K Number
K201323

Validate with FDA (Live)

Date Cleared
2022-01-25

(617 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INNO SLA Submerged Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw-retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.

Device Description

The INNO SLA Submerged Implant System offers Healing Abutment (Type A and Type B), Angulated Abutment, Temporary Abutment, Sonator S Abutment, Sonator A Abutment, and Screws. These components are used with the fixtures and cover screws cleared under K132242.

AI/ML Overview

The provided text is a 510(k) summary for the INNO SLA Submerged Implant System. It describes the device, its intended use, and argues for its substantial equivalence to previously cleared predicate devices.

However, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets those criteria, particularly in the context of an AI/ML-driven device performance study. The device described, the "INNO SLA Submerged Implant System," is a physical dental implant system and not an AI/ML-based medical device. Therefore, the questions related to AI/ML device performance (like sample size for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance) are not applicable to this document.

The document does mention "Non-Clinical Testing" which includes:

  • Surface Modification: Leveraging specifications and processes from a predicate device (K132242) for TiN coating.
  • Cytotoxicity Testing: Performed on representative abutments with TiN coating according to ANSI/AAMI/ISO 10993-5 and ISO 10993-12. This type of testing ensures the material is not harmful to cells.
  • Mechanical performance testing: Conducted according to ISO 14801. This test establishes the fatigue limit to ensure structural integrity and sufficient strength for intended use. The "worst-case scenario" was chosen based on FDA Guidance.
  • End-user sterility validation: Conducted according to ISO 17665-1 and ISO 17665-2 for a prior clearance (K132242) and leveraged for the subject devices.

These tests are standard for physical medical devices and confirm material compatibility, mechanical robustness, and effective sterilization, which are relevant "acceptance criteria" for such devices. The document explicitly states that the mechanical performance test demonstrated "sufficient strength for their intended use and are substantially equivalent to predicate devices with regard to mechanical performance."

In summary, the requested information regarding AI/ML device performance cannot be extracted from this document because it pertains to a physical dental implant system, not an AI/ML system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

Cowellmedi Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA. Inc. 1150 Roosevelt STE 200 Irvine, California 92620

Re: K201323

Trade/Device Name: INNO SLA Submerged Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: December 21, 2021 Received: December 23, 2021

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201323

Device Name INNO SLA Submerged Implant System

Indications for Use (Describe)

The INNO SLA Submerged Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K201323)

This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: 1/24/2022

1. Submitter

Cowellmedi Co., Ltd. 48, Hakgam-daero 221beon-gil, Sasang-gu, Busan, 46986, Republic of Korea

2. U.S Agent/Contact Person

Priscilla Chung LK Consulting Group USA, Inc. 1150 Roosevelt STE 200, Irvine CA 92620 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com

3. Device

  • Trade Name: INNO SLA Submerged Implant System ●
  • Common Name: Dental Implant System
  • Classification Name: Endosseous Dental Implant
  • Product Code: NHA
  • . Classification regulation: 21CFR872.3630

4. Predicate Device:

  • Primary Predicate Device: INNO SLA Submerged Implant System by Cowellmedi Co., Ltd. (K132242)
  • . Reference Predicate Device: Locator Implant Anchor Abutment for Endosseous Dental Implant by Zest Anchors, Inc. (K072878)

{4}------------------------------------------------

5. Description:

The INNO SLA Submerged Implant System offers the following components. The components below are used with the fixtures and the cover screws that were cleared under K132242.

NoComponent
1Healing AbutmentType AØ7.5 x 10.7mmØ 8.5 x 10.7 mmØ 9.5 x 10.7mm
2Type BØ 6.5 x 5.9, 6.9, 7.9, 8.9, 9.9, 11.9 mmØ 7.5 x 5.9, 6.9, 7.9, 8.9, 9.9, 11.9 mmØ 8.0 x 5.9, 6.9, 7.9, 8.9, 9.9, 11.9 mmØ 8.5 x 5.9, 6.9, 7.9, 8.9, 9.9, 11.9 mmØ 9.5 x 5.9, 6.9, 7.9, 8.9, 9.9, 11.9 mm
3Angulated AbutmentØ 4.5 x 11.9, 12.9, 13.9, 14.9mmØ 4.8 x 9.4, 10.4, 11.4mmØ 5.5 x 11.9, 12.9, 13.9, 14.9mm
5Temporary AbutmentØ 4.5 x 13.65,13.9mm
6Sonator S AbutmentØ 3.87 x 8.3, 9.3, 10.3, 11.3, 12.3, 13.3mm
7Sonator A AbutmentØ 3.87 x 7.15, 8.7mm
8ScrewØ 2.15 x 6.5, 7.5, 8.5mmØ 2.2 x 7.8, 8.8, 9.8mmØ 2.28 x 5mmØ 2.45 x 8.5mm

6. Indication for use:

The INNO SLA Submerged Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw-retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.

