K Number
K192485

Validate with FDA (Live)

Date Cleared
2019-12-23

(104 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioCode Respiratory Pathogen Panel (RPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BioCode MDx 3000 Instrument. The BioCode of the simultaneous detection and identification of nucleic acids from multiple viruses and bacteria extracted from nasopharyngeal swab (NPS) samples obtained from individuals with signs and/or symptoms of respiratory tract infection. The following pathogens and subtypes are identified using the BioCode RPP:

  • Adenovirus
  • · Coronavirus (229E, OC43, HKU1, and NL63)
  • Human Metapneumovirus A/B
  • · Influenza A, including subtypes H1, H1 2009 Pandemic, and H3
  • Influenza B
  • Parainfluenza Virus 1
  • Parainfluenza Virus 2
  • · Parainfluenza Virus 3
  • · Parainfluenza Virus 4
  • · Respiratory Syncytial Virus A/B
  • Rhinovirus/Enterovirus
  • · Bordetella pertussis
  • Chlamydia pneumoniae
  • Mycoplasma pneumoniae

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the BioCode RPP may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus, the BioCode RPP cannot differentiate them. A positive BioCode RPP Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required. The BioCode RPP detects Human Rhinovirus with reduced sensitivity. If a more accurate Rhinovirus result is required, it is recommended that specimens found to be negative for Human Rhinovirus/Enterovirus after examination using BioCode RPP be confirmed by an alternate method (e.g. FDA cleared molecular tests).

Performance characteristics for Influenza A were established when Influenza A H1 2009 Pandemic and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges.

Device Description

The BioCode® Respiratory Pathogen Panel is a respiratory pathogen multiplex nucleic acid test designed for use with the BioCode® MDx-3000 system. The BioCode® MDx-3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for multiple viral and bacterial pathogens from a single nasopharyngeal swab specimen collected in transport media. Specimens are processed and nucleic acids extracted with the NucliSens easyMAG or Roche MagNA Pure 96 automated systems. Once the PCR plate is set up and sealed, all other operations are automated on MDx-3000. The BioCode® RPP simultaneously tests for 17 pathogens and/or subtypes (see table below) from nasopharyngeal swab specimens collected in UTM or VTM. Results from the BioCode RPP test are available within about 5 hours, including off-board nucleic acids extraction.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text.

Based on the provided text, the device is an in vitro diagnostic test (BioCode Respiratory Pathogen Panel - RPP). The primary evidence for meeting acceptance criteria comes from clinical performance studies and analytical performance studies. The document does not explicitly state pre-defined acceptance criteria values (e.g., "Positive Agreement must be > X%"), but rather presents the results achieved and implicitly confirms they meet regulatory expectations for substantial equivalence.

Acceptance Criteria and Reported Device Performance

Since explicit numerical acceptance criteria were not stated, I'll present the reported performance, which implicitly met the necessary thresholds for FDA clearance. The document focuses on Positive Percent Agreement (PA) and Negative Percent Agreement (NA) compared to an FDA-cleared molecular multiplexed respiratory pathogen panel (the "comparator test").

Table of Acceptance Criteria (Implicit) and Reported Device Performance (Clinical Study - Total Specimens)

TargetImplicit Acceptance Criterion (Threshold for Clearance)Reported Device Performance (Positive Agreement (PA) % (95% CI))Reported Device Performance (Negative Agreement (NA) % (95% CI))
Adenovirus(Not explicitly defined, but achieved values were acceptable)87.2% (78.0%, 92.9%)98.4% (97.8%, 98.8%)
Bordetella pertussis(Not explicitly defined, but achieved values were acceptable)100% (34.2%, 100%)99.3% (98.9%, 99.5%)
Chlamydia pneumoniae(Not explicitly defined, but achieved values were acceptable)100% (51.0%, 100%)100% (99.8%, 100%)
Coronavirus(Not explicitly defined, but achieved values were acceptable)83.5% (76.2%, 88.8%)99.1% (98.7%, 99.4%)
Human Metapneumovirus(Not explicitly defined, but achieved values were acceptable)95.1% (90.2%, 97.6%)99.3% (98.9%, 99.6%)
Human Rhinovirus/Enterovirus(Not explicitly defined, but achieved values were acceptable)80.8% (77.0%, 84.1%)98.4% (97.8%, 98.9%)
Influenza A(Not explicitly defined, but achieved values were acceptable)96.4% (93.0%, 98.2%)99.0% (98.5%, 99.3%)
Influenza A H1No positive reference results for this subtypeN/A100% (99.9%, 100%)
Influenza A H1 2009pdm(Not explicitly defined, but achieved values were acceptable)98.1% (90.1%, 99.7%)99.7% (99.3%, 99.8%)
Influenza A H3(Not explicitly defined, but achieved values were acceptable)93.6% (88.7%, 96.5%)99.6% (99.3%, 99.8%)
Influenza B(Not explicitly defined, but achieved values were acceptable)94.4% (84.9%, 98.1%)99.5% (99.1%, 99.7%)
Mycoplasma pneumoniae(Not explicitly defined, but achieved values were acceptable)100% (82.4%, 100%)99.2% (98.8%, 99.5%)
Parainfluenza Virus 1(Not explicitly defined, but achieved values were acceptable)88.2% (65.7%, 96.7%)100% (99.9%, 100%)
Parainfluenza Virus 2(Not explicitly defined, but achieved values were acceptable)83.3% (55.2%, 95.3%)99.9% (99.7%, 100%)
Parainfluenza Virus 3(Not explicitly defined, but achieved values were acceptable)96.7% (91.9%, 98.7%)99.3% (98.9%, 99.6%)
Parainfluenza Virus 4(Not explicitly defined, but achieved values were acceptable)88.9% (67.2%, 96.9%)99.9% (99.7%, 100%)
Respiratory Syncytial Virus(Not explicitly defined, but achieved values were acceptable)98.0% (95.1%, 99.2%)99.1% (98.7%, 99.4%)

Study Details:

  1. Sample Size and Data Provenance:

    • Clinical Test Set (Prospective Study):
      • Sample Size: 2649 residual nasopharyngeal swab (NPS) specimens in VTM or UTM.
      • Data Provenance: Prospectively collected from patients suspected of respiratory tract infections at five geographically diverse clinical sites in the U.S. (August 2017 to May 2019). Specimens were either tested freshly (stored 2-8°C for no more than 7 days) or stored frozen and thawed later.
    • Clinical Test Set (Archived Specimens - Retrospective Study):
      • Sample Size: 165 clinical specimens (archived NPS in VTM or UTM).
      • Data Provenance: Retrospective, preselected archives from source laboratories, chosen because they had previously tested positive for low-prevalence pathogens or were negative. These specimens were then randomized and blinded.
    • Contrived Specimens (Analytical Performance):
      • Chlamydia pneumoniae & Influenza A H1: 50 unique negative natural NPS in VTM or UTM specimens were spiked to create 100 positive samples (2X LOD or greater) and interspersed with negative samples. A total of 110 samples were tested.
      • Reproducibility Study: 6 contrived positive samples and 1 negative sample, each extracted in triplicate, each assayed in singlet. This translates to 90 samples per concentration level per extraction type (easyMAG, MagNA Pure 96) across multiple sites/runs, total 90 (3x LoD) + 90 (1.5x LoD) + 450 (no analyte) = 630 for each virus/bacteria category listed (e.g., Adenovirus, Coronavirus, Human Metapneumovirus).
      • Limit of Detection (LoD): 20 replicates for each target at or near the presumptive LoD.
      • Analytical Reactivity/Inclusivity: Triplicate extractions for each strain/serotype.
      • Analytical Specificity/Cross Reactivity: Triplicate extractions for each off-panel and on-panel organism.
      • Interfering Substances/Microbes: Triplicate extractions for each sample/substance combination.
      • Competitive Inhibition: Triplicate extractions for each pooled sample.
      • Specimen Stability: Triplicate extractions for each time point and storage condition.
      • Matrix Equivalency: Quadruplicate extractions for each pool/matrix combination.
  2. Number of Experts and Qualifications for Ground Truth:

    • The document states that the BioCode RPP test results were compared against those from an "FDA-cleared molecular multiplexed respiratory pathogen panel" (Standard of Care/Comparator Test). This FDA-cleared comparator test serves as the primary "ground truth" for the clinical studies.
    • For discrepant results in the clinical studies, further investigation was conducted by "performing independent molecular tests, including analytically validated PCR followed by bi-directional sequencing assays and alternate NAATs." This implies the use of specialized laboratory personnel with expertise in molecular diagnostics, but the exact number or specific qualifications (e.g., "radiologist with 10 years of experience") are not specified as this is a molecular diagnostic device, not an imaging device typically read by radiologists. The ground truth for presence/absence of pathogens is established by these comparator and confirmatory molecular tests.
  3. Adjudication Method for the Test Set:

    • For the clinical studies (prospective and archived), the method for addressing discrepancies between the BioCode RPP and the comparator test was to perform "independent molecular tests, including analytically validated PCR followed by bi-directional sequencing assays and alternate NAATs." This serves as an adjudication method, where an orthogonal, high-accuracy method is used to determine the true positivity or negativity of discrepant samples. The specific "2+1" or "3+1" concensus type is not explicitly mentioned, but the retesting with confirmatory methods acts as a form of adjudication.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC study was not done. This type of study (comparative effectiveness with human readers) is typical for AI-powered imaging devices where human interpretation is a key component. The BioCode RPP is an in vitro diagnostic (IVD) molecular test where the output is directly from the instrument analysis, not reliant on human interpretation of complex images. The study focuses on comparing the device's accuracy against a legally marketed predicate device/standard of care, not on improving human reader performance.
  5. Standalone Performance (Algorithm Only):

    • Yes, the primary performance evaluation is standalone. The device ("BioCode RPP") is a qualitative multiplexed nucleic acid-based in vitro diagnostic test for use with the "BioCode MDx-3000 Instrument." The instrument integrates PCR, target capture, signal generation, and optical detection. The results are generated by the system (the "algorithm" in this context refers to the instrument's processing logic and interpretation algorithm), and the performance metrics (PA, NA) are reported for this system. There is no human interpretation component in the direct testing process that could be "assisted" by the algorithm.
  6. Type of Ground Truth Used:

    • The ground truth for the clinical test sets (prospective and archived) was established by an FDA-cleared molecular multiplexed respiratory pathogen panel (Standard of Care).
    • For discrepant results, expert molecular laboratory retesting using analytically validated PCR followed by bi-directional sequencing assays and alternate NAATs served as the confirmatory ground truth. This is a form of expert consensus/confirmatory testing using highly accurate molecular methods.
    • For the analytical studies (e.g., LoD, inclusivity, specificity, inhibition), the ground truth was based on known concentrations of purified organisms/nucleic acids or genetically characterized strains (e.g., ATCC strains, Zeptometrix controls). This is a form of analytical ground truth where the content is precisely controlled and known.
  7. Sample Size for the Training Set:

    • The document does not specify a separate "training set" sample size. For IVD devices, especially those based on molecular assays, a distinct "training set" as understood in machine learning (where an algorithm learns from labeled data) is not typically described in the same way. The device's underlying "algorithm" is the biochemical and optical detection system itself, and its performance characteristics are established through extensive analytical and clinical validation, not by training on a large dataset of patient samples in the AI sense. The design and validation are based on principles of molecular biology and traditional assay development.
  8. How the Ground Truth for the Training Set Was Established:

    • As there isn't a "training set" in the machine learning sense, this question isn't directly applicable for this type of IVD device. The development of the assay (e.g., primer and probe design) would rely on known genetic sequences of the target pathogens. The validation data (clinical and analytical studies) demonstrate that the final, developed assay meets its intended performance, rather than being used to train a model in an iterative machine learning manner.

{0}------------------------------------------------

December 23, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Applied BioCode, Inc. Robert Tullio Regulatory Consultant 10020 Pioneer Blvd Suite 102 Santa Fe Springs, California 90670

Re: K192485

Trade/Device Name: BioCode Respiratory Pathogen Panel (RPP) Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC, OZE, OEP, OEM, OOU, OTG, OZX, OZY, OZZ, NSU Dated: December 5, 2019 Received: December 6, 2019

Dear Robert Tullio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tamara Feldblyum, Ph.D. Chief Viral Respiratory and STI Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192485

Device Name BioCode Respiratory Pathogen Panel (RPP)

Indications for Use (Describe)

The BioCode Respiratory Pathogen Panel (RPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BioCode MDx 3000 Instrument. The BioCode of the simultaneous detection and identification of nucleic acids from multiple viruses and bacteria extracted from nasopharyngeal swab (NPS) samples obtained from individuals with signs and/or symptoms of respiratory tract infection. The following pathogens and subtypes are identified using the BioCode RPP:

  • Adenovirus
  • · Coronavirus (229E, OC43, HKU1, and NL63)
  • Human Metapneumovirus A/B
  • · Influenza A, including subtypes H1, H1 2009 Pandemic, and H3
  • Influenza B
  • Parainfluenza Virus 1
  • Parainfluenza Virus 2
  • · Parainfluenza Virus 3
  • · Parainfluenza Virus 4
  • · Respiratory Syncytial Virus A/B
  • Rhinovirus/Enterovirus
  • · Bordetella pertussis
  • Chlamydia pneumoniae
  • Mycoplasma pneumoniae

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the BioCode RPP may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus, the BioCode RPP cannot differentiate them. A positive BioCode RPP Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required. The BioCode RPP detects Human Rhinovirus with reduced sensitivity. If a more accurate Rhinovirus result is required, it is recommended that specimens found to be negative for Human Rhinovirus/Enterovirus after examination using BioCode RPP be confirmed by an alternate method (e.g. FDA cleared molecular tests).

Performance characteristics for Influenza A were established when Influenza A H1 2009 Pandemic and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges.

{3}------------------------------------------------

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

1.0 510(k) SUMMARY

Introduction: According to the requirements of 21 CFR 807.92, the following provides information to understand the basis for a determination of substantial equivalence.

Submitted by:

Applied BioCode®, Inc. 10020 Pioneer Blvd. Suite 102 Santa Fe Springs, CA 90670

Contact:

Robert Di Tullio Regulatory Consultant rditullio@apbiocode.com Telephone: 310 801 1235 Fax: 323 372 3816

Date Submitted: September 9, 2019

Trade Name: BioCode® Respiratory Pathogen Panel (RPP)

Classification Name and Regulation Number:

Respiratory Viral Panel Multiplex Nucleic Acid Assay (21 CFR 866.3980)

Predicate Device:

K170604 – BioFire FilmArray Respiratory Pathogen Panel 2 (RP2)

Intended Use:

BioCode® Respiratory Pathogen Panel (RPP)

The BioCode® Respiratory Pathogen Panel (RPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BioCode MDx-3000 Instrument. The BioCode RPP is capable of the simultaneous detection and identification of nucleic acids from multiple viruses and bacteria extracted from nasopharyngeal swab (NPS) samples obtained from individuals with signs and/or symptoms of respiratory tract infection. The following pathogens and subtypes are identified using the BioCode RPP:

  • Adenovirus
  • Coronavirus (229E, OC43, HKU1, and NL63)
  • Human Metapneumovirus A/B
  • Influenza A, including subtypes H1, H1 2009 Pandemic, and H3
  • Influenza B
  • Parainfluenza 1
  • Parainfluenza 2
  • Parainfluenza 3
  • Parainfluenza 4

{5}------------------------------------------------

  • Respiratory Syncytial Virus A/B
  • Rhinovirus/Enterovirus
  • Bordetella pertussis
  • Chlamydia pneumoniae
  • . Mycoplasma pneumoniae

The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the BioCode RPP may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the genetic similarity between Human Rhinovirus and Enterovirus, the BioCode RPP cannot differentiate them. A positive BioCode RPP Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required. The BioCode RPP detects Human Rhinovirus/Enterovirus with reduced sensitivity. If a more accurate HRV/EV result is required, it is recommended that specimens found to be negative for Human Rhinovirus/Enterovirus after examination using BioCode RPP be confirmed by an alternate method (e.g. FDA cleared molecular tests).

