K Number
K200336

Validate with FDA (Live)

Date Cleared
2020-11-12

(275 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tyvek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Device Description

Tyvek Sterilization Pouches with Chevron Seal are constructed from an uncoated Tyvek backing of fine, continuous, high-density polyethylene fibers, with front material consisting of a clear, laminated polyethylene terephthalate / low density polyethylene (LDPE) or LDPE-ethylene-vinyl acetate copolymer film. The pouches are used to enclose medical devices that are to be sterilized by a healthcare provider in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Following manufacturer's instructions, devices are inserted into Tyvek Sterilization Pouches and sealed. The self-seal pouch permits sealing of the pouch without heatsealing equipment, whereas the heat-sealable pouches must be heat sealed prior to the cycle. Andersen Sterilizers' AN85 EO Indicators, when placed on the outside of the sterilization pouches. indicate EO exposure and offer a convenient way to differentiate pouches that have been processed in EO sterilization cycles from unprocessed units. The color of the AN85 EO Indicators changes from yellow-green to blue after exposure to EO. After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for 6 months.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (Tyvek Sterilization Pouches with Chevron Seal), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with detailed acceptance criteria and performance metrics for a novel AI/ML-driven device.

Therefore, the requested information about acceptance criteria, study design, sample sizes, expert involvement, MRMC studies, standalone performance, and ground truth establishment for an AI/ML device is not applicable to this document.

This submission is for a physical product (sterilization pouches) and details non-clinical performance testing related to its material properties, sterilization efficacy, and shelf life, not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 12, 2020

Andersen Sterilizers, Inc. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258

Re: K200336

Trade/Device Name: Tyvek Sterilization Pouches with Chevron Seal Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: October 8, 2020 Received: October 13, 2020

Dear William Andersen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Elizabeth Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200336

Device Name

Tyvek Sterilization Pouches with Chevron Seal

Indications for Use (Describe)

Tyvek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in 3-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Critical process parameters for the cycle are summarized in Table 1.

Table 1. Critical sterilization cycle parameters in the EOGas 4 Ethylene Oxide Gas Sterilizer

Ethylene OxideTemperatureRelative HumidityEO Exposure TimeTotal Cycle Time
17.6 g ± 5%50°C ± 3°C35-70%3 hours3.5 hours
6 hours7 hours

The EOGas 4 Ethylene Oxide Gas Sterilizer 3-hour gas exposure is used for surface sterilization of medical devices, including instruments with diffusion-restricted spaces (hinges or mated surfaces), as well as for the sterilization of endoscopes working length shorter than 1100 mm as specified in the sterilizer labeling.

The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure is used for sterilization of duodenoscopes and colonoscopes with working length longer than 1100 mm as specified in the sterilizer labeling.

{3}------------------------------------------------

Device TypeMaximum LoadDevice ExamplesRequired Aeration
Metal24 lbs (11 kg)Surgical instruments,delicate sharps,including those withhinges and matedsurfacesNo additional aerationrequired;Follow pouch or wrapmanufacturer'sinstructions (Example: Tyvepouches require ≥ 6 hoursat 50°C)
Plastic7.0 lbs (3.2 kg)Reusable power cords,trocars, and similardevices24 hours at 50°C;Follow manufacturer'sinstructions
Fabric6.1 lbs (2.8 kg)Reusable cloth gowns,towels, and similar devices
≤ 1100 mmWorkingLumenLengthEndoscopesOne (1)≥ 2.0 mm ID biopsy channel≤ 1100 mm working lengthGastrovideoscopes,gastrointestinal videoscopes,and similar devices8 hours at 50°C;Follow manufacturer'sinstructions
Four (4)> 1.2 mm ID biopsy channel≤700 mm working lengthBronchoscopes,bronchovideoscopes, cystoscopes,ureteroscopes, choledocoscopes,

6-Hour EO Exposure, EOGas 4 Endo-SteriTest PCD (Gold Purge Probe)

