K Number
K193525

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-06-26

(190 days)

Product Code
Regulation Number
866.5510
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Yumizen C1200 Immunoglobulin A reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin A (IgA) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings,

Yumizen C1200 Immunoglobulin G reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin G (IgG) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings.

Yumizen C1200 Immunoglobulin M reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin M (IgM) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers. Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

This test should be used in conjunction with other laboratory and clinical findings.

Device Description

This submission consists in the Yumizen C1200 Immunoglobulin A (1300023881), Yumizen C1200 Immunoglobulin G (1300023883) and Yumizen C1200 Immunoglobulin M (1300023884) reagent for serum and plasma testing for Yumizen C1200 reagent.

The Yumizen C1200 Level 1 Protein Control (1300023944) and Yumizen C1200 Level 2 Protein Control (1300023945) for use on Yumizen C1200 Analyzer and the Yumizen C1200 Protein Cal (1300023893) for use on Yumizen C1200 Analyzer are sold separately.

AI/ML Overview

This document describes the analytical performance characteristics of the Yumizen C1200 Immunoglobulin A, G, and M reagents, intended for quantitative in vitro diagnostic determination of immunoglobulins in serum and lithium heparin plasma. The study aims to demonstrate that the device meets acceptance criteria for various analytical parameters, ensuring its safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance CriteriaYumizen C1200 Immunoglobulin A PerformanceYumizen C1200 Immunoglobulin G PerformanceYumizen C1200 Immunoglobulin M Performance
Measuring RangeAppropriateness supported by LOD, LOQ, and linearity studies.Serum: 0.10 to 7.00 g/L(Post-dilution): 7.00 to 21.00 g/L(Linearity Range): 0.21 - 6.60 g/L (Slope 1.027, R^2 0.9975)Serum: 0.75 to 30.00 g/L(Post-dilution): 30.00 to 90.00 g/L(Linearity Range): 0.82 – 29.42 g/L (Slope 0.9965, R^2 0.9986)Serum: 0.20 to 5.00 g/L(Post-dilution): 5.00 to 15.00 g/L(Linearity Range): 0.26 - 4.16 g/L (Slope 1.013, R^2 0.9994)
PrecisionWithin-Run CV: Low (≤4.5%), Middle (≤3.8%), High (≤3%)Total Precision CV: Low (≤6.0%), Middle (≤5.0%), High (≤4.0%)Total Precision (Analyzer Variability): Control Level 1: 3.7%Control Level 2: 3.1%Samples: 1.7%-3.8% Lot-to-Lot Variability: Control Level 1: 1.2%Control Level 2: 1.1%Samples: 1.2%-4.1%Total Precision (Analyzer Variability): Control Level 1: 2.9%Control Level 2: 3.3%Samples: 1.8%-3.0% Lot-to-Lot Variability: Control Level 1: 1.7%Control Level 2: 1.9%Samples: 1.4%-2.3%Total Precision (Analyzer Variability): Control Level 1: 2.1%Control Level 2: 1.8%Samples: 1.3%-2.4% Lot-to-Lot Variability: Control Level 1: 1.6%Control Level 2: 1.7%Samples: 1.0%-2.7%
InterferencesAcceptable bias: +/-10% of value without interfering substancesReported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen).Reported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen).Reported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen).
Matrix ComparisonNo significant difference between serum and heparinized plasma specimens (implied by correlation and slope close to 1).IgA: N=62 (paired serum/heparin plasma)Slope: 1.000Correlation: 0.999IgG: N=43 (paired serum/heparin plasma)Slope: 0.9929Correlation: 0.988IgM: N=43 (paired serum/heparin plasma)Slope: 1.000Correlation: 0.999
Method ComparisonDemonstrated substantial equivalence through correlation with predicate device.IgA: N=190 (native serum samples)Slope: 0.9941Correlation: 0.993IgG: N=214 (native serum samples)Slope: 1.016Correlation: 0.993IgM: N=153 (native serum samples)Slope: 1.005Correlation: 0.993
Reagent StabilityShelf life and on-board stability for opened reagents.Closed: 24 months at 2-8°COn-Board (Opened): 6 weeksClosed: 24 months at 2-8°COn-Board (Opened): 6 weeksClosed: 24 months at 2-8°COn-Board (Opened): 6 weeks
Reference RangeVerification studies support established ranges through literature.0.70 - 4.00 g/L (70 - 400 mg/dL)7 – 16 g/L (700 - 1600 mg/dL)0.40 - 2.30 g/L (40 - 230 mg/dL)

2. Sample Sizes Used for the Test Set and Data Provenance

  • Measuring Range (Linearity):
    • IgA, IgG, IgM: Samples were "spiked" to create different concentrations, then serially diluted. The exact number of initial samples for spiking is not specified, but dilutions were "assayed in quadruplicate within a single run."
  • Precision (Total Precision: analyzer variability - 20x2x2 study):
    • IgA, IgG, IgM: 5 human sera samples and 2 levels of Yumizen C1200 Protein Control. Tested with "two replicates per run, two runs per day for 20 days on each of three analyzers" (n=240 per sample).
  • Precision (Lot to Lot variability study: 3x5x2x3):
    • IgA, IgG, IgM: 5 human sera samples and 2 levels of Yumizen C1200 Protein Control. Tested in "triplicates per run, two runs per day for five days on each of three lots" (n=90 per sample).
  • Interferences: The exact sample size is not stated, but the study implies testing samples with varying concentrations of interfering substances to determine acceptable bias.
  • Matrix Comparison:
    • IgA: 62 paired samples (serum and heparinized plasma) from single donors.
    • IgG: 45 paired samples (serum and heparinized plasma) from single donors. Of these, 43 were used for correlation analysis.
    • IgM: 43 paired samples (serum and heparinized plasma) from single donors.
  • Method Comparison:
    • IgA: 190 "native samples" from human serum.
    • IgG: 214 "native samples" from human serum.
    • IgM: 153 "native samples" from human serum.

