K Number
K193496

Validate with FDA (Live)

Device Name
CharmFil
Manufacturer
Date Cleared
2020-05-06

(141 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CharmFil is indicated for following restorative applications.

  1. Class I, II, V restorations of posterior teeth
  2. Class III, IV, V restorations of anterior teeth
  3. Cervical cavities or defects involving root surfaces
Device Description

This device is light curing dental for dental surgery treatment and type 1, class 2, group 1 according to ISO 4049 classification. It is made in the form of paste and filled in syringe. When you use the product, press out the content in the syringe by screw. It is used to restore a cavity as taking out from it as much as you need with an instrument. It is dental filling material for restoring the cavities.

AI/ML Overview

The acceptance criteria and the study proving the device meets them are outlined below based on the provided text for the CharmFil device.

1. Table of Acceptance Criteria and Reported Device Performance

No.Test ItemAcceptance Criteria (Requirements)Reported Device Performance (Result)
Physical and chemical property (Polymerization time) / ISO 4049
1ColorThe exposed and unexposed halves of each of the specimens and the unirradiated specimen for any color differenceSuitable
2Appearance testShall be homogeneous and co substances are observedSuitable
3Quantity testNo less than ±5% that claimed valueSuitable
4Sensitivity to ambient lightAfter exposure for 60sec, it should be not any charge of consistencySuitable
5Polymerization depthAfter exposure for 1 minute, Opaque shade is more than 1mm, and other shades is over 1.5mmSuitable
6Flexural strength+80 MPaSuitable
7Water solubility- 40 µg/mm²Suitable
8Water sorption-7.5 µg/mm²Suitable
9Radio-opacityShall be equal to or greater than that of the same thickness of aluminum and no less than 0.5mmSuitable
10Color stabilityThe exposed and unexposed halves of each of the specimens and the unirradiated specimen for any color differenceSuitable
Elution substance test
11Character testIt should be homogeneous and appropriate. Without any foreign substance, impurity, floating substance.Suitable
12pHpH difference ≤ 1.5Suitable
13Heavy metal (Lead content)It should be less ticker than comparison liquid. (-100ppm)Suitable
14Permanganic acid calcium reduction nature substanceDifference of consumption of permanganic acid calcium≤ 2.0 mlSuitable
15Evaporation residueDifference of residues ≤ 1.0 mgSuitable
16Ultraviolet rays absorption spectrumMax. absorbance numerical value ≤ 0.1Suitable
Biological safety test (Biocompatibility, irritation)
17Cell toxicity test01 (NoneMild)Suitable
18Short period whole body toxicity test (Oral route)During the period of examination, there should not be odd or dead on it.Suitable
19Oral route impetus testCompare it with contrast product, stimulus response should not appear on mouth mucousSuitable
20Endo-toxin testAccording to the method of examination, it should be appropriate.Suitable

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each performance test. The nature of the tests (physical and chemical properties, elution substance, biological safety) suggests that material samples of the CharmFil device were tested in a laboratory setting. The data provenance is not specified beyond indicating that Dentkist, Inc. is located in Gunpo-si, Gyeonggi-do, Korea, implying that the testing likely occurred in South Korea. The study appears to be retrospective, as it's a submission for premarket notification based on completed tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a dental restorative material, and the acceptance criteria are based on objective physical, chemical, and biological safety standards (ISO 4049). There is no mention of expert readers establishing ground truth in the context of diagnostic interpretation.

4. Adjudication Method for the Test Set

Not applicable, as the acceptance criteria are objective and assessed against established international standards (ISO 4049) and biological safety guidelines. There is no qualitative assessment that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This type of study is typically used for diagnostic devices that involve human interpretation of images or other data. The CharmFil device is a dental restorative material, and its performance is evaluated based on material properties and biological safety.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in essence. The performance testing described is "standalone" in that it assesses the intrinsic properties and safety of the material itself, without direct human interaction as part of the performance evaluation (beyond conducting the tests).

