(104 days)
Indications for use of TomTec-Arena software are diagnostic review, quantification and reporting of cardiovascular, fetal and abdominal structures and function of patients with suspected disease.
TomTec-Arena is a clinical software package for reviewing, quantifying and reporting digital medical data. TomTec-Arena runs on high performance PC platforms based on Microsoft Windows operating system standards. The software is compatible with different TomTec Image-Arena™ platforms, their derivatives or third party platforms. Platforms enhance the workflow by providing the database, import, export and other services. All analyzed data and images will be transferred to the platform for archiving, reporting and statistical quantification purposes. Tom Tec-Arena consists of the following optional clinical application packages: Image-Com, 4D LV-Analysis/Function, 4D RV-Function, 4D Cardio-View, 4D MV-Assessment, Echo-Com, 2D Cardiac-Performance Analysis, 2D Cardiac-Performance Analysis MR, 4D Sono-Scan.
The provided document does not contain detailed acceptance criteria or a study proving the device meets specific performance criteria. Instead, it is a 510(k) summary for a software package, TomTec-Arena 1.0, and focuses on demonstrating substantial equivalence to predicate devices.
Here's a breakdown of what is and is not available in the provided text, in response to your requested information:
1. A table of acceptance criteria and the reported device performance
- Not available. The document states that "Testing was performed according to internal company procedures. Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted." However, it does not provide the specific acceptance criteria or the quantitative reported device performance against those criteria. It only provides a high-level summary of tests passed.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Not available. The document explicitly states: "Substantial equivalence determination of this subject device was not based on clinical data or studies." Therefore, there is no test set in the sense of a clinical performance study with patient data. The "tests" mentioned are software validation and verification.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. As indicated above, no clinical test set with patient data was used for substantial equivalence determination. Ground truth establishment by experts for clinical performance is not mentioned.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. No clinical test set.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. No MRMC study was conducted or mentioned. The device is a software package for review, quantification, and reporting, and its substantial equivalence was not based on clinical performance data demonstrating impact on human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not explicitly detailed as a standalone performance study in the context of clinical accuracy. The document confirms that "measurement verification is completed without deviations" as part of non-clinical performance testing. This suggests that the algorithm's measurements were verified, but the specifics of this verification (e.g., what measurements, against what standard, sample size, etc.) are not provided. It's a software verification, not a clinical standalone performance study.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable for clinical ground truth. For the non-clinical "measurement verification," the ground truth would likely be a known or calculated value for the data being measured, but the specific type of ground truth against which software measurements were verified is not described.
8. The sample size for the training set
- Not applicable. The document describes "TomTec-Arena" as a clinical software package for reviewing, quantifying, and reporting existing digital medical data. It is not an AI/ML device that requires a training set in the typical sense for learning patterns. Its functionality is based on established algorithms for image analysis and quantification.
9. How the ground truth for the training set was established
- Not applicable. No training set for an AI/ML model is mentioned.
Summary of available information regarding performance:
The document states that:
- "Testing was performed according to internal company procedures."
- "Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted."
- "Test results were reviewed by designated technical professionals before software proceeded to release."
- "All requirements have been verified by tests or other appropriate methods."
- "The incorporated OTS Software is considered validated either by particular tests or implied by the absence of OTS SW related abnormalities during all other V&V activities."
- The summary conclusions indicate:
- "all automated tests were reviewed and passed"
- "feature complete test completed without deviations"
- "functional tests are completed"
- "measurement verification is completed without deviations"
- "multilanguage tests are completed without deviations"
- "Substantial equivalence determination of this subject device was not based on clinical data or studies."
- A "clinical evaluation following the literature route based on the assessment of benefits, associated with the use of the device, was performed." This literature review supported the conclusion that the device is "as safe as effective, and performs as well as or better than the predicate devices."
In essence, TomTec-Arena 1.0 was cleared based on non-clinical software verification and validation, comparison to predicate devices, and a literature review, rather than a prospective clinical performance study with explicit acceptance criteria for diagnostic accuracy.
