K Number
K200107

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-06-05

(140 days)

Product Code
Regulation Number
862.1120
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.
The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, oO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
pH, pCO2, pO2 (blood gases) measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of lifethreatening acid-base disturbances.
Sodium and Potassium measurements from the epoc Blood Analysis System are treatment of diseases involving electrolyte imbalance.
lonized Calcium measurements from the epoc Blood Analysis System are treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Chloride measurements from the epoc Blood Analysis System are used in the treatment of electrolyte and metabolic disorders.
Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
Glucose measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of carbolism disorders, including diabetes mellitus and idiopathic hypodycemia, and of pancreatic islet cell tumors.
Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood),
Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment diseases and in monitoring renal dialysis.
Hematocrit measurements from the epoc Blood Analysis System are used to distinguish normal states of blood volume, such as anemia and erythrocytosis.

Device Description

The epoc® Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc® System is comprised of three (3) major subsystems: epoc® Host, epoc® Reader and epoc® BGEM Test Card.

  • epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card: single-use, device with . port for blood sample introduction which contains the sensor configurations for testing Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2).
  • . epoc® Reader: portable, battery-powered device component that measures electrical signals from the test card sensors during blood testing and transmits this sensor data wirelessly via Bluetooth to the epoc Host.
  • . epoc® Host: mobile computer-based device component for calculating test results from the sensor data sent by the epoc Reader and displaying these results on the graphical user interface. The epoc Host component can be physically connected to the Reader by a cradle component. The epoc Host also incorporates an internal laser barcode scanner for scanning patient and operator IDs. The epoc Host component currently runs on Microsoft® Windows Mobile 6.5 Operating System (OS).
AI/ML Overview

Based on the provided text, the "epoc® Blood Analysis System with NXS Host" is being submitted as a modified device, and the submission primarily focuses on hardware and software updates to the epoc Host component. The document explicitly states that "No performance data was required to evaluate the changes introduced with the alternate epoc Host component" and that there is "no change to labeled performance claims." Therefore, there isn't a comprehensive study proving the device meets new acceptance criteria. Instead, the submission argues for substantial equivalence to a previously cleared predicate device by demonstrating that the modifications do not negatively impact safety and effectiveness.

Here's an analysis based on the provided information, addressing your points where possible:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not detail specific acceptance criteria for new performance claims or provide a table of performance data because the submission states "No performance data was required to evaluate the changes introduced with the alternate epoc Host component." The device is intended to meet the same performance specifications as the predicate device. The change is in the host component (hardware and OS), not the core measurement technology or labeled performance.

The submission is essentially asserting that the "epoc® Blood Analysis System with NXS Host" (Modified Device) is substantially equivalent to the "epoc® Blood Analysis System" (Predicate Device) and thus relies on the predicate device's existing performance data and acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

Not applicable. No new clinical performance or analytical performance study with a test set is described. The submission focuses on verification and validation activities for hardware, software, and usability to support substantial equivalence due to a change in the host component.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. As no new performance study is described, there's no mention of experts establishing ground truth for a test set.

4. Adjudication Method for the Test Set

Not applicable. No new performance study is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This device is an in vitro diagnostic system for quantitative testing of blood parameters, not an imaging AI device that would typically involve a multi-reader multi-case study with human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The device is an analytical instrument. Its performance is inherent to the system (sensors, reader, software). While the software is a key component, the "standalone" concept as typically applied to AI in imaging doesn't directly map. However, the document states: "No performance data was required to evaluate the changes introduced with the alternate epoc Host component." This implies that the core analytical performance (algorithm) is considered unchanged from the predicate device and its previous clearances. Verification and validation activities were done on the new host component, but not necessarily a "standalone" re-evaluation of the core measurement algorithm's performance on a new dataset.

