Intended Use

The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:

  • · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
  • · Rheumatoid arthritis or traumatic arthritis
  • · Correction of functional deformity
  • · Non-union femoral neck fracture
  • Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.
    The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
Device Description

The present 510k submission is for a non-cemented primary hip prosthesis that consists of a 4-part total hip replacement system including femoral stem, femoral Head, acetabular poly liner, and acetabular metal (or shell) components. The femoral head component articulates within the poly acetabular component. The poly acetabular component snaps into the metal acetabular component. The design and sizing of the components correspond to natural hip anatomy to restore normal rotation, extension, and flexion.
The femoral stem component is made from forged Ti 6AL 4V, with a Ti plasma spray coating. The uni-polar femoral head component is made from CoCr, or BIOLOX® delta ceramic, in 28, 32, and 36 mm sizes.
The acetabular poly liner component is made of standard UHMWPE polyethylene in both hooded and non-hooded options. The acetabular poly liner component is offerent inner and outer diameter combinations to accept various size uni-polar femoral heads and acetabular metal components.
The acetabular metal (or shell) component is made from forged Ti 6AL 4V with a Ti porous coating. It is available in nohole, cluster-hole, and revision multi-hole styles.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Precision Casting Solutions Total Hip System." This document establishes substantial equivalence to a predicate device, the "Progressive Orthopaedic Total Hip System," based on identical design, materials, indications for use, and a reliance on preclinical performance testing previously conducted for the predicate device.

Therefore, the acceptance criteria and the study that proves the device meets the acceptance criteria are not directly reported as a standalone study for the Precision Casting Solutions Total Hip System. Instead, the substantial equivalence relies on the performance testing of the predicate device.

Here's the breakdown of the requested information based on the provided text, recognizing that it refers to the predicate device's testing:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical acceptance criteria for each test. Instead, it states that "the results confirm that all components of the Precision Casting Solutions Total Hip System exhibit the appropriate mechanical characteristics for total hip joint replacement and is substantially equivalent to the predicate devices." This implies that the performance of the predicate device in these tests met the necessary mechanical characteristics.

Test TypeReported Performance (for Predicate Device)
Fatigue Performance Test for StemAppropriate mechanical characteristics
Fatigue Performance Test for the Neck Portion of the StemAppropriate mechanical characteristics
Disassembly Force Test for Neck Taper/Femoral Head InterfaceAppropriate mechanical characteristics
Range of Motion Test for Total Hip SystemAppropriate mechanical characteristics
Burst Strength Test for Ceramic Femoral Heads (Static Compression)Appropriate mechanical characteristics
Cyclic Fatigue Test for Ceramic Femoral Heads (Cyclic Compression)Appropriate mechanical characteristics
Post-Cyclic Fatigue Burst Test for Ceramic Femoral Heads (Static Compression)Appropriate mechanical characteristics
Pull-Off Test for Ceramic Femoral HeadsAppropriate mechanical characteristics
Rotational Stability Test for Ceramic Femoral HeadsAppropriate mechanical characteristics
Torsional Properties Test for Bone ScrewsAppropriate mechanical characteristics
Driving Torque Test for Bone ScrewsAppropriate mechanical characteristics
Axial Pull-Out Strength Test for Bone ScrewsAppropriate mechanical characteristics
Lever-Out Test for Acetabular Shell/Liner AssemblyAppropriate mechanical characteristics
Torque-Out Test for Acetabular Shell/Liner AssemblyAppropriate mechanical characteristics
Push-In Test for Acetabular Shell/Liner AssemblyAppropriate mechanical characteristics
Push-Out Test for Acetabular Shell/Liner AssemblyAppropriate mechanical characteristics

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes (e.g., number of stems, heads, etc.) used for each individual performance test of the predicate device. It also does not mention the data provenance beyond stating that the tests were "preclinical."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable as the studies described are preclinical mechanical and material performance tests, not clinical evaluations requiring expert interpretation of results for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods are typically used in clinical studies involving expert assessments, not in preclinical mechanical testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a total hip replacement system, and the studies mentioned are preclinical mechanical tests, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context refers to the established mechanical and material properties required for a total hip replacement system to function safely and effectively. This is determined by engineering standards, biomechanical principles, and regulatory requirements (e.g., ISO standards for implants) that define "appropriate mechanical characteristics."

8. The sample size for the training set

This is not applicable. The described studies are preclinical mechanical tests, not machine learning or AI models with training sets.

9. How the ground truth for the training set was established

This is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square.

December 17, 2019

Precision Casting Solution, LLC Jim Moore Vice President 200031 Henrici Road OREGON CITY OR 97045

Re: K192450

Trade/Device Name: Precision Casting Solutions Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: August 30, 2019 Received: September 6, 2019

Dear Jim Moore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192450

Device Name Precision Casting Solutions Total Hip System

Indications for Use (Describe)

The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:

  • · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
  • · Rheumatoid arthritis or traumatic arthritis
  • · Correction of functional deformity
  • · Non-union femoral neck fracture

· Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.

The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the letters "P", "C", and "S" in both blue and pink colors. There is a medical symbol in the background. In the foreground, there are two DNA strands, one in blue and the other in pink.

510(k) Summary 1.

The following 510(k) Summary is provided in accordance with 21 CFR 807.92.

