K Number
K193604

Validate with FDA (Live)

Date Cleared
2021-06-16

(541 days)

Product Code
Regulation Number
866.5660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aptiva Celiac Disease IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semiquantitative determination of anti-tissue transglutaminase IgA autoantibodies and anti-deamidated gliadin peptide IgA autoantibodies in human serum. The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis. The Aptiva Celiac Disease IgA Reagent is intended for use with the Inova Diagnostics Aptiva System.

Device Description

The Aptiva Celiac Disease IgA reagent utilizes particle based multi-analyte technology (PMAT) in a cartridge format. Each analyte (tissue transglutaminase [tTG] and deamidated gliadin peptide [DGP]) in the Aptiva Celiac Disease IgA reagent is a solid phase immunoassay utilizing fluorescent microparticles. This technology allows each of the two analytes, along with a human IgA capture antibody (IgA Control Microparticle), to be coated onto three uniquely recognizable paramagnetic microparticles, which are combined into one tube.

The Aptiva instrument is a fully automated, random access analyzer. This platform is a closed system with continuous load and random-access capabilities that processes the samples, runs the reagent and reports results. It includes liquid handling hardware, optical module (OM), and integrated computer with proprietary software and touch screen user interface.

The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge under conditions that preserve the proteins in their reactive states. When the assay cartridge is ready to be used for the first time, the reagent tube seals are pierced using the cartridge lid. The reagent cartridge is then loaded onto the Aptiva instrument, where the microparticles are automatically rehydrated using buffer located within the cartridge.

A patient's serum is diluted 1:46 with Aptiva system rinse by the instrument in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and the microparticle suspension in a second cuvette, and mixed (final serum dilution: 1:230). This reaction cuvette is incubated for 9 ½ minutes at 37°C. The cuvette is then exposed to a small magnet that holds the microparticles in place. The liquid is aspirated, and the microparticles are resuspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step. After the third wash, phycoerythrin conjugated polyclonal anti-human IgA (known as PE Tracer IgA) is added to the microparticles in the cuvette, and mixed. Again, the cuvette is incubated for 9 ½ minutes at 37℃. Three wash steps, as described above, are performed on the microparticles. Following the wash steps, the microparticles are transferred to the of the instrument, where a charge coupled device (CCD) camera takes multiple images in order to identify and count the three unique microparticle regions, as well as determine the amount of conjugate on the microparticles. A third particle, coated with goat antibodies, is present in the reagent as a control to flag low concentrations of IgA in the sample as an assay verification step. The median fluorescent intensity (MFI) is proportional to the amount of PE Tracer that is bound to the human IgA, which is proportional to the amount of IgA antibodies bound to the corresponding microparticle regions.

For quantitation, the DGP IgA and tTG IgA assays (together as part of the Aptiva Celiac Disease IgA Reagent) each utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge RFID tag. Every new lot of reagent cartridge must be calibrated before first use with the reagent specific calibrators. Based on the results obtained with the calibrators included in the Aptiva Celiac Disease IgA Calibrator kit (sold separately), an instrument specific Working Curve is created for each assay, which is used to calculate reported fluorescent light units (FLU) from the median fluorescent intensity (MFI) instrument signal obtained for each sample, on each of the two assays within the reagent.

Aptiva Celiac Disease IgA Calibrators and Aptiva Celiac Disease IgA Controls are sold separately.

The Aptiva Celiac Disease IgA Reagent kit contains the following materials:

One (1) Aptiva Celiac Disease IgA Reagent Cartridge, containing the following reagents for 250 determinations:

  • a. Aptiva Celiac IgA microparticle containing 3 unique microparticle regions coated with recombinant tissue transglutaminase, deamidated gliadin peptide, or goat anti-human IgA antibody.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • PE Tracer IgA phycoerythrin (PE) labeled anti-human IgA antibody, containing buffer, C. protein stabilizers and preservative.
  • d. Rehydration Buffer - containing protein stabilizers and preservatives.
AI/ML Overview

The provided text is a 510(k) Summary for the Aptiva Celiac Disease IgA Reagent, an in vitro diagnostic device. It describes the analytical and clinical performance of the device to demonstrate its substantial equivalence to predicate devices. It does not describe an AI/ML-based device, a comparative effectiveness study with human readers, or the establishment of ground truth by expert consensus. Therefore, most of the requested information cannot be extracted from this document as it pertains to AI/ML device studies.

However, I can extract the acceptance criteria and reported performance for analytical aspects of this specific in vitro diagnostic device, as well as details regarding sample size, data provenance, and the type of ground truth used for performance evaluation.


Acceptance Criteria and Reported Device Performance

The device under review is an in vitro diagnostic (IVD) test, not an AI/ML-based medical imaging device. As such, the acceptance criteria and performance evaluation are based on typical analytical validation parameters for immunological assays, such as precision, limit of detection, linearity, interference, and clinical sensitivity/specificity against established reference methods or patient diagnoses.

Table of Acceptance Criteria and Reported Device Performance:

