K Number
K200260

Validate with FDA (Live)

Date Cleared
2021-04-26

(448 days)

Product Code
Regulation Number
876.4620
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit: The Ureteral Stent is intended to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not externally.

Percuflex™ Nephroureteral Stent System: The Percuflex Nephroureteral Stent is intended for use in Percutaneous Drainage to establish internal drainage from the ureteropelvic junction to the bladder while maintaining external access to the stent.

Amplatz Anchor™ Catheter System: The Amplatz Anchor™ Catheter System is intended for use in applications of percutaneous drainage, particularly percutaneous nephrostomy, abscess or external biliary drainage.

Device Description

The Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit is designed with the classic pigtail shape on each end and is intended to facilitate drainage internally from the renal pelvis to the bladder and stenting of the ureter without external access after placement. The ureteral stent systems contain a suture to adjust the pigtail shape or to remove the stent. These devices are constructed of a proprietary, biocompatible copolymer called Percuflex™ for patient comfort with a radiopaque resin throughout the catheter tube. Designated Percuflex™ Ureteral Stent Systems are also available with a proprietary Glidex™ hydrophilic material for the reduction of surface friction during placement or with a dissolving distal tip (Temp-Tip™). The dissolving tip internal diameter aids in placement, then dissolves soon after placement to allow a larger lumen for drainage.

The Percuflex™ Nephroureteral Stent System is also designed with the classic pigtail shape on each end and facilitates drainage from the internal ureteropelvic junction to the bladder with external access to the stent. The Nephroureteral Stent System also incorporates the Percuflex™ copolymer and a radiopaque resin throughout the catheter tube. Select models are available with a Temp-Tip™ which dissolves soon after placement. A suture is locked in place with a stopcock to retain the pigtail shape. A luer allows for connection to other drainage devices with a conical fitting.

The Amplatz Anchor™ Catheter System is designed with a unique locking Malecot diamond shape to facilitate percutaneous nephrostomy, abscess or external biliary drainage. The Amplatz Anchor™ Catheter Systems also incorporate the Percuflex™ copolymer and a radiopaque resin throughout the catheter tube. A suture is locked in place with a stopcock to retain the 2-winq design. A luer allows for connection to other drainage devices with a conical fitting.

These Percuflex™ Drainage Catheters and Stents are provided sterile, using 100% ethylene oxide (EO) gas sterilization, and are intended for hospital and single use only. Where long-term use is indicated, it is recommended not to exceed 90 days of indwelling time.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Boston Scientific Percuflex Drainage Catheters and Stents, asserting their substantial equivalence to a predicate device (K924608). The document focuses on non-clinical performance testing conducted to support new Magnetic Resonance (MR) compatibility information and cumulative internal changes to the devices.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format, nor does it specify exact numerical thresholds for "acceptance criteria." Instead, it lists the types of non-clinical tests performed and states that these tests demonstrate "the subject devices are as safe and effective as the predicate devices" and that the changes "do not raise new questions of safety or effectiveness."

However, we can infer the categories of acceptance criteria based on the tests conducted:

Acceptance Criteria CategoryReported Device Performance / Assessment
Magnetic Resonance (MR) Compatibility
Radio Frequency (RF) HeatingTesting performed per ASTM F2182-11a. The submission implies that the results are acceptable, allowing for the addition of MR compatibility information to the Directions for Use without raising new safety/effectiveness questions. Specific numerical values are not provided.
Force Measurement (Displacement)Testing performed per ASTM F2052-14. The submission implies that the results are acceptable, allowing for the addition of MR compatibility information to the Directions for Use without raising new safety/effectiveness questions. Specific numerical values are not provided.
Image ArtifactTesting performed per ASTM F2119-07. The submission implies that the results are acceptable, allowing for the addition of MR compatibility information to the Directions for Use without raising new safety/effectiveness questions. Specific numerical values are not provided.
Material Changes & Biocompatibility
Material Evaluation (Resin additive)Fourier Transform Infrared Spectroscopy (FTIR), Differential Scanning Calorimetry (DSC), Thermogravimetric Analysis (TGA), Oxidative Induction Time (OIT), and Dynamic Contact Angle (DCA) analysis performed. Results "demonstrated that both supplied resins present the same thermal transition characteristics, material type (chemistry), and composition."
Design Verification (Resin additive)Catheter Shaft Tensile, Column Strength, Kink Resistance, and Removal Force testing performed. Renal Coil Resistance-Stent Shaft >5F also performed. The submission indicates these assessed finished device properties were maintained. Specific numerical thresholds or results are not provided.
Biocompatibility (Resin additive)Testing performed per ISO 10993-1, including Cytotoxicity, Sensitization, Intracutaneous irritation, Material mediated pyrogenicity, Acute systemic toxicity, Genotoxicity (Ames and MLA assays), Muscle implantation (4-week and 13-week studies). A Toxicological Risk Assessment was also performed. The submission implies all tests showed acceptable biocompatibility.
Tensile Performance Specification Update
Dissolving Tip to Shaft BondDesign verification testing conducted on aged and unaged samples. The submission states this demonstrated "the dissolving tip to shaft bond design output continues to meet the defined design input requirements after sterilization." Specific numerical thresholds or results are not provided.

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical performance testing, primarily bench testing and material characterization. It does not mention human subject testing or clinical data. Therefore:

  • Sample Size for Test Set: Not applicable in terms of patient samples. The sample sizes for the in vitro or material tests (e.g., number of stent units tested for tensile strength, number of material samples for FTIR) are not specified in this summary.
  • Data Provenance: Not applicable in terms of country of origin for clinical data or retrospective/prospective studies, as no clinical studies are presented. The tests conducted are laboratory-based.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. Ground truth, in the context of diagnostic performance measured against clinical expert consensus or pathology, is not relevant here as this document describes non-clinical, benchtop, and material science testing of a medical device (stent and catheter systems), not a diagnostic algorithm.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human annotators or readers when establishing ground truth for diagnostic studies. Such a process is not described or relevant for the non-clinical testing presented.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document is for a physical medical device (stents and catheters), not an AI/ML-driven diagnostic or assistive device. Therefore, no MRMC study or assessment of AI assistance to human readers was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is for a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests described is inherent in the material science principles, engineering specifications, and established testing standards (e.g., ASTM, ISO 10993). For instance:

  • For MR compatibility, the "ground truth" is compliance with the limits defined by ASTM standards for RF heating, displacement force, and image artifact.
  • For material evaluation, the "ground truth" is that the new material performs identically to the old one in terms of chemical and physical properties as measured by standard analytical techniques.
  • For biocompatibility, the "ground truth" is that the material elicits no adverse biological reactions based on ISO 10993 guidelines.
  • For tensile strength, the "ground truth" is meeting predefined engineering specifications for strength and durability.

8. The Sample Size for the Training Set

Not applicable. This document describes non-clinical testing for a physical medical device, not the development or evaluation of an algorithm or AI system that requires a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set mentioned or implied for this physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 26, 2021

Boston Scientific Corporation Heidi Shearer Senior Regulatory Affairs Specialist Three Scimed Place Maple Grove, MN 55311

Re: K200260

Trade/Device Name: Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit, Percuflex™ Nephroureteral Stent System, Amplatz Anchor™ Catheter System Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD, FGE Dated: March 25, 2021 Received: March 26, 2021

Dear Heidi Shearer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the Percuflex Ureteral Stent System Kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200260

Device Name

Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit, Percuflex™ Nephroureteral Stent System Amplatz Anchor™ Catheter System

Indications for Use (Describe)

Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit:

The Ureteral Stent is intended to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not externally.

Percuflex™ Nephroureteral Stent System:

The Percuffex Nephroureteral Stent is intended for use in Percutaneous Drainage to establish internal drainage from the ureteropelvic junction to the bladder while maintaining external access to the stent.

