K Number
K192372

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-02-14

(168 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ASCEND Image Viewer is a software-only medical device intended for use by trained clinicians to display medical images. ASCEND Image Viewer provides viewing of and measurements on echocardiograms from compatible ultrasound systems that support transthoracic echocardiogram, transesophageal echocardiogram, stress echocardiogram, cath, and vascular study types.

ASCEND Image Viewer is for prescription use only.

Device Description

The ASCEND Image Viewer is a Medical Device for viewing and measuring cardiovascular echocardiograms in DICOM format.

The software has features for displaying and measuring cardiovascular images acquired from echocardiograms. Images supported include:

  • . 2D
  • Color Doppler
  • M-mode
  • . Spectral Doppler (Pulsed Wave and Continuous Wave)

The ASCEND Image Viewer provides Cine play review and control.

Standard Review supported study types include:

  • TTE
  • TEE
  • Cath
  • . Stress Echo
  • Vascular

Supported Measurements include:

  • Linear
  • . Volume (method of discs)
  • Velocity time integral
  • Pressure half-time
  • . Area
  • Time
  • Velocity
  • Angle (measurement of two lines with a common vertex drawn on the image) .

The user can make measurements which persist for later use within the ASCEND Image Viewer.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ASCEND Image Viewer, based on the provided text:

Acceptance Criteria and Device Performance

The provided document describes the ASCEND Image Viewer as a software-only medical device for displaying and measuring echocardiograms. The comparison is made against a predicate device, TomTec-Image Arena (K132544). The clinical testing and "Brief Discussion of Test Results Submitted: Bench Testing" section indicate that the device performed effectively and was found to be substantially equivalent to the predicate where functionality intersected.

However, the document does not explicitly list quantitative acceptance criteria for specific performance metrics (e.g., accuracy of measurements, speed of image display) or provide reported numerical performance values for the ASCEND Image Viewer in a table format. Instead, the acceptance is based on demonstrating substantial equivalence in functionality, intended use, technological characteristics, and typical users, along with adherence to design specifications and risk assessment.

The document states:

  • "The Graphic User Interface conforms to the ASCEND Image Viewer functional specification."
  • "The comparison ASCEND Image Viewer with the predicate devices shows that the ASCEND Image Viewer has similar functionality, intended use, technological characteristics, and typical users as the predicate devices."
  • "Verification and validation activities ensure that the design specifications are met, and that ASCEND Image Viewer does not introduce new issues concerning safety and effectiveness."

Given this, I will describe the characteristics compared rather than specific quantitative acceptance criteria and reported performance, as those are not provided in a measurable format.

Characteristic Compared to Predicate DeviceAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
Software only deviceDevice must be software-only.Yes (software-only).
Web accessibleDevice must be web-accessible.Yes (web-accessible).
Display ImagesDevice must display still images and image sequences.Yes (displays images).
Select Images for closer examinationDevice must allow selection of images for closer examination.Yes (allows image selection).
Echocardiogram analysis and review capabilityDevice must provide manual echocardiogram analysis and review.Manual (equivalent to predicate).
Data SourceDevice must handle ultrasound data.Ultrasound (predicate handled multiple modalities, but no impact on safety/effectiveness noted for this difference).
DICOM compliantDevice must be DICOM compliant.Yes (DICOM compliant).
Ultrasound mode utilized for echo analysisDevice must support 2D, Color Doppler, M-mode, Spectral Doppler images.2D, Color Doppler, M-mode, Spectral Doppler images (predicate used B mode, but no impact on safety/effectiveness noted for this difference).
Viewing mode – 2DDevice must have 2D viewing mode.Yes (2D viewing mode).
Viewing mode - 3DDevice is not required to have 3D viewing mode (distinguishes from predicate).No (does not have 3D viewing mode, acknowledged as a difference with no impact on safety/effectiveness).
General Cardiac MeasurementsDevice must provide specified measurements (Linear, Angle, Area, Volume (method of disks), Time, Velocity, VTI, PHT) and verify these differences from the predicate to assure no impact on safety and effectiveness.Provided measurements: Linear (distance), Angle, Area, Volume (method of disks), Time, Velocity, VTI, PHT. "Different features from predicate have been verified and validation to assure that there is no impact on safety and effectiveness."
Display formatDevice must support AVI, BMP, JPEG or DCM export.Yes (supports these export formats).
Operates on off the shelf hardwareDevice must operate on off-the-shelf hardware.Yes (operates on off-the-shelf hardware).
OSDevice must operate on Microsoft Windows.Microsoft Windows 7 and above.
Overall PerformanceDevice should perform effectively for its intended use, with no late-arising risks or safety concerns."ASCEND Image Viewer performed effectively its intended use. ... No late-arising risks or other safety concerns were recorded during clinical testing."

