K Number
K193349

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-14

(133 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vapocoolshot Mist is intended for topical application to skin, intact mucous membrane (oral cavity, nasal passageways, lips) and minor open wounds. The Vapocoolshot Mist is used to target and minimize cooling area for lessening pain associated with injections (venipuncture, IV starts, cosmetic procedures (such as lancing boils, incision, drainage of small abscesses and sutures) and the temporary relief of minor sports injuries (sprains, bruising, cuts, and abrasions).

Device Description

The Vapocoolshot Mist is a Vapocoolant (skin refrigerant) canister with an accessory nozzle that is intended for instant topical anesthetic. The Vapocoolshot Mist allows for the ease of use for a rapid, targeted, comfortable misting action of the refrigerant (cold like ice) onto the skin surface. Vapocoolshot Mist is prescription device designed to deliver a standard mixture 245fa (1,1,3,3-Pentafluoropropane) and 134a (1,1,1,2-Tetrafluoroethane) mist spray that will provide a transient refrigerant action. This mixture self-propels itself from the delivery system, which is designed to account for its low vapor pressure. The Vapocoolshot's nozzle, device's delivery system, controls the amount of the vapocoolant mixture that is dispensed. The mist spray configuration produces very fine droplets that create cooling at the points of contact. The Vapocoolshot produces a mist that contacts the skin surface at a targeted location. The skin is cooled through rapid evaporation of the non-medicated propellants.

AI/ML Overview

This document describes a 510(k) submission for the Vapocoolshot Mist, a topical refrigerant. The submission aims to demonstrate substantial equivalence to legally marketed predicate devices, primarily "Gebauer's Pain Ease" (K172028) and "Ouchless" (K093951).

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state quantitative acceptance criteria in a table format. Instead, it relies on comparative equivalence to predicate devices across various characteristics. The "reported device performance" is framed as demonstrating similarity or equivalence.

Characteristic / Performance MetricAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityMaterials compatible for indirect contact; gas blend approved for clinical use.Device has "Indirect Contact." Gas blend approved. Materials compatible.
Chemical CompositionIdentical aerosol profiles as predicates (95% 245fa, 5% 134a).Aerosol profile checked and verified upon receipt to ensure the same as primary predicate.
Structural and Parts CompositionCanisters, valves, caps equivalent to Gebauer; overall device equivalent to Ouchless.Engineering measurements and visual inspections confirmed equivalence to Gebauer (canisters, valves, caps) and Ouchless (overall device).
Directions for Use (DFU)Similar intended use, precautions, warnings, contraindications, and treatment regimen to predicates.All key elements of DFU similar to Gebauer and Ouchless. No significant differences.
Temperature OutputSimilar temperature output at application surface as Gebauer.Comparative performance testing demonstrated equivalent results.
Volume DispensedSimilar volume dispensed per actuation/total as Gebauer.Comparative performance testing demonstrated equivalent results.
Stability and Shelf LifeIdentity, strength, quality, and purity maintained throughout labeled dating period under controlled and accelerated conditions.Stability protocol developed; testing confirmed all requirements met established acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state sample sizes for specific test sets in a numerical fashion (e.g., number of patients or devices tested). Instead, it refers to "engineering verification measurements" and "comparative performance testing."

  • Sample Size: Not specified quantitatively.
  • Data Provenance: The studies appear to be retrospective bench testing and engineering comparisons rather than prospective clinical trials with human subjects. The details about data origin (e.g., country) are not provided, but given the submission is to the FDA, it is implied to be relevant to U.S. regulatory standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of experts to establish "ground truth" in the context of clinical outcomes or diagnostic accuracy. The testing performed is focused on comparative engineering and chemical composition, for which ground truth is established by measurement and analytical methods, not typically by expert consensus of a specific number of clinicians.

4. Adjudication Method for the Test Set

Not applicable. The reported studies are bench tests and engineering comparisons, not studies requiring adjudication of clinical or diagnostic outcomes by experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks, which is not the nature of this device or its reported testing.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The Vapocoolshot Mist is a physical device that delivers a topical refrigerant, not an algorithm or AI system.