Data Provenance: The human serum samples used for precision, matrix comparison, and method comparison studies were "anonymous remnants of human serum specimens collected from blood bank." "Spiked" samples were used for linearity studies, and "normal samples" from a blood bank were used for reference range verification. The document does not explicitly state the country of origin, but the manufacturer is based in France. The studies appear to be prospective analytical performance evaluations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

No external experts are mentioned for establishing ground truth in these analytical performance studies. The studies rely on established CLSI guidelines for evaluation and comparison with a legally marketed predicate device. The ground truth for quantitative measurements is the direct measurement by the devices themselves and comparison against predicate devices or known spiked concentrations.

4. Adjudication Method for the Test Set

Not applicable. These are analytical performance studies for an in vitro diagnostic device, not studies involving human interpretation or adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This document describes the analytical performance of an in vitro diagnostic reagent and analyzer system, not a device requiring human interpretation for diagnostic purposes where MRMC studies would typically be conducted.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, this can be considered a standalone performance study. The Yumizen C1200 system (reagents and analyzer) performs the quantitative determination of immunoglobulins. The studies evaluate the analytical capabilities of the system itself, such as accuracy (via method comparison, linearity), precision, interference, and stability, without direct human intervention in the result generation or diagnostic interpretation loop. The intent is to demonstrate the device's ability to accurately measure predefined analytes.

7. Type of Ground Truth Used

The ground truth for these analytical performance studies is established through:

  • Known concentrations: For linearity studies, spiked samples with known concentrations were used.
  • Predicate device measurements: For method comparison, results from the candidate device were compared against measurements obtained from legally marketed predicate devices (Beckman Coulter's Olympus IgA, IgG, IgM reagents on AU analyzers).
  • CLSI guidelines and established methodologies: The studies adhere to CLSI guidelines (e.g., EP05-A3 for precision, EP17-A2 for detection capability, EP06-A for linearity, EP25-A for stability, C28-A3 for reference intervals, EP-9A3 for method comparison) which define accepted methods for evaluating analytical performance and establishing performance characteristics.
  • Literature-established reference ranges: For reference range verification, the device's measurements on "normal" samples were compared against ranges cited in scientific literature (e.g., Dati et al., 1996).

8. Sample Size for the Training Set

The document does not explicitly delineate a "training set" in the context of machine learning or AI models. This device is an in vitro diagnostic reagent and analyzer system, not an AI/ML-based diagnostic algorithm that typically undergoes a distinct training phase with a dedicated dataset. The performance studies described here are for analytical validation rather than algorithm training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a "training set" in the context of an AI/ML model for this in vitro diagnostic device. The analytical evaluations described involve testing the reagent and instrument system, not training a learning algorithm.

{0}------------------------------------------------

June 26, 2020

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".

Horiba ABX SAS Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine, Rue du Caducee BP7290 Montpellier Cedex 4, 341184 France

Re: K193525

Trade/Device Name: Yumizen C1200 Immunoglobulin A, Yumizen C1200 Immunoglobulin G, Yumizen C1200 Immunoglobulin M Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A. G. M. D. and E immunological test system Regulatory Class: Class II Product Code: CZP, DEW, CFN Dated: December 17, 2019 Received: December 19, 2019

Dear Caroline Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying (Katelin) Mao, Ph.D Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K193525

Device Name

Yumizen C1200 Immunoglobulin A Yumizen C1200 Immunoglobuline G Yumizen C1200 Immunoglobulin M

Indications for Use (Describe)

Yumizen C1200 Immunoglobulin A reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin A (IgA) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings,

Yumizen C1200 Immunoglobulin G reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin G (IgG) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings.

Yumizen C1200 Immunoglobulin M reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin M (IgM) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers. Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

This test should be used in conjunction with other laboratory and clinical findings.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the HORIBA Medical logo. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter letters, is the word "Medical". There is also a number string that reads "007-510(d)".

Section 007. 510(k) summary

SECTION 007: 510(k) Summary of K193525

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the HORIBA Medical logo in blue. Below the logo is the text "Section 007. 510(k) summary" in a gray font. The text appears to be a section heading or title, possibly from a document or report.

