K Number
K201312

Validate with FDA (Live)

Date Cleared
2021-10-04

(504 days)

Product Code
Regulation Number
862.1117
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only.

For the quantitative measurement of N-terminal pro Brain Natriuretic Peptide (NT-proBNP) in human serum and plasma (K2 EDTA or Lithium Heparin) using the VITROS 3600 Immunodiagnostic System to aid in the diagnosis of heart failure. The test can also be used in the assessment of heart failure severity in patients diagnosed with heart failure.

Device Description

The VITROS NT-proBNP II test is performed using the VITROS VITROS NT-proBNP II Reagent Pack and the VITROS NT-proBNP II Calibrators on the VITROS Systems.

The VITROS NT-proBNP II test utilizes a one-step immunometric bridging assay design. A well is pushed from the pack and patient sample is dispensed into the antibody coated well. The assay reagent and the conjugate reagent are then dispensed into the well with the patient sample. NT-proBNP present in the sample binds with horseradish peroxidase (HRP)-labeled antibody conjugate which is captured by biotinylated anti-NT-proBNP capture antibody which is bound to Streptavidin coated microwells. The well is incubated for 8 minutes, before unbound materials are removed by washing.

The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrate (a luminol derivative and a peracid salt) and an electron transfer agent, is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the System. The amount of HRP conjugate bound is directly proportional to the concentration of NT-proBNP present.

AI/ML Overview

The provided document describes the analytical and clinical performance of the VITROS Immunodiagnostic Products NT-proBNP II Reagent Pack, an in vitro diagnostic device used to aid in the diagnosis and assessment of heart failure.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of acceptance criteria and the reported device performance

The document does not explicitly present a "table of acceptance criteria" in terms of pre-defined thresholds for performance metrics that the device must meet for clearance. Instead, it describes the design goals and then reports the observed performance. For clarity, I will create a table summarizing the reported performance, which implicitly indicates the criteria were met or exceeded for FDA clearance.

Test CategorySpecific Test / MetricAcceptance Criteria (Implicit/Design Goal from predicate or general IVD standards)Reported Device Performance
Analytical Performance
PrecisionRepeatability (Within-run) %CVTypically < 10% for quantitative assays; lower for low concentrations.Range: 1.0% - 2.1% (SD 0.52 to 370)
Within-calibration %CVTypically < 15%Range: 1.9% - 5.0% (SD 1.63 to 710)
Within-lab %CVTypically < 20%Range: 2.4% - 5.7% (SD 1.91 to 730)
Total Precision %CV (additional analysis)Consistent with product performance expectations.Range: 2.62% - 8.96% (SD 2.86 to 761)
Limit of Detection (LoD)Designed to be <= 30.0 pg/mL0.46 pg/mL (Observed); 0.49 pg/mL (Claimed)
Limit of Quantitation (LoQ)Designed to be <= 30.0 pg/mL at 20% CV0.46 pg/mL at 20% CV (Observed); 20.0 pg/mL (Claimed, to maintain linearity)
LinearityAcross measuring rangeLinear over the measuring rangeLinear from 20.0 to 30,000 pg/mL
Matrix ComparisonSerum vs. EDTA plasma vs. Lithium Heparin plasmaNo significant effect, meet acceptance criteria.All three tube types suitable for use (met acceptance criteria).
Analytical SpecificityBias from common substances < 10%No bias > 10% observed for most tested compounds.Specific interferents (Cefoxitin sodium, Sodium Azide) showed >10% bias.
Cross-ReactivityVarious related peptides (e.g., ANP, proBNP, BNP32)Low cross-reactivity desired.Ranges from <1.0% to 39.1% (non-glycosylated proBNP). Most are <1%.
High Dose Hook EffectNo effect up to X concentrationNo high dose hook effect up to 300,000 pg/mL.
Clinical Performance
Aid in HF Diagnosis (ED Setting)AUC (Overall)Not explicitly stated, but typically high (e.g., > 0.85 or 0.90 for this type of test)0.920 (95% CI: 0.909-0.931)
AUC (Age-stratified)Ranged from 0.904 to 0.954
AUC (Clinical Subgroups)Ranged from 0.899 to 0.945
Posttest Probability of HF (Positive result)High positive predictive value/posttest probability for rule-in.Range: 80.4% - 85.7% across age groups
Posttest Probability of non-HF (Negative result)High negative predictive value/posttest probability for rule-out.Range: 96.5% - 98.3% across age groups
Likelihood Ratio Positive (LR+)High (e.g., > 5-10 for strong rule-in)Range: 4.52 - 6.84 across age groups
Likelihood Ratio Negative (LR-)Low (e.g., < 0.1-0.2 for strong rule-out)Range: 0.01 - 0.05 across age groups
Aid in HF Diagnosis (Outpatient Setting)AUC (Overall)0.880 (95% CI: 0.822 to 0.937)
AUC (Clinical Subgroups)Ranged from 0.838 to 0.940
Sensitivity (Rule-out cutoff 125 pg/mL)High (for rule-out, e.g., >90%)91.7% (44/48)
Specificity (Rule-out cutoff 125 pg/mL)Reasonable (for rule-out, may be lower)67.2% (490/729)
NPV (Rule-out cutoff 125 pg/mL)High (for rule-out, e.g., >90%)99.2% (490/494)
PPV (Rule-out cutoff 125 pg/mL)(for rule-out, may be lower)15.6% (44/283)
Correlation with NYHAStatistical significance of relationship with HF severityStatistically significant trend.Jonckheere-Terpstra test p < 0.0001 (statistically significant correlation).

2. Sample size used for the test set and the data provenance

  • ED Setting (Diagnosis of Heart Failure):
    • Sample Size: 2200 subjects.
    • Data Provenance: Multi-center prospective study, 20 collection sites across the United States.
  • Outpatient Setting (Diagnosis of Heart Failure):
    • Sample Size: 777 subjects.
    • Data Provenance: Multi-center prospective study, 10 collection sites across the United States.
  • Correlation with NYHA Functional Classification:
    • Sample Size: 1143 subjects with heart failure.
    • Data Provenance: Not explicitly stated if this was a separate collection or a subset of the ED/outpatient studies, but given it's "subjects with heart failure," it likely derives from similar clinical populations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • For both the ED and Outpatient settings: The final clinical diagnosis (ground truth) was adjudicated by independent cardiologists or ED physicians experienced in diagnosing HF.
  • Number of Experts: Not specified. The document only mentions "independent cardiologists or ED physicians," implying more than one, but not a precise number.
  • Qualifications: "Experienced in diagnosing HF." No specific number of years of experience or board certifications are mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • The document states that the final clinical diagnosis was "adjudicated by independent cardiologists or ED physicians experienced in diagnosing HF." It does not specify a quantitative adjudication method like "2+1" or "3+1." This suggests that the adjudication process was qualitative and relied on the clinical expertise of the adjudicators to reach a consensus diagnosis for each patient, rather than a fixed number of readers and a tie-breaking rule.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a MRMC comparative effectiveness study was not done. This document describes the performance validation of an in vitro diagnostic (IVD) immunoassay kit (VITROS NT-proBNP II Reagent Pack), which measures blood biomarkers. This is not an AI/imaging device where human readers would typically be assisted by AI. The product is a laboratory test, and its performance is evaluated based on its accuracy in measuring the biomarker and its diagnostic utility compared to clinical truth.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, this is effectively a standalone performance study. The VITROS NT-proBNP II Reagent Pack provides a quantitative measurement of NT-proBNP. Its performance is assessed purely on the analytical accuracy of the measurement and its direct correlation with clinical outcomes (diagnosis of HF, NYHA classification) as determined by the expert adjudicators, without human interpretation of the device's numerical output for diagnosis (i.e., it's a direct measurement, not an interpretative AI). The output values are then interpreted against established cutoffs to aid in diagnosis, but the device itself does not involve a human-in-the-loop for its direct performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The primary ground truth for heart failure diagnosis was expert clinical judgment/adjudication. For the diagnostic studies, "The final clinical diagnosis was adjudicated by independent cardiologists or ED physicians experienced in diagnosing HF."
  • For the assessment of heart failure severity, the ground truth was New York Heart Association (NYHA) Functional Classification, which is a clinical classification based on patient symptoms and physical activity levels.