Amplatz AnchorTM Catheter System:

The Amplatz Anchor™ Catheter System is intended for use in applications of percutaneous drainage, particularly percutaneous nephrostomy, abscess or external biliary drainage.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K200260

510(K) Summary Complying with 21 CFR 807.92

I. SUBMITTER INFORMATION

Submitter Name: Boston Scientific Corporation

Submitter Address:

Three Scimed Place Maple Grove, MN 55311-1566 USA

Telephone: 763-255-0056 Fax: 763-494-2222 e-mail: Heidi.Shearer@bsci.com

Contact person name: Heidi Shearer Date Prepared: April 26, 2021

II. DEVICE INFORMATION

Tables 1 to 4 below summarize the relevant subject device information.

Product Trade NameModel Numbers
Percuflex ™ Ureteral Stent SystemM001235490
M001235500
M001235510
M001235520
M001235530
M001235570
M001235580
M001235590
M001235600
M001235610
M001245490
M001245500
M001245510
M001245520
M001245530
M001245570
M001245580

Table 1: Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit

{4}------------------------------------------------

510(k) Summary Percuflex Drainage Catheters and Stents

Product Trade NameModel Numbers
M001245590
M001245600
M001245610
Percuflex™ Ureteral Stent System KitM001245620
M001245630
M001245640
M001245650
M001245660
M001245720
M001245730
M001245740
M001245750
M001245760

Table 2: Percuflex™ Nephroureteral Stent System

Product Trade NameModel Number
Percuflex™ Nephroureteral Stent SystemM001221360
M001221370
M001221380
M001221390
M001221400
M001221410
M001221420
M001221430
M001231360
M001231370
M001231380
M001231390
M001231400
M001231410
M001231420
M001231430

{5}------------------------------------------------

Table 3: Amplatz Anchor™ Catheter System

Product Trade NameModel Number
Amplatz Anchor™ Catheter SystemM001221000
Amplatz Anchor™ Catheter SystemM001221010
Amplatz Anchor™ Catheter SystemM001221020

Table 4: Additional Device Information

Common orUsual nameClassificationNumberClassificationNameProduct CodeProduct CodeNameRegulatoryClass
Ureteral StentSystem21 CFR Part876.4620Ureteral stentFADstent, ureteralII
NephroureteralStent System21 CFR Part876.4620Ureteral stentFADstent, ureteralII
Biliary CatheterSystem21 CFR Part876.5010Biliary catheterand accessoriesFGEstents, drainsand dilators forthe biliary ductsII

III. PREDICATE DEVICE IDENTIFICATION

Name of Predicate Device

Medi-Tech Ureteral Stent System, Nephroureteral Stent System, Amplatz Anchor System with Locking Malecot (K924608. Cleared 26-Jan-1994)

Predicate devices referenced above have not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Percuflex™ Ureteral Stent System and Percuflex™ Ureteral Stent System Kit is designed with the classic pigtail shape on each end and is intended to facilitate drainage internally from the renal pelvis to the bladder and stenting of the ureter without external access after placement. The ureteral stent systems contain a suture to adjust the pigtail shape or to remove the stent. These devices are constructed of a proprietary, biocompatible copolymer called Percuflex™ for patient comfort with a radiopaque resin throughout the catheter tube. Designated Percuflex™ Ureteral Stent Systems are also available with a proprietary Glidex™ hydrophilic material for the reduction of surface friction during placement or with a dissolving distal tip (Temp-Tip™). The dissolving tip internal diameter aids in placement, then dissolves soon after placement to allow a larger lumen for drainage.

The Percuflex™ Nephroureteral Stent System is also designed with the classic pigtail shape on each end and facilitates drainage from the internal ureteropelvic junction to the bladder with external access to the stent. The Nephroureteral Stent System also incorporates the Percuflex™ copolymer and a radiopaque resin throughout the catheter tube. Select models are available with a Temp-Tip™ which dissolves soon after placement. A suture is locked in place with a stopcock to retain the pigtail shape. A luer allows for connection to other drainage devices with a conical fitting.

