K Number
K192753

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-17

(169 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeuralScan System is intended for the acquisition, display, and storage, of electrical activity of a patient's brain including electroencephalograph (EEG) and Event-related Potentials (ERP), obtained by placing two or more electrodes on the head to aid in diagnosis.

Device Description

The NeuralScan System is comprised of the NeuralScan Amplifier with embedded firmware for acquisition and transmission of physiological signals, an EEG cap, a Subject Response Device, a laptop computer preloaded with the NeuralScan Evoke software, and a charging kit (that consists of a USB cable, clip kit, and wall adapter).

The NeuralScan Evoke software runs in a Windows Operating System environment and controls the 23 channel (21 EEG and 2 bio channels) NeuralScan EEG amplifier via a USB cable or Wi-Fi connection. The software has a graphical user interface that allows the user to input patient information, create new records, conduct studies to collect EEG and ERP data, view live data streams of the laptop display, record data to a file, analyze resultant test data using standard Frequency EEG analysis and EP display methods and print the results. The software includes a mode to measure the cap electrode impedances which is useful for determining if the electrodes are making a good electrical connection with the scalp at each electrode location. Interpretation of the data is the responsibility of the physician as the software does not provide any diagnosis based on the data.

The patient contacting accessories are commercially sourced and used without modification. The device and the accessories are not sterile and are not intended to be sterilized.

AI/ML Overview

The provided text describes the regulatory clearance for the "NeuralScan System" and includes a summary of non-clinical testing performed to demonstrate its substantial equivalence to predicate devices. However, it explicitly states: "No Clinical testing was necessary to determine substantial equivalence."

Therefore, based on the provided document, the device did not undergo a clinical study involving human subjects to prove its performance against acceptance criteria for diagnostic accuracy or clinical effectiveness. The acceptance criteria and performance data are entirely based on bench testing (non-clinical data) for the device's functional integrity as an electroencephalograph (EEG) and Event-related Potentials (ERP) system.

Given this, I can only provide information derived from the non-clinical testing detailed in the document.


Acceptance Criteria and Reported Device Performance (Non-Clinical Study)

The acceptance criteria and performance measurements for the NeuralScan System are based on bench testing to confirm compliance with recognized medical device standards (primarily IEC 60601-2-26 for electroencephalographs). These tests evaluate the hardware and software's ability to accurately acquire, reproduce, and transmit physiological signals.

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

TestAcceptance Criteria (from IEC 60601-2-26)Reported Device PerformanceResult
Accuracy of signal reproductionError < 20% (for 2Hz Triangle wave, 50μV to 500μV)Accurately reproduces EEG signalsPass
Input dynamic range and differential offset voltageAmplitude error < 10% (for 1mV 6Hz Triangle wave with offset)300mV Offset: 1.5% error; 187mV Offset: 1.5% errorPass
Input NoiseMax 6µV peak-to-valleyMaximum noise: 3.4 µVPass
Frequency ResponseOutput amplitude > 71%, < 110% (for 0.5 Hz, 5 Hz, 50 Hz)0.5 Hz: 100%; 5 Hz: 105%; 50 Hz: 76%Pass
Common mode rejectionOutput amplitude < 10mm (= 100µVpp) (for 1Vrms at 60Hz)With 0 DC Offset: 67.3µVpp; With 187mV dc Offset: 68µVppPass

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable in the context of clinical patient data. The "test set" here refers to the specific configurations and conditions used for the bench tests. For each test, specific signal parameters and conditions were applied.
  • Data Provenance: Not applicable as no clinical patient data was used. The data is generated from controlled laboratory bench tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. The "ground truth" for these tests are the precise, known input signals generated by test equipment, and validation against the specified international standards (IEC 60601-2-26). No human experts were involved in establishing the ground truth for these non-clinical tests in the way they would be for diagnostic accuracy studies.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods are relevant for clinical studies where human interpretation of medical images or data is being evaluated. For bench testing, the results are objectively measured against established engineering and performance specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC clinical study was performed. The document explicitly states, "No Clinical testing was necessary to determine substantial equivalence." This device is intended for the acquisition, display, and storage of EEG/ERP data to aid in diagnosis, with interpretation being the responsibility of a physician. It does not appear to incorporate an AI component for automated diagnostic interpretation or decision support that would necessitate an MRMC study to show AI assistance to human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No standalone clinical performance testing was performed. The device's performance was evaluated through non-clinical bench testing for its technical specifications and compliance with safety and performance standards relevant to EEG/ERP systems.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth for the non-clinical tests was established by defined input signals and test methods outlined in international standards (IEC 60601-2-26). It is an objective, quantitative ground truth based on engineering principles, rather than clinical consensus or pathological findings.