7. The Type of Ground Truth Used

Not applicable for new performance data. The device measures objective chemical and physical properties of blood samples. The ground truth for such devices is typically established through reference methods and calibrated controls, not expert consensus or pathology in the same way as an imaging AI. The submission relies on the established ground truth methodologies for the predicate device.

8. Sample Size for the Training Set

Not applicable. This document describes a modification to an existing IVD device (updating hardware and operating system for the host component). It does not describe the development of a new AI algorithm or machine learning model that would involve a "training set" in the conventional sense. The "epoc Host Application Software has been modified to support the Android-based Operating System" but this is a software porting/adaptation, not a new algorithm being trained on data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a training set for a new AI algorithm.


Summary of the Study and Why it Meets (Implied) Acceptance Criteria:

The study detailed in this 510(k) submission is not a clinical performance study generating new acceptance criteria or performance data for the analytes measured. Instead, it is a Special 510(k) submission designed to demonstrate that hardware and software updates to the epoc Host component of the epoc® Blood Analysis System do not alter the safety or effectiveness of the device and thus maintain substantial equivalence to the previously cleared predicate device.

The "study" or evidence provided to meet acceptance criteria consists of:

  • Verification and Validation Activities: "All software, hardware and usability verification and validation activities were performed in accordance to relevant standards, established plans and protocols and Design Control procedures."
  • Meeting Acceptance Criteria: "Testing verified all acceptance criteria were met." (These are acceptance criteria related to software, hardware function, and usability for the new host component, ensuring it performs its intended role without impacting the core analytical performance).
  • Risk Management: A risk management process compliant with EN ISO 14971:2012 and ISO 14971:2007 was performed, concluding that "the overall residual risk of the epoc System with the epoc NXS host is acceptable."
  • Cybersecurity Information: Cybersecurity design inputs were established, risks assessed, and controls designed within the software.

The core argument for meeting acceptance criteria (by proving substantial equivalence) is that:

  • There is no change to the intended use or indications for use.
  • There is no change to the fundamental scientific technology (the epoc Reader and Test Card, which contain the sensors, remain unchanged).
  • There is no change to labeled performance claims.
  • There is no change to the principle of operation.
  • There is no change to cartridge (test card) calibrator formulation and technology.

Therefore, the "acceptance criteria" here are implicitly that the new epoc NXS Host component, through its verification and validation, functions correctly, safely, and does not introduce new risks, thereby maintaining the established performance and safety profile of the predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 5, 2020

Epocal Inc. Amy Goldberg Sr. Manager, Regulatory Affairs 2060 Walkley Road Ottawa, K1G 3P5 Canada

Re: K200107

Trade/Device Name: epoc® Blood Analysis System Regulation Number: 21 CFR 862.1120 Regulation Name: Blood Gases (PCO2, PO2) and Blood pH Test System Regulatory Class: Class II Product Code: CHL. JGS. CEM. JFP. JPI. CGA. KHP. CGL. CGZ. CDS. JFL Dated: May 1, 2020 Received: May 7, 2020

Dear Amy Goldberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, M.T., Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K200107

Device Name

epoc® Blood Analvsis System

Indications for Use (Describe)

  • The epoc Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.
  • The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, oO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
  • pH, pCO2, pO2 (blood gases) measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of lifethreatening acid-base disturbances.
  • Sodium and Potassium measurements from the epoc Blood Analysis System are treatment of diseases involving electrolyte imbalance.
  • lonized Calcium measurements from the epoc Blood Analysis System are treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
  • Chloride measurements from the epoc Blood Analysis System are used in the treatment of electrolyte and metabolic disorders.
  • Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
  • Glucose measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of carbolism disorders, including diabetes mellitus and idiopathic hypodycemia, and of pancreatic islet cell tumors.
  • Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood),
  • Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
  • Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment diseases and in monitoring renal dialysis.
  • Hematocrit measurements from the epoc Blood Analysis System are used to distinguish normal states of blood volume, such as anemia and erythrocytosis.
Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