  • 1- Date of Summary Date submitted : 26th June, 2020

2- Company

HORIBA ABX SAS HORIBA MEDICAL Parc Euromédecine Rue du Caducée – BP 7290 34184 Montpellier cedex 4 France

3- Contact person

Contact Person: Caroline Ferrer (caroline.ferrer@horiba.com) Telephone: + (33) 4 67 14 1843 Fax: + (33) 4 67 14 1517

4- Product Name

Yumizen C1200 Immunoglobulin A (1300023881) Yumizen C1200 Immunoglobulin G (1300023883) Yumizen C1200 Immunoglobulin M (1300023884)

5- Device Name and Classification

● Intended use

The devices involved by the 510(k) submission file are the following:

Device's namesIntended Use
Yumizen C1200Immunoglobulin AYumizen C1200 Immunoglobulin A reagent is intended for thequantitative in vitro diagnostic determination of Immunoglobulin A (IgA)in serum and lithium heparin plasma by immunoturbidimetry on Yumizenanalyzers. Measurement of this immunoglobulin aids in the diagnosis ofabnormal protein metabolism and the body's lack of ability to resistinfectious agents. This test should be used in conjunction with otherlaboratory and clinical findings.
Yumizen C1200Immunoglobulin GYumizen C1200 Immunoglobulin G reagent is intended for thequantitative in vitro diagnostic determination of Immunoglobulin G (IgG)in serum and lithium heparin plasma by immunoturbidimetry on Yumizenanalyzers. Measurement of this immunoglobulin aids in the diagnosis ofabnormal protein metabolism and the body's lack of ability to resistinfectious agents. This test should be used in conjunction with otherlaboratory and clinical findings.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

Section 007. 510(k) summary

Yumizen C1200Immunoglobulin MYumizen C1200 Immunoglobulin M reagent is intended for thequantitative in vitro diagnostic determination of Immunoglobulin M (IgM)in serum and lithium heparin plasma by immunoturbidimetry on Yumizenanalyzers. Measurement of this immunoglobulin aids in the diagnosis ofabnormal protein metabolism and the body's lack of ability to resistinfectious agents. This test should be used in conjunction with otherlaboratory and clinical findings.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Classification and Description ●

Trade/Proprietary Name:Yumizen C1200 Immunoglobulin A
Device Class:Class II / 510(k) required
Classification Name:§866.5510: Immunoglobulins A, G, M, D, and E immunological test systems
Product Code:CZP
Panel:Immunology (82)
Trade/Proprietary Name:Yumizen C1200 Immunoglobulin G
Device Class:Class II / 510(k) required
Classification Name:§866.5510: Immunoglobulins A, G, M, D, and E immunological test
Product Code:DEW
Panel:Immunology (82)
Trade/Proprietary Name:Yumizen C1200 Immunoglobulin M
Device Class:Class II / 510(k) required
Classification Name:§866.5510: Immunoglobulins A, G, M, D, and E immunological test system
Product Code:CFN
Panel:Immunology (82)

6- Substantial Equivalence Information

The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.

Predicate Device Name and 510(k) number:

Candidate devicePredicate devicePredicate ManufacturerPredicate510(k)number
Yumizen C1200Immunoglobulin AOlympus IgA reagent(OSR6X171)BECKMAN COULTERK073489
Yumizen C1200Immunoglobulin GOlympus IgG reagent(OSR6X172)BECKMAN COULTERK073490
Yumizen C1200Immunoglobulin MOlympus IgM reagent(OSR6X173)BECKMAN COULTERK073487

The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.

a. Yumizen C1200 Immunoglobulin A

i. Comparison with predicate Device : Similarities

ItemPredicate K073489Candidate
Device NameOSR6X171Yumizen C1200 Immunoglobulin A(1300023881)
Intended UseSystem reagent for the quantitativedetermination of IgA immunoglobulins inhuman serum and plasma on BeckmanCoulter AU analyzers.The spectrum of abnormalities in serumimmunoglobulin concentrations is broad.Abnormal concentrations range from avirtual absence of one or more of the threemajor classes of immunoglobulin (IgG,IgA, and IgM) to polyclonal increases inone or more immunoglobulins.Measurement of these immunoglobulinsaids in the diagnosis of abnormal proteinmetabolism and the body's lack of abilityto resist infectious agents.For in vitro diagnostic use.Yumizen C1200 Immunoglobulin Areagent is intended for the quantitative invitro diagnostic determination ofImmunoglobulin A (IgA) in serum andlithium heparin plasma byimmunoturbidimetry on Yumizenanalyzers. Measurement of thisimmunoglobulin aids in the diagnosis ofabnormal protein metabolism and thebody's lack of ability to resist infectiousagents. This test should be used inconjunction with other laboratory andclinical findings.
Reagent formatLiquidSame

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below "HORIBA" is the word "Medical" in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

Section 007. 510(k) summary

ItemPredicate K073489Candidate
MeasurementQuantitativeSame
Analytical RangeMeasuring Range0.1 - 7.00 g/dL10-700 mg/LMeasuring Range0.10 - 7.00 g/dL10-700 mg/L

ii. Comparison with predicate Device: Differences

ItemPredicate K073489Candidate
Device NameOSR6X171Yumizen C1200 Immunoglobulin A(1300023881)
InstrumentOlympus AU400 Clinical ChemistryAnalyzerYumizen C1200 Clinical chemistryAnalyzer
Manufactured byBECKMAN COULTERHORIBA ABX SAS
MethodTurbidimetryImmunoturbidimetry
Packaging4x14 mL (R1)4x11 mL (R2)6x20 mL (R1)6x20 mL (R2)
Product codeCFNCZP
ReagentOn board Stability60 days6 weeks
Reference range66-433 mg/dL0.66-4.33 g/L70-400 mg/dL0.70-4.00 g/L
Sample StabilityStable up to 3 days when stored 2-8°C< -20°C stability claim was not defined(specified)8 months at 20-25°C8 months at 4-8°C8 months at -20°C
Sample typeSerum,Lithium-heparin plasmaEDTA plasmaSerumLithium-heparin plasma

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for HORIBA Medical in blue font. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be a heading or title for a document or section of a document.

