K Number
K182718

Validate with FDA (Live)

Date Cleared
2019-12-23

(451 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mini and Micro Fragments Reconstruction System - GMReis is intended for fracture fixation, arthrodesis, reconstruction, and osteotomy fixation of the hand and wrist. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.

Device Description

Mini and Micro Fragments Reconstruction System – GMReis is composed of plates and screws. The bone plates are made from commercially pure titanium alloy (Ti-4Al-6V) and the bone screws are manufactured from titanium alloy only. The plates range in thickness from 0.6 to 2.0 mm, and the screws range in diameter from 1.5 to 2.7 mm. They are available on different sizes and shapes, according the implantation site and the extension of the fracture.

Mini and Micro Fragments Reconstruction System – GMReis are for single use. The devices are provided non-sterile and must being properly cleaned and sterilized before use, according the recommendations provided in the Instructions for Use.

In order to promote a correct placement of the plates and screws, GMReis has also available a range of instruments (class I exempt) to serve the surgeon such as drills, drill guides, cutting pliers, reamers, screwdrivers, among others. GMReis recommends the use of these instruments in order to ensure the compatibility with the implants and promote the success of the procedure.

The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in osteosyntheses procedures.

AI/ML Overview

This document is a 510(k) premarket notification for the "Mini and Micro Fragments Reconstruction System - GMReis". It describes the device's intended use and demonstrates its substantial equivalence to previously cleared predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes performance testing for biocompatibility and mechanical properties. It does not provide a table with specific acceptance criteria and detailed reported device performance values in a quantifiable manner (e.g., specific tensile strength values or fatigue cycles). Instead, it states that the devices meet the requirements of established standards.

Acceptance Criteria CategoryStandard/Guidance FollowedReported Device Performance (Summary)
BiocompatibilityISO 10993-1Supported by required tests
Mechanical Testing (Plates)ASTM F382Demonstrated through testing
Mechanical Testing (Screws)ASTM F543Demonstrated through testing
Sterilization ValidationISO 17665-1, ISO 17665-2Performed and validated

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes used for the biocompatibility or mechanical testing. It mentions that "tests required" and "mechanical testing" were performed according to the listed standards, implying that sufficient samples were tested to meet the requirements of those standards.

The provenance of the data is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, the manufacturer is GM dos Reis Indústria e Comércio Ltda, located in Campinas, São Paulo, Brazil. The testing would have been conducted by or for this manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not applicable to this type of device. The "Mini and Micro Fragments Reconstruction System - GMReis" is a metallic bone fixation appliance, not a diagnostic or AI-driven device that requires expert review for ground truth establishment. The performance is assessed through standardized physical and mechanical tests.

4. Adjudication Method for the Test Set:

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in diagnostic studies involving multiple human readers, not for mechanical testing of orthopedic implants.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The device is a physical implant, not an AI-assisted diagnostic tool for which human reader performance would be evaluated with and without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

No, a standalone study of an algorithm without human-in-the-loop performance was not done. This device is a physical hardware product, not a software algorithm.

7. Type of Ground Truth Used:

The "ground truth" for this device's performance is established through adherence to recognized international standards for biocompatibility (ISO 10993-1) and mechanical properties (ASTM F382, ASTM F543) and sterilization (ISO 17665-1, ISO 17665-2). The "truth" is that the device performs as expected under these standardized test conditions. It's not a diagnostic "ground truth" derived from expert consensus or pathology.

8. Sample Size for the Training Set:

Not applicable. This is a physical medical device, not a software or AI model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As this is not a software or AI model, there is no training set or associated ground truth establishment method.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

December 23, 2019

Gm Dos Reis Industria E Comerico Ltda. % Paula Oliveira Quality Manager Passarini Regulatory Affairs of America, LLC 201 S. Biscayne Blvd. Suite 1200 Miami, Florida 33131

Re: K182718

Trade/Device Name: Mini and Micro Fragments Reconstruction System - GMReis Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 18, 2019 Received: November 21, 2019

Dear Paula Oliveira:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K182718

Device Name

Mini and Micro Fragments Reconstruction System - GMReis

Indications for Use (Describe)