{0}------------------------------------------------
NOV 2 5 2013
Image /page/0/Picture/1 description: The image shows the logo for Tomtec Imaging Systems. The logo is in black and white, with the word "Tomtec" in large, bold letters. Below the word "Tomtec" is the phrase "Imaging Systems" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
K132544
Page 1 of 4
nbH
TomTec Imaging Systems, GmbH Edisonstrasse 6 85716 Unterschleissheim
Traditional 510(k) Summary
- TomTec-Arena 1.0 .
Owner's Name and Address
TomTec Imaging Systems GmbH Edisonstrasse 6 D-85716 Unterschleissheim
Contact Person
Mrs. Christine Klein +49-89-32175-830 Phone Fax +49-89-32175-750
Common, Classification & Proprietary Names
| Common Name: | Various Image Analysis System Software |
|---|---|
| Classification Name: | Picture Archiving and communications systemProgrammable diagnostic computer |
TomTec-Arena 1.0 Proprietary Name:
Predicate Devices
| Predicate Device 1 | K071232 | Image-Arena Applications, Research-Arena Applications : 4D RV-Function 1.0,4D Cardio-View 2.0, 4D LV-Function 2.0 |
|---|---|---|
| Predicate Device 2 | K090461 | Image-Arena 4.0 and Image-ArenaApplications: 2D Cardiac PerformanceAnalysis 1.0 |
| Predicate Device 3 | K103782 | 4D MV-Assessment 2.0 |
| Predicate Device 4 | K110595 | 4D Sono-Scan 1.0 |
| Predicate Device 5 | K110667 | Image-Arena and Image-ArenaApplications Image-Arena 4.5, Echo-Com4.5, Image-Com 4.5 |
| Predicate Device 6 | K110746 | 4D LV-Analysis 3.0 |
| Predicate Device 7 | K120135 | 2D Cardiac Performance Analysis MR 1.0 |
| Predicate Device 8 | K122289 | Image-Com 5.0 |
summary
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is in black and white, with the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" are the words "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
Device Description
TomTec-Arena is a clinical software package for reviewing, quantifying and reporting digital medical data. TomTec-Arena runs on high performance PC platforms based on Microsoft Windows operating system standards. The software is compatible with different TomTec Image-Arena™ platforms, their derivatives or third party platforms.
Platforms enhance the workflow by providing the database, import, export and other services. All analyzed data and images will be transferred to the platform for archiving, reporting and statistical quantification purposes.
Tom Tec-Arena consists of the following optional clinical application packages:
- Image-Com .
- . 4D LV-Analysis/Function
- 4D RV-Function .
- 4D Cardio-View ●
- . 4D MV-Assessment
- . Echo-Com
- 2D Cardiac-Performance Analysis .
- . 2D Cardiac-Performance Analysis MR
- . 4D Sono-Scan
Indications for use and Intended use
TomTec-Arena software is a clinical software package designed for review, quantification and reporting of structures and function based on multi-dimensional digital medical data acquired with different modalities.
TomTec-Arena is not intended to be used for reading of mammography images. Indications for use of TomTec-Arena software are diagnostic review, quantification and reporting of cardiovascular, fetal and abdominal structures and function of patients with suspected disease.
Technological Characteristics Comparison
The Subject Device "TomTec-Arena 1.0" contains clinical application packages (CAPs) for analysis of medical studies containing multimodal and multidimensional medical data.
The actual submission combines the advantages of the FDA cleared TomTec Imaging Systems software products:
summarv
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for TOMTEC IMAGING SYSTEMS. The logo is in black and white, with the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" is the phrase "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern, and it is likely used to represent a company that specializes in imaging technology.