7. Type of Ground Truth Used

The "ground truth" in this context is based on:

  • Chemical Analysis: Chromatographic or other analytical methods to determine the exact composition of the vapocoolant.
  • Engineering Specifications and Measurements: Physical measurements of device components, discharge characteristics (temperature, volume), and adherence to design specifications.
  • Regulatory Standards: Compliance with established safety and efficacy standards for medical devices and comparison to legally marketed predicate devices.

8. Sample Size for the Training Set

Not applicable. This device is not an AI/ML product, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 14, 2020

Vapocoolshot, Inc. % Neil Ross Chief Scientific Officer, Quality Affairs N2Pharma, LLC 1071 Nandina Dr Weston, Florida 33327

Re: K193349

Trade/Device Name: Vapocoolshot Mist Regulatory Class: Unclassified Product Code: MLY Dated: October 11, 2019 Received: December 3, 2019

Dear Neil Ross:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K193349

Device Name Vapocoolshot Mist

Indications for Use (Describe)

The Vapocoolshot Mist is intended for topical application to skin, intact mucous membrane (oral cavity, nasal passageways, lips) and minor open wounds. The Vapocoolshot Mist is used to target and minimize cooling area for lessening pain associated with injections (venipuncture, IV starts, cosmetic procedures (such as lancing boils, incision, drainage of small abscesses and sutures) and the temporary relief of minor sports injuries (sprains, bruising, cuts, and abrasions).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K193349 – 510(k) Summary

Vapocoolshot Inc. – Vapocoolshot Mist

I. SUBMITTER

Owner:Vapocoolshot Inc.1155 Main Street, Suite 105Jupiter FL, 33458(561) 427-0420 (o)(561) 420-8330 (fax)
Contact Person:Jacob Leibovici1155 Main Street, Suite 105Jupiter FL, 33458(561) 306-3434 (m)(561) 427-0420 (o)
Date Prepared:February 28, 2020
Subject Device:
Trade/Device Name:Vapocoolshot Mist (Model VM11000)
Common Name:Refrigerant, Topical (Vapocoolant)
Review panel:Product code:Physical MedicineMLY
Primary Predicate Device:Gebauer's Pain Ease (Mist Spray & Stream Spray)Gebauer Company.K172028Legally marketed medical device
Predicate 1:OuchlessOccam Design LLCK093951

{4}------------------------------------------------

II. DEVICE DESCRIPTION

The Vapocoolshot Mist is a Vapocoolant (skin refrigerant) canister with an accessory nozzle that is intended for instant topical anesthetic. The Vapocoolshot Mist allows for the ease of use for a rapid, targeted, comfortable misting action of the refrigerant (cold like ice) onto the skin surface.

Vapocoolshot Mist is prescription device designed to deliver a standard mixture 245fa (1,1,3,3-Pentafluoropropane) and 134a (1,1,1,2-Tetrafluoroethane) mist spray that will provide a transient refrigerant action. This mixture self-propels itself from the delivery system, which is designed to account for its low vapor pressure. The Vapocoolshot's nozzle, device's delivery system, controls the amount of the vapocoolant mixture that is dispensed. The mist spray configuration produces very fine droplets that create cooling at the points of contact. The Vapocoolshot produces a mist that contacts the skin surface at a targeted location. The skin is cooled through rapid evaporation of the non-medicated propellants.

Vapocoolshot Mist canister is equivalent to the Primary predicate device. Gebauer's Pain Ease. Both devices are indicated for topical application to skin, intact mucous membrane (oral cavity, nasal passageways, lips) and minor open wounds. The Vapocoolshot Mist is used to target and minimize cooling area for lessening pain associated with injections (venipuncture, IV starts, cosmetic procedures), minor surgical procedures (such as lancing boils, incision, drainage of small abscesses and sutures) and the temporary relief of minor sports injuries (sprains, bruising, cuts, and abrasions).

Vapocoolshot Mist is equivalent to the Predicate 1, Ouchless. Both Vapocoolshot and Ouchless devices are indicated for topical application of refrigerant (cold like ice) to the skin and both devices have the same exact mixture blend of vapocoolant.