Performance characteristics for Influenza A were established when Influenza A H1 2009 Pandemic and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

{6}------------------------------------------------

Device Description:

The BioCode® Respiratory Pathogen Panel is a respiratory pathogen multiplex nucleic acid test designed for use with the BioCode® MDx-3000 system. The BioCode® MDx-3000 is an automated system that integrates PCR amplification, target capture, signal generation and optical detection for multiple viral and bacterial pathogens from a single nasopharyngeal swab specimen collected in transport media. Specimens are processed and nucleic acids extracted with the NucliSens easyMAG or Roche MagNA Pure 96 automated systems. Once the PCR plate is set up and sealed, all other operations are automated on MDx-3000. The BioCode® RPP simultaneously tests for 17 pathogens and/or subtypes (see table below) from nasopharyngeal swab specimens collected in UTM or VTM. Results from the BioCode RPP test are available within about 5 hours, including off-board nucleic acids extraction.

VirusesViruses
Influenza ARespiratory Syncytial Virus A and B
• Subtype H1Human Metapneumovirus A and B
• Subtype H1 2009pdmRhinovirus/Enterovirus
• Subtype H3Coronavirus (229E, OC43, HKU1, and NL63)
Influenza BAdenovirus
Parainfluenza 1Bacteria
Parainfluenza 2Mycoplasma pneumoniae
Parainfluenza 3Chlamydia pneumoniae
Parainfluenza 4Bordetella pertussis
Internal Control (MS2)

Device Comparison:

Comparison of the BioCode RPP with the Predicate Device

CharacteristicProposed DevicePredicate
NameBioCode® Respiratory PathogenPanel (RPP)BioFire FilmArray RespiratoryPathogen Panel 2 (RP2)
Common NameRespiratory Pathogen PanelMultiplex Nucleic acid assayRespiratory Viral PanelMultiplex Nucleic acid assay
510(k) No.K192485K170604
Regulation21CFR 866.398021CFR 866.3980
Product CodeOCC, OZE, OEM, OOU, OEP,OTG, OZX, OZY, OZZ, NSUOCC, OEM, OOU, OEP, OTG,OZX, OZY, OZZ, OOI
Device ClassIIII

Similarities

{7}------------------------------------------------

Premarket Notification 510(k)

Intended Use
The BioCode® Respiratory Pathogen Panel (RPP) is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BioCode MDx-3000 Instrument. The BioCode RPP is capable of the simultaneous detection and identification of nucleic acids from multiple viruses and bacteria extracted from nasopharyngeal swab (NPS) samples obtained from individuals with signs and/or symptoms of respiratory tract infection. The following pathogens and subtypes are identified using the BioCode RPP:• Adenovirus• Coronavirus (229E, OC43, HKU1, and NL63)• Human Metapneumovirus A/B• Influenza A, including subtypes H1, H1 2009 Pandemic, and H3• Influenza B• Parainfluenza 1• Parainfluenza 2• Parainfluenza 3• Parainfluenza 4• Respiratory Syncytial Virus A/B• Rhinovirus/Enterovirus• Bordetella pertussis• Chlamydia pneumoniae• Mycoplasma pneumoniaeThe FilmArray Respiratory Panel 2 (RP2) is a multiplexed nucleic acid test intended for use with FilmArray® 2.0 or FilmArray® Torch systems for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray RP2:• Adenovirus• Coronavirus 229E• Coronavirus HKU1• Coronavirus NL63• Coronavirus OC43• Human Metapneumovirus• Human Rhinovirus/Enterovirus• Influenza A, including subtypes H1, H1-2009, and H3• Influenza B• Parainfluenza Virus 1• Parainfluenza Virus 2• Parainfluenza Virus 3• Parainfluenza Virus 4• Respiratory Syncytial Virus• Bordetella parapertussis (IS1001)• Bordetella pertussis (ptxP)• Chlamydia pneumoniae• Mycoplasma pneumoniae
The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the BioCode RPP may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence and radiography)The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and/or symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, or lower respiratory tract infection that may not be detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the FilmArray RP2 may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and)
may be necessary when evaluating apatient with possible respiratory tractinfection.Due to the genetic similarity betweenHuman Rhinovirus and Enterovirus, theBioCode RPP cannot differentiate them. Apositive BioCode RPPRhinovirus/Enterovirus result should befollowed up using an alternate method(e.g., cell culture or sequence analysis) ifdifferentiation is required. The BioCodeRPP detects Human Rhinovirus/Enteroviruswith reduced sensitivity. If a more accurateHRV/EV result is required, it isrecommended that specimens found to benegative for Human Rhinovirus/Enterovirusafter examination using BioCode RPP beconfirmed by an alternate method (e.g.FDA cleared molecular tests).Performance characteristics for Influenza Awere established when Influenza A H12009 Pandemic and A H3 were thepredominant Influenza A viruses incirculation. Performance of detectingInfluenza A may vary if other Influenza Astrains are circulating or a novel Influenza Avirus emerges. If infection with a novelInfluenza A virus is suspected based oncurrent clinical and epidemiologicalscreening criteria recommended by publichealth authorities, specimens should becollected with appropriate infectioncontrol precautions for novel virulentInfluenza viruses and sent to state or localhealth departments for testing. Viralculture should not be attempted in thesecases unless a BSL 3+ facility is available toreceive and culture specimens.radiography) may be necessary whenevaluating a patient with possiblerespiratory tract infection.Due to the genetic similarity betweenHuman Rhinovirus and Enterovirus, theFilmArray RP2 cannot reliably differentiatethem. A positive FilmArray RP2Rhinovirus/Enterovirus result should befollowed up using an alternate method (e.g.,cell culture or sequence analysis) ifdifferentiation is required.Performance characteristics for Influenza Awere established when Influenza A H1-2009,A H1, and A H3 were the predominantInfluenza A viruses in circulation.Performance of detecting Influenza A mayvary if other Influenza A strains arecirculating or a novel Influenza A virusemerges. If infection with a novel InfluenzaA virus is suspected based on current clinicaland epidemiological screening criteriarecommended by public health authorities,specimens should be collected withappropriate infection control precautions fornovel virulent Influenza viruses and sent tostate or local health departments for testing.Viral culture should not be attempted inthese cases unless a BSL 3+ facility isavailable to receive and culture specimens.
Sample TypeNasopharyngeal swab in Viral or UniversalTransport MediaSame
ControlsExternally SourcedSame
Differences
MethodologyMultiplex RT-PCR in a single reaction andprobe hybridization followed byfluorescence detection and decoding ofbarcoded magnetic beads (BMB) thatcapture biotinylated products withstreptavidin conjugateNested multiplex PCR executed in twostages. First, a single, large volume, highlymultiplexed reverse transcription PCR (RT-PCR) reaction. Second, nested PCR, isperformed in singleplex fashion in each wellof the array. Followed by fluorescentdetection of images of the PCR reactions.

{8}------------------------------------------------

{9}------------------------------------------------

Summary of Performance Characteristics of the BioCode RPP®

Clinical Performance

Study Overview

The clinical performance of the BioCode RPP was established in a multi-center study conducted during periods of the 2017-2019 respiratory illness seasons. Residual (leftover) and de-identified nasopharyngeal swab (NPS) specimens in VTM or UTM that were prospectively collected from patients suspected of respiratory tract infections at five geographically diverse clinical sites in the U.S. were enrolled and tested with the BioCode RPP at five testing sites during the prospective clinical study. The enrolled prospective specimens were tested freshly with an FDA-cleared molecular multiplexed respiratory pathogen panel as part of the Stand of Care (SOC), and were either tested freshly with the BioCode RPP (i.e., specimens that were stored in a 2-8°C refrigerator for no more than 7 days), or stored frozen and then thawed and tested with the BioCode RPP at a testing site at a later date (i.e., specimens that were initially stored in a 2-8°C refrigerator but were not able to be tested by the BioCode RPP within 7 days from specimen collection).

A waiver of the informed consent requirement was obtained from the Institutional Review Boards (IRBs) at each specimen enrollment site for the use of residual NPS in VTM or UTM specimens.

The following information was recorded on the Case Report Form (CRF) for each subject from whom a specimen was enrolled:

  • · Age and sex
  • Date and time of specimen collection
  • Standard of care (SOC) comparator test result
  • Specimen storage status, i.e., fresh or frozen

A total of 2654 residual NPS specimens in VTM that were prospectively collected at the five clinical sites from August 2017 to May 2019 were enrolled initially for the clinical study. Five specimens were withdrawn from the clinical study due to incomplete data collection and testing, resulted in a total of 2649 prospective specimens (1401 fresh and 1248 frozen specimens) that were included in the prospective clinical study.

The prospective specimens enrolled for evaluation were tested at the five testing sites by trained laboratory personnel. DNA/RNA was extracted using either the BioMerieux NucliSENS easyMAG system or Roche MagNA Pure 96 system. After extraction, the samples were tested using the BioCode RPP on the BioCode MDx-3000 System according to the instructions for use.

CharacteristicProspective Study Specimens
Total Specimens2649
Gender (n/N (%))
Male1346/2649 (50.8%)

Demographics - Prospective Samples

{10}------------------------------------------------

CharacteristicProspective Study Specimens
Female1303/2649 (49.2%)
Age Category (n/N (%))
0-5 yrs1004/2649 (37.9%)
6-21 yrs609/2649 (23.0%)
22-59 yrs531/2649 (20.0%)
60+ yrs505/2649 (19.1%)
Status (n/N (%))a
Inpatient1555/2646 (58.8%)
Outpatient1091/2646 (41.2%)

a - Inpatient/outpatient status was unavailable for 3 specimens.

Storage MethodExtraction Method
SiteSamples TestedFreshFrozeneasyMAGMagNA Pure 96
Site 015302502805300
Site 024191822370419
Site 036003662346000
Site 045503002505500
Site 055503032470550
Total:2649140112481680969

Prospective Sample Type and Test Method Breakdown

Performance of the BioCode RPP was evaluated by comparing the BioCode RPP test results with those from an FDA-cleared molecular multiplexed respiratory pathogen panel. Positive agreement was calculated as TP/(TP + FN). TP = true positive by both the comparator test and BioCode RPP; FN = false negative or negative by BioCode RPP only. Negative agreement was calculated as TN/(TN + FP). TN = true negative or negative by both the comparator test and BioCode RPP; FP = false positive or positive by BioCode RPP only. The two-sided 95% confidence interval was calculated with Score method (per CLSI EP12-A2).

Samples for which false positive and/or false negative results (i.e., discrepant results) were obtained when comparing the BioCode RPP results to the comparator test results were further investigated. The discrepancy investigation was mainly conducted by performing independent molecular tests, including analytically validated PCR followed by bi-directional sequencing assays and alternate NAATs.

Of the 2649 specimens included in the prospective clinical study, two specimens, one fresh and one frozen specimen, obtained a final "invalid" result from the BioCode RPP, and were excluded from the performance analyses for all analytes. In addition, three specimens, one fresh and two frozen specimens, obtained a final influenza A "indeterminate" result by the BioCode RPP, and two specimens, one fresh and one frozen, obtained an influenza A "equivocal" result from the comparator test. They were included in the performance analyses for all analytes but excluded from the performance

{11}------------------------------------------------

calculations for Flu A and Flu A subtypes. Furthermore, two frozen specimens obtained a valid influenza A result from the comparator test without the accompanying Flu A subtyping results. They were included in the performance analyses for all analytes but excluded from the performance calculations for Flu A subtypes.

The Prospective study results stratified by storage condition are presented in the table below.

