K Number
K200230

Validate with FDA (Live)

Date Cleared
2021-08-26

(574 days)

Product Code
Regulation Number
866.5660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semiquantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.

The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.

Device Description

The Aptiva Celiac Disease IgG reagent utilizes particle based multi-analyte technology (PMAT) in a cartridge format. Each analyte (tissue transglutaminase [tTG] and deamidated gliadin peptide [DGP]) in the Aptiva Celiac Disease IgG reagent is a solid phase immunoassay utilizing fluorescent microparticles. This technology allows each of the two analytes, along with a human IgG capture antibody (IgG Control Microparticle), to be coated onto three uniquely recognizable paramagnetic microparticles, which are combined into one tube.

The Aptiva instrument is a fully automated, random access analyzer. This platform is a closed system with continuous load and random-access capabilities that processes the samples, runs the reagent and reports results. It includes liquid handling hardware, optical module (OM), and integrated computer with proprietary software and touch screen user interface.

The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge under conditions that preserve the proteins in their reactive states. When the assay cartridge is ready to be used for the first time, the reagent tube seals are pierced using the cartridge lid. The reagent cartridge is then loaded onto the Aptiva instrument, where the microparticles are automatically rehydrated using buffer located within the cartridge.

A patient's serum is diluted 1:23 with Aptiva system rinse by the instrument in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and the microparticle suspension in a second cuvette, and mixed (final serum dilution: 1:230). This reaction cuvette is incubated for 9 ½ minutes at 37°C. The cuvette is then exposed to a small magnet that holds the microparticles in place. The liquid is aspirated, and the microparticles are resuspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step. After the third wash, phycoerythrin conjugated polyclonal anti-human lgG (known as PE Tracer IgG) is added to the microparticles in the cuvette, and mixed. Again, the cuvette is incubated for 9 ½ minutes at 37℃. Three wash steps, as described above, are performed on the microparticles. Following the wash steps, the microparticles are transferred to the of the instrument, where a charge coupled device (CCD) camera takes multiple images in order to identify and count the three unique microparticle regions, as well as determine the amount of conjugate on the microparticles. The control microparticle, a third particle, coated with goat anti-human IgG, is included in the reagent in as a control to flag low concentrations of IgG the patient serum sample as an assay verification step. The median fluorescent intensity (MFI) is proportional to the amount of PE Tracer that is bound to the human IgG, which is proportional to the amount of IgG antibodies bound to the corresponding microparticle regions.

For quantitation, the DGP IgG and tTG IgG assays (together as part of the Aptiva Celiac Disease IgG Reagent) each utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge RFID tag. Every new lot of reagent cartridge must be calibrated before first use with the reagent specific calibrators. Based on the results obtained with the calibrators included in the Aptiva Celiac Disease IgG Calibrator kit (sold separately), an instrument specific Working Curve is created for each assay, which is used to calculate reported fluorescent light units (FLU) from the median fluorescent intensity (MFI) instrument signal obtained for each sample, on each of the two assays within the reagent.

Aptiva Celiac Disease IgG Calibrators and Aptiva Celiac Disease IgG Controls are sold separately.

The Aptiva Celiac Disease IgG Reagent kit contains the following materials:

One (1) Aptiva Celiac Disease IgG Reagent Cartridge, containing the following reagents for 200 determinations:

  • a. Aptiva Celiac Disease IgG microparticle containing 3 unique microparticle regions coated with recombinant tissue transglutaminase, deamidated gliadin peptide, or goat antihuman IgG antibody.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • C. PE Tracer IgG - phycoerythrin (PE) labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservative.
  • ð. Rehydration Buffer - containing protein stabilizers and preservatives.
AI/ML Overview

This document describes the analytical and clinical performance of the Aptiva Celiac Disease IgG Reagent, an immunoassay for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies (tTG IgG) and anti-deamidated gliadin peptide IgG autoantibodies (DGP IgG) in human serum. This device is intended as an aid in the diagnosis of celiac disease and dermatitis herpetiformis.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Acceptance Criteria and Reported Device Performance

The document presents several analytical performance characteristics and their corresponding acceptance criteria, along with the reported performance values. The primary clinical acceptance criteria are related to sensitivity and specificity, and the agreement with a predicate device.

