETGLOVE CORPORATION

Cau Sat Hamlet, Lai Hung Commune, Bau Bang District ,Binh Duong Province Vietnam

Summary of Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic . toxicitv
  • ASTM D6124-06, Standard Test Method for Residual Powder on Medical Gloves .
  • ASTM D5151-06, Standard Test Method for Detection of Holes in Medical Gloves. ●
  • ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application .
  • ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes Part I: Sampling Plans ● Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

Clinical Test Conclusion

No clinical test is included in this submission.

Conclusion

The conclusion drawn from the nonclinical tests demonstrates the subject device in 510(K) submission K201428, Powder Free Black Nitrile Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K153562.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.