{5}------------------------------------------------

7. Basis for Substantial Equivalence

Substantial Equivalence Discussion

The INNO SLA Submerged Implant System is substantially equivalent to the predicate devices in terms of intended use and technical characteristics. They are made of the same material and have similar design. There are slight differences mostly in abutment size options, however, it is very minor not affecting substantial equivalence. We have performed the fatigue test to make sure the differences do not raise an issue in safety and effectiveness and the test result of the test supported substantial equivalence. Based on the information and test results provided in submission, we conclude that the subject device is substantially equivalent to the predicate devices.

Comparison Chart

Subject DevicePredicate Device
510(k) NumberK201323K132242
Device NameHealing Abutment forINNO SLA Submerged ImplantSystemHealing Abutment forINNO SLA Submerged ImplantSystemRemark
ManufacturerCowellmedi Co., Ltd.Cowellmedi Co., Ltd.
Indications forUseThe INNO SLA Submerged ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.The INNO SLA Submerged ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.same
AppearanceImage: Two healing abutmentsImage: A healing abutmentsame and similar
Diameter6.5, 7.5, 8.0, 8.5, 9.5mm4.1, 4.5, 4.9, 5.5, 5.95, 6.5mmlarge diameter
Post Height2, 5mm2, 3, 4, 5, 6, 7, 9mmsame
Gingival Height1, 2, 3, 4, 5, 7mm1, 2, 3, 4, 5, 7mmsame
ConnectionInterfaceHex, Non-hexHex, Non-hexsame
SurfacetreatmentN/AN/Asame
SterilityNon-sterile; intended for terminalsterilization via moistheat(autoclave)Non-sterile; intended for terminalsterilization via moistheat(autoclave)same
Angulationsame
MaterialTi-6Al-4V ELITi-6Al-4V ELIsame
Principle ofoperationThis component is intended toprotect the inner configuration of theimplant during the healing processand maintain, stabilize and form thesoft tissue during this phase.This component is intended toprotect the inner configuration of theimplant during the healing processand maintain, stabilize and form thesoft tissue during this phase.same

● Healing Abutment

{6}------------------------------------------------

The subject device has the same intended use, material, surface treatment, and design as the predicate device. This 510k is to add some more size options for diameter. The wider range is to meet each patient needs and larger diameter does not a concern in performance or safety.

Subject DevicePredicate Device
510(k) NumberK201323K132242
Device NameAngulated Abutment forINNO SLA Submerged ImplantSystemAngulated Abutment forINNO SLA Submerged ImplantSystemRemark
ManufacturerCowellmedi Co., Ltd.Cowellmedi Co., Ltd.
Indications forUseThe INNO SLA Submerged ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.The INNO SLA Submerged ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.same
AppearanceImage: abutmentImage: abutmentsame
Diameter4.5, 4.8, 5.5mm4.8mmadded small andlarge diameter
Post Height5, 8mm3.9, 4.9, 5.9, 6.9mmadded long height
Gingival Height1, 2, 3, 4mm1, 2, 3, 4mmsame
ConnectionInterfaceHex, Non-HexHex, Non-Hexsame
SurfacetreatmentTiN CoatingTiN Coatingsame
SterilityNon-sterile; intended for terminalsterilization via moistheat(autoclave)Non-sterile; intended for terminalsterilization via moistheat(autoclave)same
Angulation15°, 25°15°, 25°same
MaterialTi-6Al-4V ELITi-6Al-4V ELIsame
Principle ofoperationThis component is placed into dentalimplants to provide support forprosthetic restorations. It is forpermanent screw-retained abutment.This component is placed into dentalimplants to provide support forprosthetic restorations. It is forpermanent screw-retained abutment.same
Substantial Equivalence DiscussionThe subject device has the same intended use, material, surface treatment, and design as the predicate device.

● Angulated Abutment

This 510k is to add some more size options for diameter and post height. The wider range is to meet each patient needs and the difference is minor. Also the additional size was verified by performance test.