TomTec Imaging Systems, GmbH Edisonstrasse 6 85716 Unterschleissheim
| Predicate Device 1 | K071232 | Image-Arena Applications, Research-ArenaApplications: 4D RV-Function 1.0, 4DCardio-View 2.0, 4D LV-Function 2.0 |
|---|---|---|
| Predicate Device 2 | K090461 | Image-Arena 4.0 and Image-ArenaApplications: 2D Cardiac PerformanceAnalysis 1.0 |
| Predicate Device 3 | K103782 | 4D MV-Assessment 2.0 |
| Predicate Device 4 | K110595 | 4D Sono-Scan 1.0 |
| Predicate Device 5 | K110667 | Image-Arena and Image-Arena ApplicationsImage-Arena 4.5, Echo-Com 4.5, Image-Com 4.5 |
| Predicate Device 6 | K110746 | 4D LV-Analysis 3.0 |
| Predicate Device 7 | K120135 | 2D Cardiac Performance Analysis MR 1.0 |
| Predicate Device 8 | K122289 | Image-Com 5.0 |
The Subject Device incorporates all functionalities within dedicated clinical application packages (CAPs) with the goal of enhancing the workflow when analysing multimodal and multidimensional medical data. The underlying technology of the Subject Device is identical to the Predicate Devices listed above.
Discussion according non-clinical performance data testing
Testing was performed according to internal company procedures. Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted.
The test procedure was performed according to the Project Quality Plan.
Test results were reviewed by designated technical professionals before software proceeded to release.
All requirements have been verified by tests or other appropriate methods.
The incorporated OTS Software is considered validated either by particular tests or implied by the absence of OTS SW related abnormalities during all other V&V activities.
The summary conclusions state that:
- । all automated tests were reviewed and passed
- feature complete test completed without deviations -
- functional tests are completed
- measurement verification is completed without deviations -
- multilanquage tests are completed without deviations -
- all non-verified bugs have been evaluated and are rated as minor deviations. They are deferred to the next release.
summary
{3}------------------------------------------------
Discussion according clinical performance data testing
Substantial equivalence determination of this subject device was not based on clinical data or studies.
The overall product concept was clinically accepted and supports the conclusion that the device is as safe as effective, and performs as well as or better than the predicate devices.
A clinical evaluation following the literature route based on the assessment of benefits, associated with the use of the device, was performed. The clinical evaluation shows that the published data are relevant and applicable to the relevant characteristics of the device under assessment and the medical procedure for which the device is intended.
Risk analysis aspects were treated in the risk management report. Based on this document the existing applied methods in the literature and the newly described techniques of the product (which are considered in the risk analysis) were evaluated.
No further risks were identified.
Conclusion from the analysis of the literature review
- . The Risk-Benefit Assessment concludes that the benefit is superior to the risk, whereas the risk is low. The product TomTec-Arena is therefore harmless for patient and user and the advantages overbalance the probable risks of injury or illness for the patient.
- The data are sufficient to demonstrate compliance with the essential . requirements covering safety and performance of the device in question under normal conditions of use.
- The claims made in the device labeling are substantiated by the clinical data. .
Test Conclusions of non-clinical and clinical performance data
Test results support the conclusion, that the device is as safe as effective, and performs as well as or better than the predicate devices.
No reportable events or problems for the predicate devices exist.
The overall product concept was clinically accepted and the test results support the conclusion that the subject device is as safe as effective and performs as well as the predicate devices.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . د. ان الان الان القارة التأت J. Waldinger
October 31, 2013
J. Waldin-
CTO
summary
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2013
TOMTEC IMAGING SYSTEM. GmbH % CHRISTINE KLEIN QM & RA ASSISTANT EDISONSTRASSE 6 UNLERSCHLEISSHEIM 85716 GERMANY
Re: K132544 Trade/Device Name: TomTec Arena 1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 23, 2013 Received: October 28, 2013
Dear Ms. Klein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Ms. Klein
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh.7):
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K132544
Device Name TomTec-Arena 1.0
Indications for Use (Describe)
Indications for use of TomTec-Arena software are diagnostic review, quantification and reporting of cardiovascular, fetal and abdominal structures and function of patients with suspected disease.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sm.h.7).
21/1
N/A