K Number
K192679

Validate with FDA (Live)

Date Cleared
2020-04-10

(197 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterile Syringe with Safety needle for Single Use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

Sterile Syringe for Single Use is a sterile luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.

Sterile Safety Needle for Single Use is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

Device Description

Sterile Syringe with Safety needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of syringe (piston, barrel, plunger) and hypodermic needle with a safety mechanism. The proposed device is available in a variety combination of syringe volume and needle size.

The Sterile Syringe for Single Use is intended for manual and single use only, which consists of piston, barrel and plunger. The proposed device is available in a variety syringe volume. The syringe is available in luer lock, which is intended to be connected with a hypodermic needle.

Sterile Safety Needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of needle tube, retractable cartridge, jointing medium and needle hub. The proposed device is available in variety of needle gauge and needle length. According to the needle length, the safety mechanism is available in 2-part cartridge (outer retractable cartridge and inner retractable cartridge) and 3-part cartridge (outer retractable cartridge, middle retractable cartridge and inner retractable cartridge). The proposed device is compatible for use with a luer lock syringe. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

AI/ML Overview

The provided text describes a submission for K192679 (Sterile Syringe with Safety Needle for Single Use, Sterile Syringe for Single Use, Sterile Safety Needle for Single Use) to the FDA, asserting substantial equivalence to predicate device K170651.

The document discusses acceptance criteria for various tests and confirms that the proposed device met these criteria through non-clinical testing and a simulated clinical study.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Test CategorySpecific Test / StandardAcceptance CriteriaReported Device Performance
Physical, Mechanical, ChemicalClause 5 of ISO 7886-1:2017General requirementsConforms with requirements of related standards.
Clause 6 of ISO 7886-1:2017Extraneous matterConforms with requirements of related standards.
Clause 7 of ISO 7886-1:2017LubricantConforms with requirements of related standards.
Clause 8 of ISO 7886-1:2017Tolerance on graduated capacity (Syringe)Met the requirements of ISO 7886-1.
Clause 9 of ISO 7886-1:2017Graduated scaleConforms with requirements of related standards.
Clause 10 of ISO 7886-1:2017BarrelConforms with requirements of related standards.
Clause 11 of ISO 7886-1:2017Piston/plunger assemblyConforms with requirements of related standards.
Clause 12 of ISO 7886-1:2017NozzleConforms with requirements of related standards.
Clause 13 of ISO 7886-1:2017Performance (Syringe)Met the requirements of ISO 7886-1.
Clause 4.3 of ISO 7864:2016Cleanliness (Needle)Complies with ISO 7864.
Clause 4.4 of ISO 7864:2016Limits for acidity or alkalinity (Needle)Complies with ISO 7864.
Clause 4.5 of ISO 7864:2016Limits for extractable metals (Needle)Complies with ISO 7864.
Clause 4.6 of ISO 7864:2016Size designation (Needle)Complies with ISO 7864.
Clause 4.7 of ISO 7864:2016Colour coding (Needle)Complies with ISO 7864.
Clause 4.8 of ISO 7864:2016Needle hubComplies with ISO 7864.
Clause 4.9 of ISO 7864:2016Needle CapComplies with ISO 7864.
Clause 4.10 of ISO 7864:2016Needle tubeComplies with ISO 7864.
Clause 4.11 of ISO 7864:2016Needle pointComplies with ISO 7864.
Clause 4.12 of ISO 7864:2016Bond between hub and needle tubeComplies with ISO 7864.
Clause 4.13 of ISO 7864:2016Patency of lumenComplies with ISO 7864.
Clause 5.2 of ISO 9626:2016Surface finish and appearanceComplies with ISO 9626.
Clause 5.3 of ISO 9626:2016CleanlinessComplies with ISO 9626.
Clause 5.4 of ISO 9626:2016Limits for acidity and alkalinityComplies with ISO 9626.
Clause 5.5 of ISO 9626:2016Size designationComplies with ISO 9626.
Clause 5.6 of ISO 9626:2016DimensionsComplies with ISO 9626.
Clause 5.8 of ISO 9626:2016StiffnessComplies with ISO 9626.
Clause 5.9 of ISO 9626:2016Resistance to breakageComplies with ISO 9626.
Clause 5.10 of ISO 9626:2016Resistance to corrosionComplies with ISO 9626.
Sterile Barrier PackagingASTM F1886/F1886M-16Visual inspectionMaintain its integrity.
ASTM F88/F88-15Seal strengthMaintain its integrity.
ASTM F1929-15Dye penetration testMaintain its integrity.
Sterilization and Shelf LifeISO 10993-7:2008EO residue limitDid not exceed the limit.
ISO 10993-7:2008ECH residue limitDid not exceed the limit.
USP 38-NF 33 <85>Bacteria Endotoxin Limit (20EU/device)Did not exceed 20EU/device.
Physical, Mechanical, Chemical, Package Test on aging samplesMaintain its performance during the claimed shelf life.Maintained its performance during the claimed shelf life.
BiocompatibilityISO 10993-1 series (Tests for blood path, indirect, <24 hours contact)No negative impacts from materials.No negative impacts reported for Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute Systemic Toxicity, Hemolysis, Complement activation, Thromboresistance study, Pyrogen, Particulate testing.
Safety FeatureFDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908:2011Meet pre-established criteria (Simulated Clinical Study).Meet acceptance criteria (Safety Feature Test performed on both proposed and predicate).Met the pre-established criteria. (Proposed device) Met the acceptance criteria. (Proposed device and predicate device)
Connect Type (Luer lock)ISO 80369-7Performance of Luer lock.Met the requirements of ISO 80369-7.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the numerical sample sizes for each test set. It mentions tests were performed on the "proposed device" and "aging samples" for shelf-life. In the Safety Feature Test, tests were performed on "both proposed device and predicate device".