K193525 Yumizen C1200 Reagents

b. Yumizen C1200 Immunoglobulin G

Predicate K073490 Item Candidate Yumizen C1200 Immunoglobulin G Device Name OSR6X172 (1300023883) Yumizen C1200 Immunoglobulin G reagent is intended for the quantitative in System reagent for the quantitative vitro diagnostic determination of determination of IgG immunoglobulins in Immunoglobulin G (IgG) in serum and human serum and plasma on Beckman lithium heparin plasma by Coulter AU analyzers. immunoturbidimetry on Yumizen analyzers. Measurement of this Intended Use The measurement of IgG aids in the immunoglobulin aids in the diagnosis of diagnosis of abnormal protein metabolism abnormal protein metabolism and the and the body's lack of ability to resist body's lack of ability to resist infectious infectious agents. agents. This test should be used in For in vitro diagnostic use. conjunction with other laboratory and clinical findings. Reagent format Liquid Same Measurement Quantitative Same Stable up to expiry date on the label if Stable up to expiry date on the label if stored at 2-8°C.Store protected from Shelf-life stored at 2-8°C. light. Measuring Range Measuring Range Analytical Range 0.75 - 30.00 g/L 0.75 - 30.00 g/L 75-3000 mg/dL 75-3000 mg/dL

i. Comparison with predicate Device : Similarities

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for HORIBA Medical in blue font. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be a section heading or title within a larger document.

ItemPredicate K073490Candidate
Device NameOSR6X172Yumizen C1200 Immunoglobulin G(1300023883)
InstrumentOlympus AU400 Clinical ChemistryAnalyzerYumizen C1200 Clinical chemistryAnalyzer
Manufactured byBECKMAN COULTERHORIBA ABX SAS
Packaging4x22mL (R1)4x20 mL (R2)6x20 mL (R1)6x20 mL (R2)
MethodTurbidimetryImmunoturbidimetry
Product codeCFNDEW
ReagentOn board Stability90 days6 weeks
Reference range635-1741 mg/dL6.35-17.41 g/L700-1600 mg/dL7-16 g/L
Sample Stability2-8°C for up to 3 days1 week at 20-25°C3 months at 4-8°C>6 months at -20°C
Sample typeSerum,Lithium-heparin plasmaEDTASerumLithium-heparin plasma
ItemPredicate K073487Candidate
Device NameOSR6X173Yumizen C1200 Immunoglobulin M(1300023884)
Intended UseSystem reagent for the quantitativedetermination of IgM immunoglobulins inhuman serum and plasma on BeckmanCoulter AU analyzers.System reagent for the quantitativedetermination of IgM immunoglobulins inhuman serum and plasma on BeckmanCoulter AU analyzersThe spectrum of abnormalities in serumimmunoglobulin concentrations is broad.Abnormal concentrations range from avirtual absence of one or more of the threemajor classes of immunoglobulin (IgA,IgG, and IgM) to polyclonal increases inone or more immunoglobulins.Measurement of these immunoglobulinsaids in the diagnosis of abnormal proteinmetabolism and the body's lack of abilityto resist infectious agents.For in vitro diagnostic use.Yumizen C1200 Immunoglobulin Mreagent is intended for the quantitative invitro diagnostic determination ofImmunoglobulin M (IgM) in serum andlithium heparin plasma byimmunoturbidimetry on Yumizenanalyzers. Measurement of thisimmunoglobulin aids in the diagnosis ofabnormal protein metabolism and thebody's lack of ability to resist infectiousagents.This test should be used in conjunctionwith other laboratory and clinicalfindings.
Reagent formatLiquidSame
MeasurementQuantitativeSame
ProcodeCFNSame
Shelf-lifeStable up to expiry date on the label ifstored at 2-8°C.Stable up to expiry date on the label ifstored at 2-8°C.Store protected fromlight.
Analytical RangeMeasuring Range0.20-5.00 g/L20-500 mg/dLMeasuring Range0.20-5.00 g/L20-500 mg/dL
ItemPredicate K073487Candidate
Device NameOSR6X173Yumizen C1200 Immunoglobulin M(1300023884)
InstrumentOlympus AU400 Clinical ChemistryAnalyzerYumizen C1200 Clinical chemistryAnalyzer
Manufactured byBECKMAN COULTERHORIBA ABX SAS
Packaging4x14mL (R1)4x11 mL (R2)6x20 mL (R1)6x20 mL (R2)
MethodTurbidimetryImmunoturbidimetry
ReagentOn board Stability90 days6 weeks
Reference range0.45-2.81 g/L45-281 mg/dL0.4-2.30 g/L40-230 mg/dL
Sample Stability2-8°C for up to 3 days2 months at 20-25°C4 months at 4-8°C6 months at -20°C
Sample typeSerum,Lithium-heparin plasmaEDTA plasmaSerumLithium-heparin plasma

ii. Comparison with predicate Device: Differences

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is written in a simple, sans-serif font and is left-aligned.

c. Yumizen C1200 Immunoglobulin M

i. Comparison with predicate Device : Similarities

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.

ii. Comparison with predicate Device: Differences

7- Special Control/Guidance Document Referenced

a. Standards Followed :

The following standards & FDA guidance documents have been used to support this submission:

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a smaller font size compared to the logo. The overall image appears to be a document header or title section.