8. The sample size for the training set:

  • Not applicable. The document does not describe the use of machine learning or AI that would require a "training set" in the conventional sense for diagnostic algorithm development. The hardware and software verification and validation were performed on the device itself.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no mention of a training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medeia, Inc. % Daniel Lehtonen Regulatory Consultant Compliance and Regulatory Services LLC 3771 Southbrook Dr Dayton, Ohio 45430

Re: K192753

Trade/Device Name: NeuralScan System Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, GWJ, GWQ Dated: February 19, 2020 Received: February 20, 2020

Dear Daniel Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192753

Device Name

NeuralScan System

Indications for Use (Describe)

The NeuralScan System is intended for the acquisition, display, and storage, of electrical activity of a patient's brain including electroencephalograph (EEG) and Event-related Potentials (ERP), obtained by placing two or more electrodes on the head to aid in diagnosis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medeia. The word "Medeia" is written in a combination of colors, with the "M" in green and yellow, the "ede" in black, and the "ia" in blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font size.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the Requirements of Safe Medical Device Systems Act 1990 and 21 CFR Sec. 807.92

510(k) Number:

K192753

a1 APPLICANT INFORMATION:

Date Prepared:26 September 2019
Name:Medeia, Inc.
Address:7 W. Figueroa StreetSuite 300Santa Barbara, CA, 93101
Contact Person:Slav Danev
Phone Number:+1 800 433 4609
Fax Number:+1 800 433 4609
Email:danev@medeia.com

a2 NAME OF DEVICE:

Trade Name:NeuralScan System
Common Name:EEG/EP System, EEG Amplifier
Classification Name:Non-normalizing quantitative electroencephalograph software21 CFR 882.1400 (OLT);Full-montage standard electroencephalograph21 CFR 882.1400 (GWQ);Stimulator, auditory, evoked response21 CFR 882.1900 (GWJ)
Classification Panel:Neurology

аЗ PREDICATE DEVICES:

Predicate Device:K171781; eVox System
Predicate Device:K141316; CognisionTM

The FDA database for recalls was searched on 15 Aug 2019 during the preparation of the 510(k) submission and no recalls for the devices noted above were found.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Medeia. The word "Medeia" is written in a bold, sans-serif font, with the first letter "M" in green and blue gradient. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, sans-serif font.

a4 STATEMENT OF INTENDED USE:

The NeuralScan System is intended for the acquisition, display, analysis, and storage, of electrical activity of a patient's brain including electroencephalograph (EEG) and Event-related Potentials (ERP), obtained by placing two or more electrodes on the head to aid in diagnosis.

a5 DESCRIPTION OF THE DEVICE:

The NeuralScan System is comprised of the NeuralScan Amplifier with embedded firmware for acquisition and transmission of physiological signals, an EEG cap, a Subject Response Device, a laptop computer preloaded with the NeuralScan Evoke software, and a charging kit (that consists of a USB cable, clip kit, and wall adapter).

The NeuralScan Evoke software runs in a Windows Operating System environment and controls the 23 channel (21 EEG and 2 bio channels) NeuralScan EEG amplifier via a USB cable or Wi-Fi connection. The software has a graphical user interface that allows the user to input patient information, create new records, conduct studies to collect EEG and ERP data, view live data streams of the laptop display, record data to a file, analyze resultant test data using standard Frequency EEG analysis and EP display methods and print the results. The software includes a mode to measure the cap electrode impedances which is useful for determining if the electrodes are making a good electrical connection with the scalp at each electrode location. Interpretation of the data is the responsibility of the physician as the software does not provide any diagnosis based on the data.

