K Number
K152058

Validate with FDA (Live)

Date Cleared
2015-11-18

(117 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tyvek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Device Description

The Tyvek Sterilization Pouches with Chevron Seal are constructed from an uncoated Tyvek backing of fine, continuous high-density polyethylene (HDPE) fibers, with front material consisting of a clear, laminated polyethylene terephthalate (PET) / low density polyethylene (LDPE) or LDPE-ethylene-vinylacetate copolymers (EVA) film. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 3 hour sterilization cycle at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer. Devices are inserted into the pouches and sealed. The self-seal pouch permits sealing of the pouch without heat-sealing equipment, while the heat sealed pouches are heat sealed prior to the cycle. After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for at least 3 months (90 days) or until used.

AI/ML Overview

This document describes the Andersen Sterilizers, Inc. Tyvek Sterilization Pouches with Chevron Seal (K152058).

The acceptance criteria and related study details are as follows:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Seal strength performance characteristicsMaintained for the manufactured seal; Meets ISO and ASTM requirements
Microbial BarrierContents of pouches sterile when subjected to microbial aerosol challenge; Meets ISO and ASTM requirements
Burst abilityWithstood internal pressurization; Meets ISO and ASTM requirements
Peel open characteristicsMaintained; Meets ISO and ASTM requirements
Material CompatibilityPass (seal strength, microbial barrier, burst, and peel open tests)
BiocompatibilityMaterials are non-toxic, FDA compliant, meet ISO 11607-1, biological evaluations meet acceptable criteria; Pass
Shelf Life (Physical properties, microbial barrier)Physical properties and microbial barrier of processed pouches verified at end of 5 years; Pass
Sterilant Penetration (Sterility Assurance Level)Inactivated 6-Log biological indicators under worst-case half-dose conditions; Achieved 10^-6 sterility assurance level for the sterilization cycle
Maintenance of Package Integrity (Sterility)Sterility maintained for at least 3 months (90 days) after EOGas 4 processing; Pass

2. Sample size used for the test set and the data provenance:

  • The document does not explicitly state the sample sizes for each specific test mentioned (e.g., number of pouches tested for seal strength, microbial barrier, etc.).
  • The data provenance is not specified beyond indicating that "Performance testing was conducted." It does not mention the country of origin of the data or whether it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided in the document. The document describes performance testing against established standards (ISO, ASTM) and internal criteria, rather than relying on human expert assessment for ground truth.

4. Adjudication method for the test set:

  • Not applicable. The tests are objective and based on established technical standards and measurements, not subjective human assessment requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a sterilization pouch and not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is not an algorithm. The performance testing described is for the physical and functional characteristics of the sterilization pouch itself.

7. The type of ground truth used:

  • The ground truth for the performance tests is based on established industry standards and regulatory requirements, such as ISO 11607-1, ASTM standards, FDA compliance for materials, and specific biological indicator inactivation levels (e.g., 6-Log reduction for sterility assurance).

8. The sample size for the training set:

  • Not applicable. This device is not a machine learning model, so there is no training set.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized human figure in profile, with three faces overlapping to represent the department's focus on health and human well-being. The emblem is simple and modern in design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2015

Andersen Sterilizers, Inc. William Andersen, BE, MD, FAAOS President 3154 Caroline Drive Haw River, North Carolina 27258

Re: K152058

Trade/Device Name: Tyvek Sterilization Pouches With Chevron Seal Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: October 20, 2015 Received: October 22, 2015

Dear Dr. William Andersen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin I. Keith -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152058

Device Name

Tyvek® Sterilization Pouches with Chevron Seal

Indications for Use (Describe)

The Tyyek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterlized in a single pouch configuration at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Critical process parameters for the cycle are summarized below in Table 1.

Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer

EO AmountTemperatureRelative HumidityEO Exposure TimeTotal Cycle Time
17.6 g ± 5%50°C ± 3°C35-90%3 hours3.5 hours

The 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilize a load of up to 24 lbs of metal surgical instruments or 7 lbs of plastic devices (combined weight of pouched devices and tray).

The product codes for Tyvek Sterilization Pouches are listed in Table 2.