The data provenance is from non-clinical tests conducted by Gemtier Medical (Shanghai) Inc. in China (manufacturer location). The nature of these tests (bench testing, simulated clinical study) implies they are prospective in execution, meaning they were designed and performed for the purpose of this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the document describes non-clinical and simulated clinical studies for a medical device (syringes and needles), not studies involving human subjects or expert assessment of medical images/data for ground truth. The acceptance criteria are based on compliance with international standards and internal design specifications, not expert consensus on medical conditions.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. Adjudication methods are typically used in clinical studies or studies where subjective interpretations need to be resolved. The tests described here are objective performance measurements against established standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a sterile syringe and needle system, not an AI-powered diagnostic or assistive technology that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance is established by the international standards (e.g., ISO 7886-1, ISO 7864, ISO 9626, ISO 10993-7, ISO 23908, ISO 80369-7) and internal design specifications against which the device was tested. For biocompatibility, the lack of "negative impacts" is the ground truth. For the safety mechanism, meeting "pre-established criteria" of relevant guidance documents serves as the ground truth.

8. The sample size for the training set

This is not applicable. The document describes the testing and comparison of a physical medical device. There is no training set mentioned, as this is not an AI/ML device.

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

April 10, 2020

Gemtier Medical (Shanghai) Inc. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 Cn

Re: K192679

Trade/Device Name: Sterile Syringe with Safety Needle for Single Use, Sterile Syringe for Single Use, Sterile Safety Needle for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG. FMF. FMI Dated: February 26, 2020 Received: March 12, 2020

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192679

Device Name

Sterile Syringe with Safety needle for Single Use Sterile Syringe for Single Use Sterile Safety Needle for Single Use

Indications for Use (Describe)

Sterile Syringe with Safety needle for Single Use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

Sterile Syringe for Single Use is a sterile luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.

Sterile Safety Needle for Single Use is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K192679

    1. Date of Preparation: 04/09/2020
    1. Sponsor Identification

Gemtier Medical (Shanghai) Inc.