8. The sample size for the training set

  • This document describes the validation of an in vitro diagnostic device (a reagent pack and system) for measuring a biomarker, not a machine learning or AI algorithm in the typical sense that would involve a "training set" for model development. The assays (immunoassays) are based on chemical reactions rather than statistical models trained on large datasets.
  • The "training" of such a device primarily involves rigorous analytical development and characterization, ensuring the chemical processes are precise and accurate. Therefore, the concept of a separate "training set" with ground truth data, as used in AI/ML, is not directly applicable here. The document focuses on the performance validation of the developed assay.

9. How the ground truth for the training set was established

  • As noted above, the concept of a "training set" with ground truth for an AI/ML algorithm doesn't directly apply to this type of IVD immunoassay. The analytical methods and performance characteristics (precision, linearity, LoD, LoQ, analytical specificity, etc.) are established through laboratory experiments and characterization studies, not by training on a clinical dataset with ground truth in the AI/ML sense. Clinical performance is then validated against independently established clinical diagnoses (expert adjudication).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 4, 2021

Ortho Clinical Diagnostics Ann Quinn Director, Regulatory Affairs 100 Indigo Creek Drive Rochester, New York 14626

Re: K201312

Trade/Device Name: VITROS® Immunodiagnostic Products NT-proBNP II Reagent Pack Regulation Number: 21 CFR 862.1117 Regulation Name: B-Type Natriuretic Peptide Test System Regulatory Class: Class II Product Code: NBC Dated: November 6, 2020 Received: November 9, 2020

Dear Ann Quinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marianela Perez-Torres. Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201312

Device Name

VITROS® Immunodiagnostic Products NT-proBNP II Reagent Pack

Indications for Use (Describe) For in vitro diagnostic use only.

For the quantitative measurement of N-terminal pro Brain Natriuretic Peptide (NT-proBNP) in human serum and plasma (K2 EDTA or Lithium Heparin) using the VITROS 3600 Immunodiagnostic System to aid in the diagnosis of heart failure. The test can also be used in the assessment of heart failure severity in patients diagnosed with heart failure.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K201312

    1. Submitter Ortho-Clinical Diagnostics, Inc. name, 100 Indigo Creek Drive Rochester, NY 14626 address, contact P: (585) 453-4152 F: (585) 453-3113 Contact Person: Ann Quinn, Director, Regulatory Affairs
  • September 23, 2021 2. Preparation Date
    1. Device Trade or Proprietary Names: VITROS® Immunodiagnostic Products NT-proBNP II Reagent Pack name Common Name: VITROS NT-proBNP II Classification: B-Type natriuretic peptide test system (862.1117) Class II Product Code: NBC
  • Roche Elecsys® proBNP II Immunoassay, K072437 4. Predicate Device
  • Device 5. The VITROS NT-proBNP II test is performed using the VITROS description VITROS NT-proBNP II Reagent Pack and the VITROS NT-proBNP II Calibrators on the VITROS Systems.

The VITROS NT-proBNP II test utilizes a one-step immunometric bridging assay design. A well is pushed from the pack and patient sample is dispensed into the antibody coated well. The assay reagent and the conjugate reagent are then dispensed into the well with the patient sample. NT-proBNP present in the sample binds with horseradish peroxidase (HRP)-labeled antibody conjugate which is captured by biotinylated anti-NT-proBNP capture antibody which is bound to Streptavidin coated

{4}------------------------------------------------

microwells. The well is incubated for 8 minutes, before unboundmaterials are removed by washing.
The bound HRP conjugate is measured by a luminescent reaction. Areagent containing luminogenic substrate (a luminol derivative and aperacid salt) and an electron transfer agent, is added to the wells. The HRPin the bound conjugate catalyzes the oxidation of the luminol derivative,producing light. The electron transfer agent (a substituted acetanilide)increases the level of light produced and prolongs its emission. The lightsignals are read by the System. The amount of HRP conjugate bound isdirectly proportional to the concentration of NT-proBNP present.
6. DeviceintendeduseRx ONLYFor in vitro diagnostic use only.
For the quantitative measurement of N-terminal pro Brain NatriureticPeptide (NT-proBNP) in human serum and plasma (K2 EDTA or LithiumHeparin) using the VITROS 3600 Immunodiagnostic System to aid in thediagnosis of heart failure. The test can also be used in the assessment ofheart failure severity in patients diagnosed with heart failure.
    1. Compari-The following tables provide a summary of the key features of the new son to device assessed against the predicate. predicate device:
DeviceCharacteristicNew DeviceVITROS NT-proBNP IIPredicate DeviceRoche Elecsys proBNPII Immunoassay,K072437
Intended UseRx ONLYFor in vitro diagnostic use only.For the quantitative measurement of N-terminal pro BrainNatriuretic Peptide (NT-proBNP) in human serum andplasma (K2 EDTA or Lithium Heparin) using the VITROS3600 Immunodiagnostic System to aid in the diagnosis ofheart failure. The test can also be used in the assessment ofheart failure severity in patients diagnosed with heartfailure.For the quantitativedetermination of N-terminal pro-Brainnatriuretic peptide inhuman serum andplasma. Elecsys proBNPII assay is used as an aidin the diagnosis ofindividuals suspected ofhaving congestive heartfailure. The test isfurther indicated for therisk stratification ofpatients with acutecoronary syndrome andcongestive heart failure.The test may also serve
DeviceCharacteristicNew DeviceVITROS NT-proBNP IIPredicate DeviceRoche Elecsys proBNPII Immunoassay,K072437
Intended UseSee aboveas an aid in theassessment of increasedrisk of cardiovascularevents and mortality inpatients at risk for heartfailure who have stablecoronary artery disease.The electro-chemiluminescenseimmunoassay "ECLIA"is intended for use onElecsys and cobas eimmunoassay analyzers.
Basic PrincipleSandwich immunoassaySame
AntibodyMonoclonal anti-NT-proBNPSame
Sample TypeSerum and plasmaSame
MeasuringRange20-30,000 pg/mL5-35,000 pg/mL
TraceabilityStandardized against the Elecsys proBNP II assay.Standardized against theElecsys proBNP assay.<75 years: 125 pg/mL≥75 years: 450 pg/mL
ResultInterpretationFor patients presenting to the ED settings with acute orworsening dyspnea and clinical suspicion of HF, theVITROS NT-proBNP II test results should be interpreted asindicated in the table below.
VITROS NT-proBNP II Test Results(pg/mL)Age Group (Years)Interpretation of Results
<300AllNegative: Heart Failure Unlikely
≥300 to <45022-<50Gray Zone: Result Indeterminate -Consider other causes of NT-proBNP elevation
≥300 to <90050-<75
≥300 to <1800≥75
≥45022-<50Positive: Heart Failure Likely
≥90050-<75
≥1800≥75

{5}------------------------------------------------

{6}------------------------------------------------

DeviceCharacteristicNew DeviceVITROS NT-proBNP IIPredicate DeviceRoche Elecsys proBNPII Immunoassay,K072437
For ambulatory patients presenting to outpatient facilitieswith clinical suspicion of HF not previously diagnosed andat least one sign, symptom or risk factor for HF, theVITROS NT-proBNP II test results should be interpreted asindicated in the table below.
VITROS NT-proBNP II Test Results(pg/mL) Age Group Interpretation of Results <125 All Negative: Heart Failure Unlikely ≥125 All Consider Heart Failure as well as othercauses* of NT-proBNP elevation.

8. Nonclinical performance

Several nonclinical tests were performed.

Precision

Precision was evaluated consistent with CLSI document EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition. Two replicates of each of eleven fluids, eight human individual or serum pools and three controls, were tested on two separate occasions per day on at least 20 different test days. The experiment was performed using three Master Lots on one VITROS 3600 Immunodiagnostic System. The data presented are a representation of the product performance.