CLSI Guidelines:

  • CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures- Third Edition - October 2014
  • CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures - Second Edition - June 2012
  • CLSI EP06-A: Evaluation of the Linearity of Quantitative measurement Procedures A Statistical Approach - First Edition - April 2003
  • CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical laboratory- Third Edition - November 2008
  • CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic reagents- First Edition-September 2009

b. References cited

  • Valtec guideline (Vassault et al., Ann. Biol. Clin., 1986, (44), 686-745)

8- Device Description

  • a. Ig A

. Method

Immunoturbidimetric test. Endpoint determination of the concentration done by photometric measurement. It is an antigen-antibodyreaction of the antibodies with the that is present in the sample.

  • Reagent Yumizen C1200 Ig A
    Yumizen C1200 Immunoglobin A is ready-to-use.

Reagent 1 (R1): TRIS pH 7.5 100 mmol/L NaCl 150 mmol/L

Reagent 2 (R2): TRIS pH 8.0 100 mmol/L NaCl 300 mmol/L Anti-human IgA antibody (goat) < 1%

b. Ig G

. Method

Immunoturbidimetric test. Endpoint determination of the concentration done by photometric measurement. It is an antigen-antibodyreaction of the antibodies with the that is present in the sample.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the HORIBA Medical logo in blue. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be a heading or title for a document or section.

  • . Reagent Yumizen C1200 Ig G Yumizen C1200 Immunoglobin G is ready-to-use.
    Reagent 1 (R1): TRIS pH 7.5 100 mmol/L NaCl 150 mmol/L

Reagent 2 (R2): TRIS pH 8.0 100 mmol/L NaCl 300 mmol/L Anti-human IgG antibody (goat) < 1%

c. Ig M

. Method

Immunoturbidimetric test. Endpoint determination of the concentration done by photometric measurement. It is an antigen-antibody reaction of the antibodies present in the sample.

. Reagent Yumizen C1200 Ig M Yumizen C1200 Immunoglobin M is ready-to-use. Reagent 1 (R1): TRIS pH 7.5 100 mmol/L NaCl 150 mmol/L

Reagent 2 (R2): TRIS pH 8.0 100 mmol/L, NaCl 1150 mmol/L, Anti-human IgM antibody (goat) < 1%

This submission consists in the Yumizen C1200 Immunoglobulin A (1300023881), Yumizen C1200 Immunoglobulin G (1300023883) and Yumizen C1200 Immunoglobulin M (1300023884) reagent for serum and plasma testing for Yumizen C1200 reagent.

The Yumizen C1200 Level 1 Protein Control (1300023944) and Yumizen C1200 Level 2 Protein Control (1300023945) for use on Yumizen C1200 Analyzer and the Yumizen C1200 Protein Cal (1300023893) for use on Yumizen C1200 Analyzer are sold separately.

9- Analytical Performance Characteristics

9.1 Measuring Range

The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2. The reagent linearity was determined according to CLSI guideline EP06-A. The limit of quantitation and the linearity studies showed that claimed measuring range is appropriate.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for HORIBA Medical in blue text. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be part of a document or report.

Yumizen C1200 Immunoglobulin A ●

Limit ofdetectionLimit ofquantitationLinearity EvaluatedMeasuring range
Serum0.009 g/L0.08 g/L0.21 - 6.60 g/L0.10 to 7.00 g/L
SerumPost-dilutionNANANA7.00 to 21.00 g/L

> Linearity

The reagent linearity was determined according to CLSI guideline EP06-A.

Description

Samples used for this study are IgA spiked. The different concentrations in IgA were prepared based on the dilution of the highest concentration serum level with commercial depleted serum. Sample dilutions were assayed in quadruplicate within a single within the measuring range were used to determine linearity:

Range (mg/dL)Slope(95%CI)Intercept(95% CI)R2
21 – 6601.027(1.012-1.043)0.004(-0.060 – 0.051)0.9975

● Yumizen C1200 Immunoglobulin G

Limit ofdetectionLimit ofquantitationLinearity EvaluatedMeasuring range
Serum0.01 g/L0.53 g/L0.82 – 29.42 g/L0.75 to 30.00 g/L
SerumPost-dilutionNANANA30.00 to 90.00 g/L

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a simple, sans-serif font and appears to be part of a document or report.

> Linearity

The reagent linearity was determined according to CLSI guideline EP06-A.