No.18 Jianding Road, Fengjing Town, Jinshan District, Shanghai, 201502, China

Establishment Registration Number: 3009746425

Contact Person: Lenny Cao Position: Sales Manager Tel: (+86) -21-67360886 Fax: +86-21-57365666 Email: lenny.cao@gemtier.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119. Shanghai, 200120. China

Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Sterile Syringe with Safety needle for Single Use Sterile Syringe for Single Use Sterile Safety Needle for Single Use

Regulatory Information Classification Name: Syringe Antistick Classification: II; Product Code: MEG Regulation Number: 21CFR 880.5860 Review Panel: General Hospital;

Classification Name: Syringe, Piston Classification: II; Product Code: FMF Regulation Number: 21CFR 880.5860 Review Panel: General Hospital;

Classification Name: Hypodermic single lumen needle Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital

Indication for Use:

Sterile Syringe with Safety needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

Sterile Syringe for Single Use is a sterile luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.

Sterile Safety Needle for Single Use is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.

{5}------------------------------------------------

Device Description:

Sterile Syringe with Safety needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of syringe (piston, barrel, plunger) and hypodermic needle with a safety mechanism. The proposed device is available in a variety combination of syringe volume and needle size.

Needle GaugeLengthWall type
18G1", 1 1/4", 1 1/2"Thin wall
21G1/2", 5/8", 1", 1 1/4", 1 1/2"Thin wall
22G1/2", 5/8", 1", 1 1/4", 1 1/2"Thin wall
23G1/2", 5/8", 3/4", 1", 1 1/4", 1 1/2"Thin wall
25G3/8", 1/2", 5/8", 3/4", 1", 1 1/4", 1 1/2"Thin wall

Syringe volume: 1ml, 2ml, 5ml, 10ml, 20ml, 30ml, and 50ml; Needle specification:

The Sterile Syringe for Single Use is intended for manual and single use only, which consists of piston, barrel and plunger. The proposed device is available in a variety syringe volume. The syringe is available in luer lock, which is intended to be connected with a hypodermic needle. Syringe volume: 1ml, 2ml, 5ml, 10ml, 20ml, 30ml, and 50ml;

Sterile Safety Needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of needle tube, retractable cartridge, jointing medium and needle hub. The proposed device is available in variety of needle gauge and needle length. According to the needle length, the safety mechanism is available in 2-part cartridge (outer retractable cartridge and inner retractable cartridge) and 3-part cartridge (outer retractable cartridge, middle retractable cartridge and inner retractable cartridge). The proposed device is compatible for use with a luer lock syringe. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. Needle specification is same as the needle size of Sterile Syringe with Safety needle for Single Use.

  • న్. Identification of Predicate Device
    510(k) Number: K170651 Product Name: Sterile Disposable Syringe With Safety Needle; Sterile Disposable Syringe With Needle; Sterile Disposable Syringe; Sterile Disposable Safety Needle; Sterile Disposable Needle

{6}------------------------------------------------

6. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The testing included the following items:

Physical, Mechanical, Chemical testing listed in following table were performed on the proposed device. The test results show that the device conforms with the requirements of related standards.

General requirementsClause 5 of ISO 7886-1:2017
Extraneous matterClause 6 of ISO 7886-1:2017
LubricantClause 7 of ISO 7886-1:2017
Tolerance on graduated capacityClause 8 of ISO 7886-1:2017
Graduated scaleClause 9 of ISO 7886-1:2017
BarrelClause 10 of ISO 7886-1:2017
Piston/plunger assemblyClause 11 of ISO 7886-1:2017
NozzleClause 12 of ISO 7886-1:2017
PerformanceClause 13 of ISO 7886-1:2017
CleanlinessClause 4.3 of ISO 7864:2016
Limits for acidity or alkalinityClause 4.4 of ISO 7864:2016
Limits for extractable metalsClause 4.5 of ISO 7864:2016
Size designationClause 4.6 of ISO 7864:2016
Colour codingClause 4.7 of ISO 7864:2016
Needle hubClause 4.8 of ISO 7864:2016
Needle CapClause 4.9 of ISO 7864:2016
Needle tubeClause 4.10 of ISO 7864:2016
Needle pointClause 4.11 of ISO 7864:2016
Bond between hub and needle tubeClause 4.12 of ISO 7864:2016
Patency of lumenClause 4.13 of ISO 7864:2016
Surface finish and appearanceClause 5.2 of ISO 9626:2016
CleanlinessClause 5.3 of ISO 9626:2016
Limits for acidity and alkalinityClause 5.4 of ISO 9626:2016
Size designationClause 5.5 of ISO 9626:2016
DimensionsClause 5.6 of ISO 9626:2016
StiffnessClause 5.8 of ISO 9626:2016
Resistance to breakageClause 5.9 of ISO 9626:2016
Resistance to corrosionClause 5.10 of ISO 9626:2016