III. INDICATIONS FOR USE

The Vapocoolshot Mist is intended for topical application to skin, intact mucous membrane (oral cavity, nasal passageways, lips) and minor open wounds. The Vapocoolshot Mist is used to target and minimize cooling area for lessening pain associated with injections (venipuncture, IV starts, cosmetic procedures), minor surgical procedures (such as lancing boils, incision, drainage of small abscesses and sutures) and the temporary relief of minor sports injuries (sprains, bruising, cuts, and abrasions).

IV. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

The subject device Vapocoolshot Mist has similar technical characteristics as the Primary predicate device (Gebauer), including materials, design, and energy source. Refer to the following table for the comparison between the subject device Vapocoolshot Mist and the Primary predicate:

Comparison Chart – Technological Characteristics
Trade NameVapocoolshot MistK193349Subject DeviceGebauer's Pain EaseK172028Primary Predicate

{5}------------------------------------------------

TypeRxRxComparison Chart – Technological Characteristics
Product DesignPressurized dispensing container, whichincludes the vapocoolant, canister,valve, actuator, cap and an accessorynozzle.Pressurized dispensing container, whichincludes the vapocoolant, canister,valve, actuator, and cap.Trade NameVapocoolshotK193349Subject DeviceOuchlessK093951Predicate 1
Indication for UseThe Vapocoolashot Mist is intended for topical application to skin, intactmucous membrane (oral cavity, nasalpassageways, lips) and minor openwounds. The Vapocoolshot Mist is usedto target and minimize cooling area forlessening pain associated withinjections (venipuncture, IV starts,cosmetic procedures), minor surgicalprocedures (such as lancing boils,incision, drainage of small abscessesand sutures) and the temporary relief ofminor sports injuries (sprains, bruising,cuts, and abrasions).The vapocoolant (skin refrigerant) isintended for topical application to skin,intact mucous membrane (oral cavity, nasalpassageways, lips) and minor open wounds.The Vapocoolshot Mist can control painassociated with injections(venipuncture, IV starts, cosmeticprocedures), minor surgical procedures (suchas lancing boils, incision, drainage ofsmall abscesses and sutures) and thetemporary relief of minor sports injuries(sprains, bruising, cuts, and abrasions).TypeRxRx
Product Fill Volume2oz (65 mL)3.5oz (103.5mL)Product DesignPressurized dispensing container, whichincludes the vapocoolant, canister, valve,actuator, cap and accessory nozzle.Pressurized dispensing container, whichincludes the vapocoolant, canister, valve, andpush button. (No canister cap, No actuator)
VapocoolantComposition1,1,1,3,3-Pentafluoropropane (HFC-245fa) 95%and1,1,1,2-Tetrafluoroethane (HFC-134a)5%1,1,1,3,3-Pentafluoropropane (HFC-245fa) 95% and1,1,1,2-Tetrafluoroethane (HFC-134a) 5%Indication forUseThe Vapocoolashot Mist is intended for topicalapplication to skin, intact mucous membrane(oral cavity, nasal passageways, lips) and minoropen wounds. The Vapocoolshot Mist is used totarget and minimize cooling area for lesseningpain associated with injections (venipuncture,IV starts, cosmetic procedures), minor surgicalprocedures (such as lancing boils, incision,drainage of small abscesses and sutures) and thetemporary relief of minor sports injuries(sprains, bruising, cuts, and abrasions).Used like ice for the temporary relief ofminor pain.
Energy DeliveredThermal Energy via Refrigerant SprayThermal Energy via Refrigerant SprayProduct FillVolume2oz (65 mL)19mL
Energy DepositedN/AN/AVapocoolantComposition1,1,1,3,3-Pentafluoropropane (HFC-245fa) 95%and1,1,1,2-Tetrafluoroethane (HFC-134a) 5%1,1,1,3,3- Pentafluoropropane (HFC-245fa)95%and1,1,1,2-Tetrafluoroethane (HFC-134a) 5%
Vapocoolant DischargeMethodDepress the Actuator Button to releasethe vapocoolantDepress the Actuator Button to release thevapocoolantEnergyDeliveredThermal Energy via Refrigerant SprayThermal Energy via Refrigerant Spray
EnvironmentalCompatibilityNon-FlammableNon-FlammableEnergyDepositedN/AN/A
Mechanical SafetyMechanism has positive shut-offrelease.Mechanism has positive shut-off releaseVapocoolantDischargeMethodDepress the Actuator Button to release thevapocoolantDepress the Button to release thevapocoolant
EnvironmentalCompatibilityNon-FlammableNon-Flammable
MechanicalSafetyMechanism has positive shut-off release.Mechanism has positive shut- off release

