K Number
K193241

Validate with FDA (Live)

Date Cleared
2020-01-26

(62 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The uMI 550 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.

This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Device Description

The uMI 550 PET/CT system is a combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanner. This system is intended to be operated by qualified healthcare professionals for performing diagnostic imaging examinations. The spatial alignment and precise image registration between PET and CT ensure the PET and CT images of the same region can be fused accurately for reading. PET measures the distribution of PET radiopharmaceuticals inside the human body quantitatively. CT produces the anatomical information of the same scanned region, and provides accurate localization for the findings in the PET images. The attenuation information contained in the CT images can be utilized in the PET image reconstruction to ensure quantitation accuracy.

The uMI 550 PET/CT system also includes a patient table, a workstation with associated software installed. The software is used for patient management, data management, scan control, image reconstruction and image reading. All patient images produced by the system conform to the DICOM 3.0 standard.

The uMI 550 PET/CT has been previously cleared by FDA via K182237. The modifications performed on the uMI 550 PET/CT (K182237) in this submission are due to the addition of HYPER Iterative and Auto-Planbox function. Meanwhile the sensitivity specification has been updated. HYPER Iterative allows more iterations while remains the image noise at an acceptable level by incorporating a noise control term into the objective function. HYPER Iterative can achieve high image contrast and quantification accuracy. Auto-Planbox plan the scan range by recognizing body parts on CT scout image. It locates the different body parts based on anatomy characteristic. The scan range is generated to cover the whole body parts according to protocol selection. This function will simplify scanning process, which will be convenient for user operation.

AI/ML Overview

The information provided describes the acceptance criteria and study for the uMI 550 PET/CT system, focusing on the HYPER Iterative function and Auto-Planbox function, and an update to the sensitivity specification.

Here's the breakdown of the acceptance criteria and the studies that prove the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document highlights changes and comparisons to a predicate device (uMI 550, K182237). The acceptance criteria are largely implied by the "Same" remark for most specifications, indicating the device must meet the predicate's performance. The specific changes are detailed below:

FeatureAcceptance Criteria (Proposed Device)Reported Device Performance (Previous Submission / Predicate Reference)Remark
Sensitivity>=9cps/kBq @0cm, >=9cps/kBq @10cm>=10cps/kBq @0cm, >=10cps/kBq @10cmUpdated specification: lower sensitivity claimed, justified by updated activity measurement factor and manageable by slightly longer scan time without affecting safety/effectiveness.
HYPER IterativeAvailableNot availableNew function added. Performance verified through bench testing and clinical image evaluation.
Auto-PlanboxAvailableNot availableNew function added. Performance verified through bench testing.
Other PET SpecsSame as Predicate(See document for full list like Spatial Resolution, NECR, etc.)Passed non-clinical testing
CT SpecsSame as Predicate(See document for full list like Image Resolution, Image Noise, etc.)Passed non-clinical testing
Electrical SafetyConforms to ANSI AAMI ES60601-1(Implicit conformance to standard)Non-clinical testing conducted.
EMCConforms to IEC 60601-1-2(Implicit conformance to standard)Non-clinical testing conducted.
BiocompatibilityConforms to ISO 10993-5, ISO 10993-10(Implicit conformance to standard)Patient contact materials tested.
Software V&VMeets all software specificationsN/A (Internal V&V)Testing results show all specifications met.
CybersecurityConforms to guidance documentN/A (Internal implementation)Conforms through process implementation.

2. Sample Size Used for the Test Set and Data Provenance

  • HYPER Iterative Clinical Image Evaluation:

    • Sample Size: 20 retrospectively collected clinical cases.
    • Data Provenance: Retrospectively collected; country of origin is not specified but implied to be from the manufacturer's clinical sites, likely in China given the company's location.
  • Auto-Planbox Bench Testing:

    • Sample Size: 16 group scout images.
    • Data Provenance: Not explicitly stated whether these were clinical or simulated, but "scout images" are typical inputs for this function. Likely internal testing data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Clinical Image Evaluation (for HYPER Iterative):
    • Number of Experts: 3
    • Qualifications: American board-certified nuclear medicine physicians.

4. Adjudication Method for the Test Set

  • Clinical Image Evaluation (for HYPER Iterative): The document states that each image was read by the three physicians who provided an assessment. It does not explicitly mention an adjudication method for conflicting opinions (e.g., 2+1, 3+1). It describes independent assessments of image contrast and image quality.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • A reader study was done for the HYPER Iterative function. This was a comparative study between images reconstructed with OSEM (without HYPER Iterative) and HYPER Iterative.
  • Without AI vs. With AI Assistance: This study does not appear to be an AI-assisted interpretation study in the typical sense of human readers interacting with an AI output. Instead, it compares the output of two different reconstruction algorithms (OSEM vs. HYPER Iterative) as interpreted by human readers. The HYPER Iterative algorithm itself is an advanced reconstruction technique, which implicitly "assists" by improving image quality.
  • Effect Size: The study results indicated that "HYPER Iterative has better image contrast than OSEM and the image quality is sufficient for clinical diagnosis." No quantitative effect size (e.g., AUC improvement, percentage increase in diagnostic accuracy) is provided in this summary. The assessment was qualitative using 3-point and 5-point scales.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • For the HYPER Iterative function, bench testing was performed focusing on quantitative metrics and visual comparisons without human interpretation as the primary endpoint:
    • Quantification accuracy and signal to noise ratio (SNR) using the NEMA IO phantom.
    • Effectiveness for large weight patient (likely visual assessment of image quality, not human-in-the-loop diagnostic performance).
    • Image contrast improvement on brain imaging (likely quantitative metrics or visual assessment without diagnostic interpretation by experts).
  • For the Auto-Planbox function, performance evaluation was done through bench testing where the system's recognition of body parts was compared against manual annotation. This is a standalone technical validation.

7. The Type of Ground Truth Used

  • HYPER Iterative Clinical Image Evaluation: The ground truth for comparative image quality and contrast was established by the subjective assessment of 3 American board-certified nuclear medicine physicians (expert consensus/rating).
  • HYPER Iterative Bench Testing (NEMA phantom): Ground truth for quantification accuracy and SNR was established using a NEMA IO phantom, which has known physical characteristics.
  • Auto-Planbox Bench Testing: Ground truth for body part recognition was established by "manual annotation" (expert annotation).

8. The Sample Size for the Training Set

  • The document does not provide any information regarding the sample size of the training set for the HYPER Iterative or Auto-Planbox functions. These are advanced algorithms; their development typically involves large datasets. However, this specific 510(k) summary focuses on the validation of the modifications, not the initial development.

9. How the Ground Truth for the Training Set Was Established

  • As the training set sample size is not provided, the method for establishing its ground truth is also not described in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a larger font and the word "Administration" in a smaller font.

Shanghai United Imaging Healthcare Co., Ltd % Shumei Wang QM & RA VP No. 2258 Chengbei Rd., Jiading Industrial District Shanghai, 201807 CHINA

January 26, 2020

Re: K193241

Trade/Device Name: uMI 550 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: November 22, 2019 Received: November 25, 2019

Dear Shumei Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193241

Device Name uMI 550

Indications for Use (Describe)

The uMI 550 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.

This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark teal in color. The logo is simple and modern in design.

510(k) SUMMARY

    1. Date of Preparation: January 22, 2020

2. Sponsor Identification

Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China

Contact Person: Shumei Wang Position: QM&RA VP Tel: +86-021-67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com

3. Identification of Proposed Device

Trade Name: uMI 550 Common Name: Emission Computed Tomography System

Model(s): uMI 550

Regulatory Information Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK, Review Panel: Radiology

4. Identification of Predicate Device(s)

Predicate Device

510(k) Number: K182237 Device Name: uMI 550 Model(s): uMI 550

Regulatory Information Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK, Review Panel: Radiology

5. Device Description

The uMI 550 PET/CT system is a combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanner. This system is intended

1 of 12

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the text and the "U" shape is a dark gray.

to be operated by qualified healthcare professionals for performing diagnostic imaging examinations. The spatial alignment and precise image registration between PET and CT ensure the PET and CT images of the same region can be fused accurately for reading. PET measures the distribution of PET radiopharmaceuticals inside the human body quantitatively. CT produces the anatomical information of the same scanned region, and provides accurate localization for the findings in the PET images. The attenuation information contained in the CT images can be utilized in the PET image reconstruction to ensure quantitation accuracy.

The uMI 550 PET/CT system also includes a patient table, a workstation with associated software installed. The software is used for patient management, data management, scan control, image reconstruction and image reading. All patient images produced by the system conform to the DICOM 3.0 standard.

The uMI 550 PET/CT has been previously cleared by FDA via K182237. The modifications performed on the uMI 550 PET/CT (K182237) in this submission are due to the addition of HYPER Iterative and Auto-Planbox function. Meanwhile the sensitivity specification has been updated. HYPER Iterative allows more iterations while remains the image noise at an acceptable level by incorporating a noise control term into the objective function. HYPER Iterative can achieve high image contrast and quantification accuracy. Auto-Planbox plan the scan range by recognizing body parts on CT scout image. It locates the different body parts based on anatomy characteristic. The scan range is generated to cover the whole body parts according to protocol selection. This function will simplify scanning process, which will be convenient for user operation.

The modifications are listed as follow:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized, dark teal-colored symbol that resembles a shield with a white line running horizontally through the center. The logo is simple and modern, conveying a sense of strength and innovation.

  • Introduce HYPER Iterative function. >
  • Introduce Auto-Planbox function.

  • Update sensitivity specification.

uMI 550 PET/CTprevioussubmission(K182237)uMI 550 PET/CT thissubmissionRemark
Sensitivity>/=10cps/kBq @0cm>/=10cps/kBq @10cm>/=9cps/kBq @0cm>/=9cps/kBq @10cmSensitivity specificationhas been updated.
HYPERIterativeNot availableAvailableNew function has beenadded
Auto-PlanboxNot availableAvailableNew function has beenadded

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in bold, dark blue font. To the right of the words is a stylized dark blue symbol that resembles the letter "U" with a vertical line through the center.

6. Indications for Use

The uMI 550 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.

This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

7. Comparison of Technological Characteristics with the Predicate Devices

The technology characteristics of the modified uMI 550. reflected in this 510(k) submission, do not alter the scientific technology of the devices and are substantially equivalent to those of the predicate devices.

Table 1 below provides a comparison of the technological characteristics of the proposed device in comparison to the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is also in a bold, sans-serif font. The color of the logo is a dark teal.

ITEMTable 1 Comparison of Technological CharacteristicsProposed DeviceuMI 550Predicated DeviceuMI 550(K182237)Remark
General
Product CodeKPS, JAKKPS, JAKSame
Regulation No.21 CFR 892.120021 CFR 892.1200Same
ClassClass IIClass IISame
Intended UseThe uMI 550 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology andThe uMI 550 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology andSame
Specifications
PET Specifications
Scintillator materialLYSOLYSOSame
Scintillator dimensions2.76mm×2.76mm×16.3mm2.76mm×2.76mm×16.3mmSame
Number of detector rings8484Same
Number of image planes167167Same
Axial field of view24cm24cmSame
Image matrix sizes128×128, 150×150, 192×192, 256×256, 512×512, 600×600128×128, 150×150, 192×192, 256×256, 512×512, 600×600Same
Coincidence window4.0ns4.0nsSame
Spatial ResolutionAxial FWHM@1 cm</=3.5mm</=3.5mmSame
Transaxial FWHM@1 cm</=3.5mm</=3.5mm
Axial FWHM@1 0cm</=4.0mm</=4.0mmSame
Transaxial FWHM@1 0cm</=4.0mm</=4.0mm
Sensitivity>/=9cps/kBq>/=10cps/kBqNote No.1
NECR Peak Value>/=90 kcps@13kBq/cc>/=90 kcps@13kBq/ccSame
Peak True Count Rate>/=300kcps@27kBq/cc>/=300kcps@27kBq/ccSame
PET Scatter Fraction</=0.44</=0.44Same
Count Rate Bias</=±5%</=±5%Same
Image QualityAcceptance ValueSame
SphereContrast RecoveryBackground variability
10 mm>=45.0%<7.5%
13 mm>=55.0%<7.0%
17 mm>=65.0%<7.0%
22 mm$\ge 72.0%$<7.0%22 mm$\ge 72.0%$<7.0%
28 mm$\ge 65.0%$<7.0%28 mm$\ge 65.0%$<7.0%
37 mm$\ge 70.0%$<7.0%37 mm$\ge 70.0%$<7.0%
LungResidual$\le 16.0%$<7.0%LungResidual$\le 16.0%$<7.0%
CT Specifications
Scan RegimeContinuous RotationContinuous RotationSame
Scan ModesTopoAxial ScanHelical ScanTopoAxial ScanHelical ScanSame
Detector MaterialSolid-state GOSSolid-state GOSSame
Z-plane coverage22mm22mmSame
Size of detectorelement in Z-plane0.55mm0.55mmSame
Number ofelement per row864864Same
Number ofdetector row4040Same
Maximum slicesgenerated perrotation(multi-slicecapability)8080Same
Minimum slicethickness0.55mm0.55mmSame
Maximumsampling rateUp to 4800 views per360°Up to 4800 views per360°Same
Tube anodestorage capacity5.3MHU5.3MHUSame
Maximum coolingrate815 kHU/min815 kHU/minSame
Focal spot size0.5x1.0mm1.0x1.0mm0.5x1.0mm1.0x1.0mmSame
Power50 kW50 kWSame
mA Range10-420mA10-420mASame
kV Settings70, 80, 100, 120, 14070, 80, 100, 120, 140Same
Aperture700mm700mmSame
Rotation speedUp to 0.5 sec per 360°rotationUp to 0.5 sec per 360°rotationSame

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING", both in a bold, sans-serif font. To the right of the text is a stylized symbol that resembles the letter "U" with a vertical line running through the center, creating a visual effect of a shield or emblem. The color of the text and symbol is a dark teal or blue-green.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED" is stacked on top of the text "IMAGING". To the right of the text is a stylized letter "U" that is split in the middle by a horizontal line. The logo is in a dark gray color.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in bold, sans-serif font. To the right of the words is a stylized, dark blue symbol that resembles a shield with a vertical line running through the center.

16.5±1.7 lp/cm @MTF10%16.5±1.7 lp/cm @MTF10%Same
11.5±1.2 lp/cm @MTF50%11.5±1.2 lp/cm @MTF50%
Image Noise3.0±0.5 HU at 120 kV, 5 mm slice thickness,CTDIvol 28.9 mGy3.0±0.5 HU at 120 kV, 5 mm slice thickness,CTDIvol 28.9 mGySame
CT NumberDisplay Range-1024 ~+8191 HU-1024 ~+8191 HUSame
Scan Field ofViewUp to 500 mmUp to 500 mmSame
ReconstructionField of View40mm-500mm40mm-600mm withextend FOV40mm-500mm40mm-600mm withextend FOVSame
Image MatrixUp to 1024 x 1024Up to 1024 x 1024Same
Reconstructedslice thickness0.55mm,1.1mm,2.2mm,5.5mm,11mm (axial)0.55-10mm(helical)0.55mm,1.1mm,2.2mm,5.5mm,11mm (axial)0.55-10mm(helical)Same
Pitch0.1~2.00.1~2.0Same
Maximumcontinuousexposure timeUp to 100secondsUp to 100secondsSame
Others Specifications
Display24inch, 1920 x 120024inch, 1920 x 1200Same
Horizontal motionrange2160 mm2160 mmSame
Vertical motionrange550 mm-945 mm fromthe floor550 mm-945 mm fromthe floorSame
Maximumhorizontalscannable rangeStandard 1700mm withup to full 2000mm.Standard 1700mm withup to full 2000mm.Same
Table HorizontalSpeedUp to 200mm/secUp to 200mm/secSame
Vertical speedUp to 40 mm/secUp to 40 mm/secSame
Table HorizontalPosition accuracy±0.25mm±0.25mmSame
Table Maximumtable load250kg250kgSame
Safety
Electrical SafetyANSI AAMI ES60601-1:2005/(R)2012 andA1:2012,C1:2009/(R)2012 andA2:2010/(R)2012ANSI AAMI ES60601-1:2005/(R)2012 andA1:2012,C1:2009/(R)2012 andA2:2010/(R)2012Same
EMCIEC 60601-1-2 Medicalelectrical equipment -Part 1-2: Generalrequirements for basicsafety and essentialIEC 60601-1-2 Medicalelectrical equipment -Part 1-2: Generalrequirements for basicsafety and essentialSame
performance - CollateralStandard:Electromagneticdisturbances -Requirements and testsperformance - CollateralStandard:Electromagneticdisturbances -Requirements and tests
BiocompatibilityPatient ContactMaterials were testedand demonstrated nocytotoxicity (ISO10993-5), no evidencefor irritation andsensitization (ISO10993-10).Patient ContactMaterials were testedand demonstrated nocytotoxicity (ISO10993-5), no evidencefor irritation andsensitization (ISO10993-10).Same
ClinicalComparing with the predicate device, the difference of clinicalimage is due to the addition of HYPER Iterative function. Thecomparison sample images showed improved image contrast,resolution and increased SNR.
Justification
Note IDJustification
Note 1Since activity measurement factor has been updated according towell-counter vendor's notice. The sensitivity specification, whichdepends on activity measurement, shall be updated accordingly.Provide the lower system sensitivity for PET imaging whichwould be compensated by a little longer scan time. The differenceof the system sensitivity from the two devices does not affect thesafety and effectiveness.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern.

Performance Data 8.

The following performance data were provided in support of the substantial equivalence determination.

Non-Clinical Testing

Non-clinical testing including dosimetry and image performance tests were conducted for the uMI 550 during the product development.

UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the uMI 550 in accordance with the following standards:

  • A ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005. MOD)
  • A IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark gray. The logo is simple and modern.

  • IEC 60601-2-44 Edition 3.2: 2016 Medical electrical equipment Part 2-44: A Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography
  • A IEC 60825-1 Edition 2.0 2007-03 Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)]

Product Particular Standards

  • IEC 60601-1-3 Edition 2.1 2013-04 Medical electrical equipment Part 1-3: A General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
  • A IEC 61223-3-5 First edition 2004-08 Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests -Imaging performance of computed tomography X-ray equipment [Including: Technical Corrigendum 1 (2006)]
  • A NEMA XR 25-2010: Computed Tomography Dose Check
  • A NEMA XR 28-2013 Supplemental Requirements for User Information and System Function Related to Dose in CT
  • NEMA XR 29-2013: Standard Attributes on CT Equipment Related to Dose Optimization and Management

Performance Verification

  • A NEMA NU 2-2012 Performance Measurements of Positron Emission Tomographs
  • Clinical Image Evaluation for sample clinical images

  • A AEC Test for AEC performance study
  • A Performance Evaluation for HYPER Iterative function
  • A Performance Evaluation for Auto-Planbox function

Bench testing was performed to support substantial equivalence and the product performance claims. For HYPER Iterative function, the evaluation and analysis used the identical raw datasets obtained on UIH's uMI 550 system and then applied both OSEM and HYPER Iterative. The resultant images were compared for:

  • Ouantification accuracy and signal to noise ratio (SNR) using the NEMA IO phantom
  • Lesion detectability using patient data
  • Influence of regularization parameter selection using patient data
  • Effectiveness for large weight patient
  • Image contrast improvement on brain imaging

Bench testing showed that HYPER Iterative can be used in PET body and brain imaging, can provide the flexibility to achieve user-preferable image by adjusting the regularization parameter, and its images have demonstrated improved image quality, higher quantification accuracy, and improved image SNR than OSEM images.

For Auto-Planbox function, the evaluation and analysis used 16 group scout images and compared Auto-Planbox recognition with manual annotation body parts position. Bench testing results showed that Auto-Planbox can locate head center position, body

10 of 12

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is a dark teal color. The logo is simple and modern.

start position and body end position with proper accuracy and can assist operator to plan the scan range.

Software

  • NEMA PS 3.1-3.20(2011): Digital Imaging and Communications in Medicine A (DICOM)
  • A IEC 62304: Medical Device Software - software life cycle process
  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices

  • A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

Biocompatibility

  • ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices A - Part 10: Tests for irritation and skin sensitization
  • A ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity

Other Standards and Guidances

  • ISO 14971: Medical Devices Application of risk management to medical A devices
  • Code of Federal Regulations, Title 21, Part 820 Quality System Regulation A
  • Code of Federal Regulations, Title 21, Subchapter J Radiological Health A
  • A Laser Products - Conformance with IEC 60825-1 and IEC 60601-2-22; Guidance for Industry and FDA Staff (Laser Notice No. 50)
  • A Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography

Software Verification and Validation

Software documentation for a Moderate Level of Concern software per FDA' Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" is included as a part of this submission.

The risk analysis was completed and risk control was implemented to mitigate identified hazards. The testing results show that all the software specifications have met the acceptance criteria. Verification and validation testing of the proposed device was found acceptable to support the claim of substantial equivalence.

UNITED IMAGING HEALTHCARE conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modification, misuse or denial of use, or unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" is included in this submission.

Clinical Image Evaluation

The reader study used a total of 20 retrospectively collected clinical cases. The raw data from each of these cases was reconstructed with both OSEM and HYPER Iterative. Each image was read by 3 American board-certified nuclear medicine

11 of 12

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized letter "U" that is split into two sections by a vertical line. The logo is simple and modern, with a focus on the company name and a unique visual element.

physicians who provided an assessment of image contrast according to 3-point scale ("+" represents better contrast image for small lesion detection; "-"represents worse contrast image for small lesion detection; "="represents similar contrast image for small lesion detection), and image quality according to a 5-point scale (1= images are unreadable and cannot be used for diagnosis, 2=insufficient image quality and may affect diagnosis; 3=image quality is barely ok for diagnosis; 4=image quality is sufficient for diagnosis; 5=wonderful image quality for diagnosis). The results of the study indicated that HYPER Iterative has better image contrast than OSEM and the image quality is sufficient for clinical diagnosis.

Clinical Testing

No Clinical Study is included in this submission.

9. Conclusions

Based on the comparison and analysis above, the proposed device has same intended use, similar performance, equivalence safety and effeteness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.