The patient contacting accessories are commercially sourced and used without modification. The device and the accessories are not sterile and are not intended to be sterilized.

The NeuralScan System is intended for prescription use in any healthcare, medical, athletic or sports clinic, or outside of medical facilities provided they are led by qualified medical personnel.

The device is intended for use by qualified medical personnel only and qualifies for exemption per 21 CFR 801 Subpart D Prescription devices.

TECHNOLOGICAL CHARACTERISTIC COMPARISON: аб

NeuralScan product share similar device characteristics, intended use, performance, specifications, sensors and is the same in design, function, and application to the predicate devices.

The comparison table, beginning on page 5, demonstrates that the NeuralScan device is substantially equivalent to the predicate devices. The nonclinical data support the safety of the device and the hardware and software verification and validation demonstrate that the NeuralScan device should perform as intended in the specified use conditions.

Based on comparisons of device technological characteristics, features, materials, intended use and performance the NeuralScan has been shown to be substantially equivalent to the commercially available predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the first letter "M" in green and blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font.

b1 NON-CLINICAL TESTING:

Bench testing was carried out on the following characteristics:

  • . Electroencephalograph (EEG)
  • Measurement accuracy
  • Communication, data transmission and storage
  • Reliability (QoS) Wireless Quality of Service
  • Electromagnetic compatibility (EMC)
  • Electrical safety testing
  • . Wireless Coexistence Wi-Fi testing
  • . Software verification and validation testing
  • Biocompatibility verification

In addition to the above, usability testing was also conducted.

Referenced Standards and Performance Testing:

The NeuralScan device was tested and meets the requirements of following

  • IEC 60601-1- Medical Electrical Equipment Part 1: Basic safety and essential performance ● Ed3.1 2005+A1:2012
  • IEC 60601-1-2:2014 - Medical electrical equipment-basic safety and essential performance-EMC
  • . IEC 60601-2-26:2012 Medical electrical equipment - Part 2-26: Basic safety and essential performance of electroencephalographs
  • ISO 15223-1:2012 Medical devices - symbols to be used with medical device labels, labelling, and information to be supplied - part 1: general requirements
TestTest Method SummaryResults / P or F(per IEC 60601-2-26performance limits)
Accuracy of signalreproductionPurpose: verify signal isaccurately reproduced overa range of amplitudes.Test circuit and method per IEC 60601-2-26clause 201.12.1.101.1. Amplitude of 2HzTriangle wave varied from 50μV to 500μV.IEC 60601-2-26 Pass-Fail Limit: error < 20%PassTest shows thatNeuroScan systemaccurately reproduces EEGsignals over the specifiedrange
Input dynamic range anddifferential offset voltagePurpose: verify signalaccuracy in presence of dcoffset voltages.Test circuit and method per IEC 60601-2-26clause 201.12.1.101.2. 1mV 6HzTriangle wave with 300mV offset added.IEC 60601-2-26 Pass-Fail Limit: amplitudeerror < 10%300mV Offset:1.5% error: Pass187mV Offset:1.5% error: Pass

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the first letter "M" in green and blue. The rest of the letters are in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font.

TestTest Method SummaryResults / P or F(per IEC 60601-2-26performance limits)
Input NoisePurpose: verify that thesignal noise caused byamplifier does not exceedmax 6µV peak-to-valleyTest circuit and method per IEC 60601-2-26clause 201.12.1.101.3. Inputs shortedtogether.IEC 60601-2-26 Pass-Fail Limit: max 6µVpeak-to-valleyMaximum noise: 3.4 µV:Pass
Frequency ResponsePurpose: verify signal isaccurately reproduced overa range of frequenciesTest circuit and method per IEC 60601-2-26clause 201.12.1.101.4. Triangle waveAmplitude established at 5Hz, is thenadjusted to 0.5 Hz and 50Hz.IEC 60601-2-26 Pass-Fail Limit: outputamplitude > 71%, < 110%0.5 Hz: 100%: Pass5 Hz: 105%: Pass50 Hz: 76%: Pass
Common mode rejectionPurpose: verify amplifierrejects common mode noiseTest circuit and method per IEC 60601-2-26clause 201.12.1.101.5. Apply 1Vrms at 60Hzto all inputs tied together.IEC 60601-2-26 Pass-Fail Limit: outputamplitude < 10mm (= 100µVpp)With 0 DC Offset:67.3µVpp: PassWith 187mV dc Offset:68µVpp: Pass