Table 2. Product codes for Tyvek Sterilization Pouches

Product TypeSelf SealHeat SealTubing
Product CodesTLP-030TLP-031TLP-033 TLP-058 TLP-059

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

5.1 Applicant's Name and AddressAndersen Sterilizers, Inc.3154 Caroline DriveHaw River, NC 27258
5.2 Contact PersonWilliam K. Andersen, BE, MD, FAAOSPresidentPhone: 336-376-8622, Fax: 336-376-5428
5.3 Date of PreparationNovember 17, 2015
5.4 DeviceProprietary NameCommon NameClassificationProduct CodeTyvek® Sterilization Pouches with Chevron SealTLP-030, TLP-031, TLP-033, TLP-058, TLP-059Sterilization PouchClass II (21 CFR 880.6850)FRG
5.5 Predicate DeviceDevice Name510(k) numberManufacturerTyvek Sterilization Pouches with Chevron SealK831425Amcor Flexibles

This 510(k) submission modifies the indications for use of the predicate device in order to include it as a component in the EOGas 4 ethylene oxide sterilization system. No modifications were made to the technology or intended use.

ર : ર Device Description

The Tyvek Sterilization Pouches with Chevron Seal are constructed from an uncoated Tyvek backing of fine, continuous high-density polyethylene (HDPE) fibers, with front material consisting of a clear, laminated polyethylene terephthalate (PET) / low density polyethylene (LDPE) or LDPE-ethylene-vinylacetate copolymers (EVA) film. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 3 hour sterilization cycle at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer. Devices are inserted into the pouches and sealed. The self-seal pouch permits sealing of the pouch without heat-sealing equipment, while the heat sealed pouches are heat sealed prior to the cvcle. After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for at least 3 months (90 days) or until used.

The AN85/AN86 EO Indicators, when placed on the outside of the sterilization pouches, indicate ethylene oxide exposure and offer a convenient way to verify processing in the sterilization

{5}------------------------------------------------

cycle. The color of the AN85/AN86 EO Indicators changes from yellow-green to blue after exposure to ethylene oxide.

Indications for Use 5.7

The Tyvek Sterilization Pouches with Chevron Seal are intended to be used to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 5-1.

Table 5-1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer

Ethylene OxideTemperatureRelativeHumidityEthylene OxideExposure TimeTotal CycleTime
17.6 g ± 5%50°C ± 3°C35-90%3 hours3.5 hours

The 3 hour cvcle in the EOGas 4 Ethylene Oxide Gas Sterilizer has been validated to sterilize a load of up to 24 lbs of metal surgical instruments or 7 lbs of plastic devices (combined weight of pouched devices and trav).

The product codes for Tyvek Sterilization Pouches are listed in Table 5-2.

Product TypeSelf-SealHeat SealTubing
Product CodesTLP-030TLP-031TLP-033TLP-058TLP-059

Table 5-2. Product codes for Tyvek Sterilization Pouches

Device Comparison ર .8

The technological characteristics of the Tyvek Sterilization Pouches with Chevron Seal are identical to the predicate device - both are intended for the same use, use the same technology, and are designed in the same way.

The Tyvek Sterilization Pouches with Chevron Seal differ from the predicate device in the sterilization cycles used. The difference raises no issues related to safety or effectiveness of the subject device in the sterilization cycle. A comparison between the devices is listed in Table 5-3.