Conc. Units - pg/mL
NT-proBNPWithin-run*Within-cal**Within-lab***No.ObservationsNo.Days
Conc.SD%CVSD%CVSD%CV
33.60.521.51.634.91.915.78020
70.31.452.13.314.73.875.58020
97.81.451.53.904.04.624.78020
2143.91.86.43.07.93.78020
3654.41.210.93.014.03.88020
81511.61.428.73.530.53.78020
93710.11.126.62.829.63.28020
1730191.1865.0844.88020

{7}------------------------------------------------

Conc. Units - pg/mLNo.No.
NT-proBNPConc.Within-run*Within-cal**Within-lab***ObservationsDays
SD%CVSD%CVSD%CV
5830801.41552.61592.78020
113001101.02101.92702.48020
246003701.57102.97303.08020
  • Within-run (repeatability). Between Duplicate precision averaged over all runs.

** Within-calibration. Total precision with weighted components of within-run, between-run, and between-day variation.

*** Within-lab. A measure of the effect of recalibration on total precision, calculated within reagent lot, using data from at least 4 calibrations.

An additional analysis was conducted to evaluate total imprecision. The data presented are a representation of the product performance.

SystemMeanpg/mLBetween Lot*Between Day**BetweenRuns***Within Run /Residual****Total*****Observation
SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
3600322.347.320.8752.741.284.020.5451.712.868.96240
69.53.925.651.892.711.832.631.582.274.987.17240
974.654.792.652.731.891.951.561.615.896.07240
3699.142.486.321.717.752.104.901.3314.43.91240
82315.91.9411.61.4020.52.4912.51.5231.03.77240
94926.62.8117.51.8413.91.4612.21.2936.83.88240
174510.80.6244.52.5556.93.2620.31.1675.84.34240
5859751.2892.11.5755.40.9580.31.371542.62240
246863291.333551.444551.843711.507613.08240
2146.783.173.661.713.111.453.501.649.024.22240
113922061.811451.2865.20.571761.543142.76240

*Between lot: Variability of the test performance from lot to lot.

**Between day: Variability of the test performance from day to day.

*** Between run: Variability of the test performance from run to run.

*** Within Run / Residual Variability.

****Total: Variability of the test incorporating factors of Instrument, lot, day and run.

Limit of Detection

Detection studies for the VITROS NT-proBNP II test were evaluated according to CLSI document EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition.

The Limit of Detection (LoD) for the VITROS NT-proBNP II test was designed to be less than or equal to 30.0 pg/mL. The observed LoD was determined to be 0.46 pg/mL. The claimed LoD is 0.49 pg/mL. The Limit of Quantitation (LoQ) for the VITROS NT-

{8}------------------------------------------------

proBNP II test was designed to be less than or equal to 30.0 pg/mL at 20% CV. The observed LoQ at 20% CV was determined to be 0.46 pg/mL. The claimed LoQ is set at 20.0 pg/mL to maintain linearity within the measuring range.

LoDLoQ
pg/mLpg/mL
0.4920.0

Limit of Detection and Limit of Quantitation

Linearity

Linearity studies were performed according to CLSI document EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline. Three Master Lots of VITROS NT-proBNP II test were tested on the VITROS 3600 Immunodiagnostic System. A low and high sample pool was prepared and mixed to give fourteen (14) further pools of intermediate concentrations. The low and high linearity pools and the interim dilutions between the low and high linearity pools were assayed in triplicate. The VITROS NT-proBNP II test was linear over the measuring range, from 20.0 to 30,000 pg/mL for the VITROS 3600 Immunodiagnostic System.

Matrix Comparison

A total of 160 unaltered matched serum, EDTA plasma and lithium heparin plasma samples covering the measuring range of the VITROS NT-proBNP II test were used to assess sample matrix differences between serum, EDTA plasma, and lithium heparin plasma samples.

Samples were collected into each of the following tube types: serum plus clot activator plastic tube (serum), K2-EDTA plastic tube (EDTA), and lithium heparin plasma plastic tube (Li-Hep). Each serum sample was tested in duplicate while the Li-Hep plasma and EDTA plasma samples were assayed in singleton using the VITROS 3600 Immunodiagnostic System.

Overall, the results show EDTA and Li-Hep plasma plastic collection tubes did not show a significant effect on the test results when compared to serum and met the acceptance criteria. All three tube types, serum, EDTA plasma, and Li-Hep plasma are suitable for use with the VITROS NT-proBNP II test.

Analytical Specificity

The VITROS NT-proBNP II test was evaluated for interference according to CLSI guidelines EP07 3rd ed - Interference Testing in Clinical Chemistry. Commonly encountered substances were tested on three Master Lots of reagents. Of the compounds tested in the table below, none was found to cause a bias of >10% with the test at the concentrations indicated at nominal NT-proBNP concentrations of 125 pg/mL (14.8 pmol/L) and 2000 pg/mL (236 pmol/L).

{9}------------------------------------------------

CompoundConcentrationCompoundConcentration
Acetaminophen156 µg/mL1030 µmol/LIbuprofen21.9 mg/dL1.06 mmol/L
Acetylcysteine15.0 mg/dL920 µmol/LInsulin3.12 µg/dL5.37 nmol/L
Adrenaline (Epinephrine)20.0 µg/dL1.09 µmol/LIntralipid2.00 g/dLNA
Alprazolam25.8 µg/dL836 nmol/LL-dopa (Levodopa)750 µg/dL38.0 µmol/L
Amlodipine besylate10.5 µg/dL184 nmol/LLevothyroxine42.9 µg/dL552 nmol/L
Amoxicillin5.40 mg/dL148 µmol/LLidocaine1.50 mg/dL64.0 µmol/L
Ascorbic acid5.25 mg/dL298 µmol/LMethyldopa sesquihydrate2.25 mg/dL94.4 µmol/L
Atorvastatincalciumtrihydrate162 µg/dL1.34 µmol/LMethylprednisolone783 µg/dL20.9 µmol/L
Benazepril HCl44.0 µg/dL955 nmol/LMetoprolol hemitartrate150 µg/dL2.19 µmol/L
Bilirubin, conjugated40.0 mg/dL474 µmol/LMetronidazole12.3 mg/dL718 µmol/L
Bilirubin, unconjugated40.0 mg/dL684 µmol/LMolsidomine18.0 µg/dL743 nmol/L
Biotin3510 ng/mL14.4 µmol/LNaproxen sodium39.3 mg/dL1.56 mmol/L
Caffeine10.8 mg/dL556 µmol/LNicardipine HCL46.5 µg/dL901 nmol/L
Carvedilol43.2 µg/dL1.06 µmol/LNifedipine58.8 µg/dL1.70 µmol/L
Ceftriaxonedisodiumhemi(heptahydrate)100 mg/dL1510 µmol/LOmeprazole840 µg/dL24.3 µmol/L
Cholesterol400 mg/dL10.3 mmol/LOxycodone HCl32.4 µg/dL0.92 µmol/L
Clopidogrelhydrogensulfate2.40 µg/dL57.2 nmol/LPhenobarbital69.0 mg/dL2.97 mmol/L
Cotinine240 µg/dL13.6 µmol/LPhenprocoumon(Marcumar)1.50 mg/dL53.5 µmol/L
Creatinine15.0 mg/dL1.33 mmol/LPropafenone HCL72.0 µg/dL1.91 µmol/L
Cyclosporine180 µg/dL1.50 µmol/LPseudoephedrine HCl330 µg/dL16.4 µmol/L
Dextran2.40 g/dL600 µmol/LRheumatoid Factor1500 IU/mLNA
Digitoxin7.50 µg/dL98.0 nmol/LRifampicin (Rifampin)4.80 mg/dL58.3 µmol/L
Digoxin3.90 µg/dL49.9 nmol/LSalicylic acid2.86 mg/dL207 µmol/L
Diphenhydramine HCl77.4 µg/dL2.65 µmol/LSalmeterol1.65 µg/dL39.7 nmol/L
Dipyrone (as 4-methylaminoantipyrineHydrochloride)3.30 mg/dL130 µmol/LSotalol hydrochloride510 µg/dL16.5 µmol/L
Dypyridamole1.00 mg/dL19.8 µmol/LSpironolactone55.5 µg/dL1.33 µmol/L
Doxycycline hyclate1.80 mg/dL35.1 µmol/LStreptokinase150,000 U/dLNA
Enalaprilat dihydrate81.9 µg/dL2.13 µmol/LTheophylline6.00 mg/dL333 µmol/L
Ethanol600 mg/dL130 mmol/LTolbutamide54.9 mg/dL2.03 mmol/L
Fibrinogen1000 mg/dLNATotal Protein15.0 g/dLNA
Furosemide1.59 mg/dL48.1 µmol/LtPA (Alteplase)1.20 mg/dLNA
Gentamicin Sulfate3.51 mg/dL61.0 µmol/LTriglyceride1500 mg/dL16.9 mmol/L
Glycerylnitrate(Nitroglycerin)1.20 µg/dL52.8 nmol/LValproic Acid31.8 mg/dL2.21 mmol/L
HAMA (Human Anti-MouseAntibody800 µg/LNAVancomycin Hydrochloride12.3 mg/dL82.8 µmol/L
Hemoglobin1000 mg/dL155 µmol/LVerapamil Hydrochloride160 µg/dL3.26 µmol/L
Heparin (Sodium), UFH330 U/dLNAWarfarin8.00 mg/dL260 µmol/L