Description

Samples used for this study are IgG spiked. The different concentrations in IgG were prepared based on the dilution of the highest concentration serum level with commercial depleted serum. Sample dilutions were assayed in quadruplicate within a single run and samples within the measuring range were used to determine linearity:

IgG Range(mg/dL)Slope(95% CI)Intercept(95% CI)R2
82-29420.99650.25410.9986
(0.985 – 1.008)(0.07-0.44)

Yumizen C1200 Immunoglobulin M ●

Limit ofdetectionLimit ofquantitationLinearity EvaluatedMeasuring range
Serum0.01 g/L0.03 g/L0.26 - 4.16 g/L0.20 to 5.00 g/L
SerumPost-dilutionNANANA5.00 to 15.00 g/L

> Linearity

The reagent linearity was determined according to CLSI guideline EP06-A. Description

Samples used for this study are IgM spiked. The different concentrations in IgM were prepared based on the dilution of the highest concentration serum level with commercial depleted serum. Sample dilutions were assayed in quadruplicate within a single within the measuring range were used to determine linearity:

{16}------------------------------------------------

IgM Range(mg/dL)Slope(95% CI)Intercept(95% CI)R2
26 – 4161.013.(1.001 – 1.025)– 0.087(-0.115 – -0.059)0.9994

9.2 Accuracy and Precision

The acceptance criteria for the assays were:

Within Run: CV limits, for the low, middle and high level are respectively 4.5 %, 3.8 % and 3 %. Total Precision: CV limits, for the low, middle and high level are respectively 6.0 %, 5.0 % and 4.0%

. Yumizen C1200 Immunoglobulin A

Repeatability (within-run precision) and Reproducibility (total precision)

  • Total Precision: analyzer variability - 20x2x2 study
    Five human sera samples and two levels of Yumizen C1200 Protein Control were tested with two replicates per run, two runs per day for 20 days on each of three analyzers (n=240 per sample). The results are summarized below:
IgASampleNMean(g/L)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
YumizenC1200Level 1ProteinControl2401.170.82.90.72.03.7
YumizenC1200Level 2ProteinControl2403.650.92.11.41.73.1
Sample 12400.620.91.01.21.72.5
Sample 22401.120.71.01.20.01.7
Sample 32402.390.81.31.41.52.6
Sample 42404.311.02.91.32.03.8
Sample 52405.470.91.21.52.02.9

The results are within the specifications

{17}------------------------------------------------

  • Lot to Lot variability study: 3x5x2x3 .
    Five human sera samples and two levels of Yumizen C1200 Protein Control were tested in triplicates per run, two runs per day for five days on each of three lots (n=90 per sample). The results are summarized below:
IgASampleNMean (g/L)Within-Day(%CV)Between-Day(%CV)Within -Batch(%CV)Between-Batch (%CV)Total(%CV)
YumizenC1200Level 1ProteinControl901.171.00.71.20.11.2
YumizenC1200Level 2ProteinControl903.711.10.01.10.21.1
Sample 1900.631.00.51.20.01.2
Sample 2901.133.02.84.10.04.1
Sample 3902.431.00.61.20.31.2
Sample 4904.332.01.42.40.02.4
Sample 5905.631.20.21.20.31.3

The results are within the specifications.

. Yumizen C1200 Immunoglobulin G

  • Total Precision: analyzer variability - 20x2x2 study
    Five human sera samples and two levels of Yumizen C1200 Protein Control were tested with two replicates per run, two runs per day for 20 days on each of three analyzers (n=240 per sample). The results are summarized below:
IgGSampleNMean(g/L)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
YumizenC1200 Level 1Protein Control2406.591.20.81.81.72.9
YumizenC1200 Level 2Protein Control24018.851.80.42.31.43.3
Sample 12403.141.10.51.11.82.4
Sample 22405.411.40.51.00.01.8
Sample 324010.271.50.41.51.32.5
Sample 424017.211.70.21.40.92.4
Sample 524022.362.10.01.81.13.0

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for HORIBA Medical. Below the logo is the text "Section 007. 510(k) summary". The text is in a light gray color. The image is simple and contains only text and a logo.

The results are within the specifications.

Lot variability study : 3x5x2x3 ●

Five human sera samples and two levels of Yumizen C1200 Protein Control were tested in triplicates per run, two runs per day for five days on each lot (n=90 per sample). The results are summarized below:

IgGSampleNMean (g/L)Within-Day(%CV)Between-Day(%CV)Within -Batch(%CV)Between-Batch (%CV)Total(%CV)
YumizenC1200Level 1ProteinControl906.360.91.31.60.61.7
YumizenC1200Level 2ProteinControl9018.211.31.31.80.41.9
Sample 1903.061.21.72.01.02.3
Sample 2905.301.20.81.40.81.6
Sample 3909.941.00.81.31.01.6
Sample 49016.661.31.11.71.12.0
Sample 59021.841.10.91.40.41.4

The results are within the specifications.

Yumizen C1200 Immunoglobulin M ●

  • Total Precision: analyzer variability 20x2x2 study .
    Five human sera samples and two levels of Yumizen C1200 Protein Control were tested with two replicates per run, two runs per day for 20 days on each of three analyzers (n=240 per sample). The results are summarized below:
IgMSampleNMean(g/L)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
YumizenC1200Level 1LipidControl2400.880.60.91.80.02.1

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in a bold, sans-serif font and is colored in blue. Below the word "HORIBA" is the word "Medical" in a smaller, lighter font, also in blue. The logo is simple and clean, with a focus on the company name.

