K Number
K170178

Validate with FDA (Live)

Date Cleared
2017-05-26

(127 days)

Product Code
Regulation Number
878.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during laparoscopic surgery.

The Ultravision 5mm Trocar component also establishes a path of entry for instruments used in laparoscopic surgery.

Device Description

The Ultravision™ Visual Clearing System removes surgical smoke from the visual field during laparoscopic surgery by means of electrostatic precipitation. The System consists of the Ultravision Generator, the Ionwand Sterile Pack, and the Ionwand 5mm Trocar. The Ionwand is connected to the energy source and is then introduced into the body cavity near the smoke generating electrosurgical device. The Ultravision™ 5mm Trocar is a new accessory device intended for use only with the Ultravision™ Visual Field Clearing System to introduce the Ionwand while providing a pathway for laparoscopic instruments through one 5 mm trocar incision. The Ultravision™ 5mm Trocar provides an alternative to the currently available Ultravision™ Visual Field Clearing System Ionwand™ Sterile Pack for introduction of the Ionwand. The Ultravision™ 5mm Trocar is a standard dilating laparoscopic trocar compatible for use with 5mm instruments. The trocar design includes a discrete lumen that is positioned separate from the main trocar channel and which accepts the Ionwand component of the system. The trocar design allows the distal tip of the Ionwand to exit the trocar some 12mm from its distal point and at an angle to the trocar body that positions the Ionwand at a point close to but not interfering with the 5mm energy instruments that are intended to be accommodated by the trocar. The Ultravision™ 5mm trocar and Ionwand are provided packaged together in the same sterile barrier packaging. The trocar may be used with or without the Ionwand component of the system.

AI/ML Overview

The provided text describes the regulatory clearance (510(k) submission) for the Ultravision™ Visual Field Clearing System, specifically focusing on the new 5mm Trocar component. It details the device's function, comparison to predicate devices, and the nonclinical testing performed to demonstrate substantial equivalence.

Based on the provided information, the device is not an AI/ML medical device. It's a surgical device for clearing smoke during laparoscopic surgery. Therefore, many of the typical acceptance criteria and study components associated with AI/ML devices (like MRMC studies, expert consensus for ground truth on training data, etc.) are not applicable here.

However, I can extract the acceptance criteria and the study (nonclinical testing) that proves the device meets them from the provided text, interpreting them in the context of a physical medical device.

Here's the information as requested, with "N/A" for sections not applicable to this type of device or not explicitly mentioned in the document:


Acceptance Criteria and Device Performance for Ultravision™ Visual Field Clearing System (K170178)

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the successful results of the specified nonclinical tests, demonstrating equivalence to predicate devices. The reported device performance is that it "Pass"-ed all these tests and was found "Equivalent" to the predicate.

Acceptance Criteria (Test)Description of TestReported Device Performance (Result)
Performance of the system to clear the visual field of smoke and particulate matterPerformance comparison to the predicate device under simulated use conditionsEquivalent to predicate - Pass
UsabilityComplexity of use compared to the predicate and reference deviceEquivalent to predicate and reference - Pass
Ionwand removal and reinsertionRepeated insertions/removals performed and forces both quantitative and qualitative evaluations performedEquivalent to predicate and reference - Pass
Leak resistance and sealing, maintenance of pneumoperitoneum during usePressure tests under high and low pressures with and without instruments or Ionwand inserted, including multiple cycles of useEquivalent to predicate or reference - Pass
Strength, resistance to bendingTensile or pressure forces applied to shaft, trocar, and connectionsEquivalent to predicate or reference - Pass
Dimensional VerificationConfirmation of dimensional characteristics by dimensional evaluationPass
Luer fittingConfirmation of compliance with industry standard luer fittingPass
Electrical SafetyDielectric breakdown, radio-frequency applied parts, capacitive couplingPass
Biocompatibility per ISO 10993Appropriate biocompatibility based on tissue contact and durationPass
Sterilization and shelf life studiesValidated sterilization cycle according to industry standard, shelf life studies on accelerated aging conditions including product and package evaluationsPass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • The document does not explicitly state the numerical sample sizes for each nonclinical test (e.g., how many units were tested for strength, how many cycles for reinsertion). It describes the type of tests performed.
  • Data Provenance:
    • Country of Origin: Not specified, but the applicant (Alesi Surgical Ltd) is based in Cardiff, GB (United Kingdom). The tests were presumably conducted internally or by a contracted lab.
    • Retrospective or Prospective: These are laboratory/benchtop tests and simulated use conditions, not clinical studies on patients. Therefore, the terms "retrospective" or "prospective" as typically applied to patient data are not directly applicable. These were newly conducted tests to support the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: N/A. Ground truth in this context (performance of a physical device) typically relies on engineering specifications, physical measurements, and standardized testing protocols rather than expert consensus on subjective interpretations.
  • Qualifications of Experts: N/A. The qualification of personnel performing the tests would be governed by internal quality systems and standards for medical device manufacturing and testing, but not explicitly stated as "experts" for ground truth establishment in the way it's done for AI models interpreting medical images.

