Intended Use

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drug and Fentanyl Citrate (Blue) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Black) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

Device Description

Not Found

AI/ML Overview

The document describes the "Indications for Use" for two types of Nitrile Powder Free Examination Gloves (Blue and Black), specifically detailing their performance against chemotherapy drugs and Fentanyl Citrate. The study involved testing these gloves according to ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. This standard practice is the study that proves the device meets the acceptance criteria.

Here's the breakdown of the information requested:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are implicitly defined by the test results, which indicate the "Minimum Breakthrough Detection Time in Minutes" for each drug. Generally, a longer breakthrough time is desired. While the document doesn't explicitly state a minimum required breakthrough time as an acceptance criterion for all drugs (except by implication that testing was performed and results are presented), it highlights specific drugs (Carmustine and Thiotepa) with "extremely low permeation times" as a note to the user.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes (Blue Glove)Minimum Breakthrough Detection Time in Minutes (Black Glove)
Carmustine (3.3 mg/ml)23.325.0
Cisplatin (1.0 mg/ml)>240>240
Cyclophosphamide (20.0 mg/ml)>240>240
Dacarbazine (10.0 mg/ml)>240>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240>240
Etoposide (20.0 mg/ml)>240>240
Fluorouracil (50.0 mg/ml)>240>240
Methotrexate (25.0 mg/ml)>240>240
Mitomycin C (0.5 mg/ml)>240>240
Paclitaxel (6.0 mg/ml)>240>240
Thiotepa (10.0 mg/ml)58.255.7
Vincristine Sulfate (1.0 mg/ml)>240>240
Azacytidine (25.0 mg/ml)>240>240
Carboplatin (10.0 mg/ml)>240>240
Docetaxel (10 mg/ml)>240>240
Epirubicin (2.0 mg/ml)>240>240
Gemcitabine (38 mg/ml)>240>240
Ifosfamide (50 mg/ml)>240>240
Irinotecan (20 mg/ml)>240>240
Mitoxantrone (2.0 mg/ml)>240>240
Oncovin (1.0 mg/ml)>240>240
Oxaliplatin (5 mg/ml)>240>240
Vinorelbine (10 mg/ml)>240>240
Fentanyl Citrate Inj. (100 mcg/2ml)>240>240

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size (number of gloves tested) for each drug or the total sample size used in the testing. The data provenance (country of origin, retrospective/prospective) is also not specified in the provided text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a physical product performance test (permeation of gloves by corrosive chemicals), not a diagnostic or AI-assisted interpretation study. Therefore, the concept of "experts establishing ground truth for the test set" and their qualifications does not apply in this context. The "ground truth" is established by the direct measurement of breakthrough time according to the standardized ASTM D6978-05 method.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation or expert consensus for labeling data. This is a laboratory test of material properties, so such adjudication methods are not applicable. The results are quantitative measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC study was not done. This is a test of a physical device (gloves) against chemical permeation, not a study evaluating human reader performance or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, this is not an algorithm or AI performance study, so the concept of standalone performance does not apply.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" in this context is the direct measurement of permeation breakthrough time as determined by the standardized test method ASTM D6978-05 (Reapproved 2013). This is empirical, scientifically derived data from controlled laboratory testing.

8. The sample size for the training set

This is not an AI/machine learning study, so there is no training set. The term "training set" is not applicable here.

9. How the ground truth for the training set was established

Since there is no training set for an AI model, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 6, 2020

Hartalega NGC Sdn. Bhd. Nurul Kong Senior Manager - Quality Assurance Kawasan Perindustrian Tanjung Sepang, 43900 My

Re: K200019

Trade/Device Name: Nitrile Powder Free Examination Gloved Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue), Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Black) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, QDO Dated: January 10, 2019 Received: January 15, 2020

Dear Nurul Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K200019

Device Name

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drug and Fentanyl Citrate (Blue)

Indications for Use (Describe)

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drug and Fentanyl Citrate (Blue) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate,

These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs,

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)23.3
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)58.2
Vincristine Sulfate (1.0 mg/ml)>240
Azacytidine (25.0 mg/ml)>240
Carboplatin (10.0 mg/ml)>240
Docetaxel (10 mg/ml)>240
Epirubicin (2.0 mg/ml)>240
Gemcitabine (38 mg/ml)>240
Ifosfamide (50 mg/ml)>240
Irinotecan (20 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Oncovin (1.0 mg/ml)>240
Oxaliplatin (5 mg/ml)>240
Vinorelbine (10 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 23.3 minutes and 58.2 minutes respectively.

Fentanyl Citrate and ConcentrationFentanyl Citrate Injection (100 mcg/2ml)Minimum Breakthrough Detection Time in Minutes>240
----------------------------------------------------------------------------------------------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K200019

Device Name

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drug and Fentany) Citrate (Black)

Indications for Use (Describe)

Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Black) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3,3 mg/ml)25.0
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)55.7
Vincristine Sulfate (1.0 mg/ml)>240
Azacytidine (25.0 mg/ml)>240
Carboplatin (10.0 mg/ml)>240
Docetaxel (10 mg/ml)>240
Epirubicin (2.0 mg/ml)>240
Gemcitabine (38 mg/ml)>240
Ifosfamide (50 mg/ml)>240
Irinotecan (20 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Oncovin (1.0 mg/ml)>240
Oxaliplatin (5 mg/ml)>240
Vinorelbine (10 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 25.0 minutes and 55.7 minutes respectively.

Fentanyl Citrate and Concentration Fentanyl Citrate Injection (100 mcg/2ml) Minimum Breakthrough Detection Time in Minutes >240

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

FORM FDA 3881 (7/17)

{5}------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number:"

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.