The substances shown to interfere with the VITROS NT-proBNP II test are provided in the table below. These results are representative. The degree of interference at concentrations other than those listed might not be predictable from these results. Other interfering substances may be encountered in the patient population.

{10}------------------------------------------------

InterferentInterferent ConcentrationMeasured NT-proBNP Concentration*% Bias**
pg/mLpmol/L
Cefoxitin sodium697 mg/dL84.39.95-24.1
15.5 mmol/L922109-19.1
311 mg/dL10512.4-10.0
306 mg/dL80695.1-10.0
Sodium Azide100 mg/dL91.410.8-12.1
85.9 mg/dL93.611.0-10.0
15.4 mmol/L
13.2 mmol/L
  • Average test replicate determinations.

** Estimate of the maximum difference observed as a percentage.

The cross-reactivity of the VITROS NT-proBNP II test was evaluated by adding the following substances to a human serum sample containing no NT-proBNP.

Cross-ReactantConcentrationMean Result of Control SampleMean Result of Cross-Reactant Sample% Cross- Reactivity
pg/mLpmol/Lpg/mLpmol/L
ANP283.10 µg/mL1.01 nmol/L*****
proBNP (glycosylated)3000 pg/mLN/A-0.14-0.0257.66.801.9
proBNP (nonglycosylated)3000 pg/mL0.249 nmol/L-0.48-0.0686310228.8
NT-proANP1-30(preproANP25-55)3.50 µg/mL0.998 µmol/L*****
NT-proANP31-67(preproANP56-92)1.00 ng/mL0.258 nmol/L*****
NT-proANP79-98(preproANP 104-123)1.00 ng/mL0.458 nmol/L*****
BNP32 (Natrecor®)3.50 µg/mL1.01 µmol/L*****
CNP222.20 µg/mL1.00 µmol/L*****
Adrenomedullin1.00 ng/mL0.166 nmol/L*****
Aldosterone0.600 ng/mL1.66 nmol/L*****
Angiotensin I0.600 ng/mL0.463 nmol/L*****
Angiotensin II0.600 ng/mL0.574 nmol/L*****
Angiotensin III1.00 ng/mL1.07 nmol/L*****
Endothelin20.0 pg/mL8.03 pmol/L*****
Urodilatin3.50 µg/mL0.998 µmol/L*****
Arg-Vasopressin1.00 µg/mL0.922 µmol/L*****
Renin50.0 ng/mL28.4 nmol/L*****

*Not Detectable (ND). Concentration was below the measuring range of the test, 20.0-30,000 pg/mL (2.36-3,540 pmol/L).

The cross-reactivity of the VITROS NT-proBNP II test was evaluated by adding the following substances to a human serum sample containing NT-proBNP at a concentration of 125 pg/mL (14.8 pmol/L).

Cross-ReactantConcentrationMean NT-proBNPResult of ControlSampleMean NT-proBNPResult of Cross-Reactant Sample% Cross-Reactivity
pg/mLpmol/Lpg/mLpmol/L
ANP283.10 µg/mL11213.211313.3<1.0
proBNP (glycosylated)3000 pg/mL10512.418321.62.6
proBNP (nonglycosylated)3000 pg/mL11813.9129015239.1

{11}------------------------------------------------

NT-proANP1-30(preproANP25-55)3.50 µg/mL0.998 µmol/L11313.377591.5<1.0
NT-proANP31-67(preproANP56-92)1.00 ng/mL0.258 nmol/L11213.211313.3<1.0
NT-proANP79-98(preproANP104-123)1.00 ng/mL0.458 nmol/L11213.211413.5<1.0
BNP32 (Natrecor®)3.50 µg/mL1.01 µmol/L11213.211513.6<1.0
CNP222.20 µg/mL1.00 µmol/L11213.211413.5<1.0
Adrenomedullin1.00 ng/mL0.166 nmol/L11213.211413.5<1.0
Aldosterone0.600 ng/mL1.66 nmol/L11213.211713.8<1.0
Angiotensin I0.600 ng/mL0.463 nmol/L11213.211213.2<1.0
Angiotensin II0.600 ng/mL0.574 nmol/L11213.211113.1<1.0
Angiotensin III1.00 ng/mL1.07 nmol/L11213.211213.2<1.0
Endothelin20.0 pg/mL8.03 pmol/L11313.311313.3<1.0
Urodilatin3.50 µg/mL0.998 µmol/L11213.211413.5<1.0
Arg-Vasopressin1.00 µg/mL0.922 µmol/L11213.211713.8<1.0
Renin50.0 ng/mL28.4 nmol/L11213.211213.2<1.0

Cross-reactivity was expressed as the mean result obtained for the cross-reactant sample minus the mean result obtained for the control sample divided by the cross-reactant concentration in percentage terms.

Dilution

Serum or plasma (K2 EDTA or Lithium Heparin) samples with concentrations greater than the measuring range may be automatically diluted on the system up to 10-fold (1 part sample with 9 parts diluent) by the VITROS 3600 Immunodiagnostic System with the VITROS High Sample Diluent B Reagent Pack prior to testing.

High Dose Hook

The VITROS NT-proBNP II test has no high dose hook effect up to a concentration of 300,000 pg/mL.

Expected Values

It is recommended that each laboratory establish its own expected values for the population it serves. The VITROS NT-proBNP II test Reference Interval (RI) was established for six subgroups, based on age and gender from the serum of 385 female and 374 male healthy donors.

Subjects were excluded if they met any of the following exclusion criteria:

  • . Current smokers, subjects with cardiac conditions and disease, high blood pressure, kidney disease, diabetes, cancer within the last five years, stroke, and asthma or other lung disease within the last five years.
  • Subjects who have reported high cholesterol, high triglycerides, thyroid disease, and females who are pregnant.
  • Additional exclusion criteria:
    • Troponin > 99th percentile (> 0.034 ng/mL VITROS Troponin I ES assay) .
    • HbA1c > 6.5% .
    • Creatinine (eGFR ≤ 60 mL/min) ●

{12}------------------------------------------------

AgeGendernRI LowerLimit(pg/mL)RI UpperLimit(pg/mL)
22 - <50Female129<20.095.3
50 - < 75Female127<20.0221
≥75Female129<20.0296
22 - <50Male131<20.0125
50 - < 75Male120<20.0299
≥75Male123<20.0326
Overall756<20.0217

The reference interval was conducted in accordance with the CLSI EP28. Analysis at the 95% confidence level vields the ranges shown in the table.

9. Clinical performance

The clinical performance information should only be used as a guide. It is recommended that each laboratory determine and confirm the diagnostic cutoffs for the population it serves.