510(k) Owner and Registration 1.1

Owner's Name:Precision Casting Solutions, INC
Address:20031 Henrici Road., Oregon City, OR 97045
Phone Number:(503) 421-1251
Fax Number:N/A
Date Summary Prepared:August 30, 2019
Establishment RegistrationNumber:N/A

1.2 510(k) Contact

Contact:Jim Moore
Address:20031 Henrici RoadOregon City, OR 97045
Phone Number:(503) 421-1251
Fax Number:N/A
Contact Person:Jim Moore

1.3 Device Name and Classification

Device Trade Name:Precision Casting Solutions Total Hip System
Device Common Name:Total Hip Replacement
Regulation Number andDescription:21 CFR 888.3558
Device Class:Class II
Product Codes:LPH, LZO
Advisory Panel:87 (Orthopedic)

Legally Marketed Predicate 1.4

Precision Casting Solutions is utilizing the Progressive Orthopaedic Total Hip System as the predicate device (K143314/K151424). The Precision Casting Solutions Total Hip System features component designs, materials, indications, manufacturing methods, and contract manufacturer's that are identical to the Progressive Orthopaedic Total Hip System.

1.5 Device Description

The present 510k submission is for a non-cemented primary hip prosthesis that consists of a 4-part total hip replacement system including femoral stem, femoral Head, acetabular poly liner, and acetabular metal (or shell) components. The femoral head component articulates within the poly acetabular component. The poly acetabular component snaps into the metal acetabular component. The design and sizing of the components correspond to natural hip anatomy to restore normal rotation, extension, and flexion.

The femoral stem component is made from forged Ti 6AL 4V, with a Ti plasma spray coating. The uni-polar femoral

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the letters "P", "C", and "S" on both the left and right sides. The letters on the left are in blue, while the letters on the right are in pink. In the background, there is a faint image of a caduceus, a symbol associated with medicine. Below the letters and the caduceus, there is a depiction of a DNA double helix, with one strand in blue and the other in pink.

head component is made from CoCr, or BIOLOX® delta ceramic, in 28, 32, and 36 mm sizes.

The acetabular poly liner component is made of standard UHMWPE polyethylene in both hooded and non-hooded options. The acetabular poly liner component is offerent inner and outer diameter combinations to accept various size uni-polar femoral heads and acetabular metal components.

The acetabular metal (or shell) component is made from forged Ti 6AL 4V with a Ti porous coating. It is available in nohole, cluster-hole, and revision multi-hole styles.

1.6 Indications For Use

The Precision Casting Solutions Total Hip System is identical to the predicate Progressive Orthopaedic Total Hip System. The indications for use are also identical.

The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:

  • · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
  • Rheumatoid arthritis or traumatic arthritis .
  • · Correction of functional deformity
  • . Non-union femoral neck fracture
  • Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other ● techniques.

The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.

1.7 Summary of Technological Characteristics

The Precision Casting Solutions Total Hip system is identical to the predicate Progressive Orthopaedic Total Hip System. Both devices are manufactured from identical materials, possess the same the same packaging and sterilization processes. Extensive precimical testing was performed on the predicate devices per K143314 and K151424 and found substantially equivalent. The performance tests are listed below and used herein to establish substantial equivalence (Section Error! Reference source not found. Error! Reference source not found.)

Given that the subject device is identical to the subject system is substantially equivalent to the predicate systems (K143314/K151424).

1.8 Performance Testing

Extensive preclinical performance testing was conducted, and substantial equivalence determined per K143314 and K151424. The components of the subject device are identical to the predicate device, and therefore the predicate device testing demonstrates substantial equivalence for the subject device. The results confirm that all components of the Precision Casting Solutions Total Hip System exhibit the appropriate mechanical characteristics for total hip joint replacement and is substantially equivalent to the predicate devices.

  • Fatigue Performance Test for Precision Casting Solutions Stem .
  • . Fatigue Performance Test for the Neck Portion of the Precision Casting Solutions Stem
  • Disassembly Force Test for Precision Casting Solutions Neck Taper/Femoral Head Interface
  • Range of Motion Test for the Precision Casting Solutions Total Hip System
  • Burst Strength Test for Precision Casting Solutions Ceramic Femoral Heads (Static Compression)
  • . Cyclic Fatigue Test for Precision Casting Solutions Ceramic Femoral Heads (Cyclic Compression)
  • . Post-Cvclic Fatigue Burst Test for Precision Casting Solutions Ceramic Femoral Heads (Static Compression)
  • . Pull-Off Test for Precision Casting Solutions Ceramic Femoral Heads
  • Rotational Stability Test for Precision Casting Solutions Ceramic Femoral Heads
  • Torsional Properties Test for Precision Casting Solutions Bone Screws
  • Driving Torque Test for Precision Casting Solutions Bone Screws ●
  • Axial Pull-Out Strength Test for Precision Casting Solutions Bone Screws .

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters "P", "C", and "S" arranged in a specific pattern. On the left side, the letters "P", "C", and "S" are in blue, while on the right side, they are in pink. Below the letters, there is a depiction of a DNA double helix, with one strand in blue and the other in pink. In the background, there is a medical symbol, suggesting a connection to healthcare or medicine.

  • Lever-Out Test for Precision Casting Solutions Acetabular Shell/Liner Assembly .
  • Torque-Out Test for Precision Casting Solutions Acetabular Shell/Liner Assembly
  • Push-In Test for Precision Casting Solutions Acetabular Shell/Liner Assembly ●
  • Push-Out Test for Acetabular Shell/Liner Precision Casting Solutions Assembly ●

1.9 Conclusions

The Precision Casting Solutions Total Hip System is identical to the predicate Progressive Orthopaedic Total Hip System. The subject device has the same design features, materials, and indications for use as the predicate devices. The testing performed for the predicate device indicates that the Precision Casting Solutions Total Hip System is safe for clinical use.

The Precision Casting Solutions Total Hip System is substantially equivalent to the predicate device.

N/A