Positive AgreementNegative Agreement
TargetStorage(n)PA (%)95% ClNA (%)95% Cl
AdenovirusaFresh140031/40 (77.5%)(62.5%, 87.7%)1340/1360 (98.5%)(97.7%, 99.0%)
Frozen124737/38 (97.4%)(86.5%, 99.5%)1188/1209 (98.3%)(97.4%, 98.9%)
Total264768/78 (87.2%)(78.0%, 92.9%)2528/2569 (98.4%)(97.8%, 98.8%)
Bordetella pertussisbFresh14001/1 (100%)(20.7%, 100%)1387/1399 (99.1%)(98.5%, 99.5%)
Frozen12471/1 (100%)(20.7%, 100%)1239/1246 (99.4%)(98.8%, 99.7%)
Total26472/2 (100%)(34.2%, 100%)2626/2645 (99.3%)(98.9%, 99.5%)
Chlamydia pneumoniaecFresh14002/2 (100%)(34.2%, 100%)1397/1398 (99.9%)(99.6%, 100%)
Frozen12472/2 (100%)(34.2%, 100%)1245/1245 (100%)(99.7%, 100%)
Total26474/4 (100%)(51.0%, 100%)2642/2643 (100%)(99.8%, 100%)
CoronavirusdFresh140035/50 (70%)(56.2%, 80.9%)1338/1350 (99.1%)(98.5%, 99.5%)
Frozen124776/83 (91.6%)(83.6%, 95.9%)1154/1164 (99.1%)(98.4%, 99.5%)
Total2647111/133 (83.5%)(76.2%, 88.8%)2492/2514 (99.1%)(98.7%, 99.4%)
Human MetapneumoviruseFresh140089/93 (95.7%)(89.5%, 98.3%)1299/1307 (99.4%)(98.8%, 99.7%)
Frozen124746/49 (93.9%)(83.5%, 97.9%)1189/1198 (99.2%)(98.6%, 99.6%)
Total2647135/142 (95.1%)(90.2%, 97.6%)2488/2505 (99.3%)(98.9%, 99.6%)
Human Rhinovirus/EnterovirusfFresh1400221/261 (84.7%)(79.8%, 88.5%)1119/1139 (98.2%)(97.3%, 98.9%)
Frozen1247162/213 (76.1%)(69.9%, 81.3%)1020/1034 (98.6%)(97.7%, 99.2%)
Total2647383/474 (80.8%)(77.0%, 84.1%)2139/2173 (98.4%)(97.8%, 98.9%)
Influenza AgFresh1398115/120 (95.8%)(90.6%, 98.2%)1265/1278 (99.0%)(98.3%, 99.4%)
Frozen124498/101 (97.0%)(91.6%, 99.0%)1131/1143 (99.0%)(98.2%, 99.4%)
Total2642213/221 (96.4%)(93.0%, 98.2%)2396/2421 (99.0%)(98.5%, 99.3%)
Influenza A H1Fresh1398N/A+N/A+1398/1398 (100%)(99.7%, 100%)
Frozen1242N/A+N/A+1242/1242 (100%)(99.7%, 100%)
Total2640N/A+N/A+2640/2640 (100%)(99.9%, 100%)
Influenza A H1 2009pdmhFresh139829/30 (96.7%)(83.3%, 99.4%)1365/1368 (99.8%)(99.4%, 99.9%)
Frozen124223/23 (100%)(85.7%, 100%)1213/1219 (99.5%)(98.9%, 99.8%)
Total264052/53 (98.1%)(90.1%, 99.7%)2578/2587 (99.7%)(99.3%, 99.8%)
Positive AgreementNegative Agreement
TargetStorage(n)PA (%)95% CINA (%)95% CI
Influenza A H3Fresh139882/88 (93.2%)(85.9%, 96.8%)1306/1310 (99.7%)(99.2%, 99.9%)
Frozen124265/69 (94.2%)(86.0%, 97.7%)1168/1173 (99.6%)(99.0%, 99.8%)
Total2640147/157 (93.6%)(88.7%, 96.5%)2474/2483 (99.6%)(99.3%, 99.8%)
Influenza BFresh14007/7 (100%)(64.6%, 100%)1388/1393 (99.6%)(99.2%, 99.8%)
Frozen124744/47 (93.6%)(82.8%, 97.8%)1191/1200 (99.2%)(98.6%, 99.6%)
Total264751/54 (94.4%)(84.9%, 98.1%)2579/2593 (99.5%)(99.1%, 99.7%)
Mycoplasma pneumoniaeFresh14008/8 (100%)(67.6%, 100%)1381/1392 (99.2%)(98.6%, 99.6%)
Frozen124710/10 (100%)(72.2%, 100%)1228/1237 (99.3%)(98.6%, 99.6%)
Total264718/18 (100%)(82.4%, 100%)2609/2629 (99.2%)(98.8%, 99.5%)
Parainfluenza Virus 1Fresh14004/4 (100%)(51.0%, 100%)1396/1396 (100%)(99.7%, 100%)
Frozen124711/13 (84.6%)(57.8%, 95.7%)1234/1234 (100%)(99.7%, 100%)
Total264715/17 (88.2%)(65.7%, 96.7%)2630/2630 (100%)(99.9%, 100%)
Parainfluenza Virus 2Fresh14002/3 (66.7%)(20.8%, 93.9%)1396/1397 (99.9%)(99.6%, 100%)
Frozen12478/9 (88.9%)(56.5%, 98.0%)1236/1238 (99.8%)(99.4%, 100%)
Total264710/12 (83.3%)(55.2%, 95.3%)2632/2635 (99.9%)(99.7%, 100%)
Parainfluenza Virus 3Fresh140077/79 (97.5%)(91.2%, 99.3%)1312/1321 (99.3%)(98.7%, 99.6%)
Frozen124741/43 (95.3%)(84.5%, 98.7%)1196/1204 (99.3%)(98.7%, 99.7%)
Total2647118/122 (96.7%)(91.9%, 98.7%)2508/2525 (99.3%)(98.9%, 99.6%)
Parainfluenza Virus 4Fresh14001/1 (100%)(20.7%, 100%)1399/1399 (100%)(99.7%, 100%)
Frozen124715/17 (88.2%)(65.7%, 96.7%)1228/1230 (99.8%)(99.4%, 100%)
Total264716/18 (88.9%)(67.2%, 96.9%)2627/2629 (99.9%)(99.7%, 100%)
Respiratory Syncytial VirusFresh140091/93 (97.8%)(92.5%, 99.4%)1293/1307 (98.9%)(98.2%, 99.4%)
Frozen1247109/111 (98.2%)(93.7%, 99.5%)1129/1136 (99.4%)(98.7%, 99.7%)
Total2647200/204 (98.0%)(95.1%, 99.2%)2422/2443 (99.1%)(98.7%, 99.4%)

Table. Summary of Clinical Study results: Prospective specimens stratified by storage condition.

{12}------------------------------------------------

  • No positive reference results recorded

a – Adenovirus: The 10 FNs were not detected by PCR/bi-directional sequencing or alternative NAAT. Of 41 FPs, 37 were not detected and 4 were indeterminate by PCR/bi-directional sequencing.

b – Bordetella pertussis: Of 19 FPs, 4 were detectional sequencing, 3 were indeterminate and 12 were not detected by PCR/bi-directional sequencing.

c – Chlamydia pneumoniae: The 1 FP was detected by PCR/bi-directional sequencing.

d – Coronavirus: Of 22 FNs, 5 were detectional sequencing. 2 were not detected by PCR/bi-directional sequencing but detected by alternative NAAT. 12 were not detected by either alternative NAAT or PCR/bi-directional sequencing. 3 were not detected by PCR/bi-directional sequencing and were not tested by alternative NAAT. The 22 FPs were not detectional sequencing.

e – Human Metapneumovirus: Of 7 FNs, 3 were detected by PCR/bi-directional sequencing. Of 17 FPs, 7 were detected while 10 were not detected by PCR/bi-directional sequencing.

f – Human Rhinovirus/Enterovirus: Of 91 FNs, 26 were indeterminate by PCR/bi-directional sequencing. 25 were not detected by PCR/bi-directional sequencing but were detected by alternative NAAT. 11 were not detectional

{13}------------------------------------------------

sequencing or by alternative NAAT. 27 were not detectional sequencing and had insufficient volume for alternative NAAT. Of 34 FPs, 8 were detected and 26 were not detected by PCR/bi-directional sequencing.

g – Influenza A: Of the 8 FNs, 7 were not detectional sequencing. 1 had insufficient volume for follow-up testing. Of 25 FPS, 5 were detected by PCR/bi-directional sequencing, 19 were not detectional sequencing, and 1 had insufficient volume for follow-up testing.

h – Influenza A H1 2009pdm: The 1 FN had insufficient volume for follow-up testing. Of 9 FPs, 4 were detectional sequencing, and 5 were not detected by PCR/bi-directional sequencing.

i – Influenza A H3: Of 10 FNs, 7 were not detected by PCR/bi-directional sequencing. 2 had insufficient volume for followup testing. Of 9 FPs, 3 were detected, 1 was indeterminate, and 4 were not detectional sequencing. 1 had insufficient volume for follow-up testing.

j – Influenza B: The 3 FNs were not detected by PCR/bi-directional sequencing. Of 14 FPs, 1 was detected by PCR/bidirectional sequencing, and 1 had insufficient volume for follow-up testing.

k – Mycoplasma pneumonioe: Of 20 FPs, 7 were detected, and 1 was invalid by PCR/bi-directional sequencing. 2 had insufficient volume for follow-up testing.

I – Parainfluenza Virus 1: The 2 FNs were not detected by PCR/bi-directional sequencing.

m – Parainfluenza Virus 2: Of 2 FNs, 1 was detected by PCR/bi-directional sequencing. The 3 FPs were detected by PCR/bi-directional sequencing.

n – Parainfluenza Virus 3: Of 4 FNs, 2 were detected, and 1 was indeterminate, by PCR/bi-directional sequencing. Of 17 FPs, 8 were detected, 8 were not detected, and 1 was indeterminate, by PCR/bi-directional sequencing.

o – Parainfluenza Virus 4: The 2 FNs were detected by PCR/bi-directional sequencing. The 2 FPs were detectional sequencing.

p – Respiratory Syncytial Virus: The 4 FNs were not detectional sequencing. Of 21 FPs, 18 were not detected, 2 were detected, and 1 was indeterminate, by PCR/bi-directional sequencing.

Prospective prevalence as detected by BioCode® RPP, stratified by age or study site, are presented in tables below.

Table. Prospective prevalence detected by BioCode® RPP stratified by Age

AnalyteOverall(N=2647)≤5 yrs(N=1004)6-21 yrs(N=609)22-59 yrs(N=531)60+ yrs(N=503)
Adenovirus109 (4.1%)80 (8.0%)20 (3.3%)5 (0.9%)4 (0.8%)
Bordetella pertussis21 (0.8%)8 (0.8%)12 (2.0%)0 (0.0%)1 (0.2%)
Chlamydia pneumoniae5 (0.2%)1 (0.1%)3 (0.5%)1 (0.2%)0 (0.0%)
Coronavirus133 (5.0%)63 (6.3%)27 (4.4%)19 (3.6%)24 (4.8%)
Human Metapneumovirus152 (5.7%)94 (9.4%)26 (4.3%)15 (2.8%)17 (3.4%)
Human Rhinovirus/Enterovirus417 (15.8%)234 (23.3%)101 (16.6%)51 (9.6%)31 (6.2%)
Influenza A238 (9.0%)71 (7.1%)84 (13.8%)47 (8.9%)36 (7.2%)
Influenza A H10 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Influenza A H1 2009pdm62 (2.3%)24 (2.4%)15 (2.5%)15 (2.8%)8 (1.6%)
Influenza A H3157 (5.9%)45 (4.5%)61 (10.0%)26 (4.9%)25 (5.0%)
Influenza B65 (2.5%)13 (1.3%)26 (4.3%)15 (2.8%)11 (2.2%)
Mycoplasma pneumoniae38 (1.4%)9 (0.9%)21 (3.4%)6 (1.1%)2 (0.4%)
Parainfluenza Virus 115 (0.6%)4 (0.4%)1 (0.2%)4 (0.8%)6 (1.2%)
Parainfluenza Virus 213 (0.5%)6 (0.6%)4 (0.7%)3 (0.6%)0 (0.0%)
Parainfluenza Virus 3135 (5.1%)74 (7.4%)21 (3.4%)21 (4.0%)19 (3.8%)

{14}------------------------------------------------

AnalyteOverall(N=2647)≤5 yrs(N=1004)6-21 yrs(N=609)22-59 yrs(N=531)60+ yrs(N=503)
Parainfluenza Virus 418 (0.7%)12 (1.2%)1 (0.2%)2 (0.4%)3 (0.6%)
Respiratory Syncytial Virus221 (8.3%)156 (15.5%)30 (4.9%)19 (3.6%)16 (3.2%)

Table. Prospective prevalence detected by BioCode® RPP stratified by site.

AnalyteOverall(N=2647)Site 1(N=529)Site 2(N=419)Site 3(N=599)Site 4(N=550)Site 5(N=550)
Adenovirus109 (4.1%)8 (1.5%)27 (6.4%)21 (3.5%)17 (3.1%)36 (6.5%)
Bordetella pertussis21 (0.8%)0 (0.0%)19 (4.5%)1 (0.2%)0 (0.0%)1 (0.2%)
Chlamydia pneumoniae5 (0.2%)0 (0.0%)3 (0.7%)0 (0.0%)1 (0.2%)1 (0.2%)
Coronavirus133 (5.0%)36 (6.8%)21 (5.0%)14 (2.3%)26 (4.7%)36 (6.5%)
Human Metapneumovirus152 (5.7%)18 (3.4%)22 (5.3%)26 (4.3%)38 (6.9%)48 (8.7%)
Human Rhinovirus/Enterovirus417 (15.8%)48 (9.1%)83 (19.8%)65 (10.9%)125 (22.7%)96 (17.5%)
Influenza A238 (9.0%)49 (9.3%)50 (11.9%)35 (5.8%)41 (7.5%)63 (11.5%)
Influenza A H10 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Influenza A H1 2009pdm62 (2.3%)13 (2.5%)7 (1.7%)11 (1.8%)22 (4.0%)9 (1.6%)
Influenza A H3157 (5.9%)32 (6.0%)40 (9.5%)19 (3.2%)17 (3.1%)49 (8.9%)
Influenza B65 (2.5%)12 (2.3%)32 (7.6%)15 (2.5%)2 (0.4%)4 (0.7%)
Mycoplasma pneumoniae38 (1.4%)4 (0.8%)15 (3.6%)5 (0.8%)3 (0.5%)11 (2.0%)
Parainfluenza Virus 115 (0.6%)4 (0.8%)0 (0.0%)11 (1.8%)0 (0.0%)0 (0.0%)
Parainfluenza Virus 213 (0.5%)0 (0.0%)0 (0.0%)3 (0.5%)8 (1.5%)2 (0.4%)
Parainfluenza Virus 3135 (5.1%)12 (2.3%)28 (6.7%)41 (6.8%)9 (1.6%)45 (8.2%)
Parainfluenza Virus 418 (0.7%)0 (0.0%)1 (0.2%)9 (1.5%)5 (0.9%)3 (0.5%)
Respiratory Syncytial Virus221 (8.3%)21 (4.0%)35 (8.4%)25 (4.2%)49 (8.9%)91 (16.5%)

The overall success rate for initial specimen testing in the prospective study was 98.8% (2618/2649) (95% Cl: 98.3% - 99.2%); 31 tests were unsuccessful (26 tests with an invalid result and 5 tests due to low BMB count/instrument error). Upon a single retest per the instructions for use, 29 of the 31 initially unsuccessful specimens generated a valid result. The final validity rate was 99.9% (2647/2649) (95% Cl: 99.7%-100%).

{15}------------------------------------------------

There were 193 samples with mixed infections by the BioCode® RPP in the prospective clinical study (193/2649 or 7.3%). The distribution, prevalence and most common co-infections combinations detected by BioCode RPP in the prospective clinical study are summarized in the tables below.

Table. Distribution of co-infection combinations detected by BioCode® RPP from prospective clinical study

Analytes DetectedSimultaneouslyNumber of Specimen
2168 (87.0%)
324 (12.4%)
51 (0.5%)
Total Co-Infections193

Table. Prevalence of targets in co-infection combinations detected by BioCode® RPP from prospective clinical study

AnalytePrevalence in Co-Infections (N=193)
Adenovirus51 (26.4%)
Bordetella pertussis14 (7.3%)
Chlamydia pneumoniae1 (0.5%)
Coronavirus47 (24.4%)
Human Metapneumovirus35 (18.1%)
Human Rhinovirus/Enterovirus101 (52.3%)
Influenza A37 (19.2%)
Influenza B8 (4.1%)
Mycoplasma pneumoniae7 (3.6%)
Parainfluenza Virus 12 (1.0%)
Parainfluenza Virus 24 (2.1%)
Parainfluenza Virus 339 (20.2%)
Parainfluenza Virus 48 (4.1%)
Respiratory Syncytial Virus59 (30.1%)

Table. Most prevalent multiple detection combinations (5 or more instances) detected by BioCode® RPP from prospective clinical study.

