K Number
K193449

Validate with FDA (Live)

Date Cleared
2020-09-11

(273 days)

Product Code
Regulation Number
884.5160
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.

Device Description

The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are electric breast pumps that are intended for lactating women to express and collect breast milk. The Electric Breast Pump and Electric Double Breast Pump are intended for a single user and are to be used in the home environment. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053_01) are Li-ion-battery powered breast pumps that include following main components: the expressed milk collection kit, the pump unit, and tubing. The expressed milk collection kit includes a funnel cover, silicone cushion, funnel adapter, membrane, valve body, valve membrane cap, and milk bottle. The pump unit contains a chip that controls the work of the vacuum pump and solenoid valve to draw air inside of the funnel adapter to form negative pressure. The tubing has two connectors on the two ends. The user attaches the connector to the air vent of the membrane cap and the connector to the pump unit separately to ensure a tight fit.

AI/ML Overview

This document is a 510(k) summary for the Electric Breast Pump (model KP033_01) and Electric Double Breast Pump (model KP053_01). It outlines the device's characteristics and the studies conducted to demonstrate its substantial equivalence to a predicate device.

Here's the information extracted and organized as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (What was measured)Reported Device Performance (What was found)
Biocompatibility (Cytotoxicity, Sensitization, Skin Irritation) in accordance with ISO 10993-1, 10993-5, 10993-10Met the predefined acceptance criteria.
Electrical safety (IEC 60601-1:2012)Met the predefined acceptance criteria.
Electromagnetic compatibility (EMC) (IEC 60601-1-2:2014)Met the predefined acceptance criteria.
Software Verification and ValidationConducted as recommended by FDA guidance ("Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (2005)). Met the predefined acceptance criteria.
Minimum and maximum vacuum pressure (single and double pumping modes)Met the predefined acceptance criteria.
Technical specifications for stimulation mode (all cycle and vacuum settings)Met the predefined acceptance criteria.
Technical specifications for expression mode (all cycle and vacuum settings)Met the predefined acceptance criteria.
Working current, pressure, cycle rate, and noiseMet the predefined acceptance criteria.
Backflow protection mechanismMet the predefined acceptance criteria.
Battery specification validation and performanceMet the predefined acceptance criteria.
Use life testing (vacuum level and battery performance)Met the predefined acceptance criteria.
Maximum Expression Pressure247.5mmHg (Matches predicate device)
Stimulation Pressure range (KP033_01)30-120 mmHg (Matches predicate device)
Stimulation Pressure range (KP053_01)37.5-112.5mmHg (Slight difference from predicate: 30-120 mmHg)
Stimulation Velocity100 Cycles/min (Matches predicate device)
Expression Velocity26-60 Cycles/min (Difference from predicate: 20-90 cycles/min)
Expression Pressure Range120-247.5 mmHg (Difference from predicate: 75-247.5 mmHg)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (number of devices or tests) for the performance, biocompatibility, electrical safety, or EMC testing. However, it indicates that "Performance testing was conducted on the subject device," implying physical testing of the device itself.

The data provenance is not explicitly stated as country of origin, but the submitting company is "Guangdong Horigen Mother & Baby Products Co., Ltd." in China. The study appears to be prospective testing of the manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided. The testing appears to be objective, engineering, and laboratory-based, rather than requiring expert adjudication of results.

4. Adjudication method for the test set

This information is not applicable and not provided. The testing is based on objective measurements against predefined criteria.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is an electric breast pump, not an AI-assisted diagnostic or imaging device used by human 'readers'.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical product (breast pump), not an algorithm. The software component mentioned is for control of the device, not for independent diagnostic or analytical performance.

7. The type of ground truth used

For biocompatibility, electrical safety, EMC, software, and performance testing, the "ground truth" is defined by established international standards (ISO 10993-1, 10993-5, 10993-10; IEC 60601-1:2012; IEC 60601-1-2:2014) and predefined acceptance criteria based on those standards or engineering specifications. For comparison of certain parameters (e.g., maximum expression pressure), the predicate device's specifications served as a benchmark for substantial equivalence.

8. The sample size for the training set

This is not applicable. The device is not an AI/ML model that requires a "training set."