Mini and Micro Fragments Reconstruction System - GMReis is intended for fracture fixation, arthrodesis, reconstruction, and osteotomy fixation of the hand and wrist. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary
ADMINISTRATIVE INFORMATION
Sponsor/Manufacturer NameGM dos Reis Indústria e Comércio LtdaAvenida Pierre Simon de La Place 600Campinas, São Paulo, Brazil 130693-320Telephone: +55 (19) 3765-9900
Contact PersonPaula OliveiraQuality Managerqualidade@gmreis.com.brTelephone: +55 (19) 3765-9900
Date Prepared19/Dec/2019
Preparer / Alternate ContactJanine TreterRegulatory Affairs SpecialistPassarini Regulatory Affairsjanine@rapassarini.com.brTelephone +55 (47) 3804 0075
DEVICE NAME AND CLASSIFICATION
Trade/ Proprietary NameMini and Micro Fragments Reconstruction System –GMReis
Common NamesPlate, Fixation, Bone
Classification NamesSingle/multiple component metallic bone fixationappliances and accessories
Product CodesHRS, HWC
Classification Regulations21 CFR 888.3030, Class II
Review PanelOrthopedic
PREDICATE DEVICE INFORMATION
Predicate DevicesK142419 - NEOORTHO Produtos Ortopédicos S/A - NeoorthoProductos Orthopedicos S/A (Primary Predicate)K051567 - APTUS® Titanium Fixation System - Medartis, Inc.K100776 - Synthes 2.4mm12.7 mm Variable Angle LCPforefoot/. Midfoot system - Synthes (USA) LPK081546 - Small Bone Locking Plating System - DePuyOrthopaedics, Inc.K142906 - APTUS® Wrist 2.5 System - Medartis AG
Reference DeviceK180626 – Pedimax II -Pedicular Screw Spinal System - GM dosReis Indústria e Comércio Ltda

{4}------------------------------------------------

INDICATIONS FOR USE

Mini and Micro Fragments Reconstruction System - GMReis is intended for fracture fixation, arthrodesis, reconstruction, and osteotomy fixation of the hand and wrist. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.

DEVICE DESCRIPTION

Mini and Micro Fragments Reconstruction System – GMReis is composed of plates and screws. The bone plates are made from commercially pure titanium alloy (Ti-4Al-6V) and the bone screws are manufactured from titanium alloy only. The plates range in thickness from 0.6 to 2.0 mm, and the screws range in diameter from 1.5 to 2.7 mm. They are available on different sizes and shapes, according the implantation site and the extension of the fracture.

Mini and Micro Fragments Reconstruction System – GMReis are for single use. The devices are provided non-sterile and must being properly cleaned and sterilized before use, according the recommendations provided in the Instructions for Use.

In order to promote a correct placement of the plates and screws, GMReis has also available a range of instruments (class I exempt) to serve the surgeon such as drills, drill guides, cutting pliers, reamers, screwdrivers, among others. GMReis recommends the use of these instruments in order to ensure the compatibility with the implants and promote the success of the procedure.

The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in osteosyntheses procedures.

EQUIVALENCE TO MARKETED DEVICE

The subject device is substantially equivalent in indications and design principles to the following predicate devices:

K142419 - NEOORTHO Produtos Ortopédicos S/A - Neoortho Productos Orthopedicos S/A K051567 – APTUS® Titanium Fixation System - Medartis, Inc. K100776 - Synthes 2.4mm12.7 mm Variable Angle LCP forefoot/.Midfoot system - Synthes (USA) LP K081546 - Small Bone Locking Plating System - DePuy Orthopaedics, Inc. K142906 - APTUS® Wrist 2.5 System - Medartis AG

The subject device and the predicate devices have intended use and have similar technological characteristics. The subject and predicate devices are all manufactured from the same or similar materials and share similar design characteristics, including plate screw holes to accommodate locking and non-locking screws. The subject and predicate devices encompass the same range of physical dimensions, are packaged using similar materials and are to be sterilized by the same methods. Any difference in the technological characteristics do not raise new issues of safety or efficacy.

{5}------------------------------------------------

Biocompatibility of the subject devices were supported by the tests required according to its contact profile as recommended by FDA guidance entitled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".

The performance of the subject devices are demonstrated through mechanical testing of plates and screws according to ASTM F382 and ASTM F543, respectively.

The subject devices are provided non-sterile and have no expiration date defined. Steam sterilization validation was performed according to ISO 17665-1 and 17665-2.

CONCLUSION

The documentation submitted in this premarket notification demonstrates that the subject devices have comparable features and performance and, therefore, are substantially equivalent to the identified predicate device.

N/A