K Number
K192386

Validate with FDA (Live)

Device Name
KeraStat Cream
Manufacturer
Date Cleared
2020-07-16

(317 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for the management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, superficial injuries, cuts, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts.

KeraStat® Cream is not indicated for full thickness or third degree will be available by prescription.

Device Description

KeraStat® Cream is a non-sterile, non-implantable wound dressing intended to provide a moist environment in the management of a variety of partial thickness dermal wounds. KeraStat® Cream is provided in a screw top tube for multiple uses. Each tube contains 1 oz (29.6 mL) of KeraStat® Cream, which contains 5% keratin protein incorporated into a cream base.

AI/ML Overview

The provided document is a 510(k) summary for KeraStat® Cream. It focuses on demonstrating substantial equivalence to a predicate device (KeraStat® Gel) and a reference device (Biafine®), rather than presenting a study with specific acceptance criteria and performance data for the KeraStat® Cream's effectiveness in wound management.

Therefore, the document does not contain the acceptance criteria or a study that proves the device meets specific performance criteria related to its clinical efficacy in terms of wound healing endpoints.

Here's a breakdown of the requested information based on the provided text, indicating what is not available:

1. A table of acceptance criteria and the reported device performance

  • Not Available: The document does not provide a table with specific acceptance criteria for clinical efficacy (e.g., wound size reduction, healing time, infection rates) or reported device performance against such criteria. The "Performance Data Summary of Non-Clinical Tests" lists types of tests (e.g., protein content, viscosity, pH, microbial contamination, stability) performed to ensure product quality and safety, but these are not clinical acceptance criteria for wound healing efficacy.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Available for clinical efficacy: The document mentions "Radiation Dermatitis Clinical Study", "Repeat Insult Patch Test", and "Skin Prick Test" as "human clinical testing". However, it does not provide details on the sample size for these tests, their design (retrospective or prospective), or data provenance. These clinical tests appear to be primarily for safety (e.g., sensitization, irritation) or to support specific indications like radiation dermatitis, not necessarily general wound healing efficacy benchmarks.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable/Not Available: This question typically applies to studies involving interpretation of medical images or diagnostic classification, where expert consensus establishes ground truth. Since this document is for a wound dressing and does not describe such a study for clinical efficacy, this information is not relevant or provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable/Not Available: Similar to point 3, this is relevant for studies involving subjective assessments or classifications by multiple readers, which is not described here for the clinical efficacy of the wound dressing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable/Not Available: This device (KeraStat® Cream) is a medical device (wound dressing), not an Artificial Intelligence (AI) device. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable/Not Available: This device is a wound dressing, not an algorithm, so this question is not relevant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Available for clinical efficacy: For the mentioned clinical tests (e.g., Radiation Dermatitis Clinical Study), the document does not specify how "ground truth" (e.g., actual wound status, healing progress) was established or measured. For biocompatibility and performance tests, the "ground truth" is typically defined by standardized ISO or USP methods.

8. The sample size for the training set

  • Not Applicable/Not Available: This device is a wound dressing, not a machine learning model, so there is no training set in this context.

9. How the ground truth for the training set was established

  • Not Applicable/Not Available: As identified in point 8, there is no training set for this device.

In summary: The provided 510(k) summary primarily focuses on demonstrating that KeraStat® Cream is substantially equivalent to existing predicate and reference devices based on intended use, principles of operation, and technological characteristics (including biocompatibility and physical performance tests). It does not present a detailed clinical study with specific acceptance criteria and efficacy performance data in the way an AI or diagnostic device submission might. The "clinical tests" mentioned appear to be more geared towards safety and specific indications (like radiation dermatitis) rather than comprehensive efficacy endpoints for all listed indications.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.

July 16, 2020

KeraNetics Luke Burnett CEO and CSO 200 East First Street, Suite 101, Box#4 Winston Salem, North Carolina 27101

Re: K192386

Trade/Device Name: KeraStat Cream Regulatory Class: Unclassified Product Code: KGN, Dated: June 12, 2020 Received: June 15, 2020

Dear Luke Burnett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anjana Jain, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192386

Device Name KeraStat® Cream

Indications for Use (Describe)

KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for the management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, superficial injuries, cuts, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts.

KeraStat® Cream is not indicated for full thickness or third degree will be available by prescription.