Study/ParameterAcceptance Criteria (Set by Manufacturer)Reported Device Performance (as presented)
PrecisionTotal %CV: < 12%DGP IgA: All samples met criteria. Max Total %CV: 9.5% (Sample 7). tTG IgA: All samples met criteria. Max Total %CV: 8.1% (Sample 3).
Reproducibility (Between Sites)Reproducibility Between-Site %CV: < 12%DGP IgA: All samples met criteria. Max Reproducibility %CV: 11.1% (Sample 4). tTG IgA: All samples met criteria. Max Reproducibility %CV: 10.0% (Sample 1).
Reproducibility (Between Lots)Reproducibility Between-Lot %CV: < 12%DGP IgA: All samples met criteria. Max Reproducibility %CV: 9.9% (Sample 2). tTG IgA: All samples met criteria. Max Reproducibility %CV: 12.0% (Sample 6). (Note: This one is exactly at the limit)
LoQ for DGP IgATotal imprecision < 20%Final LoQ value: 0.72 FLU (set as lower limit of AMR).
LoQ for tTG IgATotal imprecision < 20%Final LoQ value: 1.02 FLU (set as lower limit of AMR).
LinearityBest fitting polynomial is linear OR difference between best-fitting non-linear and linear polynomial is < 15% or ±0.75 FLU for low level samples.DGP IgA: Samples 1 & 4 linear, Samples 2 & 3 non-linear (3rd and 2nd order polynomial, respectively). All fulfilled acceptance criteria for allowable nonlinearity. tTG IgA: All samples determined to be linear. All fulfilled acceptance criteria.
Interference85% - 115% recovery, or ± 15% of the cut-off (±0.75 FLU), whichever is greater.No interference detected for DGP or tTG IgA with bilirubin, hemoglobin, triglycerides, cholesterol, RF IgM, and human IgG within specified concentrations. All recoveries within criteria.
Sample Stability% recovery between 85-115% for positive samples, and between 80-120% for negative samples (<5.00 FLU).All samples fulfilled acceptance criteria at each time point for 48 hours at room temp, 14 days at 2-8°C, and up to 5 freeze/thaw cycles.
Reagent Shelf LifeLower and upper 95% CI of regression line between 80% and 120% recovery at day 28 (week 4) for accelerated stability.All components fulfilled acceptance criteria, allowing for a two-year preliminary expiration dating claim.
Reagent In-use (Onboard) StabilityStability claim established at actual measurement day preceding 95% CI of regression line reaching 85% or 115% recovery OR actual measurement day preceding ≥2% of recovery data (<75% or ≥125%).Onboard stability of Aptiva Celiac Disease IgA reagent cartridge set at 42 days (based on Lot 100014).
Clinical Performance (Sensitivity/Specificity)(Implicitly, to demonstrate substantial equivalence to predicate device)Aptiva DGP IgA: Sensitivity: 59.1% (51.6 – 66.2%), Specificity: 99.3% (97.5 – 99.8%) Aptiva tTG IgA: Sensitivity: 93.0% (88.1 – 95.9%), Specificity: 99.3% (97.5 – 99.8%) Dermatitis Herpetiformis: DGP IgA Sensitivity: 64.7%, tTG IgA Sensitivity: 91.2% (Specificity same as above).
Method Comparison (Agreement vs. Predicate)(Implicitly, to demonstrate substantial equivalence to predicate device)Aptiva DGP IgA vs. QUANTA Flash DGP IgA (N=200): NPA: 96.9% (89.5–99.2%), PPA: 85.2% (78.2 – 90.2%), TPA: 89.0% (83.9 – 92.6%). Aptiva tTG IgA vs. QUANTA Flash tTG IgA (N=197): NPA: 96.9% (84.3–99.4%), PPA: 98.8% (95.7 – 99.7%), TPA: 98.5% (95.6 – 99.5%).

Study Details:

  1. Sample sizes used for the test set and data provenance:

    • Precision: 9 samples for DGP IgA, 10 samples for tTG IgA (run in duplicates, twice a day, for 20 days).
    • Reproducibility (Between Sites): 7 samples for DGP IgA and 6 samples for tTG IgA.
    • Reproducibility (Between Lots): 6 samples for DGP IgA and 6 samples for tTG IgA.
    • LoB, LoD, LoQ:
      • LoB: 8 blank samples. For each assay (DGP IgA & tTG IgA), on two reagent lots, run in replicates of 5, once per day for 3 days (120 data points per lot).
      • LoD & LoQ: 4 low-level samples for each assay (DGP IgA & tTG IgA). For each assay, on two reagent lots, run in replicates of 5, twice per day for 3 days (120 data points per assay per lot).
    • Linearity: 4 human serum samples for each assay, serially diluted and assayed in duplicates.
    • Interference: 3 human serum specimens (one positive, one near cutoff, one negative).
    • Sample Stability: 8 test samples for DGP IgA, 6 test samples for tTG IgA.
    • Clinical Performance (Validation Set): A total of 495 characterized samples.
      • 171 samples from celiac disease patients.
      • 34 dermatitis herpetiformis patients.
      • 290 control samples from patients with various types of autoimmune and infectious diseases (e.g., Rheumatoid Arthritis, Ulcerative Colitis, Crohn's Disease, Hepatitis C/B, Syphilis, Sjögren's Syndrome, Systemic Sclerosis, Autoimmune Gastritis, HIV, Systemic Lupus Erythematosus, Epstein-Barr Virus).
      • The document does not explicitly state the country of origin but implies data collection from clinical settings. It describes the use of "characterized samples" and "diagnosed celiac disease (CD) patient specimens," indicating that this was likely a retrospective collection of samples with established diagnoses.
    • Expected Values (Normal Population): 120 apparently healthy blood donors.
    • Method Comparison: All 495 samples from the clinical validation study were also used for method comparison against the predicate devices.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is an IVD device, not an AI/ML imaging device. The "ground truth" for the test set (clinical validation cohort) was based on patient diagnoses (e.g., "celiac disease patients," "dermatitis herpetiformis," "control samples from patients with various types of autoimmune and infectious diseases"). Therefore, the ground truth was established by clinical diagnosis, which would typically be made by medical doctors/specialists based on relevant clinical findings and other laboratory tests, rather than by a specific number of experts reviewing image data. The document does not specify the number or qualifications of clinicians involved in establishing these diagnoses.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable, as this is an IVD test assessing biochemical markers, not an imaging device requiring expert adjudication of interpretations. The "ground truth" is the established clinical diagnosis of the patient.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML imaging device, and no MRMC study was performed or is relevant to this type of IVD test.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The "standalone performance" of this device is represented by its analytical performance characteristics (precision, linearity, LoD, etc.) and its clinical sensitivity and specificity, where the device provides a quantitative result (FLU) to aid in diagnosis. There is no "human-in-the-loop" aspect to the performance of the device itself (it's an automated analyzer), though clinicians interpret its results in conjunction with other clinical findings.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The primary ground truth for the clinical performance evaluation was established clinical diagnoses of patients ("celiac disease patients," "dermatitis herpetiformis patients," and "control samples from patients with various types of autoimmune and infectious diseases"). This implicitly relies on a combination of clinical findings, potentially other laboratory tests, and possibly biopsy results (pathology) for definitive diagnoses like celiac disease. The document states, "The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis."
  7. The sample size for the training set:

    • This device is an immunoassay (particle-based multi-analyte technology) operating on predefined Master Curves and calibrations. It does not use a "training set" in the sense of machine learning algorithms. The Master Curves and calibrations are established by the manufacturer through runs of precisely quantified standards (e.g., "in-house Master Curve Standards," "reagent specific calibrators"). The section "Quantitation and units of measure" describes how these curves are generated. For example, "Aptiva Celiac Disease IgA Master Curve Standards - DGP IgA" lists 6 standards with assigned FLU values. While these could be seen as "training data" for the device's internal quantitation function, it's not a machine learning model.
  8. How the ground truth for the training set was established:

    • Not applicable in an AI/ML context. For this IVD device, the "ground truth" for its internal calibration (analogous to a training set for an AI model) is based on manufacturer-defined standards with assigned known values ("in-house Master Curve Standards with assigned FLU values"). These standards are run multiple times to generate the 4-parameter logistic (4PL) curve that quantifies the analyte concentration.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

June 16, 2021

Inova Diagnostics, Inc. Ronda Elliott VP, Quality Systems and Regulatory Affairs 9900 Old Grove Road San Diego, California 92131

Re: K193604

Trade/Device Name: Aptiva Celiac Disease IgA Reagent Regulation Number: 21 CFR 866.5750 Regulation Name: Radioallergosorbent (RAST) immunological test system Regulatory Class: Class II Product Code: MST. MVM. NSU Dated: October 20, 2020 Received: October 21, 2020

Dear Ronda Elliott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying Mao for Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193604

Device Name Aptiva Celiac Disease IgA Reagent

Indications for Use (Describe)

The Aptiva Celiac Disease IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semiquantitative determination of anti-tissue transglutaminase IgA autoantibodies and anti-deamidated gliadin peptide IgA autoantibodies in human serum. The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis. The Aptiva Celiac Disease IgA Reagent is intended for use with the Inova Diagnostics Aptiva System.

Aptiva System is an automated particle-based multi-analyzer for in vitro diagnostic testing of clinical specimens. The system is based on digital capture of high-resolution images of the paramagnetic particles to determine the analytes in samples.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized graphic element with a red and blue design, followed by the text "Inova Diagnostics" in a clear, sans-serif font. Below the company name, the text "A Werfen Company" is displayed in a smaller font size.

510(k) Summary

Page 1 of 19

Aptiva Celiac Disease IgA Reagent

{4}------------------------------------------------

Page 2 of 19

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Administrative data
Submitter:Inova Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device
Device in the submission:Aptiva Celiac Disease IgA Reagent
Revision Date:June 4, 2021
Scientific contact:Andrea Seaman, Manager, Research and Development
Inova Diagnostics, Inc.
9900 Old Grove Road, San Diego, CA, 92131
Phone: 858-586-9900 x1395
Fax: 858-863-0025
Email: aseaman@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RA
Inova Diagnostics, Inc
9900 Old Grove Road, San Diego, CA, 92131
Phone: 858-586-9900 x1381
Fax: 858-863-0025
Email: relliott@inovadx.com
Device name (kit):Proprietary name:Aptiva Celiac Disease IgA Reagent
Common name:anti-deamidated gliadin peptide (DGP) antibodyimmunoassay, anti-tissue transglutaminaseantibody immunoassay
Classification name:DGP IgA: Radioallergosorbent (RAST)immunological test systemtTG IgA: Multiple autoantibodies immunological

test system

{5}------------------------------------------------

Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeDGP IgA: MSTtTG IgA: MVMAptiva instrument: NSU
Regulation Number866.5750, 866.5660, Aptiva instrument: 862.2570
Device Class2

Predicate device

QUANTA Flash® DGP IgA, 510(k) number: K113863. Date of clearance: September 20, 2012. QUANTA Flash® h-tTG IgA, 510(k) number: K094060. Date of clearance: October 13, 2010.

Device description

The Aptiva Celiac Disease IgA reagent utilizes particle based multi-analyte technology (PMAT) in a cartridge format. Each analyte (tissue transglutaminase [tTG] and deamidated gliadin peptide [DGP]) in the Aptiva Celiac Disease IgA reagent is a solid phase immunoassay utilizing fluorescent microparticles. This technology allows each of the two analytes, along with a human IgA capture antibody (IgA Control Microparticle), to be coated onto three uniquely recognizable paramagnetic microparticles, which are combined into one tube.

The Aptiva instrument is a fully automated, random access analyzer. This platform is a closed system with continuous load and random-access capabilities that processes the samples, runs the reagent and reports results. It includes liquid handling hardware, optical module (OM), and integrated computer with proprietary software and touch screen user interface.

The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge under conditions that preserve the proteins in their reactive states. When the assay cartridge is ready to be used for the first time, the reagent tube seals are pierced using the cartridge lid. The reagent cartridge is then loaded onto the Aptiva instrument, where the microparticles are automatically rehydrated using buffer located within the cartridge.

A patient's serum is diluted 1:46 with Aptiva system rinse by the instrument in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and the microparticle suspension in a second cuvette, and mixed (final serum dilution: 1:230). This reaction cuvette is incubated for 9 ½ minutes at 37°C. The cuvette is then exposed to a small magnet that holds the microparticles in place. The liquid is aspirated, and the microparticles are resuspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step. After the third wash, phycoerythrin conjugated polyclonal anti-human IgA (known as PE Tracer IgA) is added to the microparticles in the cuvette, and mixed. Again, the cuvette is incubated for 9 ½ minutes at 37℃. Three wash steps, as described above, are performed on the microparticles. Following the wash steps, the microparticles are transferred to the of the instrument, where a charge coupled device (CCD) camera takes multiple images in order to identify and

{6}------------------------------------------------

count the three unique microparticle regions, as well as determine the amount of conjugate on the microparticles. A third particle, coated with goat antibodies, is present in the reagent as a control to flag low concentrations of IgA in the sample as an assay verification step. The median fluorescent intensity (MFI) is proportional to the amount of PE Tracer that is bound to the human IgA, which is proportional to the amount of IgA antibodies bound to the corresponding microparticle regions.