Co-Infection CombinationNumber of Specimen
Human Rhinovirus/Enterovirus + Respiratory Syncytial Virus17
Adenovirus + Human Rhinovirus/Enterovirus13
Human Metapneumovirus + Human Rhinovirus/Enterovirus13
Human Rhinovirus/Enterovirus + Influenza A11
Adenovirus + Respiratory Syncytial Virus10
Human Rhinovirus/Enterovirus + Parainfluenza Virus 310
Coronavirus + Human Rhinovirus/Enterovirus8
Coronavirus + Respiratory Syncytial Virus7
Bordetella pertussis + Human Rhinovirus/Enterovirus6

{16}------------------------------------------------

Co-Infection CombinationNumber of Specimen
Adenovirus + Parainfluenza Virus 35
Coronavirus + Human Metapneumovirus5
Coronavirus + Influenza A5

Testing of Preselected Archived Specimens

Some of the pathogens on the BioCode RPP were of low prevalence and were not encountered in sufficiently large numbers during the prospective study to adequately demonstrate system performance. To supplement the results of the prospective clinical study, an evaluation of preselected archived retrospective specimens was performed. These specimens were archived NPS in VTM or UTM specimens that were selected because they had previously tested positive for one of the following pathogens at the source laboratory: coronavirus 229E, coronavirus HKU1, parainfluenza virus 1, parainfluenza virus 2, parainfluenza virus 4, Bordetella pertussis, Chlamydia pneumoniae, and Mycoplasma pneumoniae, or had been negative in previous laboratory testing.

A total of 165 clinical specimens were enrolled for testing in this retrospective study. The specimens were randomized such that the users performing the BioCode RPP assay were blinded to the expected test result and shipped to one of two of the testing sites participated in the prospective clinical study for testing.

A summary of the demographic information of the tested samples is provided in the following table:

CharacteristicArchived Study Specimens
Total Specimens165
Gender (n/N (%))
Male96/165 (58.2%)
Female69/165 (41.8%)
Age Category (n/N (%))
0-5 yrs77/165 (46.7%)
6-21 yrs58/165 (35.2%)
22-59 yrs15/165 (9.1%)
60+ yrs15/165 (9.1%)

Demographics of Archived Specimens

The performance of the BioCode RPP was evaluated by comparing the BioCode RPP test results with those from an FDA-cleared molecular multiplexed respiratory pathogen panel, the same panel test as the one used as the comparator in the prospective clinical study. The BioCode RPP retrospective performance data expressed as positive percent and negative percent agreements against the comparator method are presented by pathogen in the table below.

{17}------------------------------------------------

Target(n)PA (%)95% CINA (%)95% CI
Adenovirusa1657/7 (100%)(64.6%, 100%)155/158 (98.1%)(94.6%, 99.4%)
Bordetella pertussisb16510/10 (100%)(72.2%, 100%)144/155 (92.9%)(87.7%, 96.0%)
Chlamydia pneumoniae16510/10 (100%)(72.2%, 100%)155/155 (100%)(97.6%, 100%)
Coronavirusc16552/59 (88.1%)(77.5%, 94.1%)99/106 (93.4%)(87.0%, 96.8%)
Human Metapneumovirus1654/4 (100%)(51.0%, 100%)161/161 (100%)(97.7%, 100%)
Human Rhinovirus/Enterovirusd16516/23 (69.6%)(49.1%, 84.4%)141/142 (99.3%)(96.1%, 99.9%)
Influenza A165N/A+N/A+165/165 (100%)(97.7%, 100%)
Influenza A H1165N/A+N/A+165/165 (100%)(97.7%, 100%)
Influenza A H1 2009pdm165N/A+N/A+165/165 (100%)(97.7%, 100%)
Influenza A H3165N/A+N/A+165/165 (100%)(97.7%, 100%)
Influenza Be1652/3 (66.7%)(20.8%, 93.9%)162/162 (100%)(97.7%, 100%)
Mycoplasma pneumoniaef1657/7 (100%)(64.6%, 100%)153/158 (96.8%)(92.8%, 98.6%)
Parainfluenza Virus 1g16512/13 (92.3%)(66.7%, 98.6%)152/152 (100%)(97.5%, 100%)
Parainfluenza Virus 2h16519/20 (95%)(76.4%, 99.1%)144/145 (99.3%)(96.2%, 99.9%)
Parainfluenza Virus 31651/1 (100%)(20.7%, 100%)164/164 (100%)(97.7%, 100%)
Parainfluenza Virus 4i16514/15 (93.3%)(70.2%, 98.8%)150/150 (100%)(97.5%, 100%)
Respiratory Syncytial Virusj16511/12 (91.7%)(64.6%, 98.5%)152/153 (99.3%)(96.4%, 99.9%)

Table. Results from Archived Specimens tested by the BioCode® RPP with easyMAG extraction system.

  • No positive reference results recorded

a - Adenovirus: Of 3 FPs, 1 was detected by PCR/bi-directional sequencing and 2 were not detected by PCR/bi-directional sequencing.

b - Bordetella pertussis: Of 11 FPs, 7 were detected by PCR/bi-directional sequencing and 4 were not detected by PCR/bidirectional sequencing.

c - Coronavirus: Of 7 FNs, 4 were detected by PCR/bi-directional sequencing while 3 were not detectional sequencing. The 7 FPs were not detected by PCR/bi-directional sequencing. All had low MFls (<620) near the MFL cut off on initial BioCode RPP results.

d - Human Rhinovirus/Enterovirus: The 7 FNs were not detected by PCR/bi-directional sequencing. The 1 FP was not detected by PCR/bi-directional sequencing.

e - Influenza B: 1 FN was not detected by PCR/bi-directional sequencing.

f - Mycoplasma pneumoniae: Of 5 FPs, 3 were detected by PCR/bi-directional sequencing and 2 were not detected by PCR/bi-directional sequencing.

g - Parainfluenza Virus 1: 1 FN was not detected by PCR/bi-directional sequencing.

h - Parainfluenza Virus 2: 1 FN was not detected by PCR/bi-directional sequencing. 1 FP was not detected by PCR/bidirectional sequencing.

i - Parainfluenza Virus 4: 1 FN was not detected by PCR/bi-directional sequencing.

j - Respiratory Syncytial Virus: 1 FN was not detected by PCR/bi-directional sequencing. 1 FP was not detected by PCR/bidirectional sequencing.

{18}------------------------------------------------

Testing of Contrived Specimens

Some analytes are so rare that both prospective and archived specimen collection efforts were insufficient to demonstrate the clinical performance. To supplement the prospective and archived data, an evaluation of contrived specimens was performed for two pathogens: Chlamydia pneumoniae and Influenza A H1. These contrived clinical specimens were prepared using 50 unique natural NPS in VTM or UTM specimens that were previously tested negative for all BioCode RPP analytes. Contrived specimens were spiked at concentrations of 2X LOD or greater using different strains for each pathogen. The 50 positive samples of each pathogen were prepared, interspersed with negative samples and randomized before testing at one of the five testing sites participated in the prospective clinical study. A total of 110 samples, including 100 positives, were tested. The results of the BioCode RPP testing are presented in the following table:

OrganismSourceStrain/IsolateFoldLoDConcentrationPA (%)95% CINA (%)95% CI
ATCC 53592AR-39233.4 CFU/mL9/9 (100%)70.1%, 100%
10167 CFU/mL5/5 (100%)56.6%, 100%
1001670 CFU/mL4/4 (100%)51.0%, 100%
ChlamydiapneumoniaeATCC VR-1360CM-1233.4 CFU/mL8/8 (100%)67.6%, 100%60/60(100%)94.0%,100%
10167 CFU/mL5/5 (100%)56.6%, 100%
1001670 CFU/mL3/3 (100%)43.9%, 100%
ATCC VR-1310CWL-029233.4 CFU/mL8/8 (100%)67.6%, 100%
10167 CFU/mL5/5 (100%)56.6%, 100%
1001670 CFU/mL3/3 (100%)43.9%, 100%
Combined50/50 (100%)92.9%, 100%
Influenza AH1N1Zeptometrix0810036CFA/NewCaledonia/20/99230 TCID50/ mL5/5 (100%)56.6%, 100%94.0%,100%
10150 TCID50/ mL3/3 (100%)43.9%, 100%
1001500 TCID50/ mL3/3 (100%)43.9%, 100%
Zeptometrix0810247CFA/Taiwan/42/06230 TCID50/ mL5/5 (100%)56.6%, 100%
10150 TCID50/ mL3/3 (100%)43.9%, 100%
1001500 TCID50/ mL2/2 (100%)34.2%, 100%
Zeptometrix0810246CFSingapore/63/04230 TCID50/ mL5/5 (100%)56.6%, 100%60/60(100%)
10150 TCID50/ mL2/2 (100%)
1001500 TCID50/ mL2/2 (100%)
VirapurA/Denver/1/1957230 TCID50/ mL5/5 (100%)56.6%, 100%
10150 TCID50/ mL2/2 (100%)34.2%, 100%
1001500 TCID50/ mL2/2 (100%)34.2%, 100%
ATCC VR-219A/NWS/33254 TCID50/ mL5/5 (100%)56.6%, 100%
10270 TCID50/ mL3/3 (100%)43.9%, 100%
1002700 TCID50/ mL3/3 (100%)43.9%, 100%
Combined50/50 (100%)92.9%, 100%

Table. Results from contrived samples tested by the BioCode® RPP using the easyMAG extraction system.

{19}------------------------------------------------

Performance of External Controls during Clinical Trials

During clinical evaluation of the BioCode® RPP, at least one negative control (NC) was included in each run. The negative control was UTM that was taken through all steps (extraction, amplification, and detection). It was recommended that external controls consisting of 5 pools of inactivated organism from ZeptoMetrix (catalog no. NATRPP-ABC) be assayed on a rotating basis.

PoolPanel MemberStrainDilution Factor
PC-AInfluenza BB/Florida/02/061/4
Respiratory Syncytial Virus AN/A1/4
Parainfluenza virus Type 1N/A1/4
Rhinovirus 1AN/A1/4
PC-BHKU1 ConstructN/A1/4
Parainfluenza virus Type 2N/A1/4
Metapneumovirus 8Peru6-20031/4
Bordetella pertussisA6391/4
PC-CInfluenza A H1N1A/New Caledonia/20/991/4
Coronavirus NL63N/A1/4
Adenovirus Type 3N/A1/4
PC-DInfluenza A 2009 H1N1pdmA/NY/02/091/4
Parainfluenza virus Type 4N/A1/4
Coronavirus OC43N/A1/4
Chlamydia pneumoniaeCWL-0291/4
PC-EInfluenza A H3A/Brisbane/10/071/4
Parainfluenza virus Type 3N/A1/4
Coronavirus 229EN/A1/4
Mycoplasma pneumoniaeM1291/4

Table. Recommended positive control pools formulation for inactivated organism from ZeptoMetrix.

Table. Performance of controls during Clinical trials
------------------------------------------------------------------
PC-APC-BPC-CPC-DPC-ENC
easyMAG9a/108a/96/611/118/850b/53
MP964/45/54/43/33/320/20

a - Operator error NC/PC switched during loading

b - Of 3 failed NCs, 1 was due to RNA IC failure, 1 was due to operator error (NC/PC switched on loading), 1 was due to detection of unexpected target (FP).

{20}------------------------------------------------

General Performance of Assay During Clinical Trials

Table. Accounting of valid and invalid runs during clinical trials (prospective specimens)
-----------------------------------------------------------------------------------------------------------
DescriptionNumber% of Total
Valid runs with complete results5786.4%
Completed Runs with NC failures34.5%
Partially or completely invalid runs due to other failures69.1%
Incomplete runs due to instrument failures00.0%
Total66100%

Table. Summary of issues causing Invalid runs during clinical trials (prospective specimens)

Reason for failureNumber% of Total
User Error444.4%
Instrument/ Alignmenta222.2%
Negative controlb333.3%
Total invalid runs9100%

a – 2 consecutive failures for alignment error that did not happen again after adjustment.

b - Of 3 failed NCs, 1 was due to RNA IC failure, 1 was due to operator error (NC and PC switched on loading), 1 was due to detection of unexpected target (FP).

{21}------------------------------------------------

Analytical Performance

The results of the analytical studies summarized in the following paragraphs met the acceptance criteria and demonstrated the analytical performance characteristics of the proposed BioCode RPP.

Reproducibility Study

A study was performed to assess the Reproducibility of the BioCode® RPP on the BioCode® MDx-3000. Front end extraction was performed using the NucliSENS easyMAG and MagNA Pure 96 systems. This study was designed to assess intra-assay (within run), inter-assay (run-to-run), day-to-day and site-tosite reproducibility. One lot of reagents was assayed at 3 sites with easyMAG and 1 site with MP96) by 2 operators on 1 instrument per site for 5 days (total of 10 runs per site). The reproducibility panel consisted of 6 contrived positive samples and 1 negative sample, each extracted in triplicate and each assayed in singlet. The samples consisted of combinations of 12 representative targets at 1.5x LoD (Low) and 3x LoD (Medium) spiked into simulated NPS in UTM matrix (see table below).

Medium (3x LoD)Medium (3x LoD)Low (1.5x LoD)Low (1.5x LoD)Sample name
Human RhinovirusParainfluenza Virus 2Human MetapneumovirusBordetella pertussisRP1
Human MetapneumovirusBordetella pertussisHuman RhinovirusParainfluenza Virus 2RP2
Influenza BCoronavirus NL63Chlamydia pneumoniaeParainfluenza Virus 3RP3
Chlamydia pneumoniaeParainfluenza Virus 3Influenza BCoronavirus NL63RP4
Influenza A H3N2Mycoplasma pneumoniaeRespiratory Syncytial VirusAdenovirus CRP5
Respiratory Syncytial VirusAdenovirus CInfluenza A H3N2Mycoplasma pneumoniaeRP6
Simulated Negative matrix

Table. Reproducibility panel

For each target, the results were determined according to the Interpretation Algorithm. Percent positive agreement was calculated for medium and low positives separately. Samples not containing said target were used to calculate percent negative agreement for each RPP target (see data tables below).

Conclusions: Acceptance criteria were met.

