K Number
K200035

Validate with FDA (Live)

Date Cleared
2020-05-04

(117 days)

Product Code
Regulation Number
878.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during surgery, including laparoscopic surgery. The Ultravision™ 5mm Trocar component establishes a path of entry for instruments used in laparoscopic surgery.

Device Description

The Ultravision™ Visual Clearing System removes surgical smoke and particulates from the visual field by means of electrostatic precipitation. The System consists of the Ultravision Generator, the Ionwand Sterile Pack, and the Ultravision 5mm Trocar. The Ionwand is connected to the energy source and is then introduced into the body cavity near the smoke generating electrosurgical device. The Ultravision™ 5mm Trocar is intended for use only with the Ultravision™M Visual Field Clearing System to introduce the Ionwand while providing a pathway for laparoscopic instruments through one 5 mm trocar incision. The trocar may be used with or without the Ionwand component of the system. The system is powered using a rechargeable battery or through mains power. Accessories include the rechargeable battery, battery recharging station, mains converter power supply, mains converter, power supply unit, and patient return adaptor.

AI/ML Overview

The Ultravision Visual Field Clearing System is a surgical smoke precipitator. The study described in the provided text focuses on demonstrating the substantial equivalence of the updated Ultravision system to its predicate device, particularly concerning the addition of a new indication for use in open surgical procedures.

Here's an analysis of the acceptance criteria and study details based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Name of the TestPurposeAcceptance CriteriaResults
Cytotoxicity (ISO 10993-5)To analyze the potential of the test article to induce a cytotoxic effect.Must not cause cell lysis or toxicity greater than a grade of 2 (mild reactivity).Reactivity of grade 1 was observed - Pass.
Intracutaneous Irritation (ISO 10993-10)To analyze the potential of the test article to induce a local irritation response.No evidence of significant intracutaneous irritation.No evidence of intracutaneous irritation - Pass.
Systemic Toxicity (ISO 10993-11)To analyze the potential of the test article to induce a systemic response.No evidence of significant systemic toxicity or mortality after test article extracts injection.No evidence of systemic toxicity - Pass.
Maximization Sensitization (ISO 10993-10)To analyze the potential of the test article to induce a sensitization or allergic response.No evidence of induced delayed sensitization.Not a sensitizer - Pass.
Performance Test 1: Simulated UseComparison of smoke clearing characteristics of Ultravision versus other smoke clearing devices for open procedures.Must be considered at least equivalent to comparators.Results of the evaluation demonstrated that Ultravision was at least equivalent to the comparator devices - Pass. (Specifically, in terms of surgical workflow, ease of use, and risks, the Ultravision system was considered equivalent to the reference device.)
Performance Test 2: Risk Assessment (Ozone, Tissue Damage)Evaluation of device use in an oxygen-rich (open) environment considering ozone generation and tissue damage, including empirical testing on ozone generation.The risks associated with the use of Ultravision in an open procedure must not be greater than those for a laparoscopic procedure. Ozone generation must comply with 21 CFR [specific part not specified but implied standard].There were no new risks associated with the use of Ultravision in an open procedure; risks were found to be lower. The time-weighted average for ozone production was below acceptable limits - Pass.

2. Sample Size Used for the Test Set and Data Provenance

  • Biocompatibility Tests (Cytotoxicity, Intracutaneous Irritation, Systemic Toxicity, Maximization Sensitization): The sample sizes for these tests are not explicitly stated in the provided text. These are typically standard tests conducted according to ISO 10993 guidelines, which would specify the number of samples/animals required. The provenance is likely laboratory-based testing.
  • Performance Test 1 (Simulated Use): The text states a "human factors evaluation was also conducted during simulated surgery by actual users (surgeons)." The number of surgeons or simulated cases is not specified. The data provenance is from simulated surgical conditions.
  • Performance Test 2 (Risk Assessment/Ozone Generation): This involved "empirical testing on ozone generation" in a "worst-case model." The specific sample size for these empirical tests (e.g., number of test runs, duration of tests) is not provided. The data provenance is laboratory testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Biocompatibility Tests: Ground truth for these tests is established by standardized laboratory procedures and scientific assessment against predefined biological endpoints. No "experts" in the sense of clinical specialists establishing ground truth are mentioned.
  • Performance Test 1 (Simulated Use): "Actual users (surgeons)" provided feedback to compare the Ultravision device to reference devices. The number of surgeons is not specified, nor are their specific qualifications (e.g., years of experience, subspecialty), beyond being "surgeons." Their feedback likely served as the "ground truth" for perceived performance metrics like workflow, ease of use, and risks in the simulated environment.
  • Performance Test 2 (Risk Assessment): The ground truth for ozone generation is established by analytical measurements against regulatory limits (implied by "21 CFR"). The risk assessment for tissue damage risks was conducted internally by the manufacturer, comparing risks across different environments. The specific experts involved in designing or executing this risk assessment are not detailed, but it would involve experts in medical device safety, engineering, and possibly toxicology.