4. Adjudication Method for the Test Set

  • Adjudication Method: N/A. Adjudication methods (like 2+1 or 3+1) are typically used for establishing ground truth from multiple human readers in image interpretation tasks. For physical device testing, results are typically objective measurements against predefined acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • MRMC Study: No. This device is not an AI/ML system or diagnostic imaging tool that would benefit from an MRMC study. Its function is direct surgical smoke clearance and trocar access, not interpretation by human readers.
  • Effect Size of Human Readers Improvement: N/A, as no MRMC study was conducted or relevant for this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • Standalone Performance: N/A. This is a physical medical device, not an algorithm. Its "performance" is its physical and functional operation.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" for this device's performance is established by:
    • Predicate Device Performance: Demonstrating "equivalence" to the legally marketed predicate device (Ultravision™ Visual Field Clearing System, DEN150022) under simulated use conditions.
    • Industry Standards and Specifications: Adherence to standards like ISO 10993 for biocompatibility, industry standards for luer fittings, and general engineering principles for strength, sealing, etc.
    • Established Methods for Sterilization and Shelf-Life Validation: Following recognized protocols for these critical aspects.
    • Objective Measurements: Dimensional verification, pressure tests, electrical safety measurements provide objective data points.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: N/A. This device is not an AI/ML system that requires a "training set" in the computational sense. Device design and development are based on engineering principles and iterative testing, not statistical training on a data set.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training Set Establishment: N/A. As there is no training set as understood in AI/ML, there is no method for establishing ground truth for it. The development and refinement of the device would involve engineering design, prototyping, and testing against functional requirements and safety standards.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 26, 2017

Alesi Surgical Ltd. David Broderick Quality & Regulatory Manager Cardiff Medicentre, Heath Park Cardiff, CF14 4UJ GB

Re: K170178

Trade/Device Name: Ultravision™ Visual Field Clearing System Regulation Number: 21 CFR 878.5050 Regulation Name: Surgical Smoke Precipitator Regulatory Class: Class II Product Code: PQM Dated: April 26, 2017 Received: May 1, 2017

Dear David Broderick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Lori A. Wiggins -S6

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170178

Device Name

Ultravision™ Visual Field Clearing System

Indications for Use (Describe)

The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other that is created during laparoscopic surgery.

The Ultravision 5mm Trocar component also establishes a path of entry for instruments used in laparoscopic surgery.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 6

510(K) SUMMARY

{4}------------------------------------------------

510(K) SUMMARY As required by 21 CFR 807.92

Submitter Information:

Submitter's Name:Alesi Surgical Ltd
Address:Cardiff MedicentreHeath ParkCardiffCF14 4UJ
Telephone:Fax:++44 (0)2920 682106++44 (0)2920 750239
Contact Person:Telephone :Michele Lucey603-748-1374
Date Prepared:19th May 2017
Device Trade Name:Ultravision™ Visual Field Clearing System
Classification:Class II
Product Code(s):POM
Regulation Number(s):878.5050
Predicate Devices:Ultravision™ Visual Field Clearing System DEN150022
Intended Use:The Ultravision™ Visual Field Clearing System is indicated forthe clearance of smoke and other particulate matter that is createdduring laparoscopic surgery.
The Ultravision 5mm Trocar component also establishes a path ofentry for instruments used in laparoscopic surgery.
Device Description:The Ultravision™ Visual Clearing System removes surgical smokefrom the visual field during laparoscopic surgery by means ofelectrostatic precipitation. The System consists of the UltravisionGenerator, the Ionwand Sterile Pack, and the Ionwand 5mmTrocar. The Ionwand is connected to the energy source and isthen introduced into the body cavity near the smoke generatingelectrosurgical device. The Ultravision™ 5mm Trocar is a newaccessory device intended for use only with the Ultravision™Visual Field Clearing System to introduce the Ionwand while