Test CategoryAcceptance CriteriaReported Device Performance (DGP IgG)Reported Device Performance (tTG IgG)
PrecisionTotal %CV: < 12% or SD < 0.6 FLUAll samples met the criteria. For example, sample 1: 8.9% CV (SD 0.15 FLU); sample 8: 7.9% CV (SD 17.19 FLU)All samples met the criteria. For example, sample 1: 7.7% CV (SD 0.14 FLU); sample 8: 8.3% CV (SD 17.38 FLU)
Reproducibility (Between-Site)Reproducibility Between-Site %CV: < 12% or SD < 0.6 FLUAll samples met the criteria. For example, sample 1: 12.8% CV (SD 0.29 FLU); sample 7: 9.5% CV (SD 15.28 FLU)All samples met the criteria. For example, sample 1: 9.1% CV (SD 0.21 FLU); sample 7: 8.5% CV (SD 16.44 FLU)
Reproducibility (Between-Lots)Reproducibility Between-Lot %CV: < 12% or SD < 0.6 FLUAll samples met the criteria. For example, sample 1: 11.2% CV (SD 0.33 FLU); sample 6: 10.6% CV (SD 13.06 FLU)All samples met the criteria. For example, sample 1: 7.1% CV (SD 0.13 FLU); sample 7: 8.2% CV (SD 14.00 FLU)
Limit of Quantitation (LoQ)Total imprecision < 20%LoQ: 0.56 FLU (final value)LoQ: 0.82 FLU (final value)
LinearityBest fitting polynomial is linear OR difference between best-fitting non-linear and linear polynomial is <15% or ±0.75 FLU for low-level samples (allowable non-linearity).All acceptance criteria were fulfilled across the range 0.52 - 274.25 FLU.All acceptance criteria were fulfilled across the range 0.99 - 327.80 FLU.
Interference85-115% recovery, or ±20% of cut-off (±1.0 FLU) difference, whichever is greater.Less than 15% interference for bilirubin, hemoglobin, triglycerides, cholesterol, RF IgM, and human IgG. Recoveries detailed in text (e.g., bilirubin 96.0-101.3%).Less than 15% interference for bilirubin, hemoglobin, triglycerides, cholesterol, RF IgM, and human IgG. Recoveries detailed in text (e.g., bilirubin 97.7-102.5%).
Sample Stability85-115% recovery for positive samples; 80-120% for negative samples (<5.00 FLU).All samples fulfilled the acceptance criteria for storage up to 48 hours at room temperature, up to 14 days at 2-8°C, and up to 5 freeze/thaw cycles.All samples fulfilled the acceptance criteria for storage up to 48 hours at room temperature, up to 14 days at 2-8°C, and up to 5 freeze/thaw cycles.
Reagent Shelf LifeLower and upper 95% CI of regression line between 80% and 120% recovery at day 28 (week 4) of accelerated stability for 2-year preliminary dating.All components tested fulfilled the acceptance criteria, assigning a two-year expiration dating. Real-time stability data up to 25 months show 88.0-108.0% recovery (Lot 100015) and 88.6-91.9% recovery (Lot 100017).All components tested fulfilled the acceptance criteria, assigning a two-year expiration dating. Real-time stability data up to 25 months show 100.5-107.8% recovery (Lot 100015) and 97.5-98.1% recovery (Lot 100017).
In-use StabilityStability claim established at actual measurement day where 95% CI of regression line reaches 85% or 115% recovery, OR ≥2% of recovery data (<3 data points) is <75% or ≥125% recovery.Onboard stability set at 28 days for the reagent cartridge.Onboard stability set at 28 days for the reagent cartridge.
Clinical Performance (Sensitivity)Not explicitly stated but implied through comparison to established predicate performance and the need to aid diagnosis.DGP IgG: 82.2% (157/191) [95% CI: 76.2 – 87.0%] for CD (includes IgA deficient CD patients). 70.6% (24/34) [95% CI: 53.8 – 83.2%] for DH.tTG IgG: 60.7% (116/191) [95% CI: 53.7 – 67.4%] for CD (includes IgA deficient CD patients). 26.5% (9/34) [95% CI: 14.6 – 43.1%] for DH.
Clinical Performance (Specificity)Not explicitly stated but implied through comparison to established predicate performance and the need to aid diagnosis.DGP IgG: 97.9% (284/290) [95% CI: 95.6 – 99.0%] for non-CD.tTG IgG: 100.0% (290/290) [95% CI: 98.7– 100.0%] for non-CD.
Method Comparison (Positive Percent Agreement - PPA)Not explicitly stated beyond "comparison with predicate device".DGP IgG: 97.2% (141/145) [95% CI: 93.1–98.9%] with QUANTA Flash DGP IgG.tTG IgG: 91.9% (91/99) [95% CI: 84.9–95.8%] with QUANTA Flash tTG IgG.
Method Comparison (Negative Percent Agreement - NPA)Not explicitly stated beyond "comparison with predicate device".DGP IgG: 65.8% (48/73) [95% CI: 54.3–75.6%] with QUANTA Flash DGP IgG.tTG IgG: 83.7% (139/166) [95% CI: 77.4–88.6%] with QUANTA Flash tTG IgG.

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Validation Set): A total of 515 characterized samples.
    • 171 samples from celiac disease patients.
    • 20 samples from patients with IgA deficient celiac disease.
    • 34 dermatitis herpetiformis patients.
    • 290 control samples from patients with various types of autoimmune and infectious diseases.
  • Data Provenance: The document does not explicitly state the country of origin. Given it's an FDA submission, it's typically a mix of US and possibly international data, but this is not specified. The studies are retrospective as they use "characterized samples" from a "cohort."
  • Precision and Reproducibility Studies: Between 75 to 80 replicates per sample for repeatability/precision and reproducibility studies. These numbers are for analytical performance, not clinical.
  • LoB, LoD, LoQ: 120 data points were generated for each assay on each reagent lot for LoB and LoD studies. For LoQ, 120 data points per assay per reagent lot.
  • Linearity: The number of dilutions and duplicates used is stated (e.g., 4 human serum samples for DGP IgG and 3 for tTG IgG serially diluted, assayed in duplicates).
  • Interference: 3 human serum specimens (one positive, one near cut-off, one negative) were tested for each assay.
  • Sample Stability: 6 samples for DGP IgG, 7 for tTG IgG (tested in duplicates).
  • Reagent Stability: 3 lots of microparticle beads and 3 lots of PE Tracer IgG for accelerated stability. Real-time stability data from 2 different lots.
  • Reference Range/Cut-off Establishment:
    • Reference Population: 192 subjects from various autoimmune/infectious disease groups (e.g., Crohn's Disease, Autoimmune Thyroid Disease, Rheumatoid Arthritis).
    • Celiac Disease Patients: 11 diagnosed celiac disease (CD) patient specimens were assayed to aid in cut-off determination.
    • Apparently Healthy Donors (Expected Values): 120 blood donors.

3. Number of Experts and Qualifications

  • The document does not mention the number of experts used to establish the ground truth for the test set, nor their specific qualifications. It refers to "characterized samples" and "diagnosed celiac disease (CD) patient specimens," implying a pre-existing clinical diagnosis, but the process of how these characterizations were definitively made (e.g., biopsy confirmation, clinical consensus, expert review) is not detailed for the test set.

4. Adjudication Method

  • The document does not describe any specific adjudication method for the test set. The samples are described as "characterized," suggesting that their disease status was already established prior to their use in the study, likely through standard clinical diagnostic procedures, but no expert review or consensus process for this specific study's set is detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not Applicable. This is an in-vitro diagnostic (IVD) device, specifically an immunoassay for determining autoantibodies in serum. MRMC studies are typically performed for imaging diagnostic devices (e.g., AI for radiology) where human readers (e.g., radiologists) interpret images with and without AI assistance. This document describes the performance of a lab test that outputs a quantitative result (FLU) and a qualitative interpretation (positive/negative) and does not involve human interpretation of complex data in the same way an imaging AI device would.

6. Standalone Performance

  • Yes, standalone performance was done. The entire study report describes the standalone performance of the Aptiva Celiac Disease IgG Reagent without human intervention beyond performing the test and interpreting the quantitative results per the device's defined cut-offs. The sensitivity, specificity, and agreement with predicate devices are measures of its standalone performance.