Two (2) Duodenoscopes*Olympus TJF-Q180V,
≥ 2.0 mm ID biopsy channelOlympus TJF-Q160VF,6 hours at 50°C for
≤ 1250 mm working lengthOlympus TJF-Q190V,Olympus and Pentax
≥ 1.2 mm ID, ≤ 3530 mmmaximum length of anychannelOlympus PJF-160,Fujifilm ED-530XT,endoscopes in Sterisheet
Pentax ED34-i10T2,
>1100 mmPentax ED-3490TK8 hours at 50°C for
WorkingTwo (2) Colonoscopes*Fujifilm endoscopes
Lumen≥ 3.7 mm ID biopsy channelOlympus CF-Q180AL,in Sterisheet
Length≤ 1700 mm working lengthFujifilm EC-600HL,Follow manufacturer's
Endoscopes≥ 1.2 mm ID, ≤ 3530 mmmaximum length of any channelPentax EC-3490Liinstructions
  • One (1) duodenoscope may also be paired with one (1) colonoscope

After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for 6 months. Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

∑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary K200336

Applicant's Name and AddressAndersen Sterilizers, Inc.Establishment Registration Number 30046347103154 Caroline DriveHaw River, NC 27258
Contact PersonWilliam K. Andersen, BE, MD, FAAOSPresidentPhone: 336-376-8622, Fax: 336-376-5428
Date of PreparationFebruary 3, 2020
Device
Proprietary NameTyvek Sterilization Pouches with Chevron SealCommon NameSterilization PouchClassificationClass II (21 CFR 880.6850)Product CodeFRGProprietary NameTyvek Sterilization Pouches with Chevron SealCommon NameSterilization PouchClassificationClass II (21 CFR 880.6850)Product CodeFRG
Proprietary NameTyvek Sterilization Pouches with Chevron Seal
Common NameSterilization Pouch
ClassificationClass II (21 CFR 880.6850)
Product CodeFRG
Predicate Device
Device NameTyvek Sterilization Pouches with Chevron Seal510(k) numberK152058ManufacturerAmcor Flexibles, Inc.Device NameTyvek Sterilization Pouches with Chevron Seal510(k) numberK152058ManufacturerAmcor Flexibles, Inc.
Device NameTyvek Sterilization Pouches with Chevron Seal
510(k) numberK152058
ManufacturerAmcor Flexibles, Inc.

The predicate Tyvek Sterilization Pouches with Chevron Seal (K152058) were cleared as an
accessory for use in 3-hour gas exposures in the EOGas 4 Ethylene Oxide Gas SterilizationSystem manufactured by Andersen Sterilizers, Inc.

The 510(k) submission modifies the indications for use of the predicate device in order to include Tyvek Sterilization Pouches with Chevron Seal as an accessory for use in 6-hour gas exposures in the EOGas 4 Ethylene Oxide Sterilization System manufactured by Andersen Sterilizers, Inc. No modifications were made to the manufacturing method, technology, or intended use.

{6}------------------------------------------------

Device Description

Tyvek Sterilization Pouches with Chevron Seal are constructed from an uncoated Tyvek backing of fine, continuous, high-density polyethylene fibers, with front material consisting of a clear, laminated polyethylene terephthalate / low density polyethylene (LDPE) or LDPE-ethylene-vinyl acetate copolymer film. The pouches are used to enclose medical devices that are to be sterilized by a healthcare provider in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Following manufacturer's instructions, devices are inserted into Tyvek Sterilization Pouches and sealed. The self-seal pouch permits sealing of the pouch without heatsealing equipment, whereas the heat-sealable pouches must be heat sealed prior to the cycle. Andersen Sterilizers' AN85 EO Indicators, when placed on the outside of the sterilization pouches. indicate EO exposure and offer a convenient way to differentiate pouches that have been processed in EO sterilization cycles from unprocessed units. The color of the AN85 EO Indicators changes from yellow-green to blue after exposure to EO. After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for 6 months.

Indications for Use

Tyvek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 5-1.