7. The Type of Ground Truth Used

The "ground truth" for the performance tests consists of established industry standards (e.g., ISO 4049 for physical and chemical properties) and biological safety criteria. These are objective, quantitative or qualitative thresholds that the device's characteristics must meet.

8. The Sample Size for the Training Set

Not applicable. As this device is a dental restorative material and its performance is assessed against standardized material properties and biological safety, there is no "training set" in the context of machine learning or diagnostic algorithm development.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 6, 2020

Dentkist, Inc. % Peter Chung President Plus Global 300 Atwood St. Pittsburgh, Pennsylvania 15213

Re: K193496

Trade/Device Name: CharmFil Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: January 30, 2020 Received: February 6, 2020

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K193496

Device Name CharmFil

Indications for Use (Describe)

CharmFil is indicated for following restorative applications.

    1. Class I, II, V restorations of posterior teeth
    1. Class III, IV, V restorations of anterior teeth
  1. Cervical cavities or defects involving root surfaces

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect

of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Applicant

    1. Company : Dentkisk, Inc
    1. Address : (Dangjeong-dong) 3, Nongshim-ro, Gunpo-si, Gyeonggi-do, Korea
    1. Tel : 82-31-458-2822
    1. Fax : 82-31-458-1312
    1. Prepared date : May. 06, 2020
    1. Contact person : Peter Chung, 412-512-8802
    1. Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
    1. Submission date : May. 06, 2020
    1. Contact person : J. Y. Lee (jylee@dentkist.com)

2. Device Information

    1. Trade name : CharmFil
    1. Common name : Tooth Shade Resin Material
    1. Classification name : Material, Tooth Shade, Resin
    1. Product code : EBF
    1. Regulation number : 872.3690
    1. Class of device : Class II
    1. Panel : Dental
    1. Model codes : CharmFil Plus, CharmFil Plus Caps, CharmFil Flow

3. The legally marketed device to which we are claiming equivalence

Predicate device : K042124 DenFil, Vericom Co., Ltd. Reference device : K120768 EsFlow, Spident Co., Ltd.

4. Device description

This device is light curing dental for dental surgery treatment and type 1, class 2, group 1 according to ISO 4049 classification. It is made in the form of paste and filled in syringe. When you use the product, press out the content in the syringe by screw. It is used to restore a cavity as taking out from it as much as you need with an instrument. It is dental filling material for restoring the cavities.

5. Indications for Use Statement

CharmFil is indicated for following restorative applications.

    1. Class I, II, V restorations of posterior teeth
    1. Class III, IV, V restorations of anterior teeth
    1. Cervical cavities or defects involving root surfaces

{4}------------------------------------------------

6. Predicate device comparison table

1) CharmFil Plus and DenFil Flow

ManufacturerDenkist, Inc.VERICOM Co., Ltd.Gapanalysis
510(k) No.K193496K042124N/A
Indications for useStatementThe CharmFil are intended for class I, II, V ofposterior teeth, class III,IV, V of anterior teethand cervical cavities involving root surfaces.The DenFil are intended for class I, II, V ofposterior teeth, class III,IV, V of anterior teethand cervical cavities involving root surfaces.Same
ClassificationnameTooth shade resin materialTooth shade resin materialSame
Trade nameCharmFil Plus, CharmFil Plus CapsDenFilN/A
AppearanceImage: CharmFil Plus A3, part 1Image: DenFilSimilar
Image: CharmFil Plus A3, part 2
Image: CharmFil Plus A3, part 3
Image: CharmFil Plus A3, part 4
ProductconfigurationSyringe, Disposable tipSyringe, Disposable tipSame
Expiration date2 years2 yearsSame
StandardconformedISO 4049N/A
BiocompatibilityYesYesSame
Principle ofoperationManualManualSame