Section 5 - 510(k) SUMMARY

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

SUBMITTER INFORMATION 1.0

SubmitterName:Epocal Inc., a Siemens Healthineers Company2060 Walkley RoadOttawa, ON K1G 3P5Canada
CompanyContacts:Amy GoldbergSr. Manager, Regulatory AffairsEmail: amy.goldberg@siemens-healthineers.comPhone: +1 (781) 269-3544Amanda BarriageRegulatory Affairs ManagerEmail: amanda.barriage@siemens-healthineers.comPhone: +1 (613) 688 3982 ext. 2232
DatePrepared:June 4, 2020

2.0 DEVICE INFORMATION

Trade orProprietaryName:epoc® Blood Analysis System
CommonName:Blood Gases (PCO2, PO2) and Blood pH Test System

3.0 REGULATORY INFORMATION

The following table highlights the classification and regulatory information for all analytes that are measured on the epoc® Blood Analysis System, which is unchanged for this Special 510(k) submission.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.

Regulatory Information- Classification
MeasuredAnalytesClassificationNameRegulation SectionProductCodeDeviceClassClassificationPanel
pCO2, pO2, pHBlood gases (pCO2,pO2) and blood pHtest system21 CFR § 862.1120CHLIIClinicalChemistry
Sodium (Na+)Sodium test system21 CFR § 862.1665JGSII
Potassium (K+)Potassium testsystem21 CFR § 862.1600CEMII
Ionized Calcium(Ca2+)Calcium testsystem21 CFR § 862.1145JFPII
Chloride (Cl-)Chloride testsystem21 CFR § 862.1170CGZII
Glucose (Glu)Glucose testsystem21 CFR § 862.1345CGAII
Creatinine (Crea)Creatinine testsystem21 CFR § 862.1225CGLII
Hematocrit (Hct)Hematocritmeasuring device21 CFR § 862.6400JPIII
Lactate (Lac)Lactic acid testsystem21 CFR § 862.1450KHPI
Blood UreaNitrogen (BUN)Urea nitrogen testsystem21 CFR § 864.6150CDSII
Total CarbonDioxide (TCO2)Bicarbonate/carbondioxide test system21 CFR § 862.1160JFLII
Capillary bloodcollection tubeCapillary bloodcollection tube21 CFR § 864.6150GIOIHematology

4.0 DEVICE DESCRIPTION

The epoc® Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc® System is comprised of three (3) major subsystems: epoc® Host, epoc® Reader and epoc® BGEM Test Card.

  • epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card: single-use, device with . port for blood sample introduction which contains the sensor configurations for testing Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2).

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.

  • . epoc® Reader: portable, battery-powered device component that measures electrical signals from the test card sensors during blood testing and transmits this sensor data wirelessly via Bluetooth to the epoc Host.
  • . epoc® Host: mobile computer-based device component for calculating test results from the sensor data sent by the epoc Reader and displaying these results on the graphical user interface. The epoc Host component can be physically connected to the Reader by a cradle component. The epoc Host also incorporates an internal laser barcode scanner for scanning patient and operator IDs. The epoc Host component currently runs on Microsoft® Windows Mobile 6.5 Operating System (OS).

The epoc® Blood Analysis System was previously cleared for prescription use to quantitate Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, K113726 and K171247.

This Special 510(k) submission has been prepared to introduce hardware and software updates to the current epoc® Host component only.

5.0 INTENDED USE

The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or unanticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, pO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.

6.0 INDICATIONS FOR USE

pH, pCO2, p02 (blood gases) measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Sodium and Potassium measurements from the epoc® Blood Analysis System are used in diagnosis and treatment of diseases involving electrolyte imbalance.

lonized Calcium measurements from the epoc® Blood Analysis System is used in diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.

Total Carbon Dioxide measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots arranged in a circular pattern.

Glucose measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell tumors.

Lactate measurements from the epoc® Blood Analysis System are used to evaluate the acidbase status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).