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

September 11, 2020

Guangdong Horigen Mother & Baby Products Co., Ltd Jun Deng Vice President No. 18, Pingnan Industrial Zone, Mianbei Street, Chaoyang District Shantou, 515100 China

Re: K193449

Trade/Device Name: Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053_01) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: August 8, 2020 Received: August 10, 2020

Dear Jun Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193449

Device Name Electric Breast Pump (model KP033 01)

Electric Double Breast Pump (model KP053 01)

Indications for Use (Describe)

The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary - K193449

I. SUBMITTER:

510(k) Owner:

Guangdong Horigen Mother & Baby Products Co., Ltd. No. 18, Pingnan Industrial Zone, Mianbei Street, Chaoyang District, 515100 Shantou, Guangdong, PEOPLE'S REPUBLIC OF CHINA

Primary Contact Person: Jun Deng Title: Vice president Phone: +86 (754) 83613668 EXT.836 Email: dengjun@horigen.cn

Date summary prepared: September 8, 2020

II. DEVICE

Device Name: Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053 01) Common Name: Powered Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Product Code: HGX (Pump, Breast, Powered) Regulatory Class: II

PREDICATE DEVICE III.

The predicate device is the Electric Breast Pump (K182413).

This predicate device has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are electric breast pumps that are intended for lactating women to express and collect breast milk. The Electric Breast Pump and Electric Double Breast Pump are intended for a single user and are to be used in the home environment.

The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053_01) are Li-ion-battery powered breast pumps that include following main components: the expressed milk collection kit, the pump unit, and tubing. The expressed milk collection kit includes a funnel cover, silicone cushion, funnel

{4}------------------------------------------------

adapter, membrane, valve body, valve membrane cap, and milk bottle. The pump unit contains a chip that controls the work of the vacuum pump and solenoid valve to draw air inside of the funnel adapter to form negative pressure. The tubing has two connectors on the two ends. The user attaches the connector to the air vent of the membrane cap and the connector to the pump unit separately to ensure a tight fit.

DEVICE MODELS:

  • KP033_01: Single pump ●
  • KP053_01: Double pump

INDICATIONS FOR USE V.

The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.

DEVICESubject DevicePrimary Predicate Device
Product NameElectric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01)Electric Breast Pump
Indications for UseThe Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are intended for a single user.The Electric Breast Pump (K182413) is intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs.

VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICE

{5}------------------------------------------------

K1934-
Classification ProductCodeHGXHGX
Single UserYesYes
Single/Double PumpSingle or Double PumpSingle
Environment of UseHomeHome
Pump TypeDiaphragm pumpDiaphragm pump
Adjustable VacuumLevels9 Levels9 Levels
Backflow Protection(prevention ofbackflow of liquid intopump/tubing)YesYes
Maximum ExpressionPressure247.5mmHg247.5mmHg
Stimulation PressurerangeKP033_01 (30-120, mmHg)KP053_01 (37.5-112.5mmHg )30-120 mmHg
Stimulation Velocity100 Cycles/min100 cycles/min
Expression Velocity26-60 Cycles/min20 – 90 cycles/ min
Expression PressureRange120 -247.5 mmhg75-247.5mmhg

{6}------------------------------------------------

The intended use of the subject and predicate devices is the same (express and collect milk from breasts of lactating women). The subject device has different adjustable stimulation levels and a different power source as compared to the predicate device. There are minor differences in stimulation and expression pressure ranges and cycle speeds of stimulation and expression. However, these differences do not raise different questions of safety and effectiveness.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility

A biocompatibility evaluation for the subject device was conducted in accordance with the International Standard ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:

  • Cytotoxicity - (ISO 10993-5: 2009)
  • Sensitization - (ISO 10993-10:2010)
  • . Skin Irritation - (ISO 10993-10:2010)

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject device in accordance with IEC 60601-1:2012 (electrical safety) and IEC 60601-1-2:2014 (EMC).

Software

Software verification and validation for the subject device was conducted as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (2005).

Performance testing

Performance testing was conducted on the subject device, including the following assessments:

  • . Minimum and maximum vacuum pressure testing in single and double pumping mode
  • Assessment of technical specifications for stimulation mode and expression mode at ● all cycle and vacuum settings
  • . Assessment of working current, pressure, cycle rate and noise
  • Backflow protection mechanism testing
  • Battery specification validation and performance .
  • . Use life testing of vacuum level and battery performance

{7}------------------------------------------------

All of the tested parameters met the predefined acceptance criteria.

VIII. CONCLUSION

The nonclinical performance data described above demonstrate that the Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).