{22}------------------------------------------------

Results

Table. Results from the multi-site reproducibility study (viruses)
--------------------------------------------------------------------------------
ConcentrationTestedExpectedResultAgreement with Expected ResultNucliSENS easyMAGMagNA Pure 96All Sites
AnalyteSite 1Site 3Sub-TotalSite 2Sub-Total(95% CI)
Viruses
Adenovirus3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Adenovirus1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
AdenovirusNone(no analyte)NotDetected150/150100%149/15099.30%299/30099.70%150/150100%150/150100%449/45099.80%(98.8%-100%)
Coronavirus3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Coronavirus1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
CoronavirusNone(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Human Metapneumovirus3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Human Metapneumovirus1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Human MetapneumovirusNone(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Human Rhinovirus/Enterovirus3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Human Rhinovirus/Enterovirus1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Human Rhinovirus/EnterovirusNone(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Influenza A/H33x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Influenza A/H31.5x LoDDetected29/30 a96.70%30/30100%59/6098.30%30/30100%30/30100%89/9098.90%(94.0%-99.8%)
Influenza A/H3None(no analyte)NotDetected150/150100%149/150 b99.30%299/30099.70%150/150100%150/150100%449/45099.80%(98.8%-100%)
Influenza A/H1pdm09None(no analyte)NotDetected210/210100%210/210100%420/420100%210/210100%210/210100%630/630100%(99.4%-100%)
Influenza A/H1None(no analyte)NotDetected210/210100%210/210100%420/420100%210/210100%210/210100%630/630100%(99.4%-100%)
Influenza B3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)

{23}------------------------------------------------

Premarket Notification 510(k)
---------------------------------
AnalyteConcentrationTestedExpectedResultAgreement with Expected Result
NucliSENS easyMAGMagNA Pure 96All Sites(95% CI)
Site 1Site 3Sub-TotalSite 2Sub-Total
100%100%100%100%100%100%(95.9%-100%)
1.5x LoDDetected29/3096.70%30/30100%59/6098.30%30/30100%30/30100%89/9098.90%(94.0%-99.8%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Parainfluenza Virus 1None(no analyte)NotDetected210/210100%210/210100%420/420100%210/210100%210/210100%630/630100%(99.4%-100%)
3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Parainfluenza Virus 21.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Parainfluenza Virus 4None(no analyte)NotDetected210/210100%210/210100%420/420100%210/210100%210/210100%630/630100%(99.4%-100%)
3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
Parainfluenza Virus 31.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
3x LoDDetected29/30100%30/30100%59/6098.30%30/30100%30/30100%89/9098.90%(94.0%-99.8%)
Respiratory SyncytialVirus1.5x LoDDetected29/30100%30/30100%59/6098.30%30/30100%30/30100%89/9098.90%(94.0%-99.8%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Table. Results from the multi-site reproducibility study (bacteria)
AnalyteConcentrationTestedExpectedResultNucliSENS easyMAGMagNA Pure 96All Sites(95% CI)
Site 1Site 3Sub-TotalSite 2Sub-Total
Bacteria
Mycoplasmapneumoniae3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
None(no analyte)NotDetected150/150100%149/15099.30%299/30099.70%150/150100%150/150100%449/45099.80%(98.8%-100%)
Bordetella pertussis3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)
Chlamydia pneumoniae3x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
1.5x LoDDetected30/30100%30/30100%60/60100%30/30100%30/30100%90/90100%(95.9%-100%)
None(no analyte)NotDetected150/150100%150/150100%300/300100%150/150100%150/150100%450/450100%(99.2%-100%)

a – There was an indeterminate Flu A result for Influenza A H3 low positive sample RP6

b – There was an indeterminate Flu A result for Influenza A H3 negative sample

{24}------------------------------------------------

Premarket Notification 510(k)

Table. Results from the multi-site reproducibility study (bacteria)

{25}------------------------------------------------

Multiple analyte spiked samples Vs single spiked samples

This study is to assess the performance of the BioCode® RPP on the BioCode® MDx-3000 with mixed analyte samples at or near the Limit of Detection (LoD) compared to single spiked samples to justify the use of multiple analyte spiked samples during analytical validation testing. The LoD was confirmed by extracting 20 replicates of each sample type and testing each in singlet for a total of 20 replicates at or near presumptive LoD. LoD for each stock was defined as the lowest concentration with ≥95% detection of 20 replicates (19 out of 20). The single spiked LoD was then challenged by pooling all 4 organisms and testing at 1x LoD (20 replicates). If 1x LoD for each analyte in the mixed pool does not meet the 95% detection goal (19/20), the organism was retested at 3x LoD. Results within 3x LoD was considered equivalent.

Results: All organisms were within 3x LoD for single and multiple spiked samples.

StrainSourceTargetProbeSingle /MultipleSpikeConcentration
EasyMAGMP96
Influenza AH3N2/A/Wisconsin/67/05aZeptometrix0810252CFFluASingle Spike1.3 TCID50/mL1.3 TCID50/mL
FluAH3
FluAMulti Spike4.0 TCID50/mL
FluAH3
Coronavirus NL63Zeptometrix0810228CFNL63Single Spike0.013 TCID50/mL0.040 TCID50/mL
Multi Spike0.040 TCID50/mL
Mycoplasmapneumoniae (M129)Zeptometrix0801579MPNSingle Spike5.0 CCU/mL15.0 CCU/mL
Multi Spike15.0 CCU/mL
Adenovirus C (type 2)ATCC AV-846ADV1Single Spike6.0 TCID50/mL18.0 TCID50/mL
Multi Spike

Table. Results from Multiple Vs Single spiked sample study stratified by extraction system.

Conclusion: Multiple spiked (mixed analyte) samples confirmed 95% LoD at the same or within 3-fold for each organism. Based on this testing multiple spiked samples may be used for other bench studies.

Limit of Detection (LoD)

A study was performed to assess the performance of the BioCode® RPP on the BioCode® MDx-3000 at the Limit of Detection (LoD) for specimens. In this study the BioCode® RPP was tested with quantified bacteria or viral stocks spiked in simulated NPS in UTM matrix. For initial screening, four replicates of each concentration were extracted on the easyMAG and MagNA Pure 96 Systems and tested in singlet with the BioCode® RPP on the BioCode® MDx-3000 system to estimate LoD. The LoD was confirmed by extracting 20 replicates of each sample type and testing each in singlet for a total of 20 replicates at or near the presumptive LoD. LoD for each stock was defined as the lowest concentration with ≥95% detection of 20 replicates (19 out of 20).

Results: See table below.

{26}------------------------------------------------

TargetSpecies/Strain/IsolateSourceEasyMAGConcentrationDetected(n of 20)MP96ConcentrationDetected(n of 20)
Influenza A H1A/NewCaledonia/20/99Zeptometrix0810036CF15.0 TCID50/mL20/205.0 TCID50/mL20/20
Influenza A H1A/NWS/33ATCC VR-21927.0 TCID50/mL20/209.0 TCID50/mL20/20
Influenza A H12009pdmA(H1N1)/California/07/09Zeptometrix0810165CF0.4 TCID50/mL20/200.4 TCID50/mL20/20
Influenza A H3A/Wisconsin/67/05Zeptometrix0810252CF4.0 TCID50/mL20/201.3 TCID50/mL20/20
Influenza A H3A/AliceATCC VR77627.0 TCID50/mL20/209.0 TCID50/mL19/20
Influenza BFlu B/Florida/4/2006(Yamagata)Zeptometrix0810255CF0.01 TCID50/mL20/200.01 TCID50/mL20/20
Influenza BB/Hong Kong/S/1972(Victoria)ATCC VR-82348.6 TCID50/mL20/2048.6 TCID50/mL20/20
Respiratory SyncytialVirusType AZeptometrix0810040ACF0.33 TCID50/mL20/200.33 TCID50/mL20/20
HumanMetapneumovirus16; Type A1 IA10-2003Zeptometrix0810161CF15.0 TCID50/mL19/2015.0 TCID50/mL20/20
Parainfluenza Virus 1C-35/WashingtonDC/1957ATCC VR-949.0 TCID50/mL20/209.0 TCID50/mL20/20
Parainfluenza Virus 2Greer/Ohio/1955ATCC VR-921.8 TCID50/mL20/205.4 TCID50/mL20/20
Parainfluenza Virus 3N/AZeptometrix0810016CF15.0 TCID50/mL20/2015.0 TCID50/mL20/20
Parainfluenza Virus 4Type 4aZeptometrix0810060CF9.0 TCID50/mL20/209.0 TCID50/mL20/20
AdenovirusSpecies B Serotype7AZeptometrix0810021CF1.2 TCID50/mL20/201.2 TCID50/mL20/20
AdenovirusSpecies C Serotype 2ATCC VR-8466.0 TCID50/mL20/2018.0 TCID50/mL20/20
AdenovirusSpecies E Serotype 4Zeptometrix0810070CF0.04 TCID50/mL20/200.04 TCID50/mL19/20
Coronavirus 229EN/AZeptometrix0810229CF0.6 TCID50/mL20/201.8 TCID50/mL20/20
Coronavirus HKU1N/AClinicalSample4922a5.02×104copies/mL19/205.02×104copies/mL20/20
Coronavirus NL63N/AZeptometrix0810228CF0.04 TCID50/mL20/200.04 TCID50/mL20/20
Coronavirus OC43N/AZeptometrix0810024CF0.04 TCID50/mL20/200.01 TCID50/mL19/20
RhinovirusType A1Zeptometrix0810012CF1.2 TCID50/mL20/200.4 TCID50/mL19/20
EnterovirusD68Zeptometrix0810300CF3.0 TCID50/mL19/209.0 TCID50/mL20/20
Bordetella pertussisA639Zeptometrix08145915.0 CFU/mL20/2045.0 CFU/mL19/20
EasyMAGMP96
TargetSpecies/Strain/IsolateSourceConcentrationDetected(n of 20)ConcentrationDetected(n of 20)
Chlamydia pneumoniaeAR39ATCC VR-5359216.7 CFU/mL20/2033.3 CFU/mL20/20
MycoplasmapneumoniaeM129Zeptometrix080157915.0 CCU/mL20/2015.0 CCU/mL20/20

Table. Limit of Detection by extraction system

{27}------------------------------------------------

Coronavirus HKU1 clinical sample quantified (copies/mL) with Applied BioCode validated SYBR assay using an IVT RNA a. standard.

Conclusion: LoDs were comparable (within 3-fold) for each extraction system.

Analytical Reactivity/Inclusivity

A study was performed to verify Analytical Reactivity/Inclusivity of the BioCode® Respiratory Pathogen Panel (RPP). Different strains, serotypes were selected that represent various temporal, geographic, and genetic diversity for each analyte. This study tested a panel of titered stocks of relevant organisms diluted in simulated NPS in UTM matrix starting at 3x LoD. Samples not detected at 3x LoD, were tested at higher concentrations. For Influenza B strains were grouped into Yamagata, Victoria or unknown lineages. Yamagata strains were tested according to the protocol at 3x and higher of the Yamagata strain LoD (LoD - 0.01 TCIDsomL Flu B/Florida/4/2006). However, for Victoria strains, the validation LoD result was higher than expected (LoD-48.6 TCID50/mL for Flu B /Hong Kong/S/1972) which could be related to difference in titration from vendor sources. Therefore, for Victoria strains, testing for inclusivity was performed starting at 0.03x and 0.3x of the Victoria strain of unknown lineages of influenza B were tested at 3x LoD of Yamagata strain. Each sample was extracted in triplicate on the easyMAG and tested with the RPP on the BioCode® MDx-3000 system according to the instructions for use.

Assay reactivity for less common strains or serotypes that could not be tested due to unavailability was predicted using in silico analysis.

Results: The organisms listed below were detected at the concentrations indicated. See Table below.

Organism/TypeaStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
Influenza A H1N1Solomon Island/03/06Zeptometrix0810036CFN45 TCID50/mL3x
Singapore/63/04Zeptometrix0810246CF45 TCID50/mL3x
PR/8/34Zeptometrix0810245CF45 TCID50/mL3x
A/Brisbane/59/2007Zeptometrix0810244CF45 TCID50/mL3x
A/Taiwan/42/06Zeptometrix0810247CF45 TCID50/mL3x
A/New jersey/8/76bATCCVR-8977.5x103 CEID50/mL500xb
Organism/TypeaStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
A/Denver/1/1957VIRAPURNA45 TCID50/mL3x
A/FM/1/47ATCCVR-9745 CEID50/mL3x
A/Weiss/43cATCCVR-961.5x104 CEID50/mL1000xc
A/Beijing/262/95dBEINR-12277450 CEID50/mL30xd
A/Mal/302/54ATCCVR-9845 CEID50/mL3x
Influenza A H1N2Recombinant; Kilbourne F64,A/NWS/1934 (HA) xA/Rockefeller Institute/5/1957(NA)eBEINR-3682135 CEID50/mL5xe
NY/01/09Zeptometrix0810248CF1.2 TCID50/mL3x
NY/02/09Zeptometrix0810109CFN1.2 TCID50/mL3x
NY/03/09Zeptometrix0810249CF1.2 TCID50/mL3x
A/Houston/3H/2009(H1N1)pdm09fBEINR-203401.2 TCID50/mL3x
Influenza A H1N1pdm(Canada/6294/09)Zeptometrix0810109CFJ1.2 TCID50/mL3x
Influenza A H1N1pdm(Mexico/4108/09)Zeptometrix0810166CF1.2 TCID50/mL3x
Influenza A H1N1pdm09California/04/09, cell isolategBEINR-1365840 TCID50/mL100xg
A/Christ Church/16/2010hCDCN/A400 EID50/mL1000xh
A/Brisbane/02/2018CDCN/A40 EID50/mL100xi
A/Wisconsin/15/2009jBEINR-4200712 CEID50/mL3x
A/Texas/50/12Zeptometrix0810238CF12 TCID50/mL3x
A/Brisbane/10/2007Zeptometrix0810138CF12 TCID50/mL3x
A/Port Chalmers/1/73ATCCVR-810200 CEID50/mL50xk
Influenza A H3N2A/Victoria/3/1975VIRAPURNA12 TCID50/mL3x
Organism/TypeaStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
A/Victoria/361/2011lBEINR-4402212 CEID50/mL3x
A/Victoria/3/75ATCCVR-82212 CEID50/mL3x
A/Uruguay/716/07mBEINR-4200340 TCID50/mL10x
A/HK/H090-756-V1(0)/2009nBEINR-4434412 TCID50/mL3x
A/Hong Kong/8/68Zeptometrix0810250CF12 TCID50/mL3x
A/Switzerland/9715293/ 13VIRAPURNA12 TCID50/mL3x
A/Aichi/2/68ATCCVR-54712 TCID50/mL3x
MRC-2ATCCVR-77712 TCID50/mL3x
A/Perth/16/2009CDCN/A40 EID50/mL10x
A/Kansas/14/2017oCDCN/A8000 EID50/mL2000xo

Table. Influenza A isolates tested during the inclusivity study

{28}------------------------------------------------

{29}------------------------------------------------

a - In silico analysis predicts detection of Influenza A H2N3, H5N2, H5N3, H7N7, H7N9, H3N2, H3N5, H3N7, H3N8 as Influenza A. However, predicted reactivities of the subtyping assays for these influenza A strains of animal origin are variable.

b – Influenza A H1N1 – [A/New jersey/8/76]. Detected as Flu A (no subtype) at 3x LoD. Detected as dual positive A/H1 and A/H1pdm09 at 500x LoD.

c – Influenza A H1N1 [A/Weiss/43]. Detected as Flu A (no subtype) at 100x LoD. In silico analysis showed several mismatches in the forward primer for the Flu A/H1 subtyping assay which may account for the observed lower sensitivity of the Flu A/H1 subtyping assay for this strain.

d – Influenza A H1N1 [A/Beijing/262/95]. Virus obtained through BEI Resources. NIAID. NH: Influenza A. A/Beijing/262/95 (H1N1), NR-12277. Detected as Flu A (no subtype) at 10x LoD. In silico analysis showed a G-A mismatch in the 3' terminal position in the forward primer for the Flu A/H1 subtyping assay which may account for the observed lower sensitivity of the Flu A/H1 subtyping assay for this strain.

e - Influenza A H1N2 [Recombinant Virus obtained through BEI Resources, NIAID, NH: Influenza A, Kilbourne F64, A/NWS/1934 (HA) x A/Rockefeller Institute/5/1957 (NA) (H1N2), NR-3682.Detected as Flu A (no subtype) at 3x LoD.

f - Virus obtained through BEI Resources, NIAID, NIH: Influenza A, A/Houston/3H/2009 (H1N1)pdm09, NR-20340. g – Virus obtained through BEI Resources, NIAID, NH: Influenza A, A/California/04/09, cell isolate (H1N1) pdm09, NR-13658. In silico analysis of a partial sequence of this strain for the Flu A/H1pdm09 subtyping assay does nalytical reactivity. Titering inconsistency from the vendor rather than reduced reactivity due to assay design is suggested. h - Virus obtained through the CDC Influenza Divis of partial sequences of this strain for the Flu A/H1pdm09 HA subtyping assay and the Flu A matrix gene assay does not predict reactivity. Titering inconsistency from the source laboratory (EIDsp/mL vs. TCID-n/mL) rather than reduced reactivity due to assay design is suggested.

i - Virus obtained through the CDC Influenza Division. In silico analysis did not reveal any critical mismatch in the Flu A/H1pdm09 HA subtyping assay primers and probe binding regions or any mismatch in the FluA matrix gene assay primers and probe binding regions that would predict reactivity. Titering inconsistency from the source laboratory (EID50/mL vs. TCID50/mL) rather than reduced reactivity due to assay design is suggested.

j – Virus obtained through BEI Resources, NIAID, NIH: Influenza A, A/Wisconsin/15/2009 (H3N2), NR-42007.

k - In silico analysis revealed a few non-critical mismatches in the FluA/H3 subtyping HA assay probe binding region that could contribute to the observed lower reactivity. However, titering inconsistency from the than reduced reactivity due to assay design is suspected.