The Amplatz Anchor™ Catheter System is designed with a unique locking Malecot diamond shape to facilitate percutaneous nephrostomy, abscess or external biliary drainage. The Amplatz Anchor™ Catheter Systems also incorporate the Percuflex™ copolymer and a radiopaque resin throughout the catheter tube. A suture is locked in place with a stopcock to retain the 2-winq design. A luer allows for connection to other drainage devices with a conical fitting.

{6}------------------------------------------------

These Percuflex™ Drainage Catheters and Stents are provided sterile, using 100% ethylene oxide (EO) gas sterilization, and are intended for hospital and single use only. Where long-term use is indicated, it is recommended not to exceed 90 days of indwelling time.

Table 5, 6, and 7 below provide information on the differences between the Percuflex™ Drainage Catheters and Stents models, in relation to specific device features such as coating, dissolving tip, and the outside diameter and length dimensions.

Product Trade NameProduct Labeled FeaturesModel NumberDimensions(OD/Length)
Percuflex™ UreteralStent SystemPercuflex™ TempTip™ Ureteral StentSystem with Dissolving TipM0012354908F/20cm
M0012355008F/22cm
M0012355108F/24cm
M0012355208F/26cm
M0012355308F/28cm
M00123557010F/20cm
M00123558010F/22cm
M00123559010F/24cm
M00123560010F/26cm
M00123561010F/28cm
Percuflex™ UreteralStent SystemPercuflex™ Glidex™ HYDROGELCOATING Ureteral Stent SystemM0012454908F/20cm
M0012455008F/22cm
M0012455108F/24cm
M0012455208F/26cm
M0012455308F/28cm
M00124557010F/20cm
M00124558010F/22cm
M00124559010F/24cm
M00124560010F/26cm
M00124561010F/28cm
Percuflex™ UreteralStent System KitPercuflex™ Glidex™ HYDROGELCOATING Ureteral Stent System KitM0012456208F/20cm
M0012456308F/22cm
M0012456408F/24cm
M0012456508F/26cm
M0012456608F/28cm
Product Trade NameProduct Labeled FeaturesModel NumberDimensions(OD/Length)
M0012457206F/20cm
M0012457306F/22cm
M0012457406F/24cm
M0012457506F/26cm
M0012457606F/28cm

Table 5: Percuflex™ Ureteral Stent System Device Features

{7}------------------------------------------------

Table 6: Percuflex™ Nephroureteral Stent System Device Features

Product Trade NameProduct Labeled FeaturesModel NumberDimensions(OD/Length)
Percuflex™Nephroureteral StentSystemPercuflex™ Locking PigtailNephroureteral Stent SystemM0012213608F/22cm
M0012213708F/24cm
M0012213808F/26cm
M0012213908F/28cm
Percuflex™ TempTip™ Locking PigtailNephroureteral Stent System withDissolving TipM00122140010F/22cm
M00122141010F/24cm
M00122142010F/26cm
M00122143010F/28cm
M0012313608F/22cm
M0012313708F/24cm
M0012313808F/26cm
M0012313908F/28cm
M00123140010F/22cm
M00123141010F/24cm
M00123142010F/26cm
M00123143010F/28cm

Table 7: Amplatz Anchor™ Catheter System Device Features

Product Trade NameProduct Labeled FeaturesModel NumberDimensions(OD/Length)
Amplatz Anchor™ Catheter SystemAmplatz Anchor™ Locking Malecot,Amplatz Anchor Catheter SystemM0012210008F/30cm
M00122101010F/30cm
M00122102012F/30cm

{8}------------------------------------------------

V. INDICATIONS FOR USE

Predicate and subject device Intended use and Indications for Use are the same.

