K Number
K182413
Date Cleared
2019-03-07

(184 days)

Product Code
Regulation Number
884.5160
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Electric Breast Pump is intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump is intended for a single user.

Device Description

The Electric Breast Pump is designed, manufactured by the Guangdong Horigen Mother & Baby Products Co., Ltd. The device is intended for lactating women in a home setting for a single user. It comprises of a pump unit and the expression collection kit including tubing, and its raw materials include polypropylene, liquid and solid silicone, and thermoplastic elastomer. This electric breast pump has a negative pressure module that utilizes a single-chip microcomputer. It has multiple stimulation levels for breast massage, and multiple milk expression speed intensities. The keyboard of the control panel is soft. The screen is an LCD or LED screen, allowing for process viewing. The pump's electronic memory takes over, mimicking the rhythm. The electric breast pump is capable of providing vacuum levels from 30 to 247.5 mmHg with cycle rates up to 100 cycles per minute.

AI/ML Overview

The provided document is a 510(k) K182413 FDA clearance letter and summary for an Electric Breast Pump. This type of document is for a physical medical device, not a software-based AI or diagnostic device that would involve a "test set" or "ground truth" established by experts or MRMC studies.

Therefore, many of the requested elements (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, type of ground truth, training set size and ground truth establishment for training set) are not applicable to this device and the information provided in this document.

However, I can extract the acceptance criteria and the performance testing performed to demonstrate the device meets those criteria.

Here's the information based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document broadly states that "All of the tested parameters met the predefined acceptance criteria" in the Performance Testing section. While specific numerical acceptance criteria values are not explicitly stated in this public summary, the device's operational characteristics are described, which implies these fall within predefined acceptable ranges.

CategoryAcceptance CriteriaReported Device Performance
BiocompatibilityCompliance with ISO 10993-1:2009 for "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Specific tests: Cytotoxicity (ISO 10993-5: 2009), Sensitization (ISO 10993-10:2010), Skin Irritation (ISO 10993-10:2010).Performed and found to meet acceptance criteria. (Implied compliance with the standards listed).
Electrical Safety & EMCCompliance with IEC 60601-1:2012 (electrical safety) and IEC 60601-1-2:2014 (EMC).Testing was conducted, and the device "complies with the IEC 60601-1:2012, standard for electrical safety and the IEC 60601-1-2:2014 standard for EMC."
Software ValidationCompliance with FDA Guidance: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005."Software Validation was performed per the FDA Guidance." (Implied to meet acceptance criteria as the device was cleared)
Performance TestingPredefined acceptance criteria for: Working current, pressure, cycle rate, noise, backflow protection mechanism, battery specification, stability of vacuum level, and shelf life."All of the tested parameters met the predefined acceptance criteria." The document also specifies the operational parameters:Vacuum levels: 30 to 247.5 mmHgCycle rates: up to 100 cycles per minuteStimulation Phase: Fast cycles (70100 Cycles/min), low vacuum (30 ~ 112.5 mmHg), 9 levels.Expression Phase: Cycles (2490 Cycles/min), higher vacuum (75 ~ 247.5 mmHg), 9 levels.Features: Closed system with anti-backflow, LCD or LED screen, single/double pumping, dual power source. (These implicitly met their respective criteria).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is typically not included in a 510(k) summary for a physical non-AI device. The testing described (biocompatibility, electrical safety, performance) uses established laboratory methods on physical units of the device. There isn't a "test set" of clinical data in the way you'd describe for an AI model.

  • Sample Size: Not specified as a number of "cases" or "patients" but rather the physical units of the device tested in laboratory settings.
  • Data Provenance: Not applicable in the context of clinical data. The tests were conducted by the manufacturer, Guangdong Horigen Mother & Baby Products Co., Ltd., which is located in China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. "Ground truth" in this context would refer to meeting engineering specifications or validated biological/electrical safety standards, not expert clinical interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical device, not an algorithm. Its "standalone" performance refers to its ability to meet the physical and electrical specifications independently.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance testing is defined by:

  • International standards (ISO 10993, IEC 60601-1, IEC 60601-1-2) for biocompatibility, electrical safety, and EMC.
  • FDA Guidance for software validation.
  • Predefined engineering specifications (e.g., pressure ranges, cycle rates, noise levels) established by the manufacturer and accepted by the FDA as equivalent to the predicate device.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning model.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 7, 2019

Guangdong Horigen Mother & Baby Products Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Services Co.,Ltd. 8-9th Floor, R&D Building, No.26 Qinglan Street, Panyu District Guangzhou, Guangdong 510006 China

Re: K182413

Trade/Device Name: Electric Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: February 1, 2019 Received: February 4, 2019

Dear Mike Gu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason Roberts -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182413

Device Name Electric Breast Pump

Models: XN-2212M2, XN-2219M1, XN-2219M2, XN-2223M2, XN-2223M2, XN-2233M1, XN-2233M2

Indications for Use (Describe)

The Electric Breast Pump is intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump is intended for a single user.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

1. SUBMITTER

Guangdong Horigen Mother & Baby Products Co., Ltd. NO.8, Yi Road, Pingbei Industrial Zone, Chaoyang District, Shantou, 515100, Guangdong, China Phone: +86-754-83613668 Fax: +86-754-83843338