Aid in Diagnosis of Heart Failure

Emergency Department Setting

A multi-center prospective study including 20 collection sites across the United States was conducted to establish the performance characteristics of the VITROS NT-proBNP II test. Subjects 22 years and older presenting to the Emergency Department (ED) with dyspnea (acute or worsening) and clinical suspicion of heart failure (HF) were enrolled into the study. Subjects with terminal kidney failure on chronic dialysis and subjects with dyspnea clearly not secondary to HF were excluded from the study. The final clinical diagnosis was adjudicated by independent cardiologists or ED physicians experienced in diagnosing HF. Individuals in the population were African American (36.55%) and Caucasian (59.59%), withthe remaining 3.86% represented by other races. Dyspnea was acute in 44.09%, worsening in 55.45% and not specified in 0.45% of subjects.

The VITROS NT-proBNP II test results were determined from 2200 ED subjects, 1016 (46.18%) females and 1184 (53.82%) males, ranging in age from 22 to 106 years. The descriptive statistics for the VITROS NT-proBNP II test results(pg/mL) were determined within and across gender by age group and are summarized in the following tables:

{13}------------------------------------------------

All Subjects

StudyPopulationHeart FailureNon-Heart Failure
Age (years)22-<5050-<75≥75All22-<5050-<75≥75All
N11453844310951416303341105
Mean68406510781070702616941120767
SD2050094901010011400568172016901630
Median215035504850378058.4177473247
Min20.053.729520.020.020.020.020.0
Max17800086300946001780004710238001360023800

Female Subjects

StudyPopulationHeart FailureNon-Heart Failure
Age (years)22-<5050-<75≥75All22-<5050-<75≥75All
N4319421745471306185562
Mean10300683077907620196648948690
SD32200100001060013900356180012801540
Median192036204710385068.2157444226
Min20.010229620.020.020.020.020.0
Max1780008630094600178000201023800787023800

Male Subjects

StudyPopulationHeart FailureNon-Heart Failure
Age (years)22-<5050-<75≥75All22-<5050-<75≥75All
N7134422664170324149543
Mean47206330783066803267371330846
SD6780917097309180718164020801720
Median226034805020369043.6203519275
Min19553.729553.720.020.020.020.0
Max392007040086000860004710171001360017100

The area under the Receiver Operating Characteristic (ROC) curve (AUC) with a 95% confidence interval (CI) for the VITROS NT-proBNP II test within and across age groups within and across gender are presented in the table below. The AUC ranged between 0.904 to 0.954 within and across gender.

Age GroupAll SubjectsFemale SubjectsMale Subjects
AUC95% ConfidenceIntervalAUC95% ConfidenceIntervalAUC95% ConfidenceInterval
22-<50 years0.9540.928-0.9790.9520.905-0.9990.9470.912-0.983
50-<75 years0.9220.907-0.9370.9260.904-0.9480.9170.896-0.938
≥75 years0.9150.895-0.9340.9200.895-0.9460.9040.872-0.937
Overall0.9200.909-0.9310.9250.910-0.9410.9140.898-0.930

AUC analyses were also performed for relevant clinical subgroups. The AUC with a 95% confidence interval (CI) for the VITROS NT-proBNP II test across gender within the relevant clinical subgroups are presented in the table below. The AUCof the VITROS NT-proBNP II test for each of the relevant clinical subgroups for subjects with or without the condition was greater than or equal to 0.899.

{14}------------------------------------------------

Relevant Clinical Subgroups
Subjects With the ConditionSubjects Without the Condition
SubgroupsAUC95% Confidence IntervalSubgroupsAUC95% Confidence Interval
History of HF(N=1220)0.8990.881-0.918No History of HF(N=970)0.9320.917-0.947
eGFR <60*mL/min/1.73 m2(N=1012)0.9000.880-0.920eGFR ≥60mL/min/1.73 m2(N=1166)0.9230.908-0.938
BMI ≥30.0 kg/m2(N=1160)0.9030.885-0.920BMI <30.0 kg/m2(N=1025)0.9450.931-0.958
WithComorbidities**(N=1978)0.9150.902-0.927WithoutComorbidities***(N=222)0.9440.915-0.974
  • Subjects with renal disease on dialysis were excluded from the study

** Subjects with at least one of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronicobstructive pulmonary disease (COPD)

*** Subjects without any of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructivepulmonary disease (COPD)

Distribution of VITROS NT-proBNP II test results after applying the multiple cutoffs, 300 pg/mL for all age groups and 450pg/mL for 22-<50 years old, 900 pg/mL for 50-<75 years old, 1800 pg/mL for ≥75 years old, versus adjudicated diagnosiswas summarized within and across age groups across gender and is presented in the table below.

VITROS NT-proBNP II Test Results versus Adjudicated Diagnosis within and across Age Group

Age(years)VITROS NT-proBNP II Test ResultAdjudicated DiagnosisTotal
HFNon-HF
22-<50VITROS NT-proBNP II TestResult InterpretationPositive10519124
Gray Zone51318
Negative4109113
Total114141255
50-<75VITROS NT-proBNP II TestResult InterpretationPositive472115587
Gray Zone59144203
Negative7371378
Total5386301168
≥75VITROS NT-proBNP II TestResult InterpretationPositive37262434
Gray Zone69153222
Negative2119121
Total443334777
All SubjectsVITROS NT-proBNP II TestResult InterpretationPositive9491961145
Gray Zone133310443
Negative13599612
Total109511052200

{15}------------------------------------------------

The following analyses were conducted to determine clinical performance:

VITROS NT-proBNP II TestResults ClassificationAdjudicated DiagnosisTotal
HFNon-HF
Positive: Heart Failure LikelyABA+B
Gray Zone: ResultIndeterminateCDC+D
Negative: Heart FailureUnlikelyEFE+F
TotalA+C+EB+D+FA+B+C+D+E+F

Pretest Probability of HF (prevalence of HF in the study) = (A+C+E)(A+B+C+D+E+F)

Posttest Probability of HF

Posttest probability of HF for positive test results = A/(A+B)

Posttest probability of HF for Gray zone test results = C/(C+D)

Posttest Probability of non-HF

Posttest probability of non-HF for Gray zone test results = D/(C+D)

Posttest probability of non-HF for negative test results = F/(E+F)

Likelihood ratios (LR) of HF given VITROS NT-proBNP II test result category

LR positive = (A/(A+C+E))/(B/(B+D+F))

LR gray zone = (C/(A+C+E))/(D/(B+D+F))

LR negative = (E/(A+C+E))/(F/(B+D+F))

The pretest probability of HF (prevalence of HF in the study), posttest probabilities, likelihood ratios and the two-tailed 95% CIs of the VITROS NT-proBNP II test result versus adjudicated diagnosis were determined across and within gender using the agedependent rule-in (450 pg/mL for subjects 22-<50 years old; 900 pg/mL for subjects 50-<75 years old: 1800 pg/mLfor subjects >75 years old) and age-independent rule-out (300 pg/mL) cutoffs and are summarized in the following tables:

Age Group (Years)Pretest Probability of HF (Prevalence of HF in Study) (n/N)VITROS NT-proBNP II Test Result InterpretationPosttest Probability of HF (n/N)Posttest Probability of non-HF (n/N)Likelihood Ratio Positive (HF)95% CI**
Estimate (%)95% CI* (%)Estimate (%)95% CI* (%)
22-<50(N=255)44.7%(114/255)Positive84.7(105/124)77.3-90.0--6.844.48-10.42
Gray zone27.8(5/18)12.5-50.972.2(13/18)49.1-87.50.480.17-1.29
Negative--96.5(109/113)91.3-98.60.050.02-0.12
50-<75(N=1168)46.1%(538/1168)Positive80.4(472/587)77.0-83.4--4.814.06-5.69
Gray zone29.1(59/203)23.3-35.770.9(144/203)64.3-76.70.480.36-0.63
Negative--98.1(371/378)96.2-99.10.020.01-0.05