{5}------------------------------------------------

providing a pathway for laparoscopic instruments through one 5 mm trocar incision. The Ultravision™ 5mm Trocar provides an alternative to the currently available Ultravision™ Visual Field Clearing System Ionwand™ Sterile Pack for introduction of the Ionwand. The Ultravision™ 5mm Trocar is a standard dilating laparoscopic trocar compatible for use with 5mm instruments. The trocar design includes a discrete lumen that is positioned separate from the main trocar channel and which accepts the Ionwand component of the system. The trocar design allows the distal tip of the Ionwand to exit the trocar some 12mm from its distal point and at an angle to the trocar body that positions the Ionwand at a point close to but not interfering with the 5mm energy instruments that are intended to be accommodated by the trocar. The Ultravision™ 5mm trocar and Ionwand are provided packaged together in the same sterile barrier packaging. The trocar may be used with or without the Ionwand component of the system.

Nonclinical testing:

The equivalence of the Ultravision 5mm Trocar to the predicate devices has been though the following tests and assessments:

TestDescriptionResults
Performance of the system toclear the visual field of smokeand particulate matterPerformance comparison to thepredicate device undersimulated use conditionsEquivalent topredicate - Pass
UsabilityComplexity of use compared tothe predicate and referencedeviceEquivalent topredicate andreference - Pass
Ionwand removal andreinsertionRepeated insertions/removalsperformed and forces bothquantitative and qualitativeevaluations performedEquivalent topredicate andreference - Pass
Leak resistance and sealing,maintenance ofpneumoperitoneum during usePressure tests under high andlow pressures with and withoutinstruments or Ionwandinserted, including multiplecycles of useEquivalent topredicate orreference - Pass
Strength, resistance to bendingTensile or pressure forcesapplied to shaft, trocar, andconnectionsEquivalent topredicate orreference - Pass

{6}------------------------------------------------

Dimensional VerificationConfirmation of dimensionalcharacteristics by dimensionalevaluationPass
Luer fittingConfirmation of compliancewith industry standard luerfittingPass
Electrical SafetyDielectric breakdown, radio-frequency applied parts,capacitive couplingPass
Biocompatibility per ISO10993Appropriate biocompatibilitybased on tissue contact anddurationPass
Sterilization and shelf lifestudiesValidated sterilization cycleaccording to industry standard,shelf life studies on acceleratedaging conditions includingproduct and packageevaluationsPass

The results of the testing demonstrated acceptable performance for the device and substantial equivalence to the predicate device.

{7}------------------------------------------------

Feature/SpecificationUltravision™Visual FieldClearing SystemUltravision™Visual FieldClearing SystemEthicon EndopathExcel BladelessTrocar withStability Sleeve 5 -12mmComparison
Subject DevicePrimary PredicateDeviceReference Device
RegulatoryClearance/PendingDEN150022K032676N/A
Product CodePQMPQMGCJN/A
RegulationNumber878.5050878.5050876.1500Equivalent
RegulationNameSurgical SmokePrecipitatorSurgical SmokePrecipitatorEndoscopes andAccessories.Equivalent
Whereused(environment)Laparoscopic surgeryLaparoscopic surgeryAbdominal, thoracicand gynecologicalminimally invasiveproceduresEquivalent (laparoscopic surgeryincludes all the predicateanatomical sites)
AnatomicalSitesAbdominal andgynecologicalminimally invasiveproceduresAbdominal andgynecologicalminimally invasiveproceduresAs aboveEquivalent
Trocar typeBlunt trocar withdilating tipN/ABlunt trocar withdilating tipEquivalent
Trocar materialThermoplasticN/AThermoplasticEquivalent
RetentionfeaturesMicro- ridgesN/ARidgesSimilar feature but differentdimension with no impact onperformance or safety
Type of sealingemployedDuckbill valve andseparate sealN/ADuckbill valve andseparate sealEquivalent
Valve and sealmaterialElastomerElastomerElastomerEquivalent