Blood Urea Nitrogen measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.

Creatinine measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.

Hematocrit measurements from the epoc® Blood Analysis System distinguishes normal from abnormal states of blood volume, such as anemia and erythrocytosis.

7.0 SPECIAL CONDITIONS FOR USE STATEMENT(S):

  • For prescription use only. .
  • For in vitro diagnostic use only. .
  • For point-of-care (POC) or clinical laboratory settings. ●

8.0 REASON FOR SUBMISSION

This Special 510(k) submission has been prepared to introduce hardware and software updates to the current epoc Host component only. The current epoc Host hardware component is a dedicated-use Personal Digital Assistant (PDA) computer-based platform that operates on Microsoft® Windows Mobile 6.5 Operating System (OS). The hardware component is changing to a new tablet-based mobile platform running on Android 9.0 OS. The modified device consists of the existing current epoc Reader and current epoc Test Card along with a new epoc Host component. There is no change to the epoc Reader and Test Card. The epoc Host Application Software has been modified to support the Android-based Operating System. The new Host component is referred to as "epoc NXS Host" throughout this submission.

The epoc® Blood Analysis System with current Host component is the Predicate Device (K061597) for this Special 510(k) submission. Both the current epoc Host and new epoc NXS Host mobile computers are intended to have equivalent functionalities and workflow. A Special 510(k) Premarket Notification is the ideal pathway for this submission based on the following:

  • . There is no change to the intended use or indications for use.
  • . There is no change to the fundamental scientific technology.
  • There is no change to labeled performance claims. ●
  • There is no change to principle of operation. ●
  • There is no change to cartridge (test card) calibrator formulation and technology. ●

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots of varying sizes.

9.0 SUBSTANTIAL EQUIVALENCE INFORMATION

Predicate Device Information
ELEMENTPREDICATE DEVICE
Predicate Device Nameepoc® Blood Analysis System
Common NameBlood gases (PCO2, PO2) and blood pH test system
510(k) NumbersK092849, K093297, K113726, K171247
FDA Product Codes andAnalytes per 510(k)ClearanceK092849K093297K113726K171247
CHL: pH, pCO2, pO2JGS: SodiumCEM: PotassiumJFP: Ionized CalciumJPI: HematocritCGA: GlucoseKHP: LactateCGL: CreatinineCGZ: ChlorideCDS: Blood UreaNitrogenJFL: Total CarbonDioxide
Sample Types:heparinized or un-anticoagulated arterial, venous or capillary whole blood
ManufacturerEpocal Inc., a Siemens Healthineers Company

The Substantial Equivalency Table in the following pages summarizes the similarities and differences between the epoc® Blood Analysis System with NXS Host (Modified device) and the epoc® Blood Analysis System (Predicate device).

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots.