All Subjects

{16}------------------------------------------------

Positive
≥75(N=777)57.0(443/777)Positive85.7(372/434)82.1-88.74.523.60-5.68
Gray zone31.1(69/222)25.4-37.468.9(153/222)62.6-74.60.340.27-0.43
Negative98.3(119/121)94.2-99.50.010.00-0.05
AllSubjects(N=2200)49.8%(1095/2200)Positive82.9(949/1145)80.6-85.04.894.29-5.56
Gray zone30.0(133/443)25.9-34.470.0(310/443)65.6-74.10.430.36-0.52
Negative97.9(599/612)96.4-98.80.020.01-0.04
  • 95% Wilson Score Confidence Interval

** Log Method Confidence Interval

The pretest probability of HF (prevalence of HF in the study), posttest probabilities, likelihood ratios and the two-tailed 95%CIs of the VITROS NT-proBNP II test result versus adjudicated diagnosis were determined for the relevant clinical subgroups using the age-dependent rule-in (450 pg/mL for subjects 22-<50 years old; 900 pg/mL for subjects 50-<75 years old; 1800 pg/mL for subjects ≥75 years old) and age-independent rule-out (300 pg/mL) cutoffs and are summarized in the following tables:

Female Subjects
Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
37.7%(43/114)Positive88.6(39/44)76.0-95.0--12.885.50-30.15
22-<50(N=114)Gray zone11.1(1/9)2.0-43.588.9(8/9)56.5-98.00.210.03-1.59
Negative--95.1(58/61)86.5-98.30.090.03-0.26
38.8%(194/500)Positive77.0(167/217)70.9-82.1--5.274.06-6.83
50-<75(N=500)Gray zone26.4(23/87)18.3-36.673.6(64/87)63.4-81.70.570.36-0.88
Negative--98.0(192/196)94.9-99.20.030.01-0.09
≥75(N=402)54.0%(217/402)Positive85.9(176/205)80.4-90.0--5.173.68-7.27
Gray zone30.5(40/131)23.3-38.969.5(91/131)61.1-76.70.370.27-0.51
Negative--98.5(65/66)91.9-99.70.010.00-0.09

*95% Wilson Score Confidence Interval

{17}------------------------------------------------

Male SubjectsAge Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
Estimate(%)95%CI*(%)Estimate(%)95%CI*(%)
22-<50(N=141)50.4%(71/141)Positive82.5(66/80)72.7-89.3--4.652.90-7.46
Gray zone44.4(4/9)18.9-73.355.6(5/9)26.7-81.10.790.22-2.82
Negative--98.1(51/52)89.9-99.70.020.00-0.14
50-<75(N=668)51.5%(344/668)Positive82.4(305/370)78.2-86.0--4.423.54-5.51
Gray zone31.0(36/116)23.3-39.969.0(80/116)60.1-76.70.420.29-0.61
Negative--98.4(179/182)95.3-99.40.020.01-0.05
≥75(N=375)60.3%(226/375)Positive85.6(196/229)80.5-89.6--3.922.89-5.31
Gray zone31.9(29/91)23.2-42.068.1(62/91)58.0-76.80.310.21-0.46
Negative--98.2(54/55)90.4-99.70.010.00-0.09

*95% Wilson Score Confidence Interval

** Log Method Confidence Interval

Subjects with a History of HF

Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
22-<50(N=118)73.7%(87/118)Positive92.9(79/85)85.4-96.7--4.692.28-9.65
Gray zone36.4(4/11)15.2-64.663.6(7/11)35.4-84.80.200.06-0.65
Negative--81.8(18/22)61.5-92.70.080.03-0.22
50-<75(N=642)63.1%(405/642)Positive87.2(355/407)83.6-90.1--4.003.13-5.09
Gray zone38.1(43/113)29.6-47.361.9(70/113)52.7-70.40.360.25-0.51
Negative--94.3(115/122)88.6-97.20.040.02-0.08
≥75(N=460)71.1%(327/460)Positive90.5(275/304)86.6-93.3--3.862.79-5.34
Gray zone41.1(51/124)32.9-49.958.9(73/124)50.1-67.10.280.21-0.38
Negative--96.9(31/32)83.4-99.40.010.00-0.10

*95% Wilson Score Confidence Interval

{18}------------------------------------------------

Age Group (Years)Pretest Probability of HF (Prevalence of HF in Study) (n/N)VITROS NT-proBNP II Test Result InterpretationPosttest Probability of HF (n/N)Posttest Probability of non-HF (n/N)Likelihood Ratio Positive (HF)95% CI**
Estimate (%)95% CI* (%)Estimate (%)95% CI* (%)
22-<50(N=137)19.7%(27/137)Positive66.7(26/39)51.0-79.4--8.154.86-13.65
Gray zone14.3(1/7)2.6-51.385.7(6/7)48.7-97.40.680.09-5.41
Negative--100.0(91/91)95.9-100.00.00N/A
50-<75(N=521)25.3%(132/521)Positive65.2(116/178)57.9-71.8--5.514.35-6.99
Gray zone17.8(16/90)11.2-26.982.2(74/90)73.1-88.80.640.39-1.05
Negative--100.0(253/253)98.5-100.00.00N/A
≥75(N=312)36.9%(115/312)Positive75.0(96/128)66.8-81.7--5.143.70-7.13
Gray zone18.8(18/96)12.2-27.781.3(78/96)72.3-87.80.400.25-0.62
Negative--98.9(87/88)93.8-99.80.020.00-0.14

Subjects with no History of HF

*95% Wilson Score Confidence Interval

** Log Method Confidence Interval

N/A: Not applicable

Subjects with Renal Disease (eGFR <60***mL/min/1.73 m²)

Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
Estimate(%)95%CI*(%)Estimate(%)95%CI*(%)
22-<50(N=69)62.3%(43/69)Positive87.2(41/47)74.8-94.0--4.132.04-8.36
Gray zone16.7(1/6)3.0-56.483.3(5/6)43.6-97.00.120.01-0.98
Negative--93.8(15/16)71.7-98.90.040.01-0.29
50-<75(N=476)56.5%(269/476)Positive79.6(250/314)74.8-83.7--3.012.45-3.69
Gray zone24.7(18/73)16.2-35.675.3(55/73)64.4-83.80.250.15-0.42
Negative--98.9(88/89)93.9-99.80.010.00-0.06
≥75(N=467)68.1%(318/467)Positive88.6(273/308)84.6-91.7--3.652.73-4.90
Gray zone37.6(44/117)29.4-46.662.4(73/117)53.4-70.60.280.21-0.39
Negative--97.6(41/42)87.7-99.60.010.00-0.08

*95% Wilson Score Confidence Interval

{19}------------------------------------------------

*** Subjects with renal disease on dialysis were excluded from the study

Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
22-<50(N=181)38.7%(70/181)Positive82.9(63/76)72.9-89.7--7.684.58-12.88
Gray zone36.4(4/11)15.2-64.663.6(7/11)35.4-84.80.910.28-2.98
Negative--96.8(91/94)91.0-98.90.050.02-0.16
50-<75(N=679)38.7%(263/679)Positive81.0(217/268)75.8-85.2--6.735.17-8.76
Gray zone31.3(40/128)23.9-39.768.8(88/128)60.3-76.10.720.51-1.01
Negative--97.9(277/283)95.5-99.00.030.02-0.08
≥75(N=306)40.2%(123/306)Positive78.2(97/124)70.2-84.6--5.353.73-7.66
Gray zone24.0(25/104)16.8-33.176.0(79/104)66.9-83.20.470.32-0.69
Negative--98.7(77/78)93.1-99.80.020.00-0.14

Subjects Without Renal Disease (eGFR ≥60 mL/min/1.73 m²)