Summary of Technological Characteristics Demonstrating Substantial Equivalence:

{8}------------------------------------------------

Feature/SpecificationUltravision™Visual FieldClearing SystemSubject DeviceUltravision™Visual FieldClearing SystemPrimary PredicateDeviceEthicon EndopathExcel BladelessTrocar withStability Sleeve 5 -12mmReference DeviceComparison
Trocar workinglength94mmN/A100 / 102mmSimilar feature but differentdimension with no impact onperformance or safety
Trocar mainlumen5.8mmN/A5.95 / 13mmSimilar feature but differentdimension with no impact onperformance or safety
Trocar O/D10.0mmN/A8.0 / 15.2mm(Covers the range ofitems from predicatesubmission)Similar feature but differentdimension with no impact onperformance or safety
Ionwand™Mworking length129mm109mmN/ASimilar feature but differentdimension with no impact onperformance or safety
Ionwand™Moverall diameter1.4mm2.0mmN/ASimilar feature but differentdimension with no impact onperformance or safety
Ionwand™ tipmaterialImplant gradeStainless steel.annealedImplant gradeStainless steel.annealed.N/AEquivalent
Ionwand™ tip,exposed length4.0mm4.0mmN/AEquivalent
Ionwand™Minsulation wallthickness0.4mm0.7mmN/ASimilar feature but differentdimension with no impact onperformance or safety
Plug and cablelength2.5mm2.5mmN/AEquivalent
GeneratorUltravision™generator (fixedpower output)Ultravision™generator (fixedpower output)N/AEquivalent
How SuppliedSupplied sterile insingle pouchesSupplied sterile insingle blisterSupplied sterile insingle pouchesEquivalent

{9}------------------------------------------------

Feature/SpecificationUltravision™Visual FieldClearing SystemSubject DeviceUltravision™MVisual FieldClearing SystemPrimary PredicateDeviceEthicon EndopathExcel BladelessTrocar withStability Sleeve 5 -12mmReference DeviceComparison
BiocompatibilityMeets ISO 10993Part 5,10 and 11Meets ISO 10993 Part5,10 and 11UnknownAssume predicate isbiocompatible since it is anapproved device
SterilizationEtOEtOGamma IrradiationSimilar feature but differentdimension with no impact onperformance or safety
SterilitvAssurance Level10-610-610-6Assume predicate has thestandard sterility assurance levelsince it is an approved device

Conclusion:

The subject device is substantially equivalent to the predicate device.

§ 878.5050 Surgical smoke precipitator.

(a)
Identification. A surgical smoke precipitator is a prescription device intended for clearance of the visual field by precipitation of surgical smoke and other aerosolized particulate matter created during laparoscopic surgery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Adverse tissue reaction must be mitigated through the following:
(i) Chemical characterization and toxicological risk assessment of the treated surgical smoke.
(ii) Demonstration that the elements of the device that may contact the patient are biocompatible.
(2) Electrical safety and electromagnetic compatibility testing must demonstrate that the device performs as intended.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Performance data must demonstrate the sterility of the patient contacting components of the device.
(5) Performance data must support the shelf life of the sterile components of the device by demonstrating continued functionality, sterility, and package integrity over the identified shelf life.
(6) Animal simulated-use testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Device must be demonstrated to be effectively inserted, positioned, and removed from the site of use.
(ii) Device must be demonstrated to precipitate surgical smoke particulates to clear the visual field for laparoscopic surgeries.
(iii) Device must be demonstrated to be non-damaging to the site of use and animal subject.
(7) Labeling must identify the following:
(i) Detailed instructions for use.
(ii) Electrical safety and electromagnetic compatibility information.
(iii) A shelf life.