{30}------------------------------------------------

l - Virus obtained through BEI Resources, NIAID, NIH: Influenza A, A/Victoria/361/2011 (H3N2), NR-44022 m - Virus obtained through BEI Resources, NIAID, NIH: Influenza A, A/Uruguay/716/07 (H3N2), NR-42003 n - Virus obtained through BEI Resources, NIAID, NIH: Influenza A, A/HK/H090-756-V1(0)/2009 (H3N2), NR-44344 o – Virus obtained through the CDC Influenza Division. This strain was detected by BioCode RPP as Flu A (no subtype) at 50xLoD. In silico analysis did not reveal any mismatch in the Flu A matrix assay primers and probe binding regions but showed 3 mismatches in the Flu A/H3 subtyping HA assay primers binding regions that could predict reactivity, 1 mismatch at the 9th position from the 3' end of the reverse primer (mismatch at the 18th position from the 3' end of the reverse primer (mismatch at the 20th position from the 3' end of the reverse primer (mismatch #3). Wet testing data suggested that mismatch #1 is likely the root cause for the observed significant reduction in analytical reactivity for this strain. For patient samples contain a Flu A/H3 strain that harbors mismations, the BioCode RPP will likely report a Flu A (no subtype detected) result. Although estimated prevalence of a sequence variant based solely on in silico analysis may not accurately reflect the actual prevalence variant in circulation during an influenza season, based on an in silico analysis, of all the Flu A/H3 strains isolated in 2019 with published HA sequences, 73.8% of the strains harbor mismatch #1.

Organism/LineageLocation/Strain/ YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetecteda
B/Malaysia/2506/2004dBEINR-97231.458TCID50/mL0.03x
B/Malaysia/2506/2004Zeptometrix0810258CF1.458TCID50/mL0.03x
B/Ohio/01/2005eBEINR-4180114.58CEID50/mL0.3x
Influenza B (Victoriab)B/Brisbane/33/2008fBEINR-4200614.58CEID50/mL0.3x
B/Nevada/03/2011gBEINR-440231.458CEID50/mL0.03x
B/Michigan/09/2011CDCN/A14.58 EID50/mL0.3x
B/Colorado/06/2017CDCN/A14.58 EID50/mL0.3x
B/Texas/06/2011hBEINR-4402450 TCID50/mL5000xh
B/Sydney/507/2006iBEINR-365268.0 TCID50/mL800xi
B/Wisconsin/1/10Zeptometrix0810241CF0.03 TCID50/mL3x
Influenza B (Yamagata)B/Massachusetts/2/12Zeptometrix0810239CF0.03 TCID50/mL3x
B/Christchurch/33/2004jBEINR-365360.03 TCID50/mL3x
B/New Hampshire/01/2016kCDCN/A10 EID50/mL1000xk
B/Phuket/3073/2013lCDCN/A5 EID50/mL500xl
B/lee/1940Zeptometrix0810257CF0.03 TCID50/mL3x
Table. Influenza B isolates tested during the inclusivity study

{31}------------------------------------------------

Organism/LineageLocation/Strain/ YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetecteda
Influenza B (unknownlineagec)B/Taiwan/2/1962ATCCVR-2955.0 CEID50/mL500xm
Influenza B (unknownlineagec)B/Allen/45ATCCVR-1020.05 CEID50/mL5x
Influenza B (unknownlineagec)B/BrigitATCCVR-7860.03 TCID50/mL3x

a - If either FluB assay has MFI above the cutoff, the software will report as Detected for Influenza B

b - For Victoria lineage strains testing started at 0.03x LoD rather than 3x LoD.

c - Strains of unknown lineages were assayed starting at 3x LoD of Yamagata strain.

d - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Malaysia/2506/2004, NR-9723.

e - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Ohio/01/2005, NR-41801.

f - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Brisbane/33/2008, NR-42006.

g - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Nevada/03/2011, NR-44023.

h - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Texas/06/2011, NR-44024. In silico analysis does not predict reduced analytical reactivity. Titering inconsistency from the vendor reactivity due to assay design is suggested.

i - Virus obtained through BEI Resources, NIAID, NH: Influenza B, B/Sydney/507/2006, NR-36526. In silico analysis does not predict reduced analytical reactivity. Titering inconsistency from the vendor reactivity due to assay design is suggested.

j - Virus obtained through BEI Resources, NIAID, NIH: Influenza B, B/Christchurch/33/2004, NR-36536.

k - Virus obtained through the CDC Influenza Division. In silico analysis did not reveal any critical mismatch in the Flu B NS1 assay primers and probe binding regions that would predict reactivity. And In silico analysis did not reveal any mismatch in the Flu B matrix assay primers and probe binding regions. Titering inconsistency from the source laboratory (EID50/mL vs. TCID50/mL) rather than reduced reactivity due to assay design is suggested.

l - Virus obtained through the CDC Influenza Divis did not reveal any critical mismatch in the Flu B NS1 assay primers and probe binding regions that would predict reduced analytical reactivity. And In silico analysis did not reveal any mismatch in the Flu B matrix assay primers and probe binding regions. Titering inconsistency from the source laboratory (EID50/mL vs. TCID50/mL) rather than reduced reactivity due to assay design is suggested.

m - In silico analysis could not be performed due to unavailability of sequence information for this strain. Titering inconsistency from the vendor rather than reduced reactivity due to assay design is suspected.

{32}------------------------------------------------

Organism/TypeStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
Respiratory SyncytialVirus Type ATN/1998/3-2aBEINR-285290.99 TCID50/mL3x
TN/2000/3-4bBEINR-285303.3 TCID50/mL10x
TN/98/12-21cBEINR-285280.99 TCID50/mL3x
Long/Maryland/1956ATCCVR-263.3 TCID50/mL10x
9320/Massachusetts/1977ATCCVR-9550.99 TCID50/mL3x
Respiratory SyncytialVirus Type BB1dBEINR-40523.3 TCID50/mL10x
WV/14617/85ATCCVR-14000.99 TCID50/mL3x
18537/WashingtonDC/1962ATCCVR-15800.99 TCID50/mL3x
CH93(18)18Zeptometrix0810040CF0.99 TCID50/mL3x

Table. Respiratory Syncytial Virus isolates tested during the inclusivity study

a - Virus obtained through BEI Resources, NIAID, NIH: RSV, TN/1998/3-2, NR-28529

b - Virus obtained through BEI Resources, NIAID, NIH: RSV, TN/2000/3-4, NR-28530

c - Virus obtained through BEI Resources, NIAID, NIH: RSV, TN/98/12-21, NR-28528

d - Virus obtained through BEI Resources, NIAID, NIH: RSV, B1, NR-4052

Table. Human Metapneumovirus isolates tested during the inclusivity study

GenotypeSerotypeLocation/YearVendorCatalog #ConcentrationDetectedMultipleof LoDDetecteda
HumanMetapneumovirusType A19lowa3/2002Zeptometrix0810160CF45 TCID50/mL3x
HumanMetapneumovirusType A227lowa27/2004Zeptometrix0810164CF45 TCID50/mL3x
HumanMetapneumovirusType B13Peru2/2002Zeptometrix0810156CF45 TCID50/mL3x
HumanMetapneumovirusType B15Peru3/2003Zeptometrix0810158CF45 TCID50/mL3x
HumanMetapneumovirusType B24Peru1/2002Zeptometrix0810157CF45 TCID50/mL3x
HumanMetapneumovirusType B28Peru6/2003Zeptometrix0810159CF45 TCID50/mL3x
HumanMetapneumovirusType B218lowa18/2003Zeptometrix0810162CF45 TCID50/mL3x
HumanMetapneumovirusType B2UnknownTN/91-316bBEINR-2223245 TCID50/mL3x

a - If either assay has MFI above the cutoff, the software will report as Detected for Human Metapneumovirus

b - Virus obtained through BEI Resources, NIAID, NIH: Human Metapneumovirus, TN/91-316, NR-22232

{33}------------------------------------------------

Organism/SubtypeStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
Parainfluenza Virus 1FRA/27344044/2007aBEINR-4868127 TCID50/mL3x
Parainfluenza Virus 1FRA/29221106/2009bBEINR-4868027 TCID50/mL3x
Parainfluenza Virus 1UnknownZeptometrix0810014CF27 TCID50/mL3x
Parainfluenza Virus 2GreercBEINR-32295.4 TCID50/mL3x
UnknownZeptometrix0810015CF5.4 TCID50/mL3x
Parainfluenza Virus 3NIH 47885,Wash/47885/57dBEINR-323345 TCID50/mL3x
C243/WashingtonDC/1957ATCCVR-9345 TCID50/mL3x
Parainfluenza Virus 4aM-25/1958eBEINR-323727 TCID50/mL3x
Parainfluenza Virus 4bCH-19503/WashingtonDC/1962ATCCVR-137727 TCID50/mL3x
UnknownZeptometrix0810060BCF27 TCID50/mL3x

Table. Parainfluenza Virus (1-4) isolates tested during inclusivity.

a - Virus obtained through BEI Resources, NIAID, NIH: Parainfluenza Virus 1, HIPIV1/FRA/27344044/2007, NR-48681

b - Virus obtained through BEI Resources, NIAID, NIH: Parainfluenza Virus 1, HPIV1/FRA/29221106/2009, NR-48680

c - Virus obtained through BEI Resources, NIAID, NIH: Parainfluenza Virus 2, Greer, NR-3229

d - Virus obtained through BEI Resources, NIAID, NIH: Parainfluenza Virus 3, NIH 47885, NR-3233

e - Virus obtained through BEI Resources, NIAID, NIH: Parainfluenza Virus 4a, M-25, NR-3237

Table. Adenovirus isolates tested during the inclusivity study

SpeciesbSerotypeStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetectedc
Adenovirus Aa31UnknownZeptometrix0810073CF18 TCID50/mL3x
12Huie/MassachusettsATCCVR-86318 TCID50/mL3x
18Washington DC/1954ATCCVR-1918 TCID50/mL3x
Adenovirus B3GB/Maryland/1953ATCCVR-33.6 TCID50/mL3x
14UnknownZeptometrix0810108CF3.6 TCID50/mL3x
16CH.79/SaudiArabia/1955ATCCVR-173.6 TCID50/mL3x
35HoldenATCCVR-7183.6 TCID50/mL3x
Adenovirus C1UnknownZeptometrix0810050CF18 TCID50/mL3x
5UnknownZeptometrix0810020CF18 TCID50/mL3x

{34}------------------------------------------------

SpeciesbSerotypeStrain/Location/YearVendorCatalog #ConcentrationDetectedMultiple ofLoDDetectedc
Adenovirus Da6Tonsil 99/WashingtonDCATCCVR-618 TCID50/mL3x
8UnknownZeptometrix0810069CF18 TCID50/mL3x
17CH. 22/SaudiArabia/1955ATCCVR-183618 TCID50/mL3x
20UnknownZeptometrix0810115CF18 TCID50/mL3x
26UnknownZeptometrix0810117CF18 TCID50/mL3x
37UnknownZeptometrix0810119CF18 TCID50/mL3x
Adenovirus Fa40UnknownZeptometrix0810084CF18 TCID50/mL3x
41Tak/73-3544/Netherlands/1973ATCCVR-93018 TCID50/mL3x

a – Adenovirus C LoD used as LoD was not determined for Species A, D or F.

b – In silico analysis predicts detection of Adenovirus species E serotypes (Adenovirus 4 strain was tested as a part of LoD studies as well.

c – If either AdV assay has MFI above the cutoff the software will report as Detected for Adenovirus.

OrganismLocation/YearVendorCatalog #ConcentrationDetectedMultipleof LoDDetected
Coronavirus 229EUnknownATCCVR-7401.8 TCID50/mL3x
Coronavirus NL63Amsterdam/2003BEIbNR-4700.12 TCID50/mL3x
Coronavirus OC43UnknownATCCVR-15580.12 TCID50/mL3x
HKU1aUnknownClinical Sample 5016Unknown1.51x105 copies/mL3x
UnknownClinical Sample 5036Unknown1.51x105 copies/mL3x
UnknownClinical Sample 5037Unknown5.02x105 copies/mL10x

Table. Coronavirus isolates tested during the inclusivity study

a - Coronavirus HKU1 clinical samples titered with Applied BioCode validated SYBR assay using an IVT standard.

b - Virus obtained through BEI Resources, NIAID, NIH: Human Coronavirus NL63, NR-470

{35}------------------------------------------------

Organism/SpeciesSerotype/Strain/IsolateVendorCatalog #ConcentrationDetectedMultiple ofLoDDetected
Rhinovirus ASerotype 7/[68-CV11]ATCCVR-16013.6 TCID50/mL3x
Serotype 16 [1A]Zeptometrix0810285CF3.6 TCID50/mL3x
Serotype 16 [11757/DC/1960]ATCCVR-2833.6 TCID50/mL3x
Serotype 34 [137-3]ATCCVR-13653.6 TCID50/mL3x
Serotype 57 [Ch47}ATCCVR-16003.6 TCID50/mL3x
Serotype 77 [130-63]ATCCVR-11873.6 TCID50/mL3x
Serotype 80Zeptometrix0810288CF3.6 TCID50/mL3x
Serotype 85 [50-525-CV54]ATCCVR-11953.6 TCID50/mL3x
Serotype 95 [SF-998}ATCCVR-13013.6 TCID50/mL3x
Rhinovirus BSerotype 3 [FEB]ATCCVR-4833.6 TCID50/mL3x
Serotype 14Zeptometrix0810284CF3.6 TCID50/mL3x
Serotype 42Zeptometrix0810286CF3.6 TCID50/mL3x
Serotype 70Zeptometrix0810287CF3.6 TCID50/mL3x
Enterovirus 71EV 71Zeptometrix0810236CF9.0 TCID50/mL3x

Table. Human Rhinovirus and Enterovirus isolates tested during the inclusivity study

Table. Bordetella pertussis isolates tested during the inclusivity study

OrganismStrain/IsolateVendorCatalog #ConcentrationDetectedMultipleof LoDDetected
Bordetella pertussisFATCC846745 CFU/mL3x
5[17921]ATCC934045 CFU/mL3x
10-536ATCC1038045 CFU/mL3x
CNCTC Hp 12/63[623]ATCC5144545 CFU/mL3x
Tohama 1ATCCBAA-58945 CFU/mL3x
MN2531ATCCBAA-133545 CFU/mL3x
Table. Mycoplasma pneumoniae isolates tested during the inclusivity study
--------------------------------------------------------------------------------------
OrganismStrain/IsolateVendorCatalog #ConcentrationDetectedMultipleof LoDDetected
Mycoplasma pneumoniaeM129-B7ATCC2934245 CFU/mL3x
PI 1428ATCC2908545 CCU/mL3x
MacATCC1549245 CFU/mL3x
UTMB-10PATCC4989445 CCU/mL3x

{36}------------------------------------------------

OrganismStrain/IsolateVendorCatalog #ConcentrationDetectedMultipleof LoDDetected
Chlamydia pneumoniaeCM-1ATCCVR-136050.1 CFU/mL3x
CWL-029ATCCVR-131050.1 CFU/mL3x
Table. Chlamydia pneumoniae isolates tested during the inclusivity study

Conclusion: This inclusivity testing suggests that the BioCode RPP is analytically reactive for the targeted organisms. BioCode RPP assay displayed good inclusivity for related organisms with similar genetic lineage/sequence.