The Vapocoolshot Mist device has similar technical and technological characteristics as the Predicate 1 (Ouchless), including materials, design, and energy source. Refer to the following table for the comparison between the Vapocoolshot Mist device and the Predicate 1:

{6}------------------------------------------------

Equivalence:

The Primary predicate (Gebauer) is legally marketed device. The subject device as a unit has similar intended use as the Primary predicate device.

{7}------------------------------------------------

The subject device is a prescription device as well as Primary predicate and Predicate 1.

The subject device has the same technological characteristics as the Predicate 1(Ouchless). There are no technological differences, including no changes in the materials, design, energy source, or other features of the device from those of the Predicate 1.

The subject, Primary predicate, and Predicate 1 devices use identical (substantially equivalent) chemical composition by type and percent of components: 245fa (1,1,1,3,3-Pentafluoropropane) at 95% of the total and 134a (1,1,2-Tetrafluoroethane) at 5% of the total mixture.

Labeling:

The labeling of subject device has been prepared to ensure the medical professional has adequate and clear instructions for safety and usage. The canister labeling, directions for use, safety and warning statements relating to the Vapocoolant Mist are equivalent to the Primary predicate (Gebauer) and Predicate 1(Ouchless).

V. PERFORMANCE DATA.

Biocompatibility

The subject device has Indirect Contact. The gas blend is approved for this clinical use, application and methodology. The materials are compatible.

Chemical Composition Confirmation

The subject device, Primary predicate and Predicate 1 are composed of identical aerosols profiles, which is 245fa (1,1,1,3,3-Pentafluoropropane) at 95% of the total and 134a (1,1,1,2-Tetrafluoroethane) at 5% of the total mixture.

The subject device's aerosol is checked and verified upon receipt from the aerosol supplier to ensure the same chemical profile as the Primary predicate (Gebauer).

Structural and Parts Composition

Engineering verification measurements were taken, and visual inspections were made to determine the Vapocoolashot Mist canisters, valves, and caps were equivalent to the Primary predicate (Gebauer).

Engineering verification measurements were taken, and visual inspections were made to determine the Vapocoolshot was equivalent to the Predicate 1(Ouchless).

Directions for Use (Clinical Use) Application and Methodology:

All key elements of the Directions for Use (DFU) between the Primary predicate (Gebauer), Predicate 1 and the subject device are similar, intended use, precaution statements, warning statements, and contraindication statements, and treatment regimen. No significant differences exist between the Predicate 1 (Ouchless) and subject device's Directions for

{8}------------------------------------------------

Use.

Side-by-Side Temperature & Output Bench Testing

Tests were selected and performed to ensure the subject device's output performances are similar as a whole with its Primary Predicate (Gebauer).

Comparative performance testing was conducted as it related to temperature output at the application surface and volume dispensed per actuation/total and equivalent results were demonstrated.

Stability Protocol and Shelf Life Testing

A stability protocol was developed to ensure that the identity, strength, quality, and purity of the product are maintained throughout its labeled dating period. Testing assessments were conducted under controlled conditions at room temperature and under accelerated conditions. All requirements were confirmed to meet established acceptance criteria.

Summary

Based on the performance evaluations conducted, Vapocoolshot Mist was found to be safe and effective to the Primary predicate (Gebauer).

VI. CONCLUSION

Based on the information provided in the submission, it is concluded that the Vapocoolshot Mist is safe and effective for its intended use and has demonstrated similar to the Primary predicated (Gebauer). The Vapocoolshot as a unit does not raise additional questions of safety, efficacy and effectiveness different from the Predicate 1 (Ouchless).

N/A