7. Type of Ground Truth Used

  • The ground truth for the clinical validation was based on clinical diagnosis/characterization of the patient samples.
    • For celiac disease and dermatitis herpetiformis patients, they were "diagnosed" or "characterized." While not explicitly stated, the gold standard for celiac disease diagnosis usually involves intestinal biopsy with characteristic changes, alongside clinical symptoms and serology.
    • For the control group, patients were characterized with "various types of autoimmune and infectious diseases," implying a clinical diagnosis for these conditions to confirm they are not celiac disease.
    • For the cut-off determination, "diagnosed celiac disease (CD) patient specimens" were used in conjunction with a reference population.

8. Sample Size for the Training Set

  • The document does not specify a separate "training set" in the context of an AI/machine learning model. This device is an immunoassay, which relies on chemical reactions and optical detection, not on a machine learning algorithm trained on large datasets in the conventional sense. The "training" in this context refers to the development and optimization of the assay's reagents and parameters, and the establishment of master curves and cut-offs. The data used for calibration and master curve generation (e.g., "in-house Master Curve Standards with assigned FLU values run multiple times," "Calibrators included in the Aptiva Celiac Disease IgG Calibrator kit") effectively serve a similar purpose to training/calibration data in general analytical chemistry.

9. How Ground Truth for Training Set was Established

  • Given this is an immunoassay, the concept of "ground truth" for a training set (as defined for AI/ML) is not directly applicable. Instead, the assay's performance and "ground truth" are established through:
    • Calibration: Using Master Curve Standards with "assigned FLU values" (likely determined through extensive in-house characterization and reference methods).
    • Controls: Using Aptiva Celiac Disease IgG Controls with "lot specific values assigned."
    • Reference Materials: The development of the assay's antigens (recombinant tTG and deamidated gliadin peptide) and antibodies would have involved rigorous characterization against known reference materials and clinical samples during the assay development stages to ensure they correctly bind to their target autoantibodies.
    • Cut-off determination: As mentioned in point 7, this involved a reference population of 192 subjects and 11 diagnosed celiac disease patients to establish the 5.00 FLU cut-off.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

August 26, 2021

Inova Diagnostics, Inc. Andrea Seaman Manager, Research and Development 9900 Old Grove Road San Diego, California 92131

Re: K200230

Trade/Device Name: Aptiva Celiac Disease IgG Reagent Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test System Regulatory Class: Class II Product Code: MVM, MST Dated: January 27, 2020 Received: January 30, 2020

Dear Andrea Seaman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Ying Mao Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200230

Device Name Aptiva Celiac Disease IgG Reagents

Indications for Use (Describe)

The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semiquantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.

The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized checkmark in red and blue, followed by the text "Inova Diagnostics" in a sans-serif font. Below the company name is the text "A Werfen Company" in a smaller font size.

510(k) Summary

Aptiva Celiac Disease IgG Reagent

Table of Contents
Administrative data……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Predicate device……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Device description
Intended use(s)
Indications for use
Substantial equivalence
Comparison to predicate device
Analytical performance characteristics
Quantitation and units of measure
Precision
Reproducibility Studies
Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)
Analytical Measuring Range (AMR)
Auto-rerun function and reportable results
High concentration hook effect
Linearity
Interference
Sample Stability and Handling
Reagent Stability
Cut-off, reference range
Clinical performance characteristics
Clinical sensitivity, specificity
Expected values
Comparison with predicate device

{4}------------------------------------------------

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510K number is: K200230

Administrative data

Submitter:Inova Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device
Device in the submission:Aptiva Celiac Disease IgG Reagent
Revision Date:August 23, 2021
Scientific contact:Andrea Seaman, Manager, Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x1395Fax: 858-863-0025Email: aseaman@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RAInova Diagnostics, Inc9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x1381Fax: 858-863-0025Email: relliott@inovadx.com
Device name (kit):Proprietary name:Aptiva Celiac Disease IgG Reagent
Common name:anti-deamidated gliadin peptide antibodyimmunoassay, anti-tissue transglutaminaseantibody immunoassay
Classification name:DGP IgG: Radioallergosorbent (RAST)immunological test systemtTG IgG: Multiple autoantibodies immunologicaltest system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeDGP IgG: MST

{5}------------------------------------------------

tTG IgG: MVM
Regulation Number866.5750, 866.5660
Device Class2

Predicate device

QUANTA Flash® DGP IgG, 510(k) number: K113863. Date declared: October 23, 2012. QUANTA Flash® h-tTG IgG, 510(k) number: K101644. Date declared: March 23, 2011.

Device description

The Aptiva Celiac Disease IgG reagent utilizes particle based multi-analyte technology (PMAT) in a cartridge format. Each analyte (tissue transglutaminase [tTG] and deamidated gliadin peptide [DGP]) in the Aptiva Celiac Disease IgG reagent is a solid phase immunoassay utilizing fluorescent microparticles. This technology allows each of the two analytes, along with a human IgG capture antibody (IgG Control Microparticle), to be coated onto three uniquely recognizable paramagnetic microparticles, which are combined into one tube.

The Aptiva instrument is a fully automated, random access analyzer. This platform is a closed system with continuous load and random-access capabilities that processes the samples, runs the reagent and reports results. It includes liquid handling hardware, optical module (OM), and integrated computer with proprietary software and touch screen user interface.

The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge under conditions that preserve the proteins in their reactive states. When the assay cartridge is ready to be used for the first time, the reagent tube seals are pierced using the cartridge lid. The reagent cartridge is then loaded onto the Aptiva instrument, where the microparticles are automatically rehydrated using buffer located within the cartridge.