Ethylene OxideTemperatureRelativeHumidityEthylene OxideExposure TimeTotal Cycle Time
17.6 g ± 5%50°C ± 3°C35-90%3 hours3.5 hours
6 hours7 hours

Table 5-1. Critical sterilization cycle parameters for EOGas 4 Ethylene Oxide Gas Sterilizer

The EOGas 4 Ethylene Oxide Gas Sterilizer 3-hour gas exposure is used for surface sterilization of medical devices, including instruments with diffusion-restricted spaces (hinges or mated surfaces), as well as for the sterilization of endoscopes working length shorter than 1100 mm as specified in the sterilizer labeling.

The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure is used for sterilization of duodenoscopes and colonoscopes with working length longer than 1100 mm as specified in the sterilizer labeling.

{7}------------------------------------------------

Device TypeMaximum LoadDevice ExamplesRequired Aeration
3-Hour EO Exposure, EOGas 4 SteriTest PCD (Blue Purge Probe)
Metal24 lbs(11 kg)Surgical instruments,delicate sharps,including those withhinges and matedsurfacesNo additional aerationrequired;Follow pouch or wrapmanufacturer'sinstructions (Example:Tyvek pouches require≥ 6 hours at 50°C)
Plastic7.0 lbs(3.2 kg)Reusable power cords,trocars, and similardevices24 hours at 50°C;
Fabric6.1 lbs(2.8 kg)Reusable cloth gowns,towels, and similardevicesFollow manufacturer'sinstructions
≤ 1100 mmWorkingLumenLengthEndoscopesOne (1)≥ 2.0 mm ID biopsy channel≤ 1100 mm working lengthGastrovideoscopes,gastrointestinalvideoscopes, and similardevices8 hours at 50°C if inSterisheet;
Four (4)≥ 1.2 mm ID biopsy channel≤ 700 mm working lengthBronchoscopes,bronchovideoscopes,cystoscopes,ureteroscopes,choledocoscopes, andsimilar devicesFollow manufacturer'sinstructions
6-Hour EO Exposure, EOGas 4 Endo-SteriTest PCD (Gold Purge Probe)
>1100 mmWorkingLumenLengthEndoscopesTwo (2) Duodenoscopes *≥ 2.0 mm ID biopsy channel≤ 1250 mm working length≥ 1.2 mm ID, ≤ 3530 mmmaximum length of anychannelOlympus TJF-Q180V,Olympus TJF-Q160VF,Olympus TJF-Q190V,Olympus PJF-160,Fujifilm ED-530XT,Pentax ED34-i10T2Pentax ED-3490TK6 hours at 50°C forOlympus and Pentaxendoscopes inSterisheet;
Two (2) Colonoscopes ≥ 3.7 mm ID biopsy channel≤ 1700 mm working length≥ 1.2 mm ID, ≤ 3530 mmmaximum length of anychannel One (1) duodenoscope may also be paired with one (1)Olympus CF-Q180AL,Fujifilm EC-600HL,Pentax EC-3490Li8 hours at 50°C forFujifilm endoscopes inSterisheet;Follow manufacturer'sinstructions
colonoscope

Table 5- 2. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer

After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for 6 months.

{8}------------------------------------------------

Technological Characteristics

The technological characteristics of Tyvek Sterilization Pouches with Chevron Seal are identical to the predicate device (K152058) - both are intended for the same use, use the same technology, and are designed in the same way.

The subject Tyvek Sterilization Pouches with Chevron Seal differ from the predicate device only in the indicated sterilization cycles. A comparison between the devices is listed in Table 5-3.

