(262 days)
The EXPERT - Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting.
The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following:
Mini EXPERT
When used for fixation of bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as:
-
Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions;
-
Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and
-
Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the ligament at the base of the thumb metacarpal by providing stabilization between the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
EXPERT Knotless and EXPERT Knotless Dual
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
The EXPERT – Joint Fixation System is available in three models. Each model is composed of different medical devices (i.e. plates, surgical suture) assembled together for the purpose of use as a set. Each set is composed of plates from different sizes and shapes made of Titanium Alloy (ASTM F136) associated with non-absorbable surgical sutures and/or surgical meshes made of an Ultra-High Molecular Weight Polyethylene (UHMWPE).
The metallic components are manufactured by GM dos Reis Indústria e Comércio Ltda and the nonabsorbable Surgical Sutures and Surgical Suture Meshes are purchased in bulk from River Point Medical (K100006).
Each EXPERT model comes with a transient use guidewire accessory in the same package. The guidewire is made of Stainless Steel (ASTM F138) to aid in the EXPERT - Joint Fixation System placement.
In order to promote a correct placement of EXPERT – Joint Fixation System, GMReis has also available a range of instruments (class I exempt) to serve the surgeon such as drill bits, drill guides, plier, among others. GMReis recommends the use of these instruments in order to ensure the compatibility with the subject devices and promote the success of the procedure.
This document describes the EXPERT - Joint Fixation System and its substantial equivalence to predicate devices, focusing on the mechanical performance of the device rather than AI/software performance. Therefore, many of the typical acceptance criteria and study design elements found in AI/ML medical device submissions (such as ground truth establishment involving expert consensus, MRMC studies, or standalone algorithm performance) are not applicable here.
The "study" in this context refers to mechanical testing that demonstrates the device's physical performance characteristics are comparable to legally marketed predicate devices.
Here's a breakdown based on the provided document:
Acceptance Criteria and Device Performance
The acceptance criteria are implicitly defined by demonstrating substantial equivalence to existing predicate devices through mechanical testing. The document states:
- "The performance of the subject devices compared with the predicate devices are demonstrated through mechanical testing."
- "The documentation submitted in this premarket notification demonstrates that the subject devices have comparable features and performance and, therefore, are substantially equivalent to the identified predicate devices."
Since this is a 510(k) submission for a physical medical device (joint fixation system) and not an AI/ML diagnostic or prognostic tool, the "performance" is related to its physical and functional characteristics (e.g., strength, durability, fixation capability) as compared to the predicates. The document does not explicitly list quantitative acceptance criteria values or detailed reported device performance data in a table format. Instead, it relies on the conclusion that mechanical tests confirm comparability.
Table of Acceptance Criteria and Reported Device Performance (Inferred from 510(k) Context):
| Acceptance Criteria Category (Implicit) | Specific Criteria (Inferred) | Reported Device Performance (Inferred) |
|---|---|---|
| Mechanical Performance | Comparable mechanical properties (e.g., stiffness, strength, durability, fatigue resistance) to predicate devices for intended use. | Mechanically tested to demonstrate comparable features and performance to predicate devices (K090107, K061925, K043248, K111032). The submission concludes that "Any differnce in the technological characteristics do not raise new issues of safety or efficacy." Specific numerical values for the tests are not provided in this summary document, but are implied to have been submitted to the FDA and met the internal acceptance parameters for substantial equivalence. |
| Material Biocompatibility | Biocompatibility in accordance with ISO 10993-1 and AAMI/ANSI ST72. (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-mediated Pyrogenicity Testing, Pyrogenicity monitoring). | Biocompatibility confirmed for Titanium Alloy (referenced K180626, K182718), UHMWPE surgical sutures (referenced K100006), and Stainless Steel guidewire (specific test reports cited). |
| Sterilization & Shelf Life | Sterilization validity per ISO 11135 and a shelf life of 3 years. | Ethylene oxide sterilization validation performed according to ISO 11135. Device provided sterile with a 3-year shelf life. |
Study Information (Mechanical Testing for Substantial Equivalence)
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not specified in the provided summary. For mechanical testing of medical devices, sample sizes are typically determined by statistical rationale to ensure robustness and representativeness of the device's performance.