Sterile barrier packaging testing were performed on the proposed device, which include visual

{7}------------------------------------------------

inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.

Sterilization and shelf life testing listed in following table were performed on the proposed device. EO ECH residue did not exceed the limit of ISO 10993-7. Endotoxin limit did not exceed 20EU/device. Shelf life test result showed that the device can maintain its performance during the claimed shelf life.

EO residueISO 10993-7:2008
ECH residueISO 10993-7:2008
Bacteria Endotoxin LimitUSP 38-NF 33 <85>
Shelf Life EvaluationPhysical, Mechanical, Chemical, Package Testwere performed on aging samples to verify theclaimed shelf life of the device

Biocompatibility testing

The contact level of the proposed device is blood path, indirect, and the contact duration is limited contact (<24 hours). The proposed device was evaluated for the following tests. The results for the biocompatibility testing showed that there are no negative impacts from the materials that are used in the proposed device.

  • Cytotoxicity,

  • A Sensitization,
  • Intracutaneous reactivity,

  • Acute Systemic Toxicity,

  • Hemolysis.

  • Complement activation,

  • Thromboresistance study

  • A Pyrogen
  • Particulate testing

Simulated Clinical Study

A simulated clinical study was performed on proposed device according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9. 2005 and ISO 23908:2011 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.

Safety Feature Test

{8}------------------------------------------------

The safety feature test was performed on both proposed device and predicate device to determine its safety feature. The results demonstrated that both the proposed device and predicate device meet the acceptance criteria.

    1. Clinical Test Conclusion
      No clinical study is included in this submission.

{9}------------------------------------------------

8. Substantially Equivalent (SE) Comparison

Table 1 Comparison of Technology Characteristics for Sterile Syringe with Safety needle for Single Use

ITEMSubject Device K192679Predicate Device K170651Comments
Regulation No.21CFR 880.586021CFR 880.5860Same
Product CodeMEGMEGSame
ClassIIIISame
Indication for UseSterile Syringe with Safety needle for Single Use is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticksSame
Configuration and materialBarrel (Polypropylene, PP)Plunger (Polypropylene, PP)Piston (Polyisoprene, Carbon Black)Needle cap (Polypropylene, PP)Needle tube (Stainless steel)Retractable cartridge (Polycarbonate, PC)Jointing medium (Methyl methacrylate acrylonitrile butadiene styrene (MABS)Needle hub (Polycarbonate, PC)Needle tube (Stainless steel SUS304)Lubricant (Silicon plastic agent)Adhesive (UV Light Cure Adhesive)Colorants (MACROLEX Orange 3G, Heliogen Blue K 7090, BAYFERROX 318M, Heliogen Green K 8730, RanbarBarrel (Polypropylene, PP)Plunger (Polypropylene, PP)Piston (Polyisoprene)Needle hub (Polypropylene, PP)Protective cap (Polypropylene, PP)Needle tube (Stainless steel)Safety sheath (Polypropylene, PP)Needle tube (stainless steel SUS304)Needle tube (Stainless Steel SUS304)Lubricant (Silicone oil)Adhesive (UV glue)Colorants (Yellow, Red, White, Black, Blue, Green)See Comment #1
Operation ModeFor manual use onlyFor manual use onlySame
Safety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latch lock to prevent arbitrary moving.Slide over the needle to prevent from needle sticksSee Comment #2
Label/labelingConform with 21CFR Part 801Conform with 21CFR Part 801Same
Syringe Volume1ml, 2ml, 5ml, 10ml, 20ml, 30ml, 50ml1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml, 60mlSee Comment #3
Connect TypeLuer lockLuer lock and Luer slipSee Comment #4
Needle Gauge18G, 21G, 22G, 23G, 25G16G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30GSee Comment #5
Needle LengthAvailable in 3/8", 1/2", 5/8", 3/4", 1", 1 1/4", 1 1/2"Available in 5/16", 1/2", 5/8", 3/4", 1 1/4", 1 1/2"See Comment #6
Single UseSingle UseSingle UseSame
SterilizationMethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20EU20EUSame