Software Verification and Validation Testing

Software verification and validation testing were conducted following the FDA guidance document for software contained in medical devices. The software was considered to be a "moderate" level of concern since a failure or latent flaw could indirectly result in a minor injury to the patient through incorrect or delayed information or through action of the operator.

b2 CLINICAL TESTING:

No Clinical testing was necessary to determine substantial equivalence.

CONCLUSIONS DRAWN FROM TESTING: b3

Based on information obtained on the predicate device with reference to the design specification, electrical safety / EMC testing and intended use, the NeuralScan device was subjected to the same type of testing. The results support the conclusion that the NeuralScan device is substantially equivalent to the Predicate devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Medeia. The word "Medeia" is written in bold, black letters, with the "M" in green and blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in smaller, light blue letters.

Substantial Equivalence Comparison Table

Trade NameNeuralScan SystemeVox SystemCognision™Comments
510 (k) NumberK171781K141316
Product codesOLT, GWQ, GWJGWQ, GWJOLT, GWJ, OMC
Indications for UseThe NeuralScan System isintended for the acquisition,display, analysis, and storage, ofelectrical activity of a patient'sbrain includingelectroencephalograph (EEG)and Event-related Potentials(ERP), obtained by placing twoor more electrodes on the headto aid in diagnosis.The eVox System is intended forthe acquisition, display, andstorage, of electrical activity of apatient's brain includingelectroencephalograph (EEG)and Event-related Potentials(ERP) obtained by placing two ormore electrodes on the head toaid in diagnosis.The COGNISION system is foruse by qualified clinicalprofessional in private practiceoffices or small clinical settingsfor the acquisition, display,analysis, storage, reporting andmanagement ofelectroencephalograph (EEG)and auditory evoked potentials(AEP) information.Product contains multiple productcodes which are generally requiredfor a complete system.The NeuralScan System and theeVox device share the productcodes:GWQ: Full Montage SystemGWJ: Evoked response auditorystimulatorThe NeuralScan System, theCOGNISION system share theproduct code:OLT: Non-normalizing quantitativeelectroencephalograph software andGWJ: Evoked response auditorystimulator
Trade NameNeuralScan SystemeVox SystemCognision™Comments
Principle ofOperationThe NeuralScan System deviceis used for acquisition ofphysiological signals using twoor more channels ofElectroencephalography (EEG)from the scalp. It consists of aNeuralScan amplifier, a laptopcomputer (base station), apatient EEG cap, subjectresponse button, ear buds, and acharging cord. The NeuralScanamplifier and software provide ameans to:- Initiate a study, track user EEGand ERP data and enter text orquestionnaire informationacquire and save signals to thememory of the device,transmit signal data from thedevice,visually inspect the acquiredsignal.- manage event- relatedPotentialsThe eVox System device is usedfor acquisition of physiologicalsignals using two or morechannels ofelectroencephalography (EEG)from the scalp. It consists of aneVox amplifier, a laptopcomputer (base station), apatient EEG cap, subjectresponse button, ear buds, and acharging cord. The eVoxamplifier and software provide ameans to:- initiate a study, track user EEGand ERP data and enter text orquestionnaire information,acquire and save signals to thememory of the device,transmit signal data from thedevice,visually inspect the acquiredsignal.