- Data Provenance: The testing was conducted as part of the submission by GM dos Reis Indústria e Comércio Ltda, a Brazilian company. The data would be prospective, as it was generated specifically for this 510(k) submission.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not Applicable. This is a submission for a physical joint fixation system, not an AI/ML device requiring human expert annotation of medical images or clinical outcomes for ground truth. Ground truth for mechanical performance is established through standardized engineering tests and material science evaluations.
-
Adjudication Method for the Test Set:
- Not Applicable. See point 2.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- Not Applicable. This type of study is relevant for AI/ML devices that assist human readers (e.g., radiologists) in diagnostic tasks. The EXPERT - Joint Fixation System is a physical implant.
-
Standalone (Algorithm Only) Performance:
- Not Applicable. No algorithms or software are involved in the primary function or assessment of this physical device.
-
Type of Ground Truth Used:
- Engineering Specifications and Standardized Test Methods: For mechanical performance, ground truth is established by universally accepted engineering principles, material standards (e.g., ASTM, ISO), and validated test methods (e.g., fatigue testing, tensile strength). For biocompatibility, it's established by adherence to ISO 10993 series standards and AAMI/ANSI ST72.
-
Sample Size for the Training Set:
- Not Applicable. There is no "training set" in the context of mechanical testing for a physical medical device. Device design and manufacturing processes are refined through engineering iterations.
-
How the Ground Truth for the Training Set Was Established:
- Not Applicable. See point 7.
Summary of Information NOT Applicable to this Submission:
Due to the nature of the device (a physical joint fixation system) and the regulatory submission (510(k) based on substantial equivalence through mechanical testing), the following elements, commonly associated with AI/ML device evaluations, are not found in the provided document:
- Detailed quantitative performance metrics in a readily available table.
- Specific sample sizes for a "test set" in the context of clinical/image data.
- Data provenance involving patient demographics or retrospective/prospective medical data collection.
- Number/qualifications of experts for ground truth establishment.
- Adjudication methods.
- MRMC studies.
- Standalone algorithm performance.
- Training set information.
The document unequivocally states: "No clinical data were included in this submission." This further confirms that the evaluation focused on physical and mechanical properties, not clinical outcomes or human performance with AI assistance.
{0}------------------------------------------------
October 29, 2020
GM dos Reis Industria e Comercio Ltda % Graziela Brum Regulatory Affairs Specialist PR Servicos Regulatorios Administrativos Ltda Rua Alice Alem Saadi, 855/2402 Ribeirao Preto, 14096-570 Brazil
Re: K200332
Trade/Device Name: EXPERT - Joint Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: September 29, 2020 Received: October 2, 2020
Dear Graziela Brum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200332
Indications for Use (Describe)
EXPERT - Joint Fixation System
The EXPERT - Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting.
The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following:
Mini EXPERT
When used for fixation of bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as:
-
Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions;
-
Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and
-
Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the ligament at the base of the thumb metacarpal by providing stabilization between the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
EXPERT Knotless and EXPERT Knotless Dual
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ---------------------------------------------- | --------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
ADMINISTRATIVE INFORMATION
| Labels | Values |
|---|---|
| Sponsor/Manufacturer Name | GM dos Reis Indústria e Comércio Ltda Avenida Pierre Simon de La Place 600 Campinas, São Paulo, Brazil 13069-320 Telephone: +55 (19) 3765-9900 |
| Prepared by | Janine Treter and Graziela Brum Passarini Regulatory Affairs PR Serviços Regulatórios Administrativos Ltda E-Mail: graziela@rapassarini.com.br Telephone +55 (16) 3421-8488 |
| Submission date | 29/Sep/2020 |
DEVICE NAME AND CLASSIFICATION
| Labels | Values |
|---|---|
| Trade/ Proprietary Name | EXPERT - Joint Fixation System |
| Common Name | Washer, Bolt Nut |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories |
| Product Code | HTN |
| Classification Regulation | 21 CFR 888.3030, Class II |
| Review Panel | Orthopedic |
PREDICATE DEVICE INFORMATION
| Labels | Values |
|---|---|
| Predicate Devices | K090107 - Mini TightRope - Arthrex, Inc. K061925- Mini TightRopeTM Repair Kit - Arthrex, Inc. K043248 - Arthrex TightRope™ Syndesmosis Devices, Arthrex, Inc. K111032 - CMC Cable FIX - Instratek, Inc. |
| Reference Devices | K100006 - Polyester Non-absorbable Surgical Suture, Polyblend Non- absorbable Surgical Suture, Silk Non-absorbable Surgical Suture, Polypropylene Surgical Suture, Nylon Non-absorbable Surgical Suture, Riverpoint Medical K180626 - Pedimax II - Pedicular Screw Spinal System - GM dos Reis Industria E Comercio Ltda K182718 - Mini and Micro Fragments Reconstruction System - GMReis - GM dos Reis Industria E Comercio Ltda |
{4}------------------------------------------------
INDICATIONS FOR USE
The EXPERT – Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting.
