K Number
K192907

Validate with FDA (Live)

Date Cleared
2019-11-26

(42 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Advocate PTA Catheter is intended for balloon dilation of the illac, femoral, popliteal, renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Description

The Advocate PTA Catheter consists of two lumens, a compatible guidewire lumen (0.014", 0.018" or 0.035") extending from guide wire port in the manifold to the catheter distal tip, and an inflation lumen extending from the proximal inflation port in the manifold to the balloon interior. There are various size balloons in the product matrix, ranging from 2.0-12.0mm in diameter and 15-300mm in length. The catheter has two radiopaque marker bands that facilitate visibility and location during the placement and inflation of the balloon. The catheter is inserted through a 4Fr, 5Fr, 6Fr or 7Fr haemostatic or non haemostatic introducer, depending on the model selected.

AI/ML Overview

This document describes the Advocate PTA Catheter and its 510(k) premarket notification (K192907). It addresses a change in manufacturing location and an expansion of the Indications for Use. The relevant information provided focuses on the safety and performance testing related to these changes.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly provide a table of acceptance criteria with corresponding device performance metrics for the expanded indications for use. Instead, it states that "extra Validation testing was done on the device to qualify the expanded indications for use" and lists the types of tests successfully completed. This implies that the tests performed met internal acceptance criteria for demonstrating safety and performance.

Acceptance Criteria (Implied)Reported Device Performance
Device successfully navigates anatomical pathways (AVF)Deliverability (AVF Trackability) - Successfully completed
Device demonstrates sufficient force transmission for advancementDeliverability (Pushability) - Successfully completed

2. Sample Size for the Test Set and Data Provenance

The document does not specify the sample size used for the test set (i.e., the number of catheters or tests performed for Deliverability AVF Trackability and Pushability).

The data provenance is not explicitly stated in terms of country of origin. The study appears to be prospective testing conducted by Merit Medical Systems, Inc. to qualify the expanded indications for use and the manufacturing location change.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

This information is not applicable as the document describes physical performance testing of a medical device, not a diagnostic or AI-based study requiring expert ground truth establishment for a test set.

4. Adjudication Method for the Test Set

This information is not applicable as the document describes physical performance testing of a medical device, not a study requiring adjudication of expert opinions.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This information is not applicable. The document describes the testing of a medical device (PTA Catheter), not an AI algorithm. Therefore, no MRMC study or comparison of human readers with/without AI assistance was performed.

6. Standalone (Algorithm Only) Performance Study

This information is not applicable. The document describes the testing of a medical device (PTA Catheter), not an AI algorithm.

7. Type of Ground Truth Used

This information is not applicable in the context of expert consensus, pathology, or outcomes data. For the physical performance tests, the "ground truth" would be the engineering specifications and expected performance characteristics of the device, assessed through laboratory testing.

8. Sample Size for the Training Set

This information is not applicable as the document describes the testing of a medical device, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as the document describes the testing of a medical device, not an AI algorithm that requires a training set and ground truth establishment.

Summary of Key Information from the Document:

  • Device: Advocate PTA Catheter
  • Purpose of Submission (K192907): Change in Indications for Use and manufacturing location.
  • Expanded Indications for Use: Balloon dilation of iliac, femoral, popliteal, infra-popliteal, renal arteries and treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
  • Testing Performed to Qualify Expanded Indications: Deliverability (AVF Trackability) and Deliverability (Pushability).
  • Outcome of Testing: Successfully completed.
  • Manufacturing Change: Moved from a subcontractor (Arravasc) to Merit Medical in Galway, using the same qualified processes.
  • The document implies that the successful completion of the listed safety and performance tests demonstrates that the device meets the necessary acceptance criteria for the expanded indications. However, it does not provide detailed quantitative acceptance criteria or specific numerical results.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 26, 2019

Merit Medical Systems, Inc. Michael O'Sullivan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland

Re: K192907

Trade/Device Name: Advocate PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: October 10, 2019 Received: October 15, 2019

Dear Mr. O'Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192907

Device Name Advocate PTA Catheter

Indications for Use (Describe)

The Advocate PTA Catheter is intended for balloon dilation of the illac, femoral, popliteal, renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Merit Medical Systems, Inc.
Merit Advocate PTA Catheter
Traditional Premarket Notification 510(k)

GeneralProvisionsSubmitter Name:Merit Medical Systems, Inc.
Address:1600 West Merit ParkwaySouth Jordan, UT 84095
Telephone Number:(+353) 91 703700 (Ext. 3061)
Fax Number:(+353) 91 680104
Contact Person:Mark Mullaney
Registration Number:1721504
Correspondent Name:Merit Medical Ireland Ltd.
Address:Parkmore Business ParkParkmore, Galway, Ireland
Telephone Number:(+353) 91 703700 (Ext. 3223)
Fax Number:(+353) 91 680104
Contact Person:Michael O'Sullivan
Date of Preparation:3 October 2019
Registration Number:9616662
510(k) Number:K192907
SubjectDeviceTrade Name:Advocate PTA Catheter
Common/Usual Name:Percutaneous Catheter
Classification Name:Catheter, Angioplasty, Peripheral,Transluminal
Premarket Notification Predicate:
Trade Name:Advocate PTA Catheter
Classification Name:21 CFR 870.1250 Percutaneous Catheter
Premarket Notification:K173621
Manufacturer:Merit Medical
PredicateDevicePremarket Notification Reference Predicate:
Trade Name:Pirouette 018HP
Classification Name:21 CFR 870.1250 Percutaneous Catheter
Premarket Notification:K172033
Manufacturer:Arravasc

{4}------------------------------------------------

Merit Medical Systems, Inc.
Merit Advocate PTA Catheter
Traditional Premarket Notification 510(k)

ClassificationClass II21 CFR § 870.1250Product code: LITDivision of Cardiovascular Devices
Indications for useThe Advocate PTA Catheter is intended for balloon dilation of theiliac, femoral, popliteal, infra-popliteal, renal arteries and for thetreatment of obstructive lesions of native or synthetic arteriovenousdialysis fistulae.
Intended UseThe Advocate PTA catheter is intended for use to treat vascularstenoses, abnormal narrowing of a blood vessels, in variousclinical settings during endovascular procedures for adult patientpopulations
DeviceDescriptionThe Advocate PTA Catheter consists of two lumens, a compatibleguidewire lumen (0.014", 0.018" or 0.035") extending from guidewire port in the manifold to the catheter distal tip, and an inflationlumen extending from the proximal inflation port in the manifold tothe balloon interior. There are various size balloons in the productmatrix, ranging from 2.0-12.0mm in diameter and 15-300mm inlength. The catheter has two radiopaque marker bands that facilitatevisibility and location during the placement and inflation of theballoon. The catheter is inserted through a 4Fr, 5Fr, 6Fr or 7Frhaemostatic or non haemostatic introducer, depending on the modelselected.

{5}------------------------------------------------

Merit Medical Systems, Inc. Merit Advocate PTA Catheter Traditional Premarket Notification 510(k)

There are no technological differences between the subject and predicate devices. The changes to the device for this submission are as follows

  • Indications for Use Change to: The Advocate PTA Catheter is intended for balloon dilation of the iliac, femoral, popliteal, infra-popliteal, renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
  • Comparison . Removal of the 1.25mm and 1.5mm Balloon diameters from to Predicate the list of applied 0.014 devices
    • Change of Manufacturing location from Subcontractor at . Arravasc in Galway to Merit Medical in Galway, using the same processes as already qualified for the device.

The only difference between the two devices are the indications for use, the Manufacturing location and the removal of Devices with balloon diameters of 1.25mm and 1.5mm.

The existing Safety or Performance testing was not performed for the manufacturing location move as the process used is the same as already qualified by Arravasc Ltd when they acted as a subcontractor manufacturing location for the device. The processes were transferred into Merit Galway as is and were used to manufacture the devices that make up this submission. The process is the same as previously qualified under K173621.

Safety & Performance Tests

However, extra Validation testing was done on the device to qualify the expanded indications for use, that this change is introducing.

The following is a list of testing that was successfully completed:

  • Deliverability AVF Trackability ●
  • Deliverability Pushability .

{6}------------------------------------------------

Merit Medical Systems, Inc.
Merit Advocate PTA Catheter
Traditional Premarket Notification 510(k)

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).