Type of Use (Select one or both, as applicable)
---------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY--- KERASTAT® CREAM

Submitter Name:KeraNetics, Inc.
Submitter Address:200 East First Street, Box #4, Suite 102Winston Salem, NC 27101
Phone Number:1.336.202.1307
Fax Number:1.336.725.0619
Contact Person:Luke Burnett, PhDCEO and CSO, KeraNetics, Inc.
Phone Number:1.336.202.1307
Fax Number:1.336.725.0619
Email:lburnett@keranetics.com
Date Prepared:July16, 2020
Device Trade Name:KeraStat® Cream
Device Common Name:Wound Dressing
Classification Number:Unclassified
Product Code:KGN
Product Type Name:Dressing, Wound, Collagen
Predicate Device:KeraStat® GelKeraNetics Inc.K162759Cleared on June 2, 2017
Reference Device:Biafine®Valeant PharmaceuticalsK173549Cleared on August 13, 2018
Device Description:KeraStat® Cream is a non-sterile, non-implantable wounddressing intended to provide a moist environment in themanagement of a variety of partial thickness dermalwounds. KeraStat® Cream is provided in a screw top tubefor multiple uses. Each tube contains 1 oz (29.6 mL) ofKeraStat® Cream, which contains 5% keratin proteinincorporated into a cream base.

{4}------------------------------------------------

Statement of Intended Use :

KeraStat® Cream is intended to maintain a moist wound environment. KeraStat® Cream is indicated for management of a number of partial thickness skin wounds such as: partial thickness (first and second degree) burns, severe sunburns, superficial injuries, cuts, abrasions, and incisions/surgical wounds. Under the direction of a healthcare professional, KeraStat® Cream also may be used in the management of dry, light, and moderately exuding partial thickness wounds including: pressure (stage I-II) ulcers, venous stasis ulcers, ulcers caused by mixed vascular etiologies, diabetic ulcers, radiation dermatitis, donor sites, and grafts.

KeraStat® Cream is not indicated for full thickness or third degree burns. This device will be available by prescription.

Comparison to the Predicate and Reference Devices:

KeraStat® Cream and the predicate wound dressing, KeraStat® Gel, are both designed for dry, light, and moderately exuding partial thickness wounds. The wound dressings are applied to the wound and function to provide a moist environment for wound management. The operational principles of the subject and predicate device are identical, and they contain identical concentrations and composition of human derived keratin protein. The only differences between KeraStat® Cream and the KeraStat® Gel predicate device are that KeraStat® Cream is provided in an emulsion base, is nonsterile, and is intended for multiple uses.

KeraStat® Cream and the reference wound dressing, Biafine®, have the same intended use to provide a moist environment for wound management including wounds caused by radiation dermatitis. Biafine® has similar technological characteristics as KeraStat® Cream, as both are non-sterile, multi-use cream emulsions composed of a protein/amine, emollients, emulsifiers, thickeners, and preservatives.

Performance Data Summary of Non-Clinical Tests:

The following testing was performed to demonstrate substantial equivalence:

  • Biocompatibility testing: .
    • Cytotoxicity (ISO 10993-5) o
    • Sensitization (ISO 10993-10) o
    • o Irritation/Intracutaneous reactivity (ISO 10993-10)
    • Acute systemic toxicity (ISO 10993-11) O
    • Toxicology Risk Assessment (ISO 10993-17) O
    • Chemical Characterization (ISO 10993-18) o
    • Implantation Test (ISO 10993-6) o
    • o Rabbit Pyrogenicity Test (USP Chapter <151>)

{5}------------------------------------------------

  • Performance testing:
    • o Protein content
    • o Viscosity
    • o pH
    • Microbial contamination o
    • Preservative effectiveness o
    • o Size-exclusion chromatography (SEC)
    • o Stability

Summary of Clinical Tests:

The following human clinical testing was performed:

  • Radiation Dermatitis Clinical Study ●
  • Repeat Insult Patch Test ●
  • Skin Prick Test

Testing Summary:

Biocompatibility, performance, and clinical testing support the safety and effectiveness of KeraStat® Cream.

Subs tantial Equivalence :

KeraStat® Cream has the same intended use, same principles of operation, and similar technological characteristics as the predicate device KeraStat® Gel (K162759). While the subject device differs from the predicate device with respect to its additional components and multi-use design, both devices share the same mode of action in that they provide a moist environment to support wound healing. Any differences in technological characteristics outlined in Table 1 do not raise new questions about the safety and effectiveness of KeraStat® Cream.