The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following:
Mini EXPERT
When used for fixation of bone-to-bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as:
-
Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions;
-
Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and
-
Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the reconstruction of the ligament at the base of the thumb metacarpal by providing stabilization between the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
EXPERT Knotless and EXPERT Knotless Dual
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
DEVICE DESCRIPTION
The EXPERT – Joint Fixation System is available in three models. Each model is composed of different medical devices (i.g. plates, surgical suture) assembled together for the purpose of use as a set. Each set is composed of plates from different sizes and shapes made of Titanium Alloy (ASTM F136) associated with non-absorbable surgical sutures and/or surgical meshes made of an Ultra-High Molecular Weight Polyethylene (UHMWPE).
The metallic components are manufactured by GM dos Reis Indústria e Comércio Ltda and the nonabsorbable Surgical Sutures and Surgical Suture Meshes are purchased in bulk from River Point Medical (K100006).
Each EXPERT model comes with a transient use guidewire accessory in the same package. The guidewire is made of Stainless Steel (ASTM F138) to aid in the EXPERT - Joint Fixation System placement.
In order to promote a correct placement of EXPERT – Joint Fixation System, GMReis has also available a range of instruments (class I exempt) to serve the surgeon such as drill bits, drill guides, plier, among others. GMReis recommends the use of these instruments in order to ensure the compatibility with the subject devices and promote the success of the procedure.
{5}------------------------------------------------
The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in orthopedic surgeries.
EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles to the following predicate devices:
K090107 - Mini TightRope - Arthrex, Inc. K061925- Mini TightRope™ Repair Kit - Arthrex, Inc.. K043248 - Arthrex TightRope™ Syndesmosis Devices, Arthrex, Inc. K111032 - CMC Cable FIX - Instratek, Inc.
The subject device and the predicate devices have equivalent intended use and equivalent technological characteristics. The subject and predicate devices are all manufactured from identical or equivalent materials and share equivalent design characteristics. The subject and predicate devices encompass equivalent physical dimensions and are to be sterilized by identical method. Any differnce in the technological characteristics do not raise new issues of safety or efficacy.
Biocompatibility of the titanium alloy plates are supported by the reference devices K180626 and K182718. Biocompatibility of the UHMWPE surgical sutures are supported by the reference devices K100006. Biocompatibility of the stainless steel of the guide wire are supported for the following test reports: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity and Materialmediated Pyrogenicity Testing according the FDA guidance entitled Use of International Standard ISO 10993-1,"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" dated June 16, 2016 according to the device categorization and contacting profile. Pyrogenicity monitoring according to AAMI/ANSI ST72 meeting pyrogen limit specifications.
The subject devices are provided sterile and have shelf life of 3 years. Ethylene oxide sterilization validation was performed according to ISO 11135.
The performance of the subject devices compared with the predicate devices are demonstrated through mechanical testing. No clinical data were included in this submission.
CONCLUSION
The documentation submitted in this premarket notification demonstrates that the subject devices have comparable features and performance and, therefore, are substantially equivalent to the identified predicate devices.
N/A