{10}------------------------------------------------

{11}------------------------------------------------

Discussion of Differences

Comment 1

The configuration between proposed device and predicate device is different. However, the basic configurations are same, both of them include barrel, plunger, piston, needle hub, protective or needle cap. The only different is retractable cartridge, jointing medium and safety sheath, which are parts of safety mechanism, and predicate device lubricant adhesive and colorant information unknown. Additionally, the safety mechanism does affect the indication for use. Therefore, this difference does not affect substantially equivalence.

The material of configuration is different between proposed device and predicate device. However, the mainly configurations such as barrel, plunger, piston, needle/protective cap are made of same materials. The only differences are the material of needle hub, safety feature, lubricant, adhesive and colorant. And the MSDS for all related materials has been provided in this submission. Differences in materials of construction are addressed through ISO 10993-1: biocompatibility testing. Additionally, the material of configuration does not affect the indication for use. Therefore, this difference does not affect substantially equivalence.

Comment 2

The safety feature between proposed device and predicate device is different. The safety mechanism for proposed device is retractable cartridge, for predicate, that is safety sheath. Both of them are intended to prevent from needle stick injuries, and which does not affect indication for use. In addition, the safety feature for proposed device has been evaluated and the test result conforms to requirements of ISO 23908:2011 standard. Therefore, this difference does not affect substantially equivalence.

Comment 3

The predicate device has more specification of syringe volume compared predicate device. However, the different syringe volume will be selected by physician per injection requirement and this difference does not affect indication for use. Additionally, the performance of syringe has been evaluated and the test results met the requirements of ISO 7886-1. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.

Comment 4

The connect type between proposed device and predicate device is different. For proposed device, only includes Luer lock, for predicate device, the connect type includes Luer lock and Luer slip. However, the connect type for proposed device is covered by predicate, and the connect type is only used to compatible with needle, and does not affect indication for use. Additionally, the performance of Luer lock has been evaluated and the test results met the requirements of ISO 80369-7.

Comment 5

The needle gauge of predicate device is more than proposed device, and for proposed device, the needle

{12}------------------------------------------------

gauge is covered by predicate device. However, this difference is just in needle size. In addition, all the needle gauge of proposed device has been tested. The test results comply with ISO 7864 and ISO 9626. The different specifications are selected by physician according to the patient's conditions. Therefore, this difference does not affect substantially equivalence on safety and effectiveness.

Comment 6

The needle length for proposed device is covered by predicate device. The different length of needles will be selected by physician according different patient's condition, the needle performance has been evaluated and test results comply with ISO 7864 and ISO 9626. This different does not affect the performance of the syringe. Therefore, this difference is considered not to affect substantially equivalency between proposed device and predicate device.

{13}------------------------------------------------

ITEMSubject Device K192679Predicate Device K170651Comments
Regulation No.21CFR 880.586021CFR 880.5860Same
Product CodeFMFFMFSame
ClassIIIISame
Indication for UseSterile Syringe for Single Use is a sterile luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.The Sterile Disposable Syringe is a sterile luer lock or luer slip syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.See Comment #7
Configuration and materialBarrel (Polypropylene, PP)Plunger (Polypropylene, PP)Piston (Polyisoprene, carbon black)Lubricant (Silicon plastic agent)Adhesive (UV Light Cure Adhesive)Barrel (Polypropylene, PP)Plunger (Polypropylene, PP)Piston (Polyisoprene)Needle tube (Stainless Steel SUS304)Lubricant (Silicone oil)Adhesive (UV glue)See Comment #8
Operation ModeFor manual use onlyFor manual use onlySame
Label/labelingConform with 21CFR Part 801Conform with 21CFR Part 801Same
Syringe Volume1ml, 2ml, 5ml, 10ml, 20ml, 30ml, 50ml1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml, 60mlSee Comment #9
Connect TypeLuer lockLuer lock and Luer slipSee Comment #10
Single UseSingle UseSingle UseSame
Sterilization MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20EU20EUSame
Table 2 Comparison of Technology Characteristics for Sterile Syringe for Single Use
-----------------------------------------------------------------------------------------