For quantitation, the DGP IgA and tTG IgA assays (together as part of the Aptiva Celiac Disease IgA Reagent) each utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge RFID tag. Every new lot of reagent cartridge must be calibrated before first use with the reagent specific calibrators. Based on the results obtained with the calibrators included in the Aptiva Celiac Disease IgA Calibrator kit (sold separately), an instrument specific Working Curve is created for each assay, which is used to calculate reported fluorescent light units (FLU) from the median fluorescent intensity (MFI) instrument signal obtained for each sample, on each of the two assays within the reagent.

Aptiva Celiac Disease IgA Calibrators and Aptiva Celiac Disease IgA Controls are sold separately.

The Aptiva Celiac Disease IgA Reagent kit contains the following materials:

One (1) Aptiva Celiac Disease IgA Reagent Cartridge, containing the following reagents for 250 determinations:

  • a. Aptiva Celiac IgA microparticle containing 3 unique microparticle regions coated with recombinant tissue transglutaminase, deamidated gliadin peptide, or goat anti-human IgA antibody.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • PE Tracer IgA phycoerythrin (PE) labeled anti-human IgA antibody, containing buffer, C. protein stabilizers and preservative.
  • d. Rehydration Buffer - containing protein stabilizers and preservatives.

Intended use(s)

The Aptiva Celiac Disease IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgA autoantibodies and antideamidated gliadin peptide IgA autoantibodies in human serum. The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis.

The Aptiva Celiac Disease IgA Reagent is intended for use with the Inova Diagnostics Aptiva System.

Indications for use

Same as intended use.

Substantial equivalence

The Aptiva Celiac Disease IgA Reagent has the same intended use and assay principle as the predicate devices.

{7}------------------------------------------------

Comparison to predicate device

Aptiva Celiac Disease IgA Reagent – DGP IgA Assay

Similarities
ItemAptiva Celiac Disease IgA Reagent (DGP IgA)QUANTA Flash DGP IgA
Intended UseThe Aptiva Celiac Disease IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgA autoantibodies and anti-deamidated gliadin peptide IgA autoantibodies in human serum. The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis. The Aptiva Celiac Disease IgA Reagent is intended for use with the Inova Diagnostics Aptiva SystemThe QUANTA Flash DGP IgA is a chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgA anti-deamidated gliadin peptide (DGP) antibodies in human serum. The presence of IgA anti-DGP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of the gluten sensitive enteropathies: celiac disease and dermatitis herpetiformis.
Assay Methodologysolid phase (heterogeneous) immunoassaysolid phase (heterogeneous) immunoassay
Antigendeamidated gliadin peptidedeamidated gliadin peptide
Sample Typehuman serumhuman serum
Solid Phaseparamagnetic microparticlesparamagnetic microparticles
Differences
ItemAptiva Celiac Disease IgA Reagent (DGP IgA)QUANTA Flash DGP IgA
Detection/Operating Principlefluorescent immunoassaychemiluminescent immunoassay
Conjugatephycoerythrin conjugated polyclonal anti-human IgA antibodyIsoluminol conjugated monoclonal anti-human IgA antibody
Unitsfluorescent light units (FLU)chemiluminescent units (CU)
Cut-off5 FLU20 CU
Similarities
ItemAptiva Celiac Disease IgA Reagent(DGP IgA)QUANTA Flash DGP IgA
Analytical MeasuringRange0.72 FLU - 250.00 FLU5.2 CU - 2367.3 CU
ControlControls have lot specific values assigned.Control 1DGP IgA - 6.65 FLU (4.65 – 8.64)tTG IgA - 10.00 FLU (5.00-100.00)Control 2DGP IgA - 13.65 FLU (9.56–17.75)tTG IgA - 50.00 FLU (25.00 – 250.00)Controls have lot specific values assigned.Negative ControlDGP IgA - 9.7 CU (5.8 – 13.6)h-tTG IgA - 10.2 CU (6.1 – 14.3)Positive ControlDGP IgA – 50.9 CU (30.5 – 71.3)h-tTG IgA - 62.5 CU (37.5-87.5)
CalibrationLot specific Master Curve + 3calibrators (sold separately)Lot specific Master Curve + 2calibrators (sold separately)

{8}------------------------------------------------

Aptiva Celiac Disease IgA Reagent – tTG IgA Assay

Similarities
ItemAptiva Celiac Disease IgA Reagent(tTG IgA)QUANTA Flash tTG IgA
Intended UseThe Aptiva Celiac Disease IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgA autoantibodies and anti-deamidated gliadin peptide IgA autoantibodies in human serum. The presence of these autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis. The Aptiva Celiac Disease IgA Reagent is intended for use with the Inova Diagnostics Aptiva SystemThe QUANTA Flash h-tTG IgA is a chemiluminescent immunoassay (CIA) for the semi-quantitative detection of IgA anti-human tissue transglutaminase (h-tTG) antibodies in human serum. The presence of IgA anti-h-tTG antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of the gluten sensitive enteropathies celiac disease (CD) and dermatitis herpetiformis (DH).