{9}------------------------------------------------

ElementsPredicate Device: Tyvek SterilizationPouches (K152058)Subject Device: TyvekSterilization Pouches (K200336)
Intended UseTo enclose medical devices, allow sterilizationof the enclosed devices, and maintain sterilityof the enclosed devicesIdentical
Indications forUseTyvek Sterilization Pouches with Chevron Sealare intended to enclose medical devices thatare to be sterilized in a single pouchconfiguration at a healthcare facility. They areused in 3-hour gas exposures at 50°C in theEOGas 4 Ethylene Oxide Gas Sterilizer.Tyvek Sterilization Pouches withChevron Seal are intended toenclose medical devices that areto be sterilized in a single pouchconfiguration at a healthcarefacility. They are used in 3-hourand 6-hour gas exposures at 50°Cin the EOGas 4 Ethylene OxideGas Sterilizer.
DesignAdhesive laminated film is a clear, highstrength material;Uncoated Tyvek is compatible with EOsterilization, resistant to microbial penetration,and resistant to punctureIdentical
Pouch TypesSelf-seal pouch; Heat seal pouch; Heat sealtubingIdentical
DeviceConstructionSelf-seal and heat seal pouches: front and backmaterials are heat sealed together on threesides; fourth side (end) remains open forfilling; end is sealed by heat (heat sealpouches) or by removing protective liner strip,folding along the pre-fold, and pressing to thefilm (self-seal pouches).Heat seal tubing: front and back materials areheat sealed together on two sides; two ends areopen for selecting size and filling; ends aresealed by heat.Identical
MaterialsClear laminated PET/LDPE or LDPE-EVAfilm (front) and uncoated HDPE Tyvek (back)Identical
ConfigurationSingle pouch configurationIdentical
Shelf Life5 years from date of manufactureIdentical
BiocompatibilityMaterials and biological evaluations (AgarDiffusion Test, Cytotoxicity Test) meet ISO11607-1 requirementsIdentical
Maintenance ofSterilitySterility is maintained for 3 months afterprocessing in the EOGas 4 Ethylene Oxide GasSterilizerSterility is maintained for 6months after processing in theEOGas 4 Ethylene Oxide GasSterilizer
Aeration Time≥ 6 hours≥ 6 hours
PackageIntegritySeal strength, microbial barrier, burst, and peelopen characteristics meet ISO and ASTMrequirementsIdentical

Table 5-3. Device Comparison

{10}------------------------------------------------

Summary of Non-Clinical Testing

Performance Testing

Performance testing was conducted to demonstrate that Tyvek Sterilization Pouches with Chevron Seal perform as intended to allow sterilization of the enclosed medical devices. Sterilization efficacy testing demonstrated a sterility assurance level of 10 ° using the half dose validation method under worst-case conditions. Tyvek Sterilization Pouches with Chevron Seal also maintain sterility of the enclosed devices as intended. Shelf life studies demonstrated after completion of the EOGas 4 sterilization process, sterility is maintained for 6 months. The performance of Tyvek Sterilization pouches is summarized in Table 5-4.

Table 5-4. Summary of bench tests performed to demonstrate safety and effectiveness of Tyvek
Sterilization Pouches
TestDescriptionResults
Compliance to ISO 11607-1
PackageIntegritySeal strength performance characteristics were maintained forthe manufactured seal.Microbial Barrier: the contents of pouches were sterile whenthe processed pouches were subjected to the microbial aerosolchallenge test.Burst: ability to withstand the internal pressurization wasmaintained.Peel open characteristics were maintained.Meet ISO andASTMrequirements
MaterialCompatibilitySeal strength test, microbial barrier properties, burst test, andpeel open test were studied to demonstrate materialcompatibility characteristics of the TyvekSterilization PouchesPass
BiocompatibilityNot direct patient-contacting devices;Materials are non-toxic and meet ISO 11607-1 requirements;Biological and residual evaluations meet acceptable criteria;Provides reasonable assurance for safetyPass
Shelf LifePhysical properties and microbial barrier of the processedTyvek Pouches was verified at the end of shelf life of 5 years;Stability demonstrates reasonable assurance for effectivenessPass
Performance in the 6-hour Gas exposure in the EOGas 4 Ethylene Oxide Gas Sterilizer
SterilantPenetrationEO penetrated the pouch under worst-case half-doseconditions or a worst-case biological challenge, andinactivated 6-Log biological indicatorsConfirmed SALof 10-6 in 6-hourgas exposures
Maintenance ofPackageSterilitySterility was maintained for 6 months after processing in 6-hour gas exposures in the EOGas 4 Ethylene Oxide GasSterilizer.Pass

{11}------------------------------------------------

Conclusion

The conclusions drawn from the nonclinical test demonstrate that the Tyvek Sterilization Pouches with Chevron Seal is as safe, as effective, and performs as well as or better than the legally marketed Tyvek Sterilization Pouches with Chevron Seal (K152058)

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).