{14}------------------------------------------------

Discussion of Differences

Comment 7

There are some subtle differences in indication for use between proposed device and predicated device. The difference is connect type. For proposed device, only has Luer lock, while for predicate device, has Luer lock and Luer slip (See Comment 10). And the type connect of proposed device is covered by predicate device. In addition, the connect type is only used to compatible with needle, and does not affect aspiration and injection of fluids. The results demonstrated that this difference does not raise new problem on the safety and effectiveness of the proposed device. Therefore, this difference does not affect substantially equivalence.

Comment 8

The configuration between proposed device and predicate device is different. However, the basic configurations are same, both of them include barrel, plunger, and piston. The material of configuration is different between proposed device and predicate device. However, the mainly configurations such as barrel, plunger and piston are made of same materials. And the MSDS for all related materials has been provided in this submission. Differences in materials of construction are addressed through ISO 10993-1: biocompatibility testing. Additionally, the material of configuration does not affect the indication for use. Therefore, this difference does not affect substantially equivalence.

Comment 9

The predicate device has more specification of syringe volume compared predicate device. However, the different syringe volume will be selected by physician per injection requirement and this difference does not affect indication for use. Additionally, the performance of syringe has been evaluated and the test results met the requirements of ISO 7886-1. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.

Comment 10

The connect type between proposed device and predicate device is different. For proposed device, only includes Luer lock, for predicate device, the connect type includes Luer lock and Luer slip. However, the connect type for proposed device is covered by predicate, and the connect type is only used to compatible with needle, and does not affect indication for use. Additionally, the performance of Luer lock has been evaluated and the test results met the requirements of ISO 80369-7.

{15}------------------------------------------------

ITEMSubject Device K192679Predicate Device K170651Comments
Regulation No.21CFR 880.557021CFR 880.5570Same
Product CodeFMIFMISame
ClassIIIISame
Indication for UseSterile Safety Needle for Single Use is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.The Sterile Disposable Safety Needle is intended for use with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestickSee Comment #11
Configuration and materialNeedle cap (Polypropylene, PP)Needle tube (Stainless steel)Retractable cartridge (Polycarbonate, PC)Jointing medium (Methyl methacrylate acrylonitrile butadiene styrene (MABS)Needle hub (Polycarbonate, PC)Needle tube (Stainless Steel SUS304)Lubricant (Silicon plastic agent)Adhesive (UV Light Cure Adhesive)Colorants (MACROLEX Orange 3G, Heliogen Blue K 7090, BAYFERROX 318M, Heliogen Green K 8730, Ranbar red P1400)Needle hub (Polypropylene, PP)Protective cap (Polypropylene, PP)Needle tube (Stainless steel)Safety sheath (Polypropylene, PP)Needle tube (Stainless Steel SUS304)Lubricant (Silicone oil)Adhesive (UV glue)Colorants (Yellow, Red, White, Black, Blue, Green)See Comment #12
Operation ModeFor manual use onlyFor manual use onlySameSafety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latchSlide over the needle to prevent from needle sticks
Operation ModeFor manual use onlyFor manual use onlySameSafety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latchSlide over the needle to prevent from needle sticks
Operation ModeFor manual use onlyFor manual use onlySameSafety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latchSlide over the needle to prevent from needle sticks
Operation ModeFor manual use onlyFor manual use onlySameSafety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latchSlide over the needle to prevent from needle sticks
Operation ModeFor manual use onlyFor manual use onlySame
Safety FeatureThe needle is withdrawn into safety mechanism, the safety mechanism is fixed via latchSlide over the needle to prevent from needle sticksSee Comment #13
Table 3 Comparison of Technology Characteristics for Sterile Safety needle for Single Use
-------------------------------------------------------------------------------------------------