{9}------------------------------------------------

Assay Methodologysolid phase (heterogeneous)immunoassaysolid phase (heterogeneous)immunoassay
Antigenrecombinant tissuetransglutaminaserecombinant tissuetransglutaminase
Sample Typehuman serumhuman serum
Solid Phaseparamagnetic microparticlesparamagnetic microparticles

{10}------------------------------------------------

Differences
ItemAptiva Celiac Disease IgA Reagent(tTG IgA)QUANTA Flash tTG IgA
Detection/OperatingPrinciplefluorescent immunoassaychemiluminescent immunoassay
ConjugatePhycoerythrin conjugatedpolyclonal anti-human IgAantibodyIsoluminol conjugatedmonoclonal anti-human IgAantibody
Unitsfluorescent light units (FLU)chemiluminescent units (CU)
Cut-off5 FLU20 CU
Analytical MeasuringRange1.02 FLU - 600.00 FLU1.9 CU - 4965.5 CU
CalibrationLot specific Master Curve + 3calibrators (sold separately)Lot specific Master Curve + 2calibrators (sold separately)

Analytical performance characteristics

Quantitation and units of measure

For quantitation, the Aptiva Celiac Disease IgA reagent utilizes predefined lot specific Master Curves, one for tTG IgA and one for DGP IgA that is uploaded onto the instrument through the reagent cartridge RFID. The analyte specific Master Curves are generated at Inova for each reagent lot, where in-house Master Curve Standards with assigned FLU values are run multiple times. The resulting MFI values generated are used to create a unique 4 parameter logistic (4PL) curve for each of the two analytes. The IgA control bead is present in the reagent as a control to flag low concentrations of IgA in the sample as an assay verification step. This microparticle also has an in-house standard which is run each time a new reagent lot is manufactured. The MFI produced by this standard is used as the cut-off threshold for the IgA control microparticle for that reagent lot. These four parameters of the analyte curves, as well as the MFI cut-off for the IgA control microparticle are embedded in the reagent cartridge RFID.

List of Aptiva Celiac Disease IgA Master Curve Standards - DGP IgA:

MaterialAssigned Value (FLU)
Celiac IgA Master Curve Standard 10.00
Celiac IgA Master Curve Standard 24.81
Celiac IgA Master Curve Standard 316.83
Celiac IgA Master Curve Standard 458.89
Celiac IgA Master Curve Standard 5206.13
Celiac IgA Master Curve Standard 6721.45

{11}------------------------------------------------

MaterialAssigned Value (FLU)
Celiac IgA Master Curve Standard 10.00
Celiac IgA Master Curve Standard 24.55
Celiac IgA Master Curve Standard 315.93
Celiac IgA Master Curve Standard 455.76
Celiac IgA Master Curve Standard 5194.90
Celiac IgA Master Curve Standard 6685.57

List of Aptiva Celiac Disease IgA Master Curve Standards – tTG IgA

IgA Control Microparticle Standard: 1 mg/dL human IgA

Precision

The precision of the Aptiva Celiac Disease IgA reagent was evaluated on 9 samples for DGP IgA and 10 samples for tTG IgA, containing various concentrations of antibodies in accordance with CLSI EPO5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.

Data were analyzed with the Analyse-it for Excel method evaluation software, and repeatability (withinrun), between run, between day and within-laboratory precision) were calculated. Results are summarized in the two tables below.

Aptiva DGP IgA PrecisionRepeatabilityBetween RunBetween DayWithinLaboratory
SampleReplicates(N)Mean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1802.020.083.8%0.031.5%0.083.9%0.115.7%
2803.750.092.3%0.225.9%0.205.2%0.318.2%
3804.530.163.4%0.204.4%0.204.4%0.327.1%
4805.410.234.2%0.244.4%0.203.7%0.387.1%
5806.530.152.3%0.345.2%0.365.5%0.527.9%
68012.420.342.7%0.332.7%0.776.2%0.907.3%
78034.030.982.9%2.547.5%1.775.2%3.259.5%
880153.513.702.4%5.403.5%7.755.0%10.146.6%
980203.784.312.1%2.851.4%6.213.0%8.084.0%

Acceptance criteria: Total %CV: < 12%

{12}------------------------------------------------

Aptiva tTG IgA PrecisionRepeatabilityBetween RunBetween DayWithinLaboratory
SampleReplicates(N)Mean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1801.940.073.8%17.201.8%0.021.2%0.084.4%
2804.120.245.8%0.143.3%0.000.0%0.276.6%
3805.550.275.0%0.224.0%0.285.1%0.458.1%
4806.740.334.9%0.202.9%0.324.7%0.507.3%
58017.720.834.7%0.633.5%0.442.5%1.136.4%
68081.802.513.1%2.493.0%2.142.6%4.145.1%
780165.457.344.4%5.783.5%8.495.1%12.627.6%
880274.2213.554.9%5.862.1%7.172.6%16.416.0%
980398.6417.284.3%10.292.6%11.732.9%23.285.8%
1080491.0425.975.3%29.7%4.0%14.883.0%35.717.3%

Reproducibility Studies

Reproducibility between sites (instruments)

Seven samples for DGP IgA and six samples for tTG IgA were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per sample, per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision.

Acceptance criteria: Reproducibility Between-Site %CV: < 12%

Aptiva DGP IgARepeatabilityBetween-DayWithin-SiteBetween-Site/InstrumentReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1751.750.084.3%0.127.1%0.158.3%0.063.2%0.168.9%
2754.390.143.2%0.286.3%0.317.1%0.092.1%0.327.4%
3755.300.152.8%0.142.7%0.213.9%0.000.0%0.213.9%
4756.430.193.0%0.6610.2%0.6910.7%0.203.1%0.7111.1%
57538.101.052.7%2.476.5%2.687.0%1.834.8%3.258.5%
67591.852.572.8%8.959.7%9.3110.1%0.000.0%9.3110.1%
775167.903.011.8%10.146.1%10.586.3%11.556.9%15.669.4%

Results are summarized in the tables below.