4. Adjudication Method for the Test Set

  • Biocompatibility Tests: Adjudication is based on objective measurements and adherence to ISO 10993 standards and criteria.
  • Performance Test 1 (Simulated Use): Feedback from "actual users (surgeons)" was synthesized to conclude equivalence. The method of synthesizing this feedback (e.g., individual reports, consensus meeting, voting) is not specified. It's unclear if a formal adjudication process beyond simple aggregation of feedback was used.
  • Performance Test 2 (Risk Assessment): The risk assessment involves internal evaluation against predefined criteria and regulatory standards (for ozone). This is typically an internal process, so a specific external adjudication method is unlikely to have been used in the sense of a clinical read.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done. This device is a surgical smoke precipitator, not an AI diagnostic tool. The "Performance Test 1" involved human factors evaluation by surgeons, but it was not an MRMC study and did not involve AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is a physical medical device (surgical smoke precipitator) and not an algorithm or AI-powered tool. Therefore, a standalone algorithm performance study is not applicable. The device's performance is intrinsically linked to its physical operation.

7. The Type of Ground Truth Used

  • Biocompatibility Tests: Ground truth is established by standardized laboratory test results adhering to ISO 10993, which provide objective measurements of biological response (e.g., cell viability, irritation scores, systemic effects, sensitization).
  • Performance Test 1 (Simulated Use): Ground truth is based on subjective feedback and comparative assessment from "actual users (surgeons)" regarding workflow, ease of use, and perceived risks in a simulated environment.
  • Performance Test 2 (Risk Assessment): Ground truth for ozone generation is based on empirical measurement data compared against regulatory limits (21 CFR). For tissue damage assessment, it's based on an internal risk assessment methodology comparing scenarios.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set. The device was previously cleared (K182053), and the current submission is for an expanded indication, so the "development" or "training" of the device happened prior to or as part of the predicate device's clearance.

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as no training set for an AI/ML model is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

May 4, 2020

Alesi Surgical Ltd. % Michele Lucey Regulatory Affairs Advisor Lakeshore Medical Device Consulting LLC 128 Blye Hill Landing Newbury, New Hampshire 03255

Re: K200035

Trade/Device Name: Ultravision Visual Field Clearing System Regulation Number: 21 CFR 878.5050 Regulation Name: Surgical Smoke Precipitator Regulatory Class: Class II Product Code: PQM Dated: January 31, 2020 Received: February 4, 2020

Dear Michele Lucey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT Elizabeth F. Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K200035

Device Name

Ultravision™ Visual Field Clearing System

Indications for Use (Describe)

The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during surgery, including laparoscopic surgery.

The Ultravision™ 5mm Trocar component establishes a path of entry for instruments used in laparoscopic surgery.

Type of Use (Select one or both, as applicable) Z Prescription Use (Part 21 CFR 801 Subpart D)

_Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(a)fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

K200035 510(K) SUMMARY TRADITIONAL As required by 21 CFR 807.92

Submitter Information:

Submitter's Name:Alesi Surgical Ltd
Address:Cardiff MedicentreHeath ParkCardiffCF14 4UJUK
Telephone:Fax:+44 (0) 2920291022+44 (0) 2920750239
Contact Person:Michele LuceyLakeshore Medical Device Consulting LLC.128 Blye Hill Landing.Newbury,New Hampshire 03255
Telephone:603-748-1374
Date Prepared:4th May 2020
Device Trade Name:Ultravision™ Visual Field Clearing System
Common NameSurgical Smoke Precipitator
Classification Name:21CFR 878.5050
Regulatory Class:Class II
Product Code(s):PQM
Predicate Device:Ultravision™ Visual Field Clearing System (K182053)
Reference Devices:1. Smoke Evacuator pencil system comprising of Valley LabButton switch pencil E2516H (K914400) fitted with a BuffaloPenadapt tubing set (PA 2010) (K000904) and MedtronicRapidvac smoke evacuation generator (K142335)2. Medcare Yankauer Suction device (K954869)

Indications for Use:

The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during surgery, including laparoscopic surgery. The Ultravision™ 5mm Trocar component establishes a path of entry for instruments used in laparoscopic surgery.