Study Information

The document describes both Bench Testing and Clinical Testing.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a number of cases or images for either bench or clinical testing.
    • Data Provenance:
      • Bench Testing: The test plan covered "all aspects of the functionality, exercising every user interface, all menus and submenus, as well as the measurement tools." No specific data type or origin is given for the bench tests beyond general functional and DICOM compliance verification.
      • Clinical Testing: The testing occurred "in US city-hospital environments." The data used would have been clinical echocardiograms from these environments. The studies appear to be prospective in the sense that the device was evaluated during its development cycle with "clinical testers" in a real-world setting, but it's not a formal prospective clinical trial for efficacy.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Clinical Testing: The test involved "resident cardiologists, cardiology fellows, and sonographers." The number of these experts is not specified. Their role was to assess the effective performance of the device's intended use and its substantial equivalence to the predicate. The document doesn't explicitly state they established a "ground truth" for specific diagnostic findings, but rather that they evaluated the device's functional performance in a clinical context.
    • Bench Testing: No experts are mentioned for ground truth establishment during bench testing.
  3. Adjudication method for the test set:

    • Not specified. The clinical testing describes users evaluating the device's performance, rather than an adjudication process for a diagnostic outcome.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a formal MRMC comparative effectiveness study involving human readers with and without AI assistance was not conducted or described. The ASCEND Image Viewer is described as an image viewing and measurement tool, not an AI-assisted diagnostic algorithm in the sense of providing automated interpretations. The comparison was to a predicate image viewer.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device itself is a "software-only medical device intended for use by trained clinicians to display medical images." Its performance is inherently linked to human interaction (displaying images and making measurements). The bench testing would represent a form of standalone functional testing of the algorithm's capabilities (e.g., DICOM compliance, measurement calculations), but not an evaluation of its diagnostic performance in a "standalone" context separate from clinical use. The clinical testing explicitly involved "trained clinicians."
  6. The type of ground truth used:

    • For Bench Testing: "Ground truth" was conformity to the ASCEND Image Viewer functional specification and DICOM standard compliance. Risk analysis according to ISO 14971 was also a benchmark.
    • For Clinical Testing: "Ground truth" was implicitly based on the consensus or clinical judgment of the "resident cardiologists, cardiology fellows, and sonographers" during their use of the device, where they determined if it "performed effectively its intended use" and was "substantially equivalent" to the predicate. It was an evaluation of the tool in practice rather than against a predefined diagnostic gold standard (e.g., pathology report) for individual cases.
  7. The sample size for the training set:

    • Not applicable/Not provided. The ASCEND Image Viewer is described as an image viewing and measurement system, not a machine learning or AI algorithm that would typically require a "training set" in the context of image analysis. The functionality described (display, measurements, Cine play) suggests a deterministic software rather than one that learns from data.
  8. How the ground truth for the training set was established:

    • Not applicable, as no training set for a machine learning model is mentioned.

{0}------------------------------------------------

February 14, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

ASCEND HIT % Robert A. Miller Director of Quality Assurance and Regulatory Affairs 801 Warrenville Road, Suite 200 LISLE IL 60532

Re: K192372

Trade/Device Name: ASCEND Image Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 13, 2020 Received: January 16, 2020

Dear Mr. Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192372

Device Name ASCEND Image Viewer

Indications for Use (Describe)

ASCEND Image Viewer is a software-only medical device intended for use by trained clinicians to display medical images. ASCEND Image Viewer provides viewing of and measurements on echocardiograms from compatible ultrasound systems that support transthoracic echocardiogram, transesophageal echocardiogram, stress echocardiogram, cath, and vascular study types.

ASCEND Image Viewer is for prescription use only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 1

K192372

510(k) Summary

(as required by 21 CFR 807.92)

Date: 20-Jan-2020

A.Submitter's Information:
Name:ASCEND HIT
Address:801 Warrenville RoadSuite 200Lisle, Illinois 60532 USA
Telephone Number:(844) 413-2610
Contact Name:Andy Miller
Contact Email address:amiller@ascendhit.com
B.Device Information
Proprietary Name:ASCEND Image Viewer
Common or Usual Name:Image Processing System
Classification:Classification Panel:Class II 21 CFR 892.2050 LLZRadiology
Substantial Equivalence:ASCEND Image Viewer is substantially equivalent to TomTecArena (K132544)

Indications for Use:

ASCEND Image Viewer is a software-only medical device intended for use by trained clinicians to display medical images. ASCEND Image Viewer provides viewing of and measurements on echocardiograms from compatible ultrasound systems that support transthoracic echocardiogram, transesophageal echocardiogram, stress echocardiogram, cath, and vascular study types.