{13}------------------------------------------------

Aptiva tTG IgARepeatabilityBetween-DayWithin-SiteBetween-Site/InstrumentReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.440.072.8%0.239.4%0.249.8%0.052.0%0.2410.0%
2754.950.132.7%0.224.5%0.265.3%0.000.0%0.265.3%
3756.730.253.7%0.537.9%0.598.7%0.243.5%0.639.4%
47577.161.582.0%1.501.9%2.172.8%2.303.0%3.174.1%
575140.933.982.8%2.772.0%4.853.4%5.533.9%7.365.2%
675219.936.653.0%10.004.5%12.015.5%9.084.1%15.056.8%

Reproducibility between lots

Six samples for DGP IgA and six samples for tTG IgA were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, using three different lots. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per lot, 75 data points total for each sample. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between lot precision.

Acceptance criteria: Reproducibility Between-Lot %CV: < 12%

Aptiva DGP IgARepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.480.093.8%0.072.9%0.124.7%0.218.4%0.249.7%
2755.270.173.2%0.193.6%0.254.8%0.468.7%0.529.9%
3755.450.163.0%0.162.9%0.234.2%0.488.8%0.539.8%
47538.520.942.4%0.541.4%1.082.8%2.346.1%2.586.7%
575100.942.192.2%2.302.3%3.173.1%0.820.8%3.283.2%
675177.973.962.2%6.353.6%7.484.2%2.121.2%7.784.4%

Results are summarized in the tables below.

Aptiva tTG IgARepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleNMean(FLU)SD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CVSD(FLU)CV
1752.260.104.2%0.093.9%0.135.8%0.229.6%0.2511.2%
2754.040.163.9%0.112.7%0.194.7%0.286.9%0.348.3%
3755.200.142.8%0.224.2%0.265.0%0.509.5%0.5610.8%
47556.361.833.2%1.933.4%2.664.7%0.200.3%2.674.7%
575164.525.313.2%5.023.1%7.304.4%5.033.1%8.875.4%
675373.2220.795.6%16.244.3%26.387.1%36.029.7%44.6512.0%

{14}------------------------------------------------

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)

The LoB, LoD, and LoQ of the DGP IgA and tTG IgA assays in the Aptiva Celiac Disease IgA Reagent were calculated separately by a study according to CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline- Second Edition.

Study protocol for LoB:

Eight blank samples (Aptiva System Rinse) from two different lots were run in replicates of five on two reagent lots, once per day, for 3 days, with 120 data points generated on each lot. The LoB was determined for each assay, on each reagent lot separately with the Analyse-it for Excel software's Reference Interval function, at the 95th percentile, using the non-parametric method for all 4 analyses (DGP IgA on two reagent lots, and tTG IgA on two reagent lots; all having a p-value = <0.0001)

The DGP IgA LoB for both reagent lots were determined as 0.00 FLU. The final LoB value for DGP IgA is 0.00 FLU.

The tTG IgA LoB for both reagent lots were-determined as 0.01 FLU. The final LoB value for tTG IgA is 0.01 FLU.

Study protocol for LoD:

Four low level samples for each DGP IgA and tTG IgA assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoD was determined separately for each assay, on each reagent lot.

The DGP IgA limit of detection for one reagent lot was determined as 0.65 FLU, and for the second reagent lot as 0.39 FLU. The final LoD value is 0.65 FLU.

The tTG IgA limit of detection for one reagent lot was determined as 0.15 FLU, and for the second reagent lot as 0.37 FLU. The final LoD value is 0.37 FLU.

Study protocol for LoQ:

Four low level samples for each DGP IgA and tTG IgA assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoQ was determined separately for each assay, on each reagent lot. The LoQ was determined in each case by calculating the total imprecision of each sample (acceptance criteria: total imprecision <20%).

The DGP IgA limit of quantitation for one reagent lot was determined as 0.67 FLU, and for the second reagent lot as 0.72 FLU. The final LoQ value is 0.72 FLU, which has been set as the lower limit of the analytical measuring range of the DGP IgA assay.

{15}------------------------------------------------

The tTG IgA limit of quantitation for one reagent lot was determined as 0.81 FLU, and for the second reagent lot as 1.02 FLU. The final LoQ value is 1.02 FLU, which has been set as the lower limit of the analytical measuring range of the tTG IgA assay.

Analytical Measuring Range (AMR)

Within the Aptiva Celiac Disease IgA Reagent:
DGP IgA:0.72 FLU – 250.00 FLU
tTG IgA:1.02 FLU – 600.00 FLU

Auto-rerun function and reportable results

The Aptiva software has an auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result >250.00 FLU for DGP IgA or a result >600.00 FLU for tTG lgA after performing an additional 10-fold dilution, thereby bringing the measured value within the AMR. The reported result will be calculated by the software factoring the additional dilution. As the highest value that can be measured is 2500.00 FLU or 6000.00 FLU for DGP IgA or tTG IgA, respectively.

High concentration hook effect

To assess hook effect, 4 samples for DGP IgA and 7 samples for tTG IgA were tested at three increasing 2fold serial dilutions from the standard 1:46 dilution used by the Aptiva Celiac Disease IgA Reagent. All FLU values above the analytical measuring ranges of the two assays are theoretical and were mathematically calculated using the 4 parameters of their respective calibration curves. All samples showed increase in FLU values as dilution factor became more concentrated, thereby confirming that high positive specimens above the AMR do not show hook effect up to 1229.19 FLU for the DGP IgA assay and 1746.20 FLU for the tTG IgA assay (theoretical values calculated) in the Aptiva Celiac Disease IgA Reagent.

Linearity

The Linearity of the AMR was calculated separately for DGP IgA as part of the Aptiva Celiac Disease IgA Reagent.

The linearity of the AMR of DGP IgA and tTG IgA was evaluated by a study according to CLSI EPO6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. Four human serum samples for each assay with various antibody concentrations were serially diluted to obtain values that cover the entire AMR. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.

{16}------------------------------------------------

Acceptance criteria:

  • Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 15% or ±0.75 FLU for low level samples (allowable nonlinearity).

For DGP IgA, the best fitting polynomial found for samples 1 and 4 was a linear one, while third order polynomial was found for sample 2 and second order polynomial for sample 3. All acceptance criteria were fulfilled.