Device Description:

The Ultravision™ Visual Clearing System removes surgical smoke and particulates from the visual field by means of electrostatic precipitation.

{4}------------------------------------------------

The System consists of the Ultravision Generator, the Ionwand Sterile Pack, and the Ultravision 5mm Trocar. The Ionwand is connected to the energy source and is then introduced into the body cavity near the smoke generating electrosurgical device. The Ultravision™ 5mm Trocar is intended for use only with the Ultravision™M Visual Field Clearing System to introduce the Ionwand while providing a pathway for laparoscopic instruments through one 5 mm trocar incision. The trocar may be used with or without the Ionwand component of the system. The system is powered using a rechargeable battery or through mains power. Accessories include the rechargeable battery, battery recharging station, mains converter power supply, mains converter, power supply unit, and patient return adaptor.

The purpose of this submission is to add a new indication for use in open surgical procedures.

Comparison Chart
Feature/SpecificationDEVICE NAMEUltravision™ VisualField Clearing SystemPREDICATEDEVICEUltravision™ VisualField Clearing SystemComparison
Regulatory Clearance/Approval ReferenceK200035K182053N/A
Product CodePQMPQMSame
Regulation Number21 CFR 878.505021 CFR 878.5050Same
Regulation NameSurgical SmokePrecipitatorSurgical SmokePrecipitatorSame
Mechanism of ActionElectrostaticprecipitationElectrostaticprecipitationSame
Where used(environment)Operating RoomOperating RoomSame
Anatomical SitesGeneral andlaparoscopicproceduresAbdominal sites -laparoscopicproceduresSimilar
Comparison Chart
Feature/SpecificationDEVICE NAMEUltravision™ VisualField Clearing SystemPREDICATEDEVICEUltravision™ VisualField Clearing SystemComparison
Intended UseThe Ultravision™ Visual Field ClearingSystem is indicatedfor the clearance ofsmoke and otherparticulate matter thatis created duringsurgery, includinglaparoscopic surgeryThe Ultravision™5mm Trocarcomponent establishesa path of entry forinstruments used inlaparoscopic surgery.The Ultravision™ Visual Field ClearingSystem is indicatedfor the clearance ofsmoke and otherparticulate matter thatis created duringlaparoscopic surgery.The Ultravision™5mm Trocarcomponentestablishes a path ofentry for instrumentsused in laparoscopicsurgery.Similar
SoftwareYes - identical topredicateYesSame
AlarmsYes - identical topredicateYesSame
AccessoriesYes, identical topredicateYesSame
DimensionsIdenticalIdenticalSame
MaterialsIdenticalIdenticalSame
Generator Output9.8KV dc(No Change)9.8KV dc(No Change)Same
How SuppliedNon-sterile generator.Sterile Consumable(Ionwand) Non-sterileaccessories. IdenticalNon-sterile generator.Sterile Consumable(Ionwand) Non-sterileaccessories. IdenticalSame
BiocompatibilityMeets ISO 10993 Part5,10 and 11Meets ISO 10993 Part5,10 and 11Same
SterilizationEthylene OxideEthylene OxideSame
Sterility AssuranceLevel10-610-6Same

Technological Characteristics Comparison Table:

Shown below is a comparison of the subject device with the predicate device.

{5}------------------------------------------------

{6}------------------------------------------------

Summary of Non-clinical Testing:

The system has been previously updated for mains power (K182053) and accordingly new electrical safety and electromagnetic compatibility tests have been successfully completed.

There are no changes to the materials and processing of patient contact materials from the submissions in DEN150022 and K170178, consequently no additional biocompatibility testing is required.

