K Number
K193571

Validate with FDA (Live)

Date Cleared
2020-04-08

(107 days)

Product Code
Regulation Number
870.1330
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merit GO2WIRE™ is intended to facilitate the placement and exchange of devices during peripheral diagnostic and interventional procedures.

Device Description

The Merit Go2Wire guide wire will consist of 0.035" guide wire configurations available in 145, 175, 210, 260, and 300cm. The distal tip flex configurations are Floppy, Standard and Intermediate (Modified J). The distal tip shapes are either straight or modified J.

The wire is composed of a stainless-steel core wire, a PTFE coated coil, a white PTFE sleeve, a platinum-tungsten marker coil, and an optional extension system connector. The wire's proximal end is covered by a white PTFE sleeve that terminates at 100cm from the distal tip. The distal 100 cm of the wire is covered with a green PTFE pre-coated coil. The PTFE coated coil and platinum-tungsten coil are welded to the core on the very distal tip. The distal tip is shapeable and radiopaque. A torque device is included to facilitate wire steering within the vascular anatomy.

An 0.035" extension wire consisting of PTFE coated stainless steel 155cm in length to extend the length of the wire to allow initial usage with a short wire and then extendable to allow catheter exchanges when required.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the "Go2Wire Guide Wire." It is not an AI/ML device, therefore, the information requested about acceptance criteria and study proving device meets it (including multi-reader multi-case studies, standalone algorithm performance, and training/test set details for AI/ML models) is not applicable or present in this document.

The document discusses the substantial equivalence of the Go2Wire Guide Wire to a predicate device based on various performance and biocompatibility tests.

Here's the relevant information that can be extracted, addressing the spirit of the request as much as possible for a non-AI/ML device:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table with numerical acceptance criteria and reported numerical performance values. Instead, it states:

Acceptance Criterion CategoryReported Device Performance
Performance Testing-Bench"All test results were comparable to the predicate devices and the subject Merit Go2Wire Guide Wire met the predeterminded acceptance criteria."
Specifically, the device was tested for: Size Designation, Radiodetectability, Surface, Tensile Strength, Torque Strength, Torqueability, Tip Flexibility, Fracture test, Flex test, Torturous Path Lubricity, Corrosion Resistance, Tip Shape Testing Validation, Catheter and Needle Compatibility Validation, Guidewire Clinical Use in Model Validation, Extension System Clinical Use in Model Validation, Particulate.
Biocompatibility"All test results were comparable to the predicate devices and the subject Merit Go2Wire Guide Wire met the predeterminded acceptance criteria."
Specifically, the device was tested for: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Genotoxicity, Hemolysis, Thrombogenicity, Complement Activation.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for each test or the provenance of the data. The tests are described as "Performance Testing-Bench" and "Biocompatibility," indicating laboratory testing of the physical device rather than data from clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is a physical medical device (guide wire), not an AI/ML system requiring expert interpretation for ground truth. The "ground truth" for this device would be established by validated test methods and passing results according to engineering and biocompatibility standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an AI/ML study involving human readers or interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for this medical device's performance is established by:

  • Reference to industry standards and FDA guidance documents for medical device testing (e.g., ISO 11070:2014, FDA Guidance for Guidewires, ISO 10993-1:2018 for biocompatibility).
  • Predetermined acceptance criteria derived from these standards and comparison to a legally marketed predicate device (Wholey Guide Wire System [K120863]).
  • Bench testing and biocompatibility testing results.

8. The sample size for the training set

Not applicable. This device does not involve a "training set" in the context of AI/ML.

9. How the ground truth for the training set was established

Not applicable. This device does not involve a "training set" in the context of AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 8, 2020

Merit Medical System, Inc. Michael O'Sullivan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland

Re: K193571

Trade/Device Name: Go2Wire Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 5, 2020 Received: March 9, 2020

Dear Michael O'Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K193571

Device Name Go2Wire Guide Wire

Indications for Use (Describe)

The Merit GO2WIRE™ is intended to facilitate the placement and exchange of devices during peripheral diagnostic and interventional procedures.

Carefully read all instructions prior to use. Observe all warnings and cautions noted throughout these instructions. Failure to do so may result in complications.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K193571

Merit Medical Systems, Inc.
Merit Go2Wire Guide Wire
Traditional Premarket Notification 510(k)