SampleTest Range(FLU)Slope(95% CI)Y-Intercept(95% CI)Average % Recovery
159.39 - 296.960.98(0.94 to 1.02)10.65(2.20 to 19.10)0.99105.9%
219.28 - 192.820.98(0.96 to 1.00)2.72(0.44 to 5.00)1.00101.5%
33.27 - 29.021.00(0.96 to 1.04)0.61(-0.12 to 1.34)0.99104.1%
40.48 - 4.790.94(0.88 to 0.99)0.11(-0.05 to 0.26)0.99100.0%
Combined0.48 - 296.961.02(1.01 to 1.03)0.61(-0.69 to 1.92)1.00102.9%

For tTG IgA, the best fitting polynomial found for all samples was a linear one. All acceptance criteria were fulfilled.

SampleTest Range(FLU)Slope(95% CI)Y-Intercept(95% CI)Average % Recovery
169.43 - 694.331.04 (1.00 to 1.08)-17.18(-34.97 to 0.60)0.9996.7%
210.28 - 102.791.01 (0.97 to 1.04)1.12(-1.06 to 3.30)1.00102.9%
31.98 - 19.800.94 (0.87 to 1.01)-0.58(-1.38 to 0.23)0.9887.2%
40.78 - 7.761.06 (0.98 to 1.14)0.27(-0.10 to 0.65)0.98113.2%
Combined0.78 - 694.331.01 (1.00 to 1.02)-1.13(-3.58 to 1.32)1.00100.0%

These data demonstrate the linearity of the analytical measuring range (0.72 FLU – 250.00 FLU) of the DGP IgA assay and the analytical measuring range (1.02 FLU - 600.00 FLU) of the tTG IgA assay, both as part of the Aptiva Celiac Disease IgA Reagent.

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. A set of three human serum specimens, one positive, one near the cutoff and one negative sample were tested using interfering substances (bilirubin, hemoglobin, triglycerides, cholesterol, rheumatoid factor IgM and human IgG). All interferents were spiked into every serum specimen and the resulting samples were assessed in triplicates with the Aptiva Celiac Disease IgA assays. Recovery of the unit values was calculated compared to control samples.

{17}------------------------------------------------

Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 15% of the cut-off (±0.75 FLU) difference, whichever is greater.

No interference was detected for DGP or tTG IgA with bilirubin up to 1 mg/mL (recovery: from 96.4%-101.5% and from 96.2% to 99.7% for DGP IgA and tTG IgA respectively), hemoglobin up to 2 mg/mL (recovery: from 102.7% to 108.4% and from 100.4% to 111.8% for DGP IgA and tTG IgA respectively), triglycerides up to 1000 mg/dL (recovery: from 102.6% to 107.7% and from 98.2% to 110.3% for DGP IgA and tTG IgA respectively), cholesterol up to 332.5 mg/dL (recovery: from 102.4% to 103.9% or -0.43 FLU and from 99.6% to 109.5% or -0.44 FLU for DGP IgA and tTG IgA respectively), RF IgM up to 250 IU/mL (recovery: 93.1% to 99.2% or 0.30 FLU and 98.8% to 102.6% or 0.35 FLU for DGP IgA and TTG IgA, respectively) and human IgG up to 70 mg/mL (recovery: from 99.8% to 102.3% or -0.36 FLU and from 99.5 to 102.4% or 0.32 FLU for DGP IgA and tTG IgA, respectively).

Sample Stability and Handling

For the DGP IgA assay, eight test samples were tested and for tTG IgA assay six test samples were tested. The samples used for this study were achieved by combining high and low antibody level to yield their desired reactivity. Test samples covered the analytical measuring ranges of each analyte. All samples were tested in duplicates for up to 21 days while stored at 2-8℃, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 5 cycles. Results were compared to those obtained on control samples (time zero / zero cycles).

Acceptance criteria: percent recovery is between 85-115% for positive samples, and between 80-120% for negative samples (<5.00 FLU).

All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples may be stored up to 48 hours at room temperature, up to 14 days at 2-8°C and can be subjected to up to 5 freeze/thaw cycles.

Reagent Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 5 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components to establish initial stability claim:

  • Aptiva Celiac Disease IgA microparticle - 3 lots
  • . PE Tracer IgA – 3 lots
  • . Rehydration Buffer – 3 lots

Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3℃ stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to

{18}------------------------------------------------

those stored at 37 ± 3℃ (test) for 1, 2, 3, 4, and 5 weeks, where one week is equal to six months at 5 ± 3°C. Linear regression analysis was performed between recovery values and the number of days. For each component tested, linear regression analysis was performed separately on each bead in the Aptiva Celiac Disease IgA Reagent (DGP IgA, tTG IgA, IgA Control Bead).

Acceptance criteria for two-year preliminary expiration dating: With regression analysis, the lower and upper 95% Cl interval of the regression line is between 80% and 120% recovery at day 28 (week 4).

All components tested fulfilled the acceptance criteria above, therefore, two-year expiration dating was assigned to each component.

In-use (onboard) stability

Reagent Cartridge

To establish the in-use stability of the Aptiva Celiac Disease IgA reagent cartridges, one lot of reagent cartridge was tested using up to 14 human serum samples (with different reactivity levels). The specimens were tested periodically for 45 days. At day 21 the reagent cartridge was recalibrated, and a cartridge specific Working Curve was generated. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting percent recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first): - The stability claim is established at the actual measurement day proceeding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when ≥2% of the recovery data, (3 data points) is <75% or ≥125% recovery.

The onboard stability results for the Aptiva Celiac Disease IgA is as follows: Lot 100014: 45 days

Using these criteria, the in-use (onboard) stability of the Aptiva Celiac Disease IgA reagent cartridge was set at 42 days.

Real time stability

Real time stability testing has been scheduled to be performed every three or six months on the Aptiva Celiac Disease IgA Reagents kit, to verify the two-year expiration that was assigned based on accelerated stability studies. Results for the first time point at 6 months will be available in January 2020.

A negative sample, a low positive sample, and a high positive sample will be tested at each time point. - Acceptance criteria: results should fall within their respective ranges.