Primary ContactMike Gu
Person:Regulatory Affairs ManagerGuangzhou Osmunda Medical Device Technical ServiceCo., Ltd.Tel: (+86)-20-6231 6262Fax: (+86) -20-8633 0253
Secondary ContactJun Deng
Person:General ManagerGuangdong Horigen Mother & Baby Products Co., Ltd.Tel: (+86)-754-83613668-866Fax: (+86)-754-83843338
Date preparedMarch 6, 2019
2. DEVICE
Device Name:Electric Breast Pump
Common/Usual Name:Powered breast pump
Model:XN-2212M2, XN-2219M1, XN-2219M2, XN-2223M1,XN-2223M2, XN-2233M1 and XN-2233M2
Regulation number21 CFR 884.5160 Powered breast pump
Regulation Class:II
Product Code:HGX Pump, Breast, Powered

3. PREDICATE DEVICE

Predicate device: K143585, Electric Breast Pump

{4}------------------------------------------------

This predicate and the referenced have not been subject to a design-related recall. No reference devices were used in this submission.

4. DEVICE DESCRIPTION

The Electric Breast Pump is designed, manufactured by the Guangdong Horigen Mother & Baby Products Co., Ltd. The device is intended for lactating women in a home setting for a single user. It comprises of a pump unit and the expression collection kit including tubing, and its raw materials include polypropylene, liquid and solid silicone, and thermoplastic elastomer.

This electric breast pump has a negative pressure module that utilizes a single-chip microcomputer. It has multiple stimulation levels for breast massage, and multiple milk expression speed intensities. The keyboard of the control panel is soft. The screen is an LCD or LED screen, allowing for process viewing. The pump's electronic memory takes over, mimicking the rhythm. The electric breast pump is capable of providing vacuum levels from 30 to 247.5 mmHg with cycle rates up to 100 cycles per minute.

The Electric Breast Pump provides the following user features:

  • . Closed system with anti-backflow
  • . LCD or LED screen touch button operation
  • . 2-phase expression: Stimulation and Expression phases
    • Stimulation Phase: Expression pattern with fast cycles (70~100 Cycles/min) and low vacuum (30 ~ 112.5 mmHg) to start milk flowing, 9 levels
    • Phase: Expression cvcles (24~90Cycles/min) and higher vacuum (75 ~ 247.5 mmHg) to express milk, 9 levels
  • Double-pumping ability: single or double pumping
  • . Dual power source: it can be operated by A/C adapter or batteries
  • Built-in rechargeable li-ion polymer battery or additional AA battery power option

5. INDICATIONS FOR USE

The Electric Breast Pump is intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump is intended for a single user.

{5}------------------------------------------------

The indications for use statement for the subject device is similar to that of the predicate, with minor alterations in text; however, the intended use of the subject and predicate devices is the same (express and collect milk from breasts of lactating women).

{6}------------------------------------------------

Device & Predicate Device(s):K182413K143585
General Device Characteristics
Environment of UseHomeHome, institutions andhospital
Single or Multiple-UserSingleSingle
Pump Typediaphragm pumpdiaphragm pump
Adjustable Vacuum Levels9 levels9 levels
Maximum Expression Intensity247.5270
Stimulation Intensity range30-112.515.2-144.4
Stimulation Velocity70 – 100 cycles/min95-105
Expression Velocity24 – 90 cycles/min20-65
User InterfaceHardware interfaceHardware interface
Backflow Protection (preventionof backflow of liquid intopump/tubing)YesYes
Overflow Protection (stopspumping when the bottle is fullto prevent overflow)NoNo
Power SourceAC Mains 100-240 VAC,50/60Hz, 1.5A/1A4AA Alkaline batteryRechargeable Li-IonBattery (7.4 V / 1750mAh)4AA battery or AC/DCadapter

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

{7}------------------------------------------------

The subject device is different from the predicate and the referenced in suction pressure range, cycle speed, maximum suction pressure and the power source. However, these differences do not raise different questions of safety and effectiveness. The technological differences can be evaluated through the software, performance and EMC/electrical safety testing provided.

PERFORMANCE DATA 7.

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Electric Breast Pump was conducted in accordance with the International Standard ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:

  • Cytotoxicity - (ISO 10993-5: 2009)
  • Sensitization - (ISO 10993-10:2010)
  • Skin Irritation - (ISO 10993-10:2010)

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Electric Breast Pump. The device complies with the IEC 60601-1:2012, standard for electrical safety and the IEC 60601-1-2:2014 standard for EMC. Software Validation was performed per the FDA Guidance: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.

Performance testing

Performance testing was conducted on the Electric Breast Pump. Technical parameters of stimulation mode and expression mode, including working current, pressure, cycle rate and noise, backflow protection mechanism, and battery specification were evaluated in the performance testing. In addition, the stability of vacuum level and shelf life were evaluated. All of the tested parameters met the predefined acceptance criteria.

8. CONCLUSION

The indications for use statement for the subject device is similar to that of the predicate. The differences between the Electric Breast Pump and its predicate device

{8}------------------------------------------------

do not raise new issues of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the Electric Breast Pump should perform as intended in the specified use conditions.

From the results of non-clinical data including the performance testing described, Guangdong Horigen concludes that the Electric Breast Pump is as safe and as effective as the predicate device.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).