{16}------------------------------------------------

Label/labelingConform with 21CFR Part 801Conform with 21CFR Part 801Same
Connect TypeLuer lockLuer lock and Luer slipSeeComment#14
Needle Gauge18G, 21G, 22G, 23G, 25G16G, 18G, 19G, 20G, 21G, 22G,23G, 24G, 25G, 26G, 27G, 28G,29G, 30GSeeComment#15
Needle Length and walltypeAvailable in 3/8", 1/2", 5/8",3/4", 1", 1 1/4", 1 1/2"Wall type: TWAvailable in 5/16", 1/2", 5/8",3/4", 1 1/4", 1 1/2"Wall type: TW or RWSeeComment#16
Single UseSingle UseSingle Use
Sterilization
MethodEO SterilizedEO SterilizedSame
SAL10-610-6Same
Endotoxin Limit20EU20EUSame

Discussion of Differences

Comment11

The only difference for predicate device and proposed device is connect type. For proposed device, only has Luer lock, while for predicate device, has Luer slip (See Comment 14). The type connect of proposed device is covered by predicate device. In addition, the connect type is only used to compatible with syringe, and does not affect aspiration and injection of fluids. Additionally, both of them have safety feature. after withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. The results demonstrated that this difference does not raise new problem on the safety and effectiveness of the proposed device. Therefore, this difference does not affect substantially equivalence.

Comment 12

The configuration between proposed device and predicate device is different. However, the basic configurations are same, both of them include, protective or needle tube, and needle hub. The only different is retractable cartridge, jointing medium and safety sheath, which are parts of safety mechanism. Additionally, the safety mechanism does affect the indication for use and is just used to prevent from needle sticks. Therefore, this difference does not affect substantially equivalence.

The material of configuration is different between proposed device and predicate device. However, the mainly configurations such as needle tube, needle/protective cap are materials. The only differences are the material of needle hub, safety feature, adhesive and colorant. However, these configurations not contact with patient, and the materials are not novel materials. And the MSDS for all related materials has been provided in this submission. Differences in materials of construction are addressed through ISO 10993-1: biocompatibility testing. Additionally, the material of configuration does not affect the

{17}------------------------------------------------

indication for use. Therefore, this difference does not affect substantially equivalence.

Comment 13

The safety feature between proposed device and predicate device is different. The safety mechanism for proposed device is retractable cartridge, for predicate, that is safety sheath. Both of them are intended to prevent from needle stick injuries, and which does not affect indication for use. In addition, simulated clinical use testing per FDA guidance Medical Devices with Sharps Injury Prevention Features. The simulated clinical test demonstrated that the safety features were found to meet the predetermined requirements and were acceptable. .Therefore, this difference does not affect substantially equivalence

Comment 14

The connect type between proposed device and predicate device is different. For proposed device, only includes Luer lock, for predicate device, the connect type includes Luer lock and Luer slip. However, the connect type for proposed device is covered by predicate, and the connect type is only used to compatible with needle, and does not affect indication for use. Additionally, the performance of Luer lock has been evaluated and the test results met the requirements of ISO 80369-7.

Comment 15

The needle gauge of predicate device is more than proposed device, and for proposed device, the needle gauge is covered by predicate device. However, this difference is just in needle size. In addition, all the needle gauge of proposed device has been tested. The test results comply with ISO 7864 and ISO 9626. The different specifications are selected by physician according to the patient's conditions. Therefore, this difference does not affect substantially equivalence on safety and effectiveness.

Comment 16

The needle length and wall type for proposed device is covered by predicate device. The different length of needles and wall type will be selected by physician according different patient's condition. In addition, the needle performance has been evaluated and test results comply with ISO 7864 and ISO 9626. This different does not affect the performance of the syringe. Therefore, this difference is considered not to affect substantially equivalency between proposed device and predicate device.

{18}------------------------------------------------

9. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).