A patient's serum is diluted 1:23 with Aptiva system rinse by the instrument in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and the microparticle suspension in a second cuvette, and mixed (final serum dilution: 1:230). This reaction cuvette is incubated for 9 ½ minutes at 37°C. The cuvette is then exposed to a small magnet that holds the microparticles in place. The liquid is aspirated, and the microparticles are resuspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step. After the third wash, phycoerythrin conjugated polyclonal anti-human lgG (known as PE Tracer IgG) is added to the microparticles in the cuvette, and mixed. Again, the cuvette is incubated for 9 ½ minutes at 37℃. Three wash steps, as described above, are performed on the microparticles. Following the wash steps, the microparticles are transferred to the of the instrument, where a charge coupled device (CCD) camera takes multiple images in order to identify and count the three unique microparticle regions, as well as determine the amount of conjugate on the microparticles. The control microparticle, a third particle, coated with goat anti-human IgG, is included in the reagent in as a control to flag low concentrations of IgG the patient serum sample as an assay verification step. The median fluorescent intensity (MFI) is proportional to the amount of PE Tracer that is bound to the human IgG, which is proportional to the amount of IgG antibodies bound to the corresponding microparticle regions.

{6}------------------------------------------------

For quantitation, the DGP IgG and tTG IgG assays (together as part of the Aptiva Celiac Disease IgG Reagent) each utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge RFID tag. Every new lot of reagent cartridge must be calibrated before first use with the reagent specific calibrators. Based on the results obtained with the calibrators included in the Aptiva Celiac Disease IgG Calibrator kit (sold separately), an instrument specific Working Curve is created for each assay, which is used to calculate reported fluorescent light units (FLU) from the median fluorescent intensity (MFI) instrument signal obtained for each sample, on each of the two assays within the reagent.

Aptiva Celiac Disease IgG Calibrators and Aptiva Celiac Disease IgG Controls are sold separately.

The Aptiva Celiac Disease IgG Reagent kit contains the following materials:

One (1) Aptiva Celiac Disease IgG Reagent Cartridge, containing the following reagents for 200 determinations:

  • a. Aptiva Celiac Disease IgG microparticle containing 3 unique microparticle regions coated with recombinant tissue transglutaminase, deamidated gliadin peptide, or goat antihuman IgG antibody.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • C. PE Tracer IgG - phycoerythrin (PE) labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservative.
  • ರ. Rehydration Buffer - containing protein stabilizers and preservatives.

Intended use(s)

The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase lgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.

The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.

Indications for use

Same as intended use.

Substantial equivalence

The Aptiva Celiac Disease IgG Reagent has the same intended use and assay principle as the predicate devices.

{7}------------------------------------------------

Page 5 of 19

Comparison to predicate device

Aptiva Celiac Disease IgG Reagent – DGP IgG Assay

Similarities
ItemAptiva Celiac Disease IgG Reagent (DGP IgG)QUANTA Flash DGP IgG
Intended UseThe Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.The QUANTA Flash DGP IgG is a chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgG anti-deamidated gliadin peptide (DGP) antibodies in human serum. The presence of IgG anti-DGP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of celiac disease in IgA sufficient and IgA deficient patients, as well as dermatitis herpetiformis.
Assay Methodologysolid phase (heterogeneous) immunoassaysolid phase (heterogeneous) immunoassay
Antigendeamidated gliadin peptidedeamidated gliadin peptide
Sample Typehuman serumhuman serum
Solid Phaseparamagnetic microparticlesparamagnetic microparticles
Differences
ItemAptiva Celiac Disease IgG Reagent (DGP IgG)QUANTA Flash DGP IgG
Detection/Operating Principlefluorescent immunoassaychemiluminescent immunoassay
Conjugatephycoerythrin conjugated polyclonal anti-human IgG antibodyIsoluminol conjugated monoclonal anti-human IgG antibody
Unitsfluorescent light units (FLU)chemiluminescent units (CU)
Cut-off5.00 FLU20.0 CU
Analytical MeasuringRange0.56 FLU – 250.00 FLU2.8 CU – 1936.7 CU
ControlControls have lot specific valuesassigned.Control 1DGP IgG – 6.72 FLUControl 2DGP IgG – 14.12 FLU (8.47 – 19.77)Controls have lot specific valuesassigned.Negative ControlDGP IgG – 10.9 CU (6.5 – 15.3)Positive ControlDGP IgG – 52.4 CU (31.4 – 73.4)
CalibrationLot specific Master Curve + 3calibrators (sold separately)Lot specific Master Curve + 2calibrators (sold separately)

{8}------------------------------------------------

Aptiva Celiac Disease IgG Reagent – tTG IgG Assay

Similarities
ItemAptiva Celiac Disease IgG Reagent (tTG IgG)QUANTA Flash tTG IgG
Intended UseThe Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.The QUANTA Flash h-tTG IgG is a chemiluminescent immunoassay (CIA) for the semi-quantitative detection of IgG anti-human tissue transglutaminase (h-tTG) antibodies in human serum. The presence of IgG anti-h-tTG antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of the gluten sensitive enteropathy celiac disease, particularly in patients with selective IgA deficiency.
Assay Methodologysolid phase (heterogeneous) immunoassaysolid phase (heterogeneous) immunoassay
Antigenrecombinant tissue transglutaminaserecombinant tissue transglutaminase
Sample Typehuman serumhuman serum
Solid Phaseparamagnetic microparticlesparamagnetic microparticles

Page 6 of 19

{9}------------------------------------------------

ItemAptiva Celiac Disease IgG Reagent(tTG IgG)QUANTA Flash tTG IgG
Detection/OperatingPrinciplefluorescent immunoassaychemiluminescent immunoassay
ConjugatePhycoerythrin conjugatedpolyclonal anti-human IgGantibodyIsoluminol conjugatedmonoclonal anti-human IgGantibody
Unitsfluorescent light units (FLU)chemiluminescent units (CU)
Cut-off5.00 FLU20.0 CU
Analytical MeasuringRange0.82 FLU - 250.00 FLU3.75 CU - 2560.0 CU
ControlControls have lot specific valuesassigned.Control 1tTG IgA – 8.18 FLUControl 2tTG IgA - 16.39 FLU (9.83 – 22.95)Controls have lot specific valuesassigned.Negative Controlh-tTG IgA - 11.4 CU (6.8 – 16.0)Positive Controlh-tTG IgA - 50.4 CU (30.2 – 70.6)
CalibrationLot specific Master Curve + 3calibrators (sold separately)Lot specific Master Curve + 2calibrators (sold separately)

Analytical performance characteristics

Quantitation and units of measure

For quantitation, the Aptiva Celiac Disease lgG reagent utilizes predefined lot specific Master Curves, one for tTG IgG and one for DGP IgG that is uploaded onto the instrument through the reagent cartridge RFID. The analyte specific Master Curves are generated at Inova for each reagent lot, where in-house Master Curve Standards with assigned FLU values are run multiple times. The resulting MFI values generated are used to create a unique 4 parameter logistic (4PL) curve for each of the two analytes. The lgG control bead is present in the reagent as a control to flag low concentration lgG in a patient serum sample as an assay verification step. This microparticle also has an in-house standard which is run each time a new reagent lot is manufactured. The MFI produced by this standard is used as the cut-off threshold for the IgG control microparticle for that reagent lot. These four parameters of the analyte curves, as well as the MFI cut-off for the IgG control microparticle are embedded in the reagent cartridge RFID.