Zoonotic Influenza A Analytical Reactivity/Inclusivity Testing

Due to public health concerns related to zoonotic transmission of influenza A viruses to humans (primarily swine and avian lineages), serial dilutions of the following isolates of swine and avian influenza A viruses (viral nucleic acids) were also tested assessing analytical reactivity:

  • A/Japan/305/57 (H2N2)
  • · A/duck/Pennsylvania/10218/1984 (H5N2)
  • · A/turkey/Wisconsin/1/1966 (H9N2)
  • A/Anhui/1/2013 (H7N9)
  • · A/Hubei/1/2010 (H5N1)
  • A/Minnesota/11/2010 (H3N2v)

Consistent with the in silico predictions of analytical reactivity to zoonotic influenza A viruses, each nonseasonal influenza A strain tested (H2N2, H5N2, H7N9, and H5N1), except for the H3N2v strain, was detected as Influenza A (no subtype) at various concentrations tested (i.e., detected by the BioCode RPP Influenza A matrix assay only). For the H3N2v strain, it was detected as Influenza A/H3 at higher concentrations but was detected as Influenza A (no subtype) at the lowest concentration tested.

Analytical testing could not be performed to assess reactivity to H5N8, H1N2v and H1N1 swine viruses due to the lack of availability of such zoonotic influenza A strains for wet testing. Based on an in silico analysis, H5N8 (KP739416), H1N2v (MK239077), and H1N1 swine (KP822962.1) were predicted to be detected as Influenza A (no subtype), Influenza A/H1 (with reduced sensitivity for the A/H1 assay), and Influenza A (no subtype), respectively.

Analytical Specificity/Cross Reactivity

A study was performed to verify that the BioCode® Respiratory Pathogen Panel (RPP) does not detect DNA or RNA from off-panel organisms commonly found in respiratory specimens or from organisms that can cause similar clinical symptoms. In addition, on-panel organisms were tested at high concentrations to ensure there is no cross-reactivity with other on-panel targets. This study tested a panel of titered stocks of relevant organisms. Organisms that were not available for wet testing were analyzed in silico comparing the whole organism sequence against all primers to assess potential for cross reactivity. Analysis was conducted using BLASTn and Primer-BLAST programs. All testing was performed at Applied BioCode®, Inc. Microorganisms were tested at 10° CFU/mL for bacteria or fungi and 10° TCIDso/mL for

{37}------------------------------------------------

virus or higher when possible. Stocks were combined with simulated NPS in UTM matrix at the of extraction. Each organism was extracted in triplicate on the EasyMag and assayed in singlet with the RPP on the BioCode® MDx-3000 system according to the instructions for use. For each concentration tested, the number of replicates that gave valid results per the Interpretation Algorithm was determined. If any replicates were detected, testing was repeated from 5 additional extractions assayed in singlet. If detected after repeat with 5 additional replicates, serial dilutions were performed to determine the lower limit.

OrganismVendorCatalog #Titer testedCross-reactivity(Y/N)
Acinetobacter baumanniiZeptometrix8015979.67 x 106 CFU/mLN
Aspergillus flavusZeptometrix8015981.72 x 106 CFU/mLN
Bordetella bronchisepticaZeptometrix8016496.68 x 107 CFU/mLN
Bordetella holmesiiZeptometrix8014643.83 x 109 CFU/mLYa
Bordetella parapertussisZeptometrix80114611.00 x 106 CFU/mLN
Burkholderia cepaciaZeptometrix8015844.13 x 107 CFU/mLN
Candida albicansZeptometrix8015041.96 x 106 CFU/mLN
Candida glabrataZeptometrix8015351.73 x 107 CFU/mLN
Corynebacterium diphtheriaeZeptometrix8018824.57 x 106 CFU/mLN
Haemophilus influenzaeZeptometrix8016792.40 x 106 CFU/mLN
Klebsiella pneumoniaeZeptometrix8015065.10 x 107 CFU/mLN
Lactobacillus plantarumZeptometrix8015071.80 x 107 CFU/mLN
Legionella longbeachaeZeptometrix81015771.93 x 107 CFU/mLN
Legionella micdadeiZeptometrix8015762.70 x 107 CFU/mLN
Legionella pneumophilaZeptometrix8016453.17 x 107 CFU/mLN
Moraxella catarrhalisZeptometrix8015091.13 x 106 CFU/mLN
Mycobacterium tuberculosisZeptometrix8016607.23 x 106 CFU/mLN
Mycoplasma genitaliumATCC335301.00 x 103 CFU/mLN
Mycoplasma hominisATCC231142.7 x 104 CFU/mLN
Neisseria elongataZeptometrix8015101.74 x 107 CFU/mLN
Neisseria gonorrhoeaeZeptometrix8014821.26 x 107 CFU/mLN
Neisseria meningitidisZeptometrix8015112.55 x 106 CFU/mLN
Neisseria siccaZeptometrix8017541.02 x 106 CFU/mLN
Proteus vulgarisZeptometrix8018984.13 x 107 CFU/mLN
Pseudomonas aeruginosaATCC393242.11 x 106 CFU/mLN
Serratia marcescensZeptometrix8017232.06 x 107 CFU/mLN
Staphylococcus haemolyticusZeptometrix8015918.50 x 106 CFU/mLN
Streptococcus agalactiaeZeptometrix8015453.73 x 106 CFU/mLN
Streptococcus dysgalactiaeZeptometrix8015161.23 x 106 CFU/mLN
Streptococcus intermediusZeptometrix8018955.07 x 106 CFU/mLN
Streptococcus mitisZeptometrix8016955.73 x 106 CFU/mLN
Streptococcus pneumoniaeZeptometrix8014394.17 x 106 CFU/mLN
Ureaplasma urealyticumATCC276181.00 x 107 CFU/mLN
OrganismVendorCatalog #Titer testedCross-reactivity(Y/N)
SARS-CoV, formaldehyde- and UV-inactivated,purified (vaccine)BEINR-38831:100 DilutionNa
MERS-CoV genomic RNABEINR-458431.01 x 107 Copies/mLN
MERS-CoV EMC/2012, Heat-InactivatedBEINR-501712 x 105 TCID50/mlN
Coxsackievirus A10Zeptometrix0810106CF1.05 x 103 TCID50/mLYb
Coxsackievirus A21Zeptometrix0810235CF≤ 1.03 x 102 TCID50/mLYb
Coxsackievirus A24ATCCVR-5831.14 x 101 TCID50/mLYb
Coxsackievirus B2ATCCVR-295.62 x 103 TCID50/mLYb
Coxsackievirus B3Zeptometrix0810074CF1.76 x 103 TCID50/mLYb
Coxsackievirus B4Zeptometrix0810075CF1.36 x 104 TCID50/mLYb
Coxsackievirus B5Zeptometrix0810019CF≤ 5.89 x 102 TCID50/mLYb
Coxsackievirus A9Zeptometrix0810017CF1.38 x 103 TCID50/mLYb
CytomegalovirusZeptometrix0810003CF4.17 x 104 TCID50/mLN
Echovirus 11Zeptometrix0810023CF1.68 x 103 TCID50/mLYc
Echovirus 30Zeptometrix0810078CF≤ 1.95 x 102 TCID50/mLYc
Echovirus 6Zeptometrix0810076CF1.09 x 104 TCID50/mLYc
Echovirus 9Zeptometrix0810077CF1.07 x 101 TCID50/mLYc
Epstein-Barr VirusZeptometrix0810008CF3.43 x 106 TCID50/mLN
Herpes Simplex Virus Type 1Zeptometrix0810187CF9.12 x 106 TCID50/mLN
Measles VirusZeptometrix0810025CF1.31 x 105 TCID50/mLN
Mumps VirusZeptometrix0810176CF1.89 x 105 TCID50/mLN

Table. Off-panel bacteria and fungi analyzed for analytical specificity (Cross reactivity)

{38}------------------------------------------------

a - Bordetella holmesii was detected by the Bordetella pertussis (BP) assay with 2 of 3 replicates down to 3.83 x 100 CFU/mL.

Table. Off-panel Viruses analyzed for analytical specificity (Cross reactivity)

a – Inhibitory (no RNA-IC detected) at 1:10 dilution

b – The Coxsackieviruses assayed here were detected by the Human Rhinovirus/Enterovirus (HRV) assay in at least 1 or the 3 replicates down to the concentrations indicated.

c – The Echoviruses assayed here were detected by the Human Rhinovirus/Enterovirus (HRV) assay in at least 1 or the 3 replicates down to the concentrations indicated.

Table. On-panel organisms analyzed for analytical specificity (Cross reactivity).
OrganismVendorCatalog #Titer testedCross-reactivity(Y/N)
Influenza A H1N1/New Caledonia/20/99Zeptometrix0810036CF1.15 x 105 TCID50/mLN
Influenza A H1N1 /NWS/33ATCCVR-2197.40 x 105 TCID50/mLN
Influenza A H1N1 pdm09/California/07/09Zeptometrix0810165CF1.31 x 105 TCID50/mLN
Influenza A H3N2 /Wisconsin/67/05aZeptometrix0810252CF1.08 x 105 TCID50/mLN
Influenza A H3N2/AliceATCCVR-7761.43 x 106 TCID50/mLN
Influenza B/Florida/4/2006 (Yamagata)Zeptometrix0810255CF1.08 x 105 TCID50/mLN
Influenza B/Hong Kong/S/1972 (Victoria)ATCCVR-8238.57 x 105 TCID50/mLN
Respiratory Syncytial Virus (Type A)Zeptometrix0810040ACF4.57 x 105 TCID50/mLN
Human Metapneumovirus 16 (Type A1)Zeptometrix0810161CF8.51 x 105 TCID50/mLN
Parainfluenza Virus 1ATCCVR-941.60 x 105 TCID50/mLN

{39}------------------------------------------------

OrganismVendorCatalog #Titer testedCross-reactivity(Y/N)
Parainfluenza Virus 2ATCCVR-921.35 x 105 TCID50/mLN
Parainfluenza Virus 3Zeptometrix0810016CF3.39 x 105 TCID50/mLN
Parainfluenza Virus 4aZeptometrix0810060CF1.13 x 105 TCID50/mLN
Adenovirus Species B Serotype 7AZeptometrix0810021CF5.83 x 105 TCID50/mLN
Adenovirus Species C Serotype 2ATCCAV-8462.81 x 105 TCID50/mLN
Adenovirus Species E Serotype 4Zeptometrix0810070CF1.08 x 105 TCID50/mLN
Coronavirus 229EZeptometrix0810229CF1.09 x 105 TCID50/mLN
Coronavirus HKU1N/AClinicalsamplea1.92 x 105 TCID50/mLN
Coronavirus NL63Zeptometrix0810228CF1.08 x 105 TCID50/mLN
Coronavirus OC43Zeptometrix0810024CF1.08 x 105 TCID50/mLN
Human Rhinovirus Type A1Zeptometrix0810012CF1.05 x 105 TCID50/mLN
Enterovirus D68Zeptometrix0810300CF1.08 x 105 TCID50/mLN
Bordetella pertussisZeptometrix8014593.86 x 107 CFU/mLN
Mycoplasma pneumoniaeZeptometrix8015791.06 x 106 CCU/mLN
Chlamydia pneumoniae (AR-39)ATCC535921.24 x 106 CFU/mLN
Chlamydia pneumoniae (CWL-029)ATCCVR-13101.00 x 106 CFU/mLN

a - Coronavirus HKU1 clinical samples titered with Applied BioCode validated SYBR assay using an IVT RNA standard.

Conclusion: Cross reactivity was not observed with the on-panel or off-panel microorganisms at the concentrations tested in this study except the following:

  • Empirical testing and in silico sequence analysis indicate that the Rhinovirus/Enterovirus assay (HRV) may also react with other Enterovirus species (i.e., Coxsackievirus and Echoviruses; see table for testing results).
  • In silico sequence analysis indicate that the Bordetella pertussis assay (BP) may react with Bordetella holmesii and Bordetella bronchiseptica.

Interfering Substances/Microbes

A study was performed to demonstrate the accuracy of the BioCode® Respiratory Pathogen Panel on the BioCode® MDx-3000 in the presence of potentially interfering substances or microorganisms. Each member of the interfering substance panel was added to contrived samples in simulated NPS (sNPS) in UTM matrix containing representative members of the BioCode RPP at 3x LoD and a negative sample comprised of only sNPS in UTM matrix. Each sample was tested with and without potentially interfering substances. Each sample was extracted in triplicate on both the easyMAG and MP96 extraction systems and tested singly with the RPP on the BioCode® MDx-3000 system. Concentrations of interferents were determined by reviewing results of previous RP clinical trials. Substances that produce interference at the original test concentration were tested at lower concentrations.

Sample NameOrganismSource
RPP AAdenovirus B Serotype 7AZeptometrix 0810060CF
Mycoplasma pneumoniaeZeptometrix 801579

Table . Contrived samples (3x LoD in sNPS)

{40}------------------------------------------------

Influenza A H3N2 A/Wisconsin/67/2005Zeptometrix 0810252CF
Respiratory Syncytial Virus (Type A)Zeptometrix 0810040ACF
RPP BInfluenza A H1N1/California/07/09Zeptometrix 0810165CF
Human Metapneumovirus (16; type A1)Zeptometrix 0810161CF
Parainfluenza Virus 3Zeptometrix 0810016CF
RPP CCoronavirus NL63Zeptometrix 0810228CF
Influenza B/Florida/4/2006Zeptometrix 0810255CF
HRVHuman RhinovirusZeptometrix 0810012CF

Results: All targets of Samples RPP A, RPP C and HRV were detected (3/3) at the concentrations below, suggesting no interference from these potential interferents at the concentrations tested.