ASCEND Image Viewer is for use by prescription only.

Intended Use:

{4}------------------------------------------------

The Ascend device is intended for diagnostic review, quantification and reporting of cardiovascular structures and function of patients with suspected disease to support physicians in diagnosis.

Device Description:

The ASCEND Image Viewer is a Medical Device for viewing and measuring cardiovascular echocardiograms in DICOM format.

The software has features for displaying and measuring cardiovascular images acquired from echocardiograms. Images supported include:

  • . 2D
  • Color Doppler
  • M-mode
  • . Spectral Doppler (Pulsed Wave and Continuous Wave)

The ASCEND Image Viewer provides Cine play review and control.

Standard Review supported study types include:

  • TTE
  • TEE
  • Cath
  • . Stress Echo
  • Vascular

Supported Measurements include:

  • Linear
  • . Volume (method of discs)
  • Velocity time integral
  • Pressure half-time
  • . Area
  • Time
  • Velocity
  • Angle (measurement of two lines with a common vertex drawn on the image) .

The user can make measurements which persist for later use within the ASCEND Image Viewer.

Operator profile

The device is generally used by medical professionals such as doctors and sonographers in need of displaying echocardiograms for diagnostic purposes.

Intended use environment

{5}------------------------------------------------

The software is intended to be used in Health Care facilities including Office environments within hospitals and clinics.

Operating principle

  • . On desktop PCs the interaction with the software is mainly performed with mouse and/or keyboard

Primary operating functions

Diagnostic viewing of echocardiograms in ultrasound DICOM format.

  • C. Technological Characteristics of the Device as compared to Predicate Device:
Predicate DeviceTomTec-Image ArenaTomTec ImagingSystems GmbH(K132544)Subject DeviceASCEND Image ViewerAscend HITSubstantiallyEquivalent
Software only deviceYesYesYes
Web accessibleYesYesYes
Display Imagesstill images and imagesequencesYesYes
Select Images for closerexaminationYesYesYes
Echocardiogramanalysis and reviewcapabilityManual onlyManualYes
Data SourceMultiple modalitiesUltrasoundYes
DICOM compliantYesYesYes
Ultrasound modeutilized for echoanalysisB mode2D, Color Doppler, M-mode, Spectral Dopplerimages.Yes
Viewing mode – 2DYesYesYes
Viewing mode - 3DYesNoNo

{6}------------------------------------------------

General CardiacMeasurementsDistance LV distance Angle Area Elipse Disks Time Velocity Heart rate Acceleration PHT VTI PIRI SlopeLinear (distance)AngleAreaVolume (method of disks)TimeVelocityVTIPHTYes – Different featuresfrom predicate havebeen verified andvalidation to assurethat there is no impacton safety andeffectiveness.
Display formatAVI, BMP, JPEGor DCM exportYesYes
Minimum Hardware Requirements
Operates on off theshelf hardwareYesYesYes
OSMicrosoft WindowsMicrosoft Windows 7and above.Yes

D. Brief Discussion of Test Results Submitted:

Bench Testing

The test plan covers all aspects of the functionality, exercising every user interface, all menus and submenus, as well as the measurement tools.

The software was tested according to the company's Design Control process. Risks were analyzed according to ISO 14971 to find risks in the non-acceptable area. No risks have been identified in the non-acceptable area. Hence, no measures are considered essential performance characteristics. The software is also verified for compliance with the DICOM standard.

Clinical Testing

{7}------------------------------------------------

ASCEND Image Viewer was tested in US city-hospital environments throughout the development cycle. Clinical testers consisted of resident cardiologists, cardiology fellows, and sonographers. The clinical test results showed ASCEND Image Viewer performed effectively its intended use. Where the functionality of ASCEND Image Viewer intersected with the predicate device, it was found to be substantially equivalent. No late-arising risks or other safety concerns were recorded during clinical testing.

  • E. Conclusions from Test Results:
    The Graphic User Interface conforms to the ASCEND Image Viewer functional specification.

The comparison ASCEND Image Viewer with the predicate devices shows that the ASCEND Image Viewer has similar functionality, intended use, technological characteristics, and typical users as the predicate devices. Verification and validation activities ensure that the design specifications are met, and that ASCEND Image Viewer does not introduce new issues concerning safety and effectiveness. Hence ASCEND Image Viewer is substantial equivalent to the predicate device.

N/A