MaterialAssigned Value (FLU)
Celiac IgG Master Curve Standard 10.00
Celiac IgG Master Curve Standard 23.81
Celiac IgG Master Curve Standard 38.97
Celiac IgG Master Curve Standard 447.37
Celiac IgG Master Curve Standard 5250.07

List of Aptiva Celiac Disease IgG Master Curve Standards - DGP IgG:

List of Aptiva Celiac Disease IgG Master Curve Standards – tTG IgG

{10}------------------------------------------------

MaterialAssigned Value (FLU)
Celiac IgG Master Curve Standard 10.00
Celiac IgG Master Curve Standard 22.62
Celiac IgG Master Curve Standard 331.14
Celiac IgG Master Curve Standard 473.23
Celiac IgG Master Curve Standard 5386.85

Precision

The precision of the Aptiva Celiac Disease lgG reagent was evaluated on 8 samples for DGP IgG and tTG lgG, containing various concentrations of antibodies in accordance with CLSI EPO5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.

Data were analyzed with the Analyse-it for Excel method evaluation software, and repeatability (withinrun), between run, between day and within-laboratory precision) were calculated. Results are summarized in the two tables below.

DGP IgG PrecisionRepeatabilityBetween RunBetween DayWithin Laboratory
SampleReplicates(N)Mean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1801.730.16.1%0.074.0%0.095.1%0.158.9%
2804.290.173.9%0.071.6%0.194.5%0.276.2%
3804.470.173.7%0.194.3%0.143.1%0.296.5%
4805.420.234.3%0.305.5%0.224.0%0.448.1%
58016.170.382.3%0.482.9%0.815.0%1.016.3%
68031.270.812.6%1.003.2%0.902.9%1.575.0%
780131.923.252.5%6.324.8%0.940.7%7.165.4%
880216.469.764.5%7.113.3%12.235.6%17.197.9%

Acceptance criteria: Total %CV: < 12% or SD < 0.6 FLU

tTG IgG PrecisionRepeatabilityBetween RunBetween DayWithin Laboratory
SampleReplicates(N)Mean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1801.880.137.1%0.007.1%0.143.0%0.147.7%
2802.430.156.2%0.041.7%0.125.1%0.208.2%
3804.530.235.1%0.051.2%0.183.9%0.306.5%
4805.290.275.1%0.000.0%0.152.8%0.315.8%
58011.490.423.7%0.121.0%0.443.8%0.625.4%
68046.271.403.0%1.042.3%1.022.2%2.024.4%
780117.596.735.7%2.141.8%2.512.1%7.496.4%
880210.5914.777.0%0.000.0%9.174.4%17.388.3%

{11}------------------------------------------------

Reproducibility Studies

Reproducibility between sites (instruments)

Seven samples for DGP IgG and tTG IgG were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision. Acceptance criteria: Reproducibility Between-Site %CV: < 12% or SD < 0.6 FLU Results are summarized in the tables below.

Aptiva DGP IgGRepeatabilityBetween-DayWithin-SiteBetween-Site/InstrumentReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.260.2511.2%0.000.0%0.2511.2%0.146.2%0.2912.8%
2755.410.275.0%0.325.9%0.427.7%0.275.1%0.509.2%
3756.420.355.4%0.385.9%0.518.0%0.243.8%0.578.9%
47512.400.574.6%0.867.0%1.048.4%0.000.0%1.048.4%
57544.171.403.2%2.345.3%2.736.2%2.164.9%3.487.9%
675119.066.155.2%5.384.5%8.176.9%6.725.6%10.588.9%
775160.2210.826.8%4.542.8%11.747.3%9.786.1%15.289.5%
Aptiva tTG IgGRepeatabilityBetween-DayWithin-SiteBetween-Site/InstrumentReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.330.125.0%0.114.5%0.166.7%0.146.2%0.219.1%
2755.330.254.5%0.193.4%0.315.7%0.325.8%0.458.1%
3756.030.254.1%0.183.0%0.305.0%0.345.6%0.457.5%
47512.120.383.1%0.231.9%0.443.7%0.453.7%0.635.2%
57530.431.354.4%0.551.8%1.464.8%1.123.7%1.846.0%
67596.085.946.2%5.195.4%7.888.2%4.314.5%8.989.4%
775192.8713.076.8%4.582.4%13.857.2%8.864.6%16.448.5%

Reproducibility between lots

Six samples for DGP IgG and seven samples for tTG IgG were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline, using three different lots. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per lot, 75 data points total for each sample. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between lot precision.

Acceptance criteria: Reproducibility Between-Lot %CV: < 12% or SD < 0.6 FLU

{12}------------------------------------------------

Aptiva DGP IgGRepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.950.134.4%0.3010.3%0.3311.2%0.000.0%0.3311.2%
2755.790.254.3%0.457.7%0.518.9%0.407.0%0.6511.3%
3758.880.414.6%0.809.0%0.8910.0%0.262.9%0.9310.5%
47524.090.803.3%1.365.7%1.586.6%0.000.0%1.586.6%
57593.992.602.8%5.686.0%6.256.7%5.135.5%8.098.6%
675122.735.364.4%5.234.3%7.496.1%10.708.7%13.0610.6%

Results are summarized in the tables below.