Microbial InterferentBrand/SourceConcentrationInterferenceYes (Y) or No (N)
Streptococcus pneumoniaeZeptometrix1 X 106 CFU/mLN
Haemophilus influenzaeZeptometrix1 X 106 CFU/mLN
Neisseria meningitidisZeptometrix1 X 106 CFU/mLN
Staphylococcus aureusATCC1 X 106 CFU/mLN
CytomegalovirusZeptometrix1 X 105 TCID50/mLN

Table: Non-microbial Interfering substances tested for BioCode® RPP assay

Substance InterferentBrand/SourceConcentrationInterferenceYes (Y) or No (N)
Genomic DNAPromega10 ng/μlN
Mucin (MagNA Pure 96)Sigma0.6% W/VN
Mucin (EasyMAG)ªSigma0.5% W/VN
Human BloodPoplar Health1% V/VN
ZanamivirAPExBIO550 ng/mLN
OseltamivirAPExBIO142 ng/mLN
Nasal sprayEquate1% V/VN
Nasal decongestant sprayBayer1% V/VN
Nasal Allergy spray (Fluticasone)Equate1.5% V/VN
Petroleum JellyEquate1% W/VN
Analgesic OintmentVicks1% W/VN
MupirocinAlfa Aesar™2% W/VN
TobramycinMP Biomedicals,LLC0.6 mg/mLN
Bleach (10%)VWR5% V/VN
Disinfecting wipesClorox50% V/VN
Ethanol (70%)LabChem7% V/VN
Remel M4 MediaRemel90% V/VN
Remel M4-RT MediaRemel90% V/VN
Remel M5 MediaRemel90% V/VN
Remel M6 MediaRemel90% V/VN
Copan FloQ (Flocked nylon/plastic shaft)Copan1 swabN

{41}------------------------------------------------

Substance InterferentBrand/SourceConcentrationInterferenceYes (Y) or No (N)
Copan 168C (rayon/twisted aluminumshaft)Copan1 swabN
Polyester / Aluminum shaft swabPuritan/Copan1 swabN
DNAzapInvitrogen1% V/VN
RnaseOutInvitrogen1% V/VN

a - It was observed that mucin at higher concentration (0.6%) led to loss of signal for some targets (loss of analyte detection) when extracted with NucliSENS easyMAG.

Table: Nasal influenza vaccine (FluMist) tested for BioCode® RPP assay
FluMist® 2010-2011 (V/V%)Influenza AInfluenza B
H1H1N1-2009H3
10%-+++
1%-+++
0.1%-+++
0.01%-+++
0.001%-+++
0.0001%-++a+
0.00001%---+a
0.000001%----

a - 2/3 replicates detected

Conclusion: None of the substances were shown to interfere with BioCode® RPP at the concentrations tested. However, it was observed that mucin at higher concentration (0.6%) could lead to loss of signal for some targets (loss of analyte detection) when extracted with easyMAG. The effect of mucin was dependent on the concentration in the sample tested.

Flu Mist was evaluated to be reactive as predicted with the BioCode RPP assay, therefore recent administration or contamination of specimens by Flu vaccine prior to NPS collection could lead to false detection by BioCode® RPP.

Competitive Inhibition

A study was performed to evaluate the potential for inhibition in samples with mixed infections. Targets were spiked into simulated NPS in UTM matrix with one target at high concentration (≥10°CFU/mL for bacteria and ≥105 units/mL for viruses) and two targets at low concentration (≤3x LoD). Common coinfections were determined by reviewing results of previous Respiratory Panel clinical trials from 510k summaries, publications/posters and internal clinical sample testing. Each sample was extracted in triplicate on the easyMAG and each extraction tested in singlet with the RPP on the BioCode® MDx-3000 system.

Results: Results are shown in the table below.

{42}------------------------------------------------

PanelDesignationViral/Bacteria StrainSourceLevelTiter TestedResult (n of3 Detected)
CompetitiveInhibitionSample 1Adenovirus species C Serotype 2ATCC VR-846High1x105 TCID50/mL3/3
Respiratory syncytial virus Type AZeptometrix0810040ACFLow0.99 TCID50/mL3/3
Influenza A H3N2A/Wisconsin/67/05aZeptometrix0810252CFLow12 TCID50/mL3/3
CompetitiveInhibitionSample 2Respiratory syncytial virus Type AZeptometrix0810040ACFHigh1x105 TCID50/mL3/3
Influenza A H3N2A/Wisconsin/67/05aZeptometrix0810252CFLow12 TCID50/mL3/3
Adenovirus species C Serotype 2ATCC VR-846Low18 TCID50/mL3/3
CompetitiveInhibitionSample 3Influenza A H3N2A/Wisconsin/67/05aZeptometrix0810252CFHigh1x105 TCID50/mL3/3
Adenovirus species C Serotype 2ATCC VR-846Low18 TCID50/mL3/3
Respiratory syncytial virus Type AZeptometrix0810040ACFLow0.99 TCID50/mL3/3
CompetitiveInhibitionSample 4Coronavirus OC43Zeptometrix0810024CFHigh1x105 TCID50/mL3/3
Human MetapneumovirusZeptometrix0810161CFLow45 TCID50/mL3/3
Bordetella pertussisZeptometrix801459Low45 CFU/mL3/3
CompetitiveInhibitionSample 5Human MetapneumovirusZeptometrix0810161CFHigh1x105 TCID50/mL3/3
Bordetella pertussisZeptometrix801459Low45 CFU/mL3/3
Coronavirus OC43Zeptometrix0810024CFLow0.12 TCID50/mL3/3
CompetitiveInhibitionSample 6Bordetella pertussisZeptometrix801459High1x106 CFU/mL3/3
Coronavirus OC43Zeptometrix0810024CFLow0.12 TCID50/mL3/3
Human MetapneumovirusZeptometrix0810161CFLow45 TCID50/mL3/3
CompetitiveInhibitionSample 7Influenza A H1N1 pdmCalifornia/07/09Zeptometrix0810165CFHigh1x105 TCID50/mL3/3
Parainfluenza Virus 3Zeptometrix0810016CFLow45 TCID50/mL3/3
CompetitiveInhibitionSample 8Human Rhinovirus type AZeptometrix0810012CFNLow3.6 TCID50/mL3/3
Parainfluenza Virus 3Zeptometrix0810016CFHigh1x105 TCID50/mL3/3
PanelDesignationViral/Bacteria StrainSourceLevelTiter TestedResult (n of3 Detected)
CompetitiveInhibitionSample 9Influenza A H1N1 pdmCalifornia/07/09Zeptometrix0810165CFLow1.2 TCID50/mL3/3
Human Rhinovirus type AZeptometrix0810012CFNHigh1x105 TCID50/mL3/3
Influenza A H1N1 pdmCalifornia/07/09Zeptometrix0810165CFLow1.2 TCID50/mL3/3
Parainfluenza Virus 3Zeptometrix0810016CFLow45 TCID50/mL3/3
CompetitiveInhibitionSample 10Mycoplasma pneumoniaeZeptometrix801579High1x106 CCU/mL3/3
Coronavirus NL63Zeptometrix0810228CFLow1.2 TCID50/mL3/3
Influenza B/Florida/4/2006Zeptometrix0810255CFLow0.04 TCID50/mL3/3
CompetitiveInhibitionSample 11Coronavirus NL63Zeptometrix0810228CFHigh1x105 TCID50/mL3/3
Influenza B/Florida/4/2006Zeptometrix0810255CFLow0.04 TCID50/mL3/3
Mycoplasma pneumoniaeZeptometrix801579Low45 CCU/mL3/3
CompetitiveInhibitionSample 12Influenza B/Florida/4/2006Zeptometrix0810255CFHigh1x105 TCID50/mL3/3
Mycoplasma pneumoniaeZeptometrix801579Low45 CCU/mL3/3
Coronavirus NL63Zeptometrix0810228CFLow1.2 TCID50/mL3/3

Table. Competitive inhibition testing results

{43}------------------------------------------------

Conclusion: All replicates from each pooled sample were valid and detected. No competitive inhibition was observed at the concentrations tested.

Cross Contamination/Sample Carryover

Carry-over contamination studies have been performed for the BioCode® MDx-3000 system in conjunction with the easyMAG (K180041) and MagNA Pure 96 systems (K190585). Since this study is not assay-specific, no additional testing was performed for BioCode® RPP.

Specimen Stability

A study was performed to assess the Specimen Stability limitations for the optimal performance of the BioCode® Respiratory Pathogen Panel (RPP) on the BioCode® MDx-3000. Representative organisms from the BioCode® RPP were spiked into prescreened negative natural NPS in UTM matrix at 3x LoD and assayed with 3 extractions on the easyMAG at each timepoint 0). This study assessed the following storage conditions:

{44}------------------------------------------------

Nasopharyngeal Swabs in transport media
Storage tempTarget (storage time)Test time points
Room Temperature 25°C8 hours8,12 hours
Refrigerated 4°C7 days5, 7, 10 days
Fresh Vs Frozen (-70°C)90 days (2x freeze/thaw)30, 60, 90 days
Purified nucleic acids
Storage tempTarget (storage time)Test time points
Refrigerated 4°C8 hours8, 12 hours
Fresh Vs Frozen (-70°C)90 days (2x freeze/thaw)30, 60, 90 days

Table. Specimen stability conditions

Table. Contrived samples (3x LoD in prescreened negative natural matrix)

Sample NameOrganismSource
RP AAdenovirus B Serotype 7aZeptometrix 0810021CF
RP AMycoplasma pneumoniaeZeptometrix 0801579
RP AInfluenza A H3N2 A/Wisconsin/67/2005Zeptometrix 0810252CF
RP BRespiratory Syncytial Virus (Type A)Zeptometrix 0810040ACF
RP BH1N1pdm California/07/09Zeptometrix 0810165CF
RP BHuman Metapneumovirus (16; type A1)Zeptometrix 0810161CF
RP CParainfluenza Virus 3Zeptometrix 0810016CF
RP CCoronavirus NL63Zeptometrix 0810228CF
RP CFlu B/Florida/4/2006Zeptometrix 0810255CF

Results: All replicates were Valid and detected for all conditions assayed. Summary results are shown in the table below.

{45}------------------------------------------------

Premarket Notification 510(k)

Sample ASample BSample C
Sample TypeTempTime(HoursorDays)ADVMPNFlu AH3N2RSVFlu AH1N12009pdmhMPVPIV3NL63Flu B
Fresh (baseline)N/A07/77/77/77/77/77/77/77/77/7
Spiked NPSRoom Temp8 Hour3/33/33/33/33/33/33/33/33/3
12 Hour3/33/33/33/33/33/33/33/33/3
4ºC5 Day3/33/33/33/33/33/33/33/33/3
7 Day3/33/33/33/33/33/33/33/33/3
10 Day3/33/33/33/33/33/33/33/33/3
-80°C30 Day3/33/33/33/33/33/33/33/33/3
60 Day3/33/33/33/33/33/33/33/33/3
90 Day3/33/33/33/33/33/33/33/33/3
Extracted Nucleicacid4ºC8 Hour3/33/33/33/33/33/33/33/33/3
12 Hour3/33/33/33/33/33/33/33/33/3
-80°C(2x Freeze/thaw)30 Day3/33/33/33/33/33/33/33/33/3
60 Day3/33/33/33/33/33/33/33/3
90 Day3/33/33/33/33/33/33/33/3

Table. Qualitative results from Specimen Stability study

Conclusions: All replicates were Valid and detected for each specimen stability condition.

Matrix equivalency: Simulated Nasopharyngeal Swabs (sNPS) vs Native Nasopharyngeal Swabs (NPS)

Analytical validation studies are mostly performed with contrived specimens (i.e., spiking known concentrations of pathogens into negative NPS in UTM or VTM matrix). This would require a large volume of negative natural NPS to be collected and screened, while typically only between 1.0 to 2.0 mL per donor could be obtained after NPS collection and screening to confirm negative status. This makes collection from donors burdensome. An analytical study was performed to assess the performance equivalency of testing a simulated NPS (sNPS) in UTM matrix compared to testing natural NPS in UTM using the BioCode RPP, with both the NucliSENS easyMAG and the MagNA Pure 96 extraction systems. The sNPS in UTM matrix consists of 2x103 HeLa cells/mL diluted in UTM.

Samples were contrived in negative natural NPS in UTM matrix with matrix with multispiked pathogens at close to LoD levels (approximately 1.5x to 15.0xLoD). Each sample was extracted in guadruplicate on both the easyMAG and the MagNA Pure 96 extraction system and tested singly with the BioCode RPP on the BioCode MDx-3000 system.

Results: All four replicates of pathogens in three pools were Valid and detected. No unexpected targets were detected in the sNPS.

{46}------------------------------------------------

PoolsOrganismSourceConcentrationMultiple of LoD testedDetected (N of 4)
easyMAGMagNA Pure 96easyMAGMagNA Pure 96
Natural NPSsNPSNatural NPSsNPS
Pool AAdenovirus E Serotype 4Zeptometrix 0810070CF0.6 TCID50/mL15x LoD15x LoD4/44/44/44/4
Chlamydia pneumoniaeATCC 5359275.0 CFU/mL4.5x LoD2.3x LoD4/44/44/44/4
Influenza A H3N2 A/Wisconsin/67/2005Zeptometrix 0810252CF6.0 TCID50/mL1.5x LoD4.6x LoD4/44/44/44/4
Pool BInfluenza A H1N1/New Caledonia/ 20/99Zeptometrix 0810036CF67.5 TCID50/mL4.5x LoD13.5x LoD4/44/44/44/4
Respiratory Syncytial Virus (Type A)Zeptometrix 0810040ACF1.5 TCID50/mL4.5x LoD4.5x LoD4/44/44/44/4
Human Metapneumovirus (16; type A1)Zeptometrix 0810161CF67.5 TCID50/mL4.5x LoD4.5x LoD4/44/44/44/4
Pool CInfluenza B/Florida/4/2006Zeptometrix 0810255CF0.06 TCID50/mL6x LoD6x LoD4/44/44/44/4
Coronavirus OC43Zeptometrix 0810024CF0.06 TCID50/mL1.5x LoD6x LoD4/44/44/44/4
Parainfluenza Virus 4Zeptometrix 0810060CF13.5 TCID50/mL1.5x LoD1.5x LoD4/44/44/44/4
Negative ControlN/AN/AN/AN/AN/AN/AN/AN/A

Table. Results of natural NPS Vs sNPS comparison

Conclusions: No matrix effect was observed in this study. All replicates for each contrived sample assayed with the simulated matrix (sNPS- HeLa cells in UTM) and natural matrix were valid and detected for expected targets per the algorithm using both extraction systems easyMAG and MagNA Pure 96 extraction systems.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.