*95% Wilson Score Confidence Interval

** Log Method Confidence Interval

Subjects With BMI ≥30.0 kg/m²

Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
22-<50(N=188)48.9%(92/188)Positive90.2(83/92)82.4-94.8--9.625.15-17.99
22-<50(N=188)48.9%(92/188)Gray zone27.8(5/18)12.5-50.972.2(13/18)49.1-87.50.400.15-1.08
22-<50(N=188)48.9%(92/188)Negative--94.9(74/78)87.5-98.00.060.02-0.15
50-<75(N=693)46.3%(321/693)Positive80.9(266/329)76.3-84.7--4.893.89-6.16
50-<75(N=693)46.3%(321/693)Gray zone36.1(48/133)28.4-44.563.9(85/133)55.5-71.60.650.47-0.90
50-<75(N=693)46.3%(321/693)Negative--97.0(224/231)93.9-98.50.040.02-0.08
≥75(N=279)55.9%(156/279)Positive84.6(121/143)77.8-89.6--4.342.94-6.39
≥75(N=279)55.9%(156/279)Gray zone37.8(34/90)28.5-48.162.2(56/90)51.9-71.50.480.34-0.68
≥75(N=279)55.9%(156/279)Negative--97.8(45/46)88.7-99.60.020.00-0.13

*95% Wilson Score Confidence Interval

{20}------------------------------------------------

Age Group (Years)Pretest Probability of HF (Prevalence of HF in Study) (n/N)VITROS NT-proBNP II Test Result InterpretationPosttest Probability of HF (n/N)Posttest Probability of non-HF (n/N)Likelihood Ratio Positive (HF)95% CI**
Estimate (%)95% CI* (%)Estimate (%)95% CI* (%)
22-<50(N=62)35.5%(22/62)Positive68.8(22/32)51.4-82.0--4.002.34-6.84
Gray zone∞(0/0)N/A∞(0/0)N/AN/A
Negative--100.0(30/30)88.6-100.00.00N/A
50-<75(N=465)46.5%(216/465)Positive79.8(206/258)74.5-84.3--4.573.58-5.83
Gray zone14.7(10/68)8.2-25.085.3(58/68)75.0-91.80.200.10-0.38
Negative--100.0(139/139)97.3-100.00.00N/A
≥75(N=498)57.6%(287/498)Positive86.3(251/291)81.8-89.7--4.613.48-6.12
Gray zone26.5(35/132)19.7-34.673.5(97/132)65.4-80.30.270.19-0.37
Negative--98.7(74/75)92.8-99.80.010.00-0.07

Subjects With BMI <30.0 kg/m²

*95% Wilson Score Confidence Interval

** Log Method Confidence Interval

N/A: Not applicable

Subjects With Comorbidities***

Age Group (Years)Pretest Probability of HF (Prevalence of HF in Study) (n/N)VITROS NT-proBNP II Test Result InterpretationPosttest Probability of HF (n/N)Posttest Probability of non-HF (n/N)Likelihood Ratio Positive (HF)95% CI**
22-<50 (N=178)55.1% (98/178)Positive86.8 (92/106)79.0-92.0--5.363.32-8.66
Gray zone23.1 (3/13)8.2-50.376.9 (10/13)49.7-91.80.240.07-0.86
Negative--94.9 (56/59)86.1-98.30.040.01-0.13
50-<75 (N=1066)47.3% (504/1066)Positive80.2 (438/546)76.7-83.3--4.523.80-5.38
Gray zone32.1 (59/184)25.7-39.167.9 (125/184)60.9-74.30.530.40-0.70
Negative--97.9 (329/336)95.8-99.00.020.01-0.05
≥75 (N=734)57.2% (420/734)Positive86.1 (352/409)82.4-89.1--4.623.64-5.86
Gray zone31.7 (66/208)25.8-38.368.3 (142/208)61.7-74.20.350.27-0.45
Negative--98.3 (115/117)94.0-99.50.010.00-0.05

*95% Wilson Score Confidence Interval

{21}------------------------------------------------

*** Subjects with at least one of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructive pulmonary disease (COPD)

Subjects Without Comorbidities***

Age Group(Years)PretestProbabilityof HF(Prevalenceof HF inStudy) (n/N)VITROS NT-proBNP IITest ResultInterpretationPosttest Probability ofHF (n/N)Posttest Probability ofnon-HF (n/N)LikelihoodRatio Positive(HF)95% CI**
22-<50(N=77)20.8%(16/77)Positive72.2(13/18)49.1-87.5--9.914.14-23.71
Gray zone40.0(2/5)11.8-76.960.0(3/5)23.1-88.22.540.46-13.95
Negative--98.1(53/54)90.2-99.70.070.01-0.48
50-<75(N=102)33.3%(34/102)Positive82.9(34/41)68.7-91.5--9.714.82-19.59
Gray zone0.0(0/19)0.0-16.8100.0(19/19)83.2-100.00.00N/A
Negative--100.0(42/42)91.6-100.00.00N/A
≥75(N=43)53.5%(23/43)Positive80.0(20/25)60.9-91.1--3.481.60-7.55
Gray zone21.4(3/14)7.6-47.678.6(11/14)52.4-92.40.240.08-0.73
Negative--100.0(4/4)51.0-100.00.00N/A

*95% Wilson Score Confidence Interval

** Log Method Confidence Interval

*** Subjects without any of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructive pulmonary disease (COPD)

N/A: Not applicable

Outpatient Setting

The VITROS NT-proBNP II test results were also determined from 777 subjects, 391 females and 386 males ranging in age from 23 to 94 years presenting to Cardiology Clinics and other outpatient facilities with a clinical suspicion of HF (not previously diagnosed) and at least one HF sign, symptom or risk factor at 10 collection sites across the United States.

Subjects with terminal kidney failure on chronic dialysis and subjects with dyspnea clearly not secondary to HF were excluded from the study. The final clinical diagnosis was adjudicated by independent cardiologists or ED physicians experienced in diagnosing HF. The prevalence of HF in the population was 6.18% (48/777). Individuals in the population were African American (31.66%) and Caucasian (64.74%), with the remaining 3.60% represented by other races.

The AUC with a 95% CI for the VITROS NT-proBNP II test for subjects presenting to Cardiology Clinics and otheroutpatient facilities is 0.880 (CI 0.822 to 0.937).

{22}------------------------------------------------

AUC analyses were also performed for relevant clinical subgroups. The AUC with a 95% confidence interval (CI) for the VITROS NT-proBNP II test across gender within the relevant clinical subgroups are presented in the table below. The AUCof the VITROS NT-proBNP II test for each of the relevant clinical subgroups for subjects with or without the condition was greater than or equal to 0.838.

Relevant Clinical Subgroups
Subjects With the ConditionSubjects Without the Condition
SubgroupsAUC95% Confidence IntervalSubgroupsAUC95% Confidence Interval
≥75 Years old(N=139)0.8380.683–0.994<75 Years old(N=638)0.8940.775–1.000
eGFR <60*mL/min/1.73 m²(N=81)0.8430.655–1.000eGFR ≥60mL/min/1.73 m²(N=245)0.9400.888–0.993
BMI ≥30.0 kg/m²(N=430)0.8750.739–1.000BMI <30.0 kg/m²(N=346)0.9300.858–1.000
WithComorbidities**(N=667)0.8920.790–0.993WithoutComorbidities***(N=110)0.9290.799–1.000

*Subjects with renal disease on dialysis were excluded from the study

** Subjects with at least one of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronicobstructive pulmonary disease (COPD)

*** Subjects without any of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructivepulmonary disease (COPD)

Distribution of VITROS NT-proBNP II test results after applying the rule-out cutoff (125 pg/mL) versus adjudicateddiagnosis was summarized for all subjects and is presented in the table below.