Submitter Name:Merit Medical Systems, Inc.
Address:1600 West Merit ParkwaySouth Jordan, UT 84095
Telephone Number:(+353) 91 703700 (Ext. 3061)
Fax Number:(+353) 91 680104
Contact Person:Mark Mullaney
Registration Number:1721504
GeneralProvisionsCorrespondent Name:Merit Medical Ireland Ltd.
Address:Parkmore Business ParkParkmore, Galway, Ireland
Telephone Number:(+353) 91 703700 (Ext. 3223)
Fax Number:(+353) 91 680104
Contact Person:Michael O'Sullivan
Date of Preparation:20 December 2019
Registration Number:9616662
SubjectDeviceTrade Name:
Common/Usual Name:Guide Wire
Classification Name:Wire, Guide, Catheter
PredicateDeviceTrade Name:Wholey Guide Wire System
Classification Name:21 CFR 870.1330 Catheter Guide Wire
Premarket Notification:K120863
Manufacturer:Covidien LLC.
ClassificationClass II21 CFR § 870.1330Product code: DQXDivision of Cardiovascular Devices
Intended UseThe Merit GO2WIRE™ is intended to facilitate the placement andexchange of devices during peripheral diagnostic andinterventional procedures.Carefully read all instructions prior to use. Observe all warningsand cautions noted throughout these instructions. Failure to do somay result in complications.
The Merit Go2Wire guide wire will consist of 0.035" guide wireconfigurations available in 145, 175, 210, 260, and 300cm. Thedistal tip flex configurations are Floppy, Standard and Intermediate(Modified J). The distal tip shapes are either straight or modified J.
DeviceDescriptionThe wire is composed of a stainless-steel core wire, a PTFE coatedcoil, a white PTFE sleeve, a platinum-tungsten marker coil, and anoptional extension system connector. The wire's proximal end iscovered by a white PTFE sleeve that terminates at 100cm from thedistal tip. The distal 100 cm of the wire is covered with a green PTFEpre-coated coil. The PTFE coated coil and platinum-tungsten coilare welded to the core on the very distal tip. The distal tip isshapeable and radiopaque. A torque device is included to facilitatewire steering within the vascular anatomy.
An 0.035" extension wire consisting of PTFE coated stainless steel155cm in length to extend the length of the wire to allow initialusage with a short wire and then extendable to allow catheterexchanges when required.
Comparisonto PredicateThe Technological characteristics of the subject Merit Go2WireGuide Wire are substantially equivalent to those of Predicate Device,the Wholey Guide Wire System[K120863]. The Wholey Guide wirewas fully characterized by Merit and an equivalent wire wasdeveloped by Merit based upon this characterization. The subjectdevice has the same basic design as the predicate, in that they arecomposed of a stainless-steel core wire, a PTFE coated coil, a whitePTFE sleeve, a platinum-tungsten marker coil, and an optionalextension system connector.
The fundamental technology and operating principles of the subjectand the predicates are the same and while the indications for usewordings are not identical, the intended use is the same, with allwires used for the placement and exchange of devices duringperipheral diagnostic and interventional procedures.
Safety &PerformanceTestsNo performance standards have been established under section514 of the Food, Drug and Cosmetic Act for these devices. A batteryof testing was conducted, on the Merit Go2Wire Guide Wire, inaccordance with protocols based on requirements outlined inguidance's and industry standards and these were shown to meetthe acceptance criteria that were determined to demonstratesubstantial equivalence.

{4}------------------------------------------------

{5}------------------------------------------------

Where appropriate, the tests were based on the requirements of the following documents:

  • FDA Guidance "Coronary, Peripheral, and Neurovascular ● Guidewires -Performance Tests and Recommended Labelling- Guidance for Industry and Food and Drug Administration Staff" - October 2019
  • FDA Guidance "Intravascular Catheters. Wires, and ● Delivery Systems with Lubricious Coatings - Labeling Considerations" - October 2019
  • FDA guidance Coronary and Cerebrovascular Guide Wire ● Guidance January 1995.
  • ISO 11070:2014, Sterile Single-Use Intravascular Catheter ● Introducers.
  • ISO 11135:2014 Sterilization of health care products-● Ethylene oxide-: Requirements for the development, validation and routine control of a sterilization process for medical devices.
  • ASTM F1980-07 Standard Guide for Accelerated Aging of . Sterile Barrier Systems for Medical Devices
  • ISO 10993-1:2018, Biological Evaluation of Medical Devices ● Part 1: Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.

The following is a list of all testing that was successfully completed:

Performance Testing-Bench

  • Size Designation ●
  • Radiodetectability
  • Surface ●
  • Tensile Strength ●
  • Torque Strength .
  • Torqueability .
  • Tip Flexibility ●
  • Fracture test ●
  • Flex test ●
  • Torturous Path Lubricity ●
  • Corrosion Resistance
  • Tip Shape Testing Validation .
  • Catheter and Needle Compatibility Validation 0
  • Guidewire Clinical Use in Model Validation .

{6}------------------------------------------------

  • Extension System Clinical Use in Model Validation ●
  • Particulate .

Biocompatibility

  • Cytotoxicity .
  • Sensitization .
  • Irritation .
  • Acute Systemic Toxicity .
  • Pyrogenicity .
  • Genotoxicity ●
  • Hemolysis .
  • Thrombogenicity .
  • Complement Activation .

All test results were comparable to the predicate devices and the subject Merit Go2Wire Guide Wire met the predeterminded acceptance criteria. This has demonstrated that the subject device is substantially equivalent to the predicate devices.

Based on the comparisons noted, the subject Merit Go2Wire Guide Summary of Wire meets the requirements that are considered essential for its Substantial intended use and is substantively equivalent to the Predicate Equivalence Device, the Wholey Guide Wire System [K120863] manufactured by Covidien LLC

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.