PercuflexTM Drainage Catheters andStentsIndications for Use
PercuflexTM Ureteral Stent System andPercuflexTM Ureteral Stent System KitThe Ureteral Stent is intended to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
PercuflexTM Nephroureteral StentSystemThe Percuflex Nephroureteral Stent is intended for use in Percutaneous Drainage to establish internal drainage from the ureteropelvic junction to the bladder while maintaining external access to the stent.
Amplatz AnchorTM Catheter SystemThe Amplatz AnchorTM Catheter System is intended for use in applications of percutaneous drainage, particularly percutaneous nephrostomy, abscess or external biliary drainage.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The Percuflex™ Drainage Catheters and Stents incorporate substantially equivalent design, function, packaging, sterilization process, materials, fundamental technology, indications for use, and operating principles as those featured in the predicate drainage devices (K924608). The difference from the legally marketed predicate device is the addition of Magnetic Resonance (MR) compatibility information to the Directions for Use.

Cumulative changes have been made to the devices internally through Boston Scientific's Design Control System, which meets the quality system requirements of 21 CFR Part 820. These changes include a resin additive material supplier change and tensile performance specification update.

The tensile performance specification update applies to the Percuflex™ TempTip™ Ureteral Stent System with Dissolving Tip and the Percuflex™ TempTip™ Locking Pigtail Nephroureteral Stent System with Dissolving Tip devices. This change updated the dissolving tip to catheter tensile strength performance specification.

VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

In order to support the addition of Magnetic Resonance Compatibility information to the Directions for Use the following non clinical performance testing was performed:

  • Radio Frequency Heating .
  • Force Measurement ●
  • Image Artifact .

FDA guidance Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment (December 11, 2014) and the following standards were used to guide the generation of the non-clinical data:

{9}------------------------------------------------

Document NameDocument NumberDocument Version
Measurement of Radio Frequency Induced HeatingNear Passive Implants During Magnetic ResonanceImagingASTM F2182-11a
Measurement of Magnetically Induced DisplacementForce on Medical Devices in the Magnetic ResonanceEnvironmentASTM F2052- 14
Evaluation of MR Image Artifacts from Passive ImplantsASTM F2119- 07

In order to support the cumulative changes that have been made to the devices internally through Boston Scientific's Design Control System, the following non clinical performance testing was performed:

  • · Resin additive material supplier change
    • o A material evaluation consisted of Fourier Transform Infrared Spectroscopy (FTIR), Differential Scanning Calorimetry (DSC), Thermogravimetric Analysis (TGA), Oxidative Induction Time (OIT), and Dynamic Contact Angle (DCA) analysis was performed. Results demonstrated that both supplied resins present the same thermal transition characteristics, material type (chemistry), and composition.
    • o Design Verification testing including Catheter Shaft Tensile, Column Strength, Kink Resistance, Removal Force, and Renal Coil Resistance-Stent Shaft >5F was performed to assess these finished device properties.
    • · The following biocompatibility testing was performed per ISO 10993-1:
      • Cytotoxicity

      • Sensitization

      • Intracutaneous irritation testing

      • Material mediated pyrogenicity

      • Acute systemic toxicity testing

      • Genotoxicity testing Ames assay

      • Genotoxicity testing MLA

      • Muscle implantation 4-week study

      • Muscle implantation 13-week study A
      • Toxicological Risk Assessment

  • · Tensile performance specification update
    • o Design verification testing was conducted on aged and unaged samples to demonstrate the dissolving tip to shaft bond design output continues to meet the defined design input requirements after sterilization.

· CONCLUSION

Based on the intended use, technological characteristics, and non-clinical performance data provided, the Percuflex™ Drainage Catheters and Stents are substantially equivalent to the predicate devices (K924608). The Directions for Use update with magnetic resonance compatibility information for the subject devices does not raise new questions of safety or effectiveness, and the subject devices are as safe and effective as the predicate devices.

§ 876.4620 Ureteral stent.

(a)
Identification. A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.(b)
Classification. Class II (performance standards).