Substantial Equivalency
FEATUREepoc® Blood Analysis System with NXS Host(MODIFIED DEVICE)K200107epoc® Blood Analysis System(PREDICATE DEVICE)K092849, K093297, K113726,K171247
Intended UseThe epoc® Blood Analysis System isintended for use by trained medicalprofessionals as an in vitro diagnostic devicefor the quantitative testing of samples ofheparinized or un-anticoagulated arterial,venous or capillary whole blood in thelaboratory or at the point of care.The Blood Gas Electrolyte and Metabolite(BGEM) Test Card panel configurationincludes sensors that quantitate pH pCO2 pO2 Sodium Potassium Ionized Calcium Chloride Total Carbon Dioxide Glucose Lactate Blood Urea Nitrogen Creatinine HematocritSame
FDA ProductCodes andAnalytes per510(k)ClearanceK092849:CHL: pH, pCO2, pO2JGS: SodiumCEM: PotassiumJFP: Ionized CalciumJPI: HematocritCGA: GlucoseK093297:KHP: LactateK113726:CGL: CreatinineCGZ: ChlorideK171247:CDS: Blood Urea NitrogenJFL: Total Carbon DioxideSame
Substantial Equivalency
FEATUREepoc® Blood Analysis System with NXS Host(MODIFIED DEVICE)K200107epoc® Blood Analysis System(PREDICATE DEVICE)K092849, K093297, K113726,K171247
Sample TypeHeparinized or un-anticoagulated arterial,venous or capillary whole bloodSame
Sample VolumeAt least 92µLSame
MeasurementTechnologyAn electrochemical multi-sensor arrayintegrated into a single-use test that isinterpreted by a handheld reader andassociated softwareSame
SensorsContained in Test CardsSame
CalibrationCalibration using in-card calibration fluid.Same
OperatingSystemSoftwareAndroid 9.0 Operating SystemWindows 6.5 Mobile OperatingSystem
Hardwareepoc Host: Dedicated-use tablet-basedmobile platformImage: epoc NXS Host, epoc Test Card, epoc Readerepoc Test Card: sameepoc Reader: sameepoc Host: Dedicated-use PersonalDigital Assistant (PDA) mobilecomputer-based platformImage: epoc Host, epoc Test Card, epoc Readerepoc Test Card: contains an array ofsensors on a sensor module mountedin a credit-card sized fluidic housingwith a sample entry port, and a sealedcalibrator fluid reservoir.epoc Reader: Portable, battery-powered device for reading TestCards
ConnectivityWirelessBluetoothSame
Substantial Equivalency
FEATUREepoc® Blood Analysis System with NXS Host(MODIFIED DEVICE)K200107epoc® Blood Analysis System(PREDICATE DEVICE)K092849, K093297, K113726,K171247
BarcodeScannerepoc Reader: Integrated LED (1D) used toread the Test Card Barcodeepoc Host: Integrated (1D & 2D)Manufacturer: ZebraModule model #: SE4710Sensor resolution: 1280 x 800 pixelsepoc Reader: Sameepoc Host: Integrated (1D & 2D)Manufacturer: ZebraModule model #: SE4500DLSensor resolution: 752 x 480 pixels

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the text is a cluster of orange dots.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots arranged in a circular pattern.

10.0 RISK MANAGEMENT

The Risk Management was performed in compliance with EN ISO 14971:2012 and ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices. A series of risk assessments were performed in order to identify, or mitigate any potential risks associated to the design changes for the modified device. In addition, a Failure Mode, Effects and Criticality Analysis (FMECA) is performed for any identifiable risks.

The conclusion of the risk management process is that the overall residual risk of the epoc System with the epoc NXS host is acceptable.

11.0 VERIFICATION AND VALIDATION SUMMARY

All software, hardware and usability verification and validation activities were performed in accordance to relevant standards, established plans and protocols and Design Control procedures. Testing verified all acceptance criteria were met.

No performance data was required to evaluate the changes introduced with the alternate epoc Host component. Software, hardware and usability validation were adequate to support substantial equivalence.

12.0 CYBERSECURITY INFORMATION

Cybersecurity design inputs for the epoc® Blood Analysis System with NXS Host were established as part of the Quality Management procedures for software validation and risk analysis activities. Risks were identified, assessed and controls were designed within the software, to mitigate potential security threats, system vulnerabilities, viral attacks and safeguard unauthorized access to patient data.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a circular pattern.

13.0 CONCLUSION

Results from verification and validation testing confirmed that the modifications made to epoc® Host hardware, operating system, and system software do not introduce new concerns for safety and effectiveness. The modified device utilizes the same intended use, indications for use, principle of operation, workflow, scientific technology and core algorithm as the predicate device. The epoc Test Card and epoc Reader remains unchanged from the predicate device. Based on these factors, the epoc® Blood Analysis System with NXS Host (Modified device) is substantially equivalent to the previously cleared predicate device, the epoc® Blood Analysis System.

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.