Cut-off, reference range

The following cut-off is used for both the DGP IgA and tTG IgA assays in the Aptiva Celiac Disease IgA Reagent:

Negative<5.00 FLU
Positive≥5.00 FLU

{19}------------------------------------------------

The reference population for establishing the reference interval for the DGP IgA and tTG assays, within the Aptiva Celiac Disease IgA Reagent, consisted of 200 subjects:

Sample GroupN
Crohn's Disease15
Autoimmune Thyroid Disease30
Infectious Disease30
Primary Biliary Cholangitis30
Rheumatoid Arthritis40
Systemic Lupus Erythematosus20
Systemic Sclerosis20
Ulcerative Colitis15

Additionally, 12 diagnosed celiac disease (CD) patient specimens were assayed to aid in the determination of the cutoff values.

All specimens were the same matrix (human serum) as specified in the Intended Use. All specimens were unaltered. The cut-off values were established in accordance to CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Shapiro-Wilk p<0.0001), therefore the non-parametric percentile method was used.

The cut-off was established based on greater than the 95th percentile of the results on the reference subjects, along with the results of 12 samples from patients with celiac disease.

A cutoff of 5.00 FLU (78 MFI and 190 MFI for DGP IgA and tTG IgA, respectively) has been set to ensure optimal differentiation between negatives and positives samples.

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the Aptiva Celiac Disease IgA Reagent. A total of 495 characterized samples were included in this Validation Set, including 171 samples from celiac disease patients, 34 dermatitis herpetiformis and 290 control samples from patients with various types of autoimmune and infectious diseases. All samples were run on the Aptiva Celiac Disease IgA Reagent. The distribution of the cohort and the DGP and tTG positivity rate is in the Table below:

{20}------------------------------------------------

Page 18 of 19

Patient GroupNDGP IgAN PositiveDGP IgA% PositivetTG IgAN PositivetTG IgA% Positive
Rheumatoid Arthritis6911.4%22.9%
Ulcerative Colitis3100.0%00.0%
Crohn's Disease3100.0%00.0%
Hepatitis C Virus2800.0%00.0%
Hepatitis B Virus2500.0%00.0%
Syphilis2100.0%00.0%
Sjögren's Syndrome2000.0%00.0%
Systemic Sclerosis1900.0%00.0%
Autoimmune Gastritis1500.0%00.0%
Human Immunodeficiency Virus1300.0%00.0%
Systemic Lupus Erythematosus1200.0%00.0%
Epstein-Barr Virus6116.7%00.0%
Total Controls29020.7%20.7%
Celiac Disease17110159.1%15993.0%
Dermatitis Herpetiformis342264.7%3191.2%
Total495----

Clinical sensitivity and specificity for the Aptiva DGP IgA were analyzed in the table below:

Clinical Analysis (N=461)DiagnosisAnalysis
CDControlsTotal(95% confidence)
Aptiva DGP IgAPositive1012103Sensitivity: 59.1% (51.6 – 66.2%)
Negative70288358Specificity: 99.3% (97.5 – 99.8%)
Total171290461

Clinical sensitivity and specificity for the Aptiva tTG IgA were analyzed in the table below:

Clinical Analysis (N=461)DiagnosisAnalysis
CDControlsTotal(95% confidence)
Aptiva tTG IgAPositive1592161Sensitivity: 93.0% (88.1 – 95.9%)
Negative12288300Specificity: 99.3% (97.5 – 99.8%)
Total171290461

{21}------------------------------------------------

In addition to the clinical validation cohort, thirty-four samples from dermatitis herpetiformis patients were tested for the Aptiva Celiac Disease IgA assay. The sensitivity of DGP IgA in dermatitis herpetiformis patients is 64.7% (47.9 – 78.5%) and the specificity of DGP IgA is 99.3% (97.5-99.8%). The sensitivity of tTG lgA in dermatitis herpetiformis patients is 91.2% (77.0 – 97.0%) and the specificity of tTG IgA is 99.3% (97.5-99.8%).

Expected values

The expected value in the normal population is "negative". A panel of 120 apparently healthy blood donors (70 females/50 males, ages 17 to 57 years, with an average and median age of 32 and 31 years respectively) were tested on the Aptiva Celiac Disease IgA Reagent. For DGP IgA, with a cut-off of 5.00 FLU, no samples were positive, with a mean concentration of 1.08 FLU, and values ranging from 0.22 to 4.86 FLU. For tTG IgA, with a cut-off of 5.00 FLU, one sample (0.8%) was positive, with a mean concentration of 0.66 FLU, and values ranging from 0.06 to 11.88 FLU.

Comparison with predicate device

Samples for method comparison analysis included all samples (n=495) from the clinical validation study. These samples were tested on both the Aptiva Celiac Disease IgA Reagent and on their predicate QUANTA Flash DGP IgA and tTG assays.

Method Comparison (N=200)QUANTA Flash DGP IgAPercent Agreement(95% Confidence)
NegativePositiveTotal
Aptiva DGP IgANegative63265NPA: 96.9% (89.5–99.2%)
Positive20115135PPA: 85.2% (78.2 – 90.2%)
Total83117200TPA: 89.0% (83.9 – 92.6%)

Method comparison of the Aptiva DGP IgA with the predicate device. Samples within AMR.

NPA: Negative Percent Agreement; PPA: Positive Percent Agreement; TPA: Total Percent Agreement

Method Comparison (N=197)QUANTA Flash tTG IgAPercent Agreement
NegativePositiveTotal(95% Confidence)
Aptiva tTG IgANegative31132NPA: 96.9% (84.3–99.4%)
Positive1163165PPA: 98.8% (95.7 – 99.7%)
Total32164197TPA: 98.5% (95.6 – 99.5%)

Method comparison of the Aptiva tTG IgA with the predicate device. Samples within AMR.

NPA: Negative Percent Agreement; PPA: Positive Percent Agreement; TPA: Total Percent Agreement

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).