KeraStat® Cream has similar intended use, same principles of operation, and similar technological characteristics as the reference device Biafine® (K173549). Both devices share the same mode of action in that they provide a moist environment to support wound healing, including wounds due to radiation dermatitis. Biafine® has similar technological characteristics as KeraStat® Cream, as both are non-sterile, multiuse cream emulsions composed of a protein/amine, emollients, emulsifiers, thickeners, and preservatives. Any differences in technological characteristics outlined in Table 1 do not raise new questions about the safety and effectiveness of KeraStat® Cream.

{6}------------------------------------------------

Table 1. Substantial Equivalence Information
Trade NameSubject Device:KeraStat® CreamPredicate Device:KeraStat® GelReference Device:Biafine®
510(k) No.K162759K173549
Intended UseWound dressing formanagement of partialthickness woundsWound dressing formanagement of partialthickness woundsWound dressing formanagement of partial andfull thickness wounds
IndicationsDry, light, and moderatelyexuding partial thicknesswounds such as: first andsecond degree burns, severesunburns, superficial injuries,cuts, abrasion, and surgicalwounds; may also be usedunder the guidance of a healthcare professional in themanagement of pressure(stage I-II) and venous stasisulcers, ulcers caused bymixed vascular etiologies,diabetic ulcers, radiationdermatitis, donor sites andgraftsDry, light, and moderatelyexuding partial thicknesswounds such as: first andsecond degree burns, severesunburns, superficial injuries,cuts, abrasion, and surgicalwounds; may also be usedunder the guidance of a healthcare professional in themanagement of pressure(stage I-II) and venous stasisulcers, ulcers caused by mixedvascular etiologies, diabeticulcers, donor sites and graftsOTC indications:Management of superficialwounds such as minor cuts,minor scrapes, minorirritations, minorabrasions, minor blisters, 1stdegree burns includingsunburns, minor skinirritations following post nonablative laser therapyprocedures,microdermabrasion therapy orsuperficial chemical peels.May also be used for relief ofitch, pain and burning fromminor skin irritations,lacerations, abrasions andminor burns.Rx Indications: managementof full thickness wounds,pressure sores, dermal ulcersincluding lower leg ulcers,radiation dermatitis, donorsites and 2nd degree burns.May also be used for relief ofitch, pain and burning fromminor skin irritations,lacerations, abrasions andminor burns.
Mode of ActionProvides a moist environmentthat is supportive of woundhealingAbsorbs exudate and providesa moist environment that issupportive of wound healingProvides a moist environmentthat is supportive of woundhealing
TechnologicalCharacteristicsMixture of human hairderived keratin proteins in acream base intended to acreate a moist environmentand packaged as a non-Mixture of human hairderived keratin proteinsintended to absorb exudateand create a moistenvironment and packaged asTrolamine/sodiumalginate ina cream base intended to acreate a moist environmentand packaged as a non-sterile,multi-use dressing
Trade NameSubject Device:Predicate Device:Reference Device:
KeraStat® CreamKeraStat® GelBiafine®
510(k) No.K162759K173549
ComponentsPurified water, glycerin,mineral oil, keratin, caprylictriglyceride, dimethicone,sodiumpolyacrylate,hydrogenated polydecene,trideceth-6, sodiumstearoylglutamate, phenoxyethanol,and ethylhexylglycerinPurified water, keratin,phenoxyethanol, carbomer,sodiumhydroxide, andethylhexylglycerinPurified water, liquid paraffin,ethylene glycol monostearate,stearic acid, propyleneglycol, paraffin wax.squalane, avocado oil,trolamine/sodiumalginate,triethanolamine,cetylpalmitate, methylparaben(sodiumsalt), sorbic acid (aspotassiums alt ), propylparaben (sodiumsalt), and fragrance
Form of WoundDressingCreamHydrogelCream
ApplicationMethodTopicalTopicalTopical
SterilityNon-sterileSterileNon-sterile
NumberofUsesMulti-useSingle useMulti-use
PrescriptionUseYesYesYes (also OTC)

Table 1. Substantial Equivalence Information

{7}------------------------------------------------

Conclusion:

KeraStat® Cream demonstrates substantial equivalence to the predicate device, KeraStat® Gel (K162759) and there are not different questions of safety and effectiveness.

N/A