K Number
K172028

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-27

(114 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, mucous membranes and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture, IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.

Device Description

Gebauer's Pain Ease (Mist and Medium Spray) Topical Anesthetic Skin Refrigerant is a prescription device designed to deliver HFC 245fa high purity (1,1,1,3,3-Pentafluoropropane) and HFC 134a pharmaceutical grade (1,1,1,2-Tetrafluoroethane) in a mist and medium spray. This mixture self-propels itself from the delivery system. which is designed to account for its low vapor pressure. The device delivery system is specifically designed to deliver a medium and mist spray of the Gebauer's Pain Ease (Mist Spray and Stream Spray) mixture. The medium and mist spray is an appropriate mode of application when users follow directions for use, cooling the skin through rapid evaporation of the non-medicated volatile propellants. The new device, Gebauer's Pain Ease, is identical in all aspects to the predicate device, Gebauer's Pain Ease (Mist Spray and Medium Spray) 510(k) K032671, except that the product can be applied to the skin for pre-injection anesthesia by cotton swab or gauze on intact skin. (If the skin is breached use this application method only with STERILE, disposable cotton balls, cotton swabs or gauze. The cotton balls, cotton swabs or gauze are not supplied with the device.)

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray)." The purpose of the submission is to expand the approved application methods for the device. The document claims substantial equivalence to a previously cleared predicate device (K032671).

Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the traditional sense of numerical thresholds for performance metrics. Instead, it aims to demonstrate substantial equivalence to a predicate device. The performance is assessed by comparing key characteristics of the subject device (Gebauer's Pain Ease with expanded application methods) to the predicate device (Gebauer's Pain Ease, K032671).

The core "performance" reported is related to temperature and output, specifically that the subject device's profile is "similar" to the predicate.

Acceptance Criterion (Implied for Substantial Equivalence to Predicate)Reported Device Performance (Subject Device)
Formulation: Identical to predicate deviceIdentical to predicate device (HFC 245fa high purity, HFC 134a pharmaceutical grade)
Delivery System: Identical to predicate deviceIdentical to predicate device (Pressurized dispensing container, aerosol can, valve, actuator, standard aerosol nozzle technology)
Packaging: Identical to predicate deviceIdentical to predicate device
Intended Use: Same as predicate device except for expanded application methodSame as predicate device (topical application to skin, mucous membranes, minor open wounds; controls pain for minor surgical procedures, injections, minor sports injuries; Medium Spray for myofascial pain)
Product Fill Volume: Same as predicate device3.5 oz. (103.5 mL) - Same as predicate
Vapocoolant Composition: Same as predicate device1,1,1,3,3 Pentafluoropropane (HFC 245fa high purity) and 1,1,1,2 Tetrafluoroethane (HFC 134a pharmaceutical grade) - Same as predicate
Energy Delivered: Same as predicate deviceThermal energy via Refrigerant Spray - Same as predicate
Vapocoolant Discharge Method: Same as predicate deviceDepress the actuator to release the vapocoolant - Same as predicate
Environmental Compatibility: Non-FlammableNon-Flammable - Same as predicate
Mechanical Safety: Positive shut-off releasePositive shut-off release - Same as predicate
Manufacturing Environment: Controlled EnvironmentControlled Environment - Same as predicate
Microbial Limits Testing: In accordance with USP <61> and <62>In accordance with USP <61> and <62> - Same as predicate
Biocompatibility Testing: In accordance with ISO 10993 for dermal irritation, sensitization, cytotoxicity, oral mucosal irritation, acute dermal toxicityIn accordance with ISO 10993 for dermal irritation, sensitization, cytotoxicity, oral mucosal irritation, acute dermal toxicity - Same as predicate
Boiling Point: 44.6°F (7.0°C)44.6°F (7.0°C) - Same as predicate
Storage Temperature: Do not store above 50°C (120°F)Do not store above 50°C (120°F) - Same as predicate
Shelf Life: 3 years3 years - Same as predicate
Temperature & Output Profile: Similar when applied via expanded method (cotton ball/swab/gauze) compared to direct spray of predicate"safety and effectiveness profile that is similar to the predicate device" based on "temperature data generated" and output of the device

2. Sample size used for the test set and the data provenance

The document states: "Comparative testing was conducted to demonstrate equivalence between direct topical application via spraying and topical application using cotton ball, cotton swab or gauze." It refers to this as "Side-by-Side Temperature & Output Bench Testing."

  • Sample Size: The document does not specify the sample size for this bench testing. It simply states "tests were selected and performed."
  • Data Provenance: The data is from bench testing, meaning it was conducted in a controlled laboratory environment to compare the physical characteristics of the devices. No information is provided regarding the country of origin, nor whether it was retrospective or prospective in the clinical sense, as it was a bench study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the study described is a bench test comparison of physical parameters (temperature and output), not a clinical study involving human assessment or expert judgment to establish a "ground truth" for a medical condition.

4. Adjudication method for the test set

This information is not applicable for the same reason as point 3. There was no "ground truth" adjudicated by experts for a clinical outcome.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This is a 510(k) submission for a topical anesthetic, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical product (vapocoolant spray), not an algorithm or AI system.

7. The type of ground truth used

As this was a bench test comparing physical characteristics, the "ground truth" was essentially instrumental measurements of temperature and output. There was no expert consensus, pathology, or outcomes data used to establish a ground truth for a disease or condition.

8. The sample size for the training set

This is not applicable. The device is a physical product, not a machine learning model, so there is no training set in the context of AI/ML.

9. How the ground truth for the training set was established

This is not applicable as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 27, 2017

Gebauer Company Brittney Schmies Regulatory Affairs Specialist 4444 East 153rd Street Cleveland, Ohio 44128

Re: K172028

Trade/Device Name: Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray) Regulatory Class: Unclassified Product Code: MLY Dated: September 25, 2017 Received: September 26, 2017

Dear Ms. Schmies:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Vivek J. Pinto -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172028

Device Name

Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray)

Indications for Use (Describe)

Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, mucous membranes and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture, IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a black and white drawing of a spiral shape. The spiral has a hole in the center and a small extension on the right side. The spiral is filled with small dots, giving it a textured appearance. There is also a small circle with a dot in the center to the right of the spiral.

GEBAUER COMPANY

510(k) Summary - K172028

This 510(k) Summary is being submitted in accordance of 21 CFR Part 807.92.

I. SUBMITTER

Owner:Gebauer Company4444 East 153rd StreetCleveland, OH 44128(216) 581-3030
Contact Person:Brittney SchmiesGebauer Company4444 East 153rd StreetCleveland, OH 44128

Date Prepared:

October 27, 2017

(216) 581-3030

II. DEVICE

Trade Name:

Skin Refrigerant (Mist Spray and Medium Spray)

Gebauer's Pain Ease Topical Anesthetic

Refrigerant Topical, Vapocoolant

Common Name:

Classification Name:

Product Code:

MLY

Cold Spray

III. PREDICATE DEVICE:

Primary:Gebauer's Pain Ease (Mist Spray andMedium Spray)
K032671
Legally marketed medical device

{4}------------------------------------------------

IV. DESCRIPTION

Gebauer's Pain Ease (Mist and Medium Spray) Topical Anesthetic Skin Refrigerant is a prescription device designed to deliver HFC 245fa high purity (1,1,1,3,3-Pentafluoropropane) and HFC 134a pharmaceutical grade (1,1,1,2-Tetrafluoroethane) in a mist and medium spray. This mixture self-propels itself from the delivery system. which is designed to account for its low vapor pressure. The device delivery system is specifically designed to deliver a medium and mist spray of the Gebauer's Pain Ease (Mist Spray and Stream Spray) mixture. The medium and mist spray is an appropriate mode of application when users follow directions for use, cooling the skin through rapid evaporation of the non-medicated volatile propellants. The new device, Gebauer's Pain Ease, is identical in all aspects to the predicate device, Gebauer's Pain Ease (Mist Spray and Medium Spray) 510(k) K032671, except that the product can be applied to the skin for pre-injection anesthesia by cotton swab or gauze on intact skin. (If the skin is breached use this application method only with STERILE, disposable cotton balls, cotton swabs or gauze. The cotton balls, cotton swabs or gauze are not supplied with the device.) Both the new and predicate devices are indicated for use to control pain for pre-injection anesthesia, minor surgery, and minor sports injuries. The Medium Spray for the new device, like the predicate medium spray device, is also indicated for the management of myofascial pain caused by trigger points. restricted motion and muscle tension by using the Spray and Stretch® technique.

V. INDICATIONS FOR USE

The indications for Use are the same as the predicate device.

Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, intact mucous membranes (oral cavity, nasal passage ways and the lips) and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture. IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.

VI. - TECHNICAL SUMMARY

As with the predicate device, the cooling action experienced by the patient is caused by the evaporation of the chemical mixture from the patient's skin. The user applies pressure to the nozzle to dispense the aerosol product onto the skin, or saturates a

{5}------------------------------------------------

cotton ball, cotton swab or gauze with the product and applies to the skin. The material is contained in a can, filled under pressure, and dispensed using standard aerosol nozzle technology. See table below for the comparison between the subject device and the predicate:

Comparison Chart – Technological Characteristics
Trade NameGebauer's Pain EasePredicateGebauer's Pain EaseSubject Device
Product DesignPressurized dispensingcontainer which includesthe vapocoolant, aerosolcan, valve and actuator.Pressurized dispensingcontainer which includesthe vapocoolant, aerosolcan, valve and actuator.
Intended UseA vapocoolant (skinrefrigerant) intended fortopical application toskin, mucousmembranes and minoropen wounds.Gebauer's Pain Easecontrols pain associatedwith minor surgicalprocedures (such aslancing boils, incisions,drainage of smallabscesses, and sutures),injections(venipuncture, IV starts,cosmetic procedures)and the temporary reliefof minor sports injuries(sprains, bruising, cutsand abrasions). TheMedium Spray is alsointended for thetreatment of myofascialpain caused by triggerpoints, restricted motionand muscle tension.A vapocoolant (skinrefrigerant) intended fortopical application toskin, mucousmembranes and minoropen wounds.Gebauer's Pain Easecontrols pain associatedwith minor surgicalprocedures (such aslancing boils, incisions,drainage of smallabscesses, and sutures),injections(venipuncture, IV starts,cosmetic procedures)and the temporary reliefof minor sports injuries(sprains, bruising, cutsand abrasions). TheMedium Spray is alsointended for thetreatment of myofascialpain caused by triggerpoints, restricted motionand muscle tension.
Product Fill Volume3.5 oz. (103.5 mL)3.5 oz. (103.5 mL)
Vapocoolant Composition1,1,1,3,3Pentafluoropropane (HFC245fa high purity) and1,1,1,2 Tetrafluoroethane1,1,1,3,3Pentafluoropropane (HFC245fa high purity) and1,1,1,2 Tetrafluoroethane
(HFC 134a pharmaceutical grade).(HFC 134a pharmaceutical grade).
Energy DeliveredThermal energy viaRefrigerant Spray.Thermal energy viaRefrigerant Spray.
Vapocoolant DischargeMethodDepress the actuator torelease the vapocoolant.Depress the actuator torelease the vapocoolant
Method of ApplicationDirect spray application tointact skin, intact oralmucous membranes andminor open wounds.Direct spray application tointact skin, intact oralmucous membranes, andminor open wounds, orapplication via cotton ball,cotton swab or gauze onintact skin only. If the skinis breached this applicationmethod is to be used onlywith STERILE, disposablecotton balls, cotton swabsor gauze.
EnvironmentalCompatibilityNon-Flammable.Non-Flammable.
Mechanical SafetyMechanism has positiveshut-off release.Mechanism has positiveshut-off release.
ManufacturingEnvironmentControlled EnvironmentControlled Environment
Microbial Limits TestingTested in accordance withUSP <61> and <62>.Tested in accordance withUSP <61> and <62>.
Biocompatibility TestingIn accordance with ISO10993 for dermal irritation,sensitization, cytotoxicity,oral mucosal irritation,acute dermal toxicity.In accordance with ISO10993 for dermal irritation,sensitization, cytotoxicity,oral mucosal irritation,acute dermal toxicity.
Target PopulationGeneralAs the safety for use of thedevice in pediatric patientshas not been established, itis recommended the deviceshould not be used onpatients under four withoutconsultation of apediatrician.
Boiling Point44.6°F (7.0°C)44.6°F (7.0°C)
Storage TemperatureDo not store at temperaturesabove 50°C (120°F)Do not store at temperaturesabove 50°C (120°F)
Shelf Life3 years3 years

{6}------------------------------------------------

{7}------------------------------------------------

DETERMINATION OF SUBSTANTIAL EQUIVELANCE VII.

This premarket notification 510(k) is being submitted to expand the application methods in the instructions for use for Gebauer's Pain Ease, Mist and Medium Stream. There is demonstrated equivalency in basic product design and technology, in indications for use, target population, and risk factors.

As stated above, the new device is identical in formulation, delivery system and packaging to the predicate device Gebauer's Pain Ease, Mist and Medium Stream previously cleared for market under 510(k) K032671. The new device has expanded instructions for use to apply the product by either spraying directly on the skin or saturating a cotton ball, cotton swab or gauze on intact skin. (If the skin is breached use this application method only with STERILE, disposable cotton balls, cotton swabs or gauze. The cotton balls, cotton swabs or gauze are not supplied with the device.) Both the new and predicate devices are indicated for use to temporarily control pain associated with pre-injection anesthesia, minor surgery and minor sports injuries. The medium stream spray for both the predicate and new device has the identical indication for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.

VIII. PERFORMANCE DATA

The predicate and subject devices use the same materials, design and energy source and no technological differences exist. Tests were selected and performed to ensure the subject device's performance, with the new method of application, is equivalent as the predicate. To demonstrate equivalence the following test was performed:

Side-by-Side Temperature & Output Bench Testing

Comparative testing was conducted to demonstrate equivalence between direct topical application via spraving and topical application using cotton ball, cotton swab or gauze. The testing was conducted to determine the temperature and output of the predicate and subject devices

Summarv

Based on the temperature data generated, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device.

CONCLUSION IX.

Based on the information above and within this submission, it is concluded that the subiect Pain Ease device is safe and effective for its intended use and is substantially equivalent to the predicate device.

N/A