Tests carried out to support the added indication for use in open surgical procedures includes the following:

Name of the TestPurposeAcceptance CriteriaResults
CytotoxicityISO 10993-5To analyze thepotential of the testarticle to induce acytotoxic effectMust not cause celllysis or toxicitygreater than a grade of2 (mild reactivity)Reactivity of grade 1was observed - Pass
IntracutaneousIrritationISO 10993-10To analyze thepotential of the testarticle to induce alocal irritationresponseNo evidence ofsignificantintracutaneousirritationNo evidence ofintracutaneousirritation - Pass
Systemic ToxicityISO 10993-11To analyze thepotential of the testarticle to induce asystemic responseNo evidence ofsignificant systemictoxicity or mortalityafter test articleextracts injectionNo evidence ofsystemic toxicity -Pass
MaximizationSensitizationISO 10993-10To analyze thepotential of the testarticle to induce asensitization orallergic responseNo evidence ofinduced delayedsensitizationNot a sensitizer - Pass
Performance Test 1Simulated UseComparison of smokeclearingcharacteristics ofUltravision versusother smoke clearingdevicesMust be considered atleast equivalent tocomparatorsResults of theevaluationdemonstrated thatUltravision was atleast equivalent to thecomparator devices -Pass
Performance Test 2Risk Assessmentcomparing the risksassociated with openprocedure use versuslaparoscopic useEvaluation of deviceuse in an oxygen rich(open) environmentconsidering ozonegeneration, and tissuedamage, includesempirical testing onozone generationThe risks associatedwith the use ofUltravision in an openprocedure must not begreater than those fora laparoscopicprocedure. Ozonegeneration mustcomply with 21 CFRThere were no newrisks associated withthe use of Ultravisionin an open procedure;risks were found to belower. The timeweighted average forozone production wasbelow acceptablelimits - Pass

{7}------------------------------------------------

Performance Test 1

A comparison study was conducted under simulated use conditions to demonstrate that the Ultravision Visual Field Clearing System when used in open procedures provides equivalent performance when compared to the reference device. To support this, a human factors evaluation was also conducted during simulated surgery by actual users (surgeons). The users compared the Ultravision device to an electrosurgical pencil equipped with suction tubing and provided feedback comparing the Ultravision to a hand-held suction device (reference devices).

It was concluded from this feedback that in terms of surgical workflow, ease of use and risks, that the Ultravision system was considered equivalent to the reference device.

Performance Test 2

It was recognized that a potential risk of the system being used in an oxygen rich (open) environment might be the production of ozone. Accordingly, the level of ozone generation in a worst-case model was analyzed. Risk assessment was performed to address the expected relative levels of tissue damage and injury when moving from the original cleared laparoscopic environment to the proposed open environment. This was assessed over six attributes namely, principle of operation, Ultravision Settings, Ultravision components to be utilized to clear surgical technique, principal electrosurgical modalities used with Ultravision and finally the volumes and composition of smoke produced by electrosurgical devices to be used with Ultravision. This concluded that the risk from open surgery was lower than that associated with the use of Ultravision in its current cleared laparoscopic indication.

Conclusion

Based on a review of bench top assessments, comparison of the device classification, intended use, operating principle, technological characteristics, sterility, and biocompatibility the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K182053.

§ 878.5050 Surgical smoke precipitator.

(a)
Identification. A surgical smoke precipitator is a prescription device intended for clearance of the visual field by precipitation of surgical smoke and other aerosolized particulate matter created during laparoscopic surgery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Adverse tissue reaction must be mitigated through the following:
(i) Chemical characterization and toxicological risk assessment of the treated surgical smoke.
(ii) Demonstration that the elements of the device that may contact the patient are biocompatible.
(2) Electrical safety and electromagnetic compatibility testing must demonstrate that the device performs as intended.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Performance data must demonstrate the sterility of the patient contacting components of the device.
(5) Performance data must support the shelf life of the sterile components of the device by demonstrating continued functionality, sterility, and package integrity over the identified shelf life.
(6) Animal simulated-use testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Device must be demonstrated to be effectively inserted, positioned, and removed from the site of use.
(ii) Device must be demonstrated to precipitate surgical smoke particulates to clear the visual field for laparoscopic surgeries.
(iii) Device must be demonstrated to be non-damaging to the site of use and animal subject.
(7) Labeling must identify the following:
(i) Detailed instructions for use.
(ii) Electrical safety and electromagnetic compatibility information.
(iii) A shelf life.