ElementsPredicate Device: TyvekSterilization PouchesSubject Device: Tyvek SterilizationPouches
510(k) NumberK831425K152058
Intended UseTo enclose medical devices, allowsterilization of the enclosed devices,and maintain sterility of the encloseddevicesIdentical
Indications forUseTo be used in ethylene oxidesterilization systemsThe Tyvek Sterilization Pouches withChevron Seal are intended to be used toenclose medical devices that are to besterilized in a single pouch configuration ata healthcare facility. They are used in the 3hour sterilization cycle at 50°C in theEOGas 4 Ethylene Oxide Gas Sterilizer.The 3 hour cycle in the EOGas 4 EthyleneOxide Gas Sterilizer has been validated tosterilize a load of up to 24 lbs of metalsurgical instruments or 7 lbs of plasticdevices (combined weight of poucheddevices and tray).
DesignAdhesive laminated film is a clear,high strength material;Uncoated Tyvek is compatible withEO sterilization, resistant to microbialpenetration, and resistant to punctureIdentical
Pouch TypesSelf-seal pouch; Heat seal pouch; Heatseal tubingIdentical
DeviceConstructionSelf-seal and heat seal pouches: frontand back materials are heat sealedtogether on three sides; fourth side(end) remains open for filling; end issealed by heat (heat seal pouches) orby removing protective liner strip,folding along the pre-fold, andpressing to the film (self-seal pouches).Heat seal tubing: front and backmaterials are heat sealed together ontwo sides; two ends are open forselecting size and filling; ends aresealed by heat.Identical
MaterialsClear laminated PET/LDPE or LDPE-EVA film (front) and uncoated HDPETyvek (back)Identical
Configuration inLoadSingle pouch configurationIdentical
Shelf Life5 years from date of manufactureIdentical
BiocompatibilityMaterials and biological evaluations(Agar Diffusion Test, CytotoxicityTest) meet ISO 11607-1 requirementsMaterials and biological evaluations (AgarDiffusion Test, Cytotoxicity Test) meetISO 11607-1 requirements
Maintenance ofSterilityTyvek pouches are capable ofmaintaining sterility for at least 5 yearsSterility is maintained for at least 3 months(90 days) after processing in an EOGas 4 sterilizer
Aeration Time≥ 6 hours≥ 6 hours
Package IntegritySeal strength, microbial barrier, burst,and peel open characteristics meet ISOand ASTM requirementsSeal strength, microbial barrier, burst, andpeel open characteristics meet ISO andASTM requirements

Table 5-3. Device Comparison

{6}------------------------------------------------

Г

{7}------------------------------------------------

5.9 Performance Testing

Performance testing was conducted to show that the Tyvek Sterilization Pouches with Chevron Seal perform as intended to allow sterilization of the enclosed medical device. Sterilization efficacy testing demonstrated a 12-Log reduction and a sterility assurance level of 10° using the overkill method and half cycle or half dose validation methods under worst case conditions. The Tyvek Sterilization Pouches with Chevron Seal also maintain sterility of the enclosed device as intended. Shelf life studies demonstrated sterility maintenance for up to 5 years after completion of an ethylene oxide sterilization process. The performance of the Tyvek Sterilization pouches is summarized in Table 5-4.

Table 5-4. Summary of bench tests performed to demonstrate safety and effectiveness of the
Tyvek Sterilization Pouches
TestDescriptionResults
Compliance to ISO 11607-1
PackageIntegritySeal strength performance characteristics weremaintained for the manufactured seal.Microbial Barrier: the contents of pouches were sterilewhen the processed pouches were subjected to themicrobial aerosol challenge test.Burst: ability to withstand the internal pressurizationwas maintained.Peel open characteristics were maintained.Meet ISO and ASTMrequirements
MaterialCompatibilitySeal strength test, microbial barrier properties, bursttest, and peel open test were studied to demonstratematerial compatibility characteristics of the TyvekSterilization PouchesPass
BiocompatibilityNot direct patient-contacting devices;Materials are non-toxic, FDA compliant, and meetISO 11607-1 requirements;Biological evaluations meet acceptable criteria;Provides reasonable assurance for safetyPass
Shelf LifePhysical properties and microbial barrier of theprocessed Tyvek Pouches was verified at the end ofshelf life of 5 years;Stability demonstrates reasonable assurance foreffectivenessPass
Performance in the EOGas 4 Cycle
SterilantPenetrationThe sterilant penetrated the pouch under worst casehalf-dose conditions and inactivated 6-Log biologicalindicatorsAllow a sterility assurancelevel of 10-6 for thesterilization cycle
Maintenance ofPackageIntegritySterility was maintained for at least 3 months(90 days) after the EOGas 4 processing.Pass

{8}------------------------------------------------

5.10 Conclusion

Based on the intended use, technological characteristics, performance data and nonclinical tests performed, the subject Tyvek Sterilization Pouches with Chevron Seal is substantially equivalent to the predicate device - Tyvek Sterilization Pouches with Chevron Seal (K831425).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).