● Temporary Abutment

Subject DevicePredicate Device
510(k) NumberK201323K132242
Device NameTemporary Abutment forINNO SLA Submerged ImplantSystemTemporary Abutment forINNO SLA Submerged ImplantSystemRemark
ManufacturerCowellmedi Co., Ltd.Cowellmedi Co., Ltd.
Indications for UseThe INNO SLA Submerged ImplantSystem is indicated for use inThe INNO SLA Submerged ImplantSystem is indicated for use insame

{7}------------------------------------------------

partially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.partially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.
AppearanceImage: abutmentImage: abutmentsame
Diameter4.5mm4.8mmadded smalldiameter
Post Height10mm3.9, 4.9, 5.9, 6.9mmadded long height
Gingival Height1mm1, 2, 3, 4mmsame
ConnectionInterfaceHex, Non-HexHex, Non-Hexsame
SurfacetreatmentN/AN/Asame
SterilityNon-sterile; intended for terminalsterilization via moistheat(autoclave)Non-sterile; intended for terminalsterilization via moistheat(autoclave)same
Angulationsame
MaterialTi-6Al-4V ELITi-6Al-4V ELIsame
Principle ofoperationThis component is a provisionalrestoration intended for use withroot-form endosseous dentalimplants to aid in prostheticrehabilitation or provisional crownthat aids in creating an estheticemergence through the gingivaduring the final prostheticproduction.This component is a provisionalrestoration intended for use withroot-form endosseous dentalimplants to aid in prostheticrehabilitation or provisional crownthat aids in creating an estheticemergence through the gingivaduring the final prostheticproduction.same
Substantial Equivalence Discussion
The subject device has the same intended use, material, surface treatment, and design as the predicate device.

This 510k is to add some more size options for diameter and post height. The wider range is to meet each patient needs and the difference is minor.

● Sonator Abutment

Subject DevicePredicate Device
510(k) NumberK201323K072878
Device NameSonator Abutment forINNO SLA Submerged ImplantSystemLocator Implant Anchor Abutmentfor Endosseous Dental ImplantRemark
ManufacturerCowellmedi Co., Ltd.Zest Anchors, Inc.
Indications forUseThe INNO SLA Submerged ImplantSystem is indicated for use inpartially or fully edentulousmandibles and maxillae, in supportof single or multiple-unit restorationsincluding; cemented retained, screw-retained, or overdenture restorations,and terminal or intermediateabutment support for fixedbridgework.The LOCATOR Implant AnchorAbutment for Endosseous DentalImplants is appropriate for use withoverdentures or partial denturesretained in whole or in part byendosseous implants in the mandibleor maxilla.similar

{8}------------------------------------------------

Appearancesimilar
Diameter3.87, 4.5mm3.25 to 6.5mmsame
Gingival Height1, 1.5, 2, 3, 4, 5, 6mm1, 2, 3, 4, 5, 6 mmsame
ConnectionInterfaceHex, Non-hexConical, External Hex, Internal Hex,Internal Multi Lobesimilar
SurfacetreatmentTiN CoatingTiN Coatingsame
SterilityNon-sterile; intended for terminalsterilization via moistheat(autoclave)Non-sterile; intended for terminalsterilization via moistheat(autoclave)same
Angulation0, 15°0°, 10°, 20°same
MaterialTi-6Al-4V ELITi-6Al-4V ELIsame
Principle ofoperationThis component is used for implant-retained muscosa-supportedrestorations, such as dentures. Itsupports implant retained mucosssupported over dentures.This component is used for implant-retained muscosa-supportedrestorations, such as dentures. Itsupports implant retained mucosssupported over dentures.similar
Substantial Equivalence Discussion
The subject device has the same intended use, material, surface treatment, and design as the predicate device. Thewording of the indications for use is different but the overall claims are the same between the devices. The wordingof the subject device is more specific.

The size range of the predicate device encompasses that of the subject device, so it does not raise an issue in performance or safety.

8. Non-Clinical Testing

  • Surface Modification: The TiN coating for the subject abutments has the same ● specifications and processes as the predicate device cleared under K132242. The surface modification information for the TiN coating as recommended in the FDA Guidance, "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" was provided in K132242 and leveraged for the subject devices.
  • Cytotoxicity Testing provided on representative abutments with TiN coating as . cleared under K132242 with the same raw materials and manufacturing processes as the subject devices. Cytotoxicity testing was performed according to ANSI/AAMI/ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, and ISO 10993-12 Biological evaluation of medical devices -Part 12: Sample preparation and reference materials.
  • Mechanical performance testing of the subject device was performed according to ISO 14801. The worst-case scenario was chosen based on the FDA Guidance, "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The fatigue limit data demonstrated that constructs of the subject device abutments in combination with the stated compatible implants and used according to the proposed labeling, have sufficient strength for their intended use and are substantially equivalent to predicate devices with regard to mechanical performance.
  • End-user sterility validation was conducted according to ISO 17665-1 and ISO 17665-2 for the prior clearance K132242 and leveraged for the subject devices.

{9}------------------------------------------------

9. Conclusion

The subject devices and the predicate device have the same intended use and have the same technological characteristics.

Overall, the INNO SLA Submerged Implant System has the following similarities to the predicate device:

    • have the same intended use,
    • use the same operating principle,
    • incorporate similar design,
    • incorporate the same material and the sterilization method

Based on the similarities, we conclude that the INNO Sub Implant System is substantially equivalent to the predicate devices.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)