VITROS NT-proBNP II Test Results versus Adjudicated Diagnosis
---------------------------------------------------------------------
VITROS NT-proBNP II Test Result InterpretationAdjudicated DiagnosisTotal
HFNon-HF
Positive: Heart Failure Likely44239283
Negative: Heart Failure Unlikely4490494
Total48729777

The clinical performance and the two-tailed 95% Cls of the VITROS NT proBNP II test versus adjudicated diagnosis forsubjects presenting to Cardiology Clinics and other outpatient facilities was determined using the rule-out cutoff (125 pg/mL) and is summarized in the following tables:

All Subjects

GroupCutoff(pg/mL)Sensitivity(%) (n/N)95%CI*(%)Specificity(%) (n/N)95%CI*(%)NPV(%)(n/N)95%CI*(%)PPV(%)(n/N)95%CI*(%)
AllSubjects(N=777)12591.7(44/48)80.0-97.767.2(490/729)63.7-70.699.2(490/494)97.9-99.815.6(44/283)11.5-20.3

*95% Exact Confidence Interval

{23}------------------------------------------------

GroupCutoff(pg/mL)LR-95% CILR+95% CI
All Subjects(N=777)1250.120.05-0.312.802.44-3.20

The clinical performance and the two-tailed 95% CIs of the VITROS NT proBNP II test versus adjudicated diagnosis for subjects presenting to Cardiology Clinics and other outpatient facilities was also determined using the rule-out cutoff (125pg/mL) for relevant clinical subgroups and are summarized in the following tables:

Age Group

RelevantClinicalSubgroupsCutoff(pg/mL)Sensitivity(%) (n/N)95%CI*(%)Specificity(%) (n/N)95%CI*(%)NPV(%) (n/N)95%CI*(%)PPV(%) (n/N)95%CI*(%)
≥75 Years old(N=139)12593.8(15/16)69.8-99.836.6(45/123)28.1-45.797.8(45/46)88.5-99.916.1(15/93)9.3-25.2
<75 Years old(N=638)12590.6(29/32)75.0-98.073.4(445/606)69.7-76.999.3(445/448)98.1-99.915.3(29/190)10.5-21.2

*95% Exact Confidence Interval

Relevant ClinicalSubgroupsCutoff(pg/mL)LR-95% CILR+95% CI
≥75 Years old(N=139)1250.170.03-1.151.481.23-1.78
<75 Years old(N=638)1250.130.04-0.383.412.87-4.06

Renal Disease

RelevantClinicalSubgroupsCutoff(pg/mL)Sensitivity(%) (n/N)95%CI*(%)Specificity(%) (n/N)95%CI*(%)NPV(%) (n/N)95%CI*(%)PPV(%) (n/N)95%CI*(%)
eGFR <60 **mL/min/1.73m2(N=81)12584.6(11/13)54.6-98.145.6(31/68)33.5-58.193.9(31/33)79.8-99.322.9(11/48)12.0-37.3
eGFR ≥ 60mL/min/1.73m2(N=245)125100.0(13/13)75.3-100.068.5(159/232)62.1-74.5100.0(159/159)97.7-100.015.1(13/86)8.3-24.5

*95% Exact Confidence Interval

** Subjects with renal disease on dialysis were excluded from the study

Relevant ClinicalSubgroupsCutoff(pg/mL)LR-95% CILR+95% CI
eGFR <60* mL/min/1.73 m²(N=81)1250.340.09-1.251.561.13-2.14
eGFR ≥ 60 mL/min/1.73 m²(N=245)1250.00N/A3.182.63-3.84

*Subjects with renal disease on dialysis were excluded from the study

N/A: Not applicable

{24}------------------------------------------------

RelevantClinicalSubgroupsCutoff(pg/mL)Sensitivity(%) (n/N)95%CI*(%)Specificity(%) (n/N)95%CI*(%)NPV(%) (n/N)95%CI*(%)PPV(%) (n/N)95%CI*(%)
BMI ≥30.0 kg/m2(N=430)12582.6(19/23)61.2-95.076.2(310/407)71.7-80.298.7(310/314)96.8-99.716.4(19/116)10.2-24.4
BMI <30.0 kg/m²(N=346)125100.0(24/24)85.8-100.055.9(180/322)50.3-61.4100.0(180/180)98.0-100.014.5(24/166)9.5-20.7

*95% Exact Confidence Interval

Relevant ClinicalSubgroupsCutoff(pg/mL)LR-95% CILR+95% CI
BMI≥30.0 kg/m²(N=430)1250.230.09-0.563.472.68-4.48
BMI <30.0 kg/m²(N=346)1250.00N/A2.272.01-2.56

Comorbidities

BMI

Relevant ClinicalSubgroupsCutoff(pg/mL)Sensitivity(%) (n/N)95%CI*(%)Specificity(%) (n/N)95%CI*(%)NPV(%) (n/N)95%CI*(%)PPV(%) (n/N)95%CI*(%)
Withcomorbidities **(N=667)12594.9(37/39)82.7-99.465.8(413/628)61.9-69.599.5(413/415)98.3-99.914.7(37/252)10.6-19.7
Without***comorbidities(N=110)12577.8(7/9)40.0-97.276.2(77/101)66.7-84.197.5(77/79)91.2-99.722.6(7/31)9.6-41.1

*95% Exact Confidence Interval

** Subjects with at least one of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronicobstructive pulmonary disease (COPD)

*** Subjects without any of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructivepulmonary disease (COPD)

Relevant ClinicalSubgroupsCutoff(pg/mL)LR-95% CILR+95% CI
With comorbidities *(N=667)1250.080.02-0.312.772.43-3.16
Without comorbidities **(N=110)1250.290.09-0.993.272.00-5.36

*Subjects with at least one of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronicobstructive pulmonary disease (COPD)

** Subjects without any of the following: diabetes, renal insufficiency (eGFR values <60), hypertension (HTN) and/or chronic obstructivepulmonary disease (COPD)

Results of this test should be used in conjunction with clinical presentation, other diagnostic tests, and in accordance withthe appropriate clinical guidelines.

{25}------------------------------------------------

Correlation of the VITROS NT-proBNP II test results with New York Heart Association (NYHA) Functional Classification in patients diagnosed with HF

The VITROS NT-proBNP II test results were determined from samples from 1143 subjects with heart failure ranging in age from 22 to 106 years. The population consisted of 475/1143 (41.56%) females and 668/1143 (58.44%) males. The descriptive statistics for the VITROS NT-proBNP II test results (pg/mL) were determined across gender and within gender and are summarized in the following tables:

All Subjects

StatisticsNYHA Functional Classification
NYHA Class I*NYHA Class IINYHA Class IIINYHA Class IV
N8190567378
Mean4190501066808170
SD59108410977014100
5th Percentile156291490701
Median1480222036604520
IQR6360433059207810
95th Percentile16000179002310023400

*There were no NYHA Class I female subjects in the study Jonckheere-Terpstra test of trend p <. 0001

Female Subjects

StatisticsNYHA Functional ClassificationNYHA Class I*NYHA Class IINYHA Class IIINYHA Class IV
NN/A85240150
MeanN/A456066509970
SDN/A6150986020100
5th PercentileN/A272481559
MedianN/A219035104790
IQRN/A425063108580
95th PercentileN/A162002440025800

*There were no NYHA Class I female subjects in the study N/A: Not applicable

Jonckheere-Terpstra test of trend p <. 0001

Male Subjects

StatisticsNYHA Functional Classification
NYHA Class INYHA Class IINYHA Class IIINYHA Class IV
N8105327228
Mean4190538067006980
SD5910989097107760
5th Percentile156404516733
Median1480222037304280
IQR6360438057007140
95th Percentile16000179002020021600

Jonckheere-Terpstra test of trend p <.0001

The Jonckheere-Terpstra test was used to determine that there is a statistically significant relationship between the median VITROS NT-proBNP II test results and HF severity for:

{26}------------------------------------------------

  • . All Subjects NYHA Class II-IV.
  • . Female Subjects NYHA Class II-IV.
  • Male Subjects NYHA Class I–IV.

These results show that there is a relationship between the median VITROS NT-proBNP II test results and HF severity as determined by NYHA Class. The median VITROS NTproBNP II test results increase as the NYHA Classification increasesfrom Class II—IV for females and all subjects, and Class I–IV for male subjects.

10. Conclusions

The nonclinical and clinical data presented in the submission provide a reasonable assurance that the VITROS NT-proBNP II test is safe and effective for the stated intended use and is substantially equivalent to the cleared predicate device.

§ 862.1117 B-type natriuretic peptide test system.

(a)
Identification. The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.(b)
Classification. Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”