2) CharmFil Flow and EsFlow

ManufacturerDenkist, Inc.Spident Co., Ltd.Gapanalysis
510(k) No.K193496K120768N/A
Indications foruse StatementThe CharmFil are intended for class I, II, V ofposterior teeth, class III,IV, V of anterior teethand cervical cavities involving root surfaces.The EsFlow are intended for use invarious dental procedures, includingdirect anterior and posterior restorations,blocking out cavity undercuts beforefabrication indirect restorations, andrepair of porcelain / composite materials.Similar
ClassificationnameTooth shade resin materialTooth shade resin materialSame
Trade nameCharmFil FlowEsFlowN/A
AppearanceImage: CharmFil FlowImage: EsFlowSimilar
ProductconfigurationSyringe, Disposable tipSyringe, Disposable tipSame
Expiration date2 years2 yearsSame
StandardconformedISO 4049ISO 4049EN 1641N/A
BiocompatibilityYesYesSame
Principle ofoperationManualManualSame

{5}------------------------------------------------

CharmFil Plus and CharmFil Plus Caps contains the same materials with same amount. Only the using method is different because the CharmFil Plus Caps is manufactured capsule-like shape.

The CharmFil has the similar device characteristics as the predicate device, the DenFill and EsFlow; intended use, material, chemical composition, design and use concept are similar.

The differences between CharmFil and the predicate device and reference device has been subjected to performance and product validations testing prior to release. Tests have been performed to ensure the devices comply to applicable industry and US regulations.

An extensive review of literature pertaining to the safety and biocompatibility of CharmFil has been conducted. Appropriate safeguards have been incorporated in the design of CharmFil.

Although some of the materials used are different, the materials for performance are all the other materials meet biocompatibility criteria. The methods of use and physical and chemical performance are similar and all meet the appropriate criteria.

No.Test itemRequirementsResult
Physical and chemical property (Polymerization time) / ISO 4049
1ColorThe exposed and unexposed halves of each of the specimensand the unirradiated specimen for any color differenceSuitable
2Appearance testShall be homogeneous and co substances are observedSuitable
3Quantity testNo less than ±5% that claimed valueSuitable
4Sensitivity to ambient lightAfter exposure for 60sec, it should be not any charge ofconsistencySuitable
5Polymerization depthAfter exposure for 1 minute, Opaque shade is more than1mm, and other shades is over 1.5mmSuitable
6Flexural strength+80 MpaSuitable
7Water solubility- 40 µg/mm²Suitable
8Water sorption-7.5 µg/mm²Suitable
9Radio-opacityShall be equal to or greater than that of the same thickness ofaluminum and no less than 0.5mmSuitable
10Color stabilityThe exposed and unexposed halves ofeach of the specimens and theunirradiated specimen for any colordifferenceSuitable
Elution substance test
11Character testIt should be homogeneous and appropriate. Without anyforeign substance, impurity, floating substance.Suitable
12pHpH difference ≤ 1.5Suitable
13Heavy metal (Leadcontent)It should be less ticker than comparison liquid. (-100ppm)Suitable
14Permanganic acid calciumreduction naturesubstanceDifference of consumption of permanganic acid calcium≤ 2.0mlSuitable
15Evaporation residueDifference of residues ≤ 1.0 mgSuitable
16Ultraviolet rays absorptionspectrumMax. absorbance numerical value≤ 0.1Suitable
Biological safety test (Biocompatibility, irritation)
17Cell toxicity test01 (NoneMild)Suitable
18Short period whole bodyDuring the period of examinationSuitable

7. Performance testing

{6}------------------------------------------------

toxicity test (Oral route)there should not be odd or dead on it.
19Oral route impetus testCompare it with contrast product, stimulus response should not appear on mouth mucousSuitable
20Endo-toxin testAccording to the method of examination, it should be appropriate.Suitable

8. Conclusion:

The Device is investigated for function to compare the operation of function between CharmFil and predicate devices.

Comparison results demonstrate that the specifications and performance of the device are same as functional as the legally marketed predicate device.

Therefore, it is concluded that CharmFil is substantially equivalent to the legally marketed predicate device.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.