Section 007. 510(k) summary

IgMSampleNMean(g/L)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
YumizenC1200Level 2LipidControl2402.730.60.81.20.91.8
Sample 12400.560.90.82.50.32.4
Sample 22401.200.50.61.40.61.7
Sample 32401.610.50.51.00.71.4
Sample 42402.090.60.51.00.41.3
Sample 52404.430.70.80.71.31.8

The results are within the specifications.

  • Lot variability study: 3x5x2x3 ●
    Five human sera samples and two levels of Yumizen C1200 Protein Control were tested in triplicates per run, two runs per day for five days on each lot (n=90 per sample). The results are summarized below:
IgMSampleNMean (g/L)Within-Day(%CV)Between-Day(%CV)Within -Batch(%CV)Between-Batch (%CV)Total(%CV)
YumizenC1200Level 1ProteinControl900.860.81.11.40.81.6
YumizenC1200Level 2ProteinControl902.761.00.81.31.11.7
Sample 1900.531.00.81.32.42.7
Sample 2901.190.80.30.90.51.0
Sample 3901.600.70.50.90.61.1
Sample 4902.090.90.41.01.01.4
Sample 5904.441.20.81.41.01.8

The results are within the specifications.

{20}------------------------------------------------

9.3 Interferences

The Interferences were determined according to the CLSI guideline EP07-A2. The acceptable bias is defined at +/-10% of the value without interfering substances. These data in the following table represent the highest values for which no interferences higher than 10% have been observed.

Serum
Hemoglobin290 $ \u03bcmol/L $500 mg/dL
Triglycerides6.07 mmol/L531.13 mg/dL
Total Bilirubin500 $ \u03bcmol/l $29.25 mg/dL
Direct Bilirubin389 $ \u03bcmol/l $22.76 mg/dL
Ascorbic Acid340 $ \u03bcmol/L $5.98 mg/dL
AcetylsalicylicAcid3.62 mmol/L65.16 mg/dL
Ibuprofen2.43 mmol/L50.10 mg/dL
Acetaminophen1324 $ \u03bcmol/L $20 mg/dL

Yumizen C1200 Immunoglobulin A .

Yumizen C1200 Immunoglobulin G ●

Serum
Hemoglobin290 $\mu$ mol/L500 mg/dL
Triglycerides5.22 mmol/L456.75 mg/dL
Total Bilirubin668 $\mu$ mol/L39.05 mg/dL
Direct Bilirubin314 $\mu$ mol/L18.36 mg/dL
Ascorbic Acid340 $\mu$ mol/L5.98 mg/dL
AcetylsalicylicAcid3.62 mmol/L65.16 mg/dL
Ibuprofen2.43 mmol/L50.10 mg/dL
Acetaminophen1324 $\mu$ mol/L20 mg/dL

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the logo for HORIBA Medical. Below the logo is the text "Section 007. 510(k) summary". The text is in a smaller font size than the logo.

Serum
Hemoglobin145 μmol/l250 mg/dL
Triglycerides5.95 mmol/L520.63 mg/dL
Total Bilirubin477 µmol/l27.88 mg/dL
Direct Bilirubin223 µmol/l13.06 mg/dL
Ascorbic Acid340 µmol/L5.98 mg/dL
AcetylsalicylicAcid3.62 mmol/L65.16 mg/dL
Ibuprofen2.43 mmol/L50.10 mg/dL
Acetaminophen1324 µmol/L20 mg/dL

Yumizen C1200 Immunoglobulin M ●

9.4 Matrix comparison on predicate device

● Yumizen C1200 Immunoglobulin A

Description:

62 samples were evaluated in duplicate on Yumizen C1200 analyzer using Yumizen C1200 IgA reagent. For this study, each paired samples (sera and heparinized plasma) has been obtained from single donor.

Immunoglobulin A (IgA, g/L)
Passing BablokNMinMaxInterceptSlopeCorrelation
Serum620.325.430.0001.0000.999
Heparin Plasma5.44

Conclusion :

The results show there is not significant difference between serum specimens and plasma with heparin specimens.

● Yumizen C1200 Immunoglobulin G

Description:

45 samples were evaluated in duplicate on Yumizen C1200 analyzer using Yumizen C1200 IgG reagent. For this study, each paired samples (sera and heparinized plasma) has been obtained from single donor.

Immunoglobulin G (IgG, g/L)
Passing BablokNMinMaxInterceptSlopeCorrelation
Serum438.1017.060.17030.99290.988
Heparin Plasma8.2616.95

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the HORIBA Medical logo in blue. Below the logo, the text "Section 007. 510(k) summary" is displayed in a light gray font. The text appears to be a section heading or title, possibly from a document or report.

Conclusion :

The results show there is no significative difference between serum specimens and plasma with heparin specimens.

● Yumizen C1200 Immunoglobulin M

Description of Test Procedure/Method

43 samples were evaluated on Yumizen C1200 analyser using Yumizen C1200 Immunoglobulin M reagent. For this study, each paired samples (sera and heparinized plasma) has been obtained from single donor.

Immunoglobulin M (IgM, g/L)
Passing BablokNMinMaxInterceptSlopeCorrelation
Serum430.292.80-0.011.0000.999
Plasma0.292.78

Conclusion

The results show there is no significative difference between serum specimens and plasma with heparin specimens.

9.5 Method comparison with a predicate device

. Yumizen C1200 Immunoglobulin A

This study has been carried out using recommendations found in the CLSI EP-9A3 guidance. Samples: Anonymous remnants of human serum specimens collected from blood bank.

129 native samples have been assayed in duplicate, in ascendant order and descendant order on 6 working days. Only the first replicate of each method will be used for the data analysis reported below.

Passing BablokMinMaxInterceptSlopeCorrelation
َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ َ1906.910.023790.99410.993

● Yumizen C1200 Immunoglobulin G

This study has been carried out using recommendations found in the CLSI EP-9A3 guidance.

Samples: Anonymous remnants of human serum specimens collected from blood bank.

214 native samples have been assayed in duplicate, in ascendant order and descendant order on 10 working days. Only the first replicate of each method will be used for the data analysis reported below.

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a simple, sans-serif font and appears to be part of a document or report.

Passing BablokNMinMaxInterceptSlopeCorrelation - r2
(g/L)2140.9628.94-0.16291.0160.993

● Yumizen C1200 Immunoglobulin M

This study has been carried out using recommendations found in the CLSI EP-9A3 guidance. Samples: Anonymous remnants of human serum specimens collected from blood bank.

153 native samples have been assayed in duplicate, in ascendant order and descendant order on 8 working days. Only the first replicate of each method will be used for the data analysis reported below.

Passing BablokMinMaxInterceptSlopeCorrelation - 1
(g/L1 ՀՀﻝ ﻟﯿﻨﮉ ﮐﺎ ﺭﻗﺒﮧ 1 ﺍﯾﮏ ﺭﮨﺎﺋﺸﯽ ﮐﮯ ﺷﮩﺮ ﺍﯾﺮﺍﻥ ﮐﮯ ﺷﮩﺮ ﺍﯾﺮﺍﻥ ﮐﮯ ﺷﮩﺮ ﺍﯾﺮﺍﻥ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﻓﮩﺮﺳﺖ ﺁﺑﺎﺩ ﻓﮩﺮﺳﺖ ﺁﺑﺎﺩ ﺍﯾﺮﺍﻥ ﻓﮩﺮﺳﺖ ﺁﺑﺎﺩ ﻓﺮﺍﻧﺴﯿﺴﯽ ﺍﯾﺮﺍﻥ ﺍﯾﺮﺍﻥ ﺍﯾﺮﺍﻥ0.251 4.410.0051131.0050.993

9.6 Reagent Stability

Closed stability 9.6.1

The closed stability was determined according to the CLSI guideline EP25-A.

. Immunoglobulin A

Stability before opening: Stable up to the expiry date on the label if stored at 2-8°C. The Shelf Life of Yumizen C1200 Immunoglobulin A is 24 months.

Immunoglobulin G .

Stability before opening: Stable up to the expiry date on the label if stored at 2-8°C. The Shelf Life of Yumizen C1200 Immunoglobulin G is 24 months.

. Immunoglobulin M

Stability before opening: Stable up to the expiry date on the label if stored at 2-8°C. The Shelf Life of Yumizen C1200 Immunoglobulin M is 24 months.

8.6.2 Open stability

The open stability was determined according to the CLSI guideline EP25-A.

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a smaller font size than the logo and is left-aligned.

On board reagent Stability:

  • Yumizen C1200 Immunoglobulin A: The stability claim after opening, on-Board, is 6 weeks.
  • Yumizen C1200 Immunoglobulin G: The stability claim after opening, on-Board, is 6 weeks.
  • . Yumizen C1200 Immunoglobulin M: The stability claim after opening, on-Board, is 6weeks.

9.7 Reference range

The Reference Range was determined according to the CLSI guideline EP28-A3.

  • . Ig A
    58 "normal samples" (22 women + 36 men) from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.

The first replicate results for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.

0.70 - 4.00 g/l (70 - 400 mg/dl) based on CRM_DA470

Reference:

Dati F. Schumann G. Thomas L. Aguzzi F. Baudner S. Bienvenu J et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/BCR/CAP reference material (CRM 470). Eur. J. Clin. Chem. Clin. Biochem. (1996) 34: 517-20.

  • . Ig G
    44 "normal samples" (16 women + 28 men) from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.

The first replicate results for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.

7 – 16 g/l (700 -1600 mg/dl) based on CRM DA470

Reference:

Dati F, Schumann G, Thomas L, Aguzzi F, Baudner S, Bienvenu J et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/BCR/CAP reference material (CRM 470). Eur. J.Clin. Chem. Clin. Biochem. (1996) 34: 517-20.:

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for HORIBA Medical. Below the logo is the text "Section 007. 510(k) summary". The text is in a light gray font. The image is simple and contains only text and a logo.

K193525 Yumizen C1200 Reagents

● Ig M

74 "normal samples" (29 women + 45 men) from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.

The first replicate results for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.

0.40 - 2.30 g/l (40 - 230 mg/dl) based on CRM_DA470

Reference:

Dati F, Schumann G, Thomas L, Aguzzi F, Baudner S, Bienvenu J et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/BCR/CAP reference material (CRM 470). Eur. J. Clin. Chem. Clin. Biochem. (1996) 34: 517-20.

10. Proposed Labeling

The labeling is written as per the recommendations given in standard EN18113-2. It takes into account the requirements of 21 CFR Part 809.10.

11. Conclusions for Performance Testing

The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that each device is substantially equivalent to its predicate device.

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).