Aptiva tTG IgGRepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1751.870.126.2%0.063.0%0.136.9%0.031.8%0.137.1%
2755.170.214.1%0.152.9%0.265.0%0.387.4%0.468.9%
3755.680.223.8%0.183.2%0.285.0%0.335.9%0.447.7%
47511.010.403.6%0.615.5%0.736.6%0.413.7%0.837.6%
57525.831.124.3%0.993.8%1.495.8%1.425.5%2.058.0%
67573.073.514.8%6.188.5%7.119.7%0.000.0%7.119.7%
775170.4911.276.6%6.874.0%13.207.7%4.662.7%14.008.2%

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)

The LoB, LoD, and LoQ of the DGP IgG and tTG IgG assays in the Aptiva Celiac Disease IgG Reagent were calculated separately by a study according to CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline- Second Edition.

Study protocol for LoB:

Eight blank samples (Aptiva System Rinse) from two different lots were run in replicates of five on two reagent lots, once per day, for 3 days, with 120 data points generated on each lot. The LoB was determined for each assay, on each reagent lot separately with the Analyse-it for Excel software's Reference Interval function, at the 95th percentile, using the non-parametric method for all 4 analyses (DGP IgG on two reagent lots, and tTG IgG on two reagent lots; all having a p-value = <0.0001)

The DGP IgG LoB for one reagent lot was determined as 0.01 FLU (41 MFI), and for the second reagent lot as 0.02 FLU (60 MFI). The final LoB value for DGP IgG is 0.02 FLU.

The tTG IgG LoB for one reagent lot was determined as 0.00 FLU (37 MFI), and for the second reagent lot as 0.02 FLU (34 MFI). The final LoB value for tTG IgG is 0.02 FLU.

{13}------------------------------------------------

Study protocol for LoD:

Four low level samples for each DGP IgG and tTG IgG assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoD was determined separately for each assay, on each reagent lot.

The DGP IgG limit of detection for one reagent lot was determined as 0.15 FLU, and for the second reagent lot as 0.13 FLU. The final LoD value is 0.15 FLU.

The tTG IgG limit of detection for one reagent lot was determined as 0.13 FLU, and for the second reagent lot as 0.12 FLU. The final LoD value is 0.13 FLU.

Study protocol for LoQ:

Four low level samples for each DGP IgG and tTG IgG assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoQ was determined separately for each assay, on each reagent lot. The LoQ was determined in each case by calculating the total imprecision of each sample (acceptance criteria: total imprecision <20%).

The DGP IgG limit of quantitation for one reagent lot was determined as 0.53 FLU, and for the second reagent lot as 0.56 FLU. The final LoQ value is 0.56 FLU, which has been set as the lower limit of the analytical measuring range of the DGP IgG assay.

The tTG IgG limit of quantitation for one reagent lot was determined as 0.80 FLU, and for the second reagent lot as 0.82 FLU. The final LoQ value is 0.82 FLU, which has been set as the lower limit of the analytical measuring range of the tTG IgG assay.

Analytical Measuring Range (AMR)

Within the Aptiva Celiac Disease IgG Reagent:
DGP IgG:0.56 FLU - 250.00 FLU
tTG IgG:0.82 FLU - 250.00 FLU

Auto-rerun function and reportable results

The Aptiva software has an auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result >250.00 FLU for DGP IgG or tTG IgG after performing an additional 10-fold dilution, thereby bringing the measured value within the AMR. The reported result will be calculated by the software factoring the additional dilution. As the highest value that can be measured is 2500.00 FLU for DGP IgG or tTG IgG.

{14}------------------------------------------------

High concentration hook effect

To assess hook effect, 3 samples for DGP lgG and 3 samples for tTG IgG were tested at two increasing 2fold serial dilutions from the standard 1:23 dilution used by the Aptiva Celiac Disease IgG Reagent. All FLU values above the analytical measuring ranges of the two assays are theoretical and were mathematically calculated using the 4 parameters of their respective calibration curves. All samples showed increase in FLU values as dilution factor became more concentrated, thereby confirming that high positive specimens above the AMR do not show hook effect up to 723.10 FLU for the DGP IgG assay and 465.52 FLU for the tTG IgG assay (theoretical values calculated) in the Aptiva Celiac Disease IgG Reagent.

Linearity

The Linearity of the AMR was calculated separately for DGP IgG as part of the Aptiva Celiac Disease IgG Reagent.

The linearity of the AMR of DGP IgG and tTG IgG was evaluated by a study according to CLSI EPO6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. Four human serum samples for DGP IgG and three human serum samples for tTG IgG with various antibody concentrations were serially diluted to obtain values that cover the entire AMR. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.

Acceptance criteria:

  • Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 15% or ±0.75 FLU for low level samples (allowable nonlinearity).
SampleTest Range(FLU)Slope(95% CI)Y-Intercept(95% CI)Average % Recovery
154.85 - 274.251.04(1.00 to 1.09)-8.73(-16.69 to -0.77)0.9998.0%
27.13 - 71.331.04(1.02 to 1.06)-2.29(-3.27 to -1.31)1.0095.1%
34.82 - 48.201.04(1.00 to 1.08)-0.16(-1.26 to 0.94)0.99101.8%
40.52 - 5.191.02(0.99 to 1.05)-0.15(-0.24 to -0.05)1.0095.4%
Combined0.52 - 274.251.00(0.99 to 1.01)-0.50(-1.49 to 0.50)1.0097.6%

For DGP IgG, the best fitting polynomial found for samples 1, 2 and 4 was a linear one, while third order polynomial was found to provide the best fit for sample 3. All acceptance criteria were fulfilled.

{15}------------------------------------------------

SampleTest Range(FLU)Slope(95% CI)Y-Intercept(95% CI)Average % Recovery
132.78 - 327.800.93(0.88 to 0.99)13.20(1.59 to 24.82)0.99103.5%
27.48 - 74.771.00(0.96 to 1.05)1.40(-0.72 to 3.53)0.99104.8%
30.99 - 9.911.03(1.00 to 1.06)-0.22(-0.42 to -0.02)1.0096.3%
Combined0.99 - 327.800.98(0.96 to 1.00)2.32(-0.08 to 4.73)0.99101.6%

For tTG IgG, the best fitting polynomial found for samples 1, 2 and 3 was a linear one. All acceptance criteria were fulfilled.

These data demonstrate the linearity of the analytical measuring range (0.56 FLU – 250.00 FLU) of the DGP lgG assay and the analytical measuring range 0.82 FLU - 250.00 FLU) of the tTG IgG assay, both as part of the Aptiva Celiac Disease IgG Reagent.

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. A set of three human serum specimens, one positive, one near the cutoff and one negative sample were tested using the following interfering substances (bilirubin, hemoglobin, triglycerides, cholesterol, rheumatoid factor IgM and human IgG). All interferents were spiked into every serum specimen and the resulting samples were assessed in triplicates with the Aptiva Celiac Disease IgG assays. Recovery of the unit values was calculated compared to control samples. Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 20% of the cut-off (±1.0 FLU) difference, whichever is greater.

Less than 15% of interference was observed for DGP or tTG IgG with the following interferents: bilirubin up to 1 mg/mL (recovery: from 96.0%-101.3% and from 97.7% to 102.5% for DGP lgG and tTG IgG respectively), hemoglobin up to 2 mg/mL (recovery: from 97.6% to 104.2% and from 96.9% to 100.8% for DGP IgG and tTG IgG respectively), triglycerides up to 1000 mg/dL (recovery: from 93.8% to 102.5% and from 94.5% to 100.8% for DGP IgG and tTG IgG respectively), cholesterol up to 332.5 mg/dL (recovery: from 85.4% to 98.7% and from 89.2% to 100.4% for DGP IgG and tTG IgG respectively), RF IgM up to 250 IU/mL (recovery: 95.4% to 97.6 and 92.1% to 97.3% for DGP IgG and tTG IgG, respectively) and human lgG up to 35 mg/mL (recovery: from 105.0% to 114.1% or -0.80 FLU and from 101.7 to 110.8% or 0.47 FLU for DGP IgG and tTG IgG, respectively).

Sample Stability and Handling

For the DGP IgG assay, six test samples were tested and for tTG IgG assay seven test samples were tested. The samples used for this study were achieved by combining high and low antibody level to yield their desired reactivity. Test samples covered the analytical measuring ranges of each analyte. All samples were tested in duplicates for up to 21 days while stored at 2-8℃, up to 48 hours while stored at room

{16}------------------------------------------------

temperature, and after repeated freeze/thaw cycles up to 6 cycles. Results were compared to those obtained on control samples (time zero / zero cycles).

Acceptance criteria: percent recovery is between 85-115% for positive samples, and between 80-120% for negative samples (<5.00 FLU).

All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples may be stored up to 48 hours at room temperature, up to 14 days at 2-8°C and can be subjected to up to 5 freeze/thaw cycles.

Reagent Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 5 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components to establish initial stability claim:

  • . Aptiva Celiac Disease IgG microparticle (bead) - 3 lots
  • PE Tracer IgG - 3 lots

Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, 4, and 5 weeks, where one week is equal to six months at 5 ± 3°C. Linear regression analysis was performed between recovery values and the number of days. For each component tested, linear regression analysis was performed separately on each bead in the Aptiva Celiac Disease IgG Reagent (DGP IgG, tTG IgG, IgG Control Bead).

Acceptance criteria for two-year preliminary expiration dating: With regression analysis, the lower and upper 95% Cl interval of the regression line is between 80% and 120% recovery at day 28 (week 4).

All components tested fulfilled the acceptance criteria above, therefore, two-year expiration dating was assigned to each component.

In-use (onboard) stability

Reagent Cartridge

To establish the in-use stability of the Aptiva Celiac Disease lgG reagent cartridges, one lot of reagent cartridge was tested using up to 12 human serum samples (with different reactivity levels). The specimens were tested periodically for 31 days. At day 15 the reagent cartridge was recalibrated, and a cartridge specific Working Curve was generated. Percent recoveries were calculated compared to the day zero

{17}------------------------------------------------

average values, and linear regression analysis was performed by plotting percent recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first): - The stability claim is established at the actual measurement day proceeding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when ≥2% of the recovery data, (3 data points) is <75% or ≥125% recovery.

The onboard stability results for the Aptiva Celiac Disease IgG is as follows: Lot 100017: 31 days

Using these criteria, the in-use (onboard) stability of the Aptiva Celiac Disease IgG reagent cartridge was set at 28 days.

Real time stability

Real time stability testing has been scheduled to be performed every three or six months on the Aptiva Celiac Disease IgG Reagents kit, to verify the two-year expiration that was assigned based on accelerated stability studies.

A negative sample, a low positive sample, and a high positive sample will be tested at each time point. - Acceptance criteria: results should fall within their respective ranges.

The data from two different lots of Aptiva Celiac Disease IgG assay is available up to 25 months of real time stability.

For lot 100015, the percent recovery for all samples tested for DGP IgG ranges from 88.0% to 108.0%, and for tTG IgG ranges from 100.5% to 107.8%.

For lot 100017, the percent recovery for all samples tested for DGP lgG ranges from 88.6% to 91.9%, and for tTG IgG ranges from 97.5% to 98.1%.

Cut-off, reference range

The following cut-off is used for both the DGP IgG and tTG IgG assays in the Aptiva Celiac Disease IgG Reagent:

Negative <5.00 FLU Positive >5.00 FLU

The reference population for establishing the reference interval for the DGP JgG assays, within the Aptiva Celiac Disease IgG Reagent, consisted of 192 subjects:

Sample GroupN
Crohn's Disease15
Autoimmune Thyroid Disease29
Infectious Disease30
Primary Biliary Cholangitis30
Rheumatoid Arthritis40

{18}------------------------------------------------

Systemic Lupus Erythematosus13
Systemic Sclerosis20
Ulcerative Colitis15

Additionally, 11 diagnosed celiac disease (CD) patient specimens were assayed to aid in the determination of the cutoff values.

All specimens were the same matrix (human serum) as specified in the Intended Use. All specimens were unaltered. The cut-off values were established in accordance to CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The results was non-normal (Shapiro-Wilk p<0.0001), therefore the non-parametric percentile method was used.

The cut-off was established based on greater than the 99th percentile of the results obtained on the reference subjects, along with the results of 11 samples from patients with celiac disease.

A cutoff of 5.00 FLU (109 MFI and 175 MFI for DGP IgG and tTG IgG, respectively) has been set to ensure optimal differentiation between negatives and positives samples.

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the Aptiva Celiac Disease IgG Reagent. A total of 515 characterized samples were included in this Validation Set, including 171 samples from celiac disease patients, 20 samples from patients with IgA deficient celiac disease, 34 dermatitis herpetiformis patients and 290 control samples from patients with various types of autoimmune and infectious diseases. All samples were run on the Aptiva Celiac Disease IgG Reagent. The distribution of the cohort and the DGP and tTG positivity rate is in the Table below:

{19}------------------------------------------------

Page 17 of 19

Patient GroupNDGP IgGN PositiveDGP IgG% PositivetTG IgGN PositivetTG IgG% Positive
Rheumatoid Arthritis6945.8%00.0%
Ulcerative Colitis3100.0%00.0%
Crohn's Disease3100.0%00.0%
Hepatitis C Virus2800.0%00.0%
Hepatitis B Virus2500.0%00.0%
Syphilis2114.8%00.0%
Sjögren's Syndrome2000.0%00.0%
Systemic Sclerosis1900.0%00.0%
Autoimmune Gastritis1500.0%00.0%
Human Immunodeficiency Virus1300.0%00.0%
Systemic Lupus Erythematosus1200.0%00.0%
Epstein-Barr Virus6116.7%00.0%
Total Controls29062.1%00.0%
Celiac Disease17114283.0%10058.5%
IgA Deficient Celiac Disease201575.0%1680.0%
Dermatitis Herpetiformis342470.6%926.5%
Total515----

Clinical sensitivity and specificity for the Aptiva DGP IgG were analyzed in the table below:

Diagnosis
Clinical Analysis (N=481)CD*Non-CD†TotalAnalysis
Positive1576163Sensitivity: 82.2% (157/191) 95% CI: 76.2 – 87.0%
Aptiva DGP IgGNegative34284318Specificity: 97.9% (284/290) 95% CI: 95.6 – 99.0%
Total191290481
  • The study above includes 20 samples from CD patients with selective IgA deficiency: 15 out of 20 tested positive on DGP IgG for a sensitivity of 75% (53.1-88.1%).

i Non-CD does not include the DH samples

{20}------------------------------------------------

Clinical Analysis (N=481)DiagnosisAnalysis
CD**Non-CDTotal
Aptiva tTG IgGPositive1160116Sensitivity: 60.7% (116/191) 95% CI: 53.7 – 67.4%
Negative75290365Specificity: 100.0% (290/290) 95% CI: 98.7– 100.0%
Total191290481

Clinical sensitivity and specificity for the Aptiva tTG IgG were analyzed in the table below:

**The study above includes 20 samples from CD patients with selective IgA deficiency: 16 out of 20 tested positive on tTG IgG for a sensitivity of 80.0% (58.4-91.9%).

Clinical sensitivity and specificity for the Aptiva DGP IgG and Aptiva tTG IgG in diagnosis of dermatitis herpetiformis (DH) were analyzed and results are shown in the following tables:

Clinical Analysis (N=324)DiagnosisAnalysis
DHNon-DHTotal
Positive24630Sensitivity: 70.6% (24/34) 95% CI: 53.8 – 83.2%
Aptiva DGP IgGNegative10284294Specificity: 97.9% (284/290) 95% CI: 95.6 – 99.0%
Total34290324
Clinical Analysis (N=324)DiagnosisAnalysis
DHNon-DHTotal
Positive909Sensitivity: 26.5% (9/34) 95% CI: 14.6 – 43.1%
Aptiva tTG IgGNegative25290315Specificity: 100.0% (290/290) 95% CI: 98.7- 100.0%
Total34290324

Expected values

The expected value in the normal population is "negative". A panel of 120 apparently healthy blood donors (64 females/56 males, ages 17 to 57 years, with an average age of 32 years) were tested on the Aptiva Celiac Disease IgG Reagent. For DGP IgG, with a cut-off of 5.00 FLU, three samples (2.5%) were positive, with a mean concentration of 0.56 FLU, and values ranging from 0.03 to 8.06 FLU. For tTG IgG, with a cut-off of 5.00 FLU, no samples were positive, with a mean concentration of 1.26 FLU, and values ranging from 0.05 to 3.76 FLU.

{21}------------------------------------------------

Comparison with predicate device

Samples for method comparison analysis included all samples (n=515) from the clinical validation study. These samples were tested on both the Aptiva Celiac Disease IgG Reagent and on their predicate QUANTA Flash DGP IgG and tTG IgG assays.

Method ComparisonQUANTA Flash DGP IgGPercent Agreement
(N=218)NegativePositiveTotal
AptivaDGP IgGNegative48452NPA: 65.8% (48/73) 95% CI: 54.3–75.6%
Positive25*141187PPA: 97.2% (141/145) 95% CI: 93.1–98.9%
Total73145218TPA: 86.7% (48+141/218) 95% CI: 81.5–90.6%

Method comparison of the Aptiva DGP IgG with the predicate device. Samples within AMR.

NPA: Negative Percent Agreement; PPA: Positive Percent Agreement; TPA: Total Percent Agreement

  • 18 out of 25 samples tested as negative by the predicate QUANTA Flash DGP IgG, but positive by the Aptiva DGP IgG were celiac disease samples; seven (7) of the discordant samples (28%) were within+25% of the Aptiva DGP assay cut-off.

Method comparison of the Aptiva tTG IgG with the predicate device. Samples within AMR.

Method Comparison(N=265)QUANTA Flash tTG IgGPercent Agreement
AptivatTG IgGNegativePositiveTotal
Negative1398147NPA: 83.7% (139/166) 95% CI: 77.4–88.6%
Positive27**91118PPA: 91.9% (91/99) 95% CI: 84.9–95.8%
Total16699265TPA: 86.8% (139+91/265) 95% CI: 82.2–90.3%

NPA: Negative Percent Agreement; PPA: Positive Percent Agreement; TPA: Total Percent Agreement

** Amonq 27 samples tested as negative by the predicate Quanta Flash tTG IgG, but positive by the Aptiva tTG IgG Assay, 24 were celiac disease samples (six of 27 were IgA deficient celiac disease samples; and three (3) were dermatitis herpetiformis sample); eleven (11) of the discordant samples (41%) were within +25% of the Aptiva tTg assay cut-off.

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).