- manage event- relatedPotentialsThe COGNISION EEG/EPSystem is a combination devicefor reduced montage recordingand display ofelectroencephalographic (EEG)and evoked potentials (EP) testdata.The system uses elastic bandsto accurately position 10electrode pods around the head.EEG signal amplification,conditioning, and A/D conversionis performed by electroniccircuits closely coupled to theelectrode pods through shortflexible printed wires.The headset is connected by acable to a handheld control unitand data acquisition box (HCU).The HCU communicates via awireless data link to a WindowsPC to stream EEG data.Software on the PC is used tosetup the tests and view andevaluate the resultant test datausing standard EEG/EP displaymethods.The NeuralScan System has aprinciple of operation that closelymatches that of its predicates. Alldevices use skin coupling methods,either through electrodes or sensorswhich transmits patient EEG andERP from the surface of the scalp toan amplifier.The connection between the amplifierand electrodes is a wired connectionfor the devices.All devices convert the analog datainto digital data which is thentransmitted to a base station orcomputer.Once transmitted to the base station,the devices display and store thedata on the base station and allowthe user to export the data to a file.
Patient populationAll age groupsAll age groupsAdult Population
Use environmentIntended for use in anyhealthcare facility, medicalfacility, athletic or sports clinics,or outside of medical facilitiesprovided they are led by qualifiedmedical personnel.Intended for use in anyhealthcare, medical, or athleticor sports clinics, or outside ofmedical facilities such as in thesports arena under thesupervision of a physician.Also, investigations can beperformed outside of healthcarefacilities, as long as they are ledby qualified medical personnel.Physicians' Offices
FDA Device ClassClass IIClass IIClass II
Trade NameNeuralScan SystemeVox SystemCognision™Comments
BiocompatibilityPer ISO 10993-1Per ISO 10993-1Per ISO 10993-1
SystemComponentsNeuralScan System consists of:- a NeuralScan amplifier,- a laptop computer (basestation),- a patient EEG cap,- subject response button, earbuds,- and a charging cord.eVox System consists of:- an eVox amplifier,- a laptop computer (basestation),- a patient EEG cap,- subject response button, earbuds,- and a charging cord.COGNISIONTM consists of:- Head Set- Auditory stimulator- Handheld Control Unit(HCU) including InterfaceSoftware- Connecting Headset cablebetween the headset andthe handheld control unit
SterileNoNoNo
Single UseNoNoNo
Shelf lifeDurable goodDurable goodDurable good
Interface withAmplifierUSB or WiFi to PCClass 2 Bluetooth version 2.0 toPCBluetooth 2.0/4.0
Power SupplyLi-Ion Battery, with USB cablefor charging the battery.Li-Ion Battery, with USB cablefor charging the battery.Li-Ion Battery
Typical BiopotentialSignals RecordedElectroencephalography (EEG),EP/ERPElectroencephalography (EEG),EP/ERPElectroencephalography (EEG),EP/ERP
Electroencephalography (EEG), ERP
ERP StimulusModalityAuditory; VisualAuditory; VisualAuditory
ERP Paradigm(Auditory andVisual Stimuli)P300 Oddball- Single Stimulus- Single Deviant- 2 Deviant- Active and PassiveP300 Oddball- Single Stimulus- Single Deviant- 2 Deviant- Active and PassiveP300 Oddball- Single Stimulus- Single Deviant- 2 Deviant- Active and Passive
ERP TaskResponseUser ButtonsUser ButtonsUser Buttons
Skin CouplingCustom Electrode Band and GelCustom Electrode Band and GelDiscrete Electrode WiresThe NeuralScan, and eVox use thesame method of utilizing a conductivegel between electrode and the skin.The COGNISION system usesHydroDot Biosensor
Trade NameNeuralScan SystemeVox SystemCognision™Comments
Number of SignalRecordingChannelsUp to 23Up to 21Up to 10Both the NeuralScan and Evoxsystems have 2 bio-channels
EEG inputterminalsup to 21 channelsup to 19 channelsUp to 7 channelsFull montage for 10-20 system has21 channels
EEG RecordingChannels Locationand PositioningSystemsAny of the 21 EEG Channelsincluding: Fz, Cz, Pz, F3, P3, F4,P4 Utilizing elastic bands usingdistance ratios consistent withthe 10-20 SystemAny of the 19 EEG Channelsincluding: Fz, Cz, Pz, F3, P3, F4,P4 Utilizing elastic bands usingdistance ratios consistent withthe 10-20 SystemFz, Cz, Pz, F3, P3, F4, P4Utilizing elastic bands usingdistance ratios consistent withthe 10-20 SystemSame - full and reduced montagerecording capabilities
Impedance TestYesYesYes
Amplifier InputImpedance> 200 ΜΩ> 10 ΜΩ> 60 ΜΩPerformance testing per IEC 60601-2-26:2012 confirmed 10 ΜΩ issufficient to eliminate distortion in themeasured signal.
Analog to DigitalConversion24 Bit24 Bit16 Bit
Sampling Rate200, 500, 1000 Hz250 Hz125/250 HzClinically relevant humanelectrophysiology does not exceed100 Hz and therefore there is notechnical need to sample above 250Hz.
Common moderejection>110 dB>110 dB>90dB
Analysis SoftwareEmbedded, commerciallyavailable, and user defined.Embedded, commerciallyavailable, and user defined.Embedded, commerciallyavailable, and user defined.
Resolution24 bits24 bits16 bits
Band Pass0.1 to 50 Hz0.1 to 50 Hz0.4 to 40 Hz
Noise2-3 μVp-p2-3 μVp-p≤1μV RMS
Audio TypeBurst (White Noise)Burst (White Noise)UnknownThe NeuroScan system uses a bursttype noise stimulus to evoke andERP
Audio Duration100ms100ms50ms
Audio SideBothBothUnknown
Trade NameNeuralScan SystemeVox SystemCognision™Comments
Audio FrequencyRange440Hz-16kHz440 Hz-16kHzUnknownThe NeuralScan system generates asound with a frequency range of440Hz-16kHz which is a sufficientlywide enough band for humanhearing.
Intensity0 to 85dB0 to 85dBUnknownThe sound emitted by theNeuralScan system is a flat whitenoise burst, therefore the peak andaverage are the same. Maximumdecibel level is 85dB, well below125dB (no risk to user).
Audio NoisePatterns12.5% occurrence randomlydistributed over 10 minutes12.5% occurrence randomlydistributed over 10 minutes10% occurrence within 10-60minutesThe balance of this pattern is best toelicit the desired ERPs.
Input VoltageRange± 400 mV± 150 mVNot StatedDue to the 24x gain that is applied tothe A/D channels by the NeuralScanthe dynamic range is ± 400 mV forthe NeuralScan System which issufficient for its use environment.
General Specifications
Safety StandardsComplianceIEC 60601-1: 2005 +CORR. 1:2006 + CORR.2:2007 + AM1:2012 (or IEC60601-1: 2012 reprint)EN 60601-1-2:2014IEC 60601-2-26:2012IEC 60601-1: 2005 +CORR. 1:2006 + CORR.2:2007 + AM1:2012 (or IEC60601-1: 2012 reprint)EN 60601-1-2:2012IEC 60601-2-26:2012UL 60601-1:2003EN60601-1-2/A1:2007EN60601-1-2/A1:2007EN 60601-2-26IEC 60601-2-40
OperatingEnvironment10 to +45 °C, Relative humidity,30% to 70% non- condensing0 to +45 °C, Relative humidity,5% to 95% non- condensing60-90 °F
StorageEnvironment-40° to 70° C, Relative humidity,5% to 95% non- condensing-20° to 45° C, Relative humidity,5% to 95% non- condensingNot Published

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the "M" in green and yellow, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, sans-serif font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "Medeia". The logo has a stylized "M" in green and yellow, followed by the rest of the word "edeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font size.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in black font, with the "M" in green and blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller blue font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Medeia. The word "Medeia" is written in a bold, sans-serif font, with the "M" in green and yellow, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, blue font.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Medeia. The word "Medeia" is written in a bold, sans-serif font, with the "M" in green and blue and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, sans-serif font.

ConclusionThe comparison tabulated above demonstrates that the NeuralScan device issubstantially equivalent to the predicate devices. The nonclinical data support thesafety of the device and the hardware and software verification and validationdemonstrate that the NeuralScan device should perform as intended in thespecified use conditions.
Declarations• This summary includes only information that is also covered in the body of the510(k).• This summary does not contain any puffery or unsubstantiated labeling claims.• This summary does not contain any raw data, i.e., contains only summarydata.• This summary does not contain any trade secret or